Maxtec MaxO2 Oxygen Meter User manual

®
MaxO
OPERATING MANUAL &
INSTRUCTIONS FOR USE
OM-25A
2
R213M15 Rev. P
EC REP
This manual describes the function, operation and maintenance of the Maxtec Model OM-25A hand-held oxygen analyzer. A member of Maxtec's MAXO and monitors, the OM-25A utilizes the MAX-250 oxygen sensor and is engineered for fast response, maximum reliability and stable performance. The OM-25A is designed primarily for spot-checking oxygen concentration delivered by oxygen concentrator units, however, its internal sensor and adaptable sampling port make it suitable for a wide variety of oxygen analysis applications.
QNET BV Hommerterweg 286 6436 AM Amstenrade The Netherlands
®
analyzer line of oxygen analyzers
2
CLASSIFICATION
Classication: .............................................Class II medical device.
Protection against electric shock:........................ Internally powered equipment.
Protection against water: ...................................... Ordinary Equipment
Mode of Operation:.................................................. Continuous
Sterilization: ....................................................See section 6.0
Safety of application in the presence of a ammable anesthetic mixture: .....See Section 8.1
Product Disposal Instructions:
The sensor, batteries, and circuit board are not suitable for regular trash disposal. Return sensor to Maxtec for proper disposal or dispose according to local guidelines. Follow local guidelines for disposal of other components
WARRANTY
The MAXO Under normal operating conditions, Maxtec warrants the MAXO defects of workmanship or materials for a period of two (2) years from the date of shipment from Maxtec, provided that the unit is properly operated and maintained in accordance with Maxtec’s operating instructions. Based on Maxtec’s product evaluation, Maxtec's sole obligation under the foregoing warranty is limited to making replacements, repairs, or issuing credit for equipment found to be defective. This warranty extends only to the buyer purchasing the equipment directly from Maxtec or through Maxtec's designated distributors and agents as new equipment. Maxtec warrants the MAX-250 oxygen sensor in the MAXO be free from defects in material and workmanship for a period of two (2) years from Maxtec's date of shipment in a MAXO is warranted for the remainder of the original sensor warranty period. Routine maintenance items, such as batteries, are excluded from warranty. Maxtec and any other subsidiaries shall not be liable to the purchaser or other persons for incidental or consequential damages or equipment that has been subject to abuse, misuse, mis-application, alteration, negligence or accident. THESE WARRANTIES ARE EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
®
Analyzer is designed for medical oxygen delivery equipment and systems.
2
®
unit. Should a sensor fail prematurely, the replacement sensor
2
®
Analyzer to be free from
2
®
Analyzer to
2
866.4.Maxtec www.maxtecinc.com I
NOTE: In order to obtain optimum performance from your MAXO
!
and maintenance must be performed in accordance with this manual. Please read the manual thoroughly before using the analyzer and do not attempt any repair or procedure that is not described herein. Maxtec cannot warrant any damage resulting from
misuse, unauthorized repair or improper maintenance of the instrument.
®
analyzer, all operation
2
WARNINGS
FAILURE TO COMPLY WITH THESE WARNINGS AND CAUTIONS COULD RESULT IN INSTRUMENT DAMAGE AND POSSIBLY JEOPARDIZE THE WELL BEING OF THE PATIENT AND/OR HEALTH CARE PROFESSIONAL.
»
The sensor of the MAXO2 Oxygen Analyzer (MAXO2) has been tested with various
anesthesia gases including Nitrous oxide, Halothane, Isourane, Enurane, Sevourane and Desurane and found to have acceptable low interferance, the device in its entirety (including electronics) is not suitable for use in the presense of a ammable anesthetic
mixture. Only the threaded sensor face and sample inlet connection may be allowed to contact such a gas mixture.
»
To avoid explosion, do not operate the oxygen analyzer in the presence of ammable
anesthetics or in an atmosphere of explosive gases. Operating the oxygen analyzer in
ammable or explosive atmospheres may result in re or explosion.
»
Never allow an excess length tubing near the patient’s head or neck, as such could result in
strangulation.
»
Before use, all individuals who will be using the MAXO2 must become thoroughly familiar
with the information contained in this Operation Manual. Strict adherance to the operating instructions is necessary for safe effective product performance. This product will perform only as designed if installed and operated in accordance with the manufacturer’s operating instructions.
»
Use only genuine Maxtec accessories and replacement parts. Failure to do so may
seriously impair the monitor’s performance. Repair or alteration of the MAXO2 beyond the scope of the maintenance instructions or by anyone other than an authorized Maxtec service person could cause the product to fail to perform as designed.
»
Calibrate the MAXO2 weekly when in operation and if enviromental conditions change
signicantly. (ie, Temperature, Humidity, Barometric Pressure. --- Refer to Calibration
section of this manual).
»
Use of the MAXO2 near devices that generate electrical elds may cause erratic readings.
»
If the MAXO2 is ever exposed to liquids (from spills or immersion) or to any other physical
abuse, turn the instrument OFF and then ON. This will allow the unit to go through its self test and make sure everything is operating correctly.
»
Never autoclave, immerse or expose the MAXO2 (including sensor) to high temperatures
(>70°C). Never expose the device to pressure, irradiation vacuum, steam, or chemicals.
»
To protect the unit from potential leaky battery damage always remove batteries when
the unit is going to be stored (not in use for 1 month) and replace dead batteries with recognized name brand AA Alkaline batteries. Do not use rechargeable batteries.
»
This device does not contain automatic barometric pressure compensation.
»
Not for use in an MRI environment.
866.4.Maxtec www.maxtecinc.comII
TABLE OF CONTENTS
Classication .................................................I
Warranty ....................................................I
Warnings ...................................................II
Symbol Guide .............................................. IV
1.0 SYSTEM OVERVIEW ......................................1
1.1 Base Unit Description ....................................1
1.2 Components Description ..................................1
1.3 MAX-250 Oxygen Sensor .................................4
2.0 SET-UP PROCEDURE .....................................2
2.1 Battery Installation .......................................2
2.2 Calibrating the MAXO
2.2.1 Before You Begin ...................................3
2.2.2 To Calibrate the MAXO
2.2.3 Automatic Calibration to Room Air ......................5
2.2.4 Factors Inuencing Calibration .........................5
3.0 OPERATION INSTRUCTIONS ...............................6
4.0 SENSOR REMOVAL AND REPLACEMENT.....................6
5.0 PROBLEM SOLVING ......................................7
6.0 CLEANING AND MAINTENANCE ............................7
7.0 SPECIFICATIONS.........................................8
7.1 Base Unit Specications ..................................8
7.2 Sensor Specications ....................................8
®
Analyzer ...........................3
2
®
Analyzer......................3
2
8.0 APPLICATIONS...........................................9
8.1 Exposure to Anesthetic Gases..............................9
8.2 Calibration Techniques in Pressurized Systems ................9
8.3 Calibration Errors ........................................9
8.4 Atmospheres of High Humidity..............................9
9.0 SPARE PARTS AND ACCESSORIES ........................10
866.4.Maxtec www.maxtecinc.com III
Loading...
+ 11 hidden pages