M&C TechGroup MP26-H1 Datasheet

MP26-H1
Embracing Challenge
Analytical Diaphragm Pump Series MP®
Version MP26-H1 heated at 180 °C
Special Features
Heated up to 180 °C
Corrosion resistant
With excess temperature limit switch
failure
100% lubricant free
Gas tight
Maintenance free
Long service life
Application
This M&C-pump MP26-H1 is suitable for cor­rosive gases. It is constructed especially for hot gas analysis technique.
Description
The M&C-pump MP26-H1 has an electrically heated head out of PTFE. The temperature is controlled by an integrated compact capillary sensor thermostat with an excess tempera­ture limit switch and an alarm function for temperature failure. All sample contacting parts are made out of PTFE, the diaphragm mounting disc is out of nickel-plated stain­less steel.
The pump works absolutely lubricant free, so gases remain analytically unchanged. Due to a special diaphragm and valve system, the pump operates maintenance-free.
In order to guarantee an operation without interferences, it is important to pay attention to the permissible overpressure. Therefore you must decrease or regulate the quantity of gas via a needle valve on the sucking side or via a pressure control valve or bypass on the pressure side.
Technical specications and illustrations are without obligation, subject to modications. 09.96/11.06
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
10.8
Dimensions
Electrically heated pump MP26-H1
3
1
250
6
71
300
1 : 2x PG13, 1x PG9; 2 : Heater elements; 3 : Thermometer, indicator for 50 - 250 ¡C
60
Dimensions in mm
Performance characteristics MP26-H1
Electrical connections
To > +30 C
Reset
G1/4"
34
2
11
90
175
Sensor
100W 100W
M
Tsoll
0-180 C
Tu < -30 C
1 2 3 4
N L L'
Power
230V 50Hz
(ext. contact to switch pump motor Off)
9
121110
8
765
Status
contact
1,5A
On
I>
900
800
20 °C
180 °C
700
600
500
l/hr
400
300
max.
200
100
0
0,0 0,2 0,4 0,6
0,8 1,0 1,2 1,4 1,6 1,8 2,0 2,2 2,4 2,6 2,8 3,0
bar abs.
Technical Data
Diaphragm pump, heated series MP
Part number 02P1300
Capacity max. 10,0 l/min at 180 °C, without pressure
Operating pressure max. 0,3 to 2,5 bar abs.
Sample temperature -20 to +200 °C
Ambient temperature -10 to +40 °C
Storage temperature -15 to +60 °C
Operation mode 100% continous duty, start of the pump only without pressure!
Power supply Motor = 230V 50Hz*, 1,1 A, incl. protective motor switch *Standard
Degree of protection IP 44 - EN60529 *Standard
Temperature adjustment 0-180 °C, (180 °C set in the factory)
Low temperature alarm contact Alarm point: ∆T 30 °C, contact rating: 250V, 3A~, 0,25A =
Weight 6,5 kg
Sample connection G1/4" i DIN ISO 228/1
Material of sample contacting parts PTFE, (and stainless steel nickel-plated)
®
MP26-H1 10 l/min 230V 50Hz
Heater = 230V 50Hz*, 2,0 A *Standard
10.8
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
Technical specications and illustrations are without obligation, subject to modications. 09.96/11.06