
DIL-1/H
Embracing Challenge
Gas-Dilution Unit DIL-1/(H)
Special Features
Proven M&C-dilution technique
Operation at ambient temperature
Optionally heated to 180 or 320 °C
Integrated dilution preheater
No dew point problems
Dilution factor from 10:1 to 500:1
With test gas connection
Independent of the ambient
temperature
Easy maintenance and operation
Application
The non-heated or electrically heated M&C
dilution unit DIL-1 /H is used in analysis
technique to dilute the sample gas or components in the sample for example in case of
toxic gases, moisture measurement or emission measurement.
The M&C dilution unit is based on the functional dilution technique which is proved
since years in the M&C gas sample probe
SP2000-H/DIL.
Description
The M&C dilution units DIL-... are mounted on
a plate for wall mounting. The version DIL-1 is
not heated and works at ambient temperature. The version DIL-1/H is electrically heated
up to 180 °C . It is additionally equipped
with a lagged cover (version for 320 °C on
request).
A capillary thermostat controlls the temperature, adjustable from 0-180 °C.
A high temperature limiter and a low temperature alarm is included.
The connection of the heated lines is
made without cold bridges in the heated
part. Before the dilution gas enters the
dilution unit, it is heated up to operating temperature via a gas pre-heater. In
order to protect the dilution part against
contaminations, internal protective filters are installed for the sample gas and
the dilution gas stream. Calibration gas can be
feeded via the integrated connection.
A precision pressure controller with pressure
gauge is used for adjustment of the necessary
admission pressure of the dilution gas. Via a
vacuum pressure gauge, the function of the
dilution injector is controlled.
Both, pressure controller and pressure gauges
have to be ordered separately. Two versions
are available: Set -A (-A1) is mounted directly
on the mounting plate and control panel type
-S (-S1) is suitable for external 19» rack mounting. A shut off valve and a flowmeter for
adjustment of the calibration gas are included
in the version -S (-S1).
The dilution unit can realize dilution ratios
of 10:1 to 500:1. In case of high dilution
ratios, a respective small quantity of gas
is sampled from the process. Optionally
a bypass injector -B (Option I) is available
integrated right in front of the dilution unit
in order to shorten the response time in case
of operation with atmospheric pressure or
low pressure. In case of option I, an additional
pressure controller is included in both mounting versions.
The construction of the dilution unit guarantees easy maintenance and operation independent from process temperatures.
Technical specications and illustrations are without
obligation, subject to modications. 08.99/06.06
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
5.4

105
Option II
Dilution
gas in
#
Option I
500
370
450
250
482 (84TE)
132 (3HE)
450
300
300
205
PG 13,5
Option
-0,4
bar
0,0
-0,2
-0,6
-1,0
-0,8
Reset
Tsoll
To>+20 C
Tu<-20 C
0-180 C
L '
L
N
6
34
5
87
12
FI
6ø
Testgas
Open
PG 13,5
Reset
8
L
NL '
412 3
5
6
7
5,0
bar
1,0
0,0
2,0
3,0
4,0
3,0
bar
6,0
5,0
2,0
0,0
1,0
6,0
4,0
Tu<-20 C
0-180 C
Tsoll
To>+20 C
ø 6
ø 6
Dilution gasBypass injector
Open
0,0
-0,2
-0,6
bar
-1,0
-0,8
-0,4
ø 6
5,0
4,0
6,0
1,0
2,0
bar
0,0
3,0
4,0
3,0
6,0
bar
1,0
0,0
2,0
Option I
5,0
Dilution gas
in
#
Injector-vacuum
Option I
Bypass out
(diluted)
#
Option II
Sample gas
out
(non-diluted)
#
Sample gas in
(undiluted)
#
Testgas
in
#
Option I
Bypass out
(diluted)
#
Option II
Sample gas
out
(non-diluted)
#
Option I
Bypass out
(diluted)
#
Option II
Sample gas
out
(non-diluted)
#
Option I
Bypass out
(diluted)
#
Option II
Sample gas
out
(non-diluted)
#
PI
0-6 bar
PI
-1-0 bar
PI
-0,4
-0,2
0,0
-0,6
-1,0
-0,8
180 C
Option I: Bypass
5,0
3,0
2,0
0,0
6,0
4,0
4,0
3,0
6,0
1,0
0,0
2,0
5,0
1,0
Power
230V,50Hz
(115V,60Hz)
Alarm-Contact
Temp.-min.
Dilution gas
Bypass
0-6 bar
Sample gas out
(diluted)
Bypass
out
Sample gas out
(undiluted)
Sample
gas in
Testgas
Electrically heated Gas-Dilution Unit DIL-1/H with control panel -S (S.1)
Electrically heated Gas-Dilution Unit DIL-1/H with mounting set -A (-A1)
Power
230V,50Hz
(115V,60Hz)
AlarmContact
Temp.
-min.
Control panel
Bypass gas
in
#
Testgas
in
#
Sample
gas out
(diluted)
##
Under-pressure
manometer
#
Sample gas in
(undiluted)
#
DIL-1 with control panel -S (S.1)
Gas-Dilution Unit DIL-1 with mounting set -A (-A1)
## Gas conections for tube 8x 1 mm
# Gas conections for tube 6x 1 mm
Dilution
gas in
#
Bypass
gas in
#
Under-pressure
manometer
#
Sample
gas out
(diluted)
##
Dilution
gas in
#
Testgas
in
#
Sample
gas out
(diluted)
##
Sample gas in
(undiluted)
#
Testgas
in
#
Sample
gas out
(diluted)
##
Sample gas in
(undiluted)
#
Dimensions
Dimensions in mm
Connections in inch on request
5.4
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
Technical specications and illustrations are without
obligation, subject to modications. 08.99/06.06

Technical Data
M&C-Dilution unit DIL-1/ (H)
Dilution rates with the critical orifices ”a“ - ”g“ a = 500 b = 200 c = 100 d = 50 e = 30* f = 20 g = 10 : 1
Sample flow rate depending on the critical orifices
”a“- ”g“
Possibility to adapt the dilution factor with dilution gas pressure-adjustment -5% bis +30%
Dilution gas flow rate with injector version I or II I: 480 - 600 Nl/hr, optional II: 1800 – 3000 Nl/hr
Dilution gas pressure on inlet of pressure controller min. 4,5 bar g, max. 16 bar g
Bypass injector / B sample gas flow rate
gas pressure / gas flow rate
Process pressure 0,9 to 2 bar abs.
Fault caused by process temperature variations operation independent of process temperature
Fault caused by process low- or overpressure no fault as long as the differential pressure ∆P at the dilution unit is > 0,5 bar g and test gas is given to
Fault caused by atmospheric pressure variations < 1% with a variation of 50 mbar
Materials in contact with the sample gas stainless steel 316 / 316Ti, quartz glass, FPM, graphite
Power supply for DIL-1/H 230V 50Hz, 800W
Temperature controller for DIL-1/H capillary thermostat adjustable 0-180 °C , with high temperature limiter
Weight approx. 8 kg
* Standard, others to be indicated along with order, intermediate values possible.
1) approx. at 3 bar dilution gas.
2) -5% not possible for orice ”g“.
a = 1,4 b = 2,7 c = 5,5 d = 11 e = 19* f = 28 g = 55 l/hr
2)
approx. 2 bar
injector gas approx. 300 l/hr sample gas approx. 150 l/hr
the probe under process conditions
and low temperature alarm as changeover contact, voltage free
alarm point ∆T30 °C to TSet, contact rating max. 250V 3A~ 0,25A =
1)
Part-No. Type M&C-Dilution Unit DIL-1/ (H) with orifice ”e“ for dilution ratio 30 : 1 standard
20 S 4900 DIL-1 Dilution unit non heated
20 S 4905 DIL-1/H (a) Dilution unit electrically heated to 180 °C , (a) to be added to the Part-No. for power 115V 60Hz
20 S 4925 Option I Bypass injector - B
20 S 4930 Option II Additional sample gas outlet undiluted
20 S 4205 DIL-1/-A option: 1 pressure controller, 2 gauges, mounting set for dilution unit
20 S 4215 DIL-1/-A1 option: 2 pressure controllers, 3 gauges, mounting set incl. connector set for dilution unit with by-pass
20 S 4250 DIL-1/-S option: mounting set with 1 pressure controller, 2 gauges, flowmeter, 2 ball valves
20 S 4260 DIL-1/-S1 option: mounting set with 2 pressure controllers, 3 gauges, flowmeter, 2 ball valve
Other executions on request.
Technical specications and illustrations are without
obligation, subject to modications. 08.99/06.06
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
5.4