Digital Pump Drive Operating Manualii
MasterflexMASTERFLEXLIVE
Safety Precautions
Preface
SAFETY
PRECAUTIONS
DANGER: High voltages exist and are accessible. Use extreme caution when
servicing internal components.
WARNINGS: Tubing breakage may result in fluid being sprayed from pump. Use
!
appropriate measures to protect operator and equipment.
Turn drive off before removing or installing tubing. Fingers or loose clothing
could get caught in drive mechanism.
WARNINGS: Do not operate the pump drive in a manner not specified in the
documentation. Misuse of the pump drive may result in a hazard and may
compromise the safety protection built into the pump drive. If the pump drive is
damaged, turn it off and not use it until service-trained personnel can check
its safety.
Single-Phase Only. Not to be used with Split-Phase lines.
The Power switch on the Back Panel is not the main disconnect. Main
disconnect is accomplished by disconnecting the detachable power supply
cord at the appliance coupler or at the main plug. Ensure the power cord is
easily accessible and removable, in the event of an emergency, which requires
immediate disconnection.
The operator should check the detachable power supply cord condition. The
equipment should not be operated if the power supply cord is cracked or
broken. Any obvious damage to the enclosure (from a drop or fall) should be
checked by service personnel for loose or damaged parts inside.
CAUTIONS: Power must be turned off before connecting the external remote
!
control cable to prevent damage to the drive.
Do not contaminate the lubricant in the container, on the shaft or on the seal
with foreign material. Failure to observe this precaution may result in damage
to the seal and premature failure of the seal.
No foreign matter should be allowed under the gasket on the back of the front
plate or under the heads of the screws. Failure to observe this precaution may
result in leakage during wash down of the drive.
Do not block the rear panel of the pump drive. The power switch must always
be easy to access. The power cord must always be easy to disconnect.
Replace the power cord only with one of the same type and rating. The
minimum power ratings are stated on the rear panel.
The power cord set supplied with your pump drive meets the requirements of
the country where you purchased the pump drive. If you use the pump drive in
another country, you must use a power cord set that meets the requirements
of that country.
When using hazardous chemical and biological agents, take all suitable
protective measures, such as wearing protective glasses and gloves resistant
to the substances used. Follow local and/or national regulations for safe
operation and maintenance of the system.
®
Digital Pump Drive Operating ManualMasterflex
iiiMASTERFLEXLIVE
Preface
Safety Precautions
SAFETY
PRECAUTIONS
(continued)
Explanation of
Symbols
WARNING:
Production Use
CAUTION: To avoid electrical shock, the power cord protective grounding
conductor must be connected to ground. Not for operation in wet
locations as defined by EN61010-1.
CAUTIONS: Keep fingers away from rotor while pump is in operation.
Stop pump before loading or unloading tubing.
To reduce the possibility of tipping, use the stacking clip provided with
the unit.
CAUTION: Risk of Danger. Consult Operator’s manual for nature of hazard
!
and corrective actions.
CAUTION: Risk of crushing. Keep fingers away from rotor while pump is
in operation. Stop pump before loading or unloading tubing.
CAUTION: Hot Surface. Do not touch.
CAUTION: Risk of electric shock. Consult Operator’s manual for nature of
hazard and corrective actions.
This product is not designed for, nor intended for use in patient connected
applications; including, but not limited to, medical and dental use, and
accordingly has not been submitted for FDA approval.
Limitation
This product is not designed for, nor intended for use in hazardous duty
areas as defined by ATEX or the NEC (National Electrical Code); including,
but not limited to use with flammable liquids. Consult the factory for
products suitable for these types of applications.
e Digital drive controls the speed of MASTERFLEX® Pump Heads to
provide flow rates from 0.001 to 3400 mL/min for L/S drives and from
0.01 to 19 L/min for I/P drives.
Mount up to 2 (600 rpm) or 4 (100 rpm) MASTERFLEX L/S Pump
Heads and all MASTERFLEX L/S-compatible Pump Heads on L/S drives
and up to 2 MASTERFLEX I/P EASY-LOAD® or MASTERFLEX I/P
Standard Pump Heads or 1 MASTERFLEX High-Performance I/P Pump
Head. Advantages of Peristaltic Pumps:
Application Solutions
• Handle abrasive slurries and corrosive fluids with minimal wear. Ideal for
titanium dioxide or diatomaceous earth filter aid applications.
• No seals in contact with the medium pumped.
• No valves to clog.
• Inner surfaces are smooth and easy to clean.
• Fluid contacts only the tubing or tube material.
• Suction lift and priming up to 8m water column at sea level.
• Low shearing for handling the most shear sensitive of fluids like latex or
fire fighting foam.
• Capable of running dry and pumping fluids with high quantities of
entrained air, such as black liquor soap.
• High volumetric efficiency allows operation in metering or dosing
applications where high accuracy is required.
• Handle extremely viscous fluids.
• Tubing and tube materials are available that are suitable for food and
pharmaceutical use.
®
Digital Pump Drive Operating ManualMasterflex
1-1MASTERFLEXLIVE
Section 1
Introduction
General Description
e MASTERFLEX Digital Peristaltic Pump Drives offer flow rate
capacities from 0.001 mL/min to 19 L/min using MASTERFLEX Stan-
dard, EASY-LOAD® or High-Performance Pump Heads. Even lower flow
rates can be achieved with our multichannel and cartridge Pump Heads.
Features include a small footprint, plus non-stainless steel drives that are
stackable.
e MASTERFLEX digital pump provides a motor speed repeatability of
+/-0.1 percent to maximize productivity in precision liquid dosing, batch
dispensing and filling applications. A turndown ratio up to 6500-to-1,
bidirectional flow and self-priming capabilities allow for smooth, seamless
operation and an extremely broad flow range within one tubing size.
In addition to high accuracy, precision, repeatability and resolution of
speed (or flow rate), the MASTERFLEX drive features a multi-language,
intuitive, man/machine interface with a four-line graphical LCD display
providing direct readout of pump speed (rpm), flow rate (user-selected
units), number of dispenses, and menu options.
e easy-to-use keypad eliminates setpoint overshoot and provides easy
navigation through menu options that include a number of on-screen
programming features.
ese drives use high precision, no-maintenance brushless motors for
improved reliability. is, combined with its high turndown, superior
accuracy, and intuitive interface make the MASTERFLEX drives ideally
suited where ultra-precise, repeatable flow control is required. e pump
accommodates a variety of product fill volumes and batch dispensing profiles, and fluid only contacts the tubing, providing for contamination-free
pumping.
MASTERFLEX pumps are self-priming, can operate dry without damage,
are suitable for most chemicals and contain no valves or seals. See Pump Head and Tubing Guides within this CD.
®
Digital Pump Drive Operating Manual1-2
MasterflexMASTERFLEXLIVE
Before Starting Drive
!
!
Section 2 Installation and Setup
• e drive should be mounted on a flat horizontal surface, and no more
than two (2) Pump Heads should be added for 600 rpm drives or four
(4) pump Heads for 100 rpm drives.
• e ambient air temperature should not exceed 104° F (40° C) and
adequate air flow should be provided for.
CAUTION: Do not block the rear panel of the pump drive. The power
switch must always be easy to access. The power cord must
always be easy to disconnect.
• Tubing should be clean and routed so that bend radii are at a
minimum four (4) times the tube diameter and as short as possible.
WARNING: Turn drive off before removing or installing tubing. Fingers
or loose clothing could get caught in drive mechanism.
• Use a tube size of appropriate diameter for the flow rate and viscosity
required.
• To maintain the best accuracy of flow rates, re-calibrate tubing regularly.
See Tubing Calibration Section of this manual.
• For tubing selection and compatibility, see Tubing Selection Guide within
this CD.
• For Pump Head information, see Pump Head information within this CD.
• When cleaning or performing maintenance, please remove power from
the drive.
CAUTION: The power cord set supplied with your pump drive meets the
!
requirements of the country where you purchased the pump drive. If you
use the pump drive in another country, you must use a power cord set
that meets the requirements of that country.
DANGER: High voltages exist and are accessible. Use extreme caution
when servicing internal components.
®
Digital Pump Drive Operating ManualMasterflex
2-1MASTERFLEXLIVE
Section 2
Installation and Setup
Mounting the
Pump Head
!
• Mount Pump Head and load tubing (See Pump Head information within
this CD). Check to make sure that rollers are clean and free of defects.
CAUTION: When using hazardous chemical and biological agents, take
all suitable protective measures, such as wearing protective glasses
and gloves resistant to the substances used. Follow local and/or
national regulations for safe operation and maintenance of the system.
®
Digital Pump Drive Operating Manual2-2
MasterflexMASTERFLEXLIVE
Section 3 Operation
Turning On the Drive
WARNING: Do not operate the pump drive in a manner not specified in the documentation. Misuse of the pump drive may result in
a hazard and may compromise the safety protection built into the
pump drive. If the pump drive is damaged, turn it off and not use it
until service-trained personnel can check its safety.
1. Plug the power cord into the IEC Connector, located on the rear of the
drive. Plug the opposite end of the power cord into an electrical outlet.
2. Flip the power switch located on the rear of the drive.
3. Upon turning on the drive for the first time you will be prompted to
select a language. e selected language will be set as the default but
can be changed at any time by selecting “LANGUAGE” on the main
menu.
4. After selecting your language, the Main Menu will now appear on the
LCD screen. (NOTE: Each start-up after the initial will revert to the
mode of operation screen previously in use.)
5. If the language is accidently changed and the user would like to reset
it to the default language (English), press and hold the UP/DOWN
(▲/▼) keys during power up.
6. To restore drive to default settings, press and hold the LEFT/RIGHT
(◄/►) keys during power up.
CAUTION: To avoid electrical shock, the power cord protective
grounding conductor must be connected to ground. Not for operation in wet locations as defined by EN61010-1.
CAUTION: Power must be turned off before connecting the
!
external remote control cable to prevent damage to the drive.
WARNING: Tubing breakage may result in fluid being sprayed
!
from pump. Use appropriate measures to protect operator and
equipment.
®
Digital Pump Drive Operating ManualMasterflex
3-1MASTERFLEXLIVE
Section 3
Operation
Control Panel
Figure 3-1. Control Panel
• To navigate all menus on the drive use the directional pad
directly to the right of the LCD screen.
• e (ENTER) key located in the middle of the directional pad is used
to enter or select a highlighted field or option. is key is often referred to
as the ENTER key in this manual.
• e (START/STOP) key located at the top right of the control panel is
used to start and pause the drive. is key is functional only when in one of
the four operating modes: Continuous, Time Dispense, Copy Dispense, or
Volume Dispense. is key is often referred to as the START/STOP key in
this manual.
Priming the Pump
• e (PRIME) key located at the bottom right of the control panel issued
to access the prime (fast forward) function. While pressed, this key operates
the drive at the maximum allowed speed/flow rate and in the direction
shown on the display. When released, the drive returns to its original speed
or flow rate.
1. Mount Pump Head to drive.
2. Insert appropriate tubing into Pump Head.
3. Insert tube inlet into supply fluid.
4. Insert supply outlet into desired container.
5. Turn on pump using switch located on the back of the drive.
6. Press and hold the PRIME key on the drive console to prime the
pump. Priming will stop when key is released.
CAUTION: Keep fingers away from rotor while pump is in
operation. Stop pump before loading or unloading tubing.
®
Digital Pump Drive Operating Manual3-2
MasterflexMASTERFLEXLIVE
Section 3
Operation
Main Menu
CONTINUOUS MODErefer to Continuous Mode in this manual.
TIME DISPENSE MODErefer to Time Dispense Mode in this manual.
COPY DISPENSE MODErefer to Copy Dispense Mode section in this
manual.
VOLUME DISPENSE MODErefer to Volume Dispense Mode section in
this manual.
REMOTE CONTROL MODErefer to Remote Control Mode section in
this manual.
CUMULATIVE VOLUME:e drive stores and displays the cumulative
volume in units based on flow rate units (see SETUP MENU in this section). e Cumulative Volume can also be reset to zero.
NOTE:e Cumulative Volume is dependent on the Tubing Size selected.
(See SETUP MENUin this section.)
SOUNDS:An audible “beep” can be enabled to indicate a keypad press, the
end of a dispense and/or the end of a batch.
AUTOSTART:By default the drive will not restart when power is applied.
To enable this feature select AUTOSTART and then ON. e drive will
now restart when power is reapplied.
DISPLAY CONTRAST: is display can be adjusted using the
UP/DOWN( / ) arrows after selecting this menu item.
LANGUAGE:After selecting this menu, the user will be able to select one
of seven different languages.
NOTE:If the language is accidentally changed and the user would like to
reset it to the default language (English), press and hold the UP/DOWN
( / ) keys when power is reapplied.
DEFAULT SETTINGS: Selecting this menu item and pressing the
ENTER key will restore default settings. To restore drive to default settings
the user may also press and hold the LEFT/RIGHT ( / ) keys when
power is reapplied.
®
Digital Pump Drive Operating ManualMasterflex
3-3MASTERFLEXLIVE
Section 3
Operation
Tubing Calibration 1. Mount Pump Head to drive.
2. Insert appropriate tubing into Pump Head.
3. Insert tube inlet into supply fluid.
4. Insert tube outlet into desired container. Container should be a
graduated container or a container placed on a scale may be used for
increased accuracy.
If using a scale, an acceptable weight to volume conversion for water
is 1 gram = 1 mL.
5. Turn on drive using power switch located on the rear of the drive.
6. Go to the Main Menu or Mode Setup Menu by selecting the SETUP
icon and pressing the ENTER key. Use the UP and DOWN keys
to highlight TUBING CAL in the Main or Setup Menu and press
the ENTER key.
7. Set the drive for the desired flow direction, tube size, and flow rate.
Note that these settings are retained and transferred to other mode
screens when entering or leaving the TUBING CAL screen.
• e flow direction is set using the directional keypad to highlight
the directional arrow. Pressing ENTER will toggle arrow between CW and CCW.
• e tube size is set using the directional keypad to highlight the
tube size field. Press ENTER and use the UP/DOWN keys to
select the tube size. Press ENTER to SAVE the selection and
return to TUBING CAL screen.
• e estimated flow rate is set using the directional keypad to
highlight the flow rate field. Press ENTER and use the
LEFT/RIGHT keys to select the digit to be changed. Use the
UP/DOWN keys to adjust the flow rate value. Press ENTER to
SAVE the setting and EXIT field using arrow keys. e drive will
adjust this flow rate after calibration is complete.
• Note that the calibration volume is fixed and cannot be changed.
8. Press and hold the prime key on the drive console to prime the
pump. Priming will stop when key is released.
9. Place a measuring container at the pump outlet. Highlight the START
field and press the ENTER key. e drive will run based on the
default volume at the estimated flow rate selected.
®
Digital Pump Drive Operating Manual3-4
MasterflexMASTERFLEXLIVE
Section 3
Operation
Tubing Calibration
(continued)
10. Upon completion of the calibration run period, the CAL VOLUME
field will be highlighted. Press the ENTER key and adjust the CAL
VOLUME to the measured quantity. Use the LEFT/RIGHT keys to
select digit to be changed, use the UP/DOWN keys to adjust the value,
and press ENTER to SAVE setting and EXIT the field.
A lower case “c” should now be displayed when the calibrated tubing
size is selected. e volume units will depend on the flow rate units. e
flow rate unit mL/min will result in a volume unit of mL; oz/min will
result in a volume unit of oz.
Tubing Calibration Notes
• If the drive is stopped during calibration, empty the container and restart the procedure.
• Calibration time at maximum allowable flow rate (default max flow
rate) is 5-10 seconds and at minimum allowable flow rate (approximately 4% of the maximum flow rate) is 4 minutes. Select the
CUSTOM tube size for other tubing sizes or lower flow rates.
• Minimum and maximum flow rates will change after a tubing calibration due to a re-calculation of the vol/rev.
• Optimum results are best obtained after tubing has been broken in by
running in pump for at least 10 minutes. Steps 8-10 can be repeated
as necessary to optimize the accuracy of the tubing cal.
CAL RUN TIME FORMULA
60 / (flow rate [mL/min] / cal volume [mL]) = cal run time (seconds)
INVALID CAL RUN TIME EXAMPLE
• tube size 13 flow rate range is 0.006 mL/min – 36.0 mL/min
• at flow rate of 1 mL/min, cal run time calculation is as follows:
60 / (1 mL/min / 6 mL) = 360 seconds 360 seconds exceed the max
run time of 4 minutes (240 seconds)
®
Digital Pump Drive Operating ManualMasterflex
3-5MASTERFLEXLIVE