M&C TechGroup MP47-Z-BPR150 Datasheet

MP47-Z-BPR150
Embracing Challenge
Analytical Diaphragm Pump Series MP®
Version MP47-Z-BPR150 with integrated bypass admission pressure regulator
Special Features
With integrated bypass admission pres-
sure regulator
Corrosion resistant
Gas tight
Maintenance free
Long service life
Application
The diaphragm pump MP47-Z-BPR150 is suit­able for the 100 % oil-free delivery of corrosive gases with changing pressure conditions. The capacity and construction of the pump is especially designed for the problems in the analyze technique. The pump is gas-tight and operates without any maintenance.
Description
All sample contacting parts of the diaphragm pump MP47-Z-BPR150 are made of PTFE, PVDF and Viton or Kalrez. The lifted gas remains analytically pure due to the fact that the pump works absolutely without any lubri­cant. A special diaphragm and valve system guaranties a long life time and no mainte­nance is necessary. The pump is available for 230V or 115V.
One problem in the analyze technique, when lifting gases via filters, is that the filters are clogged after a longer period of time. This causes an increasing differential pressure above the filter and subsequently, a higher suction pressure for the pump is necessary. This, however, results in a decrease of the adjusted pump capacity.
In order to avoid this negative influence, the M&C diaphragm pump MP47-Z-BPR150 is equipped with a fully adequate bypass admis­sion pressure regulator inside the pump head. Independent from the changing entry pres­sure, the bypass admission pressure regula­tor keeps the pressure at the pump outlet constant, so that the adjusted pump capacity resulting thereof remains constant as well.
For example, the entry pressure may vary from 0,46 bar abs. to 1,1 bar abs. for a pump capacity of 250Nl/h.
The outlet pressure is adjusted at works to 200mbar overpressure (adjustable up to 300mbar).
Technical specications and illustrations are without obligation, subject to modications. 01.08
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
10.4
Technical Data
Diaphragm pump MP47-Z-BPR150
-Viton, 230V
MP47-Z-BPR150
-Viton, 115V
MP47-Z-BPR150
-Kalrez, 230V
MP47-Z-BPR150
-Kalrez, 115V
Part No. 02 P 1160 02 P 1160a 02 P 1165 02 P 1165a
Voltage 230V 50Hz 115V 60Hz 230V 50Hz 115V 60Hz
Material of sample
PTFE, PVDF, Viton PTFE, PVDF, Kalrez
contacting parts
Protection class IP 20 - DIN 40050
Pump capacity max. 6,6 l/min* in the range of adjustment (outlet pressure)
Operating pressure max.* 1,1 bar abs.
Operating pressure min.** 0,46 bar abs. for 250Nl/h / 0,62 bar abs. for 400Nl/h (outlet pressure 1,20 bar abs.)
Outlet pressure 1,15 – 1,3 bar abs. / at works 1,20 bar abs.
Gas temperature -30 to +80°C
Ambient temperature -10 to +40°C
Storage temperature -15 to +60°C
Power consumption 110W
Current consumption 0,75A 1,7A 0,75A 1,7A
Gas connections G1/4“ i DIN ISO 228/1
Electrical standard EN 61010 part 1
Weight 4,6 Kg
* Maximum operating pressure for a constant outlet pressure and as a result a constant ow rate of the measuring gas. ** The minimum operating pressure that still guarantees a constant pump capacity is changing according to the necessary gas volume ow adjusted behind the pump and to the outlet
pressure adjusted on the pressure regulator. The lower the necessary gas volume ow and the outlet pressure, the lower is he minimal possible operating pressure that still guarantees a constant outlet pressure and therefore a constant sample gas volume ow.
Pump head with integrated bypass admission pressure regulator
Bypass admission pressure regulator
Dimensions
G1/4"
Sample gas IN OUT
34
10.4
18 80
47
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
4xM6 10 tief (deep)
247
160
66
110
Technical specications and illustrations are without obligation, subject to modications. 01.08