M&C TechGroup PMA 100-L Operator's manual

- 1 -
Instruction manual
Microprocessor controlled oxygen analyser PMA
Version PMA100-L
9-3.12.1-ME M&C Products Analysentechnik GmbH
Table of content
1. Electrical standards 3
2. Important safety informations 3
3. Warranty 3
4. Used terms and signal indication 4
5. Introduction 4
5.1 Analyser model ............................................................................................................................ 4
5.2 Patent references ..........................................................................................................................4
5.3 Serial number ...............................................................................................................................4
5.4 Power supply ...............................................................................................................................4
5.5 Mounting system .......................................................................................................................... 4
6. Application 5
7. Description 5
7.1 Measuring principle ......................................................................................................................5
7.2 Flow diagram ................................................................................................................................6
7.3 Dimensions and weight ................................................................................................................7
7.4 Front panel .................................................................................................................................... 7
7.5 Technical data............................................................................................................................... 8
8. Supply connections 9
8.1 Medium ........................................................................................................................................ 9
8.2 Electrical .......................................................................................................................................9
8.2.1 mA output ..................................................................................................................................10
8.2.2 In- and output contacts .............................................................................................................. 10
9. Receipt and storage 11
10. Installation 11
11. Starting up 11
12. Menu description 12
13. Calibration 22
13.1 Calibration ..................................................................................................................................22
13.2 Cross-sensitivity .......................................................................................................................... 22
14. Measuring 25
14.1 Automatically range switch ........................................................................................................25
14.2 Expanded measuring range........................................................................................................ 25
15. Function of in- and output contacts and alarms 26
16. Closing down 27
17. Maintenance and repair 28
18. Trouble shooting 29
19. Spare part list 30
20. Appendix 30
- 2 -
PMA® is a registered trade mark.
7th edition: 12/2004
M&C Products Analysentechnik GmbH 9-3.12.1-ME
- 3 -
1. Electrical standards
The electrical standard corresponds to the safety regula– tions concerning the low-voltage recommendation 73/23 EWG in version 93/68 EWG and the recommendation of electromagnetic compatibility 89/336 EWG in version 93/68 EWG.
We meet the following standards:
EN 61010 part 1 / EN 50081 part 1 / EN 50082 part 1 EN 55014 / EN 60555 part 2 & 3 / EN 60335 part 1
2. Important safety informations
Please note the following basic safety procedures when using this equipment:
• Work on electrical equipment is only to be carried out by trained specialists as per the regulations currently in force.
• Attention must be paid to the requirements of IEC 364 (DIN VDE 0100) when setting high-power electrical units with nominal voltages of up to 1000 V, together with the associated standards and stipulations.
• Check the details on the type plate to ensure that the equipment is connected up to the correct mains voltage.
• Protection against touching dangerously high electrical voltages. Before opening the equipment, it must be switched and hold no voltages. This also applies to any external control circuits that are connected.
• The equipment is only to be set within the permitted range of temperatures and pressures.
• Check that the location is weatherprotected. It should not be subject to either direct rain or moisture.
• The equipment may not be operated in an area at risk from explosion.
• Installation, maintenance, monitoring and any repairs may only be done by authorised personnel with respect to the relevant stipulations.
3. Warranty
If the equipment fails, please contact M&C directly or else go through your M&C authorised dealer. We offer a one year warranty as of the day of delivery as per our normal terms and conditions of sale, and assuming technically correct operation of the unit. Consumables are hereby excluded. The terms of the warranty cover repair at the factory at no cost or the replacement at no cost of the equipment free ex user location. Reshipments must be send in a sufficient and proper protective packaging.
9-3.12.1-ME M&C Products Analysentechnik GmbH
4. Used terms and signal indication
- 4 -
SKILLED STAFF
These are persons with necessary qualification, who are familiar with installation, use and maintenance of the product.
The signals are used according to DIN 4844 and EU Recommendation 91/C53/06.
These are important informations about the product or parts of the instruction manual which require user’s attention.
5. Introduction
5.1 Analyser model
The Oxygen analyser type PMA100-L is produced by M&C Products Analysentechnik in Ratingen, Germany.
5.2 Patent references
The M&C paramagnetic measuring cell is patented in Europe and the USA under the following patent numbers:
• Germany Pat.-Nr. 36 33 750
• France Pat.-Nr. 87 13 608
• United Kingdom Pat.-Nr. 21 96 127
• The Netherlands Pat.-Nr. 188 2449
• USA Pat.-Nr. 4,807,463
5.3 Serial number
The type plate with the serial number is located at the back panel of the analyser. Whenever you call M&C regarding questions or orders for spares please give us the serial number of your PMA.
5.4 Power supply
The power supply for the oxygen analyser PMA100-L is 230V, 50Hz or 115V, 60Hz (‘a’ added to the Serial-No.). For detailed information please look at the type plate of your analyser. Variations of the power supply in a range of +10% to - 15% have no influence on the function of the analyser.
5.5 Mounting system
The analyser is build in a 19“ housing, for rack or table­mounting.
M&C Products Analysentechnik GmbH 9-3.12.1-ME
- 5 -
➡➡
6. Application
The transducer of the PMA100-L works at a stable tempe­rature of +55°C. Therefore the analyser is suitable for con­tinuous measurements of oxygen concentrations in particle­free and dry sample gases. Safe operation, reliability and minimized maintenance are the characteristic of the PMA100-L. The operation of the instrument is based upon the principle of the magneto-dynamic cell which is the most accurate and reliable cell for determining the oxygen content in gas mixtures in a range of 0 to 100 Vol.-%. The patented M&C measuring cell has been improved in order to achieve stability, minimum drift of temperature and extremely fast response time. Due to this fast response time and the negligible cross-sensitivity from other gases the PMA100-L is applicable in a wide range of processes, like:
• monitoring of flue gases,
• inerting installations,
• fermentation processes,
• process- and lab-measurements, etc.
7. Description
7.1 Measuring principle
The paramagnetic susceptibility of oxygen is significantly greater than that of other common gases, and for this reason the molecules of oxygen are attracted much more strongly by a magnetic field than the molecules of other gases. Most of the other gases are slightly diamagnetic, i.e. the molecules are then repelled by a magnetic field.
The principle of the magneto-dynamic cell is based upon Faraday’s method of determining the magnetic suscepti­bility of gas. The cell consists of two nitrogen-filled quarts spheres arranged in the form of a dumb bell. A single turn of platinum wire is placed around the dumb bell which is suspended in a symmetrical non-uniform magnetic field. When the surrounding gas contains oxygen, the dumb bell spheres are pushed out of the magnetic field by the change in the field which is caused by the relatively strong para­magnetic oxygen. The torque acting on the dumb bell will be proportional to the paramagnetism of the surrounding gas and consequently it can be used as a measure of the oxygen concentration.
N
N
2
S
Nitrogen=Diamagnetic Oxygen=Paramagnetic
Fig. 1: Magnetic susceptibility of gases
2
1
1 : Quarts sphere dumb bell 2 : Platinum wire 3 : Mirror 4 : Magnetic pole pieces
Fig. 2: The measuring cell in theory
3
N
S
4
O
2
The distortion of the dumb bell is sensed by a light-beam and projected on a mirror attached to the dumb bell where­of it is reflected to a pair of photo cells (Fig. 3). When both photo cells are illuminated equally the output will be zero. The output from the photo cells is connected to an amplifier, which in turn is fed to the feedback coil of the measuring cell. If the oxygen content of the gas sample changes,
9-3.12.1-ME M&C Products Analysentechnik GmbH
- 6 -
1
1: Measuring cell 2: „LED“ light beam 3: Photo cell 4: Feed back amplifier 5: Output amplifier 6: Meter indication
Fig. 3: Principle of operation
23 54
the corresponding output of the amplifier, which is a current and also proportional to the oxygen content, produces a magnetic field in the feedback coil opposing the forces and thereby causing the dumb bell to rotate. Since the feedback current from the amplifier is proportio­nal to the oxygen content of the gas sample, the output signals produced by the amplifier will be accurate and linear. The paramagnetic susceptibility of oxygen varies inversely as the square of the absolute temperature. There­fore, a temperature sensitive element in contact with the measuring cell assembly is included in the feedback current circuit in order to provide compensation for changes in analyser temperature.
7.2 Flow diagram
The flow can be adjusted in a range of 25 - 60Nl/h air at the flowmeter with needle valve 2 built on the front panel of the analyser. The flow sensor at the outlet of the measuring cell 3 de– tects the sample flow if it decreases under 25l/hr. We recommend a conditioning system upstream the analyser PMA100-L, f.e. consisting of a cooler and fine filter.
6
PMA100-L
1: external filter 2: flowmeter with needle valve 3: patented M&C measuring cell PMA100-L 4: flow control
PMA100-LF
1: external filter 2: flowmeter with needle valve 3: patented M&C measuring cell PMA100-L 4: flow control
Fig. 4: Flow diagram of the analyser
1234
134 2
We like to inform you about suitable M&C equipment.
M&C Products Analysentechnik GmbH 9-3.12.1-ME
- 7 -
7.3 Dimensions and weight
The analyser is build in a 19” housing, also suitable for table mounting. Fig. 5 shows the dimensions of the PMA100-L. Please take additional 60mm fitting-depth into consideration when installing the analyser. The weight of the analyser is approx. 11 kg.
483
177 (4U)
(84HP)
20,9 V0l.% O 1010 mbar
102
⇑⇓
PMA100
465
2
Alarm
440
7.4 Front panel
The following figure shows the front panel of the oxygen analyser PMA100-L. You can see the double-lined LCD display 1 , the alarm LED 2 , the flowmeter with needle valve 3 , status LEDs 4 and the six operating keys 5 . The sample flow can be adjusted at the needle valve in a range of 25-60Nl/hr. The control panel refers to NAMUR standard and is devided into:
Select key
Enter key
• Direction key
• Direction key
⇑⇑
⇑⇑ ⇓⇓
⇓⇓
CAL key
Measuring key
(for more functional description see chapter 12)
341
50
10
7
Fig. 5: Dimensions of the PMA100-L
20,9 V0l.% O
1
1010 mbar
Selec Enter ⇑⇓Cal Meas
PMA100 O2-analyser
Fig. 6: Front panel with display, operating keys
and flowmeter
2
4
5
Alarm
2
3
9-3.12.1-ME M&C Products Analysentechnik GmbH
- 8 -
7.5 Technical data
Part No. 03A3100(a): PMA100-L, power supply 230 VAC, 50 Hz, 115 VAC, 60 Hz
signal: 4-20 mA; (a)=115 V
Measuring ranges 4 linear measuring ranges free selectable, lowest span 1%,
basis parameterizing: 0-2,5; 0-10; 0-25; 0-100 Vol.% O2 *; manual, automatic or remote range control and range indication is possible
Indication, suitable in German, English and French 2 line, 16-sign. LCD-display, resolution 0,01 Vol.%O2, continuous
O2-indication and read off O2-transducer temperature, mA-signal, measur­ing range, time, date, error/alarm message
Output signals selection: isolated 0-20, 2-20, 4-20*, 4-20.5 mA for the selected range,
max. load 500 Ω;
Relay outputs, free configurable 4 potential free relay contacts NO, contact rating max. 48 VDC, 500 mA, 15 W Binary inputs, free configurable potential free, 4 x 12 - 24 VDC, max. 20 mA or internal 24 V Flow alarm caloric conductibility sensor in the outlet of the cell Status alarm for min. flow, transducer temp. < 50 °C, processor error, optional pressure
sensor: LED-indication and potential free contact output,NO, max. 48 VDC, 500 mA, 15 W and mA output signal, f.e. 22 mA
Alarm contact for underflow or exeeding of the measuring range, termination of the
calibration, external alarm, concentration alarm: LED-indication and potential free contact output, NO, max. 48 VDC, 500 mA, 15 W
Response time for 90%-FSD < 3sec at 60 NI/hr air Accuracy after calibration deviation ± 1% of 2-100% span, Reproducibility deviation < 1% of span Influence of ambient temperature no influence up to 50 °C Influence of barometric or process pressure the oxygen reading varies in direct proportion to the baromatric or process
pressure variation
Influence of sample gas flow variation in gas flow between 0 and 60 Nl/hr air will cause a difference in
reading of < 0,1 Vol.%O
2
Sample gas
- inlet pressure 0,01 up to 0,6 bar (PMA100-L requires positive pressure for adequate flow rate, no pump inside)
- outlet pressure outlet of analyser should discharge freely into atmosphere
- flow rate 25 - 60 Nl/hr air
- temperature -10 °C up to +50 °C dry gas
O2-transducer temperature fixed at +55 °C Ambient temperature -10 °C up to +50 °C Storage temperature -20 °C up to +60 °C, rel. humidity 0-90% RH Power supply internal power unit for 230 VAC or 115 VAC available, (a) +/-10%, 40-60 Hz, 35 VA Electrical connections mains supply: 3-pol. chassis plug with 2 m cable; signals: 2 x Sub-D plug Materials in contact with sample gas platinum, epoxy resin, glass, FPM, stainless steel 316, PTFE, PVDF Sample gas connection 1/8“ NPT internal thread*, option with tube connector DN 4/6 PVDF
available part no.: 05V1045
Protection / Electrical standard IP40 (EN60529) / EN61010 Housing/ 19“ rack mounting with front handles Front colour grey RAL 7032 Dimensions/ width: 84HP; hight: 4U; depth: 350 mm + approx. 60 mm installation space Weight approx. 11 kg
supply voltage
DC
* standard/basis execution
M&C Products Analysentechnik GmbH 9-3.12.1-ME
- 9 -
8. Supply connections
8.1 Medium
The oxygen analyser PMA100-L is suitable for continuous measurements of oxygen concentrations in particle-free and dry sample gases. Therefore it is recommended to use a gas conditioning system upstream the analyser, f.e. equipped with a cooler and a particle filter.
We like to inform you about suitable M&C equipment.
The following diagram shows the connections on the back panel of the PMA100-L.
Sample in Sample out
Before opening the assembly cover disconnect the electrical supply!
X1 X2 X3 X4 X5
Power in
Patented measuring cell USA patent number 4807463
For connection of the sample gas in- and outlet use 1/8” NPT male fittings.
We like to inform you about our range of tube and pipe connectors.
8.2 Electrical
The analyser PMA100-L is equipped with an internal power switch. The 2m cable with 3-pole plug at the end is part of the standard supply.
False supply voltage can damage the equipment. When connecting the equipment, please ensure that the supply voltage is identical with the information provided on the model type plate!
Fig. 7: Connections on the back panel of the
PMA100-L
Note!
For the erection of power installations with rated voltages up to 1000V, the requirements of VDE 0100 and relevant standards and specifications must be observed! The main circuit must be equipped with a fuse correspon­ding to the nominal current (over current protection); for electrical details see technical data (chapter 7.5).
9-3.12.1-ME M&C Products Analysentechnik GmbH
mA +
mA -
Fig. 8: 15-pole Sub-D socket X3
Out 4 MC
Out 3 NO
Out 2 MC
Out 1 NO
Alarm MC
Status MC
Out +24V
IN 1,2,4; GND
IN 3 (+24V)
In 1 (+24V)
In 4 (+24V)
1
9
2
10
3
11
4
12
5
13
6
14
7
15
8
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
Out 4 NO
Out 3 MC
Out 2 NO
Out 1 MC
Status NO
Out 0V
IN 3; GND
Alarm NO
In 2 (+24V)
- 10 -
8.2.1 mA output
The mA output is available on the back panel of the PMA100-L (see fig.8) at the 15-pole Sub-D socket X3. The following figure shows the configuration of the terminal.
The menu-driven handling of the mA outputs is described in chapter 12.
8.2.2 In- and output contacts
Fig. 9 shows the configuration of the terminal X4. The following connections are available at the 25-pole Sub-D plug:
• four binary inputs, In1 to In4, with 12V - 24V, max. 20mA,
• one supply power contact, Out +24V and Out 0V, with 24V, max. 20mA,
• four binary output contacts, Out1 to Out4, with max. 24V, 400mA,
• one alarm contact, Alarm MC and Alarm NO, with 48V, max. 500mA, 15W, and
• one status contact output, Status MC and Status NO, with 48V, max. 500mA, 15W.
Fig. 9: 25-pole Sub-D plug X4
The menu-driven handling of the in- and outputs is described in chapter 12.
M&C Products Analysentechnik GmbH 9-3.12.1-ME
- 11 -
9. Receipt and storage
The PMA100-L is completely pre-installed and normally delivered in one packaging unit.
• Please take the analyser and possible special acces­sories carefully out of the packaging material immediate­ly after arrival, and compare the goods with the items listed on the delivery note;
• Check the goods for any damage caused during delivery and, if necessary, notify your transport insurance com­pany without delay of any damage discovered.
Note!
The oxygen analyser PMA100-L must be stored in a whea­ther-protected frost-free area!
10.Installation
The PMA100-L is built in a 19“ housing, which is also suitable for table mounting. Accurate and proper installation of the PMA100-L analyser will not only minimize instrument breakdown, but it will also result in reliable operation of the analyser. The operator must be satisfied that the analyser installation and positioning is safe for extremes of conditions which could occur in the operating environment of the analyser. Choose installation sites which are reasonable free from vibration sources, and are not subjected to large tempera­ture fluctuations outside the analyser specifications. Without any precautions avoid any back pressure different from barometric pressure at the gas outlet of the analyser.
Note that the PMA100-L analyser is only suitable for measuring of non-hazardous gas mixtures in non-hazar­dous areas!
11.Starting up
Before using the equipment for the first time, check that the safety measures specific to the installation and process are complied with.
Before connecting the analyser to the mains, compare the mains voltage with the information on the type plate of the analyser.
Note!
The working temperature of the analyser is 55°C. After
starting, the analyser warms up. The current temperature is displayed. During the warm-up time the menu of the analyser is locked for use.
9-3.12.1-ME M&C Products Analysentechnik GmbH
12.Menu description
12.1 Table of content
Overview menu-drive and Overview operating keys ........................................................... 13
12.2 Warming up ..................................................................................................................................... 14
12.3 Alarm log-book ............................................................................................................................... 14
12.4 Parameterising level 1 ...................................................................................................................... 15
12.4.1 Language ............................................................................................................................... 15
12.4.2 Select range ........................................................................................................................... 15
12.4.3 Set range ................................................................................................................................ 15
12.4.4 Set autorange hysteresis ........................................................................................................ 15
12.4.5 Current output ....................................................................................................................... 15
12.4.6 Status error mA ...................................................................................................................... 16
12.4.7 Set date/time ......................................................................................................................... 16
12.4.8 Relay configuration ................................................................................................................ 16
12.4.9 Input configuration ................................................................................................................ 17
12.4.10 Average ................................................................................................................................. 17
12.4.11 Concentration alarm .............................................................................................................. 18
Display concentration alarm .................................................................................................. 18
Alarm range ........................................................................................................................... 18
Set alarm beep ...................................................................................................................... 18
12.5 Parameterising level 2 ...................................................................................................................... 19
12.5.1 Step into the menu................................................................................................................ 19
12.5.2 LCD contrast .......................................................................................................................... 19
12.5.3 Calibration method ............................................................................................................... 19
12.5.4 Current signal with calibration ............................................................................................... 19
12.5.5 Current pressure .................................................................................................................... 20
12.5.6 Password ............................................................................................................................... 20
Change password level 1 ...................................................................................................... 20
Change password level 2 ...................................................................................................... 20
12.5.7 Inquiry of the transducer channel .........................................................................................20
12.5.8 Read NV-Memory Editor ....................................................................................................... 20
12.6 Zero- and span calibration .............................................................................................................. 21
12.6.1 Zero gas concentration ......................................................................................................... 21
12.6.2 Span gas calibration .............................................................................................................. 21
12.6.3 Manual calibration ................................................................................................................. 21
12.6.4 Deviation from calibration ..................................................................................................... 21
12.6.5 Calibration log-book ............................................................................................................. 22
- 12 -
M&C Products Analysentechnik GmbH 9-3.12.1-ME
- 13 -
Overview menu-drive Overview operating keys
Display
Status
Alarm log-book
Entry code
Code 1
Language
Code 2
LCD contrast
Meas
CAL
Select
Enter
↑↓
Read only
see code1
• Startup of measuring menu
• Function: to leave the actual menu structure to the measuring display
• Startup of calibration
• Selection of menu points
• Selection of changeable positions
• Jump into submenu
• Confirmation of input
• Selection of submenu points if shown on the display
CAL
Zerogas concentration
Measuring range
Current output
Date/Time
Relay­configuration
Relay NC/NO
Keep alarm yes/no
Input­configuration
Average
Concentration­alarm
Calibration­method
Current pressure
Password
Baud rate
NV-Memory
Spangas concentration
Calibration
Deviation from calibration
Calibration log-book
Alarm range
Alarm signal
9-3.12.1-ME M&C Products Analysentechnik GmbH
MC Para O
TM M&C Products
2
V . . . .
- 14 -
12.2 Warming up
After switching on the analyser type PMA100-L, the warming-up begins. The following display appears for approximately 20 seconds:
M&C logo
• current O2-concentration
• Software-version implemented in the PMA100-L
• and the trademark.
press: 1005 mbar
press: 1005 mbar
heating 22.5°C
Select
alarm log-book
Select
After further 20 seconds the display changes. Via the Direc­tion keys ⇑⇓ the following current values can be indicated:
• measuring range
• temperatur of the transducer
⇑⇑
⇓⇓
⇑⇑
⇓⇓
• pressure in the measuring-cell
• date/time
• current output. The second line displays the heating procedure. The warming-up is finished by reaching a temperature of 54°C. The display shows either the current value (see above) or an error message. The second line of the display presents the current oxygen concentration.
The Select key leads to the next menu point.
12.3 Alarm log-book
The starting time and the 9 last alarms are stored. The Direction keys ⇑⇓ can be used going through the alarm log­book. The current alarm message is shown on the display
like E ....
The Select key leads to the entry of the access code.
enter to config.
Code: 0000
Enter
enter to config.
Code: 0000
Select
Enter
enter to config.
Level 1
Select
The Enter key opens the access code menu for the first, second respectively read only level. The first changeable position is underlined. With the Direc­tion keys ⇑⇓ the value of the position can be changed in between 0 and 9. A jump to the next position happens via the Select key. The entry of the complete code has to be confirmed by the Enter key and is shown in the second line of the display. The codes are preadjusted at the factory: Code:0010 level 1
⇑⇑
⇓⇓
⇑⇑
⇓⇓
Code:1000 level 2 Code:0000 read only-level;
The read only-level allows a passage through all menu points of level 1 via the Select key. In this level parameters can not be changed.
The Select key leads to level 1 - measuring parameters,
⇑⇑
⇓⇓
⇑⇑
⇓⇓
relay-, input and output configuration.
M&C Products Analysentechnik GmbH 9-3.12.1-ME
- 15 -
12.4 Parameterising level 1
12.4.1 Language
The following languages are selectable by the Direction keys ⇑⇓ :
• English
• German
• French Enter confirms the choice.
The Select key leads to the next menu point.
12.4.2 Select range
After the entry of the code for level 1 the display opens to the select range menu. Four variable ranges and the auto­range are available. The functions can be selected by the Direction keys ⇑⇓ . With the autorange always the favour­able range is used. In combination with the autorange function, unused measuring ranges are requested to be set to 100% O2 (see 12.4.3).
The Select key leads to the next menu point.
12.4.3 Set range
The desired ranges 1 to 4 or the autorange are selected by the Direction keys ⇑⇓ . The display is opened for changes by the Enter key. The first changeable position is underlined and can be changed operating the Direction keys ⇑⇓ . A jump to the next position happens via the Select key. The entry of the complete measuring range has to be confirmed by the Enter key. (lowest range: 1% span) The Select key leads to the next menu point.
language
English
Enter
Select
select range 1
r1(...4), autorange
Enter
Select Enter
Select
set range 1
xx%O2 - xxx,x%O
Enter
Select Enter
Select
⇑⇑
⇓⇓
⇑⇑
⇓⇓
⇑⇑
⇓⇓
⇑⇑
⇓⇓
2
⇑⇑
⇓⇓
⇑⇑
⇓⇓
12.4.4 Set autorange hysteresis
This adjustment is important using the autorange function. The %-value is related to the smallest span of the relevant switching ranges and determines the interval switching into the next lower range (see 14.1). The entry starts with the Enter key. The first changeable position is underlined and can be changed operating the Direction keys ⇑⇓. A jump to the next position happens via the Select key. The entry has to be confirmed by the Enter key, and the Select key leads to the next menu point.
12.4.5 Current output
Two adjustments are possible:
1. fixed output signal and
2. output signal with range of tolerance showing the underflow respectively exceeding of the signal range. The following ranges are preseted:
• 0 - 20mA 0 - 20mA + 0,45mA
• 2 - 20mA 2 - 20mA ± 0,45mA
• 4 - 20mA 4 - 20mA ± 0,45mA
set hyst. autor.
xx.xx%
Enter
Select Enter
Select
current output
0-20mA, 2-20mA, ...
⇑⇑
⇓⇓
⇑⇑
⇓⇓
⇑⇑
⇓⇓
⇑⇑
⇓⇓
9-3.12.1-ME M&C Products Analysentechnik GmbH
- 16 -
Enter
Select
status error mA
no, 0mA, 2mA, 20,5mA, ...
⇑⇑
⇑⇑
Enter
Select
set date/time
MM-DD-YY hh:mm
Enter
Select
⇑⇑
⇑⇑
A selection is done by the Direction keys ⇑⇓ and confirmed by the Enter key.
The Select key leads to the next menu point.
12.4.6 Status error mA
Using a permanent current output signal, status errors can be identified. The output signal is selected via the Direction keys ⇑⇓ and confirmed by the Enter key.
⇓⇓
⇓⇓
The following output signals are available:
no output signal
0 mA 2 mA
20,5 mA 21 mA
22 mA 22,5 mA
•23mA • 24mA The Select key leads to the next menu point.
12.4.7 Set date/time
The Enter key enables the entry of the date/time values: first month, day and year, then hour and minute. The first changeable position is underlined and can be changed operating the Direction keys ⇑⇓ .
⇓⇓
⇓⇓
A jump to the next position happens via the Select key. The entry has to be confirmed by the Enter key, and operating once more the Select key leads to the next menu point.
relay config.
relay 1
Enter
relay config.
R1: span1...4,conc. alarm1...4
Enter
Select
relay nc = 1
1:0 2:0 3:0 4:0
Enter
Select Enter
⇑⇑
⇑⇑
⇑⇑
⇑⇑
⇑⇑
⇑⇑
12.4.8 Relay configuration
This menu point determines the function of maximum four relays, R1 to R4 (see 7.5 technical data). A selection of the relays is done by the Direction keys ⇑⇓ .
⇓⇓
⇓⇓
The Enter key opens the Entry and the Direction keys ⇑⇓ select the desired function:
• allocation of the measuring ranges 1 to 4
• concentration alarm (1 to 4)
• calibration
• flow alarm
⇓⇓
⇓⇓
• external alarm (see 12.4.9). The Enter key confirms the selection. The Select key leads to the next menu point.
This menu point selects whether the relay functions as: 0 normally open contact 1 normally closed contact Alarm- and status relay contacts are always closed in the ‘ok-status’ (safety first). A configuration is not possible.
⇓⇓
⇓⇓
The entry starts with the Enter key. The first changeable position is underlined and can be changed operating the Direction keys ⇑⇓ . A jump to the next position happens via the Select key. The entry has to be confirmed by the Enter key, and the
Select
M&C Products Analysentechnik GmbH 9-3.12.1-ME
Select key leads to the next menu point.
- 17 -
An alarm situation can be kept. If yes, one input must be configured as a reset (see 12.4.9). The entry starts with the Enter key. The first changeable position is underlined and can be changed operating the Direction keys ⇑⇓ . A jump to the next position happens via the Select key. The entry has to be confirmed by the Enter key, and the
Select key leads to the next menu point.
12.4.9 Input configuration
Inputs 1 to 4 are selected by the Direction keys ⇑⇓ (see 7.5). The Enter key opens the entry. The Direction keys ⇑⇓ select one of the following input functions:
• external alarm (passing through an incoming signal)
• set span 1 to 4
• external alarm reset; Resets all alarms which are adjusted in menu point ‘keep alarm’ to 1. For the reset a pulsated signal is ne­cessary.
The choice has to be confirmed via the Enter key.
keep alarm = 1
1:0 2:0 3:0 4:0
Enter
Select Enter
Select
input config.
input 1...4
Enter
input config.
external alarm
Enter
⇑⇑
⇓⇓
⇑⇑
⇓⇓
⇑⇑
⇓⇓
⇑⇑
⇓⇓
⇑⇑
⇓⇓
⇑⇑
⇓⇓
The Select key leads to the next menu point.
12.4.10Average
This menu point opens the possibility to determine an average value over a maximum measuring time of 100 seconds. This average value is displayed. The desired value can be selected via the Direction keys ⇑⇓.
The Enter key operates the confirmation and the Select key leads to the next menu point.
Select
average
no, 1, ..., 100 seconds
Enter
Select
⇑⇑
⇓⇓
⇑⇑
⇓⇓
9-3.12.1-ME M&C Products Analysentechnik GmbH
conc. alarm 1...4
[< or >] xx.xx%
Enter
Select
Select
conc. alarm 1...4
yes/no
Select Enter
Select
Select
- 18 -
12.4.11Concentration alarm
The user is able to set maximum four threshold values in Vol.-% O2 as a minimum or maximum alarm limit. The values 1 to 4 are selected by the Direction keys ⇑⇓ . The Enter key opens the entry. The cursor jumps to the first changeable
⇑⇑
⇓⇓
⇑⇑
⇓⇓
position, the <-/>-sign for minimum respectively maximum limit. A selection is operated via the Direction keys ⇑⇓ . A
⇑⇑
⇓⇓
⇑⇑
⇓⇓
jump to the next position happens via the Select key. The entry has to be confirmed by the Enter key. The Select key leads to the next menu point.
Display concentration alarm The following menu point determines whether a concen­tration alarm is displayed as:
• alarm contact,
• status contact, or
⇑⇑
⇓⇓
⇑⇑
⇓⇓
• no. A selection happens via the Direction keys ⇑⇓, and is confirmed by the Enter key. If ‘no’ is selected, the alarm signal can be used as a relay control output (see 12.4.8). The Select key leads to the next threshold value. After reaching the 4th value the Select key leads to the next menu point.
alarm hysteresis
xx.xx%
Enter
alarm hysteresis
xx.xx%
Select
Select Enter
set beep
off/on
Enter
Select
Alarm range The alarm range is determined by a percentile of the alarm threshold value. The Enter key opens the menu for the entry of above mentioned value. The first changeable position is under–
⇑⇑
⇓⇓
⇑⇑
⇓⇓
lined and can be changed operating the Direction keys ⇑⇓ .
A jump to the next position happens via the Select key. The entry has to be confirmed by the Enter key, and opera­ting once more the Select key leads to the next menu point.
Set alarm beep
The optical alarm display can be supported by a beep alarm signal. The function beep on or off is selected via the Direction keys ⇑⇓ . The Enter key confirms the choice.
⇑⇑
⇓⇓
⇑⇑
⇓⇓
Operating the Select key leads to the beginning of the menu.
Anytime you want to start up the menu anew press the
Meas
M&C Products Analysentechnik GmbH 9-3.12.1-ME
Measuring key.
- 19 -
12.5 Parameterising level 2
12.5.1 Step into the menu
System parameters can be changed in level 2. Operating the MEAS key leads back to the configuration level. Operating two times the Select key leads from the measu­ring menu to the entry of the access code (see 12.2).
Select
Enter to config.
Code: 0000
The Enter key opens the menu. The first changeable position is underlined and can be changed via the Direction keys ⇑⇓ . Operating the Select key jumps to the next position. The entry of the complete code (1000 preadjusted at the factory) has to be confirmed by the Enter key.
The Select key leads to the next menu point.
12.5.2 LCD contrast
The display background gets darker with increasing ambient temperatures. To get a better contrast, the letters on the display can be changed to more light colour. The Enter key opens the entry display. The desired contrast is set by the Direction keys ⇑⇓ inbetween a range of 0 to 7 (light coloured letters). The Enter key confirms the entry.
The Select key leads to the next menu point.
12.5.3 Calibration method
It is possible to choose or enable the calibration method. The PMA100-L only allows manual calibration.
Enter
Enter to config.
Code: 0000
Select Enter
Select
LCD-contrast
(0...7):0
Enter
Enter
Select
cal. menu point
manual cal.
⇑⇑
⇓⇓
⇑⇑
⇓⇓
⇑⇑
⇓⇓
⇑⇑
⇓⇓
Select Enter
With Enter you step into the calibration menu. The desired function is set by the Direction keys ⇑⇓ , and confirmed by the Enter key.
Select leads to the next menu point.
manual cal.
yes/no
Select
Enter
⇑⇑
⇓⇓
⇑⇑
⇓⇓
12.5.4 Current signal with calibration
During the calibration a mA-signal can be output as:
cal. current
off mA-signal of the last measuring value on current signal of the respective calibration gas
The Select key leads to the next menu point.
9-3.12.1-ME M&C Products Analysentechnik GmbH
off/on
Select
current pressure
.....mbar
Enter
- 20 -
12.5.5 Current pressure
The Enter key opens the display to change the value for the current pressure (normal pressure at the respective location). The modified pressure value is confirmed by the Enter key.
Check current pressure value when carrying out zero and span calibration.
Select
set password 1
Code: 0001
Enter
set password 1
Code: 0001
Select Enter
Select
Select
D:x.xxx P :xxxx
T:xx.xx ADC :0...7
Enter
Enter
Select
⇑⇑
⇑⇑
⇑⇑
⇑⇑
The Select key leads to the next menu point.
12.5.6 Password
Change password level 1
The preadjusted password can be changed by a four-digit individual code. Enter opens the submenu. The first changeable position is underlined and can be changed operating the Direction keys ⇑⇓ . A jump to the next position happens via the Select key. The code must contain a figure > 0. The entry has to be confirmed by the Enter key.
⇓⇓
⇓⇓
Operating once more the Select key leads to the next menu point.
Change password level 2
Follow the steps in chapter 12.7.3 Attention! Password 1 and password 2 must be different. With Select you leave level 2 to the measuring menu.
12.5.7 Inquiry of the transducer channel
The display shows on the first line the analogue value (D) of the transducer channel in [mV] and optional the current pressure (P) in [mbar]. The second line displays the current transducer temperature (T) in [°C], the selected transducer channel (ADC), and a number to change the display from the
⇓⇓
⇓⇓
standard (1) to the configuration of the relays, alarms and inputs (0). Operating the Enter key enables the choice of the trans­ducer channel (0-7, only standard configuration). The following are functional: 6 = current analogue value ‘O2 -concentration’, and 7 = current analogue value ‘temperature’. A renewed operation of the Enter key confirms the entry, and the Select key leads to the next menu point.
12.5.8 Read NV-Memory Editor
read NV-memory
..$.........0
The NV-Memory Editor describes the basic respectively the programming level of the analyser. An access to this level is only possible with the approval of the manufacturer.
Every time it is possible to leave the actual menu structure to the measuring display operating the Measuring key on
Meas
M&C Products Analysentechnik GmbH 9-3.12.1-ME
the front board of the analyser.
- 21 -
12.6 Zero- and span calibration
Operating the Cal key leads to the display represented beside. A jump into the entry mode of the zero calibration happens automatically.
12.6.1 Zero gas concentration
Enter opens the display. The Direction keys ⇑⇓ operate the input of the zero gas concentration (cross sensitivities see chapter 13.2). The first changeable position is underlined and can be changed operating the Direction keys ⇑⇓ . A jump to the next position happens via the Select key. The entry has to be confirmed by the Enter key.
The value for the zero gas concentration must be lower than the value for the span gas concentration!
Operating once more the Select key leads to the next menu point.
12.6.2 Span gas concentration
The input of the span gas concentration happens similar to point 12.6.1.
The value for the span gas concentration must be higher than the value for the zero gas concentration!
zero conc.
xx.xx%O
Enter
zero conc.
(-)xx.xx%O
Select Enter
Select
Enter
span-conc.
00.00% O
Cal
2
2
⇑⇑
⇓⇓
⇑⇑
⇓⇓
2
Operating once more the Select key leads to the next menu point.
12.6.3 Manual calibration
The Enter key starts the calibration procedure. After reaching a stable level for the zero gas concentration (line 2 displays the current value) pressing the Enter key leads automatically to the span calibration. The deviation in %O to the last calibration is shown on the display (value „d“ ).
After reaching a stable level for the span gas concentration (line 2 displays the current value) pressing the Enter key closes the calibration procedure, and the analyser jumps automatically into the measurement level. The deviation in %O2 to the last calibration is shown on the display (value „d“ ) during the calibration procedure. A fault during the calibration procedure is represented on
the display as E ..... followed by the message: O2 calib.
The fault is also stored in the calibration log-book. The Select key leads to the next menu point.
The next menu point can be reached stepping into the cali­bration mode via the Cal key and passing the previous menu points with the Select key.
Select Enter
Select
Manual calib.
press Enter
2
Enter
0 calibration
00.00 d: (-)00.00
Enter
span calibration
00.00 d: (-)00.00
Select
Select Cal
⇑⇑
⇑⇑
⇓⇓
⇓⇓
12.6.4 Deviation from calibration
The display shows the deviation from the basic calibration
calib. diff.
carried out in the factory. The deviation from the zero point is given as Vol.%-O2 and the offset in % of the possible
ofs: (-)x.x% g: x.x%
deviation (±100%).
9-3.12.1-ME M&C Products Analysentechnik GmbH
- 22 -
Select
cal. log-book
M 10.11.99 11:o7
Select
Select leads to the following menu point.
12.6.5 Calibration log-book
It is possible to read back the data (date and time) of the last 9 calibrations via the Direction keys ⇑⇓ .
The kind of calibration is marked as follows: M manual calibration.
A fault during the calibration is displayed as:
E M 10.11.99 11:07
Select leads to the beginning of the calibration.
13.Calibration
A calibration is recommended latest every 4 weeks. The step into the calibration menu happens via the CAL key or the serial interface.
13.1 Calibration
Zero and span calibration happen sequencely. The zero gas is given to the ‘Sample in’ connection, located on the back panel of the PMA100-L oxygen analyser (see 8.1, fig. 7). A detailed description of the menu-driven handling is given in chapter 12. Faults during the calibration procedure are displayed in the main menu like:
E ... O2-calibration
The fault message is also stored in the calibration log-book with an E ... followed by the shortcut for the calibration mode, the date, and the time the calibration is carried out. If the zero and the span gas concentration deviates more than 100% to the adjusted concentration, a calibration cannot be carri­ed out. The stop at the calibration is displayed as an error message.
13.2 Cross-sensitivity
The paramagnetic measuring principle is based on the very high magnetic susceptibility of oxygen. In comparison to oxygen, other gases have such a minor susceptibility, most of them are not even worth mentioning. Exception to this are the nitrogen oxides. However, as this gas is in most cases present in a very low concentration, the error is still negligible. The next table shows a list of cross-sensitivities. All values based on a zero calibration with 100 Vol.-% N2 and a span calibration with 100 Vol.-% O2. The cross-sensitivities in the table are valid for 100 Vol.-% of the corresponding gas.
M&C Products Analysentechnik GmbH 9-3.12.1-ME
- 23 -
Gas Formula + 20°C+50°C
Argon Ar - 0,23 - 0,25
Acetylene C2H
2
Acetone C3H6O - 0,63 - 0,69
Acetaldehyde C2H4O - 0,31 - 0,34
Ammonia NH
Benzene C6H
Bromine Br
Butadiene C4H
Methyl propene C4H
n-Butane C4H
Chlorine Cl
3
6
2
6
8
10
2
Hydrogen chloride HCL - 0,31 - 0,34
Nitrous oxide N2O - 0,20 - 0,22
Diacetylene (CHCl)
Ethane C2H
Ethylen oxide C2H4O
Ethylene C2H
Ethylene glycol (CH2OH)
Ethylbenzene C8H
2
4
2
4
2
10
Hydrogen fluoride HF + 0,12 + 0,14
Furan C4H4O - 0,90 - 0,99
Helium He + 0,29 + 0,32
n-Hexane C6H
14
Krypton Kr - 0,49 - 0,54
Carbon monoxide CO - 0,06 - 0,07
Carbon dioxide CO
Methane CH
Methylen chloride CH2Cl
2
4
2
Neon Ne + 0,16 + 0,17
n-Octane C8H
18
Phenol C6H6O - 1,40 - 1,54
Propane C3H
Propylene C3H
8
6
Propylene oxide C3H6O - 0,90 - 1,00
Propylene chloride C3H7Cl - 1,42 - 1,44
Monosilane SiH
Styrene C8H
Nitrogen N
4
8
2
Nitrogen oxide NO + 42,70 + 43,00
Nitrogen dioxide NO
Oxygen O
Sulphur dioxide SO
Silphur fluoride SF
2
2
2
6
Hydrogen sulphide H2S - 0,41 - 0,43
Toluene C7H
8
Vinyl chloride C2H3Cl - 0,68 - 0,74
Vinyl fluoride CH3F - 0,49 - 0,54
Water (steam) H2O - 0,03 - 0,03
Hydrogen H
2
Xenon Xe - 0,95 - 1,02
- 0,26 - 0,28
- 0,17 - 0,19
- 1,24 - 1,34
- 1,78 - 1,97
- 0,85 - 0,93
- 0,94 - 1,06
- 1,10 - 1,22
- 0,83 - 0,91
- 1,09 - 1,20
- 0,43 - 0,47
- 0,54 - 0,60
- 0,20 - 0,22
- 0,78 - 0,88
- 1,89 - 2,08
- 1,78 - 1,97
- 0,27 - 0,29
- 0,16 - 0,17
- 1,00 - 1,10
- 2,45 - 2,70
- 0,77 - 0,85
- 0,57 - 0,62
- 0,24 - 0,27
- 1,63 - 1,80
0,00 0,00
+ 5,00 + 16,00
+100,00 +100,00
- 0,18 - 0,20
- 0,98 - 1,05
- 1,57 - 1,73
+ 0,23 + 0,26
Fig. 10: Table of cross-sensitivities
9-3.12.1-ME M&C Products Analysentechnik GmbH
- 24 -
Example :
The residual oxygen percentage should be measured in a closed 100% carbon dioxide (CO2) atmosphere. The „zero­calibration“ is done by means of Nitrogen (N2). According to the list of cross-sensitivities the error for CO at 20°C is -0,27%. In order to obtain a higher accuracy this means for the calibration that the reading should be adjusted at +0,27% with N2, in order to compensate the error of CO2. Since the values of cross-sensitivities are based on 100 Vol.-% of that particular gas, the error at 50 Vol.-% CO2 and 50 Vol.-% N2 is -0,135%.
2
M&C Products Analysentechnik GmbH 9-3.12.1-ME
- 25 -
14.Measuring
The connections for sample gas inlet and outlet are located on the back panel of the PMA100-L oxygen analyser (see
8.1).
A detailed description of the menu-driven handling is given in chapter 12.
Note! The oxygen analyser PMA100-L is suitable for continuous measurements of oxygen concentrations in particle-free and dry sample gases. Therefore it is recommended to use a gas conditioning system downstream the analyser equipped with a cooler and a particle filter.
14.1 Automatically range switch
The function of the automatically range switch is pre-selected in menu point 12.4.2 or via the serial interface. It is recom­mended to set unused measuring ranges to 100% O (see 12.4.3). The analyser uses the suitable measuring range automatically. Fig. 11 shows the switching between two overlapping measuring ranges, 0 - 50Vol.-% and 30 - 40Vol.-%. Going through the measuring ranges upward the analyser switches exact at the range limits, in this example at 30Vol.-% respectively 40Vol.-%. Going down through the measuring ranges the autorange hysteresis determines the interval switching into the next lower range. The %-value, which is set in menu point 12.4.4, is related to the smallest span of the switching relevant ranges. The resultant value in Vol.-% O2 is added to the respected range. The switching relevant range in the example (30 - 40Vol.-%) is 10Vol.-%. This means a switching interval of 1Vol.-% if the adjusted hysteresis is 10%. The switching points are at 39Vol.-% respectively 29Vol.-%. The mA-output is related to the extended measuring range 2. This means: 29Vol.-% - 0 mA and 40Vol.-% - 20 mA .
2
up
29 30 39 40 Vol.-%
0 1,82 18,12 20 mA
Fig. 11: Automatically range switch
hysteresis 10%
Meas. range 1
0 - 50Vol.-%
Meas. range 2 30 - 40Vol.-%
down
14.2 Expanded measuring range
Expanded measuring ranges are selected in menu point 12.4.3. The minimum range is 1Vol.-% O2. The selected mA-output signal (see 12.4.5) is related to the respective span. Example:
• measuring range 99 - 100Vol.-%,
• 4 - 20mA.
This means a resolution of 1/16Vol.-% O2 per mA.
9-3.12.1-ME M&C Products Analysentechnik GmbH
15.Function of in- and output contacts and alarms
The following in- and output contacts are available at the Sub-D socket X4 at the back panel of the analyser (specification see 8.2.2):
• 4 binary inputs,
• 4 relay output contacts,
• 1 common relay alarm contact, and
• 1 common relay status contact.
The binary inputs can be reserved for the following funct– ions (configuration see 12.4.9):
external alarm: an external alarm signal, f.e. a cooler- or
liquid alarm from the conditioning system upstream the analyser, can be locked on. This signal releases a respec– tiv alarm message (alarm LED and display on the front panel of the PMA100-L). If one of the output contacts is locked on with ‘external alarm’ (see 12), the input signal is passing through.
set span: the measuring ranges 1 to 4 can be chosen
externally.
reset of stored alarms: alarms which are stored after re-
lease (see 12.4.9) can be resetted with a pulsated signal.
- 26 -
The function of the output contacts are as follows (configu­ration see 12.4.9):
measuring range: this function gives a feedback about
the current active range. The ranges 1 - 4 can be selected.
conc. alarm: it is possible to lock max. 4 concentration
alarms on the relay outputs (configuration see 12.4.11).
calibration: a signal is given when the analyser works in
the calibration mode.
flow alarm: in addition to the internal signals like alarm
LED, display or status contact the flow alarm can be locked on one of the relay output contacts.
external alarm: an external alarm signal can be passed
to one of the relay output contacts (see above).
Alarms are signalised via the alarm LED and the analyser display. In addition to this a common alarm relay contact is available. An exception is the concentration alarm. The alarm limits are fixed in menu point 12.4.11. The display of the concentration alarm (alarm LED, analyser display, com­mon output or status alarm) must be confirmed separately in menu point 12.4.11. An allocation of the concentration alarm can be realised via the relay contact outputs (see above).
M&C Products Analysentechnik GmbH 9-3.12.1-ME
- 27 -
The following alarm messages are available at the common alarm output:
• concentration alarm 1 - 4 (if allocated, see above)
Status messages like:
• flow alarm,
• transducer temperature,
• hardware fault, and
• leaving the current measuring range.
are signalised via the alarm LED, the analyser display and the status contact output. Menu point 12.4.6 enables to identify a status fault via a permanent mA-output signal. It is available at the Sub-D socket X3 (see 8.2.1). The mA-measuring signal is overlaid by the status signal output.
16.Closing down
• In case of a temporary closing down of the process con– trol systems around the analyser, the AC mains supply of the analyser must remain „ON“. No further precautions are required.
• In case of a closing down of the analyser for a longer period, it is recommended to flush the analyser with clean air in order to prevent that harmful gases will be left in the instrument and damage the measuring cell.
9-3.12.1-ME M&C Products Analysentechnik GmbH
17.Maintenance and repair
Before carrying out any maintenance or service activities, the engineer concerned must ensure that the analyser is disconnected from the electrical supply of power!
• Maintenance of the analyser should be performed by qualified personnel, and the frequency of maintenance may vary as a result of operating experience.
• Any used spare parts must be specified as M&C spare parts.
• Routine maintenance of the PMA100-L oxygen analyser is limited to frequently checking the zero and span point cali-bration. If the values are not correct, the instrument should be recalibrated according to the procedure as outlined in this instruction manual.
• In the event of an incorrect reading of the analyser or a system failure make certain that the sampling system is in a good condition and that all sample handling components are working properly.
- 28 -
• Check that gas cooler, filters, condensate drain or other components are not blocked with dirt or any foreign matter. Ensure that the sample gas is connected correctly and is flowing properly through the system, and make a visual inspection of the analyser for loose or leaking connections.
Note!
It is strictly recommended that the repair of the analyser should be performed by high qualified as well as experienc­ed personnel, and that any spare part which is used should be specified as M&C spare part !
M&C Products Analysentechnik GmbH 9-3.12.1-ME
- 29 -
18.Trouble shooting
Fault description Display Solution
Measuring cell doesn‘t work E:01 resp. F:01, depending on Give air resp. nitrogen to the analyser properly at 100 Vol.-% the language and restart the analyser (suppressed measuring range 99-100 Vol.-%)
Calibration fault:span gas is Measuring value always 0 Vol.-%; Start new calibration (first calibration is switched off while proceed- no fault message on the display a ‘blind’ calibration; the displayed value ing the calibration of the calibration is registered in the is always 0 Vol.-%); after this start span. calibration log-book with another calibration under ‘normal’ con-
exeeded calibration limits ditions. (e.g. 104%).
Calibration limits of the A/D- Check the limits (see below) testing the converter exeeded. converter channel (see menu);
When limits are exeeded a calibration at the manufacturer has to be carried out.
Limits testing the converter channel (see menu):
Checking the zero point with nitrogen N
U
[V] U
min
max
[V]
2
0,005 0,147
Checking the span with ambient air (20,93 Vol.-%) at normal pressure and different voltages of the zero point:
Zero point [V] U
[V] U
min
max
[V]
0,005 0,306 0,344 0,078 0,370 0,398 0,147 0,430 0,457
Checking the span with 100 Vol.-% oxygen O2 at normal pressure and different voltages of the zero point:
Zero point [V] U
[V] U
min
max
[V]
0,005 1,440 1,580 0,078 1,470 1,610 0,147 1,500 1,630
9-3.12.1-ME M&C Products Analysentechnik GmbH
19.Spare part list
Wear, tear and replacement part requirements depend on specific operating conditions. The recommended quantities are based on experience and are not binding.
Oxygen Analyser Type PMA100-L
(C) consumable parts (R) recommended spare parts (S) Spare parts
recommended quantity PMA being
in operation
(years)
Part.-No. Indication C/R/S 1 2 3
- 30 -
90 A 0010 Measuring cell PMC1 S - - 1 90 A 0015 Flowmeter glass 7-70 Nl/hr S - - 1
20.Appendix
For further documentation: Look on our homepage:
www.muc-products.de
M&C Products Analysentechnik GmbH Rehhecke 79 • D - 40885 Ratingen
Tel: 02102 - 935 0 Fax: 02102 - 935 111
info @ muc-products.de
M&C Products Analysentechnik GmbH 9-3.12.1-ME