M&C TechGroup SP3100 Datasheet

SP3100V/RS/HEX4-135 resp. 180
Embracing Challenge
Gas Sample Probe Series SP®
Version SP3100V and SP3100 for sampling from zones with explosive gas
Special Features
Approval according to ATEX for sam-
pling from Ex-zone 0, 1 and 2
Approval according to ATEX for mount-
Optimum operational reliability
Universal applicability
Adaption to nearly all process condi-
tions due to its compact and modular design
Easy mounting and maintanance
Small volume, fast response time
Application
The M&C sample probes version SP3100V and SP3100 are used for continuous extraction of explosive gases (Ex-zone 0, 1 and 2) from dust loaden, high temperature and/or humid proc­esses. The probes can be mounted in Ex-zone 1, 2, 21 and 22. Both versions differentiate only in the sealing material of the filter housing. Version SP3100V has got a sealing out of FPM for types which are heated up to max.185°C and version SP3100 has got a sealing out of graphite for all versions heated up to more than 185 °C.
Description
The sample probes are designed for easy installation, reliable operation and trouble­free maintenance. They are versatile in appli­cation and depending on the task to be per­formed, various pre-filters series V12 or V20 with integrated volume displacer, optional with extension tubes (see page 4 and data sheet 2-1.1.0.8), not included in the scale of delivery, can be simply screwed into the mounting flange of the basic probe. For error control the sample gas flow rate has to be monitored externally.
This M&C stainless steel or ceramic filter ele­ment with a large surface and high capacity is located in the external housing. The design offers little or no stagnant space outside the process. The probe housing is covered with a protection shield which is part of the Ex-approval.
The probes are designed in such a way that changing the filter element is possible with­out the use of tools. In this operation, neither the sample probe tube nor the sample line need to be removed, thus avoiding contami­nation of the clean gas path and maintaining the integrity of the system.
The special design of the optional heat­ing of the M&C-probes version SP3100(V) permits controlled heating of the com­plete filter housing, including the mount­ing flange. This ensures reliable operation external to the process preventing the temperature falling below the dew-point.
The temperature control of the M&C-probes version SP3100(V) can be done by three dif­ferent heater types.
The first possibility is a self-regulating heater version HEX4-135 or HEX4-180 (see data sheet 2-3.2) for Ex-zone 1 and 21, for gas sampling from zone 0, 1 or 2. In dependence on the ambient temperature and the heater type, the min. temperature in the probe is 90°C resp. 120°C, the max. temperature is 120°C resp. 160°C.
The second possibility is a controlled heating version HEX1-3 for Ex-zone 1, adjustable up to 180 °C and for sampling from Ex-zone 1 or
2. The controller has to be mounted outside the Ex-zone.
The third possibility is a controlled heating version HEX1-1 for Ex-zone 1, adjustable up to 180 °C and for sampling from Ex-zone 1 or
2. The controller can be mounted inside the Ex-zone.
For back-purging the M&C pre-filter, the option RS is available with mounted buffer vessel triggered by an explosion-proof sole­noid valve. With the mounted option for back-purging type RS, gas can be sampled from zone 1 and 2. The back-purge pressure has to be monitored externally and has to be min. 1 bar higher than the process pressure. For the pressure control while back-purging, a corresponding special valve is mounted in the sample gas outlet. Thus an additional solenoid valve to shut off the sample gas outlet is not necessary. The back purge inlet is shut off by a check valve.
When sampling from Ex-zones, back purging is only allowed with a gas suit­able for the sampling point.
Technical specications and illustrations are without obligation, subject to modications. 10.04/11.12
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
2.10
Technical Data
Gas sample probe type SP3100V (up to 185 °C) SP3100 (more than 185 °C)
Part No. 20S5605 20S5600 Weather protection shield yes Filter housing material Stainless steel 316 / 316Ti Sealing materials Graphite, FPM Graphite Probe flange sealing material Graphite Pre-filter / sample tubes optional, see data sheet 2-1.1.0.6 and 2-1.1.0.8 Sample pressure max. 0,5-6 bar abs Ambient temperature -20 °C to +60 °C Permissible process gas temperature depending on the temperature class, however max. 200°C at the probe entry Filter chamber volume 120 cm Filter element, porosity F-3SS150= stainless steel*, 3 micron S-2K150= ceramic**, 2 micron Sample gas outlet connection 1x 1/4“ NPTi for max. 8mm-tube connectors Connection gas outlet at option RS 6 mm Swagelok connector Mounting flange DN65 PN6, FormB, SS316Ti* >DN or ANSI possible** Weight 7 kg Marking
Marking with option RS
Option back purge unit type RS RS
Part No. 20S5560 (a) Power supply 230V 50/60Hz 9W or 115V 50/60Hz 9W (a) Electrical connection cable 3x1mm Marking
Connection G1/2” at the buffer vessel Max. back purge pressure 6 bar abs Volume buffer vessel 2 liters Ambient temperature -20°C to 60°C
Option heating type HEX4 HEX4-135 HEX4-180
Part No. 20S5510 20S5520 Mounting controller in the Ex-zone 1, 2, 21, and 22 Control self-regulating Power supply 115V - 230V 50/60Hz Electrical connection cable gland, terminal range 7 - 12mm, terminals max. 4mm Marking
Case protection IP66; EN60529 Power 400W Max. temperature 120 °C 160 °C Min. temperature 90 °C 120 °C Ambient temperature -20 °C to +60 °C Low temperature alarm contact <60°C, 1 contact MC-NO, 230V 1,5A AC, 0,5ADC <100°C, 1 contact MC-NO, 230V 1,5A AC, 0,5ADC
Option heating type HEX1 HEX1-3
Part No. 20S9037 (a) Mounting controller outside the Ex-zone Control electronic Power supply 230V 50/60Hz or 115V 50/60Hz (a) Electrical connection 3 x 1,5mm Marking
Power 400W Case protection IP54; EN60529 Temperature 0 - 180 °C T3 or 0 - 135 °C T4 Ambient temperature -20 °C to +40 °C Low temperature alarm contact <120°C, 1 change-over contact, 230V 1,5AAC, 0,5ADC
3
II 1G/2GD -20°C ≤ Ta ≤ +60°C EXAM BVS 04 ATEX H 045X
II 2G/2GD -20°C ≤ Ta ≤ +60°C EXAM BVS 04 ATEX H 045X
2
II 2GD EEx m II 135°C
2
II 2 G EEx em T3..T4 / II 2 D IP66 135°C...180°C EXAM BVS 04 ATEX E 253
2
II 2 G EEx d ib IIC T3*, other temperature classes on request
* Standard, ** optional
2.10
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
Technical specications and illustrations are without obligation, subject to modications. 10.04/11.12