M&C TechGroup FT-H2 Datasheet

FT-...H2
Embracing Challenge
Electrically heated universal lter series FT®
Version FT-...-H2, heated at 180 °C
Special Features
Electrically heated, max. 180 °C
Consequent connection technique
No cold spots
Status alarm
Easy maintenance
Application
The M&C electrically heated universal fil­ter reliably separates off solids, in particular extremely fine particles, occurring in analysis techniques in gas filtration, using very fine, deep-acting filter elements. It is constructed especially for hot gas analysis technique at 180 °C.
Description
The universal filter out of PTFE with integrated filter elements (fineness 3 or 0,1 µm) is placed in an electrically heated 2-part aluminium oven construction.
The temperature is controlled by an inte­grated compact capillary sensor thermostat with an excess temperature limit switch with reset key at +30 °C above T an alarm function for low temperature alarm at -30 °C below TSet with voltage free status­alarm contact.
The thermostat as well as the electrical termi­nals are placed in the junction box.
The temperature of the filter is shown on a thermometer. For isolation and protection reasons, the heated filter part is covered with a cap. To prevent heating interruptions, the in- and outlet connection area is heated too. The clamps are fixed in order to support the sample lines. The filter in- and outlet can be turned 180° on the oven top position to achieve easy mounting and flexible adapta­tion to local circumstances.
SET
and
Technical specications and illustrations are without obligation, subject to modications.10.96/11.06
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
7.4
Dimensions
58
135
Ø 6,5
75
300
280
330
350
TYP. NR. VOLT
To>+30°C
Tref.
0-180°C
Tu<-30°C
12 3
VA
Reset
4
6
5
87
'L
N
L
PG 13,5
Alarm contact Temperature min.
Power 230V,50Hz (115V,60Hz)
Dimensions in mm
Technical Data
Filter series FT
Part number 01 F 3006 01 F 3008 01 F 3007 01 F 300. -a
Filter finess / material of filter element 3 µm / glass 3 µm / stainless steel 0,1 µm / glassfibre
Max. pressure 4 bar g
Sample temperature max. 180 °C (M&C-Filters to max. 350 °C on request)
Ambient temperature max. 50 °C
Stagnant space 65 cm
Filter surface 70 cm
Power supply 230/240V 50Hz, 350VA 115V 60Hz
Electrical connection 2,5 mm2 terminals, 2x PG13 cable glands
Protection classification / electrical standard IP 44-EN60529 / EN61010, EN 60519-1
Regulation of temperature thermostat # PT100 as option #
Temperature adjustment 0-180 °C, (180 °C set in the factory)
Temperature alarm contact Alarm point: ∆T 30 °C at T
Dimension 350 x 300 x 140 mm (w x h x d)
Weight 6 kg
Method of mounting wall mounting
Sample connections G1/4" i DIN ISO 228/1
Material of sample contacting parts PTFE, glass, FPM
®
FT-3G-H2 FT-3SS-H2 FT-0,1GF-H2 Option 115V version
3
2
, contact rating: 250V, 3A*, 0,25A =
SOLL
Differential pressure in mbar with clean filter elements
7.4
60
50
40
30
20
10
Differential pressure mbar
0
0 50 100 150 200 250 300 350 400 450 500
FT-3SS-H2 FT-3G-H2 FT-0,1GF-H2
air flow in l/hr at 20 °C
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
Technical specications and illustrations are without obligation, subject to modications. 10.96/11.06