1. How frequently should the CHDF GR500 eluant be replaced?
The CHDF GR500 eluant degrades with time. It should be replaced every two to three days. Check the
conductivity of the eluant in order to determine if it is suitable. The conductivity ranges for the
different GR500 types are as follows (in microSiemens per centimeter): 1X-GR500: 7-10; 2X-GR500:
22-27; 2X-GR500-LA: 14-18; 1X-GR500 pH9: 13-18. The GR500 eluant absorbs Carbon Dioxide
from the ambient which causes its pH to drop, and its conductivity to increase; this results in
measured particle sizes smaller than expected. The 1X-GR500 pH9 eluant is especially sensitive to
CO2 absorption. Bacterial growth also takes place as the GR500 ages.
2. I obtain smaller than expected CHDF particle size results when I inject calibration standards.
Most likely, your GR500 eluant is either degraded or contaminated. See Question 1 above.
3. How is the GR500 eluant prepared?
Add 950 ml of deionized water (conductivity should be less than 2 micro Siemens per centimeter) to 50 ml
of GR500 concentrate. Eluant degassing by either sonication or vacuum filtration is recommended.
4. How long does the GR500 concentrate last, and how should it be stored?
The GR500 lasts about one month. It should be kept refrigerated.
5. Does the Marker fluid need to be injected at an exact time for every run?
No. In both manual, and auto-sampler operation, the software detects when the marker is injected. The
injection time is recorded; the marker elution time is calculated as the marker exit time minus the
injection time. A marker delay time of zero seconds is recommended for auto-sampler operation (this
is set under the current method file).
6. How should the marker fluid be prepared?
The marker fluid consists of a solution of sodium benzoate in deionized water. Its concentration varies
depending on the type of fractionation cartridge used. Typical concentrations are in the range 0.1% to
0.5%. This concentration should be adjusted in order to obtain a reasonable marker peak height
(displayed on the CHDF software Raw Data graph) upon marker injection into the CHDF. Marker
peaks heights of 10 to 40 mAU’s (milli Absorbance Units) for the CHDF-2000 instrument, and 5,000
to 15,000 counts for the CHDF-1100 are acceptable.
7. Why is the pump pressure fluctuating more than 200 psi during a run?
Air bubbles are lodged in the pump pistons. Stop the pump; open the pump purge valve (located on the
lower right- hand corner of the CHDF-2000; or on the CHDF-1100 pump front panel); insert a 5-ml
plastic syringe into the purge valve, and slowly draw fluid. Air bubbles in the pump and pump-inlet
tubing should be removed. Tighten the purge valve and start the pump. If the air bubbles re-appear,
the Whatman de-gassing filter on the pump-inlet tubing should be replaced. This filter lasts about one
month. Cavitation, which results in air bubble generation, occurs if the Whatman filter needs
replacement. For the CHDF-2000, this filter is optional. You may operate your CHDF-2000 without
this filter if you so prefer.
8. Noise and drift are present in the Raw Data baseline.
This is probably due to a defective Whatman filter. Replace this filter and observe the baseline. This filter
must be filled with eluant when in operation. Use the purge knob on the filter is order to draw any air
from the filter. Another possibility is that the UV lamp may need replacement if it has been used for
more than 1000 hours (see the troubleshooting section of the hardware manual).
9. How should samples be prepared for CHDF analysis?
Only aqueous samples can be analyzed on the current CHDF setup. Dry powders must be dispersed in
GR500. Sonication should be applied in order to help suspend the particles in the liquid. Concentrated
samples should be diluted in GR500 eluant. The solids contents are typically in the range 0.1% to 5%
by volume. UV-absorbing samples can be analyzed at lower concentrations while non-UV absorbing
samples (such as colloidal silica) must be run at higher concentrations. Sonication may be used in
order to break-up particle aggregates. Reasonable sample concentrations should produce peak heights
of 20 to 50 mAU’s for the CHDF-2000, and 5,000 to 15,000 counts for the CHDF-1100. A syringe
filter may be used on the sample syringe in order to remove aggregates from samples prior to
injection. Syringe filters may only be used once per sample. Marker, as well as, Rinse filters may be
reused until leaks are observed on the filters.
10. No peaks are detected in the Raw Data graph upon sample, and marker injection.
The fractionation cartridge is likely plugged. Refer to the troubleshooting section of the hardware manual
for cartridge cleaning instructions.
11. The pump pressure is higher than normal.
The filter element in the cartridge in-line filter must be either cleaned or replaced. Refer to the
troubleshooting section of the hardware manual. The CHDF-1100 is affected by ambient temperature.
Low ambient temperatures result in high CHDF-1100 pressures.
12. What Noise Level should be used under Configure System in the CHDF software?
30-40 for the CHDF-2000; 400 for the CHDF-1100.
13. Can any wavelength be used on the CHDF UV-detector?
The CHDF UV-detector wavelength can be varied from 190 to 380 nm. However, the CHDF PC software
only contains particle extinction cross section curves for 200 nm, and 220 nm. One of these values
must be selected in the active Method file. Inaccurate results are obtained if the UV-detector, and the
software wavelengths differ.
14. What is the sharp peak in the Raw Data between the sample peak and the marker peak?
This peak corresponds to UV-absorbing molecular species present in the sample. Molecular species
include surfactants, electrolytes, acids, or bases. These molecules have a smaller size than the particles
themselves, and therefore exit the capillary after the particles.
15. A broad peak after the marker peak is displayed in the Raw Data.
This peak is likely due to non-ionic surfactant. Non-ionic surfactants can continuously adsorb and desorb
from the capillary wall as they travel down the fractionation capillary. This adsorption/desorption
retards the non-ionic surfactant from exiting the capillary.
16. Can the marker be blended with the sample so that only one injection is made per sample?
Yes, in manual operation only. Simply add maker fluid to your samples, and make only the sample
injection (do not inject the marker).
17. How long should the instrument be allowed to equilibrate (or warm up) at the beginning of the day
before starting sample analysis?
Approximately 30 to 45 minutes.
18. Can the CHDF be turned off at the end of the day, or should it be left running overnight?
The CHDF can be operated in either mode. We prefer to turn it off at the end of the day.
19. How much sample, and marker, should be Loaded into the injection valve?
100 microliters for manual operation. It is recommended that at least 400 microliters be drawn into the
syringe in order to minimize the chances of loading air bubbles into the injection valve. If using a
syringe filter make sure the filter is filled with sample before loading sample into the injection valve.
80-90 microliters is recommended for auto-sampler operation. This is set under System/Configure
Auto-Sampler.
20. Are samples loaded directly into the fractionation capillary?
No. They are loaded into a sample loop (a piece of stainless steel tubing) which is connected to a vent. As
sample is pushed into the injection valve, the sample loop becomes filled with sample. Sample exits
from the sample loop vent as more sample is loaded. The sample loop becomes connected to the
pump, and the fractionation capillary upon switching to Inject.
21. Can the CHDF software be run under Windows9X, NT, and 2000?
Yes, the CHDF2000 software can be run on any Windows operating system software.
22. What solids concentration should be used for calibration standards?
Calibration standards should not be overly concentrated. Solids concentrations should be such that Raw
Data peak heights are 10 to 30 mAU’s for calibration standards and 20-60 for regular samples.
Calibration standards larger than 300 nm should be sonicated prior to use in order to break up
aggregates.
23. How frequently should the CHDF capillary be calibrated?
Calibration shifts are related mainly to two factors: (i) changes in GR500 eluant composition; and (ii)
changes in ambient temperature (not important for the CHDF-2000). It is recommended that the
conductivity of the GR500 eluant be monitored; replace the eluant if the conductivity is outside its
normal range. Calibration standards can be run prior to regular sample analysis in order to confirm
calibration accuracy. These “control” standards may be added to the calibration curve if re-calibration
is deemed necessary.
24. How long does a CHDF cartridge (capillary) last?
Fractionation capillaries do not have an exact life span. Cartridges may last several years since no particle
adsorption occurs inside the capillary. Capillary failures are frequently due to plugging right at the
capillary inlet. Cutting-off (“snipping”)a short piece at the capillary inlet frequently removes the plug.
The capillary can also be back-flushed as described in the hardware manual, troubleshooting section.
25. How is the Weight Particle Size Distribution (PSD) calculated?
The particle weight percent is calculated as the Number percent multiplied by the particle diameter raised
to the third power: Wt1 %= Ni%(Dp)3. This assumes that the particle density is the same for all
particles in the sample. The area percent is calculated as follows: Ai% = Ni%(Dp)2
26. What particle Size Reproducibility should be expected from CHDF
About ± 5%.
27. What is the mean, 25%, median, 75%, and FWHH in the text PSD report?
The 25%, median, and 75% are the 25th, 50th, and 75th percentiles of the particle size distribution (PSD).
They are calculated as the particle size evaluated at those values of the cumulative PSD. Physically,
for the 25% for example, they indicate that 25% of the particles are below the reported particle size.
The FWHH is the PSD Full Width at Half Height. It is the width of the differential PSD at 50%
height.
28. What is the list of particle sizes at the bottom of the Text Report?
This list contains a description of all the particle size populations (also referred to as modes) automatically
detected by the software in the sample. Each mode is listed by the start and end particle size. The
percentages by Number, Area, and Weight are printed. The Area% is the average particle size of each
mode (Weight-Weighted). The mode detection sensitivity can be changed in the current method under
Optical Mode Detector. Lowering the value of Slope Sensitivity increases mode sensitivity.
29. How can the Particle Size Distribution Report be Customized or Changed?
Open your current CHDF method, and select “Report Setup”. The text report, if used, should be
placed at the bottom of the report.
30. What causes noticeable “tailing” in a fractogram?
The inlet to the fractionation capillary is likely dirty. Snip the fractionation (top) capillary (see
hardware manual section 6.6).
31. How are odd-shape particles, e.g., discs, rods analyzed?
Particles spin in the capillary tube during CHDF fractionation. As a result, particles approach the
capillary wall at different orientations. Some positions result in an exclusion layer equal to the longest
particle dimension while other positions do the shortest or intermediate dimensions. The resulting
elution time is equivalent to an average of all the particle dimensions. Research done at Lehigh
University showed that particle doublets generated from partial latex coagulation produce a CHDF
particle size equal to one particle diameter multiplied by the square root of two, or 1.41*Dp (instead
of 1.5*Dp). Triplets appear to produce one diameter times the square root of 3.
32. Does the CHDF software extrapolate beyond the two calibration standards at the limits of the
calibration curve?
The CHDF reports particle size populations in the range specified in the CHDF2000 software under
system/Configure-System. On the small particle size end, the CHDF interpolates below the smallest
standard and the data point Rf=1, Dp=0, where Rf is the separation factor for a near-zero particle size
(Dp) species such as the marker fluid. In fact, if you run the marker as a sample, you obtain Rf=1. Rf
is calculated as the particle elution time divided by the marker elution time. Rf increases with
increasing particle size as shown in the calibration curve. Elution time is peak exit time minus
injection time. For larger particles, the software does extrapolate using a curve-fitting expression of
the calibration-standard data points (up to the upper limit specified under System/Configure-System.
This curve fit can be a sigmoidal or polynomial fit. Both options have sigmoidal shapes.
33. What is Particle Frequency in the PSD graphs?
"Particle Frequency" and "Volume Percentage" are different. The CHDF produces relative PSD's.
An absolute PSD shows the actual particle volume percentage. However, absolute PSD's are
sometimes difficult to compare between different samples (overlay). The reason is that samples with
different percent solids will show different PSD-peak heights.
Relative PSD's as currently shown by the CHDF are normalized so that the tallest peak or peaks have
a height of 100 "frequency units". These are not percentage values because there can be two or more
peaks with the same 100 value. Also, next to the 100 point, there are values of 98, 92, 86, and so on. If
you add all these together, you would obtain much more than 100% for the sample.
The particle frequency can be used to compare two populations. For example, a 37% peak has half the
percentage of a 74% peak.
If you need to know the percentage of a population, you should use the cumulative distribution (CD).
You can subtract the CD values at both ends of each peak. This gives you the percentage of this peak.
See example graph below.
The CHDF 2000 (Figure 1-1) uses the principle of Capillary Hydrodynamic Fractionation
(CHDF) to measure particle sizes from 15 nanometers to 1.1 microns in diameter.
Figure 1-1 CHDF 2000
The CHDF 2000 provides the following features
• Simple front panel control of all system operations
The CHDF 2000 control panel (Figure 1-2) provides a simple icon-based interface that allows
you to directly configure, operate, and troubleshoot the CHDF 2000 . The control panel is located
on the CHDF 2000 front door. Refer to Section 1.3, Using the Control Panel, for detailed
instructions on using the CHDF 2000 control panel.
The CHDF 2000 manual injector valve (see Figure 1-2) provides a port for loading a sample and a
manual rotation handle to inject the sample into the eluant stream for analysis. The injector valve
is mounted to the right of the front door.
An autosampler option is also available for automated sample injection.
Eluant Pump
The base system pump supplies the eluant for CHDF 2000 operation. You can access the pump behind the
front door (Figure 1-3). A priming valve is connected to the pump outlet and is mounted directly below
the manual injector valve.
Outlet
check
valves
Figure 1-3 Eluant Pump and Fractionation Cartridge
Fractionation Cartridge
In the CHDF2000, a capillary column provides hydrodynamic separation of the sample
Instructions for installing the column are provided in Chapter 6.
Pressure
xducer
Access hole for
offset-fine-
Absorbance Detector
The CHDF 2000 includes a multi-wavelength, tunable, ultraviolet (UV) absorbance detector
(Figure 1-4). The UV detector measures the Ultraviolet light extinction of particles in the flow
stream from the column outlet.
A 9 pin, RS-232 connection is used to provide data transfer between the CHDF 2000 and a
personal computer. A Pentium processor and Windows ‘95 are recommended for the
personal computer.
Signal Connector Module
The signal connector module (Figure 1.5) on the CHDF 2000 rear panel provides signal
connections for external devices such as a chart recorder or integrator. You can also use these
signals to synchronize the operation of the CHDF 2000 with other devices.
System controls are just below the control panel display on the CHDF 2000 front door.
Table 1-1 identifies the function of each CHDF 2000 control.
Table 1-1 Control Panel
Control Function Description
0 – 9 Number values Used in combination to insert a value for
. Decimal point Inserts a decimal point in a value for a
CL Clear Clears an entered value.
OK Execute Selects a highlighted option. Alternative
∅AU
Home Go Home Returns the control panel display to the
Selector knob Rotate to change highlighted screen
Autozero Detector Resets the absorbance detector output
Go Back Returns the control panel display back to
I-16
a selected parameter.
parameter.
to pressing the selector knob to select a
highlighted option.
signal to the zero offset value.
Home screen.
the preceding program screen.
option or to increase or decrease a value.
Press to select a highlighted option or to
accept a displayed value.
1.3.3 Moving Within and Between Screens
The default option for a screen is highlighted when the screen first appears.
Moving Among Screen Options
To highlight an option on the screen, rotate the selector knob to move the cursor from screen
option to screen option. As you move the cursor to a new screen option, the option appears
highlighted.
For example, when you first access the Home screen, the Analyze/Monitor screen option appears
highlighted (see Figure 1.8). To highlight the Configure System option on the Home screen,
rotate the top of the selector knob to the right two times. The Configure System option appears
highlighted.
Moving Between Screens
When you choose a screen option, the control panel display shows the screen for the option you
chose. The default screen option for the new screen appears highlighted.
The CHDF 2000 determines the particle size distribution of samples suspended in aqueous solutions in the
range of 15 to 1100 nm in diameter. The size distribution is determined by the measurement of the
particle elution time in a series of capillary tubes.
The CHDF 2000 requires an IBM compatible computer, preferably equipped with a Pentium processor
and Windows ‘9X or higher. The operating software gives graphical displays and a hard copy of the
cumulative weight, area, and number distributions, the relative weight, area and number distribution, the
weight, volume, and number average particle size, and the particle size distribution standard deviation.
The measurement time, from sample injection to final calculation, is under fifteen minutes.
2.1 SCOPE
The purpose of this chapter is to provide an overall description of Capillary Hydrodynamic Fractionation.
The topics covered are:
1. Scope
2. Safety
3. Intended Use
4. Description and Specifications
5. Theory
6. Data Analysis
7. Sample Preparation
8. Eluant
2.2 SAFETY PRECAUTIONS
The pump and the detector in the CHDF 2000 use high voltage. Avoid contact with any high
voltage area. Un-plug the entire CHDF 2000 before performing service or maintenance. Do not store or
spill eluant or samples on the electrical modules. Be certain that all laboratory power receptacles are
properly grounded.
The CHDF 2000 uses ultraviolet radiation. Avoid eye contact with any ultraviolet source without
suitable eye shielding or protection. The UV source is fully contained within the CHDF 2000 casing.
The CHDF 2000 pumps a fluid through a set of capillaries at a relatively high pressure. The
compressibility of liquids is low, therefore a system component will probably leak before excessive
pressure causes other components to fracture. Always wear appropriate eye protection in the laboratory.
The actual fractionation capillaries are fully contained in a removable cartridge. The potential hazard from
a fracture or breakage in one of the capillaries is greatly reduced. Do not attempt to remove the capillaries
from the cartridge or tamper with the internal plumbing of the cartridge.
Do not attempt to perform service while the system is pressurized. Allow the system pressure to drop
below 1000 PSI before loosening any tubing fittings.
The GR-500 eluant for use in the CHDF 2000 consists of surface active agents in deionized water and to
our knowledge poses no chemical danger. There are currently three versions of the GR500 eluant: 1XGR500, 2X-GR500 (standard), and 2X-GR500-LA. Never use any eluants besides those supplied by
Matec Applied Sciences. All samples must be suspended in water.
The marker used in the CHDF 2000 is a solution of sodium benzoate (benzoic acid) in deionized water.
The marker concentration can be 0.2-1.0 wt. % depending on the kind of cartridge being used.
2.2.3 Chemical
2.3 INTENDED USE
The CHDF 2000 is designed to determine the size distribution of particles suspended in an aqueous eluant.
The particle diameter range which is covered by the standard C-202 Fractionation Cartridge and the GR500 eluant is 15 nm to 1100 nm. The fractionation resolution is low above the upper size limit. Larger
particles, up to 3 microns approximately, will not harm the fractionation capillaries. Two other
fractionation cartridges are available: high resolution (C-201), and high sensitivity (C-203).
The particle size range for the high resolution cartridge is 15 nm to 700 nm. The high resolution cartridge
offers resolution power two to three times higher than a standard resolution cartridge but at a lower upper
particle size range (700 nm). The C-203 cartridge measures particles in the range 15 nm to 1100 nm and
has about half the resolution of the C-202 cartridge. The C-203 provides particle detection sensitivity three
to four times that of the C-202 cartridge. High sensitivity may be needed for the analysis of non UV-light
absorbing materials such as colloidal silica and other inorganics.
The CHDF 2000 uses the Capillary Hydrodynamic Fractionation (CHDF) technique. This technique is
based on the rigorously described and documented behavior of particles influenced by the velocity profile
of a flowing fluid. Due to the influence of the fluid velocity profile, larger particles will have a greater
average velocity than smaller ones. The particle size distribution is determined by measuring the transit
time of the particles through a series of capillary tubes.
The CHDF 2000 can be used to analyze a wide variety of colloidal suspensions. The Fractionation
Cartridge and the GR-500 eluant are for use with either anionic or sterically stabilized dispersions.
Typical applications include:
1. Latex
2. Organic pigments
3. Carbon black
4. Emulsions
5. Liposomes
6. Inorganic Particles
For more detailed information regarding different applications, please contact Matec Applied Sciences
directly.
The CHDF 2000 achieves the separation of particles using a several different fractionation cartridges.
These are illustrated below in Figure 2.1.
Top Cartridge
Bottom
Cartridge
Filter
Out
In
Pump, Valves,
and Plumbing
After the sample is injected, the flowing eluant carries a small portion of it into the top cartridge. The
particle size separation takes place in the top cartridge. This cartridge contains a coil of glass capillary
tubing as seen in Figure 2.2. This is the cartridge that needs to be replaced on occasion. The back
cartridge contains two glass capillary coils which rarely need replacement. A large proportion of the
eluant and sample flow through the back cartridge, and then to waste. Only a small portion of any injected
sample is needed to measure a fractogram.
Figure 2.2 The inner workings of the top and bottom fractionation cartridges.
2.4.1 Computer Specifications
To run the CHDF Version 2.0 software you need an IBM or compatible PC equipped as follows:
1. Running Windows 95.
2. A Pentium processor.
3. 8 megabytes or more of random access memory (RAM).
4. 240 megabyte or larger hard disk drive.
If your computer does not meet these specifications do not install the software. Upgrade your computer,
or find a computer that meets or exceeds the specifications given above.
3. Cumulative Drift: ±1% of indicated flow rate or 10 ul/min
per 10 hours with 100 psi minimum
pressure
_____________________________________________________________________________
5. Flow stability: ±0.1% of indicated flow rate
________________________________________________________________________
2.4.4 CHDF 2000 OPERATING SPECIFICATIONS
The operating specifications of the CHDF 2000 are:
1. Particle size measurement range:
The size range of the standard C-202 Cartridge and the 2XGR-500 eluant is 15 nm to 1100 nm
particle diameter. A 15 nm to 700 nm high-resolution range is available (C-201). A high-sensitivity
cartridge for measurement of non-UV-absorbing materials is available (C-203).
2. Size Resolution
C-202 Standard Resolution Cartridge: Particle ratio = 1:1.2; High Resolution C-201 cartridge: Particle
ratio = 1:1.1 minimum; C-203 cartridge: 1:1.3, e.g.: narrow particle size distribution particles with greater
than 20% (standard resolution) or 10% (high resolution) size difference will be resolved.
3. Measurements Provided:
a) Particle size distributions by number, area, and weight
b) Cumulative particle size distribution by number, area, and weight
c) Number average particle size
d) Area average particle size
e) Weight average particle size
f) Volume average particle size
g) Absolute detector output
h) Particle size distribution standard deviation
i) Particle size distribution half height width
j) 25% percentile
k) 75% percentile
4. Measurement Time:
Twelve minutes maximum from sample injection to graphics output.
b) Tabular data of relative and cumulative size distributions
c) Graphics of relative and cumulative size distributions
d) Hard copy or terminal display only
6. Sample Size:
One milliliter at 0.1-5 volume %, pre-filtered through a 5 micron or larger pore size filter. Filters are
available (See Section 6, Replaceable Parts and Consumables). The actual sample concentration to be
used is a function of the detectability of the sample. All samples must be aqueous based.
7. Operating Temperature:
20 to 40 degrees Celsius. Fractionation column temperatures can be set to a maximum of 99°C. A column
temperature of 35°C is recommended.
8. Sample Eluant
The CHDF GR-500 eluant consists of proprietary surface active agents in deionized water supplied by
Matec Applied Sciences in concentrated form. Three types of GR-500 eluant are available: 1X-GR500,
2X-GR500, and 2X-GR500-LA. The 2X-GR500 is the standard eluant used for regular submicron particle
size analysis. The 2X-GR500 normal conductivity must be in the range 22-27 µSiemens/cm. The GR500LA conductivity must be in the range 14-18 µSiemens/cm. The 1X-GR500 conductivity should be between
6 and 10 µSiemens/cm. Contamination of the eluant, especially by electrolytes, will result in lower than
expected particle sizes. The same eluant type must be used during calibration and regular operation of the
cartridge. Changing the eluant type changes the calibration curve.
To prepare a batch of eluant, combine 1 part GR-500 concentrate with 9 parts water deionized water. The
eluant may then be sonicated or filtered under vacuum for degassing purposes.
Lower ionic strength eluants produce stronger electrical double layer repulsion between the particles and
the capillary wall. Particles traveling closer to the center of the capillary travel faster down the capillary. A
lower ionic strength produces more efficient separations but a lower practical particle size range upper
limit. See section 2.5, Theory of Operation.
9. Capillary Hydrodynamic Fractionator:
Integral, removable, disposable cartridge (two are provided per unit, one as spare).
10. Power Requirements:
115 VAC to 120 VAC, 60 Hz, or
220 VAC to 240 VAC, 50 Hz, 2.5A, single phase
11. Physical Dimensions:
Comfortably fits on a standard lab benchtop
a) Size:
15" (37.5 cm) Wide x 28" (71.1 cm) Deep x 15" (37.5 cm) High
The CHDF 2000 is factory pre-calibrated. Each fractionation cartridge is provided with a floppy disk. The
floppy disk contains calibration curve files. The eluant flow rate and type at which the cartridge must be
operated are indicated on the floppy disk label, a typical flow rate is 1.4 ml/min. Periodic field
recalibration can be performed by the user directly through the CHDF software.
13. Calibration Material:
The recommended calibration materials are monodisperse polystyrene latexes. These are available from a
wide variety of commercial sources or directly from Matec Applied Sciences.
14. Warranty:
Complete one year limited warranty, excluding replaceable CHDF cartridge.
ANY ATTEMPT TO REMOVE OR TAMPER WITH THE INTERNAL PLUMBING OF THE
FRACTIONATION CARTRIDGE WILL VOID THE WARRANTY. THE USE OF ELUANT OTHER
THAN THOSE SUPPLIED BY MATEC APPLIED SCIENCES ALSO VOIDS THE WARRANTY.
2.5 THEORY OF OPERATION
The behavior of macromolecules and particles in suspensions flowing through capillary tubes has been
studied in detail because of its importance in certain biological phenomena, such as the transport of red
blood cells through the smaller capillaries in the circulatory system (1), and for use in particle
fractionation methodology.
The separation of particles by size due to hydrodynamic effects occurring in the flow of suspensions
through capillaries was first examined by Di Marzio and Guttman in 1970 (2,3). A particle exhibiting
Brownian motion suspended in a viscous fluid undergoing Poiseuille flow within a capillary tube will
sample all accessible radial positions if the particle elution time is sufficiently long. The closest approach
of the particle center to the inner wall of the capillary will be limited to a distance equal to the particle
radius. As the particle size increases, it will become increasingly unable to sample the slowest moving
eluant streamlines near the capillary wall. As a consequence of this, the particle velocity will exceed the
average velocity of the eluant and the velocity of smaller particles. This effect is based only on the particle
size and is not a function of the particle density or composition. In addition to this exclusion effect, the
longitudinal and radial particle displacements are affected by the presence of the capillary wall, and by a
fluid inertial force.
The Capillary Hydrodynamic Fractionation CHDF 2000 instrument from Matec Applied Sciences uses
these phenomena to determine particle size and size distributions in the sub-micron range. The instrument
moves the eluant and the particles through a set of capillaries, measures the particle concentration leaving
the system as a function of the transit time, and relates this transit time to that of well characterized
monodisperse size standards used in the calibration procedure. Through a mathematical interpretation of
the raw data, the true particle size and size distribution is established.
The software uses two particle extinction cross section curves, one for particles that absorb UV light and
one for those that do not, in order to compute the sample particle numbers. These curves are plots of
particle extinction cross section against particle size. The particle size is determined from the particle
elution time. The number of particles is calculated from the following expression:
where D.O. is the detector signal for a particle of size i, Rext is the extinction cross section, and N is the
number of particles of size i. The computation of the extinction cross section by use of the Mie theory of
light scattering is currently being incorporated into the CHDF 2000 software. (D.O.) is multiplied by a
fractogram deconvolution parameter if the operator chooses to have the sample fractogram deconvoluted
into its primary components (See Software User’s Guide).
2.6 DATA ANALYSIS
The CHDF 2000 determines the concentration of particles exiting the fractionation cartridge as a function
of transit time from injection. This data is compared to a calibration curve generated by measuring the
transit times of monodisperse polystyrene latex size standards. The raw data can be mathematically
deconvoluted into component peaks. From the deconvolution and the absolute calibration using the
known size standards, the particle size and size distribution are automatically determined. The user may
also choose to not have the data deconvolved.
The CHDF 2000 Operating Software automatically calculates and displays the weight average, area
average, volume average, and number average particle size. In addition, the area, weight and number
relative and cumulative particle size distributions along with the standard deviation, half height width, and
the 25% and 75% percentiles are available on hard copy in either tabular or graphical form. The raw data,
which is the absolute detector output, is available as well.
It should be noted that both the theoretical foundation for CHDF and the calibration curve are for spherical
particles. If your particles are rod shaped or platelets, then the size measured by CHDF will be an
approximate spherical diameter. Distributions in size will still be absolutely determined, but no specific
information will be given on the aspect ratio of the particles.
2.7 SAMPLE PREPARATION
The CHDF 2000 requires 1 milliliter of sample at 0.1-5 volume %. The sample must be aqueous based.
The sample can be prepared in either of two ways; dilution of a concentrate or the addition of the disperse
phase to the eluant. It is recommended that the GR-500 be used for dilutions or making dispersions. The
concentration of the disperse phase should be high enough to obtain good resolution and a favorable signal
to noise ratio. This is particularly important when trying to identify very small sub-populations. On the
other hand, it should be kept low enough to avoid particle aggregation or injection valve clogging
problems. Do not run samples above 5 volume % or this may clog the injection valve or damage the
fractionation cartridge.
The CHDF 2000 can determine the complete particle size distribution over the entire specified size range.
Polydisperse samples do not need special pretreatment. The standard size range is 15 nm to 1100 nm
particle diameter. Other size ranges will be available in the future.
The actual state of dispersion and the stability of the sample are important considerations for any particle
characterization technique. The CHDF 2000 measures the effective hydrodynamic particle size. An
aggregate will behave as a larger particle and will not be resolved into the smaller primary particle size.
For some samples this may cause discrepancies between CHDF and transmission electron microscopy
(TEM).
The standard CHDF 2000 GR-500 eluant is designed for anionic dispersions. When diluting a
concentrate, use the GR-500 fluid as the dilutant. This will help ensure a stable, well dispersed system.
Most systems will be anionic when dispersed in GR-500. For specific systems, this can be checked using
microelectrophoresis or electro-acoustic ESA analysis. Problems may occur when diluting systems
stabilized with cationic surfactants. For specific information regarding specific sample systems, please
contact Matec Applied Sciences directly.
Most samples will need to be sonicated before analysis. For many samples prepared by the dilution of a
concentrate, placing the sample container in an ultrasonic bath will be sufficient. For dispersing powders
into liquids or when dealing with materials which may have a strong tendency to form aggregates, it will
be necessary to use an ultrasonic probe immersed in the sample itself. In both cases, the objective is to
prepare a uniform dispersion in which the particles exist in the actual state of dispersion that you wish to
characterize.
The CHDF 2000 requires that the sample be injected by syringe through an injection valve, or the sample
may be injected using an autosampler. In many cases it is necessary to filter the sample before injection to
remove coagulum or outsize particles. Five-micron filters which attach directly to the end of the syringe
are recommended. These filters are available from Matec Applied Sciences.
2.8 ELUANT
The standard eluant supplied for use with the CHDF 2000 is designated GR-500 and is for anionic
systems. Most materials will be anionic when dispersed in GR-500. The eluant consists of surface active
agents in de-ionized water. GR-500 is supplied in the form of a concentrate which must be diluted 10:1
for use, e.g., 900 ml de-ionized water to 100 ml GR500 concentrate. GR-500 should be diluted with deionized water and the final solution should be filtered under vacuum through either a 0.1 or 0.2 µm pore
size filter. The purpose of this filtration under vacuum is to remove any particulate matter and also to degas the GR-500. Care should always be taken when handling the GR-500 to avoid contamination. The
eluant should be used within 3 days of preparation. The GR-500 concentrate should be used within one
month of receipt. It should be kept refrigerated.
Matec supplies three types of GR-500 eluant: 1X-GR500, 2X-GR500, and 2X-GR500-LA. The appropriate
conductivity ranges (µSiemens/cm) of final eluant solutions are 6-9 for 1X-GR500, 14-19 for 2X-GR500LA, and 22-27 for 2X-GR500. Each of these eluants can be used with any of the three types of cartridges
available. Most commonly, 2X-GR500 is used in conjunction with a C-202 cartridge.
Both 2X- eluants produce the same resolution and use the same calibration curves. The 2X-GR500
contains 10 milligrams/l sodium azide biocide; the 2X-GR500-LA eluant contains only 0.5 milligram/l
sodium azide. As a result, the 2X-GR500-LA eluant degrades faster due to bacterial growth. The shelf-life
for these eluants is three days for the 2X-GR500-LA, and one week for the 2X-GR500. The 2X-GR500
cannot be used at 200 nm UV-detection because of the high UV-absorption by sodium azide.
The 1X-GR500 provides higher particle resolution efficiency than the 2X- formulas. The upper particle
size limit is lower with 1X-GR500. The 1X-GR500 contains 0.5 milligrams/l sodium azide. Its shelf life is
three days.
NOTE:
The CHDF 2000 and the fractionation cartridge have been specifically calibrated for use with GR-500
eluant. The use of any other eluant will give erroneous results and voids the CHDF 2000 warranty.
When you power-on the CHDF 2000, the system automatically executes a set of diagnostics to verify that
the system is ready for operation. The results of the power-on diagnostics appear in the Power-On
Diagnostics screen (Figure 3-2).
Figure 3-2 Power-On Diagnostics Screen
Table 3-1 identifies and describes the functions of the four power-on diagnostic tests.
Table 3-1 Power-On Diagnostic Tests
Name Function
EPROM Checksum Performs a checksum test of program
memory (erasable programmable readonly memory).
RAM Test Tests the validity of programmed method
data saved in RAM (random access
memory).
Cartridge Heater Determines if an optional cartridge heater
is installed.
Calibrate UV Performs calibration and check of the
absorbance detector optics. This test
takes 2 to 3 minutes to complete.
If your CHDF 2000 fails any of the power-on diagnostics, refer to Chapter 5, Diagnostics, for
troubleshooting information.
3.1.3 Checking the Control Panel Display
When you first power-on the CHDF 2000, check the control panel display for readability. If
necessary, adjust the brightness and contrast of the control panel display, as described in Section
You can initiate all CHDF 2000 control functions from the Home screen. From the Power-On
Diagnostics screen (see Figure 3-2), press the selector knob to choose Continue. The Home
screen appears (Figure 3-3).
3-4
You are now ready to set up the CHDF 2000 for normal operation.
Preparing the System for Use
From the Home screen (see Figure 3-3), choose Prepare system. The Prepare System screen
appears as shown below.
There are four choices in this screen used to prepare the system, Set Flow rate, Set Wavlngth.,
Calibrate pH, and Prime Pumps. Rotating the selector knob on the instrument alternately
highlights the different choices on the screen, and pressing the selector knob performs the
highlighted action. When any of these four options are selected, a Configure System screen
appears as shown in Figure 3.5.
Figure 3.5 The Configure System Screen
Although there are 4 choices in this menu, only one of them can be accessed at a time. For
example, if you selected Flow Rate from the menu in Figure 3.4, when the menu in Figure 3.5
appears, only the flow rate parameter can be changed. All other choices are disabled. Select
Home Screen to get back to the top menu. Here is how to set the four parameters listed in the
Configure System screen seen in Figure 3.5.
Flow Rate - This sets the rate at which eluant is pumped through the fractionation cartridge. To
change the flow rate, simply select it, then rotate the selector knob left or right until the flow rate
you desire appears on the screen. Push the selector knob to enter the value. Normally, a value of
1.4 should be selected.
Set Wavelength - This command sets the wavelength at which the UV/VIS detector will work.
The wavelength is in units of nanometers. To change the wavelength, select Set Wavelength,
rotate the selector knob left and right until the desired wavelength appears. Push the selector
knob to enter the value. The recommended value is 220 nm.
Calibrate pH - Select this option when you need to calibrate the pH sensor on the CHDF 2000.
You will need three buffer solutions, typically of pH 4.0, 7.0, and 10.0 to calibrate the pH sensor.
The sensor is calibrated by alternately immersing the sensor in the three buffers, then entering the
pH value of the buffer into the instrument. This is accomplished by first selecting Calibrate pH
from the Configure System menu. This action brings up a new screen as shown below in Figure
3.6.
Figure 3.6 The Calibrate pH screen
As stated in the screen, first highlight the pH that is going to be set by turning the selector knob.
Click the knob, then turn it to select the pH of the buffer to be used. Place the pH sensor in the
buffer, wait for the Prove mV reading to stabilize, then turn the selector knob to the right to
highlight either Accept Acid, Accept neut, or Accept base. Press the selector knob to enter the
pH value. Three buffers of different pH must be used to calibrate the sensor.
Prime Pumps - The pumps are primed by drawing eluant into the pump heads to get rid of air
bubbles. The process is described in a screen that can be accessed by selecting Prime Pumps.
The following screen appears.
of steps to follow are: 1, 2, 3, 6, 5, 4. The process will be explained here in words to avoid
confusion. The syringe attaches to a special valve in the lower right front of the instrument called
the priming valve. Attach the syringe to the priming valve, then open the valve. Draw several
mls of eluant into the syringe. Start the pump by selecting Press to START, remove the syringe
from the priming valve; wait until the pump stops, then close the valve. Please realize that this
screen only gives instructions, accessing this screen by itself will not prime the pumps.
The order of the instructions shown in this screen is not correct. The correct sequence
3.2 Reviewing and Editing Operating Parameters
The CHDF 2000 is configured with default values for 14 operating parameters that control
specific system functions. Review all operating parameters and edit them if required to customize
instrument operation for your needs.
3.2.1 Changing Operating Parameters
To review and edit the system parameter values of the CHDF 2000 choose Configure system
from the Home Screen. The first of three Configure System screens appears (Figure 3.8).
When each of these commands is selected, the display changes so that a number may be
entered. The parameters are described and set as follows:
Set Flow Rate - Sets the flow rate at which eluant is pumped through the fractionation
cartridge in units of ml/min. When selected, the display changes to “Rate = “. Turn the
selector knob to the desired value, then push the knob to enter the value. A typical flow rate is
1.4 ml/min. This is the same parameter as was set in the Configure System screen seen in
Figure 3.5.
Set Wavelength - This command sets the wavelength at which the UV/VIS detector will work.
The wavelength is in units of nanometers. To change the wavelength, select Set Wavelength.
The display changes to “Run Wavelength = “. Rotate the selector knob left and right until the
desired wavelength appears. Push the selector knob to enter the value. A typical wavelength is
220 nm.
Set Cell K - This parameter is used to calibrate the conductivity sensor that is onboard the
CHDF 2000. When this command is selected, the display changes to “Cell K:”. Rotate the
selector knob to the correct value, then push the knob to enter the value. The selected value
should be in the range 0.8 to 2.0.
Run Time - This number sets the length of the X axis on the display when data is viewed
during a CHDF run. The default value is 15 minutes. When this command is selected the
display changes to “Run Time:”. Rotate the selector knob to the desired value, then push the
knob to enter the value.
Heater - This parameter is used to set the temperature of the fractionation cartridge. The
default is 35°C. When this parameter is selected the display changes to “Cartridge Temp: “.
Rotate the selector knob to the desired value, then push the knob to enter the value.
NOTE: Changing the cartridge temperature requires the measurement of a new
calibration curve.
Next Screen - Displays the second Configure System screen, seen in Figure 3.9 below.
The second Configure System screen is displayed by selecting Next Screen in the first Configure
system screen. The second Configure System screen is shown below in Figure 3.9.
When each of these commands is selected, the display changes so that a number may be entered.
The parameters are described and set as follows:
Pressure Limit - Used to set the pressure values above or below which the pump will shut off.
Pressure is measured in units of PSI (pounds per square inch). When clicked once, the display
will change to “Low Limit”. Rotate the selector knob to the desired value (1000 is
recommended), then push the knob to enter the value. The display will then change to “High
Limit” (6000 is recommended). Rotate the selector knob to the desired value, then push the knob
to enter the value.
Prime Flow Rate - Sets the flow rate to be used during priming in ml/min. When this parameter
is selected, the display changes to “Rate ml/min:”. Rotate the selector knob to the desired value,
then push the knob to enter the value. A typical value is 8.0 ml/min.
Prime Time - Sets the amount of time the pump runs during priming (2 min. normally). Time is
measured in units of minutes. When this parameter is selected the display changes to “Prime
Time:”. Rotate the selector knob to the desired value, then push the knob to enter the value.
Pump Standby - Sets how long the pump will run before it automatically shuts off. Time is
measured in units of minutes, and a setting of “0” will cause the pump to run indefinitely. The
parameter is best used in conjunction with an autosampler. For example, it can be used to shut
down the pump after the instrument has concluded an overnight run. However, the Pump
Standby time can expire before all samples are run, so be careful when setting this parameter. For
most purposes, a value of 0 should be used and the pump can be turned off manually (or the pump
can be turned off using the autosampler sequence setup options parameters in the software).
When this parameter is selected, the display changes to “Standby:”. Rotate the selector knob to
the desired value, then push the knob to enter the value.
Lamp Standby - Is used to determine how long the lamp in the UV/VIS detector will stay on
before it is automatically shut off. The purpose of shutting off the lamp is to hopefully extend its
life. However, if the lamp is shut off before all your samples are run, you will not obtain any
data. A value of “0” will leave the lamp on indefinitely. The lamp standby value must be equal
to the pump standby value. The instrument will not allow you to set the two parameters to
different values. Once this parameter is selected, the display changes to “Standby:”. Rotate the
selector knob to the desired value, then push the knob to enter the value.
Next Screen - Selecting Next Screen brings up the third of the three configure system screens as
seen below in Figure 3.10
The third Configure System screen is displayed by selecting Next Screen in the second Configure
system screen. The third Configure System screen is shown below in Figure 3.10.
Figure 3.10 The third configure system screen.
When each of these commands is selected, the display changes so that a number may be entered.
The parameters are described and set as follows:
AU Zero Offset - Tells the instrument how many absorbance units of signal being measured by
the detector should be set equal to zero. Normally, this parameter should be left at zero. Once
this parameter is selected, the display changes to “mAU = :”. Rotate the selector knob to the
desired value, then push the knob to enter the value.
Eluant Temp - This parameter is used to calibrate the temperature probe on the CHDF 2000.
After selecting this parameter, rotate the selector knob until the current value of the eluant
temperature appears. Then push the knob to save the changes.
Auto Zero - Turns the autozeroing ability of the instrument on or off. When on, an autozero of
the detector baseline takes place each time an injection occurs. The autozero sets the detector
baseline to the mAU value defined in AU Zero Offset. Once selected, this parameter is turned on
or off by rotating the selector knob. Push the knob to record the change.
Serial Parameters - Sets the communication parameters that allow the CHDF 2000 to talk to a
personal computer. Once selected, display changes to “Baud Rate”. Rotate the selector knob the
appropriate value (normally 19200), then push the knob. This value must be the same as that set
in the software under system/configure system. Next, the display changes to “Echo” and is
followed by an “Off” or “On” (normally off). Rotate the selector knob to the desired choice, then
push the knob to save the settings.
Y Scale - Sets the maximum absorbance reading that can be displayed on the CHDF 2000 front
panel raw data graph during a fractionation. When selected, the display changes to “Y Axis AU
Max” and number appears next to this. Rotate the selector knob to the desired value, then push
the knob to save the settings. A value of 5 is recommended.
Next Screen - In this instance, selecting Next screen takes you back to the first Configure System
screen.
2. Wait for the pump pressure to be stable (about 1/2 hour). Start the CHDF Software.
Follow the software manual instructions for data collection and particle size distribution
results.
3. With the manual injector valve in the Load position, insert the prepared sample syringe in
the manual injector valve (Figure 4.5).
Note: The loop volume of the instrument is 20 µL. Load a minimum of 80 µL of sample to ensure
accurate results.
Handle
Manual
injector
valve
Figure 4.5 Loading Sample Into the Manual Injector Valve
Note: The manual injector valve must be in the Load position when you load the sample.
4. Load the sample into the manual injector valve. Do not remove the syringe until you have
completed step 4.
5. Rotate the handle of the manual injector valve to the Inject position to inject the sample
and start the selected method. The Analyze/Monitor Samples screen appears (Figure 4.5).
Remove the sample syringe.
6. Rinse the injection valve with GR-500 eluant. Repeat the above steps to inject the marker
solution. The recommended marker injection time is 60 seconds.
7. During the fractionation, the results will be displayed on the screen shown in Figure 4.4.
Also, relevant measurements such as the eluant temperature, pressure, flow rate, eluant pH,
eluant conductivity, and detection wavelength are displayed. Repeat the above steps for
new samples. Select Stop Pump or Home Screen to end your session.
8. Review your results. If they are satisfactory, you are ready to analyze the next sample. If
the results are not satisfactory, refer to Chapter 5, Troubleshooting the CHDF 2000, to
locate the source of the problem.
When you finish analyzing samples, but do not want to alter the current operating conditions, use
Standby mode to conserve eluant and lamp energy. The instrument will also go into standby
mode if the computer and instrument have ceased talking to each other for > 60 seconds. This
may happen, for example, if the computer crashes while the instrument is turned on.
Using Automatic Standby Mode
You can configure the system for two types of automatic standby mode:
• Automatic pump standby mode
• Automatic lamp standby mode
When Standby Mode is entered the following screen appears (Figure 4.6).
4-6
Using automatic pump standby mode
In automatic pump standby mode, the pump or pump shut off after a user-specified time interval
at the end of a run.
Note: The pump standby time must be either equal to or shorter than the lamp standby time.
You set the time interval by defining the value of the Pump Standby Delay operating parameter.
Refer to Chapter 3, Reviewing and Editing System Parameters, for more information.
Using automatic lamp standby mode
In automatic lamp standby mode, the detector lamp shuts off after a user-specified time interval at
the end of a method run.
Note: The lamp standby time must be equal to or longer than the pump standby time.
You set the time interval by defining the value of the Lamp Standby Delay operating parameter.
Refer to Chapter 3, Reviewing and Editing System Parameters, for more information.
This chapter provides procedures that explain how to:
• Respond to CHDF 2000 error messages
• Use troubleshooting checklists to quickly identify potential sources of system problems
• Run CHDF 2000 diagnostics
5.1 Responding to Error Messages
During operation of the CHDF 2000, conditions may occur that cause error messages to appear on
the control panel display. The error messages are:
• High pressure limit exceeded
• Low pressure limit exceeded
• Power lost
• Detector calibration failure
5.1.1 High Pressure Limit Exceeded
The High Pressure Limit error message (Figure 5-1) appears when the system pressure exceeds
the high-pressure limit for at least two seconds. You configure the high-pressure limit using the
Pressure Limit Set operating parameter (refer to Chapter 3, Reviewing and Editing Operating
Parameters).
Figure 5-1 High Pressure Limit Exceeded Error Message
Recommended response
High pressure sometimes occurs if you are priming the pump with a cartridge installed. To
prevent high pressure during pump priming, follow the procedures specified in Section 4.1,
Priming the Pump.
This message usually results from a blockage in the eluant flow path. Check the eluant flow path
and pump. Clear any blockages, if necessary. The most common cause of a blockage is a plug or
sample somewhere in the instruments tubing. There are several ways of clearing such a blockage.
They will be discussed below in order of increasing difficulty.
1. If you are analyzing latex or other polymeric based samples, a blockage may be removed by
pumping acetone or tetrahydrofuran (THF) through the system for 15 minutes. This will
hopefully dissolve the plug.
2. Change (or clean) the in-line particle filter or its element (See section 6.6).
3. You can also try backflushing the fractionation cartridge to remove a plug of material. The
idea is to pump eluant through the cartridge in a direction opposite to that which is normal.
To backflush the cartridge, open up the front door of the CHDF 2000. There is a panel on the
right hand side held in place with two screws. Remove these screws (they should come out
by hand), and you should see the following.
1
2
3
4
Cartridge
Filter
Top
Cartridge
The idea is to swap the tubes at fittings 1, and 2, and separately 3 and 4 as follows: Disconnect
the stainless steel tube from Fitting 1. Connect the tube on Fitting 2 to Fitting 1; then connect the
tube from Fitting 1 to Fitting 2. Disconnect Fitting 3. Connect the tube on Fitting 4 to Fitting 3
and Fitting 3 to Fitting 4. Run the pump for 15 minutes. Reconnect all fittings as they are
normally, and try running the pump again. If the high pressure message does not re-appear, your
backflushing was successful.
4. Snip off end of the glass capillary tube in the bottom cartridge (see Section 6.6).
5. Replace the entire bottom fractionation cartridge (see Section 6.6).
Do not restart sample analysis until you correct the cause of high system pressure.
Press the selector knob to clear the High Pressure Limit Exceeded error message.
5.1.2 Low Pressure Limit Exceeded
The Low Pressure Limit error message (Figure 5-2) appears when the system pressure falls below
the low-pressure limit for more than 110% of the value defined for the Pump Priming Time
parameter. You configure the low-pressure limit using the Pressure Limit Set operating parameter
(refer to Chapter 3, Reviewing and Editing Operating parameters).
This message often occurs when there is a leak in the injector, pump, or eluant flow path. Check
eluant flow path, injector, and pump and repair any leaks or damage.
Reprime the pump and eluant tubing. Do not restart sample analysis until you correct the cause of
low system pressure.
5-3
Press the selector knob to clear the Low Pressure Limit Exceeded error message.
5.1.3 Power Lost
The Power Lost error message (Figure 5-3) appears if the CHDF 2000 temporarily loses power
during sample analysis.
Recommended response
This message occurs when the CHDF 2000 temporarily loses power due to a power outage or
inadvertent powering-off of the system during sample analysis. Before you resume sample
analysis:
Figure 5-3 Power Lost Error Message
1. Verify that your operating parameters are correct.
2. Prepare the system by priming the pump, if necessary.
Press the selector knob to clear the Power Lost error message.
Although the system allows you to continue operation when this message appears, the results of
the analysis are not reliable. Correct the source of the detector problem before continuing
operations. Refer to Section 5.2.5, Absorbance Detector Troubleshooting, for more information.
Press the selector knob to clear the Detector Calibration Failure error message.
5.2 Quick Troubleshooting Checklists
Use the tables in this section to quickly identify corrective actions for common CHDF problems.
This section covers the following topics:
• General troubleshooting
• Fractionation troubleshooting
5.2.1 General Troubleshooting
Use Table 5-1 to quickly identify corrective actions for general system problems.
Table 5-1 General Problems – Quick Troubleshooting Checklist
Symptom Possible Cause Corrective Action Refer to
Unit will
not poweron
No power at outlet Reset circuit breaker or
Power supply fuse blown Replace fuse.
Power cord not connected Connect power cord or try
This section describes how to run the user-initiated CHDF 2000 diagnostics.
This section describes how to:
• Initiate CHDF 2000 diagnostics
• Interpret CHDF 2000 diagnostics
5.3.1 Accessing CHDF 2000 Diagnostics
To access CHDF 2000 diagnostics from the Home screen, choose Run Diagnostic. The Run
Diagnostic screen appears (Figure 5-5).
Figure 5-5 Run Diagnostics Screen
When each of these commands is selected, the display changes so that a number may be entered
or observed. The parameters are described and set as follows:
Test Start in Sig. - This displays the voltage signal sent by the CHDF 2000 to a computer so the
computer knows when samples have been injected and loaded. When selected, the displays
changes to “Inject:” and the inject signal is displayed in volts. When the selector knob is pressed
a second time the display changes to “Load:” and the load signal is displayed in volts.
Test Output Volts - This voltage is used to run a strip chart recorder or other recording device.
It is not important when a computer is used to run the CHDF 2000. When selected, the display
changes to “DAC Value =“ and a voltage is displayed.
Output Exp. Factor - Exp. Factor stands for Exponential Factor. This is another parameter used
when using a strip chart recorder or other recording device. This parameter is not used when a
computer is attached to the CHDF 2000. When selected, the display changes to “Exp Factor =“.
Rotate the selector knob to a value between 1 and 100, then push the knob to save the entry.
Test Keypad - This function is used to test whether all of the keys on the CHDF 2000 keypad are
functioning properly. Once selected, the display changes to show the series of characters that can
be typed in from the keypad. The character most recently pressed is highlighted on the display. If
a key is pressed and it does not light up on the display, the key may be malfunctioning.
Wavelength vs AU - This stands for Wavelength vs. Absorbance Units, and is used to observe
raw detector signal. When selected the display changes to:
Ref = Smp = L = AU =
Ref is reference, Smp is sample, L is wavelength, and AU is absorbance units. The reference and
sample numbers are the raw detector signals measured by the sample and reference channels on
the detector. L refers to the wavelength being measured, and AU gives the current absorbance
being measured by the detector. These parameters should be monitored over time to see if they
change. Changes may indicate that the detector cell windows are becoming dirty, or that the lamp
strength is deteriorating.
5.3.2 Initiating and Interpreting Diagnostics
The CHDF 2000 diagnostics are explained in more detail in the following sections
Test START IN Signal Diagnostic
The Test START IN Signal diagnostic provides two functions:
• Tests the START IN signal on the signal connector module
• Tests the manual injector valve
Testing the START IN signal
To test the START IN signal on the signal connector module:
1. Place the manual injector valve in the Load position.
2. Choose Test START IN Sig. LOAD should appear in the display.
3. Connect an insulated wire between pins 9 and 10 on the signal connector module. Check
that INJECT appears in the display.
4. Remove the wire connection between pins 9 and 10. LOAD should reappear in the
display.
Testing the Manual Injector Valve
To test the manual injector valve:
1. Ensure that there is no electrical connection between pins 9 and 10.
2. Choose Test START IN Sig. LOAD should appear on the display.
3. Place the manual injector valve in the Inject position.
4. INJECT should appear in the display.
Press the selector knob to return to the Run Diagnostics screen.
Test Volt. Out Diagnostic
Use the Test Volt. OUT diagnostic to test the AU OUT, PRESS OUT, and %B OUT signals on
the signal connector module.
Table 5-3 identifies the expected voltages ranges of each signal.
•L – The wavelength of the light energy (in nm) produced by the diffraction grating
in the absorbance detector.
•AU – The calculated level of absorbance based on the following equation:
AU (Absorbance) = Log K
Where K is configured as a constant of the instrument optics at the time of the Autozero
function. K is held constant throughout a given separation while the values of REF and
SMP are continuously monitored to compute the value of AU.
2. Rotate the selector knob to vary L (in 1-nm increments) while you monitor the changes in
REF, SMP, and AU values.
3. Press the selector knob to return to the Run Diagnostics screen.
The two plungers in each pump are ultra-smooth, chemically inert sapphire rods. Salt crystals can
form on a plunger and cause wear on the plunger seals and on the plunger itself.
Small pieces of a worn plunger seal may get caught in the outlet check valve of the pump head. If
the plunger seals are worn, you may have to inspect and replace the outlet check valve. To replace
an outlet check valve, see Section 6.1.3, Cleaning and Replacing the Check Valves.
Required Materials
• 5 mL syringe
• 5/16-inch open-end wrench
• 5/32-inch Allen wrench
• Brass machine screw
• Replacement seals and washers
To Replace a Plunger Seal:
1. Insert a syringe into the priming valve for the pump (see Figure 4-6).
6-3
2. Open the priming valve.
3. Place the eluant reservoir below the level of the eluant inlet tube. Remove the inlet line
from the inlet tee of the pump and allow the eluant to drain back into the reservoir.
4. Set the Priming Flow Rate parameter to 0.5 mL/min (refer to Section 3.3, Reviewing and
Editing Operating Parameters).
5. Use the Prime Pump option to turn on the pump (refer to Section 4.2, Priming the Pumps).
6. Use the syringe to extract as much eluant as possible from the pump.
7. Use the Prime Pump option to stop the pump when the indicator rod is retracted into the
pump head (this protects the plunger from breaking under the weight of the pump head).
8. Using the 5/16-inch open-end wrench, remove the outlet tube from the outlet check valve
of the pump head (see Figure 6-1).
9. Remove the inlet tube from the inlet check valve (see Figure 6-1).
10. Using a 5/32-inch Allen wrench, remove the two pump-head-assembly mounting bolts
The plungers require careful handling. Although cleaning the plungers is not difficult, it is
important to follow these instructions carefully to avoid damaging the plungers. Assemble all
materials and read the procedure thoroughly before you begin.
This section provides the following plunger maintenance procedures.
• Accessing the plunger
• Cleaning the plunger
• Inspecting the plunger
• Replacing the plunger
Required materials
• 5/16-inch open-end wrench
• 5/32-inch Allen wrench
• 9/64-inch Allen wrench
• Snap-ring pliers
• Plunger insertion tool
• Fine abrasive, such as toothpaste
Accessing the Plunger
To remove the plunger from the pump head:
1. Remove the pump head as described in Section 6.1.1, Replacing the Plunger Seals.
2. Remove the four plunger support screws with the 9/64-inch Allen wrench. Carefully slide
the head support assembly and the indicator rod off the pump and set the head support
assembly on the benchtop (Figure 6-4).
3. Set the Priming Flow Rate parameter to 0.5 mL/min (refer to Section 3.3, Reviewing and
Editing Operating Parameters).
4. Use the Prime Pump function to turn on the pump (refer to Section 4.2, Priming the
Pumps).
5. Run the pump until the plunger is fully extended, then turn off the pump.
Caution: To prevent injury, be sure to secure the plunger at all times during this procedure to
prevent it from ejecting from the pump head.
6. Use the plunger insertion tool to compress the plunger spring with one hand. With the
other hand, use the snap-ring pliers to remove the snap-ring that holds the plunger in place.
7. Remove the plunger assembly (Figure 6-5) and set it on a clean flat surface.
Cleaning the Plunger
Buffers and impurities in eluants can coat the plunger and reduce the life of the seal. Cleaning the
pump plunger periodically can prolong the life of the seal. Clean the plunger with a fine pumice
or clear abrasive toothpaste. Rinse the plunger thoroughly with water or methanol to remove all
traces of the abrasive.
The parts of the plunger assembly are shown in Figure 6-6.
Snap-ring
Inspecting the Plunger
After you clean the plunger, inspect it for damage by holding it under a bright white light and
looking down its length for nicks and scratches. It is easier to see scratches under a bright light
than to feel them with your fingers.
Figure 6-5 Removing the Plunger Assembly
Plunger
Figure 6-6 Disassembled Plunger
If the plunger is not scratched or otherwise damaged, reassemble the plunger. If the pump was
leaking, install new pump seals. If the plunger is damaged, replace the plunger and the seals.
Whether you are reinstalling a plunger that you have cleaned and inspected, or replacing a
damaged plunger with a new one, use the following procedure:
Caution: Be very careful when handling the plunger. Wear safety glasses and use the plunger
insertion tool. If you break the plunger, sharp pieces may be ejected from the
instrument.
To reinstall the plunger:
1. Assemble the plunger and use the plunger insertion tool (Figure 6-7) to reinsert the plunger
into the pump.
6-7
Plunger
Insertion
Figure 6-7 Inserting the Plunger
2. Use the Prime Pump function to turn on the pump (refer to Section 4.2, Priming the
Pumps).
3. Wait for the plunger to retract, then turn off the pump.
4. Reinstall the head support assembly with the indicator rod in the upper right corner.
Alternately tighten the four screws.
5. Wet the seal and plunger with eluant.
6. Gently slide the pump head onto the plunger and alternately tighten the two mounting
bolts. Check the head alignment visually by observing the gap between the pump head and
the pump head support. Make sure the indicator rod moves freely. If the head is
misaligned, repeat step 4.
7. Set the Priming Flow Rate system parameter to 0.3 mL/min (refer to Section 3.3,
Reviewing and Editing System Parameters).
8. Use the Prime Pump function to turn on the pump (refer to Section 4.2, Priming the
Pumps).
9. Pull out and release the indicator rod. If the rod does not snap back easily, the head is
misaligned. Loosen the pump head and repeat steps 6 and 9.
10. Reconnect the eluant lines, reposition the eluant reservoir, and prime the pump (see
This section provides the following check valve procedures:
• Cleaning the check valves
• Replacing an inlet check valve
• Replacing an outlet check valve
Cleaning the Check Valves
Clean the check valves (see Figure 6-1) every six months, or more frequently if your application
warrants. It is a good idea to keep a spare set of clean check valves on hand. If you need to
remove the check valves for a full cleaning, you can install the spare check valves immediately
and clean the dirty check valves at a convenient time.
Required materials
•If you use organic eluants – 1 L of a compatible cleaning eluant, such as a 1:1 mixture of
methanol and water with liquid detergent
•If you use aqueous buffers – 1 L of a mixture of water and liquid detergent
On-line cleaning
To clean the check valves while they are installed in the system:
1. Set the following operating parameter values (refer to Section 3.3, Reviewing and Editing
Operating Parameters:
• Priming Flow Rate to 10.0 mL/min
• Priming Pump Time to 15 min.
2. Fill the pump eluant reservoir with at least 400 mL of cleaning eluant.
3. Open the priming valve of the pump to be cleaned.
4. Position a large beaker under the priming valve.
5. Use the Prime Pumps function to pump the cleaning eluant through the pump and into the
beaker below the priming valve.
6. Repeat steps 1 to 5 if additional cleaning is required.
7. When you finish, close the priming valve.
8. Restore the Priming Flow Rate and Priming Pump Time operating parameters to their
original values.
Full cleaning
If the on-line cleaning procedure is insufficient, take the check valves off the pump (as described
in the next two sections) and immerse the valves in an upright position for 10 minutes in
methanol. If a check valve does not work properly after cleaning, replace the valve.
1. Place the eluant reservoir below the level of the eluant inlet line. Remove the inlet tube
from the inlet tee and allow the eluant to drain back into the reservoir.
2. Use the 5/16-inch open-end wrench to remove the tubing from the check valve.
3. Use the 1/2-inch open-end wrench to loosen the inlet check valve. Then remove the check
valve by hand.
4. Remove the old gasket.
5. Install a new check valve and gasket. Hand-tighten the check valve, then tighten it another
half-turn with the wrench.
6. Reconnect the eluant lines, reposition the eluant reservoir, and prime the pump (see
Section 4.2, Priming the Pumps). Check for leaks.
6-9
Replacing an Outlet Check Valve Seal and Gasket
Required materials
• 5/16-inch open-end wrench
• 1/2-inch open-end wrench
• Dentist’s pick
• Thick plastic pen cap or other blunt tool (such as the plunger insertion tool)
To replace the outlet check valve seal and gasket:
1. Use the 5/16-inch wrench to disconnect the outlet tubing.
2. Use the 1/2-inch open-end wrench to loosen the check valve a half-turn. Do not remove it.
3. Remove the pump head (refer to Section 6.1.1, Replacing the Plunger Seals, steps 1 to 7).
4. Holding the pump head upside-down in one hand with the outlet check valve facing the
floor, unscrew the outlet check valve (Error! Reference source not found.) and set the
pump head down. Set aside the outlet check valve in a safe place, ensuring that it stays
upside down (to prevent the internal parts from falling out).
This section provides procedures for replacing the deuterium lamp in the absorbance detector.
The procedure include:
• Removing the old lamp
• Installing the new lamp
• Adjusting the lamp position
Lamp Characteristics
The intensity of the deuterium source lamp varies by wavelength (Figure 6-13).
100
Relative
Intensity
(%)
50
Figure 6-13 Deuterium Lamp Intensity Profile
When to Replace the Lamp
Performance requirements and permitted tolerances vary from application to application. If the
lamp no longer provides an adequate signal-to-noise ratio for your application, replace it.
It is not unusual for lamps to show a 30 to 50 percent decrease in intensity before the noise
increases by a factor of two.
Lamp Timer
The detector lamp has a life of approximately 1000 hours. One of the leads attached to the lamp
includes a 1000-hour timer to indicate lamp usage.
Before you replace the lamp, check the lamp timer (see Figure 6-14).
190
Wavelength
656
The lamp timer is a mercury column with a scale of 0 to 10 where 10 represents 1000 hours. As
the lamp ages, the bubble in the mercury column moves toward the 10.
Save all parts (screws, nuts, washers) after removal unless otherwise indicated. The parts are
needed to install the new lamp.
6.4.2 Removing the Lamp
6-17
Caution: To prevent eye damage from ultraviolet radiation exposure, turn off the unit and strictly
When you remove the lamp housing protective cover, the UV lamp automatically turns off.
Before removing the lamp, set the wavelength to 230 nm. If the lamp is bad or fails to light and
this wavelength setting is not possible, remove the lamp and replace it as follows.
Caution: The lamp housing becomes extremely hot during operation. To prevent burns to the
1. Turn off the unit power and allow the lamp to cool for one hour.
2. Remove the front and rear system covers.
3. Loosen, but do not remove, the two screws that secure the lamp housing protective cover.
4. Remove the lamp housing by sliding it up and away from the unit.
5. Disconnect the lamp connector from the lamp interlock cable (Figure 6-14).
adhere to the following procedures while changing the lamp.
skin, allow the lamp to cool for one hour before removing.
2. With the lamp aperture facing away from you, place the new lamp on top of the adjustment
plate and secure it with the shoulder screws. The lamp leads should line up with the notch
on the lamp adjustment plate.
3. Without touching the bulb, turn the assembly so that the bulb is positioned downward.
4. Place the center adjustment screw with the hex nut and star washer in the center opening of
the adjustment plate. In preparation for maximizing the lamp energy, turn the screw until
the lamp assembly moves about 3-4 mm away from the adjustment plate.
5. Place the lamp assembly onto the lamp housing. Make sure the lamp leads slide freely into
the housing slot.
6. Secure the lamp assembly to the housing and tighten the three Allen-head screws.
7. Plug the lamp connector into the lamp power cable assembly (see Figure 6-14).
8. Replace the protective shield to prevent direct access to the lamp when instrument power is
After you install the lamp, position the lamp for maximum lamp energy. The flowchart in Figure
6-18 illustrates the procedure for positioning the lamp adjustment plate (Figure 6-16) for
maximum lamp energy.
Note: The detector does not have to be stabilized to perform this procedure.
6-21
Figure 6-18 Adjusting for Maximum Lamp Energy
To maximize lamp energy (see Figure 6-18):
1. Turn on power for the CHDF 2000. The Power-On Diagnostics screen appears (see Figure
3-2).
Caution: To prevent the possibility of exposing your eyes to ultraviolet radiation, wear eye
2. Make a note indicating the result (pass/fail) of the Calibrate UV power-on diagnostic.
protection that filters ultraviolet light and keep the lamp in the lamp housing during
operation.
3. Choose Run diagnostics. The Run Diagnostics screen appears (see Figure 5-5).
4. Choose Show Energy, AU, λλλλ. The current wavelength (λ) setting, the detected
absorbance (AU), and the current lamp sample (SMP) and reference energy (REF) values
appear (see Figure 5-6).
If the result of the Calibrate UV power-on diagnostic was Pass, go to step 10.
If the result of the Calibrate UV power-on diagnostic was Fail, continue with step 5.
The CHDF capillaries are contained in two cartridges as follows (also see section 2.4):
1. The top, or Fractionation cartridge (white).
2. The bottom cartridge (black).
The top cartridge contains the capillary responsible for particle fractionation. The bottom
cartridge contains the “make-up”, as well as the “waste” capillaries. The CHDF cartridges can be
responsible for the following two problems:
1. No peaks are detected upon injection of samples and /or marker.
This condition is frequently caused by a faulty top cartridge.
2. High pump pressure. This is usually caused by a faulty bottom cartridge, or a plugged in-line
filter.
Normally, the top cartridge is not responsible for high pressure, even if it is totally plugged. This
is because only a very minor portion of the eluant flows into the fractionation capillary. High
pressure can also be caused by a blocked cartridge in-line filter as described in section 6.6.3.
There are 3 options for fixing the top cartridge fault which is responsible for lack of sample
detection. These are snipping off the end of the capillary, backflushing the capillary (see section
5), and replacing the fractionation capillary.
Chapter 5 discusses how to flush the cartridge with solvent, or backflush it to remove a plug. If
either of these techniques are not successful, there are two techniques discussed here that will
ultimately take care of the problem. The first, snipping off the end of the capillary, assumes the
plug is in the very end of the cartridge, and that this section will be removed when snipped off.
The second method described here, changing the fractionation cartridge, guarantees that the
plugging problem will be resolved.
A third section below shows how to clean the cartridge in-line filter in order to eliminate high
pump backpressure.
6.6.1 Snipping off the End of the Capillary
The idea is to dismount the white fractionation cartridge from the CHDF2000, with the two 5inch long, 1/16” stainless steel tubes still attached to the cartridge; once on the bench, remove the
inlet SS tube, leaving the capillary exposed, then snip the capillary. See Fig. 6.20. Details are
shown below.
1. Stop the pump on the CHDF 2000, wait for the pump pressure to drop below 1,000 psi.
2. Open the front door of the instrument. The cartridge is located in the center top position, and
is shown in Figure 6.20.
3. Disconnect the cartridge fitting from the instrument at locations 1, 2, 3, and 4 shown in
Figure 6.20. CAUTION: DO NOT LOOSEN FITTINGS NUMBER 5 and 6 YET. The fusedsilica capillary stretches into the two stainless steel tubes, 7 and 8. Detaching tubes 7 and 8 at
this point will result in fractionation capillary breakage.
4. Unscrew the three screws holding the cartridge in place. Gently remove the cartridge from
the instrument.
5. Using your thumb, straighten SS tube 7 somewhat.
6. Disconnect the fitting from the cartridge at location 6 (bottom cartridge inlet) shown in figure
6.20.
7. Pull the SS tube 7 away from the cartridge to expose the small, brown, glass capillary tube.
8. Score the capillary with a ceramic disc, then break off a 1/8” piece of capillary.
9. Replace the tubing in reverse order that it was disconnected.
Replace the cartridge using the screws provided. Re-connect the four fittings. Do not over-tighten
the fittings. Tighten ¼ turn past finger-tight.
10. Run known samples or standards to see if the cartridge is working properly.
6.6.2 Replacing the Fractionation Cartridge
If the snipping procedure described in the above section does not solve your problem, the
fractionation cartridge on the CHDF 2000 will need to be replaced. Once a new cartridge is
purchased from Matec, it is a relatively simple procedure to install it. The cartridge is located
behind the front door of the instrument, in the center, top position. Changing the cartridge is
performed in the following steps.
1. Stop the pump on the CHDF 2000.
2. Open the front door of the instrument. The cartridge is located in the center top position, and
6. Capillary Inlet Fitting. Do
not disconnect during cartridge
removal/installation.
Figure 6.20 The Fractionation Cartridge. IMPORTANT NOTE: This is how your cartridge
should look upon uninstalling it from the CHDF unit. Do not remove the Inlet and Outlet tubes
until after this whole assembly has been removed from the CHDF unit. Otherwise, you will break
the capillary inside these two stainless steel tubes.
3. Disconnect the cartridge fitting from the instrument at locations 1, 2, 3, and 4 shown in Figure 6.20.
CAUTION: DO NOT LOOSEN FITTINGS NUMBER 5 and 6 YET. The fused-capillary stretches into the
two stainless steel tubes, 7 and 8. Detaching tubes 7 and 8 at this point will result in fractionation capillary
breakage.
Unscrew the three screws holding the cartridge in place.
4. Gently remove the cartridge from the instrument.
5. Remove the plastic caps from the tube ends of the new cartridge if they are present. Take the new
cartridge and place it gently into the instrument.
6. Replace the cartridge using the 3 screws provided.
7. Reconnect the cartridge tubing to the instrument. Do not over-tighten the fittings. Tighten ¼ turn past
finger-tight.
6.6.3 Cleaning the In-line Filter (High Pump Pressure)
This filter keeps large particles from plugging the outlet to waste from the fractionation cartridge.
It may plug on occasion causing high backpressure, and may need to be cleaned. It is cleaned as
follows:
1. Open the door of the instrument and locate the filter next to the fractionation cartridge as
seen in the drawing below.
In-Line filter
Fig. 6.21. Cartridge In-Line filter.
2. Using a wrench, loosen the connections on both ends of the filter, and remove the filter from
the instrument.
3. The filter consists of a small metal piece screwed into a larger metal piece. The smaller metal
piece contains the filter element. Unscrew the smaller piece and remove it from the larger piece.
4. Remove the fritted stainless steel filter element from the piece you just unscrewed.
5. If you analyze polymer samples, unplug the filter element by sonicating the element in acetone
or tetrahydrofuran for at least 15 minutes. If you analyze silica or other inorganic substances,
sonicate the element in a 50/50 (by volume) mixture of methanol and 1M NaOH for at least 15
minutes.
6. Re-install the filter in the reverse order of the steps above.
7. Cleaning the filter may also be performed by backflushing, as described in section 5.1.1.
8. If cleaning the filter element fails, you can replace the filter element with a new unit.