M&C TechGroup PMA 10S Datasheet

PMA 10S
Embracing Challenge
Oxygen Analyser Series PMA®
Portable version PMA 10S for oxygen purity measurement
Special Features
Discontinuous oxygen purity measure-
ment within the expanded range of 99 to 100 volume % O2
weight
Analogue and digital linear
indication
Physical measuring principle
Small dead volume, fast response time
High accuracy and reliability
Application
Due to the extremely fast response time of the patented M&C magneto-dynamic meas­uring cell with no stagnant volume as well as the negligible cross sensitivity from other sample gas components, the portable M&C oxygen analyser PMA 10S does not normally limit its use for a wide variety of applica­tions where oxygen purity measurements are essential.
Description
The PMA 10S oxygen purity analyser, being an extended version of the PMA 10, is provided with an electronic suppression system which enables oxygen purity measureents within the expanded 99-100 vol.% range.
This measuring range is displayed on the analogue meter whilst the digital indicator remains for the 0-100% reading.
The electronic zero suppression system may be selected by means of a push button switch which is located on the control panel of the instrument.
The PMA 10S analyser can normally be used also for its standard PMA10 functions. Oxygen purity measurements could only be made discontinuously. The measuring accuracy of the analyser can be maintained when calibra­tion of the instrument with a certified gas takes place before the measurement proce­dure in order to avoid influence of ambient temperature and pressure. A measurement of 1 vol.% can be easily realized whilst the maxi­mum value remains at 100 vol.%. Other meas­urement within this range can be performed.
– Further technical data see leaflet PMA10,
13.1. –
Measuring principle of M&C ox­ygen analyser
The PMA 10S utilises the paramagnetic principle of operation to measure oxygen concentrations. The analyser measures the paramagnetic susceptibility of the oxygen in the sample gas by means of the M&C magneto-dynamic measuring cell. The physi­cal propertys which distinguishes oxygen from other gases is its paramagnetism. It is significantly higher comparing to other com­mon gases. This operation principle is one of the most accurate and reliable procedures to determine the oxygen concentration in a gas mixture from 0 to 100 Vol.%. The robust M&C cross-flow cell has no stagnant volume. Advantages are the fast response time, the flow rate up to 60 l/hr, the small volume of 2 ml, the extremely low drift, the absolute linearity and the negligible cross sensitivity against other sample gas components. With a proper sample conditioning and pressure, the M&C cell will never need replacing. The dumbbell with a small mirror at its centre is mounted in a strong inhomogenous magnet­ic field. The paramagnetic oxygen strength­ens the forces on the diamagnetic dumbbell and causes a shifting which is detected by a system consisting of light beam, mirror and a photo cell. A compensation current is induced via the feedback coil on the dumb­bell and leads to a reset of the dumbbell into its zero-position.
The required current is linearly proportional to the oxygen concentration.
Technical specications and illustrations are without obligation, subject to modications. 05.97/12.06
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
13.2
Gas flow diagram PMA 10S
1 2
3
260
202
1. Fine lter
2. Flowmeter with needle valve
3. Oxygen measuring cell PMA
Dimensions
Dimensions in mm
Gas
OutIn
Front View
0
Suppression
Output
20
5
0
O -Analyser
2
PMA 10S
Side View
60
40
10
100
0
20,9
30
80
20
3
25
Vol.% O
30
100
2
Zero
Span
150
15
10
13.2
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
Technical specications and illustrations are without obligation, subject to modications. 05.97/12.06
Loading...