Masterflex 07575-00 Instruction Manual

OPERATING MANUAL:
PUMP DRIVE
MODE D’EMPLOI:
ENTRAÎNEMENT DE POMPE
BETRIEBSANLEITUNG:
PUMPENANTRIEB
MANUAL DE FUNCIONAMIENTO:
MANUALE DI ISTRUZIONI:
UNITÀ DI CONTROLLO POMPA
07575-00 pump drive.
Model No. N° de modèle Modellnummer Número de modelo Modello nº
07575-00
Cole-Parmer Instrument Co.
1-800-MASTERFLEX (627-8373) (U.S. and Canada only) 11 (847) 549-7600 (Outside U.S.) • (847) 549-7600 (Local) • www.masterflex.com
Barnant Company
1-800-637-3739 (U.S. and Canada only) 11 (847) 381-7050 (Outside U.S.) • (847) 381-7050 (Local) • www.barnant.com
A-1299-5083
Edition 02
2
EU Declaration of Conformity
Name of Apparatus: Masterflex®Digital Console L/S®Pump Drive Model Number: 07575-00 Description of Apparatus: Digital-controlled, variable speed, peristaltic pump motor drive
with remote I/O. Used with pump head and tubing to pump fluids.
Barnant Company declares that the above model is in conformity to the following harmonized standards and directives:
Applicable Applicable Manufacturer’s Directives Specifications Report Number
73/23/EEC EN61010-1/A2:1995 TR0389 93/68/EEC
89/336/EEC EN61326-1/A2:2001 TR0390 92/31/EEC 93/68/EEC
The last two digits of the year in which the current configuration of the above model was assessed per the Low Voltage Directive is: 03.
Manufacturer: Barnant Company Division
Cole-Parmer Instrument Company 28W092 Commercial Avenue Barrington, IL 60010-2392 USA Tel.: 847-381-7050
Manufacturer’s Signature: 30 June 2003
Scott Akins Date Director of Engineering
3
TABLE OF CONTENTS
Title Page
SAFETY PRECAUTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
CONTROL/DISPLAY FUNCTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
SETUP AND DRIVE OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Automatic Start Enable/Disable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
CALIBRATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Maximum Flow Rate (“OTHER” Tubing) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
DISPENSE/COPY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Keypad Lockout Enable/Disable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
REMOTE CONTROL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Remote Control Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
TROUBLESHOOTING AND MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Fuse Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Shaft Seal Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Replacement Parts and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
WARRANTY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
PRODUCT RETURN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
TECHNICAL ASSISTANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
SAFETY PRECAUTIONS
DANGER: High voltages exist and are accessible in the Console Drive. Use extreme caution
when servicing internal components.
WARNINGS: Tubing breakage may result in fluid being sprayed from pump. Use appropriate measures
to protect operator and equipment. Turn drive off before removing or installing tubing. Fingers or loose clothing could get caught
in drive mechanism.
CAUTIONS: Power must be turned off before connecting the external remote control cable to prevent
damage to the drive. Do not stack drives. Keep 3” minimum distance around and above drive for proper cooling.
WARNING: PRODUCT USE LIMITATION
This product is not designed for, nor intended for use in, patient-connected applications, including, but not limited
to, medical and dental use, and, accordingly, has not been submitted for FDA approval. If drive is used in a manner
not specified in this manual the protection provided by the equipment may be impaired.
4
INTRODUCTION
The Console Drive controls the speed of Masterflex®L/S®pump heads to provide flow rates from 0.10 to 3400 mL/min or 0.001 to 54 gallons/hr.
Mount up to 2 Masterflex®L/S®pump heads and all Masterflex-compatible pump heads.
CONTROL/DISPLAY FUNCTIONS
Press keys to activate function. Use INC/DEC ( ▲,▼) arrow keys to correct/change a flashing display. Press any key to enter new values.
A. DOWN ARROW (DECREMENT)—Decrease value of a flashing display. B. UP ARROW (INCREMENT)—Increase value of a flashing display. C. FLOW CONTROL—Set flow rate for selected tubing size. To change flow rate, press or arrows.
(If pump is running, its speed will change with new settings.)
D. START/STOP—Start/Stop motor. E. PRIME—Run pump at full speed to fill or clear lines. F. TOTAL VOLUME—Display cumulative dispensed volume. G.DIRECTION—To change motor direction. H. MODE SELECTINT for internal control; mA for remote current control; V for remote voltage control.
I. UNITS—Select between metric (milliliters and liters) and English (U.S. gallons) displayed units. J. DISPENSE/COPY—Select dispensed volume, copy volume, or the dispense/copy interval. K. CAL CONTROL—Refine built-in calibration using a measured volume. L. TUBING SELECT—Select tubing size.
AB C D E
F
LKJIHG
Figure 1
5
SETUP AND DRIVE OPERATION
1. Mount pump head and load tubing (see pump head manual). NOTE: The pump head may be oriented in any of four (4) positions, 90° apart. Rotation is accomplished by removing and
reinstalling the front plate, to which pumps are mounted. See “Troubleshooting and Maintenance, Shaft Seal Inspection” for directions.
2. Turn drive on and select tubing size by pressing “TUBING SELECT.”
NOTE: If CAL LED is lit, that tubing size has been previously field calibrated. If LED is not lit, the drive is operating with the built-in factory calibration. To clear a field calibration, press and hold the CAL key until the CAL light goes out. This will take about 3 seconds. To recalibrate for better accuracy, see Calibration section.
3. Select MODE (INT, mA, V).
4. Select motor DIRection (CW or CCW).
5. PRIME and CALibrate the pump (if required).
6. Press FLOW key and watch display to set the flow rate with INC/DEC (, ) keys.
7. Press START/STOP key to begin pumping. NOTE: While in INTernal mode, drive will not restart automatically after a brownout or powerout condition unless operator
changes default setting. If speed is being controlled by an external signal, drive will automatically start with a non-zero speed command in power-up.
AUTOMATIC START ENABLE/DISABLE (Internal mode only)
Press and hold START/STOP on power-up. After five (5) seconds, display will change to all dashes. Then, while holding START/STOP, press PRIME five (5) times. Display will flash “ON” or “OFF”. Use INC/DEC (▲, ▼) arrow keys to enable or disable automatic start option. Press any other key to exit. When “ON” is selected, drive will start automatically at power-up if it was “ON” when powered down.
CALIBRATION
Use only MASTERFLEX precision tubing with MASTERFLEX pumps to ensure optimum performance. Use of other tubing may void applicable warranties.
1. Select correct tubing size and flow rate.
2. Press CAL; calibration volume appears.
3. Press START/STOP; the pump will use its stored memory to dispense the specified calibration sample quantity.
The pump will stop automatically.
4. Weigh/measure the sample.
5. Use INC/DEC (, ) arrow keys to correct the flashing display. NOTE: If the adjusted calibration is too great, “Err” will appear in the display. If this occurs, press the CAL control and
repeat the calibration procedure. The microprocessor will retain one special calibration value per tubing size, even when power is turned off. The next calibration will replace the existing value.
6. Press TUBING SELECT to exit the calibration cycle.
MAXIMUM FLOW RATE (“OTHER” Tubing)
1. To set the maximum flow rate for non-standard pump heads or “OTHER” tubing sizes, press CAL, then FLOW.
The maximum flow rate will then flash on the display.
2. Use INC/DEC (, ) arrow keys to change flow rate.
3. Press TUBING SELECT to exit
6
DISPENSE/COPY
A first press of the DISP key results in the last entered dispense volume being displayed. The VOL annunciator will illuminate and flash. The INC/DEC (, ) keys are used to change the dispense volume, if desired. The START/STOP key then initiates delivery of the set volume. The amount remaining to be dispensed will be displayed during countdown. Resetting cumulative dispensed volume can be done by pressing and holding the TOTAL VOLUME key until zero is displayed (3 seconds). Display flashes “99999” when cumulative volume is greater than 99999. The dispense function is exited by the FLOW key.
A second press of the DISP key causes the COPY annunciator to illuminate and flash. The START/STOP key is then used to deliver the desired volume without the need to know the volume in specific units. A third press of the DISP key enters the volume dispensed. The COPY annunciator stops flashing. The START/STOP key is then used to initiate delivery of the copied volume. The number of copies dispensed is displayed after each dispense. The maximum number of copies is 99999. The START/STOP key is used to pause the copy dispense during dispensing; copy dispense can then be continued using the START/STOP key.
A fourth press of the DISP key results in the last entered dispense interval being displayed. The SEConds annunciator will illuminate and flash. The INC/DEC (▲, ▼) keys are used to change the time interval between dispenses, if desired, from 0 to 99999 seconds. The START/STOP key then initiates delivery for the set volume, with the drive automatically initiating a new dispense after each time out. The remaining time will be displayed during countdown. The START/STOP key is used to stop the dispense cycle. A time of 0 seconds (default) will require initiation of each dispense through the ST ART/ST OP key or the Remote START/STOP contact closure. Pressing the DISP key a fifth time exits this mode.
KEYPAD LOCKOUT ENABLE/DISABLE
Press/hold FLOW. After five (5) seconds, display changes to all dashes. While holding FLOW, press PRIME five (5) times.
REMOTE CONTROL
Selectable input (0–20 mA, 4–20 mA, 0–10V DC ) ±0.5% linearity control ±50V common mode range with earth ground START/STOP; CW/CCW; PRIME via contact closure
REMOTE CONTROL SETUP
1. Place the power switch in the off position.
CAUTION: Power must be turned off before connecting the external remote control cable to prevent
damage to the drive.
2. Connect the cable from the external remote control to the mating receptacle on the rear panel.
3. Select type of remote control input and output required as follows: a. Press and hold the MODE control while turning the power switch to the on position. After two seconds, release
the MODE control. The initial display will show: “inP”. After two seconds, the display shows either 0–20 or 4–20.
NOTE: Press the INC () or DEC () arrow keys to select between 4–20 and 0–20 for current loop control.
b. Press the MODE control again. The initial display will show: “out”. After two seconds, the display shows either
0–10, 0–20, or 4–20.
NOTE: Press the INC () or DEC () arrow keys to select between 4–20, 0–20, or 0–10 for current loop
or voltage output.
4. Press the MODE control to select mode of operation. The LEDs indicate the selected mode. Select either mA or V.
7
NOTE: If using only remote START/STOP, PRIME and/or CW/CCW, the MODE control can be set to any of the
three positions.
5. To adjust the voltage or current scaling for other than zero to full scale, Press and hold the MODE key while pressing the FLOW key. This display will show “LO” for 2 seconds and then a flow rate. Use the INC/DEC (▲, ▼) arrow keys to set the minimum control level. Press the FLOW key again. The display will show “HI” for 2 seconds and then a flow rate. Use the INC/DEC (▲, ▼) arrow keys to set the maximum control level. Press any other key to exit. The same scaling will be used for both input and output levels.
NOTE: Tubing size should be selected before adjusting the voltage or current scaling.
To provide sufficient hysteresis between minimum speed and off, a 0.2 mA and 0.1 V offset is added to the current and voltage inputs and outputs as shown in the following table. This offset is only added when the input/output scaling is set with a minimum speed of zero (default)
NOTE: Colors are those of Remote Cable 77300-32.
Speed (rpm) 4-20 mA In/Out 0-20 mA In/Out 0-10 V In/Out 0 4-4.2 mA 0-0.2 mA 0-0.1 V 1 4.2263 mA 0.233 mA 0.1165 V 600 20 mA 20 mA 10V
07575-00
Figure 2
START/STOPCW/CCWPRIME
INPUT
0-20mA; 4-20mA
INPUT 0-10V
OUTPUT
0-20mA; 4-20mA
OUTPUT
0-10V
RED/YELLOW
BLUE
GREEN
YELLOW
WHITE
ORANGE
BLACK
BROWN
VIOLET
RED
14
6
4
8
3
5
1
7
9
2
10
12
11
13
15
16
17
18
GREY
TAN
PINK
RED/GREEN
RED/BLACK
N.C.
N.C.
N.C.
TACH OUTPUT
MOTOR RUNNING
N.O. CONTACT
MOTOR RUNNING
N.C. CONTACT
8
TROUBLESHOOTING AND MAINTENANCE
FUSE REPLACEMENT
1. Place the power switch in the off position.
2. Disconnect the AC power input line cord from the attached line and receptacle.
3. Remove and check the fuse and replace if defective
4. Reconnect the AC power input line cord to the receptacle.
SHAFT SEAL INSPECTION
1. Remove any pump(s) attached to the front
of the drive. Clean any foreign material from the outside diameter of the drive shaft.
2. Remove the (4) screws (Item B) that hold the front plate assembly (Item A) to the drive, and pull the front plate
assembly off the drive. #8-32 screws may be installed in the pump-mounting holes to provide handles for pulling the plate assembly off. Retain Item B screws for Step 8. DO NOT substitute other screws.
3. Turn the front plate over so that
the seal side is visible. Wipe the elastomeric seal lips with a clean cloth to remove any grease and foreign material.
4. Inspect the elastomeric seal lips
for tears or cuts or missing material. If any of the mentioned conditions exist, replace the seal assembly using the 07575-01 replacement seal kit.
5. Wipe the exposed part of the
drive shaft with a clean cloth. Wipe from the drive outward, to remove all grease and foreign matter.
6. Inspect the shaft surface in the area touched by the seal. Look for a rough finish, or grooves parallel to the shaft
length. If the shaft end is worn or damaged, as described above, replace the gear and shaft with the 07575-02 kit. A polished groove, concentric to the outside of the shaft, is not a defect, as long as the groove is no more than
0.002 inches deep.
7. Prior to re-assembly, re-lubricate the shaft and the seal with the food-grade lubricant provided with the unit.
CAUTION: Do not contaminate the lubricant in the container, on the shaft or on the seal with foreign material. Failure to observe this precaution may result in damage to the seal and premature failure of the seal.
8. Slide the front plate assembly back over the shaft and onto the locating pins, in the orientation desired.
(4 configurations, each 90 degrees of rotation apart, are possible). Reinstall the 4 screws, removed in Step 2 above. CAUTION: No foreign matter should be allowed under the gasket on the back of the front plate or under the heads of the screws. Failure to observe this precaution may result in leakage during washdown of the drive.
A. External receptacle for
remote control
B. IEC 320 Power cord C. T3.15A Fuse CAUTION:
Do not substitute.
D. Power switch—All
settings are retained in memory.
A. Front plate assembly B. Front plate assembly
screws (4)
Figure 3
Figure 4
AB
C
D
9
SYMPTOM
A. Motor does not rotate.
Display does not light.
B. Motor does not rotate.
Display lights.
CAUSE
A. No power.
B1. Defective remote control.
B2. MODE control not
properly set.
REMEDY
1. Check fuse and replace if defective (Fig. 3).
2. Check that unit is plugged into a live line.
3. Check connection of power cord.
4. Check the line cord for continuity and replace if defective.
5. Return for servicing.
1. Place power switch in off position.
2. Check that remote cable connector is inserted fully into the receptacle (Fig. 2 and 3)
3. If motor still does not rotate, select INTernal with the MODE control and press the START/STOP control (Fig. 1).
4. If the motor rotates, replace the remote control with similar unit. If motor does not rotate, return drive for servicing.
1. Check that the MODE control is set to INT for operation with front panel control or to mA or V for operation with remote control (Fig. 1).
2. If motor still does not rotate, return for servicing.
TROUBLESHOOTING
Troubleshooting continued on page 10.
10
ERROR MESSAGE
“Err 2”
“Err 3” “Err 5” “Err 12”
“Err 7”
“Err 10” “Err 11”
“Err 14”
All other errors
CAUSE
Motor over-speed
Overload
Bad data. Operator parameters set to default values.
Voltage out of range
Over temperature
Internal error or failure.
REMEDY
1. Clear by pressing any key.
2. Check for proper tube loading and pump operation.
3. Return unit for repair if the error persists.
1. Clear by pressing any key.
2. Check for proper tube loading and pump operation.
3. Return unit for repair if the error persists.
1. Clear by pressing any key.
2. Reprogram operator parameters.
3. Return unit for repair if the error persists.
1. Clear by pressing any key.
2. Check that AC line voltage is within specified voltage ranges.
3. Return unit for repair if AC line voltage is correct and the error persists.
1. Check for heat sources or obstructions to cooling.
2. Check for proper tube loading and pump operation.
3. Allow unit to cool. Clear by pressing any key.
4. Return unit for repair if no cause for overheating is found and the error persists.
1. Clear (if possible) by turning power off and on.
2. Return unit for repair if the error persists.
TROUBLESHOOTING (Continued)
If an error message is displayed, refer to the following list for possible corrective actions you can take. If these do not correct the problem, contact your dealer.
11
CLEANING
Keep the drive enclosure clean with mild detergents. Do not immerse.
REPLACEMENT PARTS & ACCESSORIES
Description Part Number Description Part Number
Fuse—T3.15A 5 x 20 mm 77500-25 Footswitch 07595-43 Hand-held remote control 07592-83 Remote control cable, Replacement Seal Kit 07575-01 25 ft (7.62 m) 77300-32 Replacement Gear and Shaft Kit 07575-02
SPECIFICATIONS
Output:
Speed: 1 to 600 rpm Torque output, Maximum:
Continuous 180 oz-in (13 kg•cm) Start-up 540 oz-in (39 kg•cm)
Speed regulation:
Line ±0.25% F.S. Load ±0.25% F.S.
Drift ±0.25% F.S. Display: Five-digit, seven-segment LED Remote outputs: Voltage speed output (0–10V DC @ 1 kmin)
Current speed output (0–20 mA or 4–20 mA @ 0–600 ) Tach output (TTL, 10 to 6000 Hz, 50% duty cycle 10 Hz/rpm) Motor running output
(N.O. & N.C. contact closure, 1A @ 28V AC/DC)
Input:
Supply voltage limits: Dual voltage—Automatically selected
90 to 130 Vrms @ 50/60 Hz, or 200 to 260 Vrms @ 50/60 Hz Current, max.: 2.2A @ 115 Vrms, or 1.1A @ 230 Vrms Remote Inputs: START/STOP, CW/CCW, PRIME (Contact closure)
Voltage input (0–10V DC @ 10 k)
Current input (0–20 mA or 4–20 mA @ 250 )
Construction:
Dimensions (L x W x H): 14.0” x 9” x 9.3” (35 cm x 23 cm x 24 cm) Weight: 22 lb (10 kg) Enclosure Rating: IP 66 per IEC 60529/NEMA 4 X - indoor use
Environment:
Temperature, Operating: 0° to 40°C (32° to 104°F) Temperature, Storage: –25° to 65°C (–13° to 149°F) Humidity (non-condensing): 10% to 100% Altitude: Less than 2000 m Pollution Degree: Pollution Degree 3
(Indoor use—sheltered locations) Chemical Resistance: Enclosure is 316 stainless steel
Compliance: UL508C, CSA C22.2, No. 14
(For CE Mark):
EN61010-1 (EU Low Voltage Directive) and
EN61326-1 (EU EMC Directive
12
WARRANTY
Use only MASTERFLEX precision tubing with MASTERFLEX pumps to ensure optimum performance. Use of other tubing may void applicable warranties.
The Manufacturer warrants this product to be free from significant deviations from published specifications. If repair or adjustment is necessary within the warranty period, the problem will be corrected at no charge if it is not due to misuse or abuse on your part, as determined by the Manufacturer. Repair costs outside the warranty period, or those resulting from product misuse or abuse, may be invoiced to you.
The warranty period for this product is noted on the Warranty Card.
PRODUCT RETURN
To limit charges and delays, contact the seller or Manufacturer for authorization and shipping instructions before returning the product, either within or outside of the warranty period. When returning the product, please state the reason for the return. For your protection, pack the product carefully and insure it against possible damage or loss. Any damages resulting from improper packaging are your responsibility.
TECHNICAL ASSISTANCE
If you have any questions about the use of this product, contact the Manufacturer or authorized seller.
Cole-Parmer Instrument Co.
625 East Bunker Court
Vernon Hills, Illinois U.S.A. 60061-1844
1-800-MASTERFLEX (627-8373) (U.S. and Canada only)
11 (847) 549-7600 (outside U.S.)
(847) 549-7600 (Local)
FAX (847) 247-2929 (U.S. and Canada only)
11 (847) 549-1700 (Fax outside U.S.)
www.masterflex.com
e-mail: techinfo@coleparmer.com
Barnant Company
28W092 Commercial Ave.
Barrington, Illinois U.S.A. 60010-2392
1-800-637-3739 (U.S. and Canada only)
11 (847) 381-7050 (outside U.S.)
(847) 381-7050 (Local)
11 (847) 381-7053 (Fax outside U.S.)
(847) 381-7053 (Local Fax)
www.barnant.com
e-mail: barnant@barnant.com
13
Francais
Déclaration de conformité EU
Nom de l’Appareil: Masterflex®Console numérique L/S®Entraînement de Pompe Numéro de Modèle: 07575-00 Description de l’Appareil: Moteur d’entraînement de pompe péristaltique commandé
numériquement, à vitesse variable avec une interface distante. Utilisé avec une tête de pompe et des conduites pour pomper des liquides.
La Compagnie Barnant déclare que le modèle ci-dessus est en conformité avec les standards et directives harmonisées suivantes :
Directives Spécifications Numéro de Rapport Applicables Applicables du Fabricant
73/23/EEC EN61010-1/A2:1995 TR0389 93/68/EEC
89/336/EEC EN61326-1/A2:2001 TR0390 92/31/EEC 93/68/EEC
Les deux derniers chiffres de l’année au cours de laquelle la configuration courante pour le modèle ci-dessus a été évalué pour la Directive Basse Tension est : 03.
Fabricant: Barnant Company Division
Cole-Parmer Instrument Company 28W092 Commercial Avenue Barrington, IL 60010-2392 USA Tel.: 847-381-7050
Signature du Fabricant: 30 juin 2003
Scott Akins Date Directeur Technique
14
TABLE DES MATIERES
Titre Page
MESURES DE SECURITE .......................................................................................................................................... 3
INTRODUCTION........................................................................................................................................................ 14
FONCTIONS DE CONTRÔLE/AFFICHAGE.............................................................................................................. 15
CONFIGURATION ET FONCTIONNEMENT DE L’ENTRAÎNEMENT ...................................................................... 15
Démarrage Automatique Activé/Désactivé .......................................................................................................... 16
CALIBRAGE .............................................................................................................................................................. 16
Débit Maximum (Conduites “AUTRES” [“OTHER”]) ............................................................................................ 16
DISTRIBUTION/RÉPÉTITION.................................................................................................................................... 17
Verrouillage du pave numérique Activé/Désactivé .............................................................................................. 17
COMMANDE À DISTANCE........................................................................................................................................ 17
Configuration de la Commande à Distance.......................................................................................................... 17
DEPANNAGE ET MAINTENANCE ............................................................................................................................ 17
Remplacement de Fusible.................................................................................................................................... 19
Inspection de la Boîte d'étanchéité........................................................................................................................19
Dépannage .......................................................................................................................................................... 19
Nettoyage ............................................................................................................................................................ 20
Pièces détachées et accessoires ........................................................................................................................ 22
SPECIFICATIONS...................................................................................................................................................... 22
GARANTIE ................................................................................................................................................................ 23
RETOUR PRODUIT .................................................................................................................................................. 23
ASSISTANCE TECHNIQUE ......................................................................................................................................23
MESURES DE SECURITE
DANGER: Il e xiste des hautes tensions et elles sont accessibles depuis la Console de l’entraînement.
Prenez toutes les précautions requises lors de la maintenance/dépannage de composants internes.
AVERTISSEMENTS: La rupture des conduits peut entraîner la fuite de liquide depuis la pompe. Mettez en
place les mesures adéquates pour protéger l'opérateur et les équipements. Coupez l’entraînement avant de retirer ou d’installer des conduites. Les doigts ou des vêtements amples peuvent se prendre dans le mécanisme d’entraînement.
PRECAUTIONS: L’alimentation doit être coupée avant de connecter la télécommande externe afin de ne
pas endommager l’entraînement. N’empilez pas les entraînements de pompe. Gardez au moins 7 à 8 cm autour et au dessus de l’entraînement pour un refroidissement efficace.
AVERTISSEMENT : LIMITE D’UTILISATION DU PRODUIT
Ce produit n’a pas été conçu ou prévu pour être utilisé dans des applications connectées à des malades, incluant, mais ne se limitant pas à, l’utilisation médicale ou dentaire et, par conséquent, n’a pas été soumis à une approbation FDA (Food and Drug Administation - Fédération américaine des aliments et drogues). Si l’entraînement est utilisé d’une façon non spécifiée dans ce manuel, la protection fournie par cet équipement pourrait être endommagée.
15
INTRODUCTION
La Console d’Entraînement contrôle la vitesse des têtes de pompe Masterflex®L/S®pour fournir des débits de
0.10 à 3400 mL/min. Extensible à 2 têtes de pompe Masterflex
®
L/S®et à toutes les têtes de pompe compatibles Masterflex.
FONCTIONS DE CONTRÔLE/AFFICHAGE
Appuyez sur les touches pour activer une fonctionnalité. Utilisez les touches flèches INC/DEC (, ) pour corriger/modifier une valeur affichée clignotante. Appuyez sur n’importe quelle touche pour rentrer de nouvelles valeurs.
A. FLÈCHE VERS LE BAS (DECREMENTER) - Décrémente une valeur affichée clignotante. B. FLÈCHE VERS LE HAUT (INCREMENTER) - Incrémente une valeur affichée clignotante. C. CONTRÔLE DU DEBIT (FLOW CONTROL) - Règle le débit pour la taille de conduite sélectionnée. Pour modifier
le débit, appuyez sur les flèchesou. (Si la pompe est en marche, sa vitesse changera avec les nouveauxré glages.)
D. MARCHE/ARRÊT (START/STOP) - Démarre/Arrête le moteur. E. AMORCAGE (PRIME) - Fait marcher la pompe à vitesse maximale pour remplir ou vider les conduites. F. VOLUME TOTAL (TOTAL VOLUME) - Affiche le volume total distribué. G. DIRECTION (DIRECTION) - Pour changer le sens de rotation du moteur. H. SELECTION DU MODE (MODE SELECT) - INT pour contrôle interne ; mA pour contrôle de courant à distance ;
V pour contrôle de tension à distance.
I. UNITES (UNITS) - Choix des unités affichées; système métrique (millilitres et litres) ou système Anglais (gallons US). J. DISTRIBUTION/RÉPÉTITION (DISPENSE/COPY) - Sélection du volume distribué, du nombre de répétitions de
cette distribution ou de l’intervalle de distribution.
K. CONTROLE DE CALIBRAGE (CAL CONTROL) - Affine le calibrage en usine en utilisant un volume mesuré. L. SELECTION DES CONDUITES (TUBING SELECT) - Sélectionne la taille des conduits.
AB C D E
F
LKJIHG
Figure 1
16
CONFIGURATION ET FONCTIONNEMENT DE L’ENTRAÎNEMENT
1. Monter la tête de pompe et remplir les conduites (voir le manuel de la tête de pompe). REMARQUE: La tête de pompe peut être orientée dans n’importe lequel des quatre (4) points cardinaux (tous les 90°). La
rotation est effectuée en enlevant puis en réinstallant la plaque frontale sur laquelle les pompes sont montées. Voir “DEPANNAGE ET MAINTENANCE, Inspection de la Boîte d'étanchéité” pour instructions.
2. Mettez l’entraînement en marche puis sélectionnez la taille des conduits en appuyant sur “SELECTION DES CONDUITES”. REMARQUE: Si la diode électroluminescente CAL est allumée, cette taille de conduite a été préalablement calibrée sur le
terrain. Pour effacer un calibrage, appuyez et maintenez la pression sur la touche CAL jusqu’à ce que la diode CAL s’éteigne. Cela prendra environs 3 secondes. Pour re-calibrer afin d’avoir plus de précision, voir le chapitre Calibrage
3. Sélectionnez le MODE (INT, mA, V).
4. Sélectionnez le sens de rotation (DIR) du moteur CW (sens des aiguilles d'une montre) ou CCW (sens inverse des
aiguilles d'une montre).
5. AMORCEZ (PRIME) et CALibrez la pompe (si nécessaire).
6. Appuyez sur la touché DEBIT (FLOW) et observez l’affichage pour régler le débit à l’aide des touches INC/DEC (▲ ,▼).
7. Appuyez sur MARCHE/ARRÊT (START/STOP) pour commencer le pompage. REMARQUE: En mode INTerne (INTernal mode), l’entraînement ne redémarrera pas automatiquement après une chute de
tension ou une coupure de courant à moins que l’opérateur ne change les réglages par défaut. Si la vitesse est contrôlée par un signal externe, l’entraînement démarrera automatiquement avec une vitesse non nulle à la mise sous tension.
DEMARRAGE AUTOMATIQUE ACTIVE/DESACTIVE (Mode Interne seulement)
Appuyez puis maintenez la pression sur la touche MARCHE/ARRÊT (START/STOP) à la mise sous tension. Après cinq (5) secondes, l’affichage changera pour n’être qu’une suite de tirets. Puis, en restant sur la touche MARCHE/ARRÊT (START/STOP), appuyez sur AMORCAGE (PRIME) cinq (5) fois. L’affichage indiquera en clignotant “ON” ou “OFF”. Utilisez INC/DEC (▲ ,▼) pour activer ou désactiver l’option de démarrage automatique. Appuyez sur n’importe quelle autre touche pour quitter. Lorsque “ON” est sélectionné, l’entraînement démarrera automatiquement à la mise sous tension s’il était dans la position “ON” lors de la mise hors tension.
CALIBRAGE
Utilisez uniquement des conduites de précision MASTERFLEX avec des pompes MASTERFLEX pour
assurer des performances optimales. L’utilisation d’autres conduites peut annuler la garantie.
1. Sélectionnez la taille de conduites et le débit adéquats.
2. Appuyez sur CAL; le volume de calibrage apparaîtra.
3. Appuyez sur MARCHE/ARRÊT (START/STOP) ; la pompe utilisera les données en mémoire pour fournir la quantité
échantillon de calibration spécifiée.
4. Pesez/mesurez l’échantillon.
5. Utilisez INC/DEC (,)pour corriger la valeur affichée clignotante. REMARQUE: Si le calibrage choisi est trop important, “Err” apparaîtra sur l’affichage. Si cela se produit, appuyez sur CAL
et recommencez la procédure de calibrage. Le microprocesseur mémorisera un calibrage spécifique par taille de conduite, même si le courant est coupé. Le prochain calibrage remplacera la valeur actuelle.
6. Appuyez sur SELECTION DES CONDUITES (TUBING SELECT) pour sortir du cycle de calibrage.
17
DEBIT MAXIMUM (Conduites “AUTRES” [“OTHER”])
1. Pour régler le débit maximum pour des têtes de pompe non standards ou des tailles de conduites “AUTRES” (“OTHER”), appuyez sur CAL puis DEBIT (FLOW). Le débit maximum clignotera sur l’affichage.
2. Utilisez les touches INC/DEC (,) pour modifier le débit.
3. Appuyez sur SELECTION DES CONDUITES (TUBING SELECT) pour sortir.
DISTRIBUTION/RÉPÉTITION
Une première pression sur la touche DISP affiche la dernière valeur du volume distribué. L’annonciateur VOL s’allumera et clignotera. Les touches INC/DEC (▲ ,▼) permettent de modifier le volume distribué, si nécessaire. La touche MARCHE/ARRÊT (START/STOP) lance la distribution du volume pré réglé. La quantité restante à distribuer sera affichée durant le compte à rebours. Remettre à zéro la quantité totale distribuée est possible en appuyant et en maintenant la pression sur la touche VOLUME TOTAL (TOTAL VOLUME) jusqu’à ce que zéro soit affiché (3 secondes). L’affichage indique “99999” en clignotant si le volume total distribué est plus grand que 99999. On quitte la fonction de distribution en appuyant sur la touche DEBIT (FLOW).
Une seconde pression sur la touche DISP fait s’allumer et clignoter l’annonciateur COPY. La touche MARCHE/ARRÊT (START/STOP) est alors utilisée pour distribuer la quantité désirée sans avoir à connaître l’unité utilisée pour le volume. Une troisième pression sur la touche DISP rentre le volume distribué. L’annonciateur COPY arrête de clignoter. La touche MARCHE/ARRÊT (START/STOP) est alors utilisée pour lancer la distribution du volume répété. Le nombre de répétitions est alors affiché après chaque distribution. Le nombre maximum de répétitions est 99999. La touche MARCHE/ARRÊT (START/STOP) est utilisée pour mettre la distribution en pause ; la répétition de distribution peut alors être remise en route à l’aide de la touche MARCHE/ARRÊT (START/STOP).
Une quatrième pression sur la touché DISP affiche le dernier intervalle de distribution. L’annonciateur SECondes s’allumera et clignotera. La touche INC/DEC (▲ ,▼) permet alors de modifier l’intervalle de temps entre chaque distribution, de 0 à 99999 secondes. La touche MARCHE/ARRÊT (START/STOP) lance la distribution pour le volume désiré, l’entraînement initiera une nouvelle distribution à l’expiration de chaque intervalle de temps. Le temps restant est affiché durant le compte à rebours. La touche MARCHE/ARRÊT (START/STOP) permet de stopper le cycle de distribution. Un intervalle de temps de 0 secondes (valeur par défaut) nécessitera le démarrage de chaque distribution à l’aide de la touche MARCHE/ARRÊT (START/STOP) ou par la fermeture de l’interrupteur MARCHE/ARRÊT (START/STOP) situé sur la pompe. Appuyez sur la touche DISP une cinquième fois pour quitter ce mode.
VERROUILLAGE DU PAVE NUMERIQUE ACTIVE/DESACTIVE
Appuyez/maintenez la touche DEBIT (FLOW). Après cinq (5) secondes, une série de tirets seront affichés. Tout en main­tenant la pression DEBIT (FLOW), appuyez sur AMORCAGE (PRIME) cinq (5) fois.
COMMANDE À DISTANCE
Entrée sélectionnable (0–20 mA, 4–20 mA, 0–10 V c.c.) Contrôle de linéarité à ±0,5 % Fonctionnement nominal avec prise de terre de ± 50V MARCHE/ARRÊT; CW/CCW; AMORÇAGE par fermeture des contacts
CONFIGURATION DE LA COMMANDE A DISTANCE
1. Mettez l’interrupteur d’alimentation en position hors tension.
DANGER: L’alimentation doit être coupée avant de connecter la télécommande externe afin de ne pas
endommager l’entraînement.
2. Raccordez le câble venant de la télécommande à la prise femelle située sur le panneau arrière.
18
3. Sélectionnez le type d’entrée/sortie de la télécommande en procédant comme suit : a. Appuyer sur la touche MODE et maintenez la enfoncée, tout en tournant le commutateur d’alimentation en
position de mise sous tension. Relâchez la touche MODE au bout de deux secondes. “inP” (entrée) s’affiche initialement. Au bout de deux secondes, 0–20 ou 4–20 s’affiche.
REMARQUE : Appuyez sur la touche INC
() ouDEC () pour sélectionner 4–20 et0–20 pour le réglage de boucle de courant.
b. Ré appuyez sur la touche MODE. “out” (sortie) s’affiche initialement. 0–20 ou 4–20 s’affiche au bout de deux
secondes.
REMARQUE : Appuyez sur INC
() ou DEC () pour sélectionner 4–20, 0–20 ou 0–10 comme boucle de courant ou
tension de sortie.
4. Appuyez sur la touche MODE pour sélectionner le mode de fonctionnement. Les témoins indiquent le mode sélectionné.
Sélectionnez mA ou V.
REMARQUE : si seule la commande MARCHE/ARRÊT (START/STOP), AMORÇAGE (PRIME) et/ou CW/CCW à distance est utilisée, la commande MODE peut être réglée dans l’une de ces trois positions.
5. Pour régler la tension ou la graduation de courant à une valeur autre que zéro ou graduation complète, appuyez et
maintenez la touche MODE tout en appuyant sur la touche DEBIT (FLOW). Il sera affiché “LO” pendant 2 secondes et puis un débit. Utilisez la touche INC/DEC
(, ) pour régler le niveau minimum de contrôle. Appuyez sur DEBIT (FLOW) une
nouvelle fois. Il sera affiché “HI” pendant 2 secondes et puis un débit. Utilisez la touche INC/DEC
(, ) pour régler le
niveau maximum de contrôle. Appuyez sur n’importe quelle autre touche pour sortir. La même graduation sera utilisée pour les niveaux d’entrée et de sortie.
REMARQUE : La taille des conduits doit être sélectionnée avant d’ajuster la tension ou la graduation de courant. Pour fournir une hystérésis suffisante entre la vitesse minimale et l’arrêt, un décalage de 0.2 mA et 0.1V est ajouté au courant et à la tension d’entrée/sortie comme indiqué dans la table suivante. Ce décalage est ajouté uniquement si la graduation d’entrée/sortie est établie avec une vitesse minimale de zéro (défaut)
SpeVitesse (tr/min) 4-20 mA E/S 0-20 mA E/S 0-10 V E/S 0 4-4.2 mA 0-0.2 mA 0-0.1 V
1 4.2263 mA 0.233 mA 0.1165 V 600 20 mA 20 mA 10V
07575-00
Figure 2
REMARQUE: Les couleurs sont celles du câble de commande à distance 77300-32.
14
6
4
8
3
5
1
7
9
2
10
12
11
13
15
16
17
18
GRIS
BRUN
ROSE
ROUGE/VERT
ROUGE/NOIR
SORTIE COMPTE-TOURS
MOTEUR EN ROUTE CONTACT OUVERT AU REPOS
MOTEUR EN ROUTE CONTACT FERME AU REPOS
ROUGE/JAUNE
MARCHE/ARRÊT*AMORÇAGE
JAUNE
* SENS DES AIGUILLES D’UNE MONTRE/ SENS INVERSE DES AIGUILLES D’UNE MONTRE
ENTREE
0-20mA; 4-20mA
ENTREE 0-10V
SORTIE
0-20mA; 4-20mA
SORTIE
0-10V
ORANGE
MARRON
BLEU
VERT
BLANC
NOIR
VIOLET
ROUGE
Loading...
+ 42 hidden pages