SP3100V/RS/HEX4-135 resp. 180
Embracing Challenge
Gas Sample Probe Series SP®
Version SP3100V and SP3100 for sampling from
zones with explosive gas
Special Features
Approval according to ATEX for sam-
pling from Ex-zone 0, 1 and 2
Approval according to ATEX for mount-
ing in Ex-zone 1, 2 or 21, 22
Optimum operational reliability
Universal applicability
Adaption to nearly all process condi-
tions due to its compact and modular
design
Easy mounting and maintanance
Small volume, fast response time
Application
The M&C sample probes version SP3100V and
SP3100 are used for continuous extraction of
explosive gases (Ex-zone 0, 1 and 2) from dust
loaden, high temperature and/or humid processes. The probes can be mounted in Ex-zone
1, 2, 21 and 22. Both versions differentiate only
in the sealing material of the filter housing.
Version SP3100V has got a sealing out of FPM
for types which are heated up to max.185°C
and version SP3100 has got a sealing out of
graphite for all versions heated up to more
than 185 °C.
Description
The sample probes are designed for easy
installation, reliable operation and troublefree maintenance. They are versatile in application and depending on the task to be performed, various pre-filters series V12 or V20
with integrated volume displacer, optional
with extension tubes (see page 4 and data
sheet 2-1.1.0.8), not included in the scale
of delivery, can be simply screwed into the
mounting flange of the basic probe. For error
control the sample gas flow rate has to be
monitored externally.
This M&C stainless steel or ceramic filter element with a large surface and high capacity
is located in the external housing. The design
offers little or no stagnant space outside
the process. The probe housing is covered
with a protection shield which is part of the
Ex-approval.
The probes are designed in such a way that
changing the filter element is possible without the use of tools. In this operation, neither
the sample probe tube nor the sample line
need to be removed, thus avoiding contamination of the clean gas path and maintaining
the integrity of the system.
The special design of the optional heating of the M&C-probes version SP3100(V)
permits controlled heating of the complete filter housing, including the mounting flange. This ensures reliable operation
external to the process preventing the
temperature falling below the dew-point.
The temperature control of the M&C-probes
version SP3100(V) can be done by three different heater types.
The first possibility is a self-regulating heater
version HEX4-135 or HEX4-180 (see data sheet
2-3.2) for Ex-zone 1 and 21, for gas sampling
from zone 0, 1 or 2. In dependence on the
ambient temperature and the heater type,
the min. temperature in the probe is 90°C
resp. 120°C, the max. temperature is 120°C
resp. 160°C.
The second possibility is a controlled heating
version HEX1-3 for Ex-zone 1, adjustable up
to 180 °C and for sampling from Ex-zone 1 or
2. The controller has to be mounted outside
the Ex-zone.
The third possibility is a controlled heating
version HEX1-1 for Ex-zone 1, adjustable up
to 180 °C and for sampling from Ex-zone 1 or
2. The controller can be mounted inside the
Ex-zone.
For back-purging the M&C pre-filter, the
option RS is available with mounted buffer
vessel triggered by an explosion-proof solenoid valve. With the mounted option for
back-purging type RS, gas can be sampled
from zone 1 and 2. The back-purge pressure
has to be monitored externally and has to be
min. 1 bar higher than the process pressure.
For the pressure control while back-purging,
a corresponding special valve is mounted
in the sample gas outlet. Thus an additional
solenoid valve to shut off the sample gas
outlet is not necessary. The back purge inlet is
shut off by a check valve.
When sampling from Ex-zones, back
purging is only allowed with a gas suitable for the sampling point.
Technical specications and illustrations are without
obligation, subject to modications. 10.04/11.12
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
2.10
Technical Data
Gas sample probe type SP3100V (up to 185 °C) SP3100 (more than 185 °C)
Part No. 20S5605 20S5600
Weather protection shield yes
Filter housing material Stainless steel 316 / 316Ti
Sealing materials Graphite, FPM Graphite
Probe flange sealing material Graphite
Pre-filter / sample tubes optional, see data sheet 2-1.1.0.6 and 2-1.1.0.8
Sample pressure max. 0,5-6 bar abs
Ambient temperature -20 °C to +60 °C
Permissible process gas temperature depending on the temperature class, however max. 200°C at the probe entry
Filter chamber volume 120 cm
Filter element, porosity F-3SS150= stainless steel*, 3 micron S-2K150= ceramic**, 2 micron
Sample gas outlet connection 1x 1/4“ NPTi for max. 8mm-tube connectors
Connection gas outlet at option RS 6 mm Swagelok connector
Mounting flange DN65 PN6, FormB, SS316Ti* >DN or ANSI possible**
Weight 7 kg
Marking
Marking with option RS
Option back purge unit type RS RS
Part No. 20S5560 (a)
Power supply 230V 50/60Hz 9W or 115V 50/60Hz 9W (a)
Electrical connection cable 3x1mm
Marking
Connection G1/2” at the buffer vessel
Max. back purge pressure 6 bar abs
Volume buffer vessel 2 liters
Ambient temperature -20°C to 60°C
Option heating type HEX4 HEX4-135 HEX4-180
Part No. 20S5510 20S5520
Mounting controller in the Ex-zone 1, 2, 21, and 22
Control self-regulating
Power supply 115V - 230V 50/60Hz
Electrical connection cable gland, terminal range 7 - 12mm, terminals max. 4mm
Marking
Case protection IP66; EN60529
Power 400W
Max. temperature 120 °C 160 °C
Min. temperature 90 °C 120 °C
Ambient temperature -20 °C to +60 °C
Low temperature alarm contact <60°C, 1 contact MC-NO, 230V 1,5A AC, 0,5ADC <100°C, 1 contact MC-NO, 230V 1,5A AC, 0,5ADC
Option heating type HEX1 HEX1-3
Part No. 20S9037 (a)
Mounting controller outside the Ex-zone
Control electronic
Power supply 230V 50/60Hz or 115V 50/60Hz (a)
Electrical connection 3 x 1,5mm
Marking
Power 400W
Case protection IP54; EN60529
Temperature 0 - 180 °C T3 or 0 - 135 °C T4
Ambient temperature -20 °C to +40 °C
Low temperature alarm contact <120°C, 1 change-over contact, 230V 1,5AAC, 0,5ADC
3
II 1G/2GD -20°C ≤ Ta ≤ +60°C EXAM BVS 04 ATEX H 045X
II 2G/2GD -20°C ≤ Ta ≤ +60°C EXAM BVS 04 ATEX H 045X
2
II 2GD EEx m II 135°C
2
II 2 G EEx em T3..T4 / II 2 D IP66 135°C...180°C EXAM BVS 04 ATEX E 253
2
II 2 G EEx d ib IIC T3*, other temperature classes on request