Measumax 10-125, 10-126, 10-127, 10-128, MX 10-125 Operation Manual

...
www.measumax.com
Electronic Micrometer
Operation Manual
Models 10-125, 10-126, 10-127, 10-128
2. LCD Display
3. Operation
Two ways of pressing the keys are used in the following illustration:
1Press and release; (2Press and hold (for more than 2 seconds).
3.1 ON/OFF•••SET key: Press and releaseWill switch the power on/off. Press and hold (2 sec. or more) Will set the origin data for absolute measurementThe “Set” sign will be displayed on LCD. The origin data of metric measurement can be set to 0, 25, 50, 75 relative to the size of the micrometer The origin data of inch measurement can be set to 0, 1, 2, 3relative to the size of the micrometer The origin data will automatically reset to the default setting after the battery has been replaced.
Default origin data of the micrometer head is 0.
3.2 ABS/INC•••UNIT key: Absolute and incremental measuring conversion. Metric-inch conversion when pressed for a prolonged time.
Press and releaseAbsolute and incremental measuring mode conversion
INC sign will be displayed on the LCD in incremental measuring mode. When the “INC” is not displayed then
the unit is in absolute measuring mode.
Press and hold (2 sec. or more) For Metric/Inch conversion. Thein” sign will be displayed on LCD when the unit is in imperial
mode. If not displayed then the unit is in the metric measuring mode.
3.3 Output data key:
Press and releaseThe micrometer will output the displayed data and display “ ” once.
Press and hold (2 sec. or more): The micrometer will output the data and display “ ” continually until the button is pressed
4. Power
If not used for five minutes, the power will auto-shut off. The micrometer will power up when the
“ON/OFF •••SET” key is pressed or the spindle is turned. The micrometer can be switched off by pressing the “ON/OFF•••SET” key to save the battery when not in use.
Use an SR44 battery. Replace the battery when the display data is blurred or the “ ” sign is displayed on the LCD. Remove the battery cap by turning it counterclockwise with a coin or screwdr iver. The new battery must be placed with the positive (+) side up.
Front view
1 2 3 4 5 6 7
11
8 9 10
12
Part of back view
1Frame
2Anvil 3Spindle 4Locking device 5Data output key 6Friction drive 7Quick drive 8ON/OFF•••SET key 9ABS/INC•••UNIT key
10LCD display
11Frame cover 12RS232 data output 13Battery cap
13
in : Inch measuring mode.
INC: Incremental measuring mode.
ABS: Absolute measuring mode.
Set : Origin set.
: : Battery voltage is low.
: Data output Transferring
ABS SetINC
www.measumax.com
5. Data output
Data output interface is RS232C. The micrometer can be connected to a PC’s serial port, or USB port, serial port by using the correctly configured cables Remove the cap of the output connector and insert the cable. (Do not remove the cap of the cable plug as it is water resistant.
5.1 Serial port format
Baud rate
1200KB/S
Stop bit
2 Start bit 1 Parity none Data bit 7 Data logic reverse
5.2 Data output format
Order
1 2 3 4 5 6 7 8 9 10
Metric S N1 N1 N
N N N CR LX
Inch S N1
N N N N N CR LX
SMinus or Space N1Minus or Space or digit 0-9 NDigit 0-9
6. Precautions
Do not subject the instrument to blows or knocks. Do not drop it or apply excessive force. Do not disassemble the instrument. Do not press the keys with a pointed object. Only press the keys in the direction intended to insure correct sensitivity. Do not use or store the instrument under direct sunlight, or in ex treme temperatures. Do not let the instrument near strong magnetic fields and high voltage. Use a soft dry cloth to wipe stains from the instrument. Do not use organic solvents such as acetone and benzene. Wipe
measuring faces of the instrument before use it.
Remove the battery if the instrument is not to be used for a long period of time.
7. Specifications
Measuring force: 5~10N Power consumption: <=35μA
Operating temperature: 0 ~ 40 Storage temperature: -20 ~ 60
8. Trouble shooting
Failure
Causes Repairing
Display E 1 on LCD. Data overflow. Move spindle in reverse or
press the “ON/OFF•••SET” key.
Display E 3 on LCD.
1. Sensor overflow.
2. Faulty sensor.
1. Reset battery.
2. Return the micrometer for repair.
Measuring data is not correct.
1. Battery not correctly positioned.
2. Flat battery
1. Clean measuring surfaces.
2. Inspect preset data and reset it.
No display on LCD.
1. Battery not correctly positioned.
2. Flat battery
1. Reset battery.
2. Replace battery.
1. Display isn’t steady.
2. Display is illegible.
1. Battery voltage is under 1.45v.
2. Battery voltage is under 1.45v.
1. Replace battery.
2. Replace battery.
1. Display blurred.
2. The output data is wrong.
Battery voltage under 1.45v. Replace battery.
The output data failed
The cable not correctly connected
Insert the cable again and check for correct connection.
www.measumax.com
SPC DATA DOWNLOAD SYSTEM 40-400
The system is designed to connect the computer and micrometer, to make full use of the electronic micrometers precision and the computer's data processing functions. This system expands the electronic micrometers functions and is ideal for use in manufacture, to manage quality control and measurement recording. For example: In quality inspection: Pick out the qualified products, reprocess products, and the inferior product; In processing: can record the working situation of machine tools and operators.
Hardware
Connection to the computers from the electronic micrometers through the 9 pin RS232 COM port on the computer, and a small jack plug on the micrometer. The data is then transferred to the computer every time the "Data" button is presses Measumax can also supply interface units that can connect with 4 or 8 micrometers
Failure
Causes Repairing
Display E 1” on LCD. Data overflow.
Move spindle in reverse or press the “ON/OFF•••SET” key.
Display E 3” on LCD.
1. Sensor overflow.
2. Faulty sensor.
1. Reset battery.
2. Return the micrometer for repair.
Measuring data is not correct.
1. Dirty measuring surfaces.
2. Preset data isn’t correct.
1. Clean measuring surfaces.
2. Inspect preset data and reset it.
No display on LCD.
1. Battery not correctly positioned.
2. Flat battery
1. Reset battery.
2. Replace battery.
1. Display isn’t steady.
2. Display is illegible.
1. Battery voltage is under 1.45v.
2. Battery voltage is under 1.45v.
1. Replace battery.
2. Replace battery.
1. Display blurred.
2. The output data is wrong.
Battery voltage under 1.45v. Replace battery.
The output data failed The cable not correctly connected
Insert the cable again and check for correct connection.