Maxtec MAXO2+A, MAXO2+AE Instructions Manual

Page 1
MAXO
BAT
CAL
®
+
A
2
®
+
A / AE INDUSTRIAL GUIDE & OPERATING INSTRUCTIONS
2
®
+
AE
2
R217M42 Rev.K
Page 2
Maxtec TEL (800) 748.5355 6526 South Cottonwood Street FAX (801) 270.5590 Salt Lake City, Utah 84107 email: sales@maxtec.com USA website: www.maxtec.com
Classification
Protection against electric shock: Internally powered equipment. Protection against water: IPX1 Mode of Operation: Continuous
Do not throw away. Dispose of properly in accordance with local regulations.
WARRANTY
The MAXO2+ Analyzer is designed for medical oxygen delivery equipment and systems. Under normal operating conditions, Maxtec warrants the MAXO defects of workmanship or materials for a period of 2-years from the date of shipment from Maxtec, provided that the unit is properly operated and maintained in accordance with Maxtec’s operating instructions. Based on Maxtec product evaluation, Maxtec’s sole obligation under the foregoing warranty is limited to making replacements, repairs, or issuing credit for equipment found to be defective. This warranty extends only to the buyer purchasing the equipment directly from Maxtec or through Maxtec new equipment.
Maxtec warrants MAXO in material and workmanship for a period of 2-years from Maxtec’s date of shipment in a
MAXO
+ unit. Should a sensor fail prematurely, the replacement sensor is warranted for the
2
remainder of the original sensor warranty period. Routine maintenance items, such as batteries, are excluded from warranty. Maxtec and
any other subsidiaries shall not be liable to the purchaser or other persons for incidental or consequential damages or equipment that has been subject to abuse, misuse, mis­application, alteration, negligence or accident.
These warranties are exclusive and in lieu of all other warranties, expressed or implied,
LQFOXGLQJZDUUDQW\RIPHUFKDQWDELOLW\DQG¿WQHVVIRUDSDUWLFXODUSXUSRVH
+ oxygen sensor in the MAXO2+ Analyzer to be free from defects
2
‘s designated distributors and agents as
+ Analyzer to be free from
2
ii
Page 3
PREFACE
This manual describes the function, operation and maintenance of the
MAXO line of oxygen analyzers and monitors, the MAXO
+
hand-held oxygen analyzer. A member of Maxtec’s MAXO2 analyzer
2
+
utilizes the MAX-250+
2
oxygen sensor and is engineered for long life, maximum reliability and stable perfmance.
NOTE: In order to obtain optimum performance from your MAXO2+ analyzer, all operation and maintenance must be performed in accor­dance with this manual. Please read the manual thoroughly before using the analyzer and do not attempt any repair or procedure that is not described herein. Maxtec cannot warrant any damage resulting from misuse, unauthorized repair or improper maintenance of the instrument.
Thank You
Thank you for your purchase of a Maxtec MAXO appreciate the time and energy you invest in selecting the equipment best suited to your needs. As exchange, we are supplying you with a reliable, high-quality instrument that, with proper care and operation, will provide you with years of exceptional service. We also encourage your comments or sug­gestions as to how our equipment, in any way, can better serve your needs. Please feel free to write, FAX or e-mail us at the address on page ii of this manual c/o the Maxtec Marketing Department. Please visit our website www. maxtec.com for more information on our products and services.
+
oxygen analyzer. We
2
WARNING:
Never allow an excess length of tubing, lanyard, or sensor cable near a per­son’s head or neck, which may result in strangulation.
Before use, all individuals who will be using the MAXO oughly familiar with the information contained in this Operation Manual. Strict adherence to the operating instructions is necessary for safe, effective product performance. This product will perform only as designed if installed and oper­ated in accordance with the manufacturer’s operating instructions.
Use only genuine Maxtec so may seriously impair the analyzer’s performance. Repair or alteration of the MAXO
+
beyond the scope of the maintenance instructions, or by anyone
2
other than an authorized Maxtec fail to perform as designed.
Calibrate the MAXO tions change significantly. (ie. Elevation, Temperature, Pressure, Humidity — refer to Section 4.0 of this manual).
iii
+
must become thor-
2
®
accessories and replacement parts. Failure to do
®
service person, could cause the product to
+
weekly when in operation, or if environmental condi-
2
WWW.MAXTEC.COM
Page 4
Use of the MAXO erratic readings.
+
near devices that generate electrical fields may cause
2
If the MAXO
+
is ever exposed to liquids (from spills or immersion) or to any
2
other physical abuse, turn the instrument OFF and then ON. This will allow the unit to go through its self test to assure everything is operating correctly. You may need to allow the sensor time to dry out.
Never immerse or expose the MAXO
+
(including sensor) to high tempera-
2
tures (>70°C). Never expose the device to pressure, irradiation vacuum, steam, or chemicals.
NOTE: Replace the batteries with recognized high quality AA Alkaline or Lithium batteries.
NOTE: If the unit is going to be stored (not in use for 1 month), we rec­ommend that you remove the batteries to protect the unit from potential battery leakage.
FAILURE TO COMPLY WITH THESE WARNINGS AND CAUTIONS COULD RESULT IN INSTRUMENT DAMAGE AND POSSIBLY JEOPARDIZE THE WELL BEING OF THE USER.
MANUFACTURED BY MAXTEC
iv
Page 5
Safety Labeling
The following symbols and safety labels are found on the MAXO
Follow instructions for use.
On/off Button
+
:
2
Warning
Meets ETL standards Do not throw away. Follow local
guidelines for disposal. Calibration required
LOT
Lot code/Batch code Serial Number
SN
Calibration Button

Low Battery
Percent
%
Manufacturer
Catalog Number
REF
Do Not
v
WWW.MAXTEC.COM
Page 6
TABLE OF CONTENTS
Classification .....................................................................ii
Warranty ........................................................................ii
Warnings .......................................................................iii
Safety Symbols ...................................................................v
1.0 SYSTEM OVERVIEW ..........................................................1
1.1 Base Unit Description ......................................................1
1.2 Component Identification ....................................................2
1.3 Component Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
1.4 MAX-250 Oxygen Sensor ...................................................3
2.0 OPERATING INSTRUCTIONS ...................................................3
2.1 Getting Started............................................................3
2.1.1 Protect Tape........................................................3
2.1.2 Automatic Calibration.................................................3
2.2 Calibrating the MAXO
2.3 Operation with OxyKnob Adapter..............................................5
2.4 Operation with BC Adapter ..................................................6
3.0 GAS SAMPLE OPTIONS .......................................................7
4.0 FACTORS INFLUENCING ACCURATE READINGS ..................................8
4.1 Elevation Changes.........................................................8
4.2 Temperature Effects........................................................8
4.3 Pressure Effects...........................................................8
4.4 Humidity Effects...........................................................8
5.0 CALIBRATION ERRORS AND ERROR CODES.....................................9
6.0 CHANGING THE BATTERIES ..................................................10
7.0 CHANGING THE OXYGEN SENSOR.............................................11
7.1 MAXO
7.2 MAXO
+
A Model .........................................................11
2
+
AE Model ........................................................12
2
8.0 CLEANING AND MAINTENANCE ...............................................12
9.0 SPECIFICATIONS ............................................................13
9.1 Base Unit Specifications ...................................................13
9.2 Sensor Specifications......................................................14
10.0 MAXO
+
SPARE PARTS AND ACCESSORIES.....................................14
2
10.1 Included With Your Unit....................................................14
10.2 Standard Replacement Parts ................................................14
10.3 Optional Accessories ......................................................14
10.3.1 Optional Adapters ...................................................14
10.3.2 Mounting Options ...................................................15
10.3.3 Carrying Options ....................................................15
+
Oxygen Analyzer.......................................4
2
MANUFACTURED BY MAXTEC
vi
Page 7
1.0 SYSTEM OVERVIEW
1.1 Base Unit Description
The MAXO
+
analyzer provides unparalleled performance and reliability
2
due to an advanced design that includes the following features and oper­ational benefits.
Extra-life oxygen sensor of approximately 1,500,000 O
2 percent hours
(2-year warranty)
Durable, compact design that permits comfortable, hand-held operation and easy to clean.
Operation using only two AA Alkaline batteries (2 x 1.5 Volts) for
approximately 5000 hours of performance with continuous use. For extra extended long life, two AA Lithium batteries may be used.
Oxygen-specific, galvanic sensor that achieves 90% of final value in approximately 15 seconds at room temperature.
Large, easy-to-read, 3 1/2-digit LCD display for readings in the 0-100% range.
Simple operation and easy one-key calibration.
Self-diagnostic check of analog and microprocessor circuitry.
Low battery indication.
Calibration reminder timer that alerts the operator, using a calibration icon on the LCD display, to perform a unit calibration.
1
WWW.MAXTEC.COM
Page 8
1.2 Component Identification
FIGURE 1
1.3 Component Description
1. 3 1/2-Digit Display - The 3 1/2 digit liquid crystal display (LCD) provides direct readout of oxygen concentrations in the range of 0 - 105.0% (100.1% - 105.0% used for calibration determination purposes). The digits also display error codes and calibration codes as necessary.
2. Low Battery Indicator - The low battery indicator is located at the top of the display and is only activated when the voltage on the bat­teries is below a normal operating level.
3. “%” symbol - The “%” sign is located to the right of the concentration number and is present during normal operation.
4. Calibration symbol - The calibration symbol is located at the bottom of the display and is timed to activate when a calibration is necessary.
5. ON/OFF Key - This key is used to turn the device on or off.
6. Calibration Key - This key is used to calibrate the device. Holdi n g t h e key for more than three seconds will force the device to enter a calibration mode.
CAUTION: The device will assume a percent oxygen concentration
when calibrating. Be sure to apply 100% oxygen, or ambient air concentration to the device during calibration or the device will not calibrate correctly.
Sample Inlet Connection
This is the port at which the device is connected to determine oxygen concen­tration.
MANUFACTURED BY MAXTEC
2
Page 9
1.4 MAX-250+ Oxygen Sensor
The MAX-250+ oxygen sensor offers stability and extra life. The MAX-250+ is a galvanic, partial pressure sensor that is specific to oxygen. It consists of two electrodes (a cathode and an anode), a teflon membrane and an electrolyte. Oxygen diffuses through the teflon membrane and immediately reacts at a gold cathode. Concurrently, oxidation occurs electrochemically at the lead anode, generating an electrical current and providing a voltage output. Electrodes are immersed in a unique gelled weak acid electrolyte which is responsible for the sensors long life and motion insensitive characteristic. Since the sensor is specific to oxygen, the current generated is proportional to the amount of oxygen present in the sample gas. When no oxygen is present, there is no electrochemical reaction and there­fore, negligible current is produced. In this sense, the sensor is self-zeroing.
CAUTION: The Maxtec MAX-250+ oxygen sensor is a sealed device containing a mild acid electrolyte, lead (Pb), and lead acetate. Lead and lead acetate are hazardous waste constituents and should be disposed of properly, or returned to Maxtec for proper disposal or recovery.
CAUTION: Do not immerse the sensor in any cleaning solution,
autoclave or expose the sensor to high temperatures.
CAUTION: Dropping sensor can adversely affect its performance. Do not throw away. Dispose of properly in accordance with local
regulations.
2.0 OPERATING INSTRUCTIONS
2.1 Getting Started
2.1.1 Protect Tape
Prior to turning on the unit, a protective film covering the threaded sensor
face must be removed. After removing the film, wait approximately 20 minutes for the sensor to reach equilibrium.
2.1.2 Automatic Calibration
After the unit is turned on it will automatically calibrate to room air. The
display should be stable and reading 20.9%. To check the oxygen concen­tration of a sample gas: (after the unit has been calibrated).
3
WWW.MAXTEC.COM
Page 10
1. Connect the Tygon tubing to the bottom of the analyzer by threading the barbed adapter onto the oxygen sensor. (figure 2)
2. Attach the other end of the sample hose to the sample gas source and initiate flow of the sample to the unit at a rate of 1-10 liters per minute (2 liters per minute is recommended).
3. Using the “ON/OFF” key, make sure the
unit is in the power “ON” mode.
4. Allow the oxygen reading to stabilize. This will normally take about 30 seconds or more.
®
2.2 Calibrating the MAXO
+
Oxygen Analyzer
2
FIGURE 2
The MAXO
+
Analyzer should be calibrated upon initial power-up. Thereafter,
2
Maxtec® recommends calibration on a weekly basis. To serve as a reminder, a one week timer is started with each new calibration. At the end of one week a reminder icon “ ” will appear on the bottom of the LCD. Calibration is recommended if the user is unsure when the last calibration procedure was
performed, or if the measurement value is in question. Start calibration by pressing the button for more than 3 seconds. The
+
MAXO
will automatically detect if you are calibrating with 100% or 20.9%
2
oxygen (normal air). Do not attempt to calibrate to any other concentration.
100% O
• The measured O
• The measured O2 percentage in 100% O2 is above 101.0% O
verification, new calibration is required when:
2
percentage in 100% O2 is below 99.0% O
2
2.
2.
• The CAL reminder Icon is blinking at the bottom of the LCD.
If you are unsure about the displayed O2 percentage. (see Factors
influencing accurate readings.)
Compressed air (20.9% O
The measured O
The measured O2 percentage in 20.9% O2 is above 21.9% O
), new calibration is required when:
2
percentage in 20.9% O2 is below 19.9% O
2
2.
2.
The CAL reminder Icon is blinking at the bottom of the LCD.
If you are unsure about the displayed O
percentage.
2
(see Factors influencing accurate readings.)
Calibrate with the same type of circuit and flow that you will use when taking your readings.
MANUFACTURED BY MAXTEC.
4
Page 11
A simple calibration may be made with the sensor open to static Ambient air. For optimum accuracy Maxtec recommends that the Sensor be placed in a closed loop circuit where gas flow is moving across the sensor in a controlled manner.
2.3 Operation with Oxyknob Adapter (refer to figure below)
1. Attach the Oxyknob adaptor to the threaded sensor of your Maxtec
Analyzer.
2. Hand tighten until sensor o-ring seals against the Oxyknob.
3. Slowly crack open the tank. Listen for a slight hissing sound.
4. Hold the Oxyknob up to the outlet of the tank.
5. Align the small hole in the bottom or side of the Oxyknob to the hole in the
outlet of the tank.
6. Hold the analyzer in place untile the reading has stabilized on the analyzer display. (about 15 seconds or less)
7. Remove Oxyknob after reading and close valve on the tank.
WARNING:
High gas pressures may cause damage to oxygen sensor and result in incor­rect readings. Turn down the tank valve as much as possible to avoid sensor damage.
Do not cover the larger holes on the Oxyknob. Cold gas exiting a pressurized tank can cause the sensor reading to drift
if exposed for an extended period of time. Allow sensor to warm between readings if necessary.
5
WWW.MAXTEC.COM
Page 12
2.4 Operation with Optional BC Adapter (refer to figure below)
1. Attach the included Barbed Adapter to the MAXO
+
by threading it on to
2
the bottom of the sensor.
2. Connect the included tubing to the barbed adapter.
3. Attach the BC adapter (sold separately) to the other end of the tube.
4. Connect the inflator hose on the regulator to the BC adapter and tube.
5. If the MAXO
+
is not already turned on, do so now by pressing the
2
analyzer “ON” button.
6. Initiate flow of nitrox to the unit to allow the gas to saturate the sensor. The
BC adapter will regulate the optimum flow and pressure. Although a stable value is usually observed within 30 seconds, allow at least two minutes to ensure that the sensor is completely saturated with the gas.
7. The analyzer will now look for a stable sensor signal and a good reading. When obtained, the analyzer will display the oxygen percentage on the LCD.
MANUFACTURED BY MAXTEC
Inflator Hose
FIGURE 3
Nitrox Tank
6
Page 13
3.0 GAS SAMPLE OPTIONS
7
WWW.MAXTEC.COM
Page 14
4.0 FACTORS INFLUENCING ACCURATE READINGS
4.1 Elevation Changes
• Changes in elevation result in a reading error of approximately 1% of reading per 250 feet.
• In general, calibration of the instrument should be performed when elevation at which the product is being used changes by more than 500 feet.
4.2 Temperature Effects
The MAXO
+
will hold calibration and read correctly within ±3% when in ther-
2
mal equilibrium within the operating temperature range. The device must be thermally stable when calibrated and allowed to thermally stabilize after experiencing temperature changes before readings are accurate. For these reasons, the following is recommended:
• For best results, perform the calibration procedure at a temperature close to the temperature where analysis will occur.
• Allow adequate time for the sensor to equilibrate to a new ambient temperature.
CAUTION: “CAL Err St” may result from a sensor that has not reached thermal equilibrium.
4.3 Pressure Effects
Readings from the MAXO gen. The partial pressure is equal to the concentration times the absolute pressure. Thus, the readings are proportional to the concentration if the pressure is held constant. Therefore, the following are recommended:
• Calibrate the MAXO
• If sample gases flow through tubing, use the same apparatus and flow rates when calibrating as when measuring.
• The MAXO
+
oxygen sensor has been tested at pressures up to two
2
atmospheres absolute. Calibration or operation above this pressure is beyond the intended use.
+
are proportional to the partial pressure of oxy-
2
+
at the same pressure as the sample gas.
2
4.4 Humidity Effects
Humidity (non-condensing) has no effect on the performance of the MAXO
MANUFACTURED BY MAXTEC
+
2
8
Page 15
other than diluting the gas, as long as there is no condensation. Depending on the humidity, the gas may be diluted by as much as 4%, which proportion­ally reduces the oxygen concentration. The device responds to the actual oxygen concentration rather than the dry concentration. Environments where condensation may occur are to be avoided since moisture may obstruct passage of gas to the sensing surface, resulting in erroneous readings and slower response time. For this reason, the following is recommended:
• Avoid usage in environments greater than 95% relative humidity.
HELPFUL HINT: Dry sensor by lightly shaking moisture out, or flow a dry gas at two liters per minute across the sensor membrane.
5.0 CALIBRATION ERRORS AND ERROR CODES
The MAXO
+
analyzers have a self test feature built into the software to
2
detect faulty calibrations, oxygen sensor failures, and low operating voltage. These are listed below, and include possible actions to take, if an error code occurs.
E02: No sensor attached
+
MAXO
A: Open unit and disconnect and reconnect sensor. Unit should
2
perform an auto calibration and should read 20.9%. If not, contact Maxtec Customer Service for possible sensor replacement.
+
MAXO
AE: Disconnect and reconnect external sensor. Unit should perform
2
an auto calibration, and should read 20.9%. If not, contact Maxtec Customer Service for possible sensor replacement or cable replacement.
E03: No valid calibration data available
Make sure unit has reached thermal equilibrium. Press and hold the Calibration Button for three seconds to manually force a new calibration.
E04: Battery below minimum operating voltage
Replace batteries.
CAL Err St: O2 Sensor reading not stable
Wait for displayed oxygen reading to stabilize, when calibrating the device at 100% oxygen.
Wait for unit to reach thermal equilibrium (Please note that this can take up to one half hour, if the device is stored in temperatures outside the specified operating temperature range).
CAL Err lo: Sensor voltage too low
Press and hold the Calibration Button for three seconds to manually force a new calibration. If unit repeats this error more than three times,
9
WWW.MAXTEC.COM
Page 16
contact Maxtec Customer Service for possible sensor replacement.
CAL Err hi: Sensor voltage too high
Press and hold the Calibration Button for three seconds to manually force a new calibration. If unit repeats this error more than three times, con­tact Maxtec Customer Service for possible sensor replacement.
CAL Err Bat: Battery voltage too low to recalibrate
Replace batteries
6.0 CHANGING THE BATTERIES
Batteries should be changed by service personnel.
Use only brand name batteries.
Replace with two AA batteries and insert per orientation marked on the
device.
Should the batteries require changing, the device will indicate this in one of two ways:
The battery icon on the bottom of the display will begin to flash. This
icon will continue to flash until the batteries are changed. The unit will continue to function normally for approx. 200 hours.
If the device detects a very low battery level, an error code of “E04” will
be present on the display, and the unit will not function until the batter­ies are changed.
To change the batteries, begin by removing the three screws from the back of the device. A #1 phillips screwdriver is required to remove these screws.
Once the screws are removed, gently separate the two halves of the device.
The batteries can now be replaced from the back half of the case. Be sure to orient the new batteries as indicated in the embossed polarity on the back case.
NOTE: If the batteries are installed incorrectly the batteries will not make contact and the device will not operate.
Carefully, bring the two halves of the case together while
MANUFACTURED BY MAXTEC
FIGURE 4
10
Page 17
positioning the wires so they are not pinched between the two case halves. The gasket separating the halves will be captured on the back case half.
Reinsert the three screws and tighten until the screws are snug. (figure 4) The device will automatically perform a calibration and begin displaying % of
oxgen.
HELPFUL HINT: If unit does not function, verify that the screws are tight to allow proper electrical connection.
HELPFUL HINT (MAXO2+ AE): Before closing the two case halves together, verify that the keyed slot on top of the coiled cable assembly is engaged on the small tab located on the back case. This is designed to position the assembly in the correct orientation and prevent it from rotating. Improper positioning could hinder the case halves from closing and prevent operation when tightening the screws.
7.0 CHANGING THE OXYGEN SENSOR
7.1 MAXO
®
+A Model
2
Should the oxygen sensor require changing, the device will indicate this by presenting “Cal Err lo” on the display after initiating a calibration.
To change the oxygen sensor, begin by
Lock Lever
removing the three screws from the back of the device. A #1 Phillips screwdriver is required to remove these screws.
Once the screws are removed, gently separate
FIGURE 5
the two halves of the device. Disconnect the oxygen sensor from the printed circuit board by pressing the
un-lock lever first and then pulling the connector out of the receptacle
.
The oxygen sensor can now be replaced from the back half of the case.
HELPFUL HINT: Be sure to orient the new sensor by aligning the red arrow on the sensor with the arrow in the back case. A small tab is located on the back case that is designed to engage the sensor and pre­vent it from rotating within the case. (figure 5)
NOTE: If the oxygen sensor is installed incorrectly, the case will not come back together and the unit may be damaged when the screws are reinstalled.
11
WWW.MAXTEC.COM
Page 18
Note: if the new sensor has red tape over the outside, remove it and wait 30 minutes before calibrating.
Reconnect the oxygen sensor to the connector on the printed circuit board. Carefully bring the two halves of the case together while positioning the wires
to ensure they are not pinched between the two case halves. Make sure the sensor is fully inserted and in the proper orientation.
Reinsert the three screws and tighten until the screws are snug. Verify the unit operates properly.
The device will automatically perform a calibration and begin displaying % of oxygen.
7.2 MAXO2+AE Model
Should the oxygen sensor require changing, the device will indicate this by presenting “Cal Err lo” on the display.
Unthread the sensor from the cable by rotating the thumbscrew connector counterclockwise and pull the sensor from the connection.
Replace the new sensor by inserting the electrical plug from the coiled cord into the receptacle on the oxygen sensor. Rotate the thumbscrew clockwise
until snug. The device will automatically perform a calibration and begin displaying % of oxy-
gen.
8.0 CLEANING AND MAINTENANCE
Store the MAXO ment of daily use.
The instruction given below describes the methods to clean the instrument, sensor and its accessories:
+
analyzer in a temperature similar to its ambient environ-
2
Instrument
When cleaning or disinfecting the exterior of the MAXO
appropriate care to prevent any solution from entering the instrument. Do not immerse unit in fluids.
+
analyzer, take
2
Oxygen Sensor
Clean the sensor with a cloth moistened with a 65% alcohol/water
solution.
Maxtec does not recommend use of spray disinfectants because they can
MANUFACTURED BY MAXTEC
12
Page 19
contain salt, which can accumulate in the sensor membrane and impair
readings.
Accessories
The threaded barbed adapter may be cleaned by washing them with a 65% alcohol/water solution (per manufacturer’s instructions). The parts must be thoroughly dry before they are used.
Because of the variability of the cleaning processes, Maxtec cannot provide specific instructions. Therefore, we highly recommend referring to the manu­facturer’s instructions on the details of method.
9.0 SPECIFICATIONS
9.1 Base Unit Specifications
Measurement Range: 0-100% Resolution: 0.1% Accuracy and Linearity: 1% of full scale at constant temperature, R.H. and
pressure when calibrated at full scale.
Total Accuracy: ±3% actual oxygen level over full operating
temperature range.
Response Time: 90% of final value in approximately 15 seconds at
23°C. Warm-up Time: none required Operating Temperature: 15°C - 40°C (59°F - 104°F) Storage Temperature: -15°C - 50°C (5°F - 122°F) Humidity: 0-95% (non-condensing) Power Requirements: 2, AA Alkaline batteries (2 x 1.5 Volts) Battery Life: approximately 5000 hours with continuous use Low Battery Indication: “BAT” icon displayed on LCD Sensor Type: Maxtec MAX-250 galvanic fuel cell Expected Sensor Life: >1,500,000 O
typical applications A Model Dimensions: 3.0”(W) x 4.0”(H) x 1.5”(D) [76mm x 102mm x
38mm] A Weight: 0.4 lbs. (170g) AE Model Dimensions: 3.0”(W) x 36.0”(H) x 1.5”(D) [76mm x 914mm x
38mm] Height includes external cable length (retracted). AE Weight: 0.6 lbs. (285g)
percent hours minimum 2-years in
2
13
WWW.MAXTEC.COM
Page 20
9.2 Sensor Specifications
Type: Galvanic fuel sensor (0-100%) Life: 2-years in typical applications
10.0 MAXO
+
SPARE PARTS AND ACCESSORIES
2
10.1 Included With Your Unit
Part Number Item
R217M42 Industrial Guide and Operating Instructions RP76P06 Lanyard R207P17 Threaded Adapter with Tygon Tubing R110P10 Flow Diverter R218P19 Oxyknob Adapter
R217P35 Dovetail Bracket X
A Model AE Model
X X X X X X X X
10.2 Standard Replacement Parts
Part Number Item
R125P02-013 MAX-250+ Internal Oxygen Sensor R125P03-004 MAX-250E External Oxygen Sensor R115P85 MAX-250ESF External Oxygen Sensor R217P08 Gasket R217P09-002 Overlay R217P11-001 Back Assembly R217P16-002 Front Assembly (Includes Board and LCD) R217P24 Coiled Cable Assembly RP06P25 #4-40 Pan Head Stainless Steel Screw
X X X X X X X X X X X X X X
BAT
CAL
10.3 Optional Accessories
10.3.1 Optional Adapters
Part Number Item
RP16P02 “T” Adapter (15mm I.D.) RP16P09 Scuba Pro Adapter R101P01 Flow Thru Head RP11P28 BC Adapter
MANUFACTURED BY MAXTEC.
14
Page 21
10.3.2 Mounting Options (requires dovetail-R217P23)
Part Number Item
R103P01 M16 x 1 Maxtec Nut R205P86 Wall Mount R213P31 Swivel Mount
10.3.3 Carrying Options
Part Number Item
R213P02 Zipper Carrying Case with Shoulder Strap R213P56 Water Tight Carrying Case, Large (9” x 11”) R213P90-001 Water Tight Carrying Case, Small (8” x 4”) R217P22 Belt Clip and Pin Assembly R217P32 Soft Case, Tight Fit Carrying Case
Equipment in need of factory repair shall be sent to: Maxtec
Service Department 6526 South Cottonwood Street Salt Lake City, Ut 84107
15
WWW.MAXTEC.COM