ECP 1000- G
Embracing Challenge
Peltier-Gas Cooler Series ECP®
Version ECP1000 with 1x 150 Nl/h
Version ECP2000 with 2x 150 Nl/h
Version ECP3000 with 1x 350 Nl/h
Special Features
With Jet-Stream heat exchanger avail-
able in three standard materials
Ambient temperature up to +50 °C
Outlet dew point adjustable from +2 °C
to +15 °C
Dew point stability ± 0,1 °C
With Alarm contact
Small dimensions, light-weight
Optimum reliability
Self-controlling
Application
ECP gas coolers are used in gas analysis to lower the dew point of humid to
avoid condensate in the analyser. An extremely stabel gas dew point avoids water vapour
crosssensitivity and volumetric errors.
Description
The electronically controlled peltier cooler
produced by M&C incorporates the “JetStream” design of heat exchanger. This
design induces condensate formation and
guarantees optimum dew point reduction
to a stable low value, eliminating the need for
additional pre-drainage devices.
The condensate formed should be
removed with a small peristaltic pump,
simple trap or collection vessel. The simple construction employed enables heat
exchanges of different materials to be used
depending on the application.
The operating status of the cooler is displayed by three LED indicators, one each
for over or under temperature alarm and
one which pulsates showing the cooler
is functioning correctly. The alarm status
changes when the temperature deviates by
± 3 °C from the set point.
The compact lightweight design of this
device makes it ideal for use in portable
and compact continous sample conditioning
systems.
Application example for ECP
1 Heated lter sample probe SP210-H or SP2000-H
2 Heated sample line 4M4/6
3 Cooler ECP1000
4 3-way ball valve 3L/PV-1
5 Peristaltic pump SR25.2
6 Diaphragm pump MP47 or MP06/12 or N5KP
7 Fine lter FP-2T-D with liquid alarm LA1
8 Aerosol lter CLF-5 /W optional according to application
9 Flow meter FM10 or FM40, 25-250 Nl/hr
10 Analysers f. e. PMA100
Technical specications and illustrations are without
obligation, subject to modications. 10.96/06.06
10
3
ECP
sample gas IN
1 2
Alternative:
trap with float or condensate
collection vessel
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
+5°C
4
test gas
IN
condensate OUT
7
6
5
8
9
6.1

26
25
55
305
220
25
Cooler
ECP 2000
ON
<C
>C
2xPG 11
GL 25
GL 18
Dimensions
17
220
30
275
136
68
200
180
11
6
63 137
215
2xPG 11
GL 25
GL 18
<C
ON
>C
Cooler
ECP 1000
305
45
25
220
Cooler
ECP 3000
ON
<C
>C
2xPG 11
GL 25
GL 18
ECP 1000
ECP 2000
ECP 1000
°C
10
Sample inlet dew point 50°C
Sample outlet dew point
5
50
100 150 Nl/h
Gas flow rate
Gas cooler with Part-No.
A Heat exchanger out of glass 01 K 1000
B Heat exchanger out of SS316Ti 01 K 2000
C Heat exchanger out of PVDF 01 K 3000
ECP 2000
°C
10
A
C
B
ECP 3000
Sample inlet dew point 50 °C
A
C
B
Sample outlet dew point
5
50
100 150 Nl/h
Gas flow rate
Gas cooler with Part-No.
A Heat exchanger out of glass 01 K 2100
B Heat exchanger out of SS316Ti 01 K 2200
C Heat exchanger out of PVDF 01 K 2300
ECP 3000
°C
10
Sample inlet dew point 50°C
Sample outlet dew point
C
A
B
Dimensions in mm
6.1
5
50
100 150 200 250 300 350 Nl/h
Gas flow rate
Gas cooler with Part-No.
A Heat exchanger out of glass 01 K 3100
B Heat exchanger out of SS316Ti 01 K 3200
C Heat exchanger out of PVDF 01 K 3300
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
Technical specications and illustrations are without
obligation, subject to modications. 10.96/06.06

Technical Data
Gas cooler serie EC
®
ECP 1000 ECP 2000 ECP 3000
Part number with glass heat exchanger 01 K 1000 01 K 2100 01 K 3100
Part number with stainl. steel heat exchanger 01 K 2000 01 K 2200 01 K 3210
Part number with PVDF heat exchanger 01 K 3000 01 K 2300 01 K 3300
Number of heat exchanger 1 2 1
Gas flow rate per heat exchange max. 150 Nl/h** max. 350 Nl/h**
Ambient temperature +5 to +45 °C, +50 °C* +5 to +50 °C +5 to +50 °C
Storage temperature -20 to +60°C
Sample outlet dew point range of adjustment: +2 °C ..... +15 °C, factory setting: +5 °C
Dew point stability ± 0,1 °C at constant conditions
Sample inlet temperature max. 180 °C**
Sample inlet dew point max. 80 °C**
Total cooling power at +25 °C ambient 54 KJ/h 97,2 KJ/h 97,2 KJ/h
Power consumption 100 VA 110 VA 110 VA
Main connection 230 V ± 10%, 50Hz or 115 V ± 10%, 60Hz
Ready for working < 10 min.
Electrical connection Terminals 2,5 mm2, cable gland 2 x PG 11.
Status alarm: 2 changeover contacts Contact rating: 250V, 2A, 500VA, 50W, Alarm point: ∆ T ±3 °C zu T
soll
Case protection IP 22; EN 60529
Electrical standard EN 61010
Case colour RAL 9005
Method of mounting wallmounting
Dimension/Weight 5,5 kg 7 kg 7 kg
Heat exchanger
Material of heat exchanger Duranglass
Admissible gas pressure 3
®
2)
3)
/ 2
PVDF stainless steel 316Ti
3)
3 / 2
10 bar
Sample gas connection for tube ø6 mm, 8 o. 10 mm* G 1/4’’ i G 1/4’’ i, NPT*
Condensate connection for tube ø12 mm, 10 o. 8 mm* G 3/8’’ i G 3/8’’ i, NPT*
∆ P per heat exchanger at 1 mbar at 150 l/h 1 mbar at 150 l/h 5 mbar at 350 l/h
Stagnant space per heat exchanger 50 ml 50 ml 100 ml
* Option
** Maximum values in technical datas must be rated in consideration of total cooling capacity at 25 °C ambient temperature and an outlet dew point of 5 °C.
1) Others on request.
2) With GL-adapter.
3) Max. 2 bar abs. with peristatic pump SR25.1.
GL adapter and tube ttings for the connection of diererent tube diameters at the heat exchanger see chapter 11, data sheet 11.5 and 11.6.
Technical specications and illustrations are without
obligation, subject to modications. 10.96/06.06
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111