ECM-2 G/SR25.2
Embracing Challenge
Gas Cooler Series EC®
Compact Version ECM-1 and ECM-EX2-1
for 1x 250 Nl/h
Compact Version ECM-2 and ECM-EX2-2
for 2x 150 Nl/h
Special Features
Small dimension and light weight
Explosion proof version according to
ATEX, CSA, FM and NEPSI for zone 2
Both versions with approval according
to CSAUS
Gas ow 1x 250 or 2x 150 Nl/h
Jet Stream heat exchangers in 3 stand-
ard materials
Ambient temperature up to 50°C
Outlet dew point adjustable from +2 °C
to +7 °C
Dew point stability ± 0,1 °C
Digital temperature indication
Congurable status alarm contact
Compact wall mounting housing
Optimum reliability
Application
The M&C gas cooler ECM is used in gas
analysis to lower the dew point of humid
gas to avoid condensation in the analyser(s).
An extremely stable and low gas dew point
avoids water vapour cross-sensitivity and volumetric errors.
Description
Compact, maintenance-free and self-controlling. Intelligent detailed solutions provide
optimum cooling of the sample gas and
direct separation of condensate to avoid contamination effects.
The new controlled compressor cooling system and the special design of the jet stream
heat exchanger guarantee an optimum dew
point reduction to a low, stable value. An
additional pre-drainage device is not necessary in normal applications.
The condensate should be removed with
integrated small peristaltic pumps SR25.2 or
optionally by external condensate traps AD...
respectively collection vessels TG../TK..
The simple construction enables heat
exchangers of different materials to be used
depending on the application. Heat exchangers to be ordered optionally.
The digital indicator in the front panel displays the cooling temperature. The function
of the cooler can be externally controlled via
the alarm contact, configurated at <+2 °C and
>+8 °C as a standard.
The gas cooler ECM-1 (1-channel) can be
equipped with one Jet-Stream heat exchanger with a flow capacity of max. 250 Nl/hr.
The gas cooler ECM-2 (2-channel) can be
equipped with two Jet-Stream heat exchangers with a flow capacity of max. 150 Nl/hr
each.
The explosion-proof versions ECM-EX2-1 and
ECM-EX2-2 can be used in Ex-zone 2 and, they
too, may additionally be equipped with up to
2 standard peristaltic pumps SR25.2.
The compact and light weight construction
allows a simple and space-saving installation.
The ECM gas coolers are self controlling and
maintenance free in operation.
Technical specications and illustrations are without
obligation, subject to modications. 10.99/06.06
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
6.3

220
24030
50
84
11
6
46
204
270
208
270
Vent from the
compressor
ventilation
opening
Cooler
ECM
Application example for ECM-1
8
Alternative:
trap with float or condensate
collection vessel
5
condensate OUT
+5°C
4
3
10
9
7
ECM-1
1 2
sample gas IN
test gas
IN
6
1 Heated lter sample probe SP210-H or SP2000-H
2 Heated sample line 4M4/6
3 Cooler ECM-1G
4 3-way ball valve 3L/PV-1
5 Peristaltic pump SR25.2
6 Diaphragm pump MP47 or MP06/12 or N5KP
7 Fine lter FP-2T-D with liquid alarm LA1
8 Aerosol lter CLF-5 /W optional according to application
9 Flow meter FM10 or FM40, 25-250 Nl/hr
10 Analysers f. e. PMA100
Dimensions
Compact gas cooler ECM-1 / ECM-2 / ECM-EX2-1 / ECM-EX2-2
Dimensions in mm
Drawing shows ECM-2G with two heat exchanger out of glass.
Heat exchanger(s) and peristaltic pump(s) to be ordered optionally!
Sample gas outlet dew point stability
Functioning diagram of M&C
Jet-Stream heat exchanger
Sample gas - OUT
M&C Jet-Stream
heat exchanger
Cooling block
+5°C
Condensate - OUT
Sample gas - IN
10
9
8
ECM-2G
ECM-1G
7
6
5
4
Sample outlet dew point °C
3
0 50 100 150 200 250
Sample flow rate Nl/hr
Sample gas outlet dew point stability at gas inlet dew point of 60 °C.
Characteristics of heat exchanger out of PVDF or stainless steel on request.
6.3
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
Technical specications and illustrations are without
obligation, subject to modications. 10.99/06.06

Technical Data
Gas cooler series EC
Part No. for basis cooler without heat exchanger 02K7500X* (a)** 02K7510X* (a)** 02K7600X* (a)** 02K7610X* (a)**
Number of heat exchanger possibility- 1 2 1 2
Ambient temperature +10 up to +50 °C
Storage temperature -20 to +60°C
Sample outlet dew point range of adjustment: +2 °C ..... +7 °C, factory setting: +5 °C
Dew point stability at constant conditions < ±0,1 °C
Sample inlet temperature*** max. 180 °C
Sample inlet dew point*** max. 80 °C
Total cooling capacity 144 KJ/h at +10 up +50 °C ambient
Ready for working <15 min.
Main power connection / power consumption 230V AC* or 115V AC**(a) -15%/+10%, 50/60 Hz, max. 200VA
Electrical connection terminals 2,5mm2, 2x PG11 cable gland (with FM-approval conduit-hub 1/2” NPT)
Status alarm-2 contacts, potential free contact rating 250VAC, 2A, 500VA, 50W, alarm point <+2 °C and >+8 °C*
Case protection / Electrical standard IP20 EN60529 / EN 61010
Method of mounting / Case colour wall mounting / RAL 9003
Dimension / Weight 270x270x316 mm (w x h x d) / 12/13,5 kg Version 230V / 115V
®
version ECM-1 version ECM-2 version ECM-EX2-1 version ECM-EX2-2
start up current: 230V 50Hz = 2,5A / 115V 60Hz = 4,5A
II 3 G EEx nAC [L] IIC T4 ATEX
Class I, Div.2, Group A, B, C, D, T4 (FM
approval for 115V execution only)
Options
Options for basis cooler ECM-1.. and ECM-EX2-1.. ECM-2... and ECM-EX2-2
Heat exchanger type ECM-1G ECM-1PV ECM-1SS ECM-2G ECM-2PV ECM-2SS
Part No. 93K0140 93K0170 93K0160 97K0100 97K0110 97K0115
Material of heat exchanger Duran glass PVDF st. steel 316Ti Duran glass PVDF st. steel 316Ti
Max. gas flow rate per h.e. 250 Nl/hr*** 250 Nl/hr*** 250 Nl/hr*** 150 Nl/hr*** 150 Nl/hr*** 150 Nl/hr***
Gas press. max. bar abs.
Sample gas connection GL18 for
Condensate connection GL⁄25 for
∆P at max. flow rate 1 mbar 1 mbar 1 mbar 1 mbar 1 mbar 1 mbar
Stagnant space approximately 100 ml 100 ml 100 ml 40 ml 25 ml 30 ml
Peristaltic pump SR25.2 1 pc. incorporated in the cooler, compl. installed, Part No.: 01P9125 cooler weight plus 0,6 kg per pump
* Standard, other version on request.
** Option
*** Maximum values in technical data’s must be rated in consideration of total cooling capacity at 25 °C ambient temperature and 5 °C outlet dew point.
2)
With GL-Connecting adapter.
3)
With SR25.2 max. 2 bar abs.
Order example:
1 cooler ECM-2 with 2 heat exchangers out of glass ECM-2G and 2 peristaltic pumps SR25.2, power 115V 60Hz:
Part. No. 1x 02K7510Xa; 2x 97K0100; 2x 01P9125
GL-adapter and tube ttings for the connection of dierent tube diameters at the heat exchanger see chapter 11, data sheet 11.5 and 11.6.
3)
2)
2 / 3
ø 6 mm
o.d. tube*
ø 12mm tube*
ø 8mm
or ø 10mm**
3 10* 2 / 3
G1/4’’i G1/4’’i* or
1/4’’NPT**
GL18 for
ø 6 mm
o.d. tube*
G3/8’’i G3/8’’i or
3/8 NPT**
GL25 for
ø 12 mm tube*
ø 8 mm
or ø 10 mm**
2)
3 10*
Tube ø 6 mm* Tube ø 6 mm
G3/8’’i G3/8’’i or
3/8 NPT**
Technical specications and illustrations are without
obligation, subject to modications. 10.99/06.06
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
6.3