POCKET
DIGITAL MULTIMETER
INSTRUCTION
MANUAL
model no. 052-0060-2
Read and understand this instruction manual thoroughly
before using the product. It contains important
information for your safety as well as operating and
maintenance advice.
Keep this instruction manual for future use. Should this
product be passed on to a third party, then this
instruction manual must be included.
![](/html/40/4068/406861c641d7ca5322b90d4098e1660f1665e2a5cfd71a4fb7587776de56b3d8/bg3.png)
SAFETY INFORMATION
This meter has been designed according to IEC 61010 concerning
electronic measuring instruments with a measurement category
(CAT III 300V) and pollution degree 2.
2
WARNING
Do not use the meter if it is damaged. Before you use the meter,
inspect the case. Pay particular attention to the insulation
surrounding the connectors.
Inspect the test leads for damaged insulation or exposed metal.
Check the test leads for continuity. Replace damaged test leads
before you use the meter.
Do not use the meter if it operates abnormally.
Protection may be impaired. When in doubt, have the meter
serviced.
Do not operate the meter where explosive gas, vapour, or dust
is present.
Do not apply more than the rated voltage, as marked on the
meter, between terminals or between any terminal and earth
ground.
Before use, verify the meter's operation by measuring a known
voltage.
When measuring current, turn off circuit power before connecting the meter in the circuit. Remember to place the meter in
series with the circuit.
To avoid possible electric shock or personal injury, follow these
guidelines:
●
●
●
●
●
●
●
WARNING
![](/html/40/4068/406861c641d7ca5322b90d4098e1660f1665e2a5cfd71a4fb7587776de56b3d8/bg4.png)
model no. 052-0060-2 | contact us 1-800-689-9928
3
WARNING
When servicing the meter, use only specified replacement
parts.
Use caution when working with voltage above 30 V RMS, 42 V
peak, or 60 V DC. Such voltages pose a shock hazard.
When using the probes, keep your fingers behind the finger
guards on the probes.
Connect the common test lead before you connect the live test
lead. When you disconnect test leads, disconnect the live test
lead first.
Remove the test leads from the meter before you open the
battery cover or the case.
Do not operate the meter with the battery cover or portions of
the case removed or loosened.
To avoid false readings, which could lead to possible electric
shock or personal injury, replace the battery as soon as the
battery indicator ( ) appears.
To avoid electrical shock, do not touch any conductor with hand
or skin.
Do not operate this instrument if your hand, a test lead or the
instrument is wet.
Remaining endangerment:
When an input terminal is connected to dangerous live potential
it is to be noted that this potential can occur at all other terminals!
CAT III -
Measurement Category III is for measurements performed in
building installation. Examples are measurements on distribution
boards, circuit breakers, wiring, including cables, bus-bars, junction
boxes, switches, socket-outlets in the fixed installation, and equipment
for industrial use and some other equipment, for example, stationary
motors with permanent connection to the fixed installation.
Do not use the meter for measurements within Measurement Categories
IV.
●
●
●
●
●
●
●
●
●
●
●
![](/html/40/4068/406861c641d7ca5322b90d4098e1660f1665e2a5cfd71a4fb7587776de56b3d8/bg5.png)
4
CAUTION
To avoid possible damage to the meter or to the equipment under
test, follow these guidelines:
●
●
●
●
Disconnect circuit power and discharge all capacitors before
testing resistance, diode, continuity and temperature.
Use the proper terminals, function and range for your measurements.
Before measuring current, check the meter's fuses and turn off
the power to the circuit before connecting the meter to the
circuit.
Before rotating the range switch to change functions, disconnect
test leads from the circuit under test.
CAUTION
![](/html/40/4068/406861c641d7ca5322b90d4098e1660f1665e2a5cfd71a4fb7587776de56b3d8/bg6.png)
model no. 052-0060-2 | contact us 1-800-689-9928
5
INTRODUCTION
052-0060-2 multimeter is a compact 3 1/2-digit digital multimeter for
measuring DC and AC voltage, DC current, resistance, continuity,
diode, and battery.
It is easy to operate and is an ideal measurement tool.
Alternating current
Direct current
Both direct and alternating current
Caution, risk of electric shock.
Earth (ground) terminal
Fuse
Conforms to European Union directives
The equipment is protected throughout by double insulation or
reinforced insulation.
Caution, risk of danger. Refer to the operating manual before
use.
ELECTRICAL SYMBOLS
INTRODUCTION
![](/html/40/4068/406861c641d7ca5322b90d4098e1660f1665e2a5cfd71a4fb7587776de56b3d8/bg7.png)
FRONT PANEL
1.
2.
Display
3 1/2-digit LCD, with a max. reading of 1999.
" " button
Used to enter/exit Data Hold mode.
6
INTRODUCTION
![](/html/40/4068/406861c641d7ca5322b90d4098e1660f1665e2a5cfd71a4fb7587776de56b3d8/bg8.png)
3. Function / Range Switch
Used to select the desired function and range as well as to turn
on or off the meter.
To preserve battery life, set this switch to the "OFF" position
when not using the meter.
6. "COM" Terminal
Plug-in connector for the black test lead.
4. "10A" Terminal
Plug-in connector for the red test lead for current (200 mA – 10 A)
measurements.
5. "VΩmA" Terminal
Plug-in connector for the red test lead for all measurements
except current measurements 200 mA.
7
INTRODUCTION
model no. 052-0060-2 | contact us 1-800-689-9928
![](/html/40/4068/406861c641d7ca5322b90d4098e1660f1665e2a5cfd71a4fb7587776de56b3d8/bg9.png)
GENERAL SPECIFICATIONS
Display: 3 1/2-digit LCD, with a max. reading of 1999
Negative Polarity Indication: Negative sign " - " shown
on the display automatically
Sampling Rate: About 2 to 3 times/sec
Battery: 9 V battery, 6F22 or equivalent, 1 piece
Low Battery Indication: " " shown on the display
IP degree: IP20
Operating Environment: Temperature: 32 to 122ºF (0 to 50ºC)
Relative Humidity: < 75% RH
Storage Environment: Temperature: 14 to 140ºF (-10 to 60ºC)
Relative Humidity: < 85% RH
Dimensions: 5 7/16 x 2 15/16 x 1 1/4" (138 x 74 x 32 mm)
Weight: About 5 3/4 oz (163 g) (including battery)
SPECIFICATIONS
Readings are accurate for a period of one year after calibration at 64
to 82ºF (18 to 28ºC), with relative humidity < 75%.
Accuracy specifications take the form of:
± ([% of Reading]+[number of Least Significant Digits])
8
TECHNICAL SPECIFICATIONS
![](/html/40/4068/406861c641d7ca5322b90d4098e1660f1665e2a5cfd71a4fb7587776de56b3d8/bga.png)
If the voltage being measured is > 300 V, the display may show
the value of the voltage, but the measurement is dangerous.
1
DC VOLTAGE
Input Impedance: 1 MΩ
Max. Allowable Input Voltage: 300 V
RANGE
OVERRANGE
INDICATION
"1" shown on
display
±(0.8% + 5)
±(0.5% + 5)
±(1.0% + 5)
±(1.2% + 5)
RESOLUTION ACCURACY
AC VOLTAGE
Frequency Range: 40–400 Hz
Max. Allowable Input Voltage: 300 V
Response: Average, calibrated in RMS of sine wave
RANGE
OVERRANGE
INDICATION
"1" shown on
display
200 V
100 mV
1 V
300 V
RESOLUTION ACCURACY
If the voltage being measured is > 300 V, the display may show
the value of the voltage, but the measurement is dangerous.
1
200 mV
2000 mV
20 V
200 V
300 V
100 μV
1 mV
10 mV
100 mV
1 V
model no. 052-0060-2 | contact us 1-800-689-9928
9
TECHNICAL SPECIFICATIONS
![](/html/40/4068/406861c641d7ca5322b90d4098e1660f1665e2a5cfd71a4fb7587776de56b3d8/bgb.png)
Fuse, 250 mA/300 V, Fast action
Fuse, 10 A/300 V, Fast action
(For measurements > 2 A: measurement duration < 10
seconds, and interval > 15 minutes)
If the current being measured is > 10 A, the display may
show the value of the current, but the measurement is
dangerous.
DC CURRENT
Overload Protection:
Max. Open Circuit Voltage: about 2.8 V
Max. Allowable Input Current: 10 A
RANGE
OVERRANGE
INDICATION
"1" shown on
display
2000 μA
10 mA
±(1.0% + 5)
±(1.2% + 5)
±(2.0% + 5)
10 A
200 mA
20 mA
RESOLUTION ACCURACY
1 μA
10 μA
100 μA
F1:
F2:
1
RESISTANCE
RANGE
OVERRANGE
INDICATION
"1" shown on
display
±(1.0% + 5)
±(1.2% + 5)
±(1.2% + 5)
RESOLUTION ACCURACY
200 Ω
2000 Ω
20 kΩ
200 kΩ
2000 kΩ
0.1 Ω
1 Ω
10 Ω
100 Ω
1 kΩ
10
TECHNICAL SPECIFICATIONS
![](/html/40/4068/406861c641d7ca5322b90d4098e1660f1665e2a5cfd71a4fb7587776de56b3d8/bgc.png)
The working voltage of the battery will
be displayed on the LCD so that the
quality of the battery can be judged.
BATTERY
RANGE DESCRIPTION TEST CURRENT
1.5 V
9 V
12 V
about 20 mA
about 5 mA
about 4 mA
DIODE AND CONTINUITY TEST
RANGE DESCRIPTION TEST CONDITION
The approx. forward
voltage drop of the
diode is shown on the
display.
Open Circuit Voltage:
about 2.8 V
Test Current: about
1 mA
The built-in buzzer will
sound if the resistance
is less than about 30 Ω.
The buzzer may or may
not sound if the
resistance is between
30 Ω and 150 Ω.
The buzzer will not
sound if the resistance
is more than 150 Ω.
Open Circuit Voltage:
about 2.8 V
model no. 052-0060-2 | contact us 1-800-689-9928
11
TECHNICAL SPECIFICATIONS