FA2-H, 2 x F01, K-FA-H, FA-1.1
Embracing Challenge
Optical ow monitoring series FA®
Version FA1-H-.. for temperatures up to 180 °C
with 2 light guides for monostable functioning
Special Features
Easy mounting of the sensor head at the
ow meter without dismounting the
measuring glass
Also applicable for very low ow quanti-
ties
Also suitable for non-metallic or very
small oating balls
Easy adjustment of any MIN alarm set-
point
Application
The optical flow monitoring unit with light
guides is used in gas analysis on heated
flowmeters with temperatures > 60 °C. The
measuring cones must be transparent.
Thanks to optical scanning, very low flow
quantities can be recorded even in case of
flowmeters with non-metallic or very small (1
mm) floats.
Description
The M&C flow monitoring unit FA1-H-.. consists of 4 sub-assembly modules:
1. the patented sensor head FA2-H,
2. the 2 light guides FO.. for monostable
function
3. the K-FA-H pre-amplifier
4. the electronic controller FA-1.1 or
FA-1.4**
– ** see data sheet FA-.., K-FA 9.5 –
The LED emitters’ light beam passes through
the FO .. light guides and falls upon the
photo-transistor through the FM1-H
flowmeter measuring glass. As soon as the
float breaks the light beam, the photo-transistor is blacked out. The FA-1.. electronic controller analyses this changed status accordingly.
By the monostable version a MIN alarm function in the lowest sensor setting position is
guaranteed.
Technical specications and illustrations are without
obligation, subject to modications. 12.97/11.06
The sensor head FA2-H is assembled on the
FM1-H flowmeter by means of a pressure
screw with its stationary open prism to the
measuring glass. Assembly is simple and does
not require any disassembly of the flowmeasuring glass.
The FO1 light guide is supplied with a standard length of 600 mm. Light guide FO3 (900
mm long) and FO2 (1200 mm long) can be
supplied for greater length requirements.
The angled light-guide ends are fixed into the
FA2-H sensor head with one pressure screw
each. Two light-guides are needed for the
monostable function.
By use of the pre-amplifier K-FA-H a distance
of 200 meters between the flow meter and
the electronic controller FA-1.. (see 9.5) is possible without problems.
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
9.6

2
FO...
FA2-H
4
3
2
1
FA-...
K-FA-H
FA2-H
heated
FM1-H
K-FA-H
Flow monitoring FA1-H-.. on flowmeter FM1-H
Technical Data
1 piece sensor head FA2-H
2 pieces light guides FO1/FO2/FO3 for monostable function
1 piece K-FA-H pre-amplifier
1 piece electronic controller FA-1.1 or FA-1.4
Sensor Head FA2-H* Light guide FO1 Light guide FO3 Light guide FO2 pre-amplifier K-FA-H
Part Number 02 E 4002 02 E 4060 02 E 4063 02 E 4065 02 E 4010
Function monostable 1x 2 pieces 1x
Measuring tube 5-14 mm
Light guide length 600 mm 900 mm 1200 mm
Dimensions W x D x H 24 x 40 x 24 ø 6 mm 80 x 160 x 55
Weight 90 g 230 g 330 g 420 g 560 g
Material anodized aluminium Glass fibre, SS 316Ti, brass-chromed Polycarbonate
Assembly Fixed with clamping screw Wall mounting
Cable entry terminal range 5mm - 10mm
Electrical connection Terminals 2,5 mm
2
Supply voltage 12V DC internal from FA-1..
Electrical standard EN61010
Operating temperature - 25 °C to +180 °C -25 °C to +60 °C
Storage temperature - 25 °C to +70 °C
System of protection IP65 EN 60529
(* Other mounting designs for the light guides on request.)
Separate electronic controller required for the K-FA-H pre-amplifier
Electronic controller FA-1.1 FA-1.4
Part No. 230V 50/60Hz 02 E 7300* 02 E 7110
Part No. 115V 50/60Hz 02 E 7300* 02 E 7110 a
Part No. 24V DC 02 E 7300 d 02 E 7110 d
Part No. 24V AC 02 E 7300 b 02 E 7110 d
Reversing power consumption 230V 50/60Hz / 115V 50/60Hz, adjusted at works: 230V 50/60Hz
Technical data: see data sheet FA-1.1.., FA-1.4, K-FA 5-6.10.2
Complete flow monitoring unit FA1-H-.. :
Complete flow monitoring unit Type
consisting of :
Sensor FA2-H (02 E 4002) 1 piece
Light guide FO1, 600 mm long (02 E 4060) 2 pieces
Light guide FO3, 900 mm long (02 E 4063) 2 pieces
Part Number
FA1-H-6
02 E 4270
FA1-H-9
02 E 4273
FA1-H-12
02 E 4275
Light guide FO2, 1200 mm lang (02 E 4065) 2 pieces
Pre-amplifier with adapter K-FA-H (02 E 4010) 1 piece
Electronic controller FA-1.1 / 230V/115V (02 E 7300) 1 piece