Mastech MS3300, MS3302 Instruction Manual

OPERATOR'S INSTRUCTION MANUAL
01 02
AC / DC CURRENT TRANSDUCER
1. Sa fety In forma tion
The AC /DC cur rent tr ansdu cer has b een des igned acc ordin g to IEC- 1010 an d IEC10 10-2- 032 con cerni ng saf ety req uirem ents fo r elect rical m easur ing ins trume nts and h and-h eld cur rent cl amps wi th an ove r volta ge cat egory ( CAT II) and p ollut ion 2.
Fol low all i ntrod uctio ns and op erati ng requ ireme nts to ins ure tha t AC/ DC curr ent tra nsduc er is use d safel y and is ke pt in goo d opera ting co nditi on.
1.1 S ymbol s
Important safety information, refer to the operating manual.
Double insulation ( Protection classII)
2. De scrip tion
Ref er to the F igure 1 . And t o the fol lowin g numbe red ste ps to fa milia rize yo ursel f with th e AC/ DC curr ent tra nsduc er
Transformer Jaws Designed to pick up the ac. or dc current following though the conductor.
Hand Guard
Designed to protect user for safety.
Trigger Press the lever to open the transformer jaws.
DCA ZERO Adjusting the DCA ZERO knob, when the
display is not zero reading for doesn't measurement before.
Power On indication
Function Switch
Select the ac or dc current.
Output Plugs
1
2
3
4
5
6
7
03
AC / DC CURRENT TRANSDUCER AC / DC CURRENT TRANSDUCER AC / DC CURRENT TRANSDUCER
HYS 00660 2
0706
Using this appliance in an environment with a strong radiated radio-frequency electromagnetic field (approximately 3V/m) may influence its measuring accuracy. The measuring result can be strongly deviating from the actual value.
5. DC C urren t Measu remen t
5.1
5.2
5.3
Set the AC/DC current transducer function switch at desired A position.
Set a DMM at DCV function and a relevant range.
Connect the output plug of AC/DC current transducer with a DMM and combined firmly.
5.5
5.6
Adjusting the DCA ZERO knob until the DMM display show “0”, when does not measurement before.
Press the trigger to open transformer jaws and clamp one conductor only, making sure that the jaws are firmly closed around the conductor.
Read the reading on the LCD display of DMM along with the polarity of the red lead connection.
5.4
6. AC C urr ent Mea surem ent
6.1
6.2
6.3
Set the AC/DC current transducer function switch at desired A~ position.
Set a DMM at DCV function and a relevant range.
Connect the output plug of AC/DC current transducer with a DMM and combined firmly.
Press the trigger to open transformer jaws and clamp one conductor only, making sure that the jaws are firmly closed around the conductor.
6.4
6.5
Read the reading from the digital display of the DMM.
7. Ma inten ance
To keep t he inst rumen t clean , wipe th e case wi th a damp clo th and de terge nt, do no t use abr asive s or solv ents. Any a djust ment, m ainte nance a nd repa ir shal l be con ducte d by serv ice per sonne l.
Acc essor ies
Ope rator 's inst ructi on manu al
Gif t box
9 vol t batte ry. NEDA 16 04 6F22 0 06P type
0504
3. Sp ecifi catio ns
Acc uracy i s speci fied fo r a perio d of one ye ar afte r cal ibrat ion and a t 18°C to 2 8°C(6 4F to 82F ) with re lativ e hum idity t o 80. Acc uracy s pecif icati ons are g iven as : ±of r eadin g digit s
This instrument must not be used on uninsulated conductors at a voltage greater than 250V ac rms. or 250V dc
3.1 G enera l
DC CU RRENT R ANGE: 1 A dc to 1 000A dc. AC CU RRENT R ANGE: 1 A ac to 1 000A ac. R MS OUT PUT VOLTAG E: 1mV dc . per 1 A dc or 1A ac. WOR KING VO LTAGE : 100 0V CAT.II Pe r IEC 101 0 - 1 MAX IMUM ALTI TUDE: 2 000m OPE RATING TE MPERAT URE: 0°C t o 40°<C 8 0 relat ive hum idity STO RAGE TEM PERATU RE:-1 0°C to 50 °C TEM PERATU RE COEF FICIE NT:
0.2 ×(Spe c Acc 'y) /°C < 18°C or >28°C MAX IMUM JAW O PENIN G: 55mm SIZ E: 256× 104×4 7 WEI GHT: Appro x. 520g
3.2 E lectr ical sp ecifi catio n dc curr ent
Range
Measure
Output
Accuracy
1000mV1000A
± ±3.0% 5
A
Ove rload P rotec tion: 120 0A for 60 se conds m aximu m.
DC Cu rrent
AC Cu rrent
Range
Measure
Output
Accuracy
1000mV1000A
± ±3.0% 5
A
Ove rload P rotec tion: 120 0A for 60 se conds m aximu m. Fre quenc y range : 50Hz to 6 0Hz.
4. Precaution and Preparations for Measurement
4.1
4.2
4.3
Dot not applies the voltage to the output plugs of AC/DC current transducer.