FM-200K-H
Embracing Challenge
Flowmeter series FM®
Heatable up to 180 °C Version FM-2K
Electrically heated up to 180 °C,
with needle valve at the inlet
Version FM-200K-H
Special Features
Heatable version, max 180 °C
Electrically heated version, max.
180 °C
Needle valve integrated in the inlet
Large scaling
Sure connection technique
No cold bridges
Integrated thermostat
Status contact output
Easy to maintain
Optional with ow alarm FA...
Application
The M&C flowmeters FM-2K and FM-200K-H
are used in analysis technique to
control gas flow up to an operating temperature of 180 °C.
Depending on the flowmeter glass tube and
the floating ball out of glass, ranges from 16
l/h to 800 l/h can be realized.
Description
The M&C flowmeters FM-2K and FM-200K-H
consist of a vertical cone shaped glass tube
with a free movable float ball inside. The
sample gas, flowing from the bottom to the
top, lifts the float ball as far as an annular gap
appears between the tube wall and the float
ball and the forces on the ball are compensated. The position of the float ball is related
to a scale on the flow meter tube.
The FM-2K is a basic heat resistant version for installation in heated systems.
The electrically heated version FM-200K-H is
equipped with an integrated needle valve
and is fixed on a mounting plate, decoupled
from heat and covered with an insulated
enclosure. The heater consists of a heating
element with high capacity. The temperature
is adjustable on the integrated thermostat up
to 180 °C with high temperature limiter and
low temperature alarm.
The window of the heat insulated enclosure
guarantees a proper reading-out. The electrical connection box with integrated thermostat is installed outside the enclosure on the
mounting plate. A flow control is possible
consisting of fibre-optical light guides FO1,
the sensor head FA2-H and the pre-amplifier
K-FA-H which is installed on the mounting
plate as well. The electronic controller FA-1..
(option) has to be mounted externally. – See
data sheet 9.5 –
In order to avoid cold bridges, the connecting fittings are equally heated by means of
double-ended thermal conducting jaws. The
electrically heated sample lines type 3/4/5N/M/H – see data sheet 3.1 – are fixed with
mounting brackets.
Technical specications and illustrations are without
obligation, subject to modications. 03.95/11.06
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
9.3
![](/html/95/9540/95406a95c7bdf8da6125fe7a19e0e04304c41e5ff56123e96627ae50f1902486/bg2.png)
Dimensions
Version FM-200K-H
120
Sample
Sample
IN
OUT
Heatable version FM-2K
sample OUT
260
230
145
210
230
1/4"NPT
Ø 6.5
Terminals
Pre-amplifter
Part Number
.
Temperature min.
Contact output
Power ON
K-FA-H
80x160mm
02E4010
to FA-1...
480
500
Version FM-200K-H/FA
120
Sample gas
OUT IN
145
210
230
ø 6.5
K-FA-H
Pre-amplifier
Terminals
to FA-1...
80x160mm
Contact OUT
Temperature min.
Power IN
480
500
3.2.
1.
option
50
Sample IN
Optional flow monitoring
Dimensions in mm
36
50
min. 100
Flow rate in Nl/hr air and differential pressure in mbar
50
40
30
20
mbar
10
0
150 300 450 600
0
Nl/hr
750
900 1050
Optional flow monitoring consisting of:
1 Sensor head FA2-H 02E4002
2 Fiber-optical light guides FO1 02E4060 /4065 /4063
3 Pre-amplier K-FA-H 02E4010
4 Electronic controller FA-1... 02E7300 / 7300b / 7300d
02E7110 / 7110a / 7110d
9.3
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
Technical specications and illustrations are without
obligation, subject to modications. 03.95/11.06