M&C TechGroup ECM Operator's manual

Operating Manual
Electric gas cooler series EC
®
Version ECM
Gas sampling and gas conditioning technology 3-3.0-ME
2
This Operating Manual does not claim completeness and may be
subject to technical modifications. © 03/2002 M&C TechGroup Germany GmbH. Reproduction of this document or its content is not allowed without permission from M&C.
ECM® is a registered trade mark.
3rd Edition: 01/2007
Dear customer,
we have made up this operating manual in such a way that all necessary information about the
product can be found and understood quickly and easily.
Should you still have any question, please do not hesitate to contact M&C directly or go through your
appointed dealer. Respective contact addresses are to be found in the annexe to this operating
manual.
Please also contact our homepage www.mc-techgroup.com for further information about our
products. There, you can read or download the data sheets and operating manuals of all M&C
products as well as further information in German, English and French.
Gas sampling and gas conditioning technology 3-3.0-ME
3
Content
1 General information ................................................................................................................... 4
2 Declaration of conformity ......................................................................................................... 4
3 Safety instructions ..................................................................................................................... 5
4 Warranty ..................................................................................................................................... 5
5 Used terms and signal indications ........................................................................................... 6
6 Introduction ................................................................................................................................ 7
6.1 Serial number ........................................................................................................................... 7
7 Application ................................................................................................................................. 7
8 Technical Data ............................................................................................................................ 8
8.1 For basic cooler ECM without heat exchanger ......................................................................... 8
8.2 Options for basic cooler ECM ................................................................................................... 8
9 Description ................................................................................................................................. 9
10 Function ...................................................................................................................................... 9
11 Reception and storage ............................................................................................................ 10
12 Installation instructions .......................................................................................................... 11
13 Supply connections ................................................................................................................. 11
13.1 Hose connections ................................................................................................................... 11
13.2 Electrical connections ............................................................................................................. 13
14 Start-up ..................................................................................................................................... 14
15 Closing down ........................................................................................................................... 14
16 Maintenance ............................................................................................................................. 15
16.1 Adding and replacing the heat exchangers ............................................................................ 15
16.2 Cleaning the fins of the condenser ......................................................................................... 16
16.3 Maintenance of the optional mounted peristaltic pump(s), type SR25.2 ................................ 16
16.3.1 Change of the pump tube ................................................................................................ 16
16.3.2 Change of contact pulleys and springs ............................................................................ 17
16.3.3 Cleaning the pump head ................................................................................................. 18
17 Operating of the built-in electronic temperature controller ................................................. 18
17.1 Changing the set value ........................................................................................................... 18
17.2 Changing the temperature alarm window ............................................................................... 18
18 Trouble shooting ...................................................................................................................... 19
19 Spare parts list ......................................................................................................................... 20
20 Appendix ................................................................................................................................... 21
List of illustrations
Figure 1 Application example of the ECM ....................................................................................... 7
Figure 2 ECM-2G ............................................................................................................................. 9
Figure 3 Schematic diagram of the heat exchanger function ........................................................ 10
Figure 4 Position of the gas and condensate ................................................................................ 11
Figure 5 Electrical sockets alarm contact ...................................................................................... 13
Figure 6 Change of the pump tube ................................................................................................ 17
Figure 7 Front view of the temperature controller .......................................................................... 18
Figure 8 Sample outlet dew point depending on gas flow rate and sample inlet dew points
of 60°C ............................................................................................................................. 21
Figure 9 Circuit diagram ECM ....................................................................................................... 22
Gas sampling and gas conditioning technology 3-3.0-ME
4
Head Office
M&C TechGroup Germany GmbH Rehhecke 79 40885 Ratingen Germany
Telephone: 02102 / 935 - 0
Fax: 02102 / 935 - 111
E - mail: info@mc-techgroup.com
www.mc-techgroup.com
1 GENERAL INFORMATION
The product described in this operating manual has been examined before delivery and left our works
in perfect condition related to safety regulations. In order to keep this condition and to guarantee a
safe operation, it is important to heed the notes and prescriptions made in this operating manual.
Furthermore, attention must be paid to appropriate transportation, correct storage, as well as
professional installation and maintenance work.
All necessary information a skilled staff will need for appropriate use of this product are given in this
operating manual.
2 DECLARATION OF CONFORMITY
CE - Certification
The product described in this operating manual complies with the following EC directives:
EMV-Instruction
The requirements of the EC directive 2004/108/EC “Electromagnetic compatibility“ are met.
Low Voltage Directive
The requirement of the EC directive 2006/95/EC “Low Voltage Directive“ are met.
The compliance with this EC directive has been examined according to DIN EN 61010.
Declaration of conformity
The EU Declaration of conformity can be downloaded from the M&C homepage or directly requested
from M&C.
CSA - Certification
The product described in this operating manual complies with the following CSA directive:
CAN/CSA-C22.2 No.61010.1-04 UL Std.61010-1 (2 Edition).
Gas sampling and gas conditioning technology 3-3.0-ME
5
3 SAFETY INSTRUCTIONS
Please take care of the following basic safety procedures when mounting, starting up or operating this equipment:
Read this operating manual before starting up and use of the equipment. The information and warnings given in this operating manual must be heeded.
Any work on electrical equipment is only to be carried out by trained specialists as per the regulations currently in force.
Attention must be paid to the requirements of VDE 0100 (IEC 364) when setting high-power electrical units with nominal voltages of up to 1000 V, together with the associated standards and stipulations.
Check the details on the type plate to ensure that the equipment is connected to the correct mains voltage.
Protection against touching dangerously high electrical voltages: Before opening the equipment, it must be switched off and hold no voltages. This also applies to any external control circuits that are connected.
The device is only to be used within the permitted range of temperatures and pressures. Check that the location is weather-protected. It should not be subject to either direct rain or moisture. The gas cooler ECM must not be used in hazardous areas. Installation, maintenance, monitoring and any repairs may only be done by authorized personnel with
respect to the relevant stipulations.
4 WARRANTY
If the equipment fails, please contact M&C directly or else go through your M&C authorised dealer. We offer a one year warranty as of the day of delivery as per our normal terms and conditions of sale, and assuming technically correct operation of the unit. Consumables are hereby excluded. The terms of the warranty cover repair at the factory at no cost or the replacement at no cost of the equipment free ex user location. Reshipments must be send in a sufficient and proper protective packaging.
Gas sampling and gas conditioning technology 3-3.0-ME
6
DANGER!
This means that death, severe physical injuries and/or important material damages will occur in case the respective safety measures are not fulfilled.
W ARN IN G !
This means that death, severe physical injuries and/or important material damages may occur in case the respective safety measures are not fulfilled.
CARE!
This means that minor physical injuries may occur in case the respective safety measures are not fulfilled.
CAR E!
Without the warning triangle means that a material damage may occur in case the respective safety measures are not met.
ATT ENT ION !
This means that an unintentional situation or an unintentional status may occur in case the respective note is not respected.
NOTE!
These are important information about the product or parts of the operating manual which require user’s attention.
SKILLED STAFF
These are persons with necessary qualification who are familiar with installation, use and maintenance of the product.
5 USED TERMS AND SIGNAL INDICATIONS
Gas sampling and gas conditioning technology 3-3.0-ME
7
 
Sample gas inlet
+5°C
ECM-1
Alternative : Collecting vessel
Test gas IN
6 INTRODUCTION
The patented M&C ECM gas cooler unit is always to be installed in situations where there is interference from moisture in the gas to be measured. Reduction of the gas temperature inside the cooler to a stable and very low dew point effects a condensing-out of the sample gas.
6.1 SERIAL NUMBER
The type plate with the serial number is located at the side panel of the cooler housing (wall mounting version). Whenever you call M&C regarding questions or orders for the spares please give us the serial number of your ECM.
7 APPLICATION
Figure 1 shows a typical example of an application for installation of an ECM-ExII gas cooler unit.
Figure 1 Application example of the ECM
1 : Filter sample probe SP ... 2 : Heating sample line 3 : ECM cooler 4 : 3-way ball valve 5 : Peristaltic pump (option) 6 : Membrane pump 7 : Super fine filter FP ... 8 : Aerosol filter CLF-5 9 : Flow meter FM10 10 : Analyser
The gas to be measured is taken from the ECM gas cooler by a gas sample probe via a heated sample line and cooled down to a dew point of +5°C. The super-fine filter located afterwards removes solid particles. For increased operating safety of the entire system we recommend installing a super-fine filter with a liquid alarm sensor. If required an aerosol filter can be installed in front of the flow meter . The gas thus treated can now be passed into the analyser .
Gas sampling and gas conditioning technology 3-3.0-ME
8
Gas cooler series EC
version ECM-1
version ECM-2
Part No.
02K7500 (a)**
02K7510 (a)**
Number of heat exchanger possibility
prepared for 1 stream
prepared for 2 streams Sample outlet dewpoint
range of adjustment: +2 °C ..... +7 °C, factory setting: +5 °C
Dew point stability
at const. conditions: < ±0,1°C
Sample inlet temperature
***max. +180°C
Sample inlet dew point
***max. +80°C
Gas flow rate
***max. 250l/h
***max. 2x150l/h
Ambient temperature
+10°C to +50°C
Storage temperature
-20 to +60°C
Total cooling capacity
40W at 10°C - 50°C ambient temperature
Ready for working
< 15 min
Power consumption
max. 200VA (start up current at 230V=2,5A; at 120V=4,5A)
Main power connection
240V/120V, 50Hz/60Hz (a):-15%/+10%
Electrical connections
terminals: 6-12mm , cable gland: 2xM20x1,5
Status alarm
1 free configurable status alarm with 2 potential free change-over contacts, alarm point: < 2°C and > 8°C contact rating 250VAC, 2A; 500 VA; 50W
Electrical safety
EN
EN 61010
CAN/CSA-C22.2 No.61010.1-04 UL Std. No. 61010-1(2 Edition)
System of protection
IP20 (EN 60529)
Case colour
RAL 9003
Method of mounting
wall mounting
Dimensions Weight
270 x 270 x 316mm (WxHxD) 12kg/13,5kg (version: 230V/115V)
Heat exchanger type
ECM-1G
ECM-1PV
ECM-1SS
ECM-2G
ECM-2PV
ECM-2SS
Part No
93K0140
93K0170
93K0160
97K0100
97K0110
97K0115
Material of heat exchanger
Borosilicate glass
PVDF
st. steel 316Ti
Borosilicate glass
PVDF
st. steel 316Ti
Max. gas flow rate
250Nl/h
250Nl/h
250Nl/h
150Nl/h
150Nl/h
150Nl/h
Sample gas pressure abs.
max. 2bar (3bar)
1
max. 3bar
max. 10bar
max. 2bar (3bar)
1
max. 3bar
max. 10bar
Sample gas connection
GL18 for Ø6mm a.d. tube*
G1/4“i
G1/4“i
oder 1/4“
NPT**
GL18 for Ø6mm a.d. tube*
Tube Ø6mm*
Tube Ø6mm*
Condensate connection
GL25 for Ø12mm tube* Ø8mm or Ø10mm**
G3/8“i
G3/8“i o. 3/8“NPT**
GL25 for Ø12mm tube* Ø8mm or Ø10mm**
G3/8“i
G3/8“i o. 3/8“NPT**
P at max. flow
1mbar
1mbar
1mbar
1mbar
1mbar
1mbar
Stagnant space approx.
100ml
100ml
100ml
40ml
25ml
30ml
Peristaltic pump SR25.2
1 pc. incorporated in the cooler, compl. installed, part no.: 01P9125; cooler weight plus 0,6kg per pump
the sample gas pressure is limited to 2bar absolute
8 TECHNICAL DATA
8.1 FOR BASIC COOLER ECM WITHOUT HEAT EXCHANGER
8.2 OPTIONS FOR BASIC COOLER ECM
* Standard, other version on request
** Option
*** Maximum values in technical data must be rated in consideration of total cooling capacity at 25 °C ambient temperature and an outlet dew
point of 5 °C.
1
with GL connecting adapter
Gas sampling and gas conditioning technology 3-3.0-ME