M&C TechGroup FP-BF Datasheet

FP-BF
Embracing Challenge
Humidier series FP®
Version FP-BF
Special Features
Easy relling
Direct control of the liquid level
Optimum humidifying eect
Wall mounting
Application
Typically, the FP-BF humidifier is used when the humidity of the sample gas and the calibration gas differs and the analyser shows a significant cross-sensitivity against water vapour. Before entering the cooler, the zero gas passes the humidifier and leaves the cooler saturated at a stable dew point as well as the sample gas. Test gases with water­soluble components must not be passed via the humidifier.
Description
The M&C FP-BF humidifier consists of our standard filter components. With the F-120G or F-120G-D glass body, the FP-BF humidifier has only a volume of 70 ml.
The gas passes a bubbler that intensifies the humidifying effect. The glass humidifier body enables the control of the liquid level without dismantling the body. No tools are required for refilling the humidifier. The opti­mal position of the o-ring guarantees tight­ness between head and glass body.
The humidifier can be rotated about 180° in the holder for wall mounting. This allows a flexible adaptation of in- and outlet to local conditions.
For special applications, the FP-BF-GL25 can be provided with a GL25-12 connection at the bottom of the glass body.
Overpressure at the outlet or underpressure at the inlet of the humidifier must be avoided. Otherwise the liquid will be displaced to the gas inlet.
Technical specications and illustrations are without obligation, subject to modications. 09.96/11.06
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
7.10
Dimensions
Humidifier FP-BF
58
42
13
G1/4" G1/4"
182
ø 5
Humidifier FP-BF-GL25 Partial view of glass body with GL25-12 connector on the bottom
123
*
* constructional size 290
Dimensions in mm
GL18-6 connector with sealing ring for 6mm tube
Technical Data
Humidifier FP-BF FP-BF-GL25
Part number 03F5700 03F5750
Sample contacting parts: head part, connector, tube, o-ring, bubbler, glass body
Contents approx. 70 ml
Flow rate max. 100 l/hr
Operatig temperature -20 °C to +80 °C
Storage temperature -30 °C bis +110 °C
Operating pressure max. 5 bar abs. at 20 °C 2 bar abs. / 5 bar abs. with GL- connection adapter
Method of mounting wall mounting, position of mounting vertical
Absorption material standard scope of supply without absorption material
Connection G1/4" i DIN ISO 228/1 G1/4" i, GL25-12
Weight 0,3 kg
PVDF FPM, glass
7.10
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
Technical specications and illustrations are without obligation, subject to modications. 09.96/11.06