![](/html/bd/bd63/bd63cc17e70cc3f04dc484fa67f4d56a54d8c18566a93bfe3f92d24acab82350/bg1.png)
FP-BF
Embracing Challenge
Humidier series FP®
Version FP-BF
Special Features
Easy relling
Direct control of the liquid level
Optimum humidifying eect
Low stagnant space
Wall mounting
Application
Typically, the FP-BF humidifier is used when
the humidity of the sample gas and the
calibration gas differs and the analyser shows
a significant cross-sensitivity against water
vapour. Before entering the cooler, the zero
gas passes the humidifier and leaves the
cooler saturated at a stable dew point as well
as the sample gas. Test gases with watersoluble components must not be passed via
the humidifier.
Description
The M&C FP-BF humidifier consists of our
standard filter components. With the F-120G
or F-120G-D glass body, the FP-BF humidifier
has only a volume of 70 ml.
The gas passes a bubbler that intensifies
the humidifying effect. The glass humidifier
body enables the control of the liquid level
without dismantling the body. No tools are
required for refilling the humidifier. The optimal position of the o-ring guarantees tightness between head and glass body.
The humidifier can be rotated about 180° in
the holder for wall mounting. This allows a
flexible adaptation of in- and outlet to local
conditions.
For special applications, the FP-BF-GL25 can
be provided with a GL25-12 connection at the
bottom of the glass body.
Overpressure at the outlet or underpressure
at the inlet of the humidifier must be avoided.
Otherwise the liquid will be displaced to the
gas inlet.
Technical specications and illustrations are without
obligation, subject to modications. 09.96/11.06
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
7.10
![](/html/bd/bd63/bd63cc17e70cc3f04dc484fa67f4d56a54d8c18566a93bfe3f92d24acab82350/bg2.png)
Dimensions
Humidifier FP-BF
58
42
13
G1/4" G1/4"
182
ø 5
Humidifier FP-BF-GL25 Partial view of glass body
with GL25-12 connector on the bottom
123
*
* constructional size 290
Dimensions in mm
GL18-6 connector with sealing ring for 6mm tube
Technical Data
Humidifier FP-BF FP-BF-GL25
Part number 03F5700 03F5750
Sample contacting parts:
head part, connector,
tube, o-ring,
bubbler, glass body
Contents approx. 70 ml
Flow rate max. 100 l/hr
Operatig temperature -20 °C to +80 °C
Storage temperature -30 °C bis +110 °C
Operating pressure max. 5 bar abs. at 20 °C 2 bar abs. / 5 bar abs. with GL- connection adapter
Method of mounting wall mounting, position of mounting vertical
Absorption material standard scope of supply without absorption material
Connection G1/4" i DIN ISO 228/1 G1/4" i, GL25-12
Weight 0,3 kg
PVDF
FPM,
glass
7.10
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
Technical specications and illustrations are without
obligation, subject to modications. 09.96/11.06