INDICATIONS FOR USE:
The UltraMaxO2 is an oxygen analyzer for checking the oxygen concentration, flow and outlet pressure of oxygen concentrators.
WARRANTY:
Under normal operating conditions, Maxtec warrants the UltraMaxO2 to be free from defects of workmanship or materials for a period of Three (3) years from the date of shipment from Maxtec, provided that the unit is properly operated and maintained in accordance with Maxtec’s operating instructions. Based on Maxtec product evaluation, Maxtec’s sole obligation under the foregoing warranty is limited to making replacements, repairs, or issuing credit for equipment found to be defective. This warranty extends only to the buyer purchasing the equipment directly from Maxtec or through Maxtec’s designated distributors and agents as new equipment.
Routine maintenance items, such as batteries, are excluded from warranty. Maxtec and any other subsidiaries shall not be liable to the purchaser or other persons for incidental or consequential damages or equipment that has been subject to abuse, misuse, mis-application, alteration, negligence or accident.
These warranties are exclusive and in lieu of all other warranties, expressed or implied, including warranty of merchantability and fitness for a particular purpose.
CLASSIFICATIONS:
Classification:.............................................. Class II electrical device
Protection against electric shock: ...........................Internally powered equipment
Protection against water: ....................................................IPX1
Mode of operation:....................................................Continuous
Sterilization: ......................................................See section 6.0
Flammable anaesthetic mixture:......Not for use in presence of flammable anaesthetic mixtures
Power specification: .......................................1.8-3.2V 32mW10mA
Caution: Federal law restricts this device to sale by or on the order of a physician or other licensed healthcare practitioner.
NOTE: The UltraMaxO2 is for use only by trained personnel. Before use, all individuals using the UltraMaxO2 should become familiar with the information contained in this Operation Manual. Adherence to these instructions is necessary for safe, effective product performance. Thoroughly read all instructions and labeling provided with this device and any other equipment that will be used.
Maxtec TEL: (800) 748.5355 6526 South Cottonwood Street FAX: (801) 270.5590 Salt Lake City, Utah 84107 EMAIL: sales@maxtecinc.com USA WEBSITE: www.maxtecinc.com
Authorized Representative: QNET BV Hommerterweg 286 6436 AM Amstenrade The Netherlands
EC REP
Product Disposal Instructions:
The batteries and circuit board are not suitable for regular trash disposal. Follow local guidelines for proper disposal
866.4.Maxtec www.maxtecinc.comII
PRINCIPLE OF OPERATION:
The UltraMaxO2 Oxygen Analyzer measures oxygen concentration and flow using ultrasound technology and measures pressure using a piezoresistive silicon pressure sensor.
WARNINGS:
Indicates a potentially hazardous situation, if not avoided, could result in death or serious injury.
• Not for use in an MRI environment.
• Improper use of the UltraMaxO2 can cause inaccurate oxygen readings leading to
improper treatment and/or patient harm. Follow the procedures outlined in this
user manual.
• The UltraMaxO2 is for checking oxygen concentrators only.
• DO NOT use the UltraMaxO2 for continuous oxygen monitoring.
• DO NOT use the UltraMaxO2 to measure the oxygen concentration of a concentrator
when flowing at rates lower than its optimal performance as specified by the concentrator manufacturer; generally 4 LPM or less on concentrators that have a maximum flow of 10 LPM, and 1 LPM or less on concentrators that have a maximum flow of 5 LPM.
• Not for use in anesthesia applications or for measuring oxygen concentration from any sources other than conventional oxygen concentrators.
• Not for use with inhalation agents. Operating the UltraMaxO2 in flammable or explosive environments may result in fire or explosion.
• Not suitable for use in the presence of flammable anesthetic mixtures.
• Users must become thoroughly familiar with the information contained in this
Operation Manual before use. Strict adherence to the operating instructions is necessary for safe, effective product performance. This product will perform only as designed if operated in accordance with the manufacturer’s operating instructions.
• Use only genuine Maxtec accessories. Failure to do so may seriously impair the performance of the UltraMaxO2. Repair or alteration of the UltraMaxO2 by anyone other than an authorized Maxtec service representative could cause the product to fail to perform as designed.
• Use of the UltraMaxO2 near devices that generate electrical fields may cause erratic readings.
• If the UltraMaxO2 is ever exposed to liquids from spills or immersion, immediately remove the batteries and let the device dry completely. When dry, replace the batteries and check for proper operation.
• DO NOT autoclave or expose the UltraMaxO2 to high temperatures (>60°C).
• DO NOT use ethylene oxide sterilization. performance in flow and pressure readings.
!