M&C TechGroup FA1-H Datasheet

FA2-H, 2 x F01, K-FA-H, FA-1.1
Embracing Challenge
Optical ow monitoring series FA®
Version FA1-H-.. for temperatures up to 180 °C with 2 light guides for monostable functioning
Special Features
Easy mounting of the sensor head at the
ow meter without dismounting the measuring glass
ties
Also suitable for non-metallic or very
small oating balls
Easy adjustment of any MIN alarm set-
point
Application
The optical flow monitoring unit with light guides is used in gas analysis on heated flowmeters with temperatures > 60 °C. The measuring cones must be transparent.
Thanks to optical scanning, very low flow quantities can be recorded even in case of flowmeters with non-metallic or very small (1 mm) floats.
Description
The M&C flow monitoring unit FA1-H-.. con­sists of 4 sub-assembly modules:
1. the patented sensor head FA2-H,
2. the 2 light guides FO.. for monostable function
3. the K-FA-H pre-amplifier
4. the electronic controller FA-1.1 or FA-1.4**
– ** see data sheet FA-.., K-FA 9.5 –
The LED emitters’ light beam passes through the FO .. light guides and falls upon the photo-transistor through the FM1-H flowmeter measuring glass. As soon as the float breaks the light beam, the photo-tran­sistor is blacked out. The FA-1.. electronic con­troller analyses this changed status accord­ingly.
By the monostable version a MIN alarm func­tion in the lowest sensor setting position is guaranteed.
Technical specications and illustrations are without obligation, subject to modications. 12.97/11.06
The sensor head FA2-H is assembled on the FM1-H flowmeter by means of a pressure screw with its stationary open prism to the measuring glass. Assembly is simple and does not require any disassembly of the flow­measuring glass.
The FO1 light guide is supplied with a stand­ard length of 600 mm. Light guide FO3 (900 mm long) and FO2 (1200 mm long) can be supplied for greater length requirements.
The angled light-guide ends are fixed into the FA2-H sensor head with one pressure screw each. Two light-guides are needed for the monostable function.
By use of the pre-amplifier K-FA-H a distance of 200 meters between the flow meter and the electronic controller FA-1.. (see 9.5) is pos­sible without problems.
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
9.6
2
FO...
FA2-H
4
3
2
1
FA-...
K-FA-H
FA2-H
heated
FM1-H
K-FA-H
Flow monitoring FA1-H-.. on flowmeter FM1-H
Technical Data
1 piece sensor head FA2-H
2 pieces light guides FO1/FO2/FO3 for monostable function
1 piece K-FA-H pre-amplifier
1 piece electronic controller FA-1.1 or FA-1.4
Sensor Head FA2-H* Light guide FO1 Light guide FO3 Light guide FO2 pre-amplifier K-FA-H
Part Number 02 E 4002 02 E 4060 02 E 4063 02 E 4065 02 E 4010
Function monostable 1x 2 pieces 1x
Measuring tube 5-14 mm
Light guide length 600 mm 900 mm 1200 mm
Dimensions W x D x H 24 x 40 x 24 ø 6 mm 80 x 160 x 55
Weight 90 g 230 g 330 g 420 g 560 g
Material anodized aluminium Glass fibre, SS 316Ti, brass-chromed Polycarbonate
Assembly Fixed with clamping screw Wall mounting
Cable entry terminal range 5mm - 10mm
Electrical connection Terminals 2,5 mm
2
Supply voltage 12V DC internal from FA-1..
Electrical standard EN61010
Operating temperature - 25 °C to +180 °C -25 °C to +60 °C
Storage temperature - 25 °C to +70 °C
System of protection IP65 EN 60529
(* Other mounting designs for the light guides on request.)
Separate electronic controller required for the K-FA-H pre-amplifier
Electronic controller FA-1.1 FA-1.4
Part No. 230V 50/60Hz 02 E 7300* 02 E 7110
Part No. 115V 50/60Hz 02 E 7300* 02 E 7110 a
Part No. 24V DC 02 E 7300 d 02 E 7110 d
Part No. 24V AC 02 E 7300 b 02 E 7110 d
Reversing power consumption 230V 50/60Hz / 115V 50/60Hz, adjusted at works: 230V 50/60Hz Technical data: see data sheet FA-1.1.., FA-1.4, K-FA 5-6.10.2
Complete flow monitoring unit FA1-H-.. :
Complete flow monitoring unit Type
consisting of :
Sensor FA2-H (02 E 4002) 1 piece
Light guide FO1, 600 mm long (02 E 4060) 2 pieces
Light guide FO3, 900 mm long (02 E 4063) 2 pieces
Part Number
FA1-H-6 02 E 4270
FA1-H-9 02 E 4273
FA1-H-12 02 E 4275
Light guide FO2, 1200 mm lang (02 E 4065) 2 pieces
Pre-amplifier with adapter K-FA-H (02 E 4010) 1 piece
Electronic controller FA-1.1 / 230V/115V (02 E 7300) 1 piece
9.6
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
Technical specications and illustrations are without obligation, subject to modications. 12.97/11.06