M&C TechGroup SP2000H-DIL-BR Operator's manual

Instruction Manual
Gas Sample Probes Series SP
Versions SP2000H/DIL, SP2000H/DIL/B, SP2000H/DIL/BR
®
Gas sampling and gas conditioning technology 2-1.1.7-ME
2
This Operating Manual does not claim completeness and may be sub­ject to technical modifications. © 01/2003 M&C TechGroup Germany GmbH. Reproduction of this document or its content is not allowed without permission from M&C. 4th Edition: 04/2004
Dear customer,
we have made up this operating manual in such a way that all necessary information about the prod­uct can be found and understood quickly and easily. Should you still have any question, please do not hesitate to contact M&C directly or go through your appointed dealer. Respective contact addresses are to be found in the annexe to this operating man­ual. Please also contact our homepage www.mc-techgroup.com for further information about our prod­ucts. There, you can read or download the data sheets and operating manuals of all M&C products as well as further information in German, English and French.
Gas sampling and gas conditioning technology 2-1.1.7-ME
3
Content
1 General information ...................................................................................................................... 5
2 Declaration of conformity ............................................................................................................ 5
3 Safety instructions ........................................................................................................................ 6
4 Warranty ........................................................................................................................................ 6
5 Used terms and signal indications .............................................................................................. 7
6 Introduction ................................................................................................................................... 8
7 Description .................................................................................................................................... 8
7.1 Variations ............................................................................................................................... 12
8 Technical Data ............................................................................................................................. 13
9 Dilution principle ......................................................................................................................... 14
10 Dimensions .................................................................................................................................. 15
11 Receipt and storage .................................................................................................................... 15
12 Installation information .............................................................................................................. 15
13 Installation ................................................................................................................................... 16
13.1 Installation of the sample probe ............................................................................................. 16
13.2 Connection of diluted sample outlet tubing ............................................................................ 17
13.3 Connection of supply tubes for dilution bypass gas ............................................................... 18
13.4 Connection of calibrating gas ................................................................................................ 19
13.5 Electrical connection .............................................................................................................. 19
13.5.1 Types with internal capillary tube hermostat .................................................................. 20
13.5.2 Types with external temperature controller ................................................................... 20
14 Initial use ..................................................................................................................................... 21
14.1 Calibration .............................................................................................................................. 23
15 Decommissioning ....................................................................................................................... 23
16 Maintenance ................................................................................................................................ 24
16.1 Replacement of filter element and seals ................................................................................ 25
16.2 Dismantling the dilution cross ................................................................................................ 27
16.2.1 Change and cleaning of the critical orifice (180°C version) ............................................ 28
16.2.2 Changing the packing at the critical orifice ( 320°C-Version) ......................................... 29
16.2.3 Change and cleaning of the injector ............................................................................... 30
16.2.4 Change of the o-ring at the injector (180°C-Version) ..................................................... 30
16.2.5 Change the packing at the injector (320°C-Version) ...................................................... 30
16.3 Removing the Bypass-T using option ..B/..BR (180°C version) ............................................. 31
16.3.1 Change and cleaning of the bypass injector (180°C version) ......................................... 32
16.3.2 Change of the O-ring at the bypass injector (180°C-Version) ........................................ 32
16.3.3 Change of the graphite packing at the bypass injector (320°C version) ......................... 33
17 Spare parts .................................................................................................................................. 34
18 Appendix ...................................................................................................................................... 35
Gas sampling and gas conditioning technology 2-1.1.7-ME
4
List of illustrations
Figure 1 Probe type SP2000H/DIL/2x wit second sample outlet (undiluted) ................................... 8
Figure 2 Probe type SP2000H/DIL/VA/B ......................................................................................... 9
Figure 3 Probe type SP2000H/DIL/BR .......................................................................................... 10
Figure 4 Installation set and control panel ..................................................................................... 11
Figure 5 Gas flow pattern of available probe variants ................................................................... 12
Figure 6 Dilution principle .............................................................................................................. 14
Figure 7 Dimensions (mm) of SP2000H/DIL probe ....................................................................... 15
Figure 8 SP2000H/DIL/B with control panel DIL/S1 ...................................................................... 18
Figure 9 Extract from an injector data sheet .................................................................................. 22
Figure 10 Bypass injector data sheet ........................................................................................... 22
Figure 11 Cross-sectional drawing SP2000-H ............................................................................. 25
Figure 12 Removing the new filter housing lid ............................................................................. 26
Figure 13 Exploded drawing of dilution unit ................................................................................. 27
Figure 14 Crosspiece with critical orifice and o-ring seals (180°C version) ................................. 28
Figure 15 Dilution cross with graphite packing (320°C-Version) .................................................. 29
Figure 16 Injector 320°C version with graphite packing ............................................................... 30
Figure 17 Bypass-T with Injector and o-ring (180°C version) ...................................................... 31
Figure 18 Bypass Injector unit with o-ring ( 180°C-Version ) ....................................................... 32
Figure 19 Bypass Injector unit with graphite packing (320°C version) ......................................... 33
Gas sampling and gas conditioning technology 2-1.1.7-ME
5
Head Office
M&C TechGroup Germany GmbH Rehhecke 79 40885 Ratingen Germany
Telephone: 02102 / 935 - 0 Fax: 02102 / 935 - 111 E - mail: info@mc-techgroup.com
www.mc-techgroup.com
1 GENERAL INFORMATION
The product described in this operating manual has been examined before delivery and left our works in perfect condition related to safety regulations. In order to keep this condition and to guarantee a safe operation, it is important to heed the notes and prescriptions made in this operating manual. Fur­thermore, attention must be paid to appropriate transportation, correct storage, as well as professional installation and maintenance work. All necessary information a skilled staff will need for appropriate use of this product are given in this operating manual.
2 DECLARATION OF CONFORMITY
CE - Certification
The product described in this operating manual complies with the following EC directives:
EMV-Instruction
The requirements of the EC directive 2004/108/EC “Electromagnetic compatibility“ are met.
Low Voltage Directive
The requirement of the EC directive 2006/95/EC “Low Voltage Directive“ are met. The compliance with this EC directive has been examined according to DIN EN 61010.
Declaration of conformity
The EU Declaration of conformity can be downloaded from the M&C homepage or directly requested from M&C.
Gas sampling and gas conditioning technology 2-1.1.7-ME
6
3 SAFETY INSTRUCTIONS
Please take care of the following basic safety procedures when mounting, starting up or oper­ating this equipment:
Read this operating manual before starting up and use of the equipment. The information and warn­ings given in this operating manual must be heeded.
Any work on electrical equipment is only to be carried out by trained specialists as per the regulations currently in force.
Attention must be paid to the requirements of VDE 0100 (IEC 364) when setting high-power electrical units with nominal voltages of up to 1000 V, together with the associated standards and stipulations.
Check the details on the type plate to ensure that the equipment is connected to the correct mains voltage.
Protection against touching dangerously high electrical voltages: Before opening the equipment, it must be switched off and hold no voltages. This also applies to any external control circuits that are connected.
The device is only to be used within the permitted range of temperatures and pressures. Check that the location is weather-protected. It should not be subject to either direct rain or moisture. The device must not be used in hazardous areas. Installation, maintenance, monitoring and any repairs may only be done by authorized personnel with
respect to the relevant stipulations.
4 WARRANTY
If the equipment fails, please contact M&C directly or else go through your M&C authorised dealer. We offer a one year warranty as of the day of delivery as per our normal terms and conditions of sale, and assuming technically correct operation of the unit. Consumables are hereby excluded. The terms of the warranty cover repair at the factory at no cost or the replacement at no cost of the equipment free ex user location. Reshipments must be send in a sufficient and proper protective packaging.
Gas sampling and gas conditioning technology 2-1.1.7-ME
7
DANGER!
This means that death, severe physical injuries and/or important ma­terial damages will occur in case the respective safety measures are not fulfilled.
WAR NI NG !
This means that death, severe physical injuries and/or important ma­terial damages may occur in case the respective safety measures are not fulfilled.
CARE!
This means that minor physical injuries may occur in case the re­spective safety measures are not fulfilled.
CARE!
Without the warning triangle means that a material damage may oc­cur in case the respective safety measures are not met.
AT TEN TI ON!
This means that an unintentional situation or an unintentional status may occur in case the respective note is not respected.
NOTE!
These are important information about the product or parts of the operating manual which require user’s attention.
SKILLED STAFF
These are persons with necessary qualification who are familiar with installation, use and maintenance of the product.
5 USED TERMS AND SIGNAL INDICATIONS
Gas sampling and gas conditioning technology 2-1.1.7-ME
8
Sample out
diluted
Pressure gauge
-1…0bar
Cal gas
Dil. gas
6x1mm o.
Tube 8x1mm * (Inj. I)
Sample out not diluted
option 2x
1/4"NPT
Dilution gas in
Tube
6x1mm o.
Tube
Dilution cross
6 INTRODUCTION
Dilution probes are used wherever dilution of the sample gas is necessary for the measurement of one or several sample gas components. Examples are the measurement of toxic gas components, mois­ture measurements or adjustment of the sample gas concentration to the analyser measuring range.
M&C dilution probes type SP2000H/DIL... are based on the modular probe type SP2000-H. The di­verse variants, filter techniques and materials available ensure optimal adjustment to the particular process conditions.
7 DESCRIPTION
The probe types SP2000H/DIL/2x, .../DIL/VA/B, .../DIL/BR are shown below.
Figure 1 Probe type SP2000H/DIL/2x wit second sample outlet (undiluted)
Gas sampling and gas conditioning technology 2-1.1.7-ME
9
Sample out diluted
Pressure gauge
-1...0bar
Cal gas
Dil gas
Check valve
Pre-heater
Filter
Crit. orifice
Bypass gas
View "A"
Bypass out
Bypass gas
Ball valve
Option VA
Filter
Bypass-T
A
A
Figure 2 Probe type SP2000H/DIL/VA/B
Gas sampling and gas conditioning technology 2-1.1.7-ME
10
Sample out
diluted
Pressure gauge
-1…0bar
Cal gas
Dil gas
Check valve
Pre-heater
Filter
Filter
Crit. orifice
Bypass back to
The process
Bypass gas
View "A"
A
A
Figure 3 Probe type SP2000H/DIL/BR
To prevent cooling below dew point at the dilution point, the dilution unit with critical orifice is in­stalled temperature-stable in the heated part of the gas sample probe directly in the clean gas outlet. A gas pre-heater heats the dilution gas to probe temperature. With the option ..B/..BR wit bypass in­jector the bypass gas is pre-heated as well (double pre-heater). The probe is available with heating to maximum 180°C or 320°C. Test gas can be supplied to the probe for analyser calibration via the integrated test gas feed connec­tion .
Gas sampling and gas conditioning technology 2-1.1.7-ME
11
Pressure regulator
-A
-A1-A1
-A
-1...0 bar
measuring point
SP2000H/DIL...
0-6bar PI
Dilution gas
0-6barPI
PI
Underpressure
Bypass Dilution gas
Vacuum pressure
-1...0 bar
132 (3HE)
19" (84TE)
SP2000H/DIL...
-S1
-S
-S
Open
Test gas Inj.-low pressure
Open
Dilution gasBypass-Injector
-S
-S1
Bypass gas
measurement
PI
0-6 bar
15-150Nl/h
-1..0 bar
0-6 bar
FI
PI PI
Low pressure
Test gas Dilution gas/Bypass
Dilution gas
The probe can also be provided with an optional hand-operated, heated shut-off ball valve in the inlet ­VA (see Fig. 2) for the purpose of sealing off the filter space from the sampling process during filter replacement. Another option is the second sample outlet for undiluted gas (..2x, see Fig.1)
A precision pressure regulator with pressure gauge serves for adjustment of the necessary dilution gas supply pressure. The function of the dilution injector is monitored via a vacuum pressure gauge. The pressure regulator and pressure gauge necessary for the dilution function must be ordered sepa­rately. Set -A is mounted directly on the probe. Control panel -S is designed for external installation and is additionally provided with a shut-off valve and flow meter for test gas volume adjustment. The options -A1 and -S1 include an additional pressure regulator for the bypass injector –B or bypass injector with process return –BR (see Fig. 1).
Figure 4 Installation set and control panel
The dilution probe can be used for standard dilution factors from 10:1 to 500:1 (Injector I). Higher dilu­tion rates in a ratio of 50:1 to 2000:1 are possible (Injector II).
Where a correspondingly small sample gas volume is taken from the process with large dilution fac­tors, an optional heated bypass injector that is integrated directly upstream of the dilution part is avail­able for reducing the response time in atmospheric respectively small vacuum operation. The bypass option is available without gas return –B or with gas return to the process –BR . In vacuum op­eration, a bypass needle valve –BV serves for bypass volume adjustment (see Fig. 3). The dilution unit is designed for problem-free operation independent of the process temperature and process pressure (see technical data) as well as for easy maintenance.
Gas sampling and gas conditioning technology 2-1.1.7-ME
12
Dilution gas
measuring point
sample gas out
Underpressure
180° C
Test gas
Underpressure measuring point
Test gas
320°C
Dilution gas
sample gas out
180° C
Test gas Dilution gas
measuring point
Sample gas out
Underpressure
not present with type *
-B
-BR
Bypass gas in
*
Sample gas Dilution gas
Underpressure measuring point
180°C
Sample gas out
Underpressure measuring point
Sample gas
180°C
Dilution gas
Sample gas out
Bypassgas out
7.1 VARIATIONS
The gas flow patterns of the available probe variants are shown in the following figure.
SP2000-H/DIL SP2000-H320/DIL
SP2000-H/DIL-VA SP2000-H/DIL-VA-B SP2000-H/DIL-B* (-BR*)
SP2000-H/DIL-VA-BV
Figure 5 Gas flow pattern of available probe variants
Gas sampling and gas conditioning technology 2-1.1.7-ME
8 TECHNICAL DATA
Technical data SP series
®
Dilution probe version SP2000H/DIL ..
Weather protective cover
Yes
Electrical connection
Terminals; max. 4 mm²
Degree of protection of terminal box
IP54 EN60529
Mains supply
230V 50/60Hz, 800W or 115V 60Hz, 800W (fuse 10A)
Material of medium contacted parts
Stainless steel 1.4571, 1.4404, quartz glass, FPM, graphite
Ambient temperature
Thermostat: +5°C to 60°C PT100, thermocouple: +5°C to 80°C
Operating temperature
0-180°C* alternatively 0-320°C
Ready
after 2 h
Temperature status alarm
/30°C to T
NOMINAL
*
Alarm contact rating
250V 3A AC, 0:25A DC
Sample gas outlet connection
1/4"-NPT internal, for max. 10mm pipe union for version /H320 = pipe union øi 6mm*
Test gas back-flushing connection
for version /R pipe 6mm o.d.*
Filter space volume
120 ml
Weight
about 20 kg
Orifice type
a b c d e f g
Dilution factors1)
500:1
200:1
100:1
50:1
30*:1
20:1
10:1
Sample gas volume [Nl/hr]2)
1.4
2.7
5.5
11
19*
28
55
Dilution factor adjustment range
with dilution gas supply pressure setting -5% - +30%3)
Dilution gas volume injector version I or II
I: 480-600 Nl/hr*, optional II: 1800–3000 Nl/hr
Dilution gas supply pressure upstream of pres­sure regulator
Min. 4.5 bar, max. 16 bar
Bypass injector/B/BR: Supply pressure – gas consumption – sample gas volume
about 2 bar : Propellant ca. 300 l/hr : Sample gas about 150 l/hr
Process pressure
0.9 to 2 bar abs. at constant pressure
Influence of process temperature variations
No influence due to operation independent of process tem­perature
Influence of pressure change
Influence negligible at pressure changes 200mbar other­wise proportional to pressure changes. Calibration of probe must be performed under process con­ditions or pressure correction must be implemented. Op­tional connection for pressure measurement / compensation 1/4“ NPT i.
13
* Standard, specify other values when ordering, intermediate values can also be specified.
1)
Values with injector I (higher dilution rates with injector II possible),
2)
at about 3 bar (downstream of supply pressure regulator) dilution gas,
3)
-5% however, not for orifice “g”
Gas sampling and gas conditioning technology 2-1.1.7-ME
14
Orifice type
a b c d e f g
Dilution ratio
500:1
200:1
100:1
50:1
30:1
20:1
10:1
Volume flow through the critical orifice [Nl/h]
1.4
2.7
5.5
11
19
28
55
9 DILUTION PRINCIPLE
The functional principle of the dilution probe is based on ultrasonic flow through a critical orifice (see Fig. 4). The flow through the orifice is constant when the differential pressure via the orifice is higher than 500mbar. For the atmospheric inlet pressure (Poin = 1020mbar), this means a pressure at the orifice outlet (P The necessary vacuum at the orifice outlet is produced by an injector operated with dilution gas.
Depending on the critical orifice selected, dilution rates can be between 10:1 and 500:1. The table be­low gives an overview of the dilution factor and sample gas volume using the injector I (480 – 600Nl/h):
* With injector II 50:1 up to 2000:1
) of less than 520mbar absolute.
out
Figure 6 Dilution principle
Checking the dilution rates and exact adjustment of the pressure conditions are described in Chapter
14.1.
Gas sampling and gas conditioning technology 2-1.1.7-ME
15
Power
230V,50Hz
(115V,60Hz)(a)
Low temperature
alarm contact
Sample out
Dil gas
8x1mm o
Tube
Tube
6x1mm o
Tube
Sample diluted
Pressure gauge
-1…0bar
Cal gas
Dil gas
Tube 8x1mm * (Inj. I)
Sample out
undiluted
1/4"NPT
Dil gas in
6x1mm o
Tube
6x1mm o
260
345
340
1) Implementation of the cal gas valve only with 180°C version; With 320°C version assembling in the angular sheet.
2) Tube connection 8mm (3/8”) o; With injector I connection DN6/8mm or DN8/10mm and with injector II DN8/10mm-DN12/14mm.
1)
2)
NOTE!
The probe should be stored in a protected, frost-free room.
10 DIMENSIONS
The dimensions of the SP2000H/DIL/2x probes are shown below.
Figure 7 Dimensions (mm) of SP2000H/DIL probe
11 RECEIPT AND STORAGE
Carefully remove the probe and any accessories from the transport packaging immediately upon
receipt and check the delivery against the delivery note.
Check the contents for possible transport damage and immediately notify the transport insurer of
any damage.
12 INSTALLATION INFORMATION
The safety rules and regulations for the prevention of accidents must be observed during installation and also subsequent operation. The information in Chapter 3‚ “Important safety information” must be observed.
The following also applies: Select the optimal sampling point according to the generally applicable directives or co-ordinate
with the responsible departments.
Gas sampling and gas conditioning technology 2-1.1.7-ME
16
NOTE!
The probe must be checked for its suitability for use with the available oper­ating parameters prior to installation (see type plate).
NOTE!
Check pipe unions for tightness.
Place the sampling point so that sufficient space is available for installation and removal of the
probe, taking into account the insertion length of the sample pipe.
Ensure good access to the probe for ease of maintenance.  Dimension the local sample connection so that the temperature of the connection is always above
the acid dew point to prevent corrosion and blockage problems. If this is not possible, a heated sample pipe type SP35/SP30 is recommended for a cold orifice.
If the ambient temperature in the connection area is >60°C due to the radiated heat, a radiant heat
reflection plate must be provided locally for protection of the probe.
The mounting flange connection of the orifice should be DN65 PN6 or 3“ANSI (115V version). For
other required connection dimensions, an optional adapter intermediate flange is available. In­stead of fitting a flanged orifice, the probe can also be fitted with an available R2" connection adapter to an appropriate pipe coupling connection. The necessary, minimum flange size or minimum connection diameter depends on the sample pipe used or pre-filter diameter.
It is recommended to fit the probe horizontally with an angle of inclination of 10° towards the proc-
ess.
13 INSTALLATION
M&C SP2000H/DIL... probes are designed for stationary use. With correct selection of the sample
point and proper installation, they will give many years of service with minimum maintenance required.
13.1 INSTALLATION OF THE SAMPLE PROBE
Remove probe cover after opening two toggle-type fasteners.  Turn stirrup bolt at the top end of the filter holder several times anticlockwise until the retaining
bracket can be turned laterally to the left.
Remove filter holder from the probe and check that the filter element is screwed tight. Subse-
quently refit filter holder.
Bring retaining bracket into previous position and tighten stirrup bolt hand-tight.  Remove heat conducting plates at sample gas outlets after loosening knurled screw. For connec-
tion of the sample pipe, screw in an appropriately dimensioned pipe union with 1/4"-NPT threaded connection using PTFE sealing tape.
Screw supplied sample pipe or pre-filter directly or using an extension pipe into the 3/4" internal
thread in the probe flange with 3/4" flat gasket and tighten. If the heated sample pipe type SP30/35 or ceramic pre-filter type V12 is used, the probe must be bolted to its flange (with welded threaded studs). The flange seal must previously be placed be­tween both flanges.
Gas sampling and gas conditioning technology 2-1.1.7-ME
17
NOTE!
When using the heated sample pipe or pre-filter V12 in conjunction with probes with the bypass return option ..BR, a short-circuit between the by­pass gas and sample inlet in the centre of the probe flange must be pre­vented or low sample gas readings will be obtained. Make sure that the gas­ket for the mounting flange does not block the bypass return connection or the probe will not function properly.
NOTE!
For the preferred probe mounting position, the sample gas outlet points downward (unnecessary for perfect functioning).
It is recommended to install the probe with a slight downward inclination towards the process. This is essential for sampling, e.g. downstream of wet scrubber (sample pipe type SP32), so that produced drops flow back into the
process.
NOTE!
Excess backpressure on the sample outlet of the probe can result in poor performance. The maximum recommended tube length for common inside tube diameters are as follows: Injector type I:
i
6mm max. 50m
i
8mm max. 150m
Injector type II:
i
8mm max. 15m
i
10mm max. 40m
i
12mm max. 80m
NOTE!
When connecting hose pipes to stainless steel fittings, a supporting sleeve must always be used.
The connection must be checked for tightness.
If the sample connection does not correspond with the flange connection DN65 PN6, the supplied
optional adapter flange must be fitted to the probe in the same way.
Position flange seal on sample connection.  Insert process-internal sample part of complete probe unit into sample connection.  Bolt mating and probe flange with supplied nuts and bolts.
13.2 CONNECTION OF DILUTED SAMPLE OUTLET TUBING
Provided on the probe side for connection of the sample outlet tubing is an elbow union with the di-
mension DN6/8mm (3/8“
for 115V version; optional also DN 8/10mm) in the space below the insu-
o
lating cover.
The sample tube is installed as follows:
Loosen toggle-type fasteners of insulating cover and remove the cover.  Insert tube through corresponding opening in base plate of the probe.  Connect tube to union.
Refit probe cover and lock with toggle-type fasteners.
Gas sampling and gas conditioning technology 2-1.1.7-ME
18
Test gas
njector-vacuum
Dilution gas
Bypass injector
Open
Open
500 l/h
Check valve
Calibration gas
Dilution gas
Vacuum pressure gauge
Bypass
Control panel DIl/S1
Dilution gas
Cal gas
Sample out
Bypass
Analyser
Back pressure max. 1,2 bar abs.
Option –A/A1
When ordering option –A or -A1, the sets, consisting of pressure regulators and pressure gauges, are mounted directly on the probe (see Fig.2). Connection of the sample pipe takes place at the elbow union of the probe (see above).
Option –S/S1 For option –S or - S1, the necessary pressure regulators, pressure gauge, shut-off valves and flow
meter are installed externally in a 19“ control panel (see Fig. 4). Connection of the sample or supply
pipes to the probe and downstream analyser system must be established by the customer. Appropri­ately marked connections are provided at the rear of the 19“ unit.
Figure 8 SP2000H/DIL/B with control panel DIL/S1
13.3 CONNECTION OF SUPPLY TUBES FOR DILUTION BYPASS GAS
Available for connection of the supply tubes are probe-sided tube unions with the dimension
DN4/6mm (1/4“
for 115V version). The tube unions are located outside the probe cover and can
o
be connected without disassembly. When using option –A (only dilution gas) or –A1 (dilution and bypass gas), the supply tubes are con-
nected to the unions of the precision pressure regulator. When using the 19“ control panels, corresponding connections for the supply gases are provided at
the rear of the unit.
Gas sampling and gas conditioning technology 2-1.1.7-ME
19
WAR NI NG !
An incorrect mains voltage can damage the unit. Check the rating plate for the correct voltage prior to connection!
The probe must be installed so that contact with live parts is ex­cluded!
In any case we recommend the use of temperature resistant cable ! The alarm contact for under-temperature must be displayed! In case of an under-temperature alarm (failure of probe heating or
sensor), the dilution gas and bypass gas supply (if equipped) must be interrupted or serious probe damage will result. We recommend that external solenoid valves to perform this function be wired into
the probe’s under-temperature alarm circuitry automatically. Damage the probe’s internal components from plugging or corro-
sion resulting from the lack of use of the probe’s under-
temperature alarm are not covered by warranty.
NOTE!
For the erection of power installations with rated voltages up to 1000V, the requirements of VDE 0100 and its associated standards and specifications must be observed. A main switch must be provided externally.
The supply circuit of the unit must be protected by a fuse with the correct rating (over-current protection); the electrical values are shown in the technical data.
When installing the capillary regulator or PID controller at the sam­ple point, the maximum permissible ambient temperature must be observed (see 8.). If this limit is exceeded, a PID controller must be installed externally and outside the temperature critical zone.
13.4 CONNECTION OF CALIBRATING GAS
The dilution heated probe is provided with a check valve as standard, opening at pressures above
0.7bar.
In both versions, a pipe union DN4/6mm (1/4“
for 115V version) is available for connection of the
o
calibrating gas.
13.5 ELECTRICAL CONNECTION
Temperature setting of the SP2000H/DIL... probes takes place with a capillary regulator as standard. The probe can also be provided with an optional PT100 or thermocouple. This requires the connection of an external temperature controller.
Gas sampling and gas conditioning technology 2-1.1.7-ME
20
WAR NI NG !
For connecting use temperature resistant cable!
WAR NI NG !
Use corresponding compensating lines with thermocouples !
Power 230V 50Hz or 115V 60Hz
Alarm output
1
2
1
2
+30°C
-30°C
Power 230V 50Hz or 115V 60Hz
PT100 or thermocouple
1
2
1
2
13.5.1 TYPES WITH INTERNAL CAPILLARY TUBE HERMOSTAT
Remove cover of connection box. The electrical wiring
diagram is located in the cover.
Insert mains cable (min. 3x1.5mm2) through cable gland
and connect to appropriate terminals.
Insert signal cable (under-temperature alarm) through ca-
ble gland and connect to appropriate terminals (the contact Tu shows the alarm event).
Screw cover back in place.
13.5.2 TYPES WITH EXTERNAL TEMPERATURE CONTROLLER
Remove cover of connection box. The electrical
wiring diagram is located in the cover.
Insert mains cable (min. 3x1.5mm2) through cable
gland and connect to appropriate terminals.
Insert temperature sensor cable through cable
gland and connect to appropriate terminals.
Screw cover back in place.
Gas sampling and gas conditioning technology 2-1.1.7-ME
21
WAR NI NG !
The air supply to the injectors should not be turned on until the probe is at operating temperature (see. technical data). Or damage to the probe or/and inconsistent initial dilution ratio can result.
Prior to initial use, it must be ensured that the mains voltage corre­sponds with the voltage on the rating plate!
Caution: Do not touch the probe surface during use as this can cause burns due to the high surface temperatures. Protective gloves must be worn and the probe protected against unauthorised access!
NOTE!
If the set temperature on the capillary controller should be reduced by more than 30°C in one step during operation, the thermostat over-temperature cut-out will operate (for re-starting press the re­set button).
NOTE!
The total heating time is about two hours. After about one hour, the lower alarm threshold value (30°C below set value) is exceeded.
NOTE!
For safety reasons, the vacuum pressure gauge must show a vac­uum reading of >-0.6 bar. With a vacuum reading of < -0.6 bar, the critical orifice will not function properly. If the necessary vacuum is not reached, the dilution gas supply pressure must be increased.
14 INITIAL USE
Prior to initial use, system and process-specific safety measures must be observed. For the media to be transported, the relevant safety requirements and measures must be taken into account.
The following step-by-step procedure is recommended: If available, close ball valve via pneumatic actuator or manually (turn twist handle fully to the right)
Check set temperature on installed thermostat or external controller.
Switch on the mains voltage.
After heating the complete probe (about two hours), open the ball valve for the dilution gas. If a ball valve is installed operate the pneumatic actuator or turn the twist handle fully to the left manually. Set the precision pressure reducer to the pressure specified on the enclosed injector data sheet (see Fig. 6).
Gas sampling and gas conditioning technology 2-1.1.7-ME
22
Injektortyp
I
Injektor-Nr.
689
Für Sonden-Nr.
8652/222835
Typ
SP2000-H/DIL/BR
Injector type
Injector No.
For probe No.
Type
Betriebsdruck
Operating pres-
sure [bar]
Durchfluss
Flow
[l/h]
Unterdruck ohne kritische Düse
Vacuum without critical orifice
[bar]
Unterdruck mit kritischer Düse Vacuum with critical orifice 5,1l/h
[bar]
2,4
415
-0,62
-0,61
2,6
435
-0,65
-0,64
2,8
460
-0,68
-0,67
3,0
490
-0,80
-0,77
3,2
510
-0,79
-0,77
3,4
535
-0,79
-0,77
3,6
560
-0,78
-0,76
3,8
585
-0,77
-0,76
4,0
605
-0,77
-0,75
Überprüfung des Verdünnungsfaktors Check of the dilution ratio
Messgasdruck atmosphärisch Sample gas pressure atmospheric
Kritische Düse Critical nozzle
Verd.gas Dilution gas
Verdünnungsgasdruck Dilution gas pressure
Messgas Sample
Verdünnung Dilution
Messwert d. verd. Gases Meas. value of the dil. gas
5,1 l/h
100% N2
3,1 bar
100% O2
100:1
1,0 % O2
NOTE!
The gas pressure must be higher than 0.7 bar, as the vacuum valve mounted on the probe has an opening pressure of 0. 7 bar.
When using the control panel, the respective ball valve for the test
gas supply must be opened and the flow rate set on the flow meter.
Bypass- Injektor
I
Ser.-Nr.
526
Für Sonden-Nr.
8652/222835
Typ
SP2000-H/DIL/BR
Bypass injector
Ser. No.
For probe No.
Type Bypassgas / Bypass gas
Prozessgas / Sample gas
Betriebsdruck
Operating pressure
[bar]
Durchfluss
Flow
[l/h]
Ansaugvolumenstrom bei 1bar
abs. Suction flow at 1bar abs.
[l/h]
Ansaugvolumenstrom bei 0,9bar abs.
Suction flow at 0,9bar abs.
[l/h]
0,5
110
45
-
1,0
155
115
-
1,5
190
200
65
2,0
235
250
135
2,5
270
300
200
3,0
310
350
250
3,5
355
370
270
4,0
395
390
305
4,5
430
425
350
An extract from an injector data sheet is shown below. Injektor-Datenblatt / Injector data sheet
Figure 9 Extract from an injector data sheet
For the operation of an installed bypass injector, the necessary pressure must be set on the pressure regulator provided (on left next to precision pressure regulator).
Figure 10 Bypass injector data sheet
Gas sampling and gas conditioning technology 2-1.1.7-ME
23
WAR NI NG !
In the event of under-temperature (failure of probe heating), the di­lution gas supply must be interrupted!
NOTE!
The gas pressure must be higher than 0.7 bar, as the vacuum valve mounted on the probe has an opening pressure of 0. 7 bar.
When using the control panel, the respective ball valve for the test
gas supply must be opened and the flow rate set on the flow meter.
NOTE!
Calibrate systems analysers directly with zero gas and low concen­tration sample gas.
14.1 CALIBRATION
Calibration of the downstream analyser system or checking the dilution factor must always take place under process conditions.
An appropriate calibration gas can be supplied via the cal gas valve.
The procedure for calibration or checking the dilution ratio is as follows:
Supply test gas of a known concentration to probe.
Check dilution ratio on downstream analyser and correct pressure on dilution gas supply pressure
regulator.
Test gas supply without bypass injector
To ensure that sufficient test gas is available in the filter space of the probe, the To test gas volume should be minimum three times the flow through the critical orifice (see injector data sheet).
Test gas supply with bypass injector
The test gas volume must be about 10% higher than the volume flow through the probe produced by the bypass and the volume flow via the critical orifice (see Fig. 7).
15 DECOMMISSIONING
Prior to decommissioning, i.e. switching off the heating, the probe should be purged with inert gas or air to prevent the condensation of aggressive components from the process gas.
Gas sampling and gas conditioning technology 2-1.1.7-ME
24
WAR NI NG !
Aggressive condensation possible. Safety goggles and appropri­ate protective clothing should be worn!
Caution: Do not touch the probe surface during use as this can cause burns due to the high surface temperatures. Protective gloves must be worn and the probe protected against unauthorised access!
Before carrying out maintenance work on electrical components, the mains voltage must be disconnected in all poles. This applies similarly to any connected alarm and control circuits!
NOTE!
The probe does not need to be removed for repairs or maintenance. The probe will need to be removed for replacement of the process-
sided sample pipe.
16 MAINTENANCE
Prior to maintenance and repairs, system and process-specific safety measures must be observed.
No recommendations for maintenance intervals can be given. Maintenance intervals must be deter­mined depending on the particular process conditions and specific application.
Probe maintenance is limited mainly to replacement of the filter elements, inspection of the seals and maintenance of the dilution systems.
Gas sampling and gas conditioning technology 2-1.1.7-ME
25
WAR NI NG !
Before carrying out any maintenance or repairs, it must be ensured that no health-endangering contaminants remain in the probe. A suitable measure is, e.g. purging the probe with inert gas.
The bypass respectively primary injector air supply should be turned off before the probe filter lid assembly is removed. Other­wise the critical orifice will be blocked by dust !
16.1 REPLACEMENT OF FILTER ELEMENT AND SEALS
For change of the filter element respectively gaskets carry out the following steps: Close ball valve (if available).
Close dilution gas supply and, if available, bypass gas supply. Remove protective cover after loosening toggle-type fasteners. Loosen stirrup bolt, swing retaining bracket to the left, grip filter holder by the ring and withdraw.
From serial no. 10283:
Figure 11 Cross-sectional drawing SP2000-H
The following steps are recommended removing the filter housing lid:
 Turn toggle screw „A“ approx. 1 turn counter-clockwise, that the filter housing lid will lift up;  Turn handle „C“ in position „E“;Turn clamp „B“ counter-clockwise in direction „G“; Take toggle screw „A“ an pull out the filter housing lid;
Gas sampling and gas conditioning technology 2-1.1.7-ME
26
NOTE!
It is now also possible to clean the sample pipe to remove depos­its.
Caution ceramic probe tubes. Increased risk of breakage!
NOTE!
Graphite gaskets in combination with the highly heated probe can only be used once. To be noted is that after inserting the filter holder, the stirrup bolt must tightened hand-tight in a cold condi­tion and retightened after reaching operating temperature.
The following pictures should explain the above mentioned steps.
Figure 12 Removing the new filter housing lid
Unscrew filter knurled screw, inspect filter element and replace if necessary. Inspect filter element seals and replace if necessary. Inspect O-rings (for /320H.. flat graphite gasket, for /7aT PTFE sealing rings) in cover and re-
place if necessary.
Clean filter space.
Turn clamp „B“ clockwise and turn with the handle „C“ the ringsscrew „D“ in position „E“, that
the clamp „B“ will latch into the ringscrew „D“ and the threaded bolt „H“. It could be neccessary to move the filter housing lid a little bit forward and backward;
Turn handle „C“ in position „F“ and fasten the filter housing lid by turning the toggle screw „A“
clockwise by hand.
Fit protective cover. Open ball valve (if available).
Gas sampling and gas conditioning technology 2-1.1.7-ME
27
Nut 1
Nut 2
Nut 3
Nut 4
Nut 5
Nut 6
Critical orifice
Mark
Injector
Mark
Mark
Tube sample gas out
Block
180°C Version
O-rings
WAR NI NG !
Do not remove nut 2 until the dilution cross is removed from the probe or the orifice will be broken !
16.2 DISMANTLING THE DILUTION CROSS
Figure 11 shows an exploded drawing of the dilution unit.
Figure 13 Exploded drawing of dilution unit
The following step-by-step procedure is recommended:
Remove probe insulating cover (see 16.1). Loosen knurled screw for fixing heat conducting plates. Remove heat conducting plates. Remove all pipe connections on dilution unit (crosspiece). These are:
- vacuum pressure gauge nut 6
- dilution gas inlet nut 4
- sample gas outlet nut 5 (remove sample gas out pipe at upper elbow union. Do not remove pipe connected to crosspiece)
- nut 1.
The dilution cross is now ready for further maintenance and repair work.
Gas sampling and gas conditioning technology 2-1.1.7-ME
28
WAR NI NG !
Do not clean the critical orifice mechanically. Cleaning should take place in an ultrasonic bath!
Jet pipe
2
1
Orifice
Sucking pipe
Pos. 2:
Sealing set for critical orifice consisting of 2 o-rings
Part No.: 93 S 4009
Orifice
Seat
O-rings
Mounting tool
Part No.: 93S4040
3
Nut 3
Injector
WAR NI NG !
The Swagelok® fittings must be carefully tightened to avoid damag­ing the internal components. Tightening too far will damage the cross.
Use a mating gauge to make sure that the fittings are not too tight. If a fitting is believed to be leaking, do not tighten the fitting to re-
move the leak. Disassemble the piece completely and reassemble it making sure that the nut is not tightened too far.
16.2.1 CHANGE AND CLEANING OF THE CRITICAL ORIFICE (180°C VERSION)
Figure 14 shows the position of the orifice in the crosspiece and the o-ring seals (180°C version).
Figure 14 Crosspiece with critical orifice and o-ring seals (180°C version)
Please dismantle as follows:
Loosen union nut 3 and remove the jet pipe; Push the injector out with the mounting tool (pos. 3, Fig. 14). Carry it out from the opposite
side;
Push the critical orifice, orifice seat and o-rings carefully out of the crosspiece (with the tool
from the opposite side) ;
Check the o-rings and change them if necessary; Re-fit the new respectively cleaned orifice into the o-rings (180°C-Version) up to block. Re-install the dilution unit in the opposite way.
Gas sampling and gas conditioning technology 2-1.1.7-ME
29
Jet pipe
2
1
Orifice
Sucking pipe
Pos. 2:
Packing for critical orifice consisting of 4 graphite seals 3,2/ 6 – 1,5mm thickness plus plain washer 3,3/ 6
Part No.: 93 S 4005
Orifice
Seat
Graphite rings
Plain washer
Mounting tool
Part No.: 93S4040
3
Nut 3
Injector
WAR NI NG !
The Swagelok® fittings must be carefully tightened to avoid damag­ing the internal components. Tightening too far will damage the cross.
Use a mating gauge to make sure that the fittings are not too tight. If a fitting is believed to be leaking, do not tighten the fitting to re-
move the leak. Disassemble the piece completely and reassemble it making sure that the nut is not tightened too far.
16.2.2 CHANGING THE PACKING AT THE CRITICAL ORIFICE ( 320°C-VERSION)
When the orifice is removed please change the packing (320°C- version) as well. Figure 15 shows the dilution cross with the orifice and the packing (320°C version).
Figure 15 Dilution cross with graphite packing (320°C-Version)
The stepwise change of the graphite packing is as follows:
Remove the dilution cross as described in 16.2; Loosen nut 3 and dismount the jet pipe; Push out the injector with the mounting tool (Pos.3, Fig. 15) from the jet pipe side Push out the orifice, the orifice seat and the graphite packing carefully from the opposite side
using the mounting tool (pos. 3, Fig. 15);
Remove the graphite sealing; Push 4 new graphite rings and the plain washer on the orifice (dimension see Fig.15); Place the orifice seat in the crosspiece; Put the orifice in the seat; Push the graphite packing with the tool into the cross up to block and press; Re-fit the jet pipe.
Gas sampling and gas conditioning technology 2-1.1.7-ME
30
NOTE!
For cleaning the injector it is not necessary to remove it out of the dilution cross. Clean the nozzle mechanically. For the 180°C­version the complete cross can be cleaned up in an ultrasonic bath.
WAR NI NG !
The complete cross of the 320°C-version can not be cleaned in an ultrasonic bath. Otherwise the graphite packing will be damaged !
Before cleaning the nozzle with compressed air remove the critical orifice first (s. 16.3) !
WAR NI NG !
The Swagelok® fittings must be carefully tightened to avoid damag­ing the internal components!
Thrust collar
Nozzle
Graphite rings
Part No.: 93S4002
Injector complete, DIL-H320
16.2.3 CHANGE AND CLEANING OF THE INJECTOR
16.2.4 CHANGE OF THE O-RING AT THE INJECTOR (180°C-VERSION)
Please change the o-ring as follows:
Loosen union nut 3 and remove the jet pipe; Loosen nut 4 and remove the dilution tube connection; Push out the injector with the tool (Pos. 3) from the opposite side (jet pipe); Remove the o-ring from the nozzle; Push on the o-ring on the nozzle; Re-install the injector with the tool. Push it up to block and press it tightly.
16.2.5 CHANGE THE PACKING AT THE INJECTOR (320°C-VERSION)
All the time when the injector is removed please change the packing (320°C- version). Figure 14 shows the injector with the graphite packing.
Figure 16 Injector 320°C version with graphite packing
Gas sampling and gas conditioning technology 2-1.1.7-ME
31
WAR NI NG !
The Swagelok® fittings must be carefully tightened to avoid damag­ing the internal components!
Injector nozzle
O-ring
Bypass-T
Jet pipe
Union nut 2
Nut 1
Nut 3
Sample gas
Dilution gas
Please change the packing as follows:
Loosen union nut 3 and remove the jet pipe; Loosen nut 4 and remove the dilution tube connection; Push out the injector with the tool (Pos. 3) from the opposite side (jet pipe); Remove the thrust collar from the nozzle; Remove the graphite rings; Push on the nozzle and the 3 new graphite rings; Replace the thrust collar; Re-install the injector with the tool. Push it up to block and press it tightly.
16.3 REMOVING THE BYPASS-T USING OPTION ..B/..BR (180°C VERSION)
For maintenance and cleaning the Bypass-T can be dismantled completely. The figure below shows the scheme of the Bypass-T.
Figure 17 Bypass-T with Injector and o-ring (180°C version)
Dismantle the Bypass-T as follows:
Remove probe insulating cover (see 16.1); Remove all pipe connections on Bypass-T. These are:
- Bypass inlet union nut 1
- union nut 2 (sample outlet side)
- union nut 3 (sample inlet).
The Bypass-T is now ready for further maintenance and repair.
Gas sampling and gas conditioning technology 2-1.1.7-ME
32
NOTE!
For cleaning the injector of the 180°C-vesion it is not necessary to remove it out of the Bypass-T. Clean the nozzle mechanically or put the complete cross into an ultrasonic bath. For the 320°C-version do not clean the com­plete cross in an ultrasonic bath because the graphite packing could be de­stroyed.
A cleaning with compressed air has to be carried out from the sample outlet side (jet pipe).
nozzle
O-Ring
Part No.: 93S4000
Injector complete, SP2000-H/DIL
Mounting tool
Part No.: 93 S 4040
Jet-pipe
Bypass-T
Pos.1: O-Ring for nozzle
16.3.1 CHANGE AND CLEANING OF THE BYPASS INJECTOR (180°C VERSION)
We recommend the following steps for the change of the bypass injector:
Unscrew the jet pipe (pos. 2) from the Bypass-T; Push out the injector nozzle from the opposite side (tool pos. 3, Fig. 18) and change it with
sealing;
Push the injector into the T up to the block and press it tightly. Use the mounting tool.
16.3.2 CHANGE OF THE O-RING AT THE BYPASS INJECTOR (180°C-VERSION)
Figure 18 shows the injector in the 180°C-version.
Figure 18 Bypass Injector unit with o-ring ( 180°C-Version )
We recommend the following steps:
Unscrew the jet pipe (pos. 2) from the Bypass-T; Push out the injector nozzle from the opposite side (tool pos. 3, Fig. 18); Remove the thrust collar and the o-ring; Push new o-ring on the nozzle; Put on the thrust collar; Push the injector into the T up to the block and press it tightly. Use the mounting tool.
Gas sampling and gas conditioning technology 2-1.1.7-ME
33
WAR NI NG !
The Swagelok® fittings must be carefully tightened to avoid damag­ing the internal components!
Thrust collar
Nozzle
Graphite rings
Part No.: 93S4003
Injector complete, DIL-H320-BR
Mounting tool
Part No.: 93 S 4040
Jet pipe
Bypass-T
Pos.1: Packing for injector nozzle consisting of 3 graphite rings 4/ 6 – 1,5mm thickness Part NO.: 93 S 4010
WAR NI NG !
The Swagelok® fittings must be carefully tightened to avoid damag­ing the internal components!
16.3.3 CHANGE OF THE GRAPHITE PACKING AT THE BYPASS INJECTOR (320°C VERSION)
When changing the injector the packing (320°C version) must be changed as well. Figure 16 shows the bypass injector unit of a 320°C version.
Figure 19 Bypass Injector unit with graphite packing (320°C version)
We recommend the following steps:
Unscrew the jet pipe (pos. 2) from the Bypass-T; Push out the injector nozzle from the opposite side (tool pos. 3); Remove the thrust collar and the graphite rings; Push 3 new rings on the nozzle; Put on the thrust collar; Push the injector into the T up to the block and press it tightly. Use the mounting tool.
Gas sampling and gas conditioning technology 2-1.1.7-ME
34
Recommended Spare Parts
Article
Description
90 S 0020
Spare filter element S-2K150, ceramic, 2µm, 150mm
93 S 0045
Spare flat packing (30), Viton , for filter element S-2K150
93 S 0020
Spare O-ring (39), Viton , for lid SP2000H
93 S 0025
Spare O-ring (55), Viton , for lid SP2000H
93 S 0055
Spare flat packing (30), graphite, for filter element S-2K150, 320°C
93 S 0030
Spare flat packing (69), graphite, for lid SP2000H,320°C
90 S 2077
Flange packing DN65 PN6B (67), for SP2000
93 S 0010
Spare thermostat 0-180°C
93 S 0011
Spare thermostat 0-320°C
93 S 0015
Spare heater cartridge HLP, 230V 630W
93 S 0017
Spare heater cartridge HLP, 115V 630W
93 S 0018
Heat conductivity paste for heater cartridge
20 S 4300
Complete set SP2000H/DIL critical orifice incl. 2x O-ring. Material: glass # please indicate ratio with the order.
20 S 4301
Complete set SP2000H/DIL critical orifice incl. 2x O-ring Dilution Ratio 100:1, Flow rate 6,2L/h, Material: glass
93 S 4000
Eductor for SP2000-H/DIL incl. O-ring.
93 S 4002
Eductor for SP2000-H320/DIL, incl. gasket.
93 S 4003
Eductor for SP2000-H320/DIL-BR, incl. gasket.
93 S 4006
Dilution cross 180°C/320°C empty, without any inserts, Material SS316
93 S 4010
Gasket set for eductor of probe SP2000-H320/DIL, 3x graphite-ring. 1x SS-ring 4/6x1,5
93 S 4015
Eductor for SP2000-H/DIL-B, incl. O-ring.
93 S 4016
Bypass T-piece 180°C/320°C empty, without any inserts, Material SS316
93 S4005
Sealing set for critical orifice of SP2000-H320/DIL: 4 pcs. Graphite ring 3/6x1,5mm, 1pc. washer 3,3/6x0,5mm
93 S 4009
Gasket set for critical orifice of probe SP2000-H/DIL and SP2006-H/DIL, Viton.
93 S 4020
Dilution cross complete for SP2000-H/DIL. DIL-ratio to be specified with the order.
93 S 4022
Dilution cross complete for sample probe SP2000-H320/DIL. DIL-ratio to be specified
93 S4025
Bypass-T-piece complete for SP2000-H/DIL-BR.
93 S 4027
T-piece complete for SP2000-H320/DIL-BR.
17 SPARE PARTS
Wear, tear and replacement part requirements depend on specific operating conditions. The following table gives an extract of recommended spares for the dilution probes type
SP2000H/DIL... .
Gas sampling and gas conditioning technology 2-1.1.7-ME
18 APPENDIX
More product documentation is available on our Internet catalogue:
www.mc-techgroup.com.
35
Gas sampling and gas conditioning technology 2-1.1.7-ME
Loading...