mbnet Engineering GmbH has a network of customer service, sales and advisory centres.
Contact your nearest mbnet Engineering GmbH subsidiary to obtain the address of your
local distributor.
You can find a full list of all subsidiaries and distributors on our website: www.mbnet.de
Sales information can also be obtained from: info@mbnet.de
Ergo vac bears the mark (IEC 60601-1, Class I, Type BF without defibrillation protection), indicating its compliance
with the essential requirements regarding safety, functionality and labelling of Annex I of the Medical Device Directive
93/42/EEC. The requirements apply to patients, users and third persons who come into contact with this device within
the scope of its intended use.
Date of issue: 30.05.2019
mbnet Engineering GmbH
Kirschauer Straße 37a
D-02681 Callenberg
Telephone +49 (0)3592 34 83 0 u. 54 25 47
Fax +49 (0)3592 34 34 4 u. 54 25 49
E-Mail info@mbnet.de
Internet www.mbnet.de
The device must be used only by qualified physicians or trained medical professionals.
The responsibilities of the staff for operating and maintaining the device must be speci-
fied by the operator.
Ensure that the staff have read and understood the user guide. This applies in particular
to this section Safety notes.
The device must not be stacked at any moment.
Damaged or missing parts must be replaced immediately.
The safety, reliability and performance of the device can only be guaranteed when the
maintenance intervals as stated in Chapter 5: “Maintenance and Care” are observed.
Warning
Do not modify this equipment without authorization of the manufacturer.
1.2 Organisational Measures
Before using the device, ensure that a medical product representative has
explained its functions as well as the safety requirements.
Keep this user guide in an accessible place for reference purposes.
Make sure that it is always complete and legible.
Observe the operating and maintenance instructions.
1.3 Indications for Use
The device is an ECG vacuum and is operated in combination with normal ECG
devices. The device is suitable for both recording resting as well as exercise ECG and
is used for patients of both genders as well as all ancestries and age groups (preferably
as of the age of seven, also dependent on body size).
The device is only to be operated in a professional healthcare environment.
The device is suitable for use inside hospitals, cardiology centres,
outpatient clinics and medical practices.
The device can safely be used with pacemaker patients.
Always operate the device in line with the technical data indicated.
The device is not intended for sterile use or use outdoors.
This is a device of type BF. It is not defibrillation protected. As a safety precaution,
remove the electrodes before defibrillation!
1.4 Contra-indication
The device is not intended for sterile use.
The device must not be used in potentially explosive areas or in the presence of flam-
mable gases such as anaesthetic agents.
The device is not for direct cardiac application.
The device is not for use in an MRI suite.
Seite 4
1.5 Safety-conscious Operation
Make sure that the staff have read and understood the operating instructions, in par-
ticular this section Safety Notes.
Do not touch the housing of the device during defibrillation.
To ensure patient safety, none of the electrodes, including the neutral electrode, nor the
patient or any person with simultaneous patient contact, must come in contact with
conductive parts, even when these are earthed.
Immediately report any changes that impair safety (including operating behaviour) to
the responsible person.
Only use accessories and disposables recommended or supplied by mbnet
Engineering GmbH. The use of accessories or disposables from other manufacturers
may result in injury, inaccurate information and/or damage to the unit.
User guide
ergo vac
1.6 Safe Use with Electronics
Operating the device without the correctly rated fuse or with defective cables
constitutes a danger to the life and limb of the patient or the operator!
Therefore take note of the following:
diately.
original.
The device must not be used if the power cable is damaged or
suspected of being damaged.
Damaged cable connections and connectors must be replaced imme-
Electrical safety devices, such as fuses, must not be modified.
Fuses must only be replaced with the same type and rating as the
1.7 Operation with other Devices
If the device is part of a medical system, only the original suction hoses from mbnet
Engineering GmbH must be connected to the device.
Portable communication devices, HF radios and devices labelled with the symbol:
(non-ionic electromagnetic radiation) can affect the operation of this device.
1.8 Maintenance
Danger of electric shock - do not open the device! It contains no parts, which can be
repaired by the user. Servicing must only be performed by qualified technicians
authorised by mbnet Engineering GmbH.
Switch off the device before cleaning and disconnect it from the mains.
Do not use high-temperature sterilisation processes (such as autoclaving). Do not use
e-beam or gamma radiation sterilisation.
Do not use aggressive or abrasive cleaners.
Do not, under any circumstances, immerse the device or cable assemblies in cleaning
liquid.
Seite 5
ergo vac
User Guide
1.9 Terms of Warranty
Your device is warranted against defects in material and manufacture, as stated in the
Terms and Conditions. Excluded from this warranty is damage resulting from negligence
or improper use. The warranty covers the free replacement of the defective part. Any liability for subsequent damage is excluded. The warranty is void if unauthorised or unqualified
persons attempt to make repairs.
In case the device is defective, send it to your local mbnet Engineering GmbH representative or directly to the manufacturer. The manufacturer can only be held responsible for the
safety, reliability and performance of the apparatus if:
assembly operations, extensions, readjustments or repairs are carried out by persons
authorized by the manufacturer, and
the device and approved attached equipment is used in accordance
with the manufacturer's instructions, and
the maintenance intervals as stated in Chapter 5: “Maintenance and Care” have been
complied with.
No further guarantees are assumed. mbnet Engineering GmbH makes no warranty of
merchantability or fitness for a particular purpose with respect to the product or parts
thereof.
1.10 Symbols and Pictograms
1.10.1 Symbols Used in this Document
The safety level is classified according to ISO 3864-2. The following overview shows the
safety symbols and pictograms used in this user guide.
For general safety notes as listed in this section.
For electrical hazards, warnings or precautionary measures when dealing with
electricity.
For possibly dangerous situations which could lead to damage to property or system
failure. Important or helpful user information.
Reference to other instructions.
DANGER
Warning
For a possibly dangerous situation which could lead to severe personal injury or to
death.
For a direct danger which could lead to severe personal injury or to death.
Seite 6
Caution
For a possibly dangerous situation which could lead to slight personal injuries. This symbol
is also used to indicate possible damage to property.
1.10.2 Symbols Used on the Device
BF-symbol, no protection against defibrillation current.
Dispose of as electronic waste.
Attention: consult accompanying documents.
Manufacturer
User guide
ergo vac
Date of manufacture
CE label
Follow the manufacturer’s instructions
Seite 7
ergo vac
User Guide
2 Introduction
The ergo vac is an ECG suction device for recording heart potentials during resting and
stress ECGs and sending the data to the ECG device. The built-in control panel allows for
easy operation and efficient configuration of the device.
2.1 Elements of the Suction Device
2.1.1 Overview
suspension for suction leads
suction electrode
leads with colour
code
height-adjustable
telescopic arm
internal cable
feed
fixation pump module
Seite 8
pump module:
small, quiet, powerful
suction
Trolley optional
(standard version w/o trolley)
2.1.2 Scope of Delivery
Standard Model
Resting and stress ECGs
Mains operation
Options
Battery operation
Trolley (with grounding cable)
Fixation pump module on trolley
2.2 Control Box with Control Panel
What makes the control box stand out is its optimal user ergonomics. It consists of a
control panel, control electronics and defibrillation protection. The control panel features
white and green backlighting as well as push buttons. The control panel is easy to operate
and to clean.
User guide
ergo vac
ON / OFF
Stop ECG
Clean
2.3 Suction Electrode Leads
The ten shielded, interference-free electrode lines are trouble-free and stand out by virtue of
their low abrasion and high flexibility.
Control
Box
Suction Power
Caution
Please pay attention on the careful handling of suction leads (see 7.2.2 page 23)
Seite 9
ergo vac
User Guide
2.4 Cable Arm
The special feature of the cable arm is its hidden cable routing as well as its movable and
horizontal telescopic arm.
Control
height-adjustable
telescopic arm
Joint
2.5 Pump Module
Power
Power Switch
Fuse
Caution
The unit may only be connected to the following devices:
ECG devices that meet IEC 60601-1 standards.
ECG monitors that meet IEC 60601-1 standards.
The unit must not be connected to Class B ECG devices.
Any connection to unauthorized hardware is at your own risk.
It may also void the warranty.
pump module
Seite 10
2.6 Fixation module pump
Bracket for pump module ergo vac with Trolley vexio-cart.
Comment: We have a large selection of brackets (bracket can be
created according to customer requirements), incl. table and wall
mounts
2.7 Serial Number
fixation
pump module
User guide
ergo vac
cable arm
pump module
pump module
2.8 Supplied accessories for ergo vac (mains) or (battery)
Cable arm with patient module
Electrode suction hose (6 x 1.30 m / 4 x 1.50 m)
Spacers (2 pcs with 3 rows / 2 pcs with 2 rows)
Pump unit, large
Power cable for mains connection
ECG Spray 250 ml
Standard bracket
Screw set consisting of:
2 x cable ties 200mm
1 x D-sub connection kit
1 x cable clip, small
1 x 5 mm and 1 x 2.5 mm Allen key
4 x each of M6x12 / M6x16 / M6x20 Allen screw
4 x each of M6 nut / M6 self-locking / M6 square nut
4 x 6.3 washer / 4 x 6.37 spring washer / 4 x rubber feet
Operation manual
with accumulator
Seite 11
ergo vac
User Guide
3 Operation
3.1 Getting Started
HAZARD
Electric shock hazard. The device must not be used if it is not properly grounded or
the power cable is damaged or suspected to be damaged.
Location
The device must not be stored or operated in a wet, humid or dusty area.
It must also not be exposed to direct sunlight or heat from other sources.
The device must not come into contact with acids or acidic fumes.
The device should not be placed in the vicinity of X-ray, hf surgical equipment,
3.2 On / Off
The unit is switched on and off with the
the electrode can be sucked to the skin of the patient by applying light pressure with
the fingers.
diathermy units, large transformers or electric motors.
button. When pressing the ON button,
Warning
ON / OFF
Turn off the unit when not in use for a longer time.
Never apply suction when the electrodes are in a cleaning fluid.
When pressing the
5 seconds and then turns off automatically.
button, the suction pump creates a vacuum for about
Seite 12
3.3 Power Supply
3.3.1 Displays for Mains or Battery Power
The unit can be connected to the mains or be supplied from the built-in battery (option).
In both modes the
Battery Life (Option):
continuous use.
Battery being charged:
(when power is off). If the device is switched on during charging, the
charging time may be longer.
The battery is charged when the device is connected to the mains. The
device can remain connected to the power grid without causing damage
to the device or battery.
The device can remain connected to the power grid without risk for
the battery.
Recharging Times:
not in use
button on the control panel will light up.
The internal battery provides power for up to two hours of
A completely depleted battery takes about 3.5 hours to fully charge
Battery recharging time if depleted: about 3.5 hours if the unit is
Charging time when in use: about 10 hours
User guide
ergo vac
3.3.2 Isolation from Power Grid
To isolate the device and the power supply from the mains, pull the power plug.
4 Controls
The controls are located on the front of the control box and control the entire
suction system.
clean / blow out air
suction levels
regulate suction power
Seite 13