PSS-10/1
Embracing Challenge
Portable Gas Conditioning Unit
series PSS®
Version PSS-10/1 for 480 Nl/hr gas ow rate
Special Features
Low maintenance and self-monitoring
Outlet dew point adjustable from +2 °C
to +15 °C
Dew point stability ± 0,1 °C
Operational in 20 minutes
Compact construction
Light weight
Optimum reliability
Jet Stream heat exchangers in 3 stand-
ard materials
Universal equipment possible
Application
The portable gas conditioning unit PSS-10/1
is designed to carry out precise gas analyses
at different locations. The entire gas conditioning unit is housed in a robust aluminiumframe case to enable you to carry out your gas
analysis rapidly, safely and with little maintenance work. The PSS-10/1 gas conditioning
unit is suitable for variable, discontinuous use
as for continuous operation. The components
built into the PSS-10/1 gas conditioning unit
are envisaged for “standard use“. Obviously
our extensive Product Range provides additional components.
Description
The M&C PSS-10/1 gas conditioning unit is
fitted with an two stage ECP20-2 gas cooler
which cools the sample gas to a constant +
5°C irrespective of the ambient temperature.
As soon as the operating temperature of
<+8°C is reached after start up, the gas pump
N9KPE is switched on automatically by the gas
cooler status contact. The SR 25.1 peristaltic
pumps ensures a constant condensate drain,
making long-term measurement possible without problems. The corresponding
particle filtration is carried out by the FP-2T
micron filter. This all makes the portable gas
conditioning unit a complete gas conditioning system for most gas-analysis devices.
Technical specications and illustrations are without
obligation, subject to modications. 10.96/12.11
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
14.6

Flow scheme PSS-10/1
1 Gas sample probe
2 Sample line, 3 m Viton hose
3 Gas cooler ECP 20-2
4 Fine lter FP-2T, lter porosity 2 µm
5 Gas diaphragm pump N9KPE
6 optional owmeter FM 40
7 Pre-lter PF2
8 Peristaltic pump SR25.1
Dimensions
Sample Gas IN
21
ECP20-2
7
8
+5°C
Condensate OUT
63
4
5
Dimensions in mm
power IN
sample
OUT
condensate
OUT
290
555
sample
GL 18
IN
4
3
5
8
GL 25
420
14.6
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
Technical specications and illustrations are without
obligation, subject to modications. 10.96/12.11

Technical Data
Gas Conditioning Unit series PSS® version PSS-10/1
Part No. for 230V 50Hz version 01G5000
Part No. for 115V 60Hz version 01G5000a
Sample outlet dew point range of adjustment: +2 °C ..... +15 °C, factory setting: +5 °C
Dew point stability at constant conditions < ±0,1 °C
Sample inlet temperature **max. 80 °C optional: **max.180 °C with stainless steel bulkhead union
Sample inlet dew point **max.+80 °C
Gas flow rate **max. 480 Nl/hr
Ambient temperature **+5 °C up to +40 °C
Storage temperature -25 °C up to +65 °C
Pressure 0,7 bar up to 1,4 bar abs.
Total cooling capacity **max. 80 kJ/hr
Number of gas inlets 1
Number of gas outlets 1 optional: max. 4
Medium connections tube connection 4/6 mm
Material of sample contacting parts stainless steel, glass, PVDF, Viton®, Novoprene
Ready for operation approx. 10 min.
Power supply 230V 50Hz or 115V 60Hz
Power consumption max. 350VA; with option temperature controller and heated sample line 230V: max.1730VA 115V:max. 1040VA
Fuse protection 4A t, 5x 20 mm, with option temperature controller: 6,3A t
Electrical connection Cold appliance plug with 2 m of cable
Case protection IP20 (DIN 40050. IEC 529)
Housing portable aluminium-frame case
Housing dimensions 555 mm x 460 mm x 290 mm (H x W x D)
Electrical equipment standard EN61010
Weight approx. 24 kg
®
Options Type Part No.
Further sample gas outlet Parallel samplegas outlet, tubing via T-piece on lateral PVDF bulkhead fitting, DN4/6,
01G9065
max. 4 pieces
Flowmeter, max. 4 pieces FM40 7-70 l/hr air, mounted in sample gas outlet
FM40 15-150 l/hr air , mounted in sample gas outlet
FM40 25-250 l/hr air , mounted in sample gas outlet
FM40 50-500 l/hr air, mounted in sample gas outlet
01G9070
01G9075
01G9080
01G9085
Fittings out of PVDF Fittings out of PVDF instead of PP and 3m FPM sample tube DN4/6 01G9025
Sample tube Sample tube out of Kanthal® ø 6 mm, length 1 m, sampling temperature max. 1300 °C 01G9030
Liquid alarm Liquid alarm LA 1/1.4, consisting of: liquid alarm sensor LA1, controller LA1.4, filter glass
01G9035
F120G-D with GL connection incl. mountage/wiring. In case of condensate inrush,
the sample gas pump is automatically switched of.
3-way ball valve 3L/PV-1 for switching over from test gas to sample gas, in the inlet of the sample gas
01G9046
conditioning unit, mounted with mounting brackets, fitting PVDF
5-way ball valve 5L/PV-1 for switching over from test gas to sample gas, in the inlet of the sample gas
01G9045
conditioning unit, mounted with mounting brackets, fitting PVDF
Electronic temperature controller for max. 12 m
heated sample line 100W/m
Electronic temperature controller for max. 12 m
701 control range 0-200 °C, inlet PT100, power 230V 50Hz,
Contact capacity 250V AC max. 4A, completely mounted incl.
7-pin plug 10A
701 control range 0-200 °C, inlet PT100, power 115V 60Hz,
Contact capacity 250V AC max. 4A, completely mounted incl. 7-pin plug 10A
01G9055
01G9055a
heated sample line 100W/m
Connecting adapter DN4/6 for
heated sample line
PSS-10 connecting adapter with bending protection for rigid mountage
of heated sample line with interchangeable PTFE tube DN4/6, consisting of
01G9062
stiffening plate, Swagelok fitting with 4mm cartridge, material: SS316Ti.
Connecting adapter DN6/8 for
heated sample line
PSS-10 connecting adapter with bending protection for rigid mountage
of heated sample line with interchangeable PTFE tube DN6/8, consisting of
01G9063
stiffening Swagelok fitting with 4mm cartridge, material: SS316Ti.
PP = Polypropylene
PTFE = Polytetrauoroethylene (Teon®)
PVDF = Polyvinylidenuoride
** Maximum values in technical data’s must be rated in consideration of total cooling capacity at 25 °C ambient temperature and 5 °C outlet dew point.
Other versions on request.
Technical specications and illustrations are without
obligation, subject to modications. 10.96/12.11
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
14.6