MDT Bovie Specialist Operator's Manual

OPERATORS
MANUAL
Specialist
~lectrosur~ical
MDT Diagnostic Company
-
Unit
-
Technology for Life
Rev.
C
OPERATORS MANUAL Rev. 2 (12/15/81) Rev.
C
(
9/14/88) Retyped and MDT
500807
Logo
and Address Added
CAUTION
DESCRIPTION OF SYMBOLS
The following symbols with related notes appear
"Warning" notes alert the user to the possibility of personal injury.
"Caution" notes alert the user to the possibility of damage to the equipment.
"Notes" alert the user to pertinent facts and conditions.
Users must read and understand the contents of this manual before using the equipment.
&
NOTES IN MANUAL
in
this manual.
The
BOVlE
only for use by qualified physicians and surgeons. Misuse or mis­application of this device may result in serious physical harm to the patient andlor user. It is imperative that this manual be read and understoodprior to use of the equipment.
Specialist is a sophisticated medical device intended
RETAIN THIS MANUAL FOR FUTURE REFERENCE!
-MDT,
i
i
BOVlE
and CHUCK-IT are registered trademarks.
TABLE
OF
CONTENTS
...
PAGE
....
4
1
4
3
4
4
5
6 6 9
9 9
10
19 11
11
12
13
SECTION
"
.
1 . INTRODUCTION
Preface to Electrosurgery Introduction to the Specialist 1 Unpacking Power Requirements Location in Office 1
2 . OPERATION
Operating Controls Description Circuitsentry Initial Check-Out Procedure Features Accessories Checking Cords. Connections Electrosurgical Currents 9 Electrosection (Cutting Current) Electrocoagulation Desiccation and Fulguration
3
.
SAFETY AND PRECAUTIONS
General Precautions Anesthetic Prevention of Burns Patient with Cardiac Pacemaker 12 Grounding the Unit 12 Preventing High-Frequency Skin Burns Review of Action Steps for Safety
............................................................
...........................................................
.......................................................
.......................................................
..........................................................
........................................................
..................................................................
....................................................
.................................................
........................................................
................................
...................
...................
..
........................
...
..............
....
....
..
....
.....
..................
.....................................................
........................
..
...
...‘..........
..
..........................................................
...........................
....................
..
.........................................................
.........................................................
..............................................
..........................................................
................................
.............................................
4
.
SUGGESTED POWER SETTINGS
Power Required Suggested Power Settings Output Power vs . Dial Settings Electrical Specifications 19 Mechanical Specifications 19 Limitations of Equipment 19
5 . EXPERIMENTS IN ELECTROSURGERY
Preliminary Experimental Work 21 Experiment with Electrosection (Cutting) 22 Experiments with Coagulation 23 Experiments with Fulguration and Desiccation Summary
6
.
ACCESSORIES & OPTIONS
........................................................
......................
.........
.. ..
................
...............................................
.....................................................
...................................................
....................................................
...............................................
.....................................
................................................
................................
...................................................................
..............................................
15 16
18
26 26
27
iii
BOVlE SPECIALIST ELECTROSURGICAL UNIT
AND Sf ANDARD ACCESSORlES
MONO-POLAR 81-POLAR HAND/ FOOT DISPERSIVE
OUTPUT
P'
f/
CHUCK HANULE
1
r
\
FORCEPS
MANUALS-OPERATOR'S
7
SERVICE
BLADE
ELECTRODE
\
h
POINT
NEEDLE
LOOP
HAND
ADAPTER
CONTROL
~nn~
.
--.
CONTROL
DISPERSIVE
ELECTRODE
BOVIE SPECIALIST ELECTROSURGICAL UNIT
INTRODUCTION
1.
Introduction
PREFACE TO ELECTROSURGERY
Electrosurgery has become an important modal­ity in modern medicine. The BOVIE Specialist is a versatile and safe unit providing a unique coagulation /hemostasis waveshape with superior clinical per­formance. The availability of accessories and a wide selection of electrodes will provide many surgeons with an instrument that can be utilized effectively in specialized minor surgery.
The Specialist is designed with a graphic around the Coagulating Power Control Knob which depicts the power settings to be used for various types of electrodes. The graphic is based on bi-terminal techniques utilizing a dispersive electrode. It is an exact setting for any procedure. However, it can be used as an aid for safe approximate settings, thereby reducing the chance of electrode damage or excessive tissue destruction.
Electrosurgical units provide currents essential to office practice in proctology, dermatology, oral surgery, gynecology and are also valuable in major surgery. The final choice of currents is based on the surgeon's knowledge and experience in obtaining an incision current (Cut), a destructive current
(Coagulation/Fulguration)
rent (Hemostasis).
The nature of this equipment as a low power electrosurgical unit makes it unadaptable (Transurethral Resection of the Prostate) and some other major surgical procedures.
or a combination cur-
not
to TURP
INTRODUCTION TO THE SPECIALIST
The BOVIE Specialist, like all electrically oper­ated equipment, must be properly installed to avoid any chance of electrical shock. It is a source of high frequency energy, and it must be used properly to prevent excessive tissue destruction. The informa­tion contained in the following sections is intended to provide a basic understanding of the factors involved in the use and service of this equipment.
For user convenience, the manual is physically divided into an Operator's Section and a Service Section. These may be separated and placed in an operatory and a service area at the user's option.
Both sections, Operator's and Service, are in­tended to assist the user in proper BOVIE operation. A Parts List is included in the Service Section.
The Operator's Section includes a review of
"Experiments in Electrosurgery." However, the
not
review is only as a common point of reference for clinical applications.
intended as a clinical training manual,
UNPACKING
Carefully remove the Specialist generator and other partsfrom theshipping container. Ensure that no parts are discarded with the packing material. Inspect the components for shipping damage. Report any such damage immediately.
Use the following packing list to confirm receipt of all parts:
PACKING LIST
1.
Specialist Generator Assembly
2.
Active Chuck Handle
3.
Box of Five Electrodes: Ball, Point, Loop, Blade
and Needle
4.
Foot Control
5. Dispersive Cable
6.
Dispersive Electrode and Tube of Liquid Gel
7.
Hand Control Adapter and Sample Hand
Control
8.
Operator's Manual
9.
Warranty Card
POWER
50/60 Hertz power source furnishing a maximum of 130 volts. The volt-ampere requirements are below conventional wiring ratings.
REQUIREMENTS
The Specialist is designed to operate from a
LOCATION IN THE OFFICE
The specialist generator can be placed desk, wall or cart. The outlet must be of the wire grounding type.
cut off or by-passed with a non-grounded adapter ("Cheater plug") to avoid use of
If so used, the Specialist would not be properly grounded and will void the warranty. If there is any doubt about the ground connection, consult your electrical contractor and make any recommended outlet changes.
Use the handle on the back of the generator to carry the Specialist.
The grounded leg must not be
a
three-wire outlet.
on
three-
a
NOTES
BOVlE SPECIALIST ELECTROSURGICAL UNIT
OPERATION
OPERATING CONTROLS
CUTTING POWER (yellow)
To vary the amount of
CUTTING
TO
forms. Pure Cut has minimal Hemostasis. useful in desiccation. power for Coagulation. Cut with Hemostasis has a moderate amount.
TYPE
select cutting wave AMPLITUDE CONTROL (blue)
(yellow) POwerfor Cutting.
Makes cut with hemostasis
I
COAGULATING POWER To vary the amount of PULSE CONTROL
Gives excellent fulguration.
2.
Operation
ELECTRODE SYMBOLS (Coagulating) Symbols are approxi­mations for Bi-terminal Procedures.
POWER (white) 81-POLAR FORCEPS OUTPUT ON-OFF Switch tor A white and a Receptacle for main power to the unit. Indicator llght when switch ON. MONO-POLAR OUTPUT
will
is
White receptacle for Chuck Handles or Adapter. Control Accessories.
tjlack receptacle for Bi-polar Forceps plugs. Cable.
(Foot actuated only).
CONTROLS, OUTLETS
AND
\
HANDIFOOT CONTROL Receptacle tor Foot Control or Hand
MONITORS
DISPERSIVE ELECTRODE Dispersive Electrode
RF
OUTPUT (amber) Monitor Unit
is
Outout Terminals.
Monitor will light when dispersive electrode circuit is NOT complete. Output power cannot be activated when light
will light when
is
activated. Power
available at the
is
ON.
BOVlE SPECIALIST ELECTROSURGICAL UNIT
2.
Operation
DESCRIPTION
Electrosurgical units are alternating current generators of high frequency. Passage of these cur­rents through living tissue ulation. The BOVlE Specialist operates at a funda­mental frequency of that pass through tissue with no stimulating effect except for the normal heating produced by all trosurgical currents.
Controls - Provide the operator with a selection of
two cutting modes and adjustment of output levels for Cutting and Coagulating.
Receptacles - On the Front Panel supply output
power to electrodes and receives signals from Footswitch pedals or hand actuated controls.
Coagulating - Will
attempt is made to energize both simultaneously. Coagulating being the safest or least destructive mode of operation.
Monitors - Lights on the front panel indicate nor-
mal or faulty operation of the equipment.
The
amber
monitor indicates actuation of either
the Cutting or Coagulating mode.
will not cause nervestim-
1.75
megahertz, and at ranges
always
override Cutting if an
elec-
The
red
monitor indicates an incomplete disper-
sive electrode circuit and the disabling of the out-
put power. This is the normal operation of the
Circuit Sentry Monitor.
Audible Signal
-The specialist contains a two tone audible signal Po indicate when the output power is activated and in which mode the unit is operating.
A
higher pitched tone indicates the Cutting mode and a lower pitched tone indicates the Coagulating mode. Volume of the audible signal is adjusted by turning the level control, accessible through the bottom of the unit.
CIRCUIT SENTRY
A
D.C. monitor signal flows from the specialist through one wire of the dispersive electrode cable to the dispersive electrode. The signal returns through the other cable wire, and the Specialist is ready to operate.
light will be off.
If the circuit path is interrupted by more than a
200
Ohm resistance, the
on, and the unit will automatically shut off.
The red Circuit Sentry monitor
RED
monitor light will go
--__F___t-
DISPERSIVE ELECTRODE RECEPTACLE
CIRCUIT SENTRY
DISPERSIVE ELECTRODE
BOVlE SPECIALIST ELECTROSURGICAL UNIT
INITIAL CHECK-OUT PROCEDURE
Turn the three control knobs to the full counter-
1.
clockwise position.
Insert the AC Power Plug into a three-wire
2.
115
grounded type
3.
Without
white power will light will accessory connection.
Insert the Dispersive Electrode Cable into the
4.
DISPERSIVE ELECTRODE receptacle. use of the
5.
Attach the blue plastic cable connector to the
Dispersive Electrode. Use a straight pull on the
grip surface of the connector. This connection will complete the circuit and
thered CIRCUIT SENTRY monitor will goOFF. The BOVlE Specialist is now ready to actuate. Connect the footswitch adapter into the HAND/
6. FOOT CONTROL receptacle. Rotate plug until
the keyways mesh, then turn knurled portion clockwise to secure pl
attaching any accessories, depress the
ON/OFF button to ON. The button
light and thered CIRCUIT SENTRY monitor
go
volt/60
ON.
-rhe unit is
Hz.
service outlet.
now
ready
for
7.
8.
9.
procedures are attempted. Technique experiments
are
tion..
skilled surgeon to develop an operational feel for the surgical equipment.
iar with the "General Safety and tion of this manual. These brief statements will to mind many of the concerns and potential hazards inherent to electrosurgery.
2.
Operation
Actuate the footswitch in the CUT mode. The
Amber RF OUTPUT monitor will light and the
high pitch audible signal will sound. Actuate the footswitch in the COAG mode. The
amber RF OUTPUT monitor will light and the
low pitch audible signal will sound.
Selecttheapplicableactiveaccessor~andelec-
trode for the required surgical procedure. The surgeon should become familiar with the
BOVlE Specialist before any actual
reviewed
of
unique>haracteristicsof thi~'~articular electro-
In addition, the surgeon should become
this
in
the
manual
will
Experiments
prove
helpful
Precaut~ons" sec-
Set-
to
fam~l-
br~ng
ACTIVE ELECTRODE
ACTIVE CHUCK
HANDLE AND CORD
DISPERSIVE
ELECTRODE CABLE
DISPERSIVE ELECTRODE
GENERATOR AND ACCESSORIES
BOVlE SPECIALIST ELECTROSURGICAL UNIT
2.
Operation
FEATURES ACCESSORIES
The safety and versatility of the BOVlE Special-
ist include the following features:
e
Circuit Sentry monitoring (Dispersive Electrode).
e
Coagulation overrise when inadvertent actuation of Cut and Coagulate is simultaneous.
e
Coagulation Electrode Symbols.
a
Output only to the selected active electrode.
e
Distinctive separate audio tones. Bipolar outlet. following illustration.
Active Electrode Outputs
Allactive accessories leads should be foot ing efficiency.
Active electrodes can be energized by either a
foot control or a hand control. This is shown in the
in
length to obtain maximum operat-
10
TYPE
OF
ACTUATION
,P-
CONTROL MONOPOLAR FORCEPS HAND
CONTROL
MONOPOLAR
BIPOLAR BIPOLAR FORCEPS
MONOPOLAR CHUCK HANDLE WITH
ACTIVE ELECTRODES
Reusable Chuck Handle
This type of active accessory is the most versa-
tile of all shapes and configurations for surgeon
preference. The cord plugs directly into the MONO­POLAR OUTPUT receptacle.
Insert the electrode into the open end of the
chuck handle. The electrode may not seat com-
C)
INSERT AND SEAT
ACCESSORY CHUCK HANDLE WITH
/-
SINGLE ELECTRODE
SINGLE ELECTRODE
pletely until the chuck jaws inside the chuck handle are fully opened. To open, hold the barrel of the chuck handle while turning the cable end
counter­clockwise. Seat the electrode fully and close chuck jaws by turning the end of the chuck handle clock-
wise.
CLOSE
ELECTRODE
HANDLE
INSERTING ELECTRODE INTO HANDLE
Optional Disposable Chuck Handle (Chuck-it)@
A disposable chuck handle is used with the addition of a directly into the chuck handle; there is no locking
.Some chuck handles are supplied with the
twist. electrode built into the chuck handle.
BOVlE adapter. The electrodes slip
DISPOSABLE CHUCK HANDLE ELECTRODE
BOVlE SPECIALIST ELECTROSURGICAL UNIT
2.
Operation
1.
Plug the insert terminal into the adapter.. It will
be secured by a spring latch. Plug the adapter into the white MONO-POLAR
2.
OUTPUT receptacle.
I
TO WHITE MONOPOLAR
OUTPUT RECEPTACLE
___)
1
DISPOSABLE CHUCK HANDLE
Optional Hand Control
The BOVIE Specialist can accommodate a hand control with the use of a plug adapter. Hand controls
provide fingertip control of the cutting or coagulat­ing currents. They are used in place of the conven-
tional footswitch control.
HAND CONTROL
3-PIN CONNECTOR
ADAPTER
The hand adapter plugs into the same cle as the foot control. When the hand control adap­ter is connected, the foot control output circuits are
disconnected.
TO SPECIALIST HAND/FOOT CONTROL RECEPTACLE
recepta-
-
HAND CONTROL
HAND CONTROL ADAPTER
.
BOVIE SPECIALIST ELECTROSURGICAL
2.
Operation
UNIT
Footswitch Control
Power output from the white monopolar recep­tacle can be obtained only with a footswitch acces­sory. This outlet allows any monopolar accessory to produce cut or coagulation currents.
Po connect the footswitch plug, slowly turn the
1.
flexible connector portion of the plug until the
pins mesh with the receptacle holes.
Turn the knurled portion of the plug clockwise
2.
to lock the plug into the receptacle.
Forceps
Forceps are used to grip tissue between the blades or tips of the electrode. Coagulation current should be applied after the forceps grasp the tissue. Effectiveness of forceps coagulation depends on
the
size
Of
the
blades tissue gripped. Larger blades require more power, also, as the volume of the tissue being gripped increases, the power requirement increases.
There are two distinct modes of forceps opera­tion; monopolar and ical effects which should be observed by the sur­geon. Monopolar forceps coagulate at the gripped
tissue first and then deeper into the tissue as the application of current continues. Bipolar forceps coagulate across the gripped tissue.
Monopolar Forceps
niques require footswitch activation, power being available at the white monopolar receptacle.
and
the
surface
dipolar. Each have distinct clin-
-
Monopolar forceps tech-
area
Of
the
Coagu-
lating currents with this adapter require bi-terminal techniques since both bladis of thi monopolar for­ceps are active. Coagulating currents must return to
the
dispersive
sis. A coagulating power setting can be estimated by comparing electrode pictorial size on the gra-
phics
to
m~n~~~larf~r~ep~ blade size being used.
smaller area of contact would require less Power.
Bipolar Forceps
quire coagulation currents to flow from one blade of the forceps through a small area of tissue to the other forceps blade. Because of this small area of contact, power must be radically reduced. The coagulation graphics as presented on the Specialist do because technique depends on various forceps blade sizes, tissue grip and areas of contact. Lower settings than geon acquires a "feel" for the Specialist and its uni­que capabilities. Surgical procedures using bipolar coagulation can be referenced only by skill and experience similar to that obtained for cutting currents.
Depending on the accessory used, bipolar for­ceps may be actuated by either footswitch or hand control. The dispersive electrode is always con­nected and need no actuation. The black outlet for
the indifferent electrode acts as a current return path in place of the dispersive electrode.
Both cutting and coagulation currents are
available at these outlets, and determination of coagulating power settings are at the discretion of the surgeon.
electrode,
for
effective tissue
necro-
A
-
Bipolar forceps techniques re-
BOVIE
not
represent suggested settings,
normal should be used until the sur-
MONOPOLAR
ACTIVE
/cl
FORCEPS COAGULATION
ACTIVE
BIPOLAR
///
,
FLOW
BOVlE SPECIALIST ELECTROSURGICAL UNIT
2.
Operation
Dispersive Electrodes
The reusable stainless steel dispersive elec­trode must be clean to ensure a safe and good con­tact. Always clean after each use. To clean, discon-
nect the electrode cord from the electrode and wash in hot water with mild detergent, rinse and dry thor-
oughly with
a
towel.
Disposable dispersive electrode should be
highlyflexible, have medical grade adhesive, and be applied with a liberal amount of electrosurgical conductive medium (BOVIE Gel Conductor). The dispersive electrode should be large enough for the full output of the generator. The recommended min-
imum area for the Specialist is 50cm2
(7.8
in.2). Fol­low all manufacturer's instructions and recommen­dations when using disposable dispersive elec­trodes.
CHECKING CORDS, CONNECTIONS
It is important that all cords and plugs be fre­quently checked for breaks in insulation or electri­cal contact. When plugs are pulled out of recepta­cles,
grasp
cords may pull the connections loose.
the
plug,
not
the
cord.
Jerking on the
The Pure Cut current is almost devoid of any dehy­drating effect.
Its usefulness therefore, is predomi­nantly in nonvascular areas or where fast cutting is the only requirement.
It has advantages in confined areas where the utilization of the cold scalpel becomes impractical. Application of the cutting current of the BOVlE Specialist is
always
bi-terminal (dispersive elec­trode) and requires more understanding and skill than the dehydrating currents.
Cut with Hemostasis
Clinical requirements for cutting with a dehy­drating effect requires a blend of cutting and coagu­lating currents with the result being a cutting capa­ble of sealing small vessels during electrosection. The BOVlE Specialist provides this hemostasis by the surgeon's selection of cutting type. This stasis cutting current could be a more desirable method on some electrosection since a moderate amount of dehydration will occur.
This mode can also be used very effectively in desiccation coagulation where a precise fine spark is desired.
hemo-
Age and repeated sterilizations can cause
cracks in insulation which might result in burns to patient or surgeon.
Disposable Chuck-It active cables and disper­sive electrode cables are completely tested for con­tinuity at the factory. This eliminates the need for any test before use.
ELECTROSURGICAL CURRENTS1
The BOVlE Specialist, as indicated earlier, is designed to provide high-frequency energy for the purpose of cutting, coagulating, desiccation or fulguration.
ELECTROSECTION (Cutting Current) Pure Cut
The waveform of the Pure Cut current is essen­tially sinusoidal. Its effect is to heat tissue cells so quickly that they volatilize or explode in the path of the electrode. Tissue parting is similar to a sharp cutting blade. The heat generated by the electrode arcing also causes a slight zone of dehydration at the severed edges. This will seal off minute vessels.
ELECTROCOAGULATION
Coagulation is always bi-terminal (dispersive or indifferent electrode is used) and tends to produce more necrosis of tissue than desiccation. The des­truction effects are not as quickly limited by a zone of dehydration. Therefore, this current is more effective on large or deep growths. Active elec­trodes such as needles, pointed rods, or small balls are always used in contact with the tissue; either firmly pressed against the tissue to be coagulated, or inserted into it. The tissue area of contact being constant, the amount of coagulation around the electrode is dependent on the amount of current and length of time it is applied.
Continued
'Adapted from Principals of Minor Electrosurgery, John
Jr.,
M.D.,
Liebel-Flarsheim Company,
1957.
Pages
8-1
. .
1
F.
Otto,
BOVlE SPECIALIST ELECTROSURGICAL UNIT
2.
Operation
For effective coagulation on a large growth, length of time is more important than the amount of current. The Specialist graphic indicates suggested power trode indicated. Emphasis again is 'placed on the many procedures where variations in power set­tings depend largely on the experience and judg­ment of the surgeon. The process of coagulation is dependent on areas of highest current concentra-
tion where dehydration and carbonization will occur first. Skilled surgeons in their procedures have found short times with high power are not as effec­tive as low power and longer times if deeper pene­tration is desired.
trode, coagulation begins at its point, this being the area of highest current density. Dehydration occurs and loss of conductivity at this point forces current to dehydrate successive areas of less penetration. While the current is left on, the ascending area of coagulum, broadens in a cone-like fashion until it becomes evident at the surface. Ball or flat blade electrodes in contact with tissue for coagulaton purposes depend on area of contact, power setting and time to produce the clinical requirement. The effect is a dehydration of descending nature towards the dispersive electrode.
adjustinents or settings for the type of elec-
In applications of a penetrating needle elec-
DESICCATION AND FULGURATlON
Desiccation is usually a bi-terminal technique with dispersive electrode. Many surgeons find the amplitude controlled cut with hemostasis waveform ideal for this precise form-of coagulation. You will find that spark intensity diminishes as the cut power control is turned to lower settings. the short fine spark appropriate for dermatology and facial surgery.
In the process of fulguration, the coagulation current jumps from a pointed active electrodeto the large mass of the patient. Fulguration being a sur­face dehydrator, becomes self-limiting, since dehy­drated areas are non-conductive to currents. The same effect happens if the active electrode becomes coated with coagulum.
The effectiveness of desiccation and tion are superficial, and therefore are preferred when cosmetic results are important.
Phis produces
fulgura-
BOVIE SPECIALIST ELECTROSURGICAL UNIT
3.
Safety & Precautions
SAFETY
&
PRECAUTIONS
GENERAL PRECAUTIONS
Certain general precautions are neces­sary in the use of any electrosurgical unit. Read and observe them carefully.
The BOVIE Specialist Electrosurgical unit is sold only for use by qualified physicians and surgeons. The observance of safe
establishedmedicalpractices
and tial to its proper use, otherwise, there are possibilities of injury to patients or oper­ators.
Operator and patient safety are of prime impor-
1.
tance in electrosurgical applications. Due to the
necessity of high voltages in any cal instrument, added precautions have been taken in the BOVIE Specialist.
In bi-terminal techniques, proper application of
2.
the dispersive plate with conductive gel usually will require less power for good clinical results.
The active cable, regardless of accessory used,
3.
should NOT be coiled, but let droop away from the patient and operator and preferably away from large metal masses.
--
The safe and effective use of elecfro-
surgery is dependent, to a large extent,
upon factors under the control of the operator and not entirely controllable by
the design of this equipment. It is impera­tive that the instructions supplied with this equipment be read, understood and followed in order that safety and effectiveness be enhanced.
is essen-
electrosurgi-
With present-day techniques of administering spinal, intramuscular, oral and rectal anesthetics, the surgeon or anesthesiologist has a wide choice from which to select an appropriate anesthesia for any electrosurgical procedure.
Useof local anesthetic may require an increase in power if the operating field contains a heavy fluid content of injectable anesthetic. It is generally bet­ter to inject an anesthetic a little distance from the lesion itself, allowing it to serve more as a block anesthesia rather than injecting directly under the lesion.
PREVENTION OF BURNS Accidental Ignition
When a flammable fluid, such as alcohol or ether, is used to cleanse the field preparatory to surgery, it is well to remember that there is a possi­bility of igniting any gas or residual liquid by a spark from the electrode. When flammable fluids or sol­vents are used, allow sufficient time for complete evaporation and be sure that drapes, dressings, coverings, clothing, etc., surrounding the field are
not saturated with the liquid.
of
Flammable Fluids
Accidental Ignition of Flammable Gases
The risk of ignition of flammable gases or other
materials is an in surgery which cannot device design. Precautions must be taken by
user to restrict flammable gases and liquids whether from anesthetic or skin preparation agents, or present in surgical drapes.
Disposable anesthesia hoses are conductive
and will burn if touched by an active electrode.
herent hazard in the use of Electro-
be entirely eliminated by
the
-
ANESTHETIC
The use of an electrosurgical apparatus imposes some limitations on the type of anesthetics which can be safely used. Choice of anesthesia should be made with full consideration of the danger of using electrical sparks in the presence of explosive gases.
all
The best rule is to avoid and it should be'recognized that many commonly
used inhalant anesthetics
cially so when used with oxygen.
flammable anesthetics;
are
flammable - espe-
Avoidance of Internal Body Gases
Another precaution which must be observed is
the avoidance of any flammable gases within the
body. This is especially true in body cavities. It should also be kept in mind that colonic gases are flammableand should bedispersed with nitrogen or carbon dioxide before applying energy in the colon.
BOVIE SPECIALIST ELECTROSURGICAL UNIT
3.
Safety
PATIENT
a pacer since high-frequency with its action. This is particularly true of the demand type pacer which depends on theabsence of signal for its triggering. Interaction with pacers and other heart contact
by placement of the dispersive electrode so that the high frequency energy path is remote from the area surrounding the heart and any heart contacting devices.
(h
Precautions
WITH CARDIAC PACEMAKER
It isvery important to know if the patient is using
energy can interfere
devices can be minimized
GROUNDING THE UNIT
Be sure that the chassis of the unit is grounded
1.
in accordance with grounding instructions,
-
otherwise, there is possibility
-
of shock to patient or surgeon. Grounding
instructions are identified in the "Introduction"
section.
Do not increase power output before checking
2.
for obvious defects or misapplication. Apparent
low power output or failure of the cal equipment to function correctly at other­wise normal settings may indicate faulty appli­cation of the dispersive electrode, or failure of an electrical lead.
Electrosurgical leads should be positioned so
3.
they cannot contact the patient or other leads connected to the patient; nor should they run
parallel, close to other leads.
remote but real
Electrosurgi-
PREVENTING HIGH-FREQUENCY SKIN BURNS
Follow directions and recommended practices
3.
for the preparation, placement, surveillance,
removal and use of any dispersive electrodes. The use and proper placement of a dispersive electrode is a key element in the safe and effec­tive use of Electrosurgery, particularly in the
prevention of burns.
4.
Because of the risk of burns, needles should not be used as dispersive electrodes.
To minimize the possibility of localized burns
5.
resulting from stray electrosurgical currents, the patient should not be allowed to come into contact with grounded metal parts or other large conductive surfaces.
Physiological monitoring electrodes should be
6.
placed as far as possible from the cal site and the dispersive electrode. This will reduce the risk of burns at the monitoring site. Protective impedance incorporated into the monitoring leads may further reduce the risks of such burns. Needles should not be used as monitoring electrodes during such procedures.
The use of Safety Glasses or other eye
protection is recommended to protect the
surgeon from sparks is used in conjunction with Endoscopic
instruments.
ELECTROSURGERY DALITY FOR CIRCUMCISION PROCEDURES.
IS
NOT AN APPROPRIATE
electrosurgi-
when the Specialist
MO-
Accidental burns are possible from the disper-
1.
sive electrode (if improperly prepared or ap­plied) or from the active electrode if it is care­lessly handled or laid on the patient when not in use.
The dispersive electrode must be kept clean
2.
and smooth and should always be applied with a generous application of Conductor. Wet towels or soap lather should not be used, as conductive agents, as they might dry out, resulting in a skin burn to the patient. Under no circumstan­ces should the dispersive electrode be applied over matted hair or hard scar tissue.
BOVIE
Liquid or Gel
highifrequency
BOVlE SPECIRLIST ELECTROSURGICAL UNIT
3.
Safety & Precautions
REVIEW OF ACTION STEPS FOR SAFETY
These steps should be observed for the safety of the patient and operating person­nel.
1.
Explosion Hazards
a. Avoid flammable anesthetic. b. Avoid flammable prep solutions. c. Remove gas from colon prior to internal
surgery.
2.
Shock Hazard
a. Ground equipment, not the patient. b. If electrosurgery is essential on a patient
with a pacemaker, keep RF out of area of pacer. Be prepared to shunt a demand pacer with magnet and closely monitor patient activity.
3.
Patient Burn
a. Use adequate area dispersive electrode for
bi-terminal procedures.
b. Avoid patient contact with extraneous
ground points when using mono-terminal techniques.
c. Avoid hair and scar tissue under dispersive
electrode.
d. Do not place dispersive electrode at bony
protrusion.
e. Make skin contact uniform by using BOVlE
Liquid or BOVlE Gel Conductor.
f. Keep reusable plates clean. g. Keep dispersive electrode as close to sur-
gery site as practical but on opposite side of trunk or limb rather than adjacent so current flows through body, not on skin.
R.
Do not
active electrode is in contact with tissue to be cut or coagulated.
i.
Do not
patient tissue containing metal implants. Localized heating may occur.
4.
Surgeon Burn
a. Wear eye protection when using
scope type instruments. b. Avoid metal eyeglass frames. c. Avoid metal eye pieces on scopes. d. Use heavy duty rubber gloves. e. Do not activate ESU unless active electrode
is in contact with tissue to be operated on.
The BOVlE Specialist method of activating only one accessory function at a time eliminates the pos­sibility of inadvertently laying down an active elec­trode that may become energized when another accessory is used.
A hemostat, used in conjunction with coagulat­ing currents, should be insulated at the point of operator contact.
activate Electrosurgical unit unless
pass electrosurgical currents through
RF
on
NOTES
BOVlE SPECIALIST ELECTROSURGICAL
4.
Common Power Settings
UNIT
COMMON
POWER SETTINGS
POWER REQUIRED
The Cutting and Coagulating controls must be set at the required power. The amount of power required, either for cutting or coagulating, involves a number of variables. To determine the best set-
tings, experience plus familiarity with the Specialist
and its electrodes is required. For a general rule and
guide, see the Power Setting Chart in this section. With the fundamental principles understood, little
difficulty should be experienced in making the cor­rect settings for different types of work. The degree of hemostasis required with the cutting current is
selected by setting the Cutting Type selector.
Cutting Power Required Varies According to:
Nature of tissue to be sectioned. Fat or cartilage
1,
requires more power than skin or muscle. Scle-
rotic, fibrotic or cicatrical tissues will require
more power than softer tissue. Depth of incision. A deep incision (with the
2.
same electrode and a given speed of cut)
requires more power than a shallow one.
2.
Type of electrode used. A heavy electrode, ball or disc, will require a higher power setting than pointed electrodes, due to the larger area of tissue contact.
3.
Character of surrounding media. More power will be required for coagulation under water or in a bloody field than for csagulation in a rea­sonably dry field.
Coagulating Graphic
This graphic provides the skilled surgeon the
opportunity to recognize the approximate power
settings for the type of electrode used.
If the lowest possible power setting is used, which will allow free cutting at a desired depth, there will be a minimum of sparking and flashing at the electrode. This prolongs the life of the operating instruments and prevents charring and excessive coagulation of the wound edges.
COAGULATING
Rate of speed at which cutting electrodes is
3.
moved. Fast cuts require more power than slow. Slow cuts produce more hemostasis.
Type of electrode used. For a given depth of
4.
incision with equivalent speed of cutting. Power required will depend on the size of the electrode
used. The thinner the electrode, the less power
required.
Coagulation Power Required Varies According to:
Length of time current is applied. Contrary to
1.
what might be expected, a lower power for a
longer time will give greater depth of coagula­tion than will a high power setting for a shorter time.
COAGULATING POWER GRAPHIC
BOVlE SPECIALIST' ELECTROSURGICAL UNlT
4.
Common Power
Settings
With these fundamentals in mind, it can be seen that power settings are not fixed,
unalterable figures, but are subject to individual variation. Therefore, it should be recognized that best results are attained when the power settings are altered to meet
the immediate or existing conditions.
COMMON POWER
CUTTlffi COAGULATING
f
PURE
Cur
\
Power settings listed are
approximate.
Keratoses
1
Aborting Boils Sebaceous Cyst
Plantar Warts
Skin Tumors
NOSE AND THROAT SURGERY
Tonsil Tags
Incision of Abscesses
7
Epistaxis
Nasal Polyps (electrocutting)
TYPE
CUT
WlTH
l€haxTrns
I I
PURE CUT
CUT/HEMO
CUT/HEMO
1
'0
POWER
10'
\
J
4-5
3
3
SETTINGS
Remove lesions with sharp curet. Treat base with fulgurating or desic­cative current.
3
3
3
4-5
3-4
Insert coarse needle into center of
necrotic area.
Insert needle into center of cyst.
Use loop. Cut
rotate loop
ulate.
I
Use loop electrode.
Use heavy pointed rod, straight or curved. Use ball for more superficial destruction.
Use coarse needle. Slow cut produces more coagulation of wound edges and prevents premature closure.
Use ball electrode in contact with sur­face. Avoid contact with metal soeculum.
Place loop over polyp at Apply power and cut through light pressure. If necessary, lightly coagulate base.
180"
ELECTROSURGICAL
into center of wart and
Use needle to coag-
UNlT
~ts pedicle.
uslng
BQVIE SPECIALIST ELECTROSURGICAL UNIT
4.
Common Power Settings
COMMON
f
WRE
CUT
\
Power settings listed
arn
gnnrn~imnt~
U,"
L.p,p,,"
PROCEDURE
Nasal Polyps (Coagulation)
PROCTOLOGY
Hemorrhoidectomy
Anal Fissurectomy
Polypectomy
TYPE
,.,..,-.-.
CU~
NG
CUT
WITH
PSIS
CUTTING TYPE
CUT/HEMO
CUT/HEMO
CUT/HEMO
POWER
POWER
CUTIPOWER
5
5
4
SETTINGS (Cont'd.)
COAGULATlNG
ELECTROSURGICAL
COAGIPOWER
1-2
6
5
4
COMMON PROCEDURE*
Use needle electrode. May also be used with wire snare method of removal.
Use blade electrode. Cut with blade. Use clamp to seal
bleeder. Touch electrode to clamp
to coagulate.
Use blade electrode for subcuta­neous tissue.
UNIT
The blank chart below offers the surgeon space to keep notes on special
procedures not listed above. The operator is encouraged to use these notes
for future reference.
'See References, Page
500007
26.
17
BOVlE SPECIALIST ELECTROSURGlCAL UNIT
4.
Common Power Settings
OUTPUT POWER VS.
DIAL
SETTING
TYPICAL POWER (AT
4QQ
OHMS LOAD)
NOTE,
The above power chart is a typical output characteristic the BOVlE Specialist. Actual output power measurements may vary with type of test equipment used and methods of circuit connection. See Service Manual for output power calibration.
of
BOVlE SPECIALIST ELECTROSURGlCAL UNIT
4.
Common Power Settings
ELECTRICAL SPECIFICATIONS
This equipment is intended to operate at an
input voltage and frequency range as indicated on
the nameplate attached to the back of the unit.
lnput Voltage:
115
Rated Input Operational Range
Input Current 3.2A
Operating Frequency 1.75 MHz
Open Circuit Voltage
Output Power'
.................
.....
90-130 Vac, 50/60 Hz.
.....................
................
(Crystal Controlled)
.......
................
(True RMS Power) into a 400 Ohm Non-
Inductive Load (Pure Cut Model)
+
10% Watts (True RMS Power)
50
+
10% Watts (True RMS Power)
36
4000V Peak to Peak
into a 400 Ohm Load
(Cut with Hemostasis)
into a 400 Ohm Load
Vac, 60 Hz.
@
115V
+20% -10%
75
+
10% Watts
(Coagulation)
MECHANlCAL SPECIFICATIONS
Dimensions:
Width 10.00" Height 4.75" Depth 14.25"
Weight 20 lbs. net
Length of Power Cord 10' Length of Footswitch Cord 10'
................................
...............................
................................
.............................
(Unit and Accessories)
......................
.................
LIMITATIONS OF EQUIPMENT
This unit has been designed to operate indefi­nitely at 20% operational duty cycle 40
sec. OFF) at half power (5 Dial Setting) pure cut
into a 400 Ohm load.
(10 sec. ON and
*All output power ratings are for
volts/60
Hz.
input supply potential.
115
MOTES
EXPERIMENTS IN ELECTROSURGERY2
PRELIMINARY EXPERIMENTAL WORK
BOVlE SPECIALIST ELECTROSURGICAL UNIT
5.
Experiments
in
Electrosurgery
Before attempting actual electrosurgery, it is advisable to experiment with use of the cialist on a piece of meat. This is an excellent way for
the physician to acquaint himself with
the different currents, the control settings, manipu­lation of the footswitch and use of the various electrodes.
It is suggested that a moist piece of meat (pref­erably flank steak due to its longitudinal grain) be placed on the dispersive plate with the dispersive plate plug inserted. The meat should be allowed to
1-
NEEDLE ELECTRODE
BOVlE Spe-
the action of
attain room temperature and should be fresh and moist. If necessary, it may be moistened by injection of
a
moderate amount of physiologic saline solu­tion. The moisture content is important in order to simulate living tissue as closely as possible, although dead meat cannot be entirely comparable.
Electrosurgical techniques are not necessarily
simple; experience and care are essential for optimal
results. It is beyond the scope of this manual to suggest specific surgical procedures, the nature of which require skilled surgeons.
LOOP ELECTRODES
STRAIGHT BLADE ELECTRODE
-
-
BALL ELECTRODE
c
DISC ELECTRODE
STAINLESS STEEL COAGULATION ELECTRODE
ANGULATED BLADE ELECTRODE
TYPES OF ELECTRODES
?ibid.
Page
11
-15
BOViE SPECIALIST ELECTROSURGICAL
5.
Experiments in Electrosurgery
EXPERIMENTS
WITH ELECTROSECTION
UNIT
(Cutting)
Thecutting current will prove to be avery useful
modality in minor electrosurgical procedures, but
the successful use of this current requires more
understanding and skill than the purely dehydrating
currents. The physician will wisely familiarize him­self with the "feel" of the cutting current, the syn­chronization of hand and foot in the manipulation of the electrode and footswitch, the proper selection and application of electrodes, and the selection of
power settings suitable to the technique being employed.
Application of the cutting current is always terminal. In preparation for cutting experiments, therefore, the dispersive electrode is connected to the dispersive electrode receptacle of the and the meat placed upon it. Set the Selector to PURE CUT. Insert theelectrode into the chuck han­dle and connect to the white active receptacle of the Specialist.
bi-
BOVlE
Pure Cut
CUT.
Place the tip of the electrode lightly in contact with the meat and as the footswitch is activated, draw the electrode through the tissue at about the samespeed you would use for cutting with
Note that the current does the cutting, carrying the electrode through the tissue with almost complete lack of the resistant "feel" of a scalpel.
Lift your foot at the exact moment you complete the cut and lift the electrode from the tissue. This synchronization is important. If the current is turned off too soon, the cut will be incomplete; if too late, electrical arcing will follow the electrode from the tissue, causing extra dehydration at that point. Sim­ilarly, unless the electrode starts to move imme­diately after the current is turned on, there will be a
wider depth of dehydration at
incision.
At this point, several cuts made at different depths should be practiced with lower and higher settings used to determine lowest power settings
required for desired results; increase the power set-
ting for deepest cutting, use less power for shallow
incisions-and try stopping and starting to practice hand and foot synchronization.
the start of the
ascalpel.
Blade Electrode
flat blade electrode. The flat blade should always be used for a "clean" incision which will heal by prim­ary union if you do not cut too slowly. Set the CUT­TING control at
-
For the first experiment, use a
5
and the CUTTING TYPE to PURE
CUT WITH BLADE ELECTRODE
Changing speed of the electrode drawn through the meat will provide various amounts of sis. A clean incision with minimum hemostasis
requires a faster cut with area of contact and type of
tissue dictating setting of cut power.
hemosta-
BOVIE SPECIALIST ELECTROSURGICAL UNIT
5.
Experiments in Electrosurgery
Needle Electrode
ments, substituting a needle electrode for the flat
blade. The needle, presenting a thicker cutting edge, will produce greater dehydration on the edges of the incision, with more resistance and, conse­quently, slower cutting at the same power setting. A
needle electrode is preferred for incising abscesses or cysts where drainage is desirable and healing by
primary union is to be prevented by choice. Try short cross incisions at a CUTTING power of 2, a type of incision used particularly for an abscess or cvst. A areater hemostatic effect will probably be
noticed difficult to perform small incisions with great speed.
kith the needle electrode since it is most
Loop Electrode
tissue or looping off simulated neoplasms with a
loop electrode. Depending on the size of loop and clinical effects desired, CUTTING power should be set at 2-3 for scalloping, with higher settings re­quired for larger or deeper areas. Loop and needle electrodes tend to build up coagulum easily. After several trials, cleaning or replacing the loop would give more indicative performance. A small area of the loop should be utilized for initial contact since a
larger area of starting contact could cause coagula­tion to occur, crusting the loop and making inci­sions clinically unpredictable.
-
Now, repeat the above experi-
-
Next, try "scalloping" out bits of
Cut with Hemostasis
The experiments carried out in electrosection Pure Cut may be repeated with the CUTTING type switch turned to CUT WlTH HEMOSTASIS. Cut with Hemostasis is a blend of pure cut currents and coagulating currents, and may require a slightly higher power setting for each electrode used since the hemostasis current must, of necessity, be a pulsed waveform. Pulsing will reducecutting ability unless the power control is increased to a setting compatible with clinical requirements.
Comparison of parallel incisions on the meat in
PURE CUT and in the CUT WlTH position will verify the ease of cutting in PURE CUT and the tendency to drag slightly in CUT WlTH HEMOSTASIS. Use one power control setting for both modes of electrosection and compare the clean incision of Pure Cut with the dehydration occurring in Hemostasis.
HEMOSTASIS
EXPERIMENTS WlTH COAGULATION
Coagulation and fulguration are probably the most used techniques with electrosurgical instru­ments in office practice. The BOVIE Specialist is the
forerunner of a new coagulation current waveform which appears to be not only a cooler method, but also has greater desiccation capabilities with less
power requirement. In coagulation, power levels,
timing, electrode size and skill of application are
requirements for good clinical results.
CUT WlTH LOOP ELECTRODE
Contact the tissue lightly with the loop before turning on the current. Step on the footswitch and start movement of the loop imme-
diately. If too much of the loop contacts the tissue or it you wait too long with current on before moving
the loop, coagulation will retard or prevent cutting. Cut into, through, and out of the tissue. It is quite an art to bring the loop away from the
piece completely excised. The final thin tab of tissue
will resist severance unless the loop is handled just
right. Practice makes perfect.
The wire loop of the electrodes will bend to various angles or curve to required shapes other than round. There is no need to be equipped with loops of varied shapes or angulation.
asmall section of
tissue with the
Needle Electrode
The uniqueness of the Specialist coagulating current waveform can be appreciated if the first experiments are initiated at low power levels. Insert the needle electrode in the chuck handle, set COAGULATING power at electrode to a
noted that a ring of coagulated tissue will form at the
surface of the meat. If power is kept applied, spark-
ing will occur. Sparking under these conditions indicates loss of conductive tissue and the coagula-
tion voltage now is applied to a higher resistance or
impedance. Unless superficial dehydration is the
clinical requirement, timing is important so that
necrosis is accomplished without excessive arcing.
1/4
inch depth. Apply power. It will be
1.5
and insert needle
Continued
. . . .
. .
.
BOVlE SPECIALIST ELECTROSURGICAL
5.
Experiments
in
Electrosurgery
UNIT
Turn off the current by releasing the footswitch and withdraw the electrode. Cut a section through the blanched circle and observe the tissue change which has taken place. You will find that you have
produced an inverted cone of coagulum surround­ing and extending beyond the depth of electrode insertion.
Insert different sizes of needles to different depths, varying the amount of power from and noting the time required to produce a ring of
blanched tissue around the electrode. With exami­nation of the sections, you will soon gain an under-
standing of the amount of coagulation to be ex-
pected from various amounts of power with different depths of insertion of heavier or finer electrodes for varied periods of time.
Next, try applying low power for a longer time. Set the COAGULATING power at trode to the same depth and apply the current until blanching is obvious. At this low power, fifteen sec­tions or more will probably be required. Upon sec­tioning the meat this time, you will seea more exten­sive mass of coagulum extending farther away from the electrode and penetrating more deeply.
You have now demonstrated two methods of electrocoagulation; first, the application of high power for a short time to produce limited, accurately controlled tissue destruction and, second, the application of low power for a longer time to pro­duce more massive coagulation with a single inser-
3,
insert the elec-
1.5
to 4,
tion of the electrode. Obviously, the first method is generally preferable in the interest of conservatism, accuracy and time saving. For coagulation of mas­sive areas, multiple punctures at high power are considered usually more desirable than a single or fewer punctures at low power and for a longer time.
Blade Electrode
Blade electrodes commonly used for cut or hemostasis are frequently used for coagulation and fulguration. The ease of using the same electrode for cutting and coagulation becomes evident since electrodes do not have to be changed. The only requirement is to actuate in the coagulation mode. Here again, depending on the surgical procedure,
the Specialist and its coagulating graphic
the surgeon. Place a clean blade electrode in the chuck handle and set COAGULATING power at Place blade electrode on the moist surface of the meat with about
When power is applied, a period of will elapse before a coagulated area around the blade appears. Coagulation time and power should be determined by experimentation on the meat to become familiar with its depth of coagulation abil­ity. A cut through a coagulated area with a scalpel will further acquaint the surgeon depth of coagulation. Visual inspection of coagu­lated areas will help the surgeon relate the results to actual surgery.
1/8
to 1/4 inch area of contact.
wtth the Spec~al~st
will assist
1-3
seconds
4.
COAGULATE
WITH NEEDLE ELECTRODE
BOVlE SPECIALIST ELECTROSURGICAL UNIT
5.
Experiments in Electrosurgery
Ball Electrode
Placing the ball in contact with the meat and setting COAGULATING POWER at areas and depths of necrosis which are related to
P
time applied and power setting. Hold the ball trode in contact with the meat and apply power for seconds. Move to a fresh area and apply power for
10
seconds. Finally, apply power in another loca­tion, in line with the other two test spots, for second. Cut through the three coagulated areas
with a scalpel. Observe the difference in coagula-
6
will produce
elec-
tion. Longer time applications cause larger areas of tissue destruction.
To achieve controlled superficial destruction,
try "brushing" the surface of the meat with a ball
5
electrode over a predetermined area. Section and
note the depth of penetration as evidenced by the blanching. Vary the amount of power and speed of
1
moving the ball, noting on sections, how the depth of destruction can be varied from afew thousandths of an inch to any moderate depth required.
COAGULATE
Monopolar Forceps or Hemostat
In techniques involving the use of monopolar forceps or hemostat, use the coagulating graphic as a guide if the blade sizes compare in size to the pictorial. Micro forceps must have the power greatly reduced. Use the needle electrode setting as a start-
.
ing point. Monopolar forceps usually use bi-terminal technique. Experiment with tissue grip, power and time variations to become familiar with their use.
Hemostat techniques are similar to using
polar forceps. Use a hemostat, insulated to the operator, with bi-terminal operation. Again, exper­iment until the required power setting is found for the blade size used.
mono-
WITH BALL ELECTRODE
Bipolar Forceps
Footswitch controlled bipolar forceps can be
used with the Specialist. Bipolar forceps construc­tion is such that one blade is the active electrode, the other blade the indifferent electrode. This arrangement has a small coagulation current path
and therefore, less resistance to current flow, lower
power is usually required. The white jack is for the active electrode, the black jack is for the indifferent electrode.
BOVlE SPECIALIST ELECTROSURGICAL UNIT
5.
Experiments in Electrosurgery
EXPERIMENTS
WITH FULGURATION AND
DESICCATION
Using a pointed electrode (preferably the heavy
skin needle), allow a stream of sparks one or two
mm. long to play from the point to the meat. Try this
with and without the dispersive electrode. Hold the
meat in your left hand to add the capacitance of your
body which will simulate actual treatment condi­tions. The spray of sparks will dehydrate the super­ficial layers of tissue quite rapidly and you will find that the depth of destruction is self-limiting because complete surface dehydration will insulate the cur­rent against further penetration. Some variation in depth of penetration can be accomplished by vary­ing the power and you will note that you will use about one-half the power setting for bi-terminal guration than will be required for mono-terminal applications.
Using a needle electrode, try the CUT WITH HEMOSTASIS current for desiccation. Touch the meat with the needle and note the effect. At low power settings, you will see a more precise effect than possible with pulse controlled coagulation current.
ful-
Do not cut or coagulate your hand. Do not touch ground while unit is energized.
Your experience with these two applications of high frequency dehydration will demonstrate their particular usefulness in skin conditions.
SUMMARY
Operator skill, experience, and observations establish confidence with any electrosurgical in­strument. With continued use and expanded adapt­ability, the operator will find the superior operatory help in the practice of surgery.
BOVlE Specialist a
PART NUMBER
ACCESSORIES
BOVlE SPECIALIST ELECTROSURGICAL UNIT
6.
ACCESSORIES & OPTIONS
ACCESSORIES
Electrode, Needle Electrode, Rod Electrode, Loop Electrode, Ball Electrode, Blade Dispersive Electrode Cable Assembly, Dispersive
Footswitch Assembly, Specialist Accessory Package, Electrode Active Electrode, Control Handle Adapter Conductor, Gel
Handle, Chuck, 10 Ft. Cord
Accessories & Options
#
REQ.
OPTIONAL ACCESSORIES
4
6861
Tri-Mode (Bipolar) Adapter.
Plugs into front of Specialist unit. Provides normal and reduced-spark monopolar output plus completely isolated bipolar output.
Disposable Dispersive Electrode
Disposable Electrode Adapter
Wall Bracket
Mobile Cart,
Adapter, Hand Control
Disposable Active Electrode, CHUCK
Cord
&
Banana Plug
4
Caster
Plug for Footswitch
,IT Handle with Blade
'These parts are not field serviceable.
Loading...