McQuay IM-701 Installation Manual

Page 1
Installation and Maintenance IM-701
DesignFlow
TM
Precision Outdoor Air Measurement System
Group:
Part Number:
Date:
Applied Systems
April 2001
© 2001 McQuay International
Page 2
General
DesignFlow™ airflow measurement stations are located inside the louvered outdoor air intake doors between the intake louver and outside air dampers. Essentially, they con­sist of a vane that is repositioned by airflow, the amount of rotation indicating the amount of airflow. They have been precision calibrated at the factory and no further calibration is required. However, a leveling adjustment is required. The DesignFlow unit must be accurately leveled so that when operating it will be orientated the same as when it was cali­brated.
The rotational position of the DesignFlow unit vane is trans­lated into CFM by the microprocessor in the MicroTech II control system. The position of the vane is determined by two things - the force of the airflow impacting the vane and the gravitational effect on the vane. Gravity is the only factor at the lower CFM end of the range. On a correctly leveled unit, this gravitational effect will be the same as when the unit was calibrated in the factory.
Both the right-hand and left-hand DesignFlow stations must be accurately leveled. To accurately level a station, a precise mechanical force is applied against the vane. That force should cause the vane to move to a specific position if the DesignFlow unit is correctly leveled. If the vane does not move to the correct position with the force applied, the DesignFlow unit is not properly leveled and gravity is caus­ing the difference.
The DesignFlow unit is mounted such that it pivots at the top. A lock nut, a slotted hole, and a threaded adjuster allow posi­tioning the bottom of the assembly so it can be correctly lev­eled. When the lock nut is loosened, the threaded adjuster allows precise, controlled movements.
3. Remove the covers from the access opening in the bottom blade of the outdoor air intake louvers (See Figure 1). The leveling procedure must be accomplished with the louvered door in the closed position. These openings provide access for leveling.
Figure 1
ACCESS OPENING
COVER
4. The leveling procedure is performed with the fans off and the outdoor air dampers closed. If there is a wind, the out­door air louvers should be covered with poly film, cardboard, etc. to prevent adverse readings due to wind.
5. Tape the fulcrum Alignment Plate to the vane (See Figure
2). Locate the Alignment Plate so the bottom edge of its notches are flush with the bottom edge of the vane, one side of one notch is even with the bend near the outer edge of the vane, and the plate is flat against the outer surface of the vane.
Figure 2
DesignFlow Airflow Measurement Station Startup:
Note: Before starting the station startup procedure, locate the leveling components kit that can be found in the main unit con­trol box.
Note: Before starting the station startup procedures, varify that power has been supplied to the unit’s MicroTech II con­trol system. The DesignFlow startup procedure cannot be completed without use of the MicroTech II controls.
VANE
DUCT TAPE THIS EDGE FLUSH WITH BEND ON VANE
THESE EDGES FLUSH WITH BOTTOM OF VANE
ALIGNMENT PLATE
FULCRUM DUCT TAPE
1. The swinging vanes on the measurement stations have been locked in place for shipment. The stations are accessi­ble when the louvered outdoor air intake doors are opened. Remove the two shipping screws from each vane. One screw is located one inch from the top of the vane. The other is one inch from the bottom of the vane. Both are about eight inches in from the outer edge of the vane.
2. Examine the station for shipping damage. Manually rotate the vane and verify that it does not rub against anything. Close and latch the louvered intake door.
2 IM-701
6. Install the fulcrum that is used in the leveling procedure (See Figure 2). The fulcrum will be held in place with duct tape. Wipe the bottom of the louver door where the fulcrum will be located so the tape will stick. Pre-apply the duct tape to the top surface of the bottom flat of the fulcrum, having it extend about one inch beyond the edges on three sides. Once the fulcrum is positioned, these extensions can be pressed
Page 3
down to tape the fulcrum in place. Use the Alignment Plate to locate the fulcrum. With the Alignment Plate taped to the vane and the vane in the zero airflow position, the fulcrum is located parallel to and against the Alignment Plate. The zero airflow position is when the vane is swung away from the back wall and resting against its stop. Remove the Align­ment Plate after locating the fulcrum.
7. Place the leveling weight on the fulcrum (See Figure 3). The vane has a small marker hole one inch up from the bot­tom and .50 inch in from the bend near the outer edge. There is a vertical mark intersecting the hole. The leveling weight should contact the vane along this vertical mark.
Figure 3
WIDE OPENING
VERTICAL MARK
FORCE AGAINST VANE
ZERO CFM POSITION
FULCRUM
CONTACT POINT
VANE
VANE
LEVELING WEIGHT
WEIGHT
PIVOTS ON TOP EDGE OF FULCRUM
TAPED IN PLACE
LEVELING WEIGHT
Figure 4
THREADED ADJUSTER ASSEMBLY
LONG ADJUSTER NUT
JAM NUTS
TO INCREASE LEVEL POSITION READING
RIGHT HAND ADJUSTER
Figure 4b
Figure 4a
LARGE
BOLT
LOCKNUT
TO REDUCE LEVEL POSITION READING
LEFT HAND ADJUSTER
Figure 4c
10. If the value indicated by the LH Lvl Pos=(or RH Lvl Pos=) parameter is not within the range of 23.64 to 24.08%,
a leveling adjustment is required (See Figure 4). Leveling adjustments are accomplished by shifting the bottom of the measuring station to the right or left. If the LH Lvl Pos=(or RH Lvl Pos=) parameter indication is higher than specified, the bottom of the DesignFlow frame must be moved closer to the outdoor air damper. If the LH Lvl Pos=(or RH Lvl Pos=) parameter indication is lower than specified, the bottom of the DesignFlow frame must be moved farther from the out­door air damper.
FULCRUM
NOTE: The leveling weight must be perpendicular to the vane.
8. Set up the leveling test. While holding the weight so it stays on the fulcrum, manually rotate the vane to the wide­open position, manually return it to the zero CFM position, and gently release the vane. Locate the Leveling Weight so its contact point is against the vertical mark on the vane. While the weight teeters on the fulcrum, gently rap the base frame to slightly vibrate the assembly and encourage the vane to seek its equilibrium point.
9. Read the current vane leveling position through the unit keypad/display. This is done by viewing the LH Lvl Pos= or
RH Lvl Pos= parameter in the DesignFlow Setup menu. The LH Lvl Pos= parameter indicates the current position of the
vane for the left-hand DesignFlow station. The RH Lvl Pos= parameter indicates the current position of the vane for the right-hand DesignFlow station. These parameters will vary from 20% to 80% depending on the position of the Design­Flow vane. For detailed information regarding operation and navigation through the unit keypad menu structure, refer to operation manual OM137 (discharge air control units) or OM138 (zone control units).
To make an adjustment, loosen the .25-20 NC lock nut at the slotted hole at the bottom of the frame. The threaded adjuster facilitates small accurate movements. Reposition the two .25-20 NC jam nuts on the threaded rod to make large adjust­ments, then use the long adjuster nut to make fine adjust­ments. Rotate the long adjuster nut to move the bottom of the station. Unscrewing the long adjuster nut from the large bolt makes the assembly longer and reduces the LH Lvl Pos=(or RH Lvl Pos=) parameter indication. Screwing in the long adjuster nut from the large bolt makes the assembly shorter and increases the LH Lvl Pos=(or RH Lvl Pos=) parameter indication.
After an adjustment is made, retighten the lock nut, gently rap the base frame to slightly vibrate the assembly and encourage the vane to seek its equilibrium point. Confirm that the contact point on the fulcrum is on the vertical mark on the damper. If the contact point is not on the vertical mark, reposition the fulcrum contact point to line up with vertical mark (See Figure 3) and take another reading.
Important: After an adjustment is made, wait until the LH Lvl Pos=(or RH Lvl Pos=) value on the keypad/display stabilizes before making further adjustments. This can take several seconds.
11. After leveling, remove fulcrum and replace the access opening covers.
IM-701 3
Page 4
13600 Industrial Park Boulevard, Minneapolis, MN 55441 USA (612) 553-5330
Loading...