1.1 Important instructions for safe usage ....................................................................................................................................... 5
1.2 Principal installation instructions .............................................................................................................................................. 5
1.3 Principal operating instructions ................................................................................................................................................ 5
1.4 Responsibilities of the user ....................................................................................................................................................... 6
1.4.1 Only allow installation and use by specialist technicians .......................................................................................................... 6
1.4.2 Correct use ................................................................................................................................................................................ 6
1.4.3 Consideration of local conditions .............................................................................................................................................. 6
1.4.4 Safekeeping of the documentation ........................................................................................................................................... 6
2.1.3 Certification for explosion-endangered areas ........................................................................................................................... 7
2.4 Scope of delivery ..................................................................................................................................................................... 10
3Installation of the Measuring Probe ........................................................................................................................................................... 11
3.1 Selecting the installation location for the measuring probe ................................................................................................... 11
3.1.1 Requirements for the measuring liquid at the installation location ........................................................................................ 11
3.1.2 Requirements for the installation location.............................................................................................................................. 11
3.2 Preparing the piping system .................................................................................................................................................... 12
3.3 Installation of the measuring probe with immersion pipe ...................................................................................................... 12
4Installation of the Display Unit .................................................................................................................................................................... 14
4.1 Mounting the display unit ....................................................................................................................................................... 14
4.1.1 Criteria for the mounting location ........................................................................................................................................... 14
4.1.2 Securing the display unit without a mounting bracket ........................................................................................................... 14
4.1.3 Securing the display unit with a mounting bracket ................................................................................................................. 14
4.2 Connecting the potential equalisation .................................................................................................................................... 15
4.3 Opening the housing ............................................................................................................................................................... 15
4.4 Installing the cables ................................................................................................................................................................. 16
4.4.1 General information about installing the cables ..................................................................................................................... 16
4.4.2 Connecting the probe cable .................................................................................................................................................... 16
4.4.3 Connecting the measurement reading outputs ...................................................................................................................... 16
4.4.4 Connecting the supply voltage ................................................................................................................................................ 17
4.4.5 Closing the cable glands and housing ...................................................................................................................................... 17
5.1 Putting the MLA1000 into operation ...................................................................................................................................... 18
5.2 Checking the operating status ................................................................................................................................................. 18
5.2.1 Indications for a safe operating status .................................................................................................................................... 18
5.2.2 Indications for an unsafe operating status .............................................................................................................................. 18
5.3.1 Screen after switching on ........................................................................................................................................................ 18
6.1 Removing the immersion pipe (with measuring probe) .......................................................................................................... 21
6.2 Testing the measuring function ............................................................................................................................................... 21
6.3 Cleaning the measuring probe ................................................................................................................................................ 22
6.4 Reinstalling the measuring probe............................................................................................................................................ 22
7.1 If the MLA1000 does not function at all ... .............................................................................................................................. 23
7.2 If the measurement readings are obviously incorrect... ......................................................................................................... 23
8.1 Technical data ......................................................................................................................................................................... 24
8.3 Disposal Information ............................................................................................................................................................... 25
1.1 Important instructions for safe usage
Only use the MLA1000 in explosion-endangered areas if permitted by the specifications for the
zone, explosion group and temperature class (refer to the type plate on the display unit and
measuring probe).
Pay attention to and heed the "special conditions" stated on the type examination certificate.
Installation, commissioning, maintenance and testing may only be carried out by technicians who
possess requisite knowledge of the rules and regulations for explosion-endangered areas, e.g.
zoning
types of ignition protection
installation regulations, e.g. "Ordinance on electrical systems in explosion-endangered
locations (ElexV)"
Do not alter the product. Repairs may only be carried out by the manufacturer or by trained and
authorised technicians.
If the product was delivered with special information about it: Pay preferential attention to any
special information.
1.2 Principal installation instructions
IMPORTANT:
► Install the measuring probe so that the liquid flows past the probe in the required
direction (→ p.14, Fig. 3).
► Only use measuring probes and probe cables that have been approved by the
manufacturer for the intended measuring liquid.
► Only put the MLA1000 into operation when the entire housing and all the cable
glands have been correctly sealed.
1.3 Principal operating instructions
Intended usage → p.7, section 2.1.
Do not use the measurement readings if the requirements for the installation location
of the measuring probe have not been fulfilled, in particular if the flow velocity of the
measuring liquid is not within the specified range (→ p.7, section 2.1.1).
IMPORTANT:
►In explosion-endangered areas: Only open the terminal box for the display unit
(bottom part of the housing) after the supply voltage and all external power
supplies have been switched off.
►If the measuring probe could become soiled during operation: Clean the probe at
regular intervals (→ p.22, section 6.3).
►If the measurement readings cannot be correct: Check whether the measuring
1.4.1 Only allow installation and use by specialist technicians
►Make sure that the MLA1000 is only installed and used by specialist technicians who, on account
of their specialist training and knowledge and their familiarity with the applicable regulations, are
able to assess the work assigned to them and recognise any hazards.
This applies in particular to:
– safety in explosion-endangered areas
– welding work
– handling of combustible substances
1.4.2 Correct use
►Only use the product as described in these operating instructions. The manufacturer will bear no
responsibility for any deviating usage.
► Carry out the described maintenance work.
► Do not carry out any work or repairs on the product which are not described in these operating
instructions.
►Do not remove, add or alter any components belonging to the product unless stipulated in official
information provided by the manufacturer. Otherwise,
– the manufacturer's warranty will become void,
– the product could become dangerous,
– the certification for use in explosion-endangered areas will become void.
1.4.3 Consideration of local conditions
►All local legislation, technical rules and operating regulations that apply at the location where the
product is to be used must be observed in addition to these operating instructions.
1.4.4 Safekeeping of the documentation
►These operating instructions and all the supplied documentation must be kept available for
► If the product was delivered with special information about it:
► Pay preferential attention to any special information.
2.1.1 Intended scope of application
Measuring function
The MLA1000 is designed for continuously measuring the electrical conductance and
temperature of suitable flowing measuring liquids. Suitable measuring liquids: kerosene, light
fuels, lubricants, coolants, various chemical liquids.
It can only be guaranteed that the MLA1000 will function as intended when used in
flowing light fuel at a flow velocity of at least 0.2 m/s.
Application purpose
Liquids can become electrostatically charged during filling, pumping or filtration. In the
presence of an ignitable gas mixture in the ambient environment, there is also a risk that the
gas mixture will be ignited (i.e. explode) by discharge sparks. This risk is assessed by
measuring the electrical conductance of the liquid (see DIN 51 412-T02-79 "Determining
electrical conductance, field procedure").
The conductance is stated using the physical unit "pS/m" (picosiemens per meter).
The unit "c.u." (conductivity unit), which is still frequently used in the oil industry, has
an equal value:
1 pS ⁄ m = 1 c.u.
2.1.2 Application restrictions
► Only use the MLA1000 for measuring suitable measuring liquids.
► The measuring liquid must meet certain requirements (→ p.11, section 3.1.1).
► Heed the supplied manufacturer's information regarding restrictions and suitability.
Unsuitable liquids (e.g. acids, chlorinated hydrocarbon solvents) can damage the measuring
probe and probe cable.
WARNING: Risk of incorrect measurements
The measuring liquid must fulfil the stipulated requirements for the measuring probe
at the installation location (→ p.11, section 3.1.1) as otherwise the readings will not be
reliable.
► Make sure that the measuring fluid at the installation location of the measuring
probe fulfils the stipulated requirements.
► If the conductance measurement reading is being used to monitor the safe usage
of the measuring fluid (in particular for explosion protection purposes): Make sure
that the measurement readings are recognised as being invalid if the measuring
liquid at the installation location of the measuring probe does not fulfil the
stipulated requirements. If necessary, install appropriate safety mechanisms.
Unsafe operating conditions could otherwise arise in the monitored system.
2.1.3 Certification for explosion-endangered areas
●The MLA1000-S measuring probe (→ p.9, section 2.3.1) is certified for installation in containers (e.g.
piping systems) holding combustible liquids in Zone 0 hazardous areas.
●The MLA1000-A display unit (→ p.9, section 2.3.2) is certified for use in Zone 1 hazardous areas.
Numbers of the type examination certificates → p.24, section 8.2
● The measuring probe of the MLA1000 is installed in the container through which the measuring
liquid flows (e.g. a piping system). The measuring liquid must meet certain requirements (→ p.11,
section 3.1.1).
●The probe measures the conductance by sending an electrical current through the measuring
liquid by means of low auxiliary voltage. The greater the conductance, the stronger is the electrical
current. The electronics in the probe convert the measured current into a digital measurement
signal.
●The display unit supplies the probe with auxiliary voltage and processes the measurement signals
using a microprocessor. The measurement readings are displayed numerically and outputted as
an analogue signal.
2.3.1 MLA1000-S measuring probe
The measuring probe consists of two parts:
● Probe body (hexagonal) with core electrode and connection for the probe cable
● Sheathed electrode (cylindrical) with inlet and outlet opening
The sheathed electrode can be unscrewed for cleaning purposes.
The probe body has a screw thread for attaching the measuring probe.
The temperature sensor and probe electronics are integrated in the sensor body.
● Technical data → p.24, section 8.1.1.
● Functional test with ring magnet → p.21, section 6.2.
2.3.2 MLA1000-A display unit
The display unit housing consists of two parts:
● Top housing: Display section – contains the electronic components.
● Bottom housing: Terminal box – contains the connection terminals.
The top and bottom parts of the housing are separated by a gas-tight seal.
The electronic connection to the measuring probe is intrinsically safe.
IMPORTANT:
The connection between the housing parts must be supported (certification requirement).
► Secure both housing parts of the display unit on a solid subsurface such as a
installation location as homogeneously
and evenly as possible.
► If possible, install the measuring probe in
a long, straight piece of tubing.
► After a pipe bend, select a distance
equivalent to at least five times the
inside diameter of the pipe (5d).
The measuring liquid at the installation
location should have an homogeneous
composition (fully mixed).
► Do not install the measuring probe
immediately after the mixing stations.
The flow velocity must lie between 0.2
and 7 m/s (→ p.24, section 8.1).
► Specifications → p.24, section 8.1.
Temperature and pressure must be
within
Criterion
Action
The probe cable between the measuring
probe and display unit can have a
maximum length of 24 m.
► Check the length of the supplied probe
cable.
The immersion pipe may have to be
removed after a certain number of
operating hours for cleaning or replacing
the measuring probe.
► Take into account the space required for
the piping system.
►As required: Install auxiliary devices for
disassembling the probe while the piping
system is in operation (sliding sleeve[1],
ball valve[1]).
WARNING: Risk of incorrect measurements
The measuring liquid must fulfil the stipulated requirements at the
measuring probe as otherwise the readings will not be reliable.
►Make sure that the measuring fluid at the installation location of the
measuring probe fulfils the stipulated requirements.
►If the conductance measurement reading is being used to monitor the
safe usage of the measuring fluid (in particular for explosion
protection purposes): Make sure that the measurement readings are
recognised as being invalid if the measuring liquid at the installation
location of the measuring probe does not fulfil the stipulated
requirements. If necessary, install appropriate safety mechanisms.
Unsafe operating conditions could otherwise arise in the monitored system.
3 Installation of the Measuring Probe
3.1 Selecting the installation location for the measuring probe
3.1.1 Requirements for the measuring liquid at the installation location
► Prepare the installation location so that the work can take place in a safe manner.
► On the piping system, mount the devices that are required for installing the immersion pipe.
► If delivery includes a ball valve: Install the ball valve in the piping system as described in the directions for
use provided by the manufacturer of the ball valve (separate document).
If delivery includes the optional immersion pipe, the measuring probe is installed in
the pipe.
WARNING: Danger of explosionIn explosion-endangered areas:► Before beginning with the installation, carry out all the measures that are
necessary to safeguard against explosion during the installation work.
WARNING: Danger of explosion/fire hazardBefore welding:
► Completely empty the piping system.
► Remove all remaining measuring liquid from the piping system.
Vapours can otherwise occur during welding work that will produce an explosive
gas mixture; residues of the measuring liquid could catch fire.
3.3 Installation of the measuring probe with immersion pipe
Preparations for the installation → p.12, section 3.2.
IMPORTANT:
► Install the measuring probe so that the required direction of flow for the probe is
given.
Otherwise the measurement readings will not be reliable.
Marking the direction of flow
► Set the marker arrow on the immersion pipe so that it points in the direction of flow for the
Not applicable if the display unit and immersion pipe were delivered already assembled.
4.1.1 Criteria for the mounting location
●The display unit may be installed in a Zone 1 explosion-endangered area. An installation in Zone
0 is not permitted.
●The possible distance between the display unit and measuring probe is dependent upon the
length of the supplied probe cable (maximum length: 24 m).
● The display unit must be attached to a wall or a stable mounting.
● The display should be visible while the unit is being operated (for checking the operating status).
4.1.2 Securing the display unit without a mounting bracket
1Secure the top housing of the display unit with 2 screws (Ø 8 mm) to a firm, flat subsurface (e.g. metal
wall, metal plate, smooth wall surface of a building).
2Secure the bottom housing with at least 2 screws (Ø 6 mm) to the same subsurface.
4.1.3 Securing the display unit with a mounting bracket
The mounting bracket is a delivery option. If the delivery includes a mounting bracket, the display unit
is already mounted on it.
► Affix the mounting bracket with the unit on the immersion pipe (→ p.13, section 3.3).
Connecting terminals for the probe cable (→ p.17, section 4.4.2)
3
Connecting terminals for the supply voltage (→ p. 18, section 4.4.4) and measurement
reading outputs
(→ p.17, section 4.4.3)
4
Connection for the potential equalisation
2
3
1
4.2 Connecting the potential equalisation
►Join the two connections for the potential equalisation (→ Fig. 5) directly with the main potential (earthing).
4.3 Opening the housing
Fig. 5 Connecting terminals
WARNING: Risk of explosion in explosion-endangered areas
►Before opening the terminal box (bottom part of the display unit): The display unit
must be disconnected from the power supply.
Otherwise there could be a risk of explosion when the terminal box is opened.
1If the display unit is already connected to the power supply: Switch off or disconnect the
power supply.
2 Open the bottom housing of the display unit.
1 2 3 4 5 6 7 1 2 3 4
1089
pS/m
20.1 °C
V0.95
5-2000
The top housing of the display unit does not need to be opened. It may only be
opened by authorised technicians for repair purposes.
Measurement readings cannot be generated.
Possible causes:
– The true physical value is greater than the
end value of the measuring range
(overflow).
4.4 Installing the cables
4.4.1 General information about installing the cables
► Use suitable cable materials → p.24, section 8.1.2.
► Run all the cables through the cable glands in the bottom housing of the display unit (→ Fig. 2).
When connecting the conductors:
► Strip the conductor insulation to a length of 7 mm.
► Tighten the clamping screw on the connecting terminal with a torque of at least 0.3 Nm.
WARNING: Danger of explosion
The cable glands are subject to approval for explosion-endangered areas.
► Only use cable materials with a suitable outside diameter.
► If the MLA1000 is being used in an explosion-endangered area: Do not replace
the cable glands with cable glands of a different type. Otherwise the certification
for explosion-endangered areas will become void and there will be a risk of
explosion in such areas.
4.4.2 Connecting the probe cable
►Connect the probe cable to the blue 4-pole terminal block (→ Table 1).
4.4.3 Connecting the measurement reading outputs
Table 1 Connecting terminals for the probe cable
The measurement reading outputs continually give out the actual measured values. The physical
measuring range is outputted in the signal range between 4 and 20 mA.
►Connect the measurement reading outputs as shown in the terminal diagram (→ Table 2).
Table 2 Output signals for the measurement reading outputs
► Connect the power supply as shown in the terminal diagram (→ Table 3).
► Make sure that the current fed into the power supply lead cannot exceed 5 A.
Specification for the power supply → p.24, section 8.1
4.4.5 Closing the cable glands and housing
Table 3 Connecting terminals for the measurement reading output and power supply
After installing the cables:
► Seal the cable glands.
► Close the housing.
WARNING: Danger of explosion
Before putting the unit into operation in explosion-endangered areas:
► Seal all the cable glands so that they are "flame-tight" (practically gas-tight).
► Either close off any unused cable glands with a closure plug or replace them
completely with sealing caps.
– Closure plugs: Select suitable closure plugs for the permitted cable
diameters and install them instead of a cable.
– Sealing caps: Select sealing caps with thread M20 x 1.5 which are certified
for usage in explosion-endangered areas. Apply suitable adhesive to the
threads and sealing surfaces.
► Tightly close the display unit housing.
Otherwise there could be a risk of explosion.
Explanation:
MLA1000 is starting up (approx. 3 seconds).
Actions:
► Wait until the operating screen appears (→ section 5.3.2).
► If the operating screen does not appear: Refer to the
information about the malfunction screens (→ section 5.3.3).
1089 pS/m
20.1 °C
V0.95 5-2000
Explanation:
MLA1000 is in the normal operating status (measuring mode).
1 = actual conductance measurement reading (sample value)
2 = actual temperature measurement reading (sample value)
3 = version number of the firmware (example)
Actions:
► Check/ensure that the requirements for safe operation are
fulfilled (see warning note).
1
2
5 Operation
5.1 Putting the MLA1000 into operation
1 Make sure that the display unit housing is tightly closed.
2 Switch on the supply voltage.
3 Wait for the display to light up after switching it on (→ p.19, section 5.3.1).
4 Check the operating status (→ section 5.2).
5.2 Checking the operating status
5.2.1 Indications for a safe operating status
1The measuring liquid fulfils the requisite operating conditions at the measuring probe
(→ p.11, section 3.1.1, e.g. flow velocity).
2 The display on the display unit shows the measurement readings (→ section 5.3.2).
3 Measurement readings are returned at the measurement reading outputs.
Testing the measuring function → p.21, section 6.2.
5.2.2 Indications for an unsafe operating status
Nothing is shown on the display of the display unit.
A malfunction is shown on the display (→ p.20, section 5.3.3).
At least one measurement reading output returns "0 mA".
At least one measurement reading output returns "24 mA".
The measuring liquid does not have the required flow velocity at the measuring probe
(→ section 5.3.2).
The measuring liquid is not within the permitted temp. range at the measuring probe (→ section 5.3.2).
Measurement readings are being displayed that cannot be correct.
At least one measurement reading output returns readings that cannot be correct.
5.3 Display screens
5.3.1 Screen after switching on
5.3.2 Operating screen
WARNING: Requirements for safe operation
● The flow velocity of the measuring liquid at the measuring probe must be between 0.2
and 7 m/s.
● The measuring liquid temperature must be between -20 and +60 °C. Otherwise the
measurement readings will not be reliable.
6.1 Removing the immersion pipe (with measuring probe)
● The MLA1000 does not have to be shut down for this work.
● If the MLA1000 remains in operation, the measuring function can be tested using
simple means (→ section 6.2).
CAUTION: Possible danger to healthIf the measuring liquid could represent a health risk:►Before removing the immersion pipe, take the appropriate protective measures suitable
for the measuring liquid (e.g. safety goggles, protective clothing, respirator).
For installations without a ball valve
1 Take the piping system out of operation (stop the liquid flow).
2 Remove the immersion pipe from the unit.
For installations with a ball valve
► Follow the instructions given by the manufacturer of the ball valve (separate document).
6.2 Testing the measuring function
● This procedure tests the electronic functionality of the measuring probe and
display unit.
● The supplied ring magnet is required for this procedure.
● Adjustment (electronic adjustment of the measuring probe and display unit)
are not necessary during operation.
A measuring probe that is clean, dry and functioning correctly the device displays
"low" in atmosperetic air.
1 Make sure
– that the sheathed electrode is firmly screwed onto the measuring probe,
– that the probe is clean and dry,
– that the probe is properly connected to the display unit (probe cable, plug-in connection on the
probe).
2 Place the ring magnets on the surface of the measuring probe opposite the labelled area.
For the test value for electrical conductivity, please refer to the test certificate.
►If the test value is not within the tolerance range: Notify the manufacturer's customer service or
return the MLA1000 to the manufacturer for repair.
WARNING: Hazard in explosion-endangered areas► Do not perform cleaning work in explosion-endangered areas.
IMPORTANT: Risk of damage from rough handling
● Exerting mechanical pressure on the probe body can destroy the plastic
moulding in the body (cause it to rupture).
● Deformations will lead to incorrect measuring results.
● Dirt in scratches can falsify the measurement readings.
► Do not deform the probe body or sheathed electrode.
► Do not clamp the measuring probe in a vice - neither the probe body nor the
sheathed electrode.
► Only use tools with a soft surface.
► Make sure that the plastic moulding is not scratched.
Fig. 6 MLA1000-S measuring probe
1 Unscrew the sheathed electrode.
If the sheathed electrode cannot be unscrewed by hand: Use a tool with a soft surface. Only apply
pincer-like tools at the bottom-most rim (closed end) of the sheathed electrode.
2 Carefully clean all the surfaces of the probe body and sheathed electrode using a soft cloth moistened with a
suitable solvent (→ Table 5).
The cleanliness of the parts determines the quality of the conductance measurements.
Table 5 Cleaning agents
IMPORTANT: Risk of damage from solvents
► Only use suitable cleaning agents. Other substances can damage the
measuring probe.
WARNING: Danger due to damaged measuring probes► Do not use damaged measuring probes in explosion-endangered areas.
6.4 Reinstalling the measuring probe
1 Reassemble the measuring probe and immersion pipe (→ p.22, section 6.4).
2 Reinstall the immersion pipe (→ p.13, section 3.3).
► Check the mains power supply (e.g. external master
switch, external fuse, connecting cable).
The operating temperatures are
incorrect.
► Check the ambient temperatures (permitted ranges
→ p.24, section 8.1).
The internal software is not
functioning.
► Switch off the supply voltage to the MLA1000 and
then switch it back on again (reset) after a few
seconds.
Note: Can only occur in conjunction with complex
internal malfunctions or extreme external influences (e.g.
strong electromagnetic interference pulse).
The internal overcurrent fuse
(PTC resettable fuse) has been
tripped.
1 Switch off the supply voltage to the MLA1000.
2 Wait for several minutes.
3 Put the MLA1000 back into operation
(→ p.19, section 5.1).
If the overcurrent fuse is tripped again:
The MLA1000 is defective.
► Have it repaired.
Possible cause
Remarks
The MLA1000 is not ready for
operation.
► Check the operating status (→ p.19, section 5.2).
The requirements for the
measuring liquid are not being
fulfilled (e.g. flow velocity).
► Check the operating conditions at the measuring
probe (requirements → p.11, section 3.1.1).
If only the case at the
measurement reading output:
The output load is too great.
► Make sure that the internal resistance of the
connected devices does not exceed the permitted
output load (→ p.24, section 8.1).
The measuring probe is soiled.
► Visually check the condition of the measuring probe.
► Clean the probe if necessary (→ p.22, section 6.3).
The measuring probe is
defective.
► Visually check the condition.
► Test the measuring function (→ p.21, section 6.2).
The probe cable is damaged.
► Check the probe cable.
► Check the probe cable connections.
7 Troubleshooting
7.1 If the MLA1000 does not function at all ...
7.2 If the measurement readings are obviously incorrect...