T6-H1
Embracing Challenge
Electrically heated connectors
series T..-H1, V..-H1
Version T6-H1, T8-H1, tee tube union, electrically
heated to 180 °C
Version V6-H1, V8-H1, straight tube union, electrically
heated to 180 °C
Special Features
Corrosion resistant
Compact design
Gas tight
Maintenance free
Easy to operate
Easy mounting
Controlled position as option
Application
In analysis technique, often temperatures
must be kept above the sample gas dew
point. Therefore, it is absolutely necessary to
avoid cold bridges.
In order to ensure this, the temperature regulated M&C hose-/tube connectors T..-H1 and
V..-H1 are used to connect heated sample
lines type 3/4/5-N/M/H with an operating
temperature up to 180 °C.
Description
The M&C hose-/tube connectors T..-H1 and
V..-H1 are fixed on a mounting plate, decoupled from heat and covered with an insulated
enclosure.
The heater consists of a heating element
with high capacity. The temperature is adjustable on the integrated thermostat up to 180
°C with high temperature limiter and low
temperature alarm. The connection box with
integrated thermostat is installed outside the
enclosure on the mounting plate.
The bushings for the heated sample lines in
the enclosure are insulated with sockets of
heat resistant silicone. In order to avoid cold
bridges, the connector is completely heated
by means of a double-ended thermal conducting jaw. For fixing the electrically heated
sample lines type 3/4/5-N/M/H – see data
sheet 3.1 –, mounting brackets are integrated.
Technical specications and illustrations are without
obligation, subject to modications. 08.98/11.06
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
11.4

Dimensions
Electrically heated union tee tube connector version T6-H1/T8-H1
TYP.
NR.
VOLT
VA
Caution! Hot surface
To>+30°C
Reset
Tsoll
0-180°C
320 300
ø 57
ø 6,5
Tu<-30°C
1
23 4
NLL
6
5
8
7
'
PG 13,5
Alarm contact
Temperatur Min.
Power
230V, 50Hz
(115V, 60Hz)
330
350
Electrically heated straight tube connector version V6-H1/V8-H1
TYP.
NR.
VOLT
VA
Caution! Hot surface
To>+30°C
Reset
Tsoll
320 300
ø 6,5
ø 57
0-180°C
Tu<-30°C
1 2
4
3
6
5
87
'L
N L
PG 13,5
Alarm contact
Temperatur Min.
Power
230V, 50Hz
(115V, 60Hz)
60
120
Dimensions in mm
11.4
330
350
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
60
120
Technical specications and illustrations are without
obligation, subject to modications. 08.98/11.06

Technical Data
T6-H1 T8-H1 V6-H1 V8-H1
Part-No. 10B1001 (a)** 10B1101 (a)** 10B1011 (a)** 10B1111 (a)**
Connector version union tee union tee straight connector straight connector
Connector dimensions 3x ø 6 mm 3x ø 8 mm 2x ø 6 mm 2x ø 8 mm
Fittings system tube connectors of stainless steel type Swagelok
Operating pressure max. 200 bar
Operating temperature max. +180 °C
Ambient temperature -25 to +60 °C
Storage temperature -25 to +80 °C
Temperature controller capillary thermostat with high temperature limiter and low temperature alarm
integrated in electrical connection box
Operating temperature adjustable from 0 to 180 °C, set at factory at 180 °C
Low temperature alarm contact alarm point ∆T 30 °C, change over contact, voltage free; contact rating 250V, 3A~, 0,25A =
Power supply 230V50Hz, 100VA **optionally 115V 60Hz (a)
Electrical connections teminals 4 mm2, 2x cable gland PG13
Degree of protection IP 54 EN60529
Electrical standard EN 61010, EN 60519-1
Dimensions 350 x 320 x 120 mm (w x h x d)
Weight 4,5 kg
Material of sample contacting parts stainless steel 316Ti
Mounting Wall mounting
Technical specications and illustrations are without
obligation, subject to modications. 08.98/11.06
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
11.4