![](/html/e1/e1ec/e1ec0a742a2056a532b81455c65999455db02c311c5e6034541264e7e71481b8/bg1.png)
Embracing Challenge
Air clean up supply series BA
Version BA, BA-P4
BA
Special Features
High degree of purity:
Cn Hm <10ppb,
H2O dew point <-10 °C
Easy lter and adsorber change at the
front without tools
Long service life
Minimum operating costs
Status self-monitoring
Compact, easy-to-use and easy-to-serv-
ice 19 inch plug-in housing
Application
When measuring hydrocarbons with a FID,
hydrocarbon free burner air is necessary. The
M&C-burner air BA conditioning has been
developed in order to achieve proper burner
air supply, independent of gas cylinders.
The BA unit can also be used as a zero gas
generator when calibrating I.R.-analysers for
immission monitoring.
Description
The M&C-burner air conditioning versions
BA and BA-P4 have been designed as compact, easy-to-use and easy-to-service 19 inch
modules.
The M&C burner air conditioning version BA is
designed without pump.
With version BA-P4 ambient air is taken in by
means of the integrated pendulum reciprocating pump via the rear-mounted air suction
filter and is directed to particle and condensate separation by means of a high-performance filter (0.01 micron of filter fineness) with
an integrated, floating condensate drainer.
Drying the condensate-free compressed
air to <-10 °C dew point takes place in a
membrane dryer.
There upon the compressed air is cleaned
as a result of catalytic oxidation at 500 °C by
means of a platinum/palladium catalyst of
hydrocarbon traces.
A downstream adsorber cartridge with
molecular sieve and activated carbon guarantees optimum purity of burner air.
The catalyst and adsorber material can, if
required, be replaced at the front without
tools.
Technical specications and illustrations are without
obligation, subject to modications. 09.96/01.07
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
12.2
![](/html/e1/e1ec/e1ec0a742a2056a532b81455c65999455db02c311c5e6034541264e7e71481b8/bg2.png)
Dimensions
Fl4P2 YD1
Adsorber
Regulator
177 (4K)
Rearside
102
Outlet
Inlet
Burner air supply
BA
Catalyst
!
Hot
465
435
Dimensions in mm
Functional diagram BA-P4
483 (84TE)
B1S2S1Fl3P1
23 1Fl14YA1X2X1
YD2
472
360
44
2 Vent
3 Condensate outlet
4 Air inlet
for BA compressed air IN
B1 Temperature controller
FI1 Air inlet filter
FI2 Membrane dryer
FI3 Catalyst
FI4 Adsorber
M1 Compressor
1 Burner air outlet
10
7
for BA solenoid valve
P1 Inlet pressure gauge
P2 Outlet pressure gauge
S1 Pump ON switch
for BA solenoid valve ON
S2 Power ON switch
S3 Pressure switch
X1 Power supply
X2 Status contact
Y1 Solenoid valve
YA1 Condensate drainer
YD1 Pressure regulator
YD2 Check valve
YN1 Needle valve
P1
M1
4
Y1
YA1FI1
YN1
FI3
FI2
S3
YD1
P2
FI4
1
2
3
Technical Data
Burner air conditioning series BA version BA version BA-P4
Part No. 60A2000 60A2200
Inlet pressure compressed air, 4 - 5,8 bar internal pump
Outlet pressure / Burner air flow 0-4 bar adjustable / max. 3 Nl/min
Ambient temperature +5 °C to +40 °C
Start up time approx. 45 min.
Catalyst temperature 500 °C
Burner air purity <10 ppb CnHm, H2O-dew point <-10 °C
Catalyst poisons Halogens, silicon, lead, materials containing phosphorous
Storage temperature -25 °C to +65 °C
Relative humidity <75 %, avoid condensation
Inlet gas connection compressed air in G1/4"i air suction filter
Outlet gas connections Hose connection DN 4/6 mm* Standard
Condensate connection G1/8 i DIN ISO 228/1
Power supply 230V 50Hz 480VA 230V 50Hz 560VA
Electrical connections Plug with 2 m cable for power in
9-pin sub-D connector for status contact
Status signal contact for pressure and temperature 1 changeover contact - potential free, contact rating 24V, 1A
Case protection IP 20 EN60529
Housing dimensions / protection 19" plug-in unit 4U, 84HP, depth 430 mm / IP 20 EN60529
Weight approx. 14,5 kg approx. 16 kg
Electrical equipment standard EN 61010
12.2
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
Technical specications and illustrations are without
obligation, subject to modications. 09.96/01.07