M&C PSS, PSS-5, PSS-5/3 Instruction Manual

Portable Gas Conditioning Unit Series PSS
PSS-5, PSS-5/3
®
Version 1.00.03
Dear customer,
This Instruction Manual does not claim completeness and may be subject to technical modifications. © 0
Germany GmbH. Reproduction of this
document or its content is not all
PSS
Version: 1.00.0
Thank you for buying our product. In this manual you will find all necessary information about this M&C product. The information in the manual is fast and easy to find, so you can start using your M&C product right after you have read the manual. If you have any question regarding the product or the application, please don’t hesitate to contact M&C or your M&C authorized distributor. You will find all the addresses in the appendix of this instruction manual. For additional information about our products, please go to M&C’s website www.mc-techgroup.com There you can find the data sheets and manuals of our products in German and English.
.
5/2019 M&C Techgroup
®
is a registered trade mark.
3
2 PSS-5 | 1.00.03 www.mc-techgroup.com
owed without permission from M&C.
List of Contents
1 General information .................................................................................................................. 4
2 Declaration of conformity ......................................................................................................... 4
3 Safety instructions .................................................................................................................... 5
4 Warranty .................................................................................................................................... 5
5 Used terms and signal indications .......................................................................................... 6
6 Introduction ............................................................................................................................... 8
7 Application ................................................................................................................................ 8
8 Technical data ........................................................................................................................... 9
9 Description .............................................................................................................................. 10
10 Receipt of goods and storage ............................................................................................ 12
11 Installation instructions ...................................................................................................... 13
12 Supply connections ............................................................................................................ 14
12.1 Tube connections ............................................................................................................... 14
12.1.1 Connecting the heated sample line with special adapter (option) ................................ 16
12.2 Electrical connections ........................................................................................................ 16
13 Commissioning ................................................................................................................... 18
14 Closing down ....................................................................................................................... 19
15 Maintenance ......................................................................................................................... 19
16 Trouble shooting ................................................................................................................. 21
17 Spare parts list .................................................................................................................... 23
18 Appendix .............................................................................................................................. 25
List of Figures
Figure 1 PSS-5 and PSS-5/3 gas flow diagram .............................................................................. 8
Figure 2 Design of the conditioning units PSS-5 and PSS-5/3 ...................................................... 10
Figure 3 Sample gas connection .................................................................................................. 14
Figure 4 Heated sample line connection with special adapter....................................................... 16
Figure 5 Electrical connection and main switch ............................................................................ 17
Figure 6 Circuit diagram PSS-5 and PSS-5/3, 115 V and 230 V ................................................... 26
www.mc-techgroup.com PSS-5 | 1.00.03 3
HEAD OFFICE
M&C TechGroup Germany GmbH Rehhecke 79 40885 Ratingen Germany
Telephone: 02102 / 935 – 0 Fax: 02102 / 935 – 111 E - mail: info@mc-techgroup.com
www.mc-techgroup.com

1 GENERAL INFORMATION

The product described in this operating manual has been examined before delivery and left our works in perfect condition related to safety regulations. In order to keep this condition and to guarantee a safe operation, it is important to heed the notes and prescriptions made in this operating manual. Furthermore, attention must be paid to appropriate transportation, correct storage, as well as professional installation and maintenance work. All necessary information a skilled staff will need for appropriate use of this product are given in this operating manual.

2 DECLARATION OF CONFORMITY

CE - Certification
The product described in this operating manual complies with the following EU directives:
EMC-Instruction
The requirements of the EU directive 2014/30/EU “Electromagnetic compatibility“ are met.
Low Voltage Directive
The requirement of the EU directive 2014/35/EU “Low Voltage Directive“ are met. The compliance with this EU directive has been examined according to DIN EN 61010.
RoHS Directive
The requirements of the RoHS2 (‘Restriction of Hazardous Substances 2’) directive 2011/65/EU and its annexes are met.
Declaration of conformity
The EU Declaration of conformity can be downloaded from the M&C homepage or directly requested from M&C.
4 PSS-5 | 1.00.03 www.mc-techgroup.com

3 SAFETY INSTRUCTIONS

Please take care of the following basic safety procedures when mounting, starting up or operating this equipment:
Read this operating manual before starting up and use of the equipment. The information and warnings given in this operating manual must be heeded.
Any work on electrical equipment is only to be carried out by trained specialists as per the regulations currently in force.
Attention must be paid to the requirements of VDE 0100 (IEC 364) when setting high-power electrical units with nominal voltages of up to 1000 V, together with the associated standards and stipulations.
Check the details on the type plate to ensure that the equipment is connected to the correct mains voltage.
Protection against touching dangerously high electrical voltages: Before opening the equipment, it must be switched off and hold no voltages. This also applies to any external control circuits that are connected.
The device is only to be used within the permitted range of temperatures and pressures.
Check that the location is weather-protected. It should not be subject to either direct rain or moisture.
The gas conditioning systems PSS-5 and PSS-5/3 must not be used in hazardous areas.
Installation, maintenance, monitoring and any repairs may only be done by authorized personnel with respect to the relevant stipulations.

4 WARRANTY

In case of a device failure, please contact immediately M&C or your M&C authorized distributor. We have a warranty period of 12 months from the delivery date. The warranty covers only appropriately used products and does not cover the consumable parts. Please find the complete warranty conditions in our terms and conditions. The warranty includes a free-of-charge repair in our production facility or the free replacement of the device. If you return a device to M&C, please be sure that it is properly packaged and shipped with protective packaging. The repaired or replaced device will be shipped free of delivery charges to the point of use.
www.mc-techgroup.com PSS-5 | 1.00.03 5

5 USED TERMS AND SIGNAL INDICATIONS

DANGER!
This means that death, severe physical injuries and/or important
This means that death, severe physical injuries and/or important
in case the respective safety
CAUTION!
This means that minor physical injuries may occur in case the
CAUTION!
Without the warning triangle means that a material damage may
ATTENTION
This means that an unintentional situation or an unintentional status
These are important information about the product or parts of the
QUALIFIED PERSONNEL
These are persons with necessary qualification who are familiar with
High voltages!
Corrosive!
Wear protective gloves!
objects or extremely high
Wear safety glasses!
objects.
material damages will occur in case the respective safety measures are not fulfilled.
material damages may occur
WARNING!
measures are not fulfilled.
respective safety measures are not fulfilled.
occur in case the respective safety measures are not met.
NOTE!
may occur in case the respective note is not respected.
operating manual which require user’s attention.
installation, use and maintenance of the product.
Protect yourself and others against damages which might be caused by high voltages.
These substances destroy living tissue and equipment upon contact. Do not breathe vapors; avoid contact with skin and eyes.
Working with chemicals, sharp temperatures requires wearing protective gloves.
Protect your eyes while working with chemicals or sharp Wear safety glasses to avoid getting something in your eyes.
6 PSS-5 | 1.00.03 www.mc-techgroup.com
Wear protective clothes!
ly high
Working with chemicals, sharp objects or extreme temperatures requires wearing protective clothes.
www.mc-techgroup.com PSS-5 | 1.00.03 7

6 INTRODUCTION

The portable gas conditioning systems PSS-5 and PSS-5/3 have been especially designed, so that precise gas analysis can be carried out in any place and at any time.
The entire gas conditioning system is housed in a compact and robust aluminium framed protective case which ensures that the components can be removed easily, and gas analysis carried out quickly, safely and with a minimum amount of maintenance.

7 APPLICATION

The gas conditioning system is ideally suited for both intermittent and continuous operation.
The components of the PSS-5 and PSS-5/3 systems are intended for "standard use." We also provide a wide range of additional equipment and other components if special measurements are required.
Gas sample probe, stainless steel 316, Ø 4/6 mm, length 0.5 m Gas sample line, PVC hose, Ø 4/6 mm, length 3 m Gas cooler ECP1000 or ECP3000 Fine filter FP-2T, filter element fineness 2 µm Sample gas diaphragm pump N3KPE or N9KPE Pre-filter PF 2 Peristaltic pump SR25.1 for continuous removal of condensate
Figure 1 PSS-5 and PSS-5/3 gas flow diagram
8 PSS-5 | 1.00.03 www.mc-techgroup.com

8 TECHNICAL DATA

Gas Conditioning Type
PSS-5
PSS-5/3
Gas outlet dew point
Range of adjustment: +2 to +15 °C [35.6 to 59 °F], factory Gas outlet dew point stability
At const. conditions: < ±0.1 °C [±0.18 °F]
Gas inlet temperature
Max. 80 °C* [176 °F*], optional: max. 180 °C* [356 °F*] with Gas inlet water vapor saturation
Max. +80 °C* [176 °F*]
Gas flow rate
Max. 150 Nl/h*
Max. 350 Nl/h*
Ambient temperature
+5 to +40 °C* [41 to 104 °F*]
Storage temperature
-25 °C to +65 °C [-4 to 149 °F]
Pressure
0.7 to 1.4 bar abs.*
Total cooling capacity
Max. 50 kJ/h
Max. 90 kJ/h
Number of gas inlets
1
Number of gas outlets
1, optional: max. 4
Material of sample contacting parts
Stainless steel, glass, PPH, PVC, PVDF, PTFE, Novoprene® Ready for operation
Approx. 10 min.
Mains power supply
230 V/50 Hz ±10 % or 115 V/60 Hz ±10 %
Power consumption
Max. 240 VA Fuse protection
4 AT, 5 x 20 mm
10 AT, 5 x 20 mm
Electrical connection
Cold appliance plug with 2 m [6.6 ft] cable
Case protection
IP20 (EN 60529)
Case type
Portable aluminium framed protective case
Case dimensions (H x W x D)
440 x 540 x 255 mm [approx. 17.3“ x 21.3“ x 10“]
Weight
Approx. 18.5 kg [40.8 lbs]
Approx. 20.0 kg [44.1 lbs]
Options
Electrical equipment standard
EN 61010
setting: +5 °C [41 °F]
stainless steel bulkhead union
Medium connections Tube connections DN 4/6
Temperature controller:
Flow meter (optionally):
Optional: Viton® for gas sample line, Part No. 01G9025
Option temperature controller and heated sample line: 230 V max. 1620 VA
115 V max. 930 VA
With option temperature controller and heated sample line:
Range of control: 0 to 200 °C [32 to 392 °F]
PT100
Input: 7 to 70 Nl/h air, 15 to 150Nl/h air, 25 to 250 Nl/h air, 50 to 500 Nl/h air Quantity: max. 4
PPH=Polypropylene, PTFE=Polytetrafluoroethylene (Teflon®), PVC=Polyvinyl chloride, PVDF=Polyvinylidenfluoride Viton®, Teflon® are registered Trademarks of DuPont Performance elastomers
* Maximum values in technical data must be rated in consideration of total cooling capacity at 25 °C [77 °F] ambient
temperature and an outlet dew point of 5 °C [41 °F].
www.mc-techgroup.com PSS-5 | 1.00.03 9

9 DESCRIPTION

Cooler
ECP
>°C
ON
<°C
EIN
Meßgas
AUS
AUS
Kondensat
max. 1800VA
GL18
GL25
11
255
540
440
PSS-5...
Type of cooler
Max. gas flow [Nl/h]
Sample gas pump
PSS-5
ECP1000
150
N 3 KPE
PSS-5/3
ECP3000
350
N 9 KPE
Figure 2 Design of the conditioning units PSS-5 and PSS-5/3
All components of the gas conditioning system are built into a portable case and are freely accessible. The case door can be opened easily to the left by loosening the tension locks mounted on the side and top of the case.
The installation of the gas cooler and a corresponding diaphragm measuring gas pump depends on the required maximum gas volume flow. The possible combinations are summarised in the following table:
The minimum amount of flow is determined by the sample gas pump (see chapter 8). If the required minimum total flow rate is not reached, excessive overpressure can lead to premature destruction of the pump diaphragm.
The gas cooler is equipped with a Duran glass heat exchanger as standard. Heat exchangers in PVDF or stainless steel are optionally available.
The FP-2T fine filter (2 µm filter porosity) installed upstream of the sample gas pump provides the necessary solids separation.
The overtemperature alarm contact (+8 °C [46.4 °F]) of the cooler automatically regulates the switching on and off of the sample gas pump.
10 PSS-5 | 1.00.03 www.mc-techgroup.com
The resulting condensate is continuously discharged by a peristaltic pump type SR25.1 . A pre-filter type PF2 is installed in the condensate line between the heat exchanger and the peristaltic pump. This protects the pump from particle contamination in the condensate.
The 4/6 mm tube connections for the condensate and sample gas lines
11 are located on the right
side of the case (see Figure 2 and Figure 5).
A stainless steel sample tube (length 0.5 m, Ø 6 mm) and 3 m PVC sample tubing (4/6 mm) are included as standard.
The ventilation grids in the lid and in the left side wall of the case provide sufficient convex forced ventilation.
Options:
The PSS-5 and PSS-5/3 sample gas conditioning unit can be equipped at the factory with a maximum of four sample gas outlets. Each sample gas outlet can be controlled according to the specified volume flow range (see table on page 9) by the optional installation of a flow meter type FM40 with needle valve. Unused mounting holes for sample gas outputs or flow meters are closed by blind caps.
To protect the downstream analyzers against liquid ingress and to increase the operational reliability of the entire system, we recommend the installation of a liquid alarm sensor type LA1S. For this purpose, the FP-2T fine filter installed as standard is replaced at the factory by the FP-2T-D fine filter with built-in liquid alarm sensor. The LA1.4 electronic controller is located on the terminal support rail (Figure 2), in the upper part of the case. The LA electronics automatically shuts off the sample gas pump in the event of a liquid alarm. The alarm is indicated by a red LED. If there is no alarm, a green LED will be on.
The PSS-5 and PSS-5/3 sample gas conditioning units can optionally be equipped with a sample gas inlet (see Figure 3, Part No. 01G9060) for connecting a heated sample line. The existing anti-kink protection must only be used for heated sample lines of connection type "C" (Part No. 03B1012). Assembly instructions can be found in the appendix.
It is also possible to connect the heated sample line Part No. 01B4036 in connection with the gas sample probe PSP 4000.
The temperature controller 701 (Part No. 01G9055) required to control the heated line is factory­installed on the terminal support rail (Figure 2).
A 3-way ball valve (Part No. 01G9046) or a 5-way ball valve (Part No. 01G9045) can optionally be installed in the inlet of the gas conditioning unit for test gas feed or sample gas switching.
www.mc-techgroup.com PSS-5 | 1.00.03 11

10 RECEIPT OF GOODS AND STORAGE

The gas conditioning and sampling systems PSS-5 and PSS-5/3 are completely pre-installed units.
Immediately after arrival take the gas conditioning system and possible special accessories carefully out of the packaging material.
Compare the goods with the items listed on the delivery note;
Check the goods for any damage caused during delivery and, if necessary, notify your transport
insurance company without delay of any damage discovered.
The gas conditioning unit should be stored in a protected
NOTE!
frost-free area!
12 PSS-5 | 1.00.03 www.mc-techgroup.com

11 INSTALLATION INSTRUCTIONS

The case should be placed on an even horizontal surface to
to the cooler. Condensate pre-separation is then not necessary.
ensure a secure and stable position.
NOTE!
The operating position is exclusively vertical. Only in this case is the proper condensate separation and discharge in the heat exchanger of the cooler guaranteed.
The gas conditioning case should be set up away from heat sources and freely ventilated so that no unwanted heat accumulation occurs.
For outdoor installation, adequate protection against direct sunlight and moisture must be provided. In winter, the installation site must be frost-free; observe the protection class of the case.
To ensure the operational safety of the portable gas conditioning unit and the downstream analyzers and to avoid false alarms, the sample gas conditioning unit must not be used outside the specified temperature range.
Downstream analyzers must always be operated at temperatures well above the specified gas output dew point of +5 °C. The temperature range of the downstream analyzers must not be exceeded. This avoids any condensation of the gas in the connecting lines to the analyzers.
Unheated gas sample lines must be installed with a gradient down
www.mc-techgroup.com PSS-5 | 1.00.03 13

12 SUPPLY CONNECTIONS

NOTE!
Do not swap tube connections; connections are marked
Option: Mounting of max. 4 FM40 flowmeters possible
EIN
Meßgas
AUS
AUS
Kondensat
1
1
2 3 4
2
3
4
Condensate
Outlet
Sample gas Outlet (Option: max. 4)
Sample gas
Inlet
Option: Electrical connection for
heated sample line
max. 6A, (230V/50Hz or 115V/60Hz)

12.1 TUBE CONNECTIONS

accordingly. After connecting all lines, the tightness must be checked.
Figure 3 shows the possible medium connections. These are located recessed in a special mounting frame on the right side of the sample gas conditioning case.
Figure 3 Sample gas connection
All tube connections are equipped with 4/6 mm sealing ring threaded hose couplings made of polypropylene (PP) for gas input temperatures of up to a maximum of 80 °C [176 °F] (see chapter 8). If heated sample lines are used, whereby the gas input temperatures are increased up to a maximum of 180 °C [356 °F], additional bulkhead unions made of stainless steel are recommended.
Dimension 4/6 mm connecting tubes are used as standard.
14 PSS-5 | 1.00.03 www.mc-techgroup.com
The sample gas tubes or condensation tubes, are to be assembled as follows:
NOTE!
The tightness of the connections can only be guaranteed if the
Remove the union nut from the sealing ring couplings by turning it anti-clockwise. The nut should be removed from the thread with great care so as to ensure that the loose sealing ring in the nut is not lost.
Place the union nut over the connecting tube.
Place the sealing ring over the connecting hose with the thicker bead towards the nut.
Place the hose over the nipple on the thread.
connecting tube has a straight rim (hose cutter).
The union nut is to be screwed tight by hand.
The tube will no longer be able to slip off, and is now compression-proof.
The tubes are to be removed in the reverse order.
WARNING!
Aggressive condensate is possible.
Wear protective glasses and proper protective clothing!
www.mc-techgroup.com PSS-5 | 1.00.03 15

12.1.1 CONNECTING THE HEATED SAMPLE LINE WITH SPECIAL ADAPTER (OPTION)

the equipment. When
the supply voltage
is identical with the information provided on the model type
For the erection of power installations with rated voltages up to 1000V, the requirements of VDE 0100 and relevant standards and
The main circuit is equipped with a fuse corresponding to the
technical data.
Sample gas IN
Bulkhead union
Back ferrule 6
mm Ø
ID
Back ferrule
10 mm Ø ID
Sealing ring 10 mm Ø ID
Sealing ring
6 mm Ø ID
Support sleeve
4 mm Ø OD
Union nut 10 mm Ø ID
Heated sample line
Adapter
Figure 4 Heated sample line connection with special adapter
Place the special adapter on the PTFE tube according to the drawing seen above;
Place the support sleeve into the PTFE tube;
Insert the Teflon tube as far as possible into the 'Sample gas IN' bulkhead fitting and hand-tighten
the adapter;
Tighten the adapter 1 1/4 turns with a wrench (SW 14), while holding the lock nut of the Schott screw connection with a wrench (SW 15);
Insert the 10mm pipe of the heating cable into the adapter as far as possible and hand-tighten with the union nut;
Tighten the union nut 1 1/4 turns with the wrench (SW 19), holding the adapter in place with the wrench;
The screw connection is now cut gas-tight and can be loosened as often as required.

12.2 ELECTRICAL CONNECTIONS

False supply voltage can demage
WARNING!
connecting the equipment, please ensure that
plate!
specifications must be observed!
NOTE!
nominal current (over current protection); for electrical details see
16 PSS-5 | 1.00.03 www.mc-techgroup.com
The PSS-5 and PSS-5/3 gas conditioning systems are available with either 230V/50 Hz or with
Main switch and cold appliance plug socket
0
Ventilation grids
115V/60 Hz (for circuit diagram see Appendix). A 4 A fuse is used on all models as fuse protection. The fuse is located on the clamp mounting rail (see Figure 2). When optionally using a temperature controller with heated line, the overload protection increases to 10 A.
The electrical connection is made via a cold appliance plug with 2 m cable on the left side of the case. The power socket is equipped with a two-pole main switch.
Figure 5 Electrical connection and main switch
Option “ heated sample line”:
For the electrical supply of a heated sample line with PT-100 sensor and/or the connection of other heated components, e.g. heated sample gas probe or heated filter, a connection socket as shown in Figure 3 is available. The maximum connected load is 6 A, 1380 W for the 230 V sample gas conditioning or 6 A, 690 W for the 115 V version.
The maximum length of the usable heated sampling line is calculated as follows:
Max. connected load [W] - Power consumption of heated components (e.g. sample probe) [W]
L(m)=
Power consumption heated line per meter [W/m]
www.mc-techgroup.com PSS-5 | 1.00.03 17

13 COMMISSIONING

NOTE!
When operating the sample gas conditioning system with a
heated sample gas line, the temperature must be checked at the
The following minimum gas flow rates result from the requirement
side load of the sample gas pumps N3
If the required minimum total flow rate is not reached, excessive overpressure can lead to premature destruction of the pump
term measurements with a high dust content in the
sampling probe must be provided to
protect the sampling line from blockages.
Before commissioning, the plant-specific and process-specific safety measures must be observed.
Before connecting the equipment to the supply voltage, the main switch must be in position “0”.
The following steps must be carried out before initial commissioning:
Plug the power plug of the supplied power cable into the power socket;
Connect the heated cable (optional);
WARNING!
temperature controller.
Connect the mains plug to the mains;
Switch the main switch to position 'I'.
Set the desired temperature on the temperature controller.
Type 701: The digital display of the controller shows the actual value of the heated sample line after switching on the sample gas conditioning unit. The controller is set to 0 °C [32 °F] on delivery. Briefly press the P key to change the set point. SP appears in the display and then the display changes to the adjusted set point. Use the arrow keys to set the desired value. After 60 seconds, the display automatically changes to the actual value.
The sample gas cooler is ready for operation after approx. 10 minutes. The overtemperature alarm contact of the cooler switches on the sample gas pump automatically when the cooler reaches the temperature of +8°C [46.4 °F].
of the maximum pressure­KPE and N9 KPE of 1.4 bar abs:
NOTE!
N 3 KPE approx. 60l/h air, N 9 KPE approx. 200l/h air.
diaphragm.
For long­sample gas, a suitable gas
18 PSS-5 | 1.00.03 www.mc-techgroup.com

14 CLOSING DOWN

NOTE!
The installation site of the gas conditioning unit must remain frost­free even when the unit is switched off.
No special measures are to be taken in the event of short-term shutdowns of the gas conditioning system.
In the case of long-term shutdowns, for example after a completed series of measurements, it is recommended to purge the gas conditioning system with fresh air or inert gas. A flushing time of 3 to 5 minutes is sufficient under normal conditions. Condensate residues must also be removed from the system.
WARNING!
Aggressive condensate is possible.
Wear protective glasses and proper protective clothing!

15 MAINTENANCE

Before carrying out maintenance work, the plant-specific and process-specific safety measures must be observed!
Dangerous voltage.
WARNING!
The maintenance cycles depend on the process conditions and must therefore be determined for each specific application.
All parts to be serviced are easily accessible and installed in the sample gas conditioning case. These are (see Figure 2):
The filter element of the preliminary filter FP-2T
Before carrying out any work on the gas conditioning unit, move the main switch to position “0” and pull out the mains plug!
.
www.mc-techgroup.com PSS-5 | 1.00.03 19
NOTE!
In order to protect downstream analyzers, the wet filter element
must always be replaced after a condensate ingress.
Peristaltic pump pre-filter PF2 ; The pre-filter must be replaced at regular intervals if the condensate is loaded with particles. The "disposable filter" is inserted into the pump tube on the suction side (see Figure 2) and can be easily replaced;
Check the tubing of the SR25.1
condensate pump every six months and replace it if necessary
(see operating instructions in the appendix);
Check the diaphragm of the gas feed pump N3 KPE or N9 KPE
every six months and replace
if necessary.
20 PSS-5 | 1.00.03 www.mc-techgroup.com

16 TROUBLE SHOOTING

Indication
Problem
Possible Cause
Check/Solution
Upper LED of
Interruption of
No voltage;
Needle valve closed.
Check supply voltage with model type plate;
Remove sample gas hose on analyser side and
Momentary overloading of the cooler due to
The following table aims to point out possible operational problems and offer solutions to such problems (not applicable during the starting procedure).
the cooler lights up red
Middle LED on cooler is green;
Alarm LED on the LA electro­nics is red (see
4);
gas flow;
Cooler in operation but gas flow interrupted;
Cooler does not function. Cooler alarm registers ‘excess temperature’. Cooler turns gas measuring pump of automatically.
Membrane pump faulty;
Contamination of the membrane pump;
Sample probe/hose clogged up or lead squashed;
Sample gas hose to analyser clogged up or squashed;
Optional liquid alarm sensor: Sensor turns measuring pump off automatically;
Optional flowmeter(s):
OK? Check if supply voltage plug is inserted correctly and if the main switch is turned in position “1” OK? Check fine fuse on connector block  (Figure 2); OK?
Ambient temperature too high. OK? Free convection in case impaired  case temperature too high; OK? Cooler error (check instruction manual ECP1000/3000). OK? Check voltage on clamps X1/11 and X1/8;
OK? Remove the hoses at head of pump and check; OK?
Clean pump if necessary; OK? Remove tube at gas inlet (see chapter 12.1); Gas flow? Clean contaminated lead or replace;
No gas flow?
check via hose thread if sample gas flows;
No gas flow? Clean contaminated leads or replace; Gas flows?
excessive amount of condensate; OK? Check hoses for condensate removal; OK? Check Pre-filter and if necessary replace; OK? Check peristaltic pump tube (see manual SR25.1); OK? Check peristaltic pump (see manual SR25.1); OK? Check cooler according to instruction manual;
Adjust needle valve(s) to the desired flow.
www.mc-techgroup.com PSS-5 | 1.00.03 21
Indication
Problem
Possible Cause
Check/Solution
Middle LED on
Cooler and
Pre-filter  clogged
pump off.
Remove filter from condensate lead;
cooler is green;
LED of the LA electronics is green (see 4).
gas pump in operation; condensate in sample gas lead;
up;
Pump hose faulty;
Peristaltic pump SR25.1 faulty;
Unsufficient drying of sample gas;
Optional liquid alarm sensor: Sensor has not turned
Pump delivering? Change filter; Pump not delivering?
Change pump hose (see manual peristaltic pump SR25.1); OK?
Check peristaltic pump (see manual peristaltic pump SR25.1); OK?
Check cooler (see manual ECP1000/3000);
Check the LA electronics and if necessary replace.
22 PSS-5 | 1.00.03 www.mc-techgroup.com

17 SPARE PARTS LIST

Portable Sampling System Versions PSS-5, PSS-5/3
recommended quantity
operation [years]
C/R/S 1 2
3
Fine filter FP-2T: (see Fig. 2)
90F0002
Filter element F-2T, PTFE, 2 µm
C 6 12
20
90F0040
Viton O-ring, 26 for FP-
R 1 1
1
90F0056
PVDF filter element clamp F-P
S - -
1
90F0012
Filter body F-120G of glass
R 1 1
1
Fine filter FP-2T with Option LA1S:
90F0015
Filter body F-120G-D of glass with GL25 condensate connection thread
R 1 1
1
90F0020
Union nut GL 25
R 1 1
1
90F0025
PTFE sealing ring GL 25 - 12mm Ø
R 1 1
1
Peristaltic pump SR25.1:
90P1007
SR25 pump hose with PVDF tube connectors DN 4/6 mm
C 1 2
4
Diaphragm pump type N3 KPE/KP18; N5 KPE/KP18
90P2100
Square cap type D3, 1/8“i for N3/N5 KPE/KP18 Material: PVDF
S - -
1
90P2120
Diaphragm type S3, for N3/N5 KPE/KP18, Material: Viton, PTFE coated
C 1 2
3
90P2111
Valve reed type V3 with O-ring type O3, for N3­N5, 1 pc, material: Viton (2 pieces required)
C 2 4
6
90P2105
Intermediate plate type Z3, for N3/N5 PE/KP18 Material: PVDF
S - -
1
Diaphragm pump type N9 KPE/KP18
90P2200
Square cap type D9, 1/8“i for N9 KPE/KP18, Material: PVDF
S - -
1
90P2220
Diaphragm type S9, for N9 KPE/KP18, Material: Viton, PTFE coated
C 1 2
3
90P2211
Valve plate with seal for N9 KPE, 1 pc., material: Viton. (2 pcs./pump)
C 2 4
6
90P2205
Intermediate plate type Z9, for N9 KPE/KP18, Material: PVDF
S - -
1
Option flowmeter FM40:
90A0015
Flowmeter glass for FM40 range 7-70 l/h air
S - 1
1
Wear, tear and replacement part requirements depend on specific operating conditions. The recommended quantities are based on experience and are not binding. For spare parts of components which are not presented in the following list please see the specific instruction manuals or leaflets added in the appendix.
(C) consumable parts, (R) recommended spare parts, (S) spare parts
PSS-5 and PSS-5/3 being in
www.mc-techgroup.com PSS-5 | 1.00.03 23
Portable Sampling System Versions PSS-5, PSS-5/3
(C) consumable parts, (R) recommended spare parts, (S) spare parts
recommended quantity
operation [years]
C/R/S 1 2
3
94F0010
Flowmeter glass for FM40 range 15-150 l/h air
S - 1
1
94F0015
Flowmeter glass for FM40 range 25-250 l/h air
S - 1
1
94F0020
Flowmeter glass for FM40 range 50-500 l/h air
S - 1
1
90A0018
Viton O-ring 9 for flowmeter glass FM40
R 2 4
6
Diverse:
90G0006
Pre-filter PF 2 for condensate pump SR25.1
C 5 10
15
90K6030
Fine fuse 4 A T, 5 x 20 mm for PSS-5, PSS-5/3
R 5 5
5
90G0020
Fine fuse 10 A T, 5 x 20 mm for PSS-5, PSS-5/3
with option temp. controller and heated sample line
R 5 5
5
Hose and hose fittings:
05V3230
Bulkhead union SV-PP DN 4/6
PSS-5 standard PPH = Polypropylene
R 2 2
2
05V3215
Bulkhead union SV-PVDF DN 4/6
PSS-5 optional PVDF = Polyvinylidenfluoride
R 2 2
2
05V6500
Sealing ring 4/6 PP
R 5 10
10
05V6600
Sealing ring 4/6 PVDF
R 5 10
10
05V6505
Union nut M10-4/6 PP
R 5 10
10
05V6605
Union nut M10-4/6 PVDF
R 5 10
10
01T4000
Hose PVC DN 4/6 (meters)
S 3 6
9
01T1000
Hose Viton DN 4/6 (meters)
S 1 2
3
01T2000
Hose Novoprene DN 3,2/6,4 (per meter)
S 1 2
3
02B1000
Hose PTFE DN 4/6 (per meter)
S 1 2
3
10T1000
Hose cutter
S 1 1
1
PSS-5 and PSS-5/3 being in
see above
see above
see above
see above
24 PSS-5 | 1.00.03 www.mc-techgroup.com

18 APPENDIX

Circuit diagram PSS-5 and PSS-5/3
More product documentation is available in our Internet catalogue:
www.mc-techgroup.com
Instruction manual electric gas cooler ECP 1000, ECP 3000
Data sheet: Universal-Filters FP, FT, FS, FSS
Document: 7.1
Instruction manual diaphragm pump Series N
Instruction manual peristaltic pump SR25.1, SR25.1-G
Data sheet: Liquid alarm sensor LA1S and electronic controller type LA1.4
Document: 8.1
• Data sheet: Flow meter FM40 Document: 9.2
• Data sheet: Ball valves L/PV-1 Document: 11.1
Data sheet: Temperature controller 701 Document: 4.2
www.mc-techgroup.com PSS-5 | 1.00.03 25
Figure 6 Circuit diagram PSS-5 and PSS-5/3, 115 V and 230 V
26 PSS-5 | 1.00.03 www.mc-techgroup.com
Loading...