Mbv MAS-100 Operator’s Manual

Operator’s Manual MAS-100
Professional Microbial Air Monitoring System
for the Microbiological Testing of Air
MAS-100
MBV AG Tel. +41 44 928 30 80 MAS-100_Usermanual_blank_010606e.doc Microbiology and Bioanalytic Fax +41 44 928 30 89 Date: 02.2006 Laubisrütistrasse 24 Mail: info@mbv.ch Version: 01 Revision: 06 CH-8712 Stäfa www.mbv.ch www.mbv.ch
from Firmware Version 5.0x
Table of contents
1 Introduction 4 2 Basic principle 4 3 Most important advantages of the MAS-100™ 5
3.1 First time use of MAS-100™ 5
3.2 New: SQS- Sequential air sampling 5
4 General functions 6
4.1 Overview 6
4.2 First time use of MAS-100 = Activation of calibration reminder 6
4.3 Switching ON and OFF the MAS-100 6
4.4 Calibration reminder 7
4.5 Selectabel Mode 3 7
4.6 Last volume used 7
4.7 Select other volumes 7
4.8 Delay function 8
4.9 User menu 8
4.10 Start sampling 8
4.11 Adjusting Petri dish 9
4.12 Adjusting RODAC plates 9
5 Perform a “Manual” air sampling 10
5.1 Switch the MAS-100 „ON” for manual sampling 10
5.2 Sampling interrupted; failed 11
5.3 Switching off the MAS-100 11
5.4 Interrupting a test 11
5.5 Airflow blocked 12
6 Perform air sampling in “Template mode” 13
6.1 Switch the MAS-100 „ON” for template sampling 13
7 Perform air sampling in “Selectable-Mode” 15 8 Instrument settings “Parameter Mode 4” 16
8.1 Clear protocol data 16
8.2 Connecting a Printer 16
8.3 Print protocol 16
8.4 Back to main menu 17
8.5 Set time and date 17
8.6 Selecting date format (EU or US) and changing time and date 18
8.7 Battery discharge? 18
8.8 Anemometer mode 18
8.9 Change volumes 18
8.10 Delay menu (Change / Activate / Deactivate ) 19
8.11 Select operation mode 19
8.12 Manual mode 19
8.13 Selectable mode 19
8.14 Template mode 19
8.15 Select SQS mode 20
8.16 User menu 20
8.17 Back to main menu 20
MBV AG, CH-8712 Stäfa +41 44 928 30 80 Version: 01 page 2 of 34 Released and approved : Hzi Revision: 06
02.2006 / MAS-100_Usermanual_blank_010606e.doc
9 Sampling head 21
9.1 Adjustment of Petri dish 21
9.2 Sampling head 400 x 0,7 21
9.3 Sterilization of sampling head 21
10 Software for MAS-100™ 21
10.1 Software download new MAS-100 C&C 21
11 Interpreting results 22 12 Charging the MAS-100™ 22
12.1 Battery charging 22
12.2 Battery discharge 22
12.3 Battery status display 22
12.4 Battery replacement 23
13 Calibration 23
13.1 Protected mode 23
13.2 Calibration reminder 23
14 Technical specifications 24
14.1 Environmental conditions 24
15 Components check list 25 16 Care and maintenance 25 17 Warranty 25 18 Accessories and spare parts 26 19 References 26 20 Appendix 26
20.1 Table of statistical corrections according to Feller for SH 400 x 0.7mm 27
20.2 Firmware Flow Diagram MAS-100: Manual Mode 1 28
20.3 Firmware Flow Diagram MAS-100: Template Mode 2 29
20.4 Firmware Flow Diagram MAS-100: Selectable Mode 3 30
20.5 Firmware Flow Diagram MAS-100: Parameters Mode 4 31
20.6 Firmware Flow Diagram MAS-100: Calibration reminder Mode 5 33
20.7 CE Conformity declaration 34
MBV AG, CH-8712 Stäfa +41 44 928 30 80 Version: 01 page 3 of 34 Released and approved : Hzi Revision: 06
02.2006 / MAS-100_Usermanual_blank_010606e.doc

1 Introduction

There is increasing need for effective air sampling in all areas where airborne micro-organisms may contaminate or otherwise affect industrial products and processes. Monitoring is especially important in:
the pharmaceutical industry
the food industry
hospitals
the cosmetics industry
the environment
Today a number of monitoring methods are available, most of which are based on the impaction principle. During development of the MAS-100 eliminated, and their best aspects retained, leading to the creation of a user-friendly, dependable and precise air sampler for standard Petri dishes. The MAS-100
aspirates 100 liters of air per minute, allowing up to 2000 liters to be sampled per cycle. This performance enables sampling to be carried out according to the strict conditions of monitoring in sterile and „clean” areas (see point 1 in chapter
The impaction velocity (speed at which airborne micro-organisms hit the surface of the agar) is approximately 11 meters per second, corresponding to stage 5 of the Andersen Sampler (see point 3 in chapter
19). This velocity guarantees that all particles > 1 µm are collected. Comparisons with other methods of air sampling ( see point 2 in chapter numbers of micro-organisms collected.
Note:
The MAS-100 ISO 14698-1:2003(E)
, all the shortcomings of existing sampling systems were
19).
19) have not revealed any significant differences in the
corresponds to International Standard
1

2 Basic principle

The MAS-100 sampler (see point 3 in chapter is directed onto a standard agar poured plate. After the collection cycle, the Petri dish is incubated and the colonies are counted. The MAS-100 aspirated volume is continuously monitored. The system measures the inflow of air and regulates the aspirated volume to a constant value of 100 liters per minute. The MAS-100 this volume if there are external factors preventing a steady flow of air, or if overfilled Petri dishes are blocking or impeding flow. For measurement in pharmaceutical industry the recommended sampling volume – according to international guidelines – should be 1000 liters.
1
ISO 14698-1:2003(E) Cleanrooms and associated controlled environments - Biocontamination control
Part1: general principles and methodes; chapter 3.1.13; 5.3;
MBV AG, CH-8712 Stäfa +41 44 928 30 80 Version: 01 page 4 of 34 Released and approved : Hzi Revision: 06
02.2006 / MAS-100_Usermanual_blank_010606e.doc
is a high-performance instrument that is based on the principle of the Andersen air
19), which aspirates air through a perforated plate. The resultant airflow
operates with a high-performance suction device, and the
automatically regulates
3 Most important advantages of the MAS-100™
new: individual activation of calibration reminder
new: Adapter-Kit for RODAC-Plates
new: Validated Firmware MAS-100 V5.0X by Novatek for „Environmental Monitoring” software
new: SQS-Cycle (see point 5 in chapter
volume may be split up to 20 sequences over a period of total 24 hours
Uses standard 90-100mm Petri dishes
Data communication for error-free results and automatic record-keeping
Manual and/or template mode
Build in anemometer for automatic volume compensation: Differences in Petri dish fills cannot affect
airflow
Indication of low airflow
Real time clock and calendar
Compact and easy to use
Easy to calibrate
Adjustable for different Petri dish sizes
100/240V, 50-60kHz power unit, or battery operation
According to international Guidelines: Isokinetic measurement, 0.45 m/sec, 100 liters / minute
Sampling volume adjustable from 1 – 2000 liters
3.1 First time use of MAS-100
If the MAS-100
4.2 and 20.6)
(see
is used for the first time or after a calibration the “calibration reminder" has to be set.

3.2 New: SQS- Sequential air sampling5

A newly developed control system ( Software V 5.0x) allows a measurement with the MAS-100 prolonged time with a single agar plate. Hereby, without any microbial loss, up to 20 air samples can be accumulated on the same Petri dish over a period of several hours. The main advantage of this method is that the risk of secondary contamination by the operators during the exchange of plates is reduced to a minimum. This is of very great importance for ambient air monitoring in sterile zones. In addition, by collecting sequential samples over a prolonged period of time the randomization is improved and thus a more reliable result obtained. Moreover, instead of several 90mm standard Petri dishes, only one dish has to be evaluated, thus reducing the expenditure on equipment and labour.
19) may be selected. If this cycle is activated, the desired
over a
5
SQS-Sequential air sampling; Validation of the microbial count in the air using the MAS-100; Swiss
Pharama 26 (2004) No.11,9-11
MBV AG, CH-8712 Stäfa +41 44 928 30 80 Version: 01 page 5 of 34 Released and approved : Hzi Revision: 06
02.2006 / MAS-100_Usermanual_blank_010606e.doc

4 General functions

4.1 Overview

The MAS-100 „Modify?” by pressing the „yes” or „no” button. The program works its way automatically through all the required parameters. Once you have set the aspirated volume and delay, the program goes to the „Start ?” prompt. When you press „yes”, the MAS-100 green light indicates that the instrument is operating. The sampled volume and time remaining are continuously displayed. Once the pre-set volume has been sampled, the red „stop” LED lights.
4.2 First time use of MAS-100 = Activation of calibration reminder
MAS100 V5.XX S/N: XXXXX
Set calibration reminder Y/N
Set reminder to 12 month Y/N
Are you sure to set xx month Y/N
Calibration valid: XX month
4.3 Switching ON and OFF the MAS-100
MAS100 V5.XX S/N: XXXXX
Calibration
valid: XX months
XXXXX litres left to aspirate
Volume:XXXX Y/N?
Select other volumes Y/N? _
has built-in dialog software. To accept a parameter or command, answer the question
starts to sample the pre-set volume of air, and a
Press „yes“ to switch on the MAS-100
for the first time. The firmware version is shown on the display. The calibration reminder has to be set by the first user. (see
20.6)
If you select „no” the instrument will go off immediately. If you select „yes“ next window will be displayed.
The reminder is set by the manufacturer to 12 month. If you would like to change , push „no” as many times until you get the time for recalibration between 1 and 12 month. If you select 0, the calibration reminder is deactivated. Push „yes” to accept the displayed value.
If you select „no” the instrument will go off immediately. If you push „yes“ the selected time for recalibration starts to count down. The last two month will be displayed as 60 days. After this the menu will proceed as written in (
The time until next calibration is shown . The last 2 month will be displayed in days.
Press „yes“ to switch on the MAS-100
The firmware version V5.XX and the serial number of the instrument is shown on the display. (see
20.2)
After a few seconds the display will indicate the remaining time until next calibration.
Again a few seconds after the display will show the total amount of liters which can be sampled with the battery charge.
The last used volume is displayed. Accept by pushing „yes”. If you push „no” next step is displayed.
With „yes” you can select other volumes. Accept a volume with „no” the menu changes into the start menu.
4.3).
MBV AG, CH-8712 Stäfa +41 44 928 30 80 Version: 01 page 6 of 34 Released and approved : Hzi Revision: 06
02.2006 / MAS-100_Usermanual_blank_010606e.doc
dd.mm.yy hh:mm
Start? (Y/N)
With „yes“ start the sampling. With „no” the menu starts with the last used volume.
If no further manipulations are made on the Instrument and no run is started, the instrument will switch off automatically after 5 minutes.
Exception:
With the calibration reminder being activated and the calibration deadline reached, the display first shows „Calibration time expired” and then „Please recalibrate unit”. If you press „yes“ or „no“, you will see the date on which the MAS-100
was last calibrated. The program then moves to the next prompt.
Note:
The MAS-100
is permanently on if connected to the battery charger.
Note:
After a collection cycle the instrument will display the last volume collected. It will shut off after 30 minutes unless you press „yes“ or „no“„ to activate the 5-minute automatic shut off.

4.4 Calibration reminder

Calibration time
expired!
Please
recalibrate unit
A calibration reminder is displayed once the calibration expiry date has been reached. The calibration reminder is set from the manufacturer to 12 month. The last two month will be displayed as 60 days. The reminder may be set from 0 up to 12 month. 0 month means deactivation of calibration reminder.
Last calibration Volume:XXXX Y/N?
If yes or no is pushed the last calibration date is displayed.
Note:
To program the calibration reminder see Mode 5 (see
4.2 and 13.2).

4.5 Selectabel Mode 3

In parameter Mode 3 it is possible to select between „manual mode”, „template mode” or „selectable mode” (see
The „manual mode” is selected if no PC is connected to the MAS-100
20.4)
The „selectable mode” is selected if only a few data are stored and transferred to the PC
The „template mode” is used if all information is transferred from and to the PC

4.6 Last volume used

After the MAS-100 is started always the last volume or SQS-cycle used will be displayed and may be used again without changing any parameters.

4.7 Select other volumes

Volume:XXXX Y/N?
If the volume is displayed and „no” is pressed the MAS-100 volume selection menu. If you accept with „yes” the start menu is displayed.
changes to the
MBV AG, CH-8712 Stäfa +41 44 928 30 80 Version: 01 page 7 of 34 Released and approved : Hzi Revision: 06
02.2006 / MAS-100_Usermanual_blank_010606e.doc
Select other volumes Y/N ?
The MAS-100
has 5 pre-set sampling volumes: 50, 100, 250, 500 and
1000 liters. These can be found in parameter Mode 5 under „Change volumes?” Each of these volumes can be modified between 0 and 2000 liters. Setting a volume to 0 will blank it from the display in the main menu. Each volume can be re-activated and be displayed again. Accept a displayed volume by pressing „no“.

4.8 Delay function

Delay: 0:00 Modify ? (Y/N)
If the delay function is activated in Parameter Mode 4, the display shows „Change delay? (Y/N)“. By choosing „yes“ you may set a start delay between 1 and 60 minutes. This function makes sense if the room in which the sampling takes place has to be left during the sampling process.

4.9 User menu

User X change user? Y/N
If the user menu is activated it is possible to choose between 10 user names. The user names may be programmed with the MAS-100 C&C software (see
10.2)

4.10 Start sampling

dd.mm.yy hh:mm To start sampling, press „yes“. If a delayed start has been programmed, the Start? (Y/N)
sampling cycle only starts when the delay time has elapsed.
Volume [l]: XXXX Time: h.mm
During the run the display shows the sampled volume and counts down the time until the total sampling volume is reached.
MBV AG, CH-8712 Stäfa +41 44 928 30 80 Version: 01 page 8 of 34 Released and approved : Hzi Revision: 06
02.2006 / MAS-100_Usermanual_blank_010606e.doc

4.11 Adjusting Petri dish

Before using the MAS-100 for the fist time you will need to adjust the blue Petri dish support. Lift off the perforated lid and place a Petri dish on the dish support. Use a 3 mm Allen key to adjust the three blue jaws until the Petri dish is securely supported and there is no play. Check that the Petri dish is properly supported by turning the sampling head to the horizontal position. The Petri dish support is now adjusted. The procedure need only be repeated at most if you change Petri dish manufacturer.
Note:
Do not over-tighten the blue jaws. If you do, the Petri dish may break and shed plastic debris in the aspiration unit and block the motor.

4.12 Adjusting RODAC plates

New : The MAS-100 may be used with commercially available contact plates. The use of a special Adapter No. 1.09214.0001 and Sampling head N0. 1.09213.0001 is necessary.
Take the allen key and exchange the 3 existing blue jaws by removing the allen screw and replace with the 3 new blue jaws.
Place the positioning template and mark the centre (see picture) with a marker pen.
Remove the positioning template and put the Adapter ring as shown in the picture bellow.
Put now your contact plate in the adaptor and tighten the 3 allen screws. The marked point should be in the centre of the contact plate as shown in the picture.
Note:
Tighten the blue jaws and make sure that the stainless steel adapter holds firmly .
MBV AG, CH-8712 Stäfa +41 44 928 30 80 Version: 01 page 9 of 34 Released and approved : Hzi Revision: 06
02.2006 / MAS-100_Usermanual_blank_010606e.doc

5 Perform a “Manual” air sampling

5.1 Switch the MAS-100 „ON” for manual sampling
To work in manual mode you just push the yes button to switch on the MAS-100
Note:
The MAS-100
must be in the manual mode. If it is in the selectable or in
the template mode change to manual mode (see
Place the MAS-100
Open the perforated lid (with attached dust cover) by rotating counter clock wise.
Place a closed standard Petri dish
on a firm support.
2
filled with agar on top of the dish support.
Take the lid off the Petri dish.
Close the MAS-100 perforated lid.
The sampling head can be adjusted to any angle from horizontal to vertical airflow direction.
Program the MAS-100 as follows
MAS100 V5.XX S/N: XXXXX
Press „yes“ to switch on the MAS-100 The firmware version V5.XX and the serial number of the instrument is shown on the display.
Calibration
valid XX months
After a few seconds the display will indicate the remaining time until next calibration.
Exception:
With the calibration reminder being activated and the calibration deadline reached, the display first shows „Calibration time expired” and then „Please recalibrate unit”. If you press „yes“ or „no“, you will see the date on which the MAS-100
was last calibrated. The program then moves to the next prompt.
XXXXX litres left to aspirate
Again a few seconds after the display will show the total amount of liters which can be sampled with the current battery capacity.
Volume:XXXX Y/N?
The last used volume is displayed. Accept by pushing „yes”. If you push „no” next step is displayed.
Select other With „yes” you can select other volumes. Accept a volume with „no” the volumes Y/N? _
menu changes into the start menu.
User X change user ? Y/N
Delay: mm:ss Modify? Y/N
If the user menu is activated (see 8.16) it is possible to change the user with „yes“ and accept it wit „no“.
Modify „Delay” with „yes” and accept with pressing „no“. If „Delay” is not active, the start menu will be displayed.
dd.mm.yy hh:mm Start? (Y/N)
Date and time are displayed. To start the sampling press „yes“. To change some of the parameters press „no“ and change the desired parameters. If no further manipulations are made on the instrument and no run is started, the instrument will shut down automatically after 5 minutes.
.
4.5 and 6).
2
Petri dish or if the contact plate adapter is used contact plate.
MBV AG, CH-8712 Stäfa +41 44 928 30 80 Version: 01 page 10 of 34 Released and approved : Hzi Revision: 06
02.2006 / MAS-100_Usermanual_blank_010606e.doc
Note:
The MAS-100
is permanently on if connected to the battery charger.
Volume [l]: XXXX Time: h.mm
During the run the display shows the sampled volume and counts down the time until the total sampling volume is reached.
passed
test completed
After the sampling cycle, the message „passed test completed” indicates that the total volume has been sampled without any disturbances and the test is valid.
Note:
After a collection cycle the instrument will display the last volume collected. Press „yes“ or „no“ to activate the 5-minute automatic shut-off.
Volume [l]: XXXX At the end of the sampling the total volume sampled is displayed and the Time: 0:00
time shows 0:00 and the red (stop) LED is on.

5.2 Sampling interrupted; failed

failed! If „failed repeat test” is displayed the test was interrupted by pressing „no”
repeat test!
during sampling.
Air flow blocked
If the air flow was blocked and the MAS-100 could not get 100 liter/min “air flow blocked” will be displayed and the red LED is blinking. Quit with „yes” or „no”
Note:
Make sure that the dust cover was removed before start was pushed, and there is not to much agar in the Petri dish which could block the wholes on the sampling head.
Open the sampling head, cover the Petri dish with the lid and remove the Petri dish from the dish support.
The Petri dish is now ready for incubation.
5.3 Switching off the MAS-100
Volume [l]: XXXX Time: 0:00
After sampling, the red (stop) LED will light up and the total volume collected will be displayed. In this position the MAS-100 after 30 minutes. In all other menu positions the MAS-100
will automatically shut down
switches off
automatically after 5 minutes.
Note:
After a collection cycle the instrument will display the last volume collected. Press „yes“ or „no“ to activate the 5-minute automatic shut-down.

5.4 Interrupting a test

failed!
repeat test!
If the „no“ button was pressed during sampling, the instrument will display „failed! repeat test!”. Repeat the test by using a new Petri dish filled with media.
MBV AG, CH-8712 Stäfa +41 44 928 30 80 Version: 01 page 11 of 34 Released and approved : Hzi Revision: 06
02.2006 / MAS-100_Usermanual_blank_010606e.doc
Note:
If the MAS-100 will not be registered by the software, unless the total volume has been collected without interruption.

5.5 Airflow blocked

failed!
repeat test!
If the built-in sensor does not attain 100 liters of air per minute, the message „failed! repeat test!” and „Air flow blocked” appears on the display and the red LED flashes. Sampling is stopped. Remove the Petri dish and start a new test. Press „yes“ or „no“ to cancel the error message. Repeat the test.
Air flow blocked
If the air flow was blocked and the MAS-100 could not get 100 liter/min air flow blocked will be displayed and the red LED is blinking. Quit with „yes” or „no”
Note:
Make sure that the dust cover was removed before start was pushed, there is not to much agar in the Petri dish which could block the wholes on the sampling head.
was activated in the „template mode”, the interrupted test
MBV AG, CH-8712 Stäfa +41 44 928 30 80 Version: 01 page 12 of 34 Released and approved : Hzi Revision: 06
02.2006 / MAS-100_Usermanual_blank_010606e.doc

6 Perform air sampling in “Template mode”

6.1 Switch the MAS-100 „ON” for template sampling
To work with templates the calibration MAS100 C&C software is needed. It is possible to download the software from internet: provided by the local distributor. More details are given in the Software manual.
Note:
Place the MAS-100
Open the perforated lid (with attached dust cover) by rotating counter clock wise.
Place a closed standard Petri dish filled with agar on top of the dish support.
Take the lid off the Petri dish.
Close the MAS-100
The sampling head can be adjusted to any angle from horizontal to vertical airflow direction.
Program the MAS-100
MAS100 V5.XX S/N: XXXXX
Calibration
valid XX months
Exception:
XXXXX litres left to aspirate
Manual mode modify? (Y/N)
Template mode
modify? (Y/N)
Template name Modify? Y/N?
www.mbv.ch . A special communication cable between PC and MAS-100™ is
The MAS-100 must be in the selectable or in the template mode
4.5 and 6)
(see
on a firm support.
perforated lid.
as follows
Press „yes“ to switch on the MAS-100
The firmware version V5.XX and the serial number of the instrument is shown on the display.
After a few seconds the display will indicate the remaining time until next calibration.
With the calibration reminder being activated and the calibration deadline reached, the display first shows „Calibration time expired” and then „Please recalibrate unit”. If you press „yes“ or „no“, you will see the date on which the MAS-100
was last calibrated. The program then moves to the next prompt.
Again a few seconds after the display will show the total amount of litres which can be sampled with the battery charge.
With „yes“ you can modify from Manual to Template mode. With „no” you change to next menu point. volume is accepted.
To accept Template mode push „no” . The first Template name will be displayed.
Choose the Template you like to work with and accept by pressing „no”. Example: Template name = Insulin 123
MBV AG, CH-8712 Stäfa +41 44 928 30 80 Version: 01 page 13 of 34 Released and approved : Hzi Revision: 06
02.2006 / MAS-100_Usermanual_blank_010606e.doc
Location Device
Each template may contain up to 10 sampling places defined with Location and Device. Push „no” to choose the position you would like to take the sample. If the position has already been measured and the rum was not interrupted the position is hidden and appears only after having transferred the data to a PC or if the values have been erased in the MAS-100 (see Clear protocol data
8.1)
Example:
1. Location name = Fill_Line 123_A, Device = Position 123_A
2. Location name = Fill_Line 123_B, Device = Position 123_B
Delay: mm:ss Modify? Y/N
If delay menu is activated modify „Delay” with „yes” and accept with pressing „no“. If „Delay” is not active, the start menu will be displayed.
User X change user ? Y/N
If the user menu is activated (see „yes“ and accept it with „no“.
8.16) it is possible to change the user with
Delay: mm:ss Modify? Y/N
The „Delay” may be accepted by pressing yes and changed by pressing „no“. If „Delay” is not active, the start menu will be displayed.
dd.mm.yy hh:mm Start? (Y/N)
Date and time are displayed. To start the sampling press „yes“. To change some of the parameters press „no“ and change the desired parameters. If no further manipulations are made on the instrument and no run is started, the instrument will shut down automatically after 5 minutes.
Note:
The MAS-100
is permanently on if connected to the battery charger.
Volume [l]: XXXX Time: h.mm
During the sampling the aspirated volume and the remaining sampling time are displayed.
passed
test completed
After the sampling cycle, the message „passed test completed” indicates that the total volume has been sampled without any disturbances and the test is valid.
Note:
After a sampling run the last used volume is displayed. Push „yes“ or „no“ to shut off the MAS-100 after 5 minutes.
Volume [l]: XXXX Nach dem Sammelvorgang wird das total gesammelte Volumen angezeigt Time: 0:00
und die verbleibende Zeit ist 0:00 und die rote (stop) LED leuchtet auf.

6.2 Interrupted sampling

failed!
repeat test!
If „failed repeat test” is displayed the test was interrupted by pressing „no” during sampling.
Air flow blocked
If the air flow was blocked and the MAS-100 could not get 100 liter/min air flow blocked will be displayed and the red LED is blinking. Quit with „yes” or „no”
Open the sampling head, cover the Petri dish with the lid and remove the Petri dish from support.
The Petri dish is now ready for incubation.
MBV AG, CH-8712 Stäfa +41 44 928 30 80 Version: 01 page 14 of 34 Released and approved : Hzi Revision: 06
02.2006 / MAS-100_Usermanual_blank_010606e.doc

7 Perform air sampling in “Selectable-Mode”

Selectable mode is used if the user wants both, to work with individual samples and with preset values in the „Template mode”. In this mode the display asks before each sampling run, whether sampling is to be performed manually or at a location designated by the template. In „Template mode”, the user is prompted to select the protocol and location before sampling is started.
MBV AG, CH-8712 Stäfa +41 44 928 30 80 Version: 01 page 15 of 34 Released and approved : Hzi Revision: 06
02.2006 / MAS-100_Usermanual_blank_010606e.doc

8 Instrument settings “Parameter Mode 4”

Pressing „yes“ and „no“ simultaneously for about 5 seconds after you Volume XXXX Y/N?
Note:

8.1 Clear protocol data

Clear protocol data? (Y/N)

8.2 Connecting a Printer

activated the MAS-100 or in any other menu position, will take you to the instrument settings „Parameter Mode 4”. This mode allows to erase data in the memory if the „template mode” was used. It is also possible to connect a printer. All other user-specific data such as date, time, volume, start delay etc. can be set.
If the MAS-100 is working in the Protected mode (see
13.1) the access without the MAS-100 C&C Software, is only possible to the following menu points:
Print protocol
Battery discharge
Anemometer mode
Press „yes” to erase protocol data. If you are in the selectable mode you can choose between erasing the manual or the template data. With the function „Clear all prot.” It is possible to erase all data collected in the PC mode. With the function „Modify? (Y/N) an individual protocol may be selected and the data erased. During this process the red and green LED’s are flashing and a message „Clear data in progress” is displayed. Erasing all the protocol data together may take up to 2 minutes.
The output is configured for a portable thermo printer*
(model on picture: CUSTOM s’print
®
)
A connection cable is available from your supplier. (* ask MAS-100
supplier for other printer models)

8.3 Print protocol

Print Protocol ? (Y/N)
It is possible to connect a printer directly to the MAS-100 configured for a portable thermo printer (model CUSTOM s’print A connection cable is available from your supplier.
. The output is
®
)
Protocol 1..10 You can choose one of the protocol or print all protocols. Modify ? (Y/N)
MBV AG, CH-8712 Stäfa +41 44 928 30 80 Version: 01 page 16 of 34 Released and approved : Hzi Revision: 06
02.2006 / MAS-100_Usermanual_blank_010606e.doc
no data present, empty memory !
If no data are present and you select „yes” nothing will happen. When you push „no” the display will show you that no data are present.
No printer
device connected
If „Print protocol” is activated and no printer is connected the display shows „no“ printer device connected. Push „no“ to go to the next step. With „yes“ back to main menu with „no“ to the user parameter.
Change online print mode? Y/N
If you want to change the print mode push „yes”. Press as many times „yes” until you have found the number of samples after which you would like to have a print out. After the total possible stored amount of samples the next message will be displayed.
Online print If „yes” is pressed the print mode will be activated or deactivated. If activated deactivated
next step will be displayed.
Print after xx samplings Y/N _
If „yes” is pressed the number of samples may be increased before the samples must be printed out. When the instrument is delivered the max. number is 50. With the MAS-100 C&C Software this value may be set up to
100. Quit to next menu with „no”.
Print calib. parameters (Y/N)
It is possible to print the calibration parameters. (see example) With „no” it is possible to go back to the main menu. With „yes” you continue in the parameter setting menu.
Example of a calibration parameter printout :

8.4 Back to main menu

Back to main With „yes“ you return to the Start-Menu with „no” you continue in the Menu? (Y/N)
instrument settings-menu.

8.5 Set time and date

Set time and
Push „no“ to proceed. Push „yes” to choose date format.
date? (Y/N)
MBV AG, CH-8712 Stäfa +41 44 928 30 80 Version: 01 page 17 of 34 Released and approved : Hzi Revision: 06
02.2006 / MAS-100_Usermanual_blank_010606e.doc

8.6 Selecting date format (EU or US) and changing time and date

Date: EUR Modify? (Y/N)
The first item in the Parameter menu is the „Date mode“. With „yes“ you can switch between US and European date format. By pressing „no“ you accept the displayed format.
Time
hh.mm.ss.
Then the time is displayed. Pressing „no“ places a cursor underneath the first digit. Press „no“ until the required number is displayed. When you press „yes“, the cursor moves on to the next digit. Continue until you have set the correct time.
Date
dd.mm.yy
Now the date is displayed. The cursor moves to the first digit. Pressing „no“ changes the numerical value and „yes“ places the cursor below the next digit. You return to the start of the „Set time and date” menu. Clicking „no“ takes you to the next prompt. ( Please note that US format is: mm-dd-yy, EUR format is: dd.mm.yy)

8.7 Battery discharge?

Battery For battery maintenance it is possible to discharge the battery by pressing discharge? (N)
„yes”. Push „no” to proceed to next position.
Batt. discharge: XXXXX litres_
For battery maintenance it is possible to discharge the battery by pressing „yes”. The batteries will discharge while the display shows a decreasing number of liters and the run and stop LED are blinking. If you push „no” during discharge the process of discharge will be interrupted.
Note:
The discharge process may last several hours depending on the actual battery charge. Do recharge the MAS-100™ after discharge for at least 3 hours.

8.8 Anemometer mode

Anemometer This mode is used to check the anemometer function. After a start up mode? (Y/N)
phase.., the anemometer goes to 100 L/min.
Anemometer mode Displays the measured liters. Please wait..x
Anemometer
If the anemometer has a problem Anemometer error is displayed.
error !!!

8.9 Change volumes

Change Each of the 5 pre-set volumes (standard volumes 50, 100, 250, 500, 1000 volumes? (Y/N)
liters) may be customized. A value between 0 and 2000 liters can be specified.
MBV AG, CH-8712 Stäfa +41 44 928 30 80 Version: 01 page 18 of 34 Released and approved : Hzi Revision: 06
02.2006 / MAS-100_Usermanual_blank_010606e.doc
Volume 1 = 50
Modify? (Y/N)
Press „yes“ to increment the displayed value by 1. By holding down „yes“ you can scroll quickly to the required value. Pressing „no“ takes you to the next volume. Setting a volume to 0 will delete it from the screen. For example, you could display just two volumes by setting the other three to 0. They are activated again as soon as you assign them a value between 1 and 2000. Continue through the volumes until you get to the „change volumes?” display. Quit with „no” to go to next menu step.

8.10 Delay menu (Change / Activate / Deactivate )

Change delay? (Y/N)
The MAS-100
has a start menu that enables you, if you wish, to set a delay
time between 1 and 60 minutes before the start of a sampling cycle . The last used delay time will be displayed. By pressing „yes“, you can set a delay time between 0 and 60 minutes. If you enter 0 as delay time, no delay is activated. The MAS-100
starts as soon as you press „yes“ at the „Start?”
prompt.
activated Modify ? (Y/N)
If you never use the delay function, you can suppress the „Delay?” prompt by deactivating the delay in this menu (to exit the program answer „no“ to „deactivate delay menu?”). You can reactivate the delay at any time (to exit the program answer „no“ to „activate delay menu?”).

8.11 Select operation mode

XXXX mode Modify? (Y/N)
Select manual or Template mode? Use this menu to define how to work with the MAS-100
. There are 3 options, as briefly described below.

8.12 Manual mode

Manual mode Modify ? (Y/N)
Use Manual mode when you do not wish to establish a PC connection. For example, when your work does not follow a regular pattern and/or there is no PC available, it is recommended to work in manual mode.

8.13 Selectable mode

Selectable mode Modify? (Y/N)
Selectable mode is used when only sometimes data are sent to the PC and flexibility in the use of the MAS-100
must be guaranteed. In this mode the display asks before each sampling run whether sampling is to be performed manually or at a location designated by the PC. In PC mode the user is prompted to select the report and location before sampling is started.

8.14 Template mode

Template mode
Modify? (Y/N)
The MAS-100 as freeware on (
is equipped for data communication. A software is available
www.mbv.ch).
Note:
A communication cable to transfer data between MAS-100 which comes with user instructions for the MAS-100
MBV AG, CH-8712 Stäfa +41 44 928 30 80 Version: 01 page 19 of 34 Released and approved : Hzi Revision: 06
02.2006 / MAS-100_Usermanual_blank_010606e.doc
separately from your MAS-100 distributor.
and the PC,
software, is available

8.15 Select SQS mode

Select SQS Mode? (Y/N)
The SQS mode allows you to divide a given volume of air into sub volumes, over a given period of time. Select „yes” to activate this function
SQS: total time SQS: total time will be displayed if this mode is active. Modify? (Y/N)
deactivated Deactivated will be displayed if the SQS mode is inactive. Mode? (Y/N)
SQS: cycles: xx Choose between 1 and 20 cycles. The number of cycles will divide the total modify? (Y/N)
volume into corresponding sub volumes
SQS time: h:mm Choose between 1 and 24 hours and minutes. Be sure to choose a total modify? (Y/N)
time in witch the Petri dish does not dry out. Up to 8 hour have been successfully tested by the manufacturer but have to be validated by the customer.
[l] CYCLE TIME After setting the relevant parameters the next two windows are displayed XXXX XX hh:mm
showing the selected volume, the number of cycles and the total sampling period.
XX.XX.XX hh:mm Push„yes” to start sampling in SQS-mode. If you want to change any of the Start? (Y/N)
parameters push „no”. You will be directed to the selection menu.
Note:
To work again in normal sampling mode deactivate the SQS-Cycle in the Parameter 4 Mode.

8.16 User menu

Select user Menu? (Y/N)
If you press „yes” it is possible to activate or deactivate the user menu. If activated, it is possible to allocate 1 out of 10 user names to the relevant air sampling tests.
activated Modify? (Y/N)
The default names are User 1 etc. To define proper user names you need the C&C communication and calibration Software. ( see
10.2)

8.17 Back to main menu

Back to main Menu? (Y/N)
With „yes” the MAS-100
will go to the start menu. With „no” the MAS-100
starts from first menu point of the selected Mode.
MBV AG, CH-8712 Stäfa +41 44 928 30 80 Version: 01 page 20 of 34 Released and approved : Hzi Revision: 06
02.2006 / MAS-100_Usermanual_blank_010606e.doc

9 Sampling head

9.1 Adjustment of Petri dish

Before using the MAS-100 Take off the perforated lid and place a Petri dish on the dish support. Use a 3 mm Allen key (supplied) to adjust the three blue jaws symmetrical, until the Petri dish is securely supported and it is fixed in place. Check that the Petri dish is properly supported by turning the sampling head to the horizontal position. The Petri dish support is now adjusted. This procedure has not to be repeated unless you change the Petri-dish manufacturer.
Note:

9.2 Sampling head 400 x 0,7

The MAS-100
runs with 100 liters/min. The calculated impaction speed is approx. 11 m/sec with the
sampling head with 400 holes with a diameter of 0,7mm.
Note:

9.3 Sterilization of sampling head

The sampling head may be sterilized at 121°C for 20 minutes.
the first time you will need to adjust the blue Petri dish support.
Do not over-tighten the blue jaws. The Petri dish may break and plastic parts could block the unit.
Be sure to use 400 x 0.7 mm Feller conversion table(see
20.1)
10 Software for MAS-100™
10.1 Validated Software for MAS-100
The MAS-100
may be connected to a validated LIMS Software from the Canadian Company Novatek. With this Software it is possible to exchange data between the MAS-100 information you will find over:
http://www.ntint.com/prod-envmon.shtml
10.2 Software download new MAS-100 C&C
The MAS-100 MAS-100
C&C Software is a terminal software which enables the user to communicate with the
in an easy way. The user gets all information about the settings in the MAS-100 but is not authorized to change any calibration parameters. Please find the software on the internet button „Software”. Choose MAS-100 installed automatically on your PC.
Note:
To communicate between MAS-100 communication cable. The cable is part of the MAS-100 C&C communication kit.
MBV AG, CH-8712 Stäfa +41 44 928 30 80 Version: 01 page 21 of 34 Released and approved : Hzi Revision: 06
02.2006 / MAS-100_Usermanual_blank_010606e.doc
and a computer. More
C&C Software Version 2.7x or higher. The software will be
www.mbv.ch,
and a PC you need a special

11 Interpreting results

First the total microbial count is corrected based on the statistical correction table according to Feller (see Appendix 20.1), then it is converted to contaminant particles per cubic meter. The correction table is based on the presumption that, as the number of organisms being impinged on a given plate during sampling rises, the chance that several organisms will enter the same hole in the perforated lid is also increasing.
Pr = N [1/N + 1/N-1+ 1/N-2........1/N-r+1]
Pr = Probable statistical total N = Number of holes in the perforated lid (400)
r = Number of colony-forming units counted on standard Petri dish
For each colony count „r” the table gives a statistically corrected total count „Pr”.
For easy application of this formula, please refer to the table (see Appendix
(A table of statistical corrections according to Feller for the MAS-100
in a plastified cover is provided
20.1)
separately with each sampler).
Note:
The MAS-100 automatically converts them to CFU/m
software corrects the results according to Feller and
3
.
4
12 Charging the MAS-100™

12.1 Battery charging

Connect the battery charger on the handle. „Battery charging” appears on the display and the red light is on. A complete recharging cycle takes about 3.5 hours. With a fully charged battery, the instrument has a sampling capacity of approximately 7.0 hours. .(value for new batteries)

12.2 Battery discharge

See description in chapter

12.3 Battery status display

Low battery! Show capacity?
XXXXX litres left to aspirate
Note:
MBV AG, CH-8712 Stäfa +41 44 928 30 80 Version: 01 page 22 of 34 Released and approved : Hzi Revision: 06
02.2006 / MAS-100_Usermanual_blank_010606e.doc
8.7
If the battery capacity decreases the MAS-100
will display the total amount of litres which can still be sampled without recharging the batteries. The Low battery display appears from a total sampling volume below 10’000 litres. Press yes to see the total amount of volume which is possible to sample. If you press “no” the MAS-100
will start the sampling.
Recharge the batteries as soon as possible (see chapter 12.1).
Use only the MAS-100 charging circuitry of the MAS-100
battery charger to recharge the NiMH battery. The
prevents overcharging of the battery.

12.4 Battery replacement

If the batteries have to be changed, loosen the bottom screw and replace the batteries with an original battery pack from the manufacturer. We advise that this is done by a service engineer who has been instructed by the supplier.
Note:
After changing the battery pack recharge the MAS-100
for at least 3.5
hours.

13 Calibration

The MAS-100 provided in the calibration certificate which is delivered with each instrument.
Note:

13.1 Protected mode

If you use the MAS-100 C&C software it is possible to block the calibration value and/or all settings in the „Parameter Mode 4” see point “Protected mode “ in the Instruction-manual of the software.
Note:

13.2 Calibration reminder

The MAS-100 month is selected the calibration reminder is deactivated. If 6 months is set, the reminder will add 6 months to the actual date in the MAS-100 show the following message:
Calibration time expired! Please recalibrate unit. When pressing down the „yes“ or „no“ button, the display will show the date of the last calibration. After releasing the button the menu proceeds to the start line.
Each time the MAS-100 and the reminder is set to the corresponding value.
is calibrated in the factory for a mass flow of 100 liters/min. Detailed information are
Only authorized person may exchange any mechanical parts in the instrument or change any of the software settings in the „service” menu.
To communicate between MAS-100
and a PC you need a special communication cable. The cable is part of the MAS-100C&C communication kit.
has an integrated calibration reminder. It can be set between 0 and 12 months. If 0
. If after 6 months the instrument is turned on, the display will
is turned on, this message will be displayed until the instrument is recalibrated
MBV AG, CH-8712 Stäfa +41 44 928 30 80 Version: 01 page 23 of 34 Released and approved : Hzi Revision: 06
02.2006 / MAS-100_Usermanual_blank_010606e.doc

14 Technical specifications

General:
Flow rate: 100 Liters 100 liters / min. ± 2.5%
Standard sampling volumes: 50, 100, 250, 500, 1000 liter
Freely definable sampling volumes: 1 to 2000 liters ,volumes individually selectable
Sampling head :
Diameter: 110 mm
Weight: 190 g
Material: Anodized aluminum
Autoclavable: 20 minutes at 121°C
MAS-100
with handle
Height: 260 mm
Width incl. handle: 180 mm
Weight: 2.2 kg
Case for storage plastic
Material: Anodized aluminum
Additional information:
Tripod screw: For use with normal camera tripod
Battery pack: 9 cells NiMH, 10.8 V
Motor: 6 V
Battery charger input: 100-240 Volt, 47-63 Hz
Charger output: 24 V DC, 1.5 A, 36 W
Display: Alphanumeric liquid crystal display, 32 characters
Lifetime RTC Battery: Approx. 10 years
Driving motor: PWM frequency for driving motor
Processor: Type : 80C552
CE approval: EN 55022 Class B, EN 61000-4-2, ENV 50140,

14.1 Environemental conditions

Altitude: Up to 2000 m or above if specified by manufacturer Temperature: 0 to 40°C Maximum relative humidity: 80 percent for temperatures up to 31°C,
Mains supply-voltage: Fluctuations must not exceed ± 10 percent of the
Transient over-voltage: Acc. to INSTALLATION CATEGORIES
Technical modification possible without prior notice.
MBV AG, CH-8712 Stäfa +41 44 928 30 80 Version: 01 page 24 of 34 Released and approved : Hzi Revision: 06
02.2006 / MAS-100_Usermanual_blank_010606e.doc
between 0 and 2000 liters.
ENV 50204, EN 61000-4-4, ENV 50141
decreasing linearly to 50 percent relative humidity at 40°C;
nominal voltage; other supply voltage: fluctuations as stated by the manufacturer
(OVERVOLTAGE CATEGORIES) I,II, and III and normal category is II;POLLUTION DEGREE 1 or 2 in accordance with IEC 664

15 Components check list

The MAS-100
1 MAS-100 1 battery charger 1 allen key 3 mm (for centering the Petri dish) 1 mains cable 1 plastified „Table of statistical corrections according to Feller” 1 calibration certificate 1 calibration certificate for DA-100 1 MAS-100 Operator’s Manual CD
is delivered in an plastic case containing the following:
, complete system with sampling head (400 x 0.7)

16 Care and maintenance

The MAS-100 The sampling head is autoclavable (20min/121°C). When the MAS-100 area, we suggest to wipe it with a suitable disinfectant. Between sampling cycles, the lid may also be cleaned with disinfectant. Make sure that the holes in the perforated lid are not clogged.
should be periodically calibrated (see 13).
is moved to a sterile or „clean”

17 Warranty

With the purchase of the MAS-100 malfunction of electrical and electronic components (see invoice date).
The MAS-100
is a laboratory instrument and should only be used by qualified personnel.
Note:
The MAS-100 is calibrated and a screw in the sampling head is sealed. If the seal is removed by a unauthorized person the warranty will decay.
you will receive a 1-year warranty against any failure resulting from
MBV AG, CH-8712 Stäfa +41 44 928 30 80 Version: 01 page 25 of 34 Released and approved : Hzi Revision: 06
02.2006 / MAS-100_Usermanual_blank_010606e.doc

18 Accessories and spare parts

Accessories:
Calibration unit DA-100 1.09228 PC-cable / C&C Software package 1.09226 Quick adapter (for tripod) 1.09223 Tube adapter 1.09224 Lightweight Shoulder bag, blue 1.09126 MAS-100
plastic case
Spare parts:
Perforated lid (400x0.7) 1.09088 Dustcover 1.09084 Mains power supply unit 1.09085 Battery pack NiMH 1.09229 Perforated lid (400x0.7) for “contact” plates 1.09213 Adapter for “contact” plates 1.09214
1.09225

19 References

[1] The Rules Governing Medicinal Products In The European Community Vol. IV: Guide to Good Manufacturing practice for medicinal products, 1992
[2] MAS-100, Qualification of air sampler systems: The MAS-100 Meier R. und Zingre H., (Swiss Pharma 1-2/00)
[3] New Sampler for the Collection, Sizing, and Enumeration of Viable Airborne Particles, A. Andersen, U.S. Army Chemical Corps Proving Groung, Dugway, Utha., 1958
[4] Feller, W. An introduction to the probability theory and its application, p. 175. John Wiley and sons, Inc., New York, 1950
[5] SQS-Sequential air sampling; Validation of the microbial count in the air using the MAS-100;
Swiss Pharama 26 (2004) No.11,9-11

20 Appendix

20.1 Table of statistical corrections according to Feller for SH 400 x 0.7mm
20.2 Firmware Flow Diagram MAS-100: Manual Mode 1
20.3 Firmware Flow Diagram MAS-100: Template Mode 2
20.4 Firmware Flow Diagram MAS-100: Selectable Mode 3
20.5 Firmware Flow Diagram MAS-100: Parameters Mode 4
20.6 Firmware Flow Diagram MAS-100: Calibration reminder Mode 5
20.7 CE Conformity declaration
MBV AG, CH-8712 Stäfa +41 44 928 30 80 Version: 01 page 26 of 34 Released and approved : Hzi Revision: 06
02.2006 / MAS-100_Usermanual_blank_010606e.doc

20.1 Table of statistical corrections according to Feller for SH 400 x 0.7mm

r = Number of colony forming units counted on standard Petri dish Pr = Probable statistical total
r Pr r Pr r Pr r Pr r Pr r Pr r Pr r Pr
1 1 51 54 101 116 151 189 201 279 251 394 301 557 351 836 2 2 52 56 102 118 152 191 202 281 252 397 302 561 352 844 3 3 53 57 103 119 153 193 203 283 253 400 303 565 353 853 4 4 54 58 104 120 154 194 204 285 254 402 304 569 354 861 5 5 55 59 105 122 155 196 205 287 255 405 305 573 355 870 6 6 56 60 106 123 156 197 206 289 256 408 306 578 356 879 7 7 57 61 107 124 157 199 207 291 257 411 307 582 357 888 8 8 58 63 108 126 158 201 208 293 258 413 308 586 358 897
9 9 59 64 109 127 159 202 209 295 259 416 309 591 359 907 10 10 60 65 110 128 160 204 210 297 260 419 310 595 360 917 11 11 61 66 111 130 161 206 211 299 261 422 311 599 361 927 12 12 62 67 112 131 162 207 212 301 262 425 312 604 362 937 13 13 63 68 113 133 163 209 213 304 263 428 313 608 363 947 14 14 64 70 114 134 164 211 214 306 264 431 314 613 364 958 15 15 65 71 115 135 165 212 215 308 265 433 315 618 365 969 16 16 66 72 116 137 166 214 216 310 266 436 316 622 366 981 17 17 67 73 117 138 167 216 217 312 267 439 317 627 367 992 18 18 68 74 118 140 168 218 218 314 268 442 318 632 368 1005 19 19 69 76 119 141 169 219 219 317 269 445 319 637 369 1017 20 20 70 77 120 142 170 221 220 319 270 449 320 642 370 1030 21 22 71 78 121 144 171 223 221 321 271 452 321 647 371 1043 22 23 72 79 122 145 172 224 222 323 272 455 322 652 372 1057 23 24 73 80 123 147 173 226 223 325 273 458 323 657 373 1071 24 25 74 82 124 148 174 228 224 328 274 461 324 662 374 1086 25 26 75 83 125 150 175 230 225 330 275 464 325 667 375 1102 26 27 76 84 126 151 176 232 226 332 276 467 326 673 376 1118 27 28 77 85 127 153 177 233 227 335 277 471 327 678 377 1134 28 29 78 87 128 154 178 235 228 337 278 474 328 684 378 1152 29 30 79 88 129 156 179 237 229 339 279 477 329 689 379 1170 30 31 80 89 130 157 180 239 230 342 280 480 330 695 380 1189 31 32 81 90 131 158 181 241 231 344 281 484 331 701 381 1209 32 33 82 92 132 160 182 242 232 346 282 487 332 706 382 1230 33 34 83 93 133 161 183 244 233 349 283 491 333 712 383 1252 34 35 84 94 134 163 184 246 234 351 284 494 334 718 384 1276 35 37 85 95 135 164 185 248 235 353 285 497 335 724 385 1301 36 38 86 97 136 166 186 250 236 356 286 501 336 730 386 1327 37 39 87 98 137 167 187 252 237 358 287 504 337 737 387 1356 38 40 88 99 138 169 188 254 238 361 288 508 338 743 388 1387 39 41 89 101 139 171 189 255 239 363 289 511 339 749 389 1420 40 42 90 102 140 172 190 257 240 366 290 515 340 756 390 1456 41 43 91 103 141 174 191 259 241 368 291 519 341 763 391 1496 42 44 92 104 142 175 192 261 242 371 292 522 342 769 392 1541 43 45 93 106 143 177 193 263 243 373 293 526 343 776 393 1591 44 47 94 107 144 178 194 265 244 376 294 530 344 783 394 1648 45 48 95 108 145 180 195 267 245 378 295 534 345 791 395 1715 46 49 96 110 146 181 196 269 246 381 296 537 346 798 396 1795 47 50 97 111 147 183 197 271 247 384 297 541 347 805 397 1895 48 51 98 112 148 185 198 273 248 386 298 545 348 813 398 2028 49 52 99 114 149 186 199 275 249 389 299 549 349 820 399 2228 50 53 100 115 150 188 200 277 250 391 300 553 350 828 400 2628
MBV AG, CH-8712 Stäfa +41 44 928 30 80 Version: 01 page 27 of 34 Released and approved : Hzi Revision: 06
02.2006 / MAS-100_Usermanual_blank_010606e.doc
20.2 Firmware Flow Diagram MAS-100: Manual Mode 1
Manual Mode 1
yes no
MAS-100 V 5.XX S/N: XXXXX
calibration valid xx month (days)
Calibration time
expired !
Please recalibrate
unit
yes no
Last calibration
XX.XX.XX
XXXXX litres left to aspirate
Volume: XXXX Y/N?
yes no
Select other volumes Y/N ?
yes no
software version and
serial number
if calibration time is not
expired it displays the time
until calibration
If calibration reminder is activated and calibration
time is expired
last volume selected
continues to next active menu / user or delay or
start
XX.XX.XX XX:XX
Start?
yes no
Volume [l]: XXXX
(Delay:)Time : XX:XX
yes or no
failed!
repeat test !
Delay: X:XX Modify? (Y/N)
Volume: XXXX Y/N?
yes no
passed
test completed
If delay time is activated
Delay time will be
displayed first
after run, back to start
during run forward to
failed! repeat test!
If delay time is activated
max. delay 60 min.
after 3 sec.display
changes to Volume: Time:
Press yes to start next
sample.
After 30 Min automatic
shut off
XXXX
YES=inc. litres
yes no
Select volume:
XXXX modify Y/N?
yes no
User 1 Change user? Y/N
yes no
Delay: X:XX Modify? (Y/N)
yes no
defined volumes are
displayed
If delay time is activated
max. delay 60 min.
MAS100_Mode1_01.vsd
MBV AG, CH-8712 Stäfa +41 44 928 30 80 Version: 01 page 28 of 34 Released and approved : Hzi Revision: 06
02.2006 / MAS-100_Usermanual_blank_010606e.doc
20.3 Firmware Flow Diagram MAS-100: Template Mode 2
Template Mode 2
yes no
MAS-100 V 5.XX S/N: XXXXX
calibration valid xx month (days)
Calibration time
expired !
Please recalibrate
unit
yes no
Last calibration
XX.XX.XX
XXXXX litres left to aspirate
Mr. XXXXXX Change user? Y/N
yes no
Protocol XX..XX Modify? (Y/N)
yes no
Room XX Location XX
yes no
Delay: X:XX Modify? (Y/N)
software version and
serial number
if calibration time is not
expired it displays the time
until calibration
If calibration reminder is
activated and calibration
time is expired
no will change to the next
possible sampling place
If delay time is activated
max. delay 60 min.
XX.XX.XX XX:XX
Start?
yes no
Volume [l]: XXXX
(Delay:)Time : XX:XX
yes no during run
yes no after run
Room XX Location XX
Mr. XXXXXX Change user? Y/N
failed!
repeat test !
passed
test completed
All protocol positions done!
yes starts the sampling
If delay time is activated
Delay time will be
after 3 sec.display
changes to Volume: Time:
Press yes to start next
After 30 Min automatic
If all positions in a
protocoll are sampled the
All protocol positions
The next protocol will be
The protocols must be
cleared in the Mode 5 in
order to use the same
displayed first
sample.
shut off
display shows:
done!
displayed.
protocol again
yes no
MAS100_Mode2_01.vsd
MBV AG, CH-8712 Stäfa +41 44 928 30 80 Version: 01 page 29 of 34 Released and approved : Hzi Revision: 06
02.2006 / MAS-100_Usermanual_blank_010606e.doc
20.4 Firmware Flow Diagram MAS-100: Selectable Mode 3
Selectable Mode 3
yes no
MAS-100 V 5.XX S/N: XXXXX
Calibration valid xx month (days)
Calibration time
expired !
Please recalibrate
unit
yes no
Last calibration
XX.XX.XX
XXXXX litres left to aspirate
Volume: XXXX Y/N?
yes no yes no
If Manual mode was
selected
continue in Mode 1
User XX Change user? Y/N
software version and
serial number
If calibration time is not
expired it displays the time
until calibration
If calibration reminder is activated and calibration
time is expired
If Template Mode 2 was
selected
Protocol XX..XX Modify? Y/N?
yes no
Room XX Location XX
shows the next position in
the protocol which has not
been sampled
MAS100_Mode3_01.vsd
If all positions in all protocolls are done the
display shows
All protocolls done!
The values in protocols must be cleared in
Mode 4
Clear protocol data? (Y/N)
XX.XX.XX XX:XX
Start? (Y/N)
yes no
Template mode
modify? (Y/N)
yes no
To change to Manual or
Template mode go to
Start display and push no
To change to Manual or
Template mode push no
MBV AG, CH-8712 Stäfa +41 44 928 30 80 Version: 01 page 30 of 34 Released and approved : Hzi Revision: 06
02.2006 / MAS-100_Usermanual_blank_010606e.doc
20.5 Firmware Flow Diagram MAS-100: Parameters Mode 4
Parameters Mode 4
XXXXXXXXX XXXXXXXXX
yes no
Clear protocol data? (Y/N)
yes no
Clr. man. prot. Modify ? (Y/N)
yes no
clear manual mode protocols
yes no
Print protocol ? (Y/N)
yes no
prints data if printer is connected
Change online print mode? Y/N
yes no
Print after XX samplings Y/N
push no and yes button for 5 seconds and release after this
time
if Template mode is
active "no" clears the
selected template
if Template mode is
active no clears
selected template
Online print
deactivated is
displayed if no number
is selected
Back to main menu?
(Y/N)
yes no
Set time and date ? (Y/N)
yes no
Date mode:EUR/USA Modify? (Y/N)
yes no
Time
xx:xx:xx
yes no
Date
xx.xx.xx
yes no
Battery discharge? (Y/N)
yes no
Anemometer mode? (Y/(N)
yes no
yes goes back to start
of selected Mode
push no first. Then
change the value above the cursor with no. Accept new value
with yes
push no first. Then
change the value above the cursor with no. Accept new value
with yes
yes discharge
batteries LED are
blinking
if yes is pushed
MAS-100 runs at
100 L/min for
anemometer test
yes no
Print after XX samplings Y/N
yes no
Print calib. parameters (Y/N)
yes no
prints data if printer is
connected
MAS100_Mode4_01.vsd
Change volumes? (Y/N)
yes no
Volume 1 - 5 = XXXX Modiy ? (Y/N)
yes no
Change delay? (Y/N)
yes no
1-5 Volume may be
modified from 0 to
2000 Litres, if 0 volume is selected it will not be displayed
with yes activate or
deactivate delay
function
continue next page
MBV AG, CH-8712 Stäfa +41 44 928 30 80 Version: 01 page 31 of 34 Released and approved : Hzi Revision: 06
02.2006 / MAS-100_Usermanual_blank_010606e.doc
Continue: Firmware Flow Diagram MAS-100
: Parameters Mode 4
Select operation mode? (Y/N)
yes no
Select SQS mode? (Y/N)
yes no
deactivated Modify? (Y/N)
SQS: total time Modify? (Y/N)
yes no
Select users menu (Y/N)
yes no
Choose Manual
Selectableor Template
Mode
total time over the
SQS-cycle is active
User menu: on/off Modify (Y/N)
yes no
Back to main menu?
(Y/N)
yes no
goes back to start of
selected mode
start to first menu line
clear protocol data?
MAS100_Mode4_01.vsd
MBV AG, CH-8712 Stäfa +41 44 928 30 80 Version: 01 page 32 of 34 Released and approved : Hzi Revision: 06
02.2006 / MAS-100_Usermanual_blank_010606e.doc
20.6 Firmware Flow Diagram MAS-100: Calibration reminder Mode 5
Calibration reminder Mode 5
Initial activation
yes no
MAS100 V5.XX S/N: XXXXX
Set calibration reminder Y/N
calibration reminder;
only the first time in
use
yes no
Set reminder to 12 month? Y/N
yes no
Are you sure to set xx month Y/N
yes no
Calibration valid: XX month
if you push no the
MAS will go off
with no change month from 0-12 . "0" means
no reminder is active.
12 month is the
default setting
if you push no the
MAS will go off
program continues in
normal running mode
MAS100_Mode5_01_Cal_Rem.vsd
MBV AG, CH-8712 Stäfa +41 44 928 30 80 Version: 01 page 33 of 34 Released and approved : Hzi Revision: 06
02.2006 / MAS-100_Usermanual_blank_010606e.doc

20.7 CE Conformity declaration

CE Konformitätserklärung CE Conformity Declaration
MBV AG Tel.: +41 44 928 30 80 Laubisrütistr. 24 Fax: +41 44 928 30 89 CH-8712 Stäfa www.mbv.ch
Das bezeichnete Gerät entspricht der einschlägigen grundlegenden Anforderungen der aufgeführten EG-Richtlinien und Normen. Bei Änderungen am Gerät verliert diese Erklärung ihre Gültigkeit.
The device named below fulfills the relevant fundamental requirements of the EC directives and standards listed. In the case of unauthorized modification to the device, this declaration becomes invalid.
Gerätebezeichnung Gerätetyp
MAS-100 Mikrobiologischer Luftkeimsammler
Device name: Device type
MAS-100 Microbiological Air Sampler
EMV: EG-Richtlinien/Normen Relevant directives/standards:
EN 55022 Class B, EN 61000-4-2, ENV 50140, ENV 50204, EN 61000-4-4, ENV 50141
MBV AG
H. Zingre A. Barmettler
Director Director
Staefa, 4. Februar 2003
MBV AG, CH-8712 Stäfa +41 44 928 30 80 Version: 01 page 34 of 34 Released and approved : Hzi Revision: 06
02.2006 / MAS-100_Usermanual_blank_010606e.doc