M&C TechGroup PMA 10 Datasheet

Page 1
PMA 10
Embracing Challenge
Oxygen Analyser Series PMA®
Portable version PMA 10 for mobile oxygen measurement
Special Features
Portable and rugged instrument, light-
weight
Analogue and digital linear indication
Small stagnant volume, fast response
time
High accuracy and reliability
Approved according to DIN EN 14181
resp. to 13th and 17th BImSchV and TÜV certicate TA-Luft as option
Application
Due to the extremely fast response time of the M&C magneto-dynamic measuring cell with no stagnant volume as well as the negligible cross sensitivity from other sample gas components, the portable M&C oxygen analyser PMA 10 has a wide variety of applica­tions. The analyser is a suitable and reliable instrument for monitoring oxygen concentra­tions in various gas analytical control applica­tions including flue gas-, inert gas-, fermen­tation processes-, food packing machines-, ambient air- and laboratory process control measurements.
Description
The portable M&C oxygen analyser PMA 10 is an non-thermostated instrument which has been designed for continuous and discontin­uous measurement of oxygen concentrations in dry and particle-free sample gases.
The PMA 10 is a reliable and easy-to-operate instrument and immediately operable. The light-weight instrument is built into a port­able housing. The four measuring ranges are displayed on the analogue meter with 30%/100% scale and the 100% O2 range also on the digital meter. One output signal 0-1V is available as standard. Sample gas connections as well as a connector for the output signal are located on the front panel of the analyser. The connectors for incoming power supply and optional O2 alarm contact are located at the rear panel. The sample gas enters the analyser via an internal protec­tive fine-filter. The required flow rate can be adjusted at the flowmeter with needle valve, mounted on the front panel upstream the M&C measuring cell.
Options: mA-output signal, O2 alarm, rechargeable battery, internal pump and TÜV certificate.
Measuring principle of M&C ox­ygen analyser
The PMA 10 utilises the paramagnetic principle of operation to measure oxygen concentrations. The analyser measures the paramagnetic susceptibility of the oxygen in the sample gas by means of the M&C magneto-dynamic measuring cell. The physi­cal property which distinguishes oxygen from other gases is its paramagnetism. It is signifi­cantly higher comparing to other common gases. This operation principle is one of the most accurate and reliable procedures to determine the oxygen concentration in a gas mixture from 0 to 100 Vol.%. The robust M&C cross-flow cell has no stagnant volume. Advantages are the fast response time, the flow rate up to 60 l/hr, the small volume of 2 ml, the extremely low drift, the absolute linear­ity and the negligible cross sensitivity against other sample gas components. With a proper sample conditioning and pressure, the M&C cell will never need replacing. The dumbbell with a small mirror at its centre is mounted in a strong inhomogenous magnetic field. The paramagnetic oxygen strengthens the forces on the diamagnetic dumbbell and causes a shifting which is detected by a system con­sisting of light beam, mirror and a photo cell. A compensation current is induced via the feedback coil on the dumbbell and leads to a reset of the dumbbell into its zero-position. The required current is linearly proportional to the oxygen concentration.
Technical specications and illustrations are without obligation, subject to modications. 05.97/08.09
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
13.1
Page 2
Gas flow diagram PMA10
1 2 4 3
O -Analyser PMA 10
2
Alarm
Output
30
10
3
100
0
Span
Zero
0
0
Vol.% O
2
30
10
5
20
15
40
20
60
25
80
100
20,9
OutIn
Gas
202
150
260
1. Fine filter
2. Flowmeter with needle valve
3. Oxygen measuring cell PMA
4. Option internal pump
Dimensions
Dimensions in mm
Front View
Side View
13.1
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
Technical specications and illustrations are without obligation, subject to modications. 05.97/08.09
Page 3
Technical Data
Version PMA 10 portable oxygen analyser
Part No. 01A1000 : PMA 10, power supply 230V 50Hz, output signal 0-1 V; 01A1000a = 115V 60Hz Measuring ranges selectable for 0-3, 0-10, 0-30 and 0-100 vol.% O2, linear Indication analogue / digital meter:
analogue meter selectable for each range with a scale of 0-30 and 0-100% digital meter, 3
Output signal 0-1V DC, non-isolated, load > 100 KΩ, for each selected range;
option: 0-20mA* or 4-20mA* for each selected range, non-isolated, max. load 300Ω Part no. 01A9000 Response time for 90% FSD < 3 seconds at 60 Nl/hr air Accuracy after calibration analogue = ± 1% of span / digital = ± 0,1 vol.% O2 error of precision Reproducibility analogue = < 1% of span / digital = ± 0,1 vol.% O2 error Influence of ambient temperature zero point ± 0,02 vol.% O Influence of barometric pressure The oxygen reading varies in direct proportion to changes of the barometric pressure. Influence of sample gas flow variation in gas flow between 0-60 Nl/hr air will cause a difference of < 0,1 vol.% O Sample gas inlet pressure 0,01 up to 1 bar g, (PMA 10 required admission pressure for competent flow rate, no pump inside)
option: PMA 10 with internal pump, capacity 0,9 Nl/min. without pressure, Part no. 01A9100 Sample gas outlet pressure Outlet of analyser must discharge freely into atmosphere. Flow rate of sample gas max. 60 Nl/hr air, adjustable with needle valve on the flowmeter 7-70 Nl/hr Temperature of sample gas -10 °C up to +40 °C, dry gas Analyser temperature according to ambient temperature, non heated version Ambient temperature -10 °C up to +55 °C Storage temperature -20 °C up to +60 °C, relative humidity 0-90% RH Power supply
internal power unit for 230VAC standard or 115VAC available (a)* +/-10%, 40-60Hz, 8VA
option: rechargeable battery, recharged by the internal power unit, Part no. 01A9050
battery capacity 10 hr, recharge time of battery 14 hr with instrument off Electrical connections mains supply: 3-pole chassis plug with 2m of cable; signal: 3-pole plug Materials in contact with sample gas Platinum, Glass, Polypropylene, Stainless Steel 316, FPM, Epoxy resin Sample gas connection PP-hose nipple for DN 11-4 mm tube Protection / electrical standard IP 41 EN 60529 / EN 61010 Housing / colour portable plastic housing / gray Dimension / weight height 150 mm, width 202 mm, depth 260 mm / approx. 3 kg Alarms
option: O2 alarm low* or high* adjustable 0-100% O2, relays contact NC/NO Part no. 01A9150
additional with acoustic signal and auto reset after 30 sec. Part no. 01A9155
additional with acoustic signal and manual reset, Part no. 01A9156 Certificate option: approved according to DIN EN 14181 resp. to 13th and 17th BImSchV and TA-Luft,
Part no. 01A9160
* Please specify with order. In case the options alarm and pump are both required, a rechargeable battery is automatically obliged. In case the option “acoustic alarm” required, the option “O2 alarm” is obliged. Option “TÜV approved” includes 0*/4*-20mA signal output, as further options only rechargeable battery and internal pump available.
1
/
digit 9 mm high LCD for 0-100 % O2 reading, selectivity 0,1 vol.% O
2
/ °C; sensitivity ± 0,1 vol.% O
2
/ °C
2
2
.
2
W A R N I N G ! I M P O R T A N T !
An external fine filter must always be used at the gas inlet of the analyser. Depending on the composition of the sample gas, it may be necessary to use a sample conditioning system. Without precautions, the analyser is only suitable for measuring of non-hazardous gases or gasmixtures in non-hazardous areas.
Technical specications and illustrations are without obligation, subject to modications. 05.97/08.09
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
13.1
Loading...