Maxwell 25304 User Manual

User manUal
Használati Utasítás
manUal de Utilizare
Užívateľská prírUčka
DIGITAL MULTIMETER
DIGITÁLIS MULTIMÉTER
MULTIMETRU DIGITAL
Product code / Termékkód / Cod produs / Kód produkta:
25304
HU
RO
SK
i
VC9808+
EN
EN
User Manual
General description
Display
4 digit liquid crystal display max. value 9999
Polarity
Automatic negative polarity display
Nullication automatic
Measurement process
A/D converter
Overload display only „OL” is displayed
Low battery power
The( ) symbol appears on the screen
Safety prescriptions
CE EMC/LVD. Device complies to IEC1010 standard
Protection category
II, double insulation
Operating environment
temperature: 0…40 °C, relative humidity: < 80%
Storage environment
temperature: -20…60°C, relative humidity: < 80%
Battery 1 x 9V IEC 6F22 Battery
Dimensions
190 mm x 88,5 mm x 27,5 mm
Weight 422 g (with battery)
Accesories
Instructions, connection cable (red and black, 1 pair), hitting pocket protector, K-type thermometer probe
electrical cHaracteristics
Accuracy +/- (% of displayed value + number of digits) at 23 +/-5 °C, < 75% relative humidity
DCV
Range Accuracy Resolution
200 mV
+/- (0,5% + 3)
0,1 mV
2 V 1 mV
20 V 10 mV
200 V 100 mV
1000 V +/- (0,8% + 10) 1 V
Input Impedance
10 MΩ
Overload protection
(Range: 200 mV)
1000 VDC or AC peak value
250 VDC or AC peak value
ACV
Range Accuracy Resolution
200 mV
+/- (0,8% + 5)
0,1 mV
2 V 1 mV
20 V 10 mV 200 V 100 mV 750 V +/- (1,2% + 10) 1 V
Input Impedance 10 MΩ
Frequency response
40 Hz - 1 KHz (sine wave and triangular wave)
40 Hz - 200 Hz (other waves)
DCA
Range Accuracy Resolution
2 mA
+/- (0,8% +10)
1 μA
20 mA 10 μA
200 mA 100 μA
20 A +/- (2% + 5) 10 mA
Overload protection
20 A / 250 V fuse with "F" mark
200 mA / 250 V fuse with "F" mark
User Manual
EN
ACA
Range Accuracy Resolution
2 mA
+/- (0,8% + 10)
1 μA
20 mA 10 μA
200 mA 100 μA
20 A +/- (2% + 5) 10 mA
Overload protection
20 A / 250 V fuse with "F" mark
200 mA / 250 V fuse with "F" mark
Frequency response
40 Hz - 1 KHz (sine wave and triangular wave)
40 Hz - 200 Hz (other waves)
Capacitance
Range Accuracy Resolution
20 nF
+/- (3,5% + 20)
10 pF
200 nF 100 pF
2 μF 1 nF
20 μF 10 nF
200 μF
+/- (5% + 10)
100 nF
2000 μF 1 μF
Overload protection
DC 250V or AC peak value
Transistor hFE-Test
Function Description Test state
h
FE
The amplication
factor of the transis-
tor is measured
(0-1000)
(NPN or PNP)
Base current
appr. 10 μA
VCE appr.
3 V
Resistance
Range Accuracy Resolution
200 Ω +/- (0,8% + 5) 0,1 Ω
2 kΩ
+/- (0,8% + 3)
1 Ω
20 kΩ 10 Ω
200 kΩ 100 Ω
2 MΩ 1 kΩ
20 MΩ
+/- (1,0% + 25)
10 kΩ
Overload protection
250 V DC or AC peak value
The reading be displayed slowly when the measurement is more than 1 MΩ. Please wait it to be stable.
Temperature (with a K type temperature sensor)
Range Accuracy Resolution
-20 °C - 1000 °C +/- (1,5%+15)
1°C
0 °F - 1832 °F 1 °F
Diode and Continuity Testing
Function Reading Condition
Forward voltage
drop of diode
Forward DCA is
approx. 1 mA, the backward
voltage is
approx. 3 V
Buzzer makes a
long sound while
resistance is less
than
70 Ω +/- 20 Ω
Opening volta-
ge appr. 3 V
Overload protection
250 V DC or AC peak value
EN
User Manual
Frequency
Range Accuracy Resolution
10 Hz
+/-(0,1%+3)
0,01 Hz
100 Hz 0,1 Hz
1 kHz 1 Hz
10 kHz 10 Hz
100 kHz 100 Hz
1 MHz/10 MHz 1 kHz/10 kHz
Overload protection
250 V DC or AC peak value (for maximum 10 seconds)
Input sensitivity 1 V RMS
Inductance
Range Accuracy Resolution
2 mH
+/-(2,5%+30)
1 μH
20 mH 10 μH
200 mH 100 μH
2 H 1 mH
20 H 10 mH
Overload protection DC 36 V
Warning: Do not connect external voltage source to the connectors!
1
2-2 2-1
7 6
5
4
3
2-4
2-3
1. LCD: display the measuring value and unit
2. Function keys 2-1. Power switch: turn on/o the power 2-2. Hold and backlight key: short press to the
measured value is held on LCD and "HOLD" symbol displays. Press it again "HOLD" symbol disappears and meter is exited the holding mode. Long press to turn on/o the backlight.
2-3. Transistor test socket. 2-4. Press "REL/MAX/MIN", then you can test the
relative magnitude. If you hold the "REL/MAX/ MIN" for more than 2s, you can test the Max
and Min value.
3. Range knob: to select measuring function and
range.
4. Voltage, resistance, frequency, capacitance, diode port - „V / Ω
5. Common measuring port (COM)
6. Port for measuring current less than 200 mA
and inductivity
7. Port for measuring current 20 A
Usage
• Check the 9 V battery and press the "POWER"
button. If the battery is depleted, the battery symbol appears on the screen.
• The signs near the connectors warn about not
exceeding the inward voltage or current. This is to prevent damage to the inside circuits.
• Before measuring set the function switch to the
desired position.
• If you are unsure about the magnitude of the
result, set the switch to the highest possible range and go backwards until you reach the correct setting.
DC and AC voltage measuring
• Connect the black measuring wire to the „COM”,
and the red one to the „V / Ω” connector.
• Set the function switch to the correct "V"
position and connect the wires to the voltage source paralelly.
Note:
• If you are unsure about the magnitude of the result, set the switch
to the highest possible range and go backwards until you reach the correct setting.
• Set the DC / AC switch to the desired (DC or AC) function.
• If only an „OL” is displayed on the screen, it signals overload.
Switch to a higher range.
• Do not connect voltage exceeding DC 1000 V / AC 700 V to the
connectors. Results may be displayed at higher voltages also, but this may lead to damage to the inside circuits.
• Do not touch the high voltage circuits while measuring.
User Manual
EN
DC and AC current measuring
• Connect the black wire to the „COM” and the
red one to the „mA” connector (max. 200 mA) or the „20A” connector for measuring 200 mA or 20 A.
• Set the function switch to the proper range.
• Connect the the wires to the circuit serially.
• For measuring current between 200 mA and 20
A follow the instructions above, but connect the red wire to the „20A” connector.
Note:
• If you are unsure about the magnitude of the result,
set the switch to the highest possible range and go backwards until you reach the correct setting.
• If only an „OL” is displayed on the screen, it signals
overload. Switch to a higher range.
• The maximum inward current is 200 mA or 20 A
depending on the connector. (For maximum 15 seconds.) Too much currency overheats the fuse which needs to be replaced. The 20 A measuring range is not protected.
• Maximum overload voltage: 200 mV.
Resistance measuring
• Connect the black wire to the „COM”, and the red
one to the „V / Ω” connector.
• Set the function switch to the desired resistance
range.
• Touch the wires to the measured circuit paralelly.
Warning: make sure that the measured circuit is not under power!
• The maximum overload capability for the
connector is 250 V RMS (for maximum 10 seconds).
Note:
• If the measured resistance value is over the maximum
threshold and the device signals overload („OL”), select a higher range. When measuring above 1 MW it may take a few seconds for the device to stabilize the displayed value. This is completely normal for such resistors.
• If you do not connect a resistor to the connector (or there is
a break) the display will show „OL”.
• If you are measuring a resistor that is part of a circuit make
sure that the circuit is not under power and all capacitors are discharged.
• Do not connect inward voltage to the device as it may lead to
damage to the inside circuits.
• The outward voltage on connectors not under power is
approximately 3 V.
Capacity measuring
• Set the function switch to "2 mF" position.
• Connect the measured capacitor to the „V /
Ω” and "COM" connector. Pay attention to the correct polarity.
Note:
• If the measured value exceeds the maximum value of the
selected range and the device signals overload („OL”).
• Unit: 1 nF=10
-3
μF or 1000 pF
• Do not connect external voltage or current to the connector!
Turn o the power and discharge the capacitors before measuring. Also discharge the electrolyte capacitors before measuring!
• Before measuring discharge the electrolyte capacitors
more times.
Diode and continuity test
• Connect the black measuring wire to the
COM”, the red one to „V / Ω” connector. (Note: the polarity of the red wire is: +)
• Set the function to „ ” position.
• Touch the wires to the diode's connectors.
The screen shows the opening voltage of the diode.
• Connect the wires to two points of the circuit. If
the resistance is below 90 Ω the device beeps.
Note:
• If you do not connect a unit to the connectors (or there is a
break) the display will show „OL”.
• During the test 1 mA current goes through the diode.
• The display shows the opening voltage of the diode in mV,
and overload if the diode is connected with the polarity reversed
Temperature measuring
• Set the function switch to "°C/°F" position.
• Connect the black wire of the temperature
measuring sensor to the „COM” connector and the red one to the „V/Ω” connector. Put the sensor to the place measured. The display shows the measured value in °C or °F (press "HOLD" button to select).
Note:
• The device functions with a special temperature measuring
sensor.
• If the sensor is not connected to the connector, the device
displays the temperature of the environment.
• Do not connect outward voltages to the connectors, if the
device is in temperature measuring function.
Transistor hFE test
• Set the function switch to "hFE" position.
• Determine if the transistor is NPN or PNP and
place it in the proper connector.
• The amplication factor of the transistor is
displayed on the screen.
• IB = 10 μA, VCE = 2,8 V
EN
User Manual
Frequency measuring
• Connect the measuring wire or the shielded
cable to the „COM” and „V / Ω” connectors.
• Set the function switch to „20MHz” position
and touch the wires to the signal source.
Note:
• Do not measure frequency at a voltage exceeding 250
V(RMS).
• In noisy environment it is advised to use a shielded cable for
measuring small values.
• Avoid touching the circuit when measuring high voltage!
• Frequency measuring changes range automatically.
Inductance measuring
• Set the function swith to the desired measuring
range.
• Connect the measured inductor to the "mA"
and "COM" terminal.
Remark:
• If the measured inductance is not known, start measuring
at H range until overload signal turns o.
• Inductors with low inductance function as short circuits. If
possible, do not connect such inductors to the connectors.
Auto Power O
• After stop working for 15 minutes, the meter
will be into sleep mode. Press "POWER" key can back to work.
• Press "REL/MAX/MIN" key and at the same time
with the "POWER" key the "APO" symbol will disappear, now you already closed the function of auto power o.
WarninG
• When measuring voltage make sure that the
wires are not connected to a current measuring connector and that the function switch is not in resistance or diode measuring measuring mode. Always check if you have connected the wires into the proper connectors.
• Be careful when measuring voltage above 50 V,
especially with high voltage equipment.
• Avoid connecting to „live” circuits.
• Turn o the power in the circuit before
connecting the multimeter to measure current and do not measure above 20 A.
• Before measuring resistance and diode make
sure the circuit is not under power.
• Always set the function and measuring range
appropriate for the measuring. If in doubt,
set the maximum measuring range and go backwards.
• Make sure that the wire and its isolation is not
damaged.
• Only replace fuses to the same type and value.
• When replacing the fuse or battery make sure
all external power sources are turned o and that the multimeter is turned o as well.
care and maintenance
Operation
• Keep the multimeter dry. If it gets wet, dry it up
immediately. The liquid corrodes the electric circuit.
• The multimeter should be stored and used
only at normal temperature. High temperature can shorten the lifetime of electronic devices, damage to the batteries and distort, melt the plastic parts.
• Handle the multimeter carefully. If it is dropped
that can cause damage to the electric circuit and to the case which leads to the improper operation of the multimeter.
• Keep away the multimeter from dust and
dirt which results in premature wear of the components.
• The multimeter can be cleaned with wet
cloth. Do not use chemicals, solvents or strong detergents for cleaning.
• Servicing
• Battery replacement (1 piece 9V)
• Turn o all external circuits. Switch o the
multimeter and pull out the test lead of the socket.
• Unscrew the screws and remove the bottom
cover.
• Remove the low battery and replace a same
type of it.
Fuse replacement
• Turn o all external circuits. Set the switch to
OFF position and pull out the test lead of the socket.
• Unscrew the screws and remove the bottom
cover.
• Replace the blown out fuse to a same type and
valued (5x20 mm, 200 mA / 250 V „F” signed.)
Használati utasítás
HU
általános leírás
Kijelző
4 digites folyadékkristályos max. kijelezhető érték 9999
Polaritás
automatikus negatív polaritás kijelzés
Nullázás automatikus Mérési eljárás A/D konverter
Túlterhelés kijelzés
csak az „OL” felirat látható
Alacsony telepfeszültség
az elem( ) szimbólum megjelenik a kijelzőn
Biztonsági előírás
CE EMC/LVD. A műszer megfelel az IEC1010 szabványnak
Érintésvédelmi osztály
II. Kettős szigetelésű
Üzemi környezet hőmérséklet (relatív páratartalom)
0-40 °C (< 80%)
Tárolási környezet hőmérséklet (relatív páratartalom)
-20 °C- +60 °C (< 80%)
Elem
1 db 9V-os IEC 6F22 típusú elem
Méretek
190 mm x 88,5 mm x 27,5 mm
Tömeg 422 g (elemmel)
Tartozékok
használati utasítás, műszerzsinór (piros és fekete,1 pár), ütésvédő tok, K tipusú hőmérő szonda
elektromos jellemzők
A pontosság +/- (kijelzett érték %-a + digitek száma) 23 +/-5 °C-on, 75%-nál kisebb relatív páratartalom esetén
DCV
Méréshatár Pontosság Felbontás
200 mV
+/- (0,5% + 3)
0,1 mV
2 V 1 mV
20 V 10 mV
200 V 100 mV
1000 V +/- (0,8% + 10) 1 V
Impedancia 10 MΩ
Túlterhelés elleni védelem
(tartomány: 200 mV)
1000 VDC vagy AC csúcsérték
250 VDC vagy AC csúcsérték
ACV
Méréshatár Pontosság Felbontás
200 mV
+/- (0.8% + 5)
0,1 mV
2 V 1 mV
20 V 10 mV 200 V 100 mV 750 V +/- (1.2% + 10) 1 V
Impedancia 10 MΩ
Frekvenciaátvitel
40 Hz - 1 KHz (szinuszjel és háromszögjel)
40 Hz - 200 Hz (egyéb jel)
DCA
Méréshatár Pontosság Felbontás
2 mA
+/- (0,8% + 10)
1 μA
20 mA 10 μA
200 mA 100 μA
20 A +/- (2% + 5) 10 mA
Túlterhelés elleni védelem
20 A / 250 V "F" jelzésű biztosíték
200 mA / 250 V "F" jelzésű biztosíték
Loading...
+ 14 hidden pages