McDantim Beer Check Quick Start Manual

Page 1
Connect the adaptor and tube to the
Beer Check by twisting the Luer Lock fittings clockwise until snug. (The tubing should be pushed on the adaptor.)
Connect to the gas system: Read
Caution on the Back Page! The threads on
the adaptor are 7/8” - 14 threads like the
of the Secondary regulator to 5 to 10 psi.
Disconnect the gas supply from the coupler and attach to the adaptor.
Turn on the Beer Check by touching the screen. It will open in
the CO2/N2 mode within 10 seconds. Turn on the gas to the Beer
Check.
Take a reading: The most accurate reading is when the numbers
first stop flashing for more than a few seconds.
Turn off and disconnect the Beer Check: The Beer Check will
turn itself off but you’ll increase times between battery charges if
you turn it off after you have taken a reading.
Turning the Beer Check Off: Touch “Main” and then “Off”.
Gas Supply ON
Gas Supply OFF
Using the Beer Check: Quick Start Guide
Page 2
Uncouple the keg coupler.
Remove the gas line
at the coupler by loosening the nut (keep
track of the washer).
Aach the nut to the adaptor (use the washer).
Coupled (turned ON)
Uncoupled (turned OFF)
Gas Line Nut
0.011 Ø Hole here limits flow (1 to 2 scfh @ 5 to 10 psig).
7/8-14 Thread
(Same thread as keg couplers) attach gas line nut here.
1/4 hose barb attach
Beer Check tube here.
0.022” Ø Hole here allows flow of (8 to 15 scfh @ 5 to 10 psig)
Page 3
Notes
The Beer Check has a rechargeable battery that is not replaceable in the field. Remember to keep it charged. (Baery-
shaped indicator in upper left corner of the screen.)
The operating principles of the Beer Check are such that it will only be accurate when the gases in the mix are the only ones present. For example if you were to sample the gas from a beer keg the gas would also contain water vapor and alcohol vapor making the results inaccurate.
The screen which is also a touch pad works best when tapped with a fingernail or plastic stylus.
The Beer Check will turn itself off when left untouched for 5
minutes. To turn it off manually touch the Main icon and then the
Off icon.
The Beer Check cannot be calibrated in the field. It is ideal to keep a known, accurate calibration gas to test the unit from time to time, ensuring the accuracy of your readings.
The Beer Check is affected by altitude, if you will be using it above 4,000 please discuss this with McDantim.
If there is more than one blend, repeat the process for a keg on the other blend.
If you have any questions or need assistance, contact us
Monday through Friday, 8:00 a.m. - 5:00 p.m. Mountain time.
888-735-5607
Page 4
Caution
Never leave the adaptor connected and flowing. The gases in beer systems, CO2 and N2 can cause a person to pass out without warning which can be fatal. Always have another person outside the cooler keeping track of you when working in a beer cooler. This is true whenever working on a gas supply system.
The Beer Check is intended to work in the modes described in the Beer Check Instructions. It will not work accurately with other gas combinations. We suggest checking the Beer Check periodically with a Calibration gas and having it tested once a year for accuracy.
Warranty
McDantim, Inc. will extend the following warranty to the original
purchaser of the Beer Check (MGA-01 / MV019):
If the unit becomes inoperable within the first year from the date of delivery for any reason other than unusual or abusive use or handling.
McDantim, Inc. will at its option, have the unit repaired or replaced.
If a claim is made after one year from date of delivery to the original purchaser and the unit is inoperable all costs involved will be paid by the purchaser.
In the event the unit becomes inoperable the purchaser should promptly notify McDantim, Inc. for instructions. Any return
without prior authorization is at the purchaser’s expense and risk.
McDantim, Inc. does not assume any other liability than stated above.
Phone 888.735.5607 • mcdantim.com
750 Shephard Way, Helena, MT 59601
Loading...