Maxtec MaxO2 Plus Instructions Manual

OPERATING MANUAL &
INSTRUCTIONS FOR USE
R217M72 Rev. E
MaxO
2
+
®
2
®
+ AE Industrial
BAT
CAL
I
866.4.Maxtec www.maxtec.com
Maxtec TEL (800) 748.5355 2305 South 1070 West FAX (801) 973.6090 Salt Lake City, Utah 84119 email: sales@maxtec.com USA website: www.maxtec.com
CLASSIFICATION
Protection against electric shock: ..........................Internally powered equipment.
Protection against water: ....................................................IPX1
Mode of Operation: ...................................................Continuous
Sterilization: ......................................................See section 7.0
Flammable anesthetic mixture: Not suitable for use in presence of a flammable anesthetic mixture
Product Disposal Instructions:
The sensor, batteries, and circuit board are not suitable for regular trash disposal. Return sensor to Maxtec for proper disposal or dispose according to local guidelines. Follow local guidelines for disposal of other components.
WARRANTY
Under normal operating conditions, Maxtec warrants the MAXO2+ Analyzer to be free from defects of workmanship or materials for a period of 2-years from the date of shipment from Maxtec, provided that the unit is properly operated and maintained in accordance with Maxtec’s operating instructions. Based on Maxtec product evaluation, Maxtec’s sole obligation under the foregoing warranty is limited to making replacements, repairs, or issuing credit for equipment found to be defective. This warranty extends only to the buyer purchasing the equipment directly from Maxtec or through Maxtec 's designated distributors and agents as new equipment.
Maxtec warrants MAXO2+ oxygen sensor in the MAXO2+ Analyzer to be free from defects in material and workmanship for a period of 2-years from Maxtec’s date of shipment in a MAXO2+ unit. Should a sensor fail prematurely, the replacement sensor is warranted for the remainder of the original sensor warranty period.
Routine maintenance items, such as batteries, are excluded from warranty. Maxtec and any other subsidiaries shall not be liable to the purchaser or other persons for incidental or consequential damages or equipment that has been subject to abuse, misuse, mis-application, alteration, negligence or accident.
These warranties are exclusive and in lieu of all other warranties, expressed or implied, including warranty of merchantability and fitness for a particular purpose.
EC REP
QNET BV Kantstraat 19 NL-5076 NP Haaren The Netherlands
II
866.4.Maxtec www.maxtec.com
WARNINGS
Indicates a potentially hazardous situation, if not avoided, could result in death or serious injury.
» Improper use of this device can cause inaccurate oxygen readings which can lead to improper
treatment, hypoxia or hyperoxia. Follow the procedures outlined in this user manual.
»
Device specified for dry gas only.
»
Before use, all individuals who will be using the MAXO2+ must become thoroughly
familiar with the information contained in this Operation Manual. Strict adherence to the operating instructions is necessary for safe, effective product performance. This product will perform only as designed if installed and operated in accordance with the manufacturer’s operating instructions.
»
Use only genuine Maxtec accessories and replacement parts. Failure to do so may seriously impair the
analyzer’s performance. Repair of this equipment must be performed by a qualified service technician experienced in repair of portable hand held equipment.
»
Calibrate the MAXO2+ weekly when in operation, or if environmental conditions change significantly.
(ie, Elevation, Temperature, Pressure, Humidity — refer to Section 3.0 of this manual).
»
Use of the MAXO2+ near devices that generate electrical fields may cause erratic readings.
»
If the MAXO2+ is ever exposed to liquids (from spills or immersion) or to any other physical abuse, turn
the instrument OFF and then ON. This will allow the unit to go through its self test to assure everything is operating correctly.
»
Never autoclave, immerse or expose the MAXO2+ (including sensor) to high temperatures (>70°C).
Never expose the device to pressure, irradiation vacuum, steam, or chemicals.
»
This device does not contain automatic barometric pressure compensation.
»
Although the sensor of this device has been tested with various gases including nitrous oxide,
Halothane, Isoflurane, Enflurane, Sevoflurane and Desflurane and found to have acceptably low interference, the device in entirety (including electronics) is not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide. Only the threaded sensor face, flow diverter, and “T” adapter may be allowed to contact such a gas mixture.
»
NOT for use with inhalation agents. Operating the device in flammable or explosive atmospheres may
result in fire or explosion.
CAUTION: Indicates a potentially hazardous situation, if not avoided, could result in minor or moderate injury and property damage.
»
Replace the batteries with recongnized high quality AA Alkaline or Lithium batteries.
DO NOT use rechargeable batteries. »
If the unit is going to be stored (not in use for 1 month), we recommend that you remove the
batteries to protect the unit from potential battery leakage.
»
The Maxtec MAX-250 oxygen sensor is a sealed device containing a mild acid electrolyte, lead
(Pb), and lead acetate. Lead and lead acetate are hazardous waste constituents and should be disposed of properly, or returned to Maxtec for proper disposal or recovery.
DO NOT use ethylene oxide sterilization. DO NOT immerse the sensor in any cleaning solution, autoclave or expose the sensor to high temperatures.
»
Dropping sensor can adversely affect its performance.
»
The device will assume a percent oxygen concentration when calibrating. Be sure to apply
100% oxygen, or ambient air concentration to the device during calibration or the device will not calibrate correctly.
NOTE: This product is latex free.
!
III
866.4.Maxtec www.maxtec.com
TABLE OF CONTENTS
Classification ............................................I
Warranty ...............................................I
Warnings .............................................. II
1.0 SYSTEM OVERVIEW ....................................1
1.1 Base Unit Description .....................................1
1.2 Symbol Guide ...........................................1
1.3 Component Identification...................................2
1.4 Component Description ....................................2
1.5 MAX-250+ Oxygen Sensor..................................3
2.0 OPERATING INSTRUCTIONS..............................3
2.1 Getting Started ..........................................3
2.1.1 Protect Tape ...........................................3
2.1.2 Automatic Calibration ....................................3
2.2 Calibrating the MAXO2+ Oxygen Analyzer ......................4
2.2.1 In-line Calibration (Flow Diverter - Tee Adaptor).................4
2.2.2 Direct Flow Calibration (Barb) ..............................5
3.0 FACTORS INFLUENCING ACCURATE READINGS ...............5
3.1 Elevation Changes ........................................5
3.2 Temperature Effects.......................................5
3.3 Pressure Effects .........................................5
3.4 Humidity Effects .........................................6
4.0 CALIBRATION ERRORS AND ERROR CODES ..................6
5.0 CHANGING THE BATTERIES ..............................7
6.0 CHANGING THE OXYGEN SENSOR .........................8
6.1 MAXO2+AE Model ........................................8
7.0 CLEANING AND MAINTENANCE ...........................8
8.0 SPECIFICATIONS ......................................9
8.1 Base Unit Specifications ...................................9
8.2 Sensor Specifications......................................9
9.0 MAXO2+ SPARE PARTS AND ACCESSORIES ..................9
9.1 Included with your Unit ....................................9
9.2 Standard Replacement Parts ...............................10
9.3 Optional Accessories .....................................10
9.3.1 Optional Adaptors......................................10
9.3.2 Mounting Options ......................................10
9.3.3 Carrying Options.......................................10
Loading...
+ 11 hidden pages