7 Trouble shooting
Problem Cause Corrective action
The tool does not work The tool has not been connected to the air
connection
The air supply closing valve is still closed
There is insufcient air pressure
Connect the tool to the air connection
Open the air supply closing valve
Use the correct air pressure 5-7 bar
Air is coming out of the
safety valve
The air pressure is too high Use the correct air pressure 5-7 bar
There is no or insufcient
extraction
The extraction has not been turned on
There is insufcient air pressure
The mandrel collector is full
The tool is blocked by rest mandrels
Check the position of the mandrel
collector
Use the correct air pressure 5-7 bar
Empty the mandrel collector
Remove the rest mandrel
The trigger does not work There is insufcient air pressure Use the correct air pressure 5-7 bar
The blind rivet cannot
be placed into the nose
piece
The incorrect nose piece has been mounted
The tool is blocked by rest mandrels
Mount the correct nose piece
Remove the rest mandrel
The blind rivet is not set
correctly
Contaminated or worn clamping jaws
There is insufcient air pressure
The capacity of the tool has been exceeded
Clear or replace the clamping jaws
Use the correct air pressure
Use the correct tool
The rest mandrel does
not release from the nose
piece
The incorrect nose piece has been mounted
The tool is blocked by rest mandrels
Mount the correct nose piece
Remove the rest mandrel
During setting the rivet
mandrel does not break
There is insufcient air pressure
The capacity of the tool has been exceeded
Use the correct air pressure
Use the correct tool
The rest mandrel is
not extracted into the
mandrel collector
The incorrect nose piece has been mounted
The tool is blocked by rest mandrels
The mandrel collector is full
Mount the correct nose piece
Remove the rest mandrel
Empty the mandrel collector
The air supply unit cannot
be turned 360°
The tool is still under air pressure Close the air supply closing valve and
depressurize the tool by turning on the
extraction or by operating the trigger
The tool does not perform
well consistently
Contact a service centre