Mastech MS2208 Instruction Manual

Page 1
I
Contents
Safety Instructions .................................................. 2
Safety Sign ............................................................... 3
General Description................................................. 3
Features .................................................................... 4
Knob Switch Operations ......................................... 8
Button Switch Operations .................................... 10
LCD DISPLAY ......................................................... 12
Hz ............................................................................ 13
Instruction Manual ................................................ 15
AC voltage (V) measurement .............................. 15
AC CURRENT measurement ................................ 17
Test of single-phase circuit .................................. 19
cosΦ, sinΦ, and phase angle measurement ..... 22
Single-phase three-line circuit ............................. 22
Input Voltage and Current ................................... 23
During power measurement, if input voltage is over
600 V (RMS) or current over 200 A (RMS), “OL”
symbol will be displayed and bar symbol shown in
full scale. ................................................................. 23
Page 2
II
Backlight ................................................................. 23
Auto Power Off ...................................................... 24
Power curve diagram ............................................ 24
Battery-low Indication .......................................... 26
General specification ............................................ 27
Technical specification .......................................... 29
Accessories ............................................................. 32
Page 3
1
Please carefully read the instruction manual before
using the tester, and pay special attention to
“Warning” content. Please follow instructions under
“Warning”.
1. Please be very careful when test voltage is
higher than AC 30 V, and do keep in mind that
your finger shall not exceed the hand-shielding
part of the test probe.
2. Do not measure voltage which is higher than the
allowed input limit.
3. Before use, please check the meter and test
probe; do not carry out testing in case the test
probe is naked, tester housing is damaged, or
there is no LCD display, etc..
4. It meets requirements of safety standards only
when the meter is used together with the
supplied test probes. In case the test probe is
damaged and needs replacement, it is required to
replace it with a test probe of the same model
Page 4
2
and identical electrical specifications.
5. Please never carry out any voltage measurement
whenever the test probe is inserted in any
current outlet.
6. Please do not expose the meter to strong light,
high temperature, or dampness.
Before use, please carefully read this
instruction manual. Especially safety
contents!
Safety Instructions
The digital power meter is designed and
manufactured in accordance with international
standard, IEC61010-1, and international safety
specification, IEC1010-2-032, and the meter strictly
follows the safety standard of double-insulation AC
600 V CAT III.
Page 5
3
Safety Sign
Important safety signs; please refer to instruction manual
High voltage hazard
Earthing
Double insulation (Category-II safety equipment)
Battery low Indicator
General Description
The digital power meter is a hand-held intelligent
harmonic power tester, with both functions of digital
current testing and power testing. The tester is
comprised of three channels including voltage,
current, and power as well as a micro single chip
system, and it is equipped with a powerful software
for measurement and data processing functions; it can
Page 6
4
measure, calculate, and display voltage, current,
active power, power factor, apparent power, passive
power, frequency, harmonic parameters, with stable
performance and operation convenience. The meter is
especially suitable for the measurement and
examination of on-site power equipment and
power-supplying circuits; with hand-held clamp
structure, small volume, and light weight, it can be
easily carried by the user, which makes it easy and
fast for doing measurement. For measurement of
sing-phase power, The meter is your ideal choice.
Features
1. The meter can be used for testing power, voltage,
current, peak value, phase, frequency, power
factor, phase angle, and reaction factor, etc. of
single-/three-phase circuit; automatic phase
sequence testing is possible for 3-phase
measurement.
Page 7
5
2. True effective value measurement: accurate
measurement is possible even with serious
distortion in current waveform.
3. Low-consumption high-speed single-chip
microprocessor is employed and sophisticated
algorithm is applied, as a result, results can be
obtained rapidly and precisely, and up to 20
harmonics and distortion value thereof can be
measured.
Hand-held, clamp-type structure, with light weight,
convenient for carry-on.
Page 8
6
Appearance
1
2
3
4
5
7
8
m
6
Page 9
7
1. Current clamp size:Ф26mm
2. Function-switching knob :Rotation knob for
selecting different measuring function
3. Function-selection button: Button for operating
the measuring functions
4. Input terminal
T
erminal
Function
FunctionFunction
Function
V
Input terminal for measuring the first phase; use yellow test probe for connection
COM Input terminal for measuring the 2nd
phase; use black test probe for
connection. Common terminal: ground
input terminal (earthing) for all
measuring functions; use black test probe
for connection.
5. LCD display :4-digit digital display; 7-section
Page 10
8
LCD for displaying measurement operation
function, test result, and unit sign.
6. H/
键:HOLD button :DATA HOLD
button; press down HOLD button, and the last
reading will be held and displayed on the display,
and “HOLD” symbol will be shown; press
HOLD button again, and the meter will switch
back to normal measurement mode.
7. Trigger :Press down the trigger, and the clamp
will open; release it, and the clamp will close.
8 LED Lamp
In power and current mode ,press
button
open or close Lamp
Knob Switch Operations
The function-switching knob is used for powering-on and for switching to any measurement function in the following table.
Page 11
9
Knob position description
Sign
Knob position
Functions
OFF Powering-off
position
For powering-off
KW
Active power
position
For measuring ac
tive power, etc.
W
power position For measuring
active power, etc.
V~~
AC-voltage har
monics test posi
tion
For measuring A
C-voltage harmon
ics, etc.
A~~
AC-current har
monics test posi
tion
For measuring A
C-current harmon
ics, etc.
mA~~
AC-voltage har
monics test posi
tion
For measuring A
C-voltage harmon
ics, etc.
Page 12
10
Note:
When the meter is automatically powered off, be sure to switch the knob to “OFF” position; turn on the meter after 5 seconds.
Button Switch Operations
Button descriptions
SN Function-selection button
1 MODE Test-mode switching button
2 Reverse-search button
3 Forward-search button
4 MIN/MAX
MAX/MIN Value
5 H/
HOLD/LIGHT
The following functions can be realized through
button operations:
1MODE
Under KW/W test mode, you can press MODE
button to switch the display of active power and
passive power; under A/V~ test mode, you can switch
Page 13
11
the display among total harmonic distortion rate F, r,
and harmonic percentage.
2.▲:
press button to change view MAX value,MI
N value,measuring time; During testing harmonics
, you can change the times of harmonics. press button to search backward thestored dat
a and show it on LCD. With every press of the
button, the searching cursor will move one step backward to the previous data.
3.▼
press button to change view MAX value,MIN
value,measuring time; During testing harmonics,
you can change the times of harmonics. press button to search foreward thestored data
and show it on LCD. With every press of the
button, the searching cursor will move one step f oreward to the previous data.
4MAX/MIN:::
In test mode ,press MAX/MIN button to view
Page 14
12
the max /min Power,current,voltage.
5HOLD/
::::
After measurement, press this button to hold this
data on LCD; after powering-off, data will display
Press this button longer than 1S,Open or close the
back light.
LCD DISPLAY
Page 15
13
LCD symbol Descripti
on
LCD symbol
Descripti on
AutoPowe r Off
HOLD
Hold
LAG
Phase angle lag
LEAD Phase
angle
lead Battery power indication
o Phase
angle
(degree)
SINФ
Phase angle (degree)
COSФ Power
factor
AC symbol
PEAK Peak
value
Hz
Frequency
Negative
symbol
MIN
Minimum value
MAX Maximu
m value
USED
Used READ
read
W
watt A current
VAr
Passive power
VA
Apparent
power
PHASE phase
TRMS
TRMS
%
Harmonic percentag e
V Voltage
Page 16
14
LCD symbol Descripti
on
LCD symbol
Descripti
on
%THD Total harmonics distortion ratio H01F Total harmonics distortion ratio F
(relative to base wave)
H01r Total harmonics distortion ratio r
(relative to real effective value)
Page 17
15
Instruction Manual AC voltage (V) measurement
MAS TE CH
TL88-1
UL
MA ST EC H
TL88-1
UL
AC
V
m
Page 18
16
1. According to the connection mode as above
Table , switch the function switching knob to V~,
select corresponding sockets from V,COM
terminal, and insert the test wire.
2. Connect the two test probes V1, COMto the
power source or load to be tested. The meter will
automatically test and display the result, and the
present harmonics percentage will be shown on
the following line.
3. Under voltage test mode, press MAX/MIN
button to view Max or Min value
4. Press MODE button to show harmonics
percentage on LCD, and the total harmonic
distortion ratio F and R will be cyclically
displayed. Press▲/▼ button to display value
of each measurement of the harmonic.
Page 19
17
AC CURRENT measurement
3
2
1
m
N
Page 20
18
1. Switch the function knob to A~ position;
2. Pull the trigger to open the clamp, and then clip
a wire which is to be tested; the measured
current value will be automatically shown on
LCD
3. Press MODE button to show harmonics
percentage on LCD, and the total harmonic
distortion ratio F and r will be cyclically
displayed.
4. Press ▲/▼ button to display value of each
measurement of the harmonic.
5. press MAX/MIN button to view Max or Min
value
Note:
1. You can select 50/60-Hz FIXED/AUTO
frequency test (AUTO). When input waveform
fluctuates, displayed harmonic values can be
kept stable if 50/60-Hz FIXED mode is selected.
2. Under AUTO frequency test mode, FFT
Page 21
19
calculation is performed only when base-wave
frequency is between 45 and 65 Hz; harmonic
analysis is not performed when base-wave
frequency exceeds this range.
Test of single-phase circuit
1. Clip the clamp on the test wire of the power
supply or load. If the user needs to measure a
certain phase of the 3-phase circuit, then the
clamp should clip on the wire of the same phase.
2. switch the function switching knob to
KW/Wposition, select corresponding input
sockets from V1 or COM terminal and insert the
test wire
3. After it is correctly connected, you can measure
single-phase power (active power, power factor,
apparent power, passive power, voltage, current,
phase angle, peak value of voltage and current,
and frequency):
4. The meter will carry out automatic measurement
and display active power, and voltage/current
Page 22
20
value of the load being tested will be displayed
on the bottom line of LCD; press down
MODE button, Var value of passive power will
be displayed on LCD; press WATT button to
display apparent power and power factor (cos Φ);
negative power factor signifies that the load
being tested is a load with capacitor
characteristics.
5. The maximum measurement range of active
power kW is 120 kW; if this range is exceeded,
“OL” symbol will be displayed beyond this
range. If voltage being tested is greater than 600
V, or current being tested greater than 200 A,
“OL”
symbol will be displayed on LCD.
6. The min. input voltage is 10 V and the min.
input current is 10mA; if active power value is
smaller than this limit, “0.00 kW” will be
displayed in stead of active power value.
7. Press Max/Min value to view Max/Min Value
Page 23
21
8. Passive power is a value not directly measured;
equation for kVAr is kVAr
2
=kVA2-kW2; its value is
calculated by software based on the measured
voltage, current and active power, and displayed
on LCD.
Power supply side
Load
Red
Black
m
Page 24
22
cosΦ, sinΦ, and phase angle measurement
1. Switch the function-switch knob to KW/W position,
and the test wire is inserted to V1/COM input
terminals.
2. The meter will automatically measure and
display power factor, voltage value and current
value.
3. Press MODE button to display phase angle,
power factor (cos Φ), and sinΦ; negative power
factor signifies that the load being tested is a
load with capacitor characteristics.
4. Press MAX/MIN Button to view Max value,Min
value,measuring time.
Single-phase three-line circuit
the process for measuring power and power fact
or for single-phase three-line circuit is the same
as that for single-phase two-line circuit, where th
e black clip is connected to the middle wire, and
Page 25
23
the red clip and clamp-type sensor are simultan
eously connected to all test wires.
Input Voltage and Current
During power measurement, if input voltage is over
600 V (RMS) or current over 200 A (RMS), “OL”
symbol will be displayed and bar symbol shown in
full scale.
Backlight
Press button longer than 1s, the backlight
will be lit up, and it will then be automatically turned
off after about 20 seconds.
Power supply side
Red
Black
Black
Red
m
m
Page 26
24
Auto Power Off
1. If there is no function change or button press fo
10 minutes ,the meter will automatically turn
power off , When the meter is automatically
powered off, be sure to switch the knob to
“OFF” position; turn on the meter after 5
seconds.
2. Holding the button
down while turning
the meter on,Disables automatic power -off
Power curve diagram
(Active power=apparent power × PF)
(PF=1)
2000W 2000W
+E +I +P
360
O
-E
-I
-P
ACTIVE POWER
1000W
PEAK VALUE
E =141V
PEAK VALUE OF POWER
POWER
=14A
PEAK VALUE
180
O
90
O
MAXI
MAX
PF=kW/kVA
Page 27
25
+V +I +P
P
+i
MAX
I
MAX
v
+v
-i
-V
X
-E
-I
-P
360
O
180
O
0
-
0
-
+V +I +P
P
0
-
-E
-I
-P
0
-
O45O
O
45
O
45
O
O
90
O
180
O
225
O
270
O
360
O
kW=I R
2
ACTIVE POWER
500W
MAX
I
MAX
v
Page 28
26
Battery-low Indication
If battery voltage is low, “ ” symbol will be
displayed on the upper right corner of LCD. Then, it
is needed to replace new batteries.
Battery Replace
Warning
1. Before opening the back lid to replace batteries,
please make sure the meter is turned off and no
test probe is connected to any test wire so as to
avoid electrical shock; before using the meter,
please make sure the back lid is tightly closed.
Only batteries of identical model or electrical
specification can be used.
2. If “ ” symbol is shown on LCD, it signifies
that battery voltage with load is lower than the
minimum voltage for ensuring measurement
error limits, and the meter will prompt you to
change new batteries. Please follow the steps
below to replace batteries:
Page 29
27
3. Disconnect test probes from test circuit, and
rotate function-switching knob to “OFF”, and
then take off test probes from the input sockets.
4. Open battery cove, Take out old batteries, and
replace them with 3new 1.5 V batteries. New
batteries shall not be used together with old ones.
Properly close the battery cover.
General specification
Complies with IEC/EN 61010-1 1000V CAT II ,600
V CAT III
1. Max. common-mode voltage: 600V AC RMS
2. Mode of display: LCD display; Max. reading:
6000
3. Range selection: Fully automatic range selection
4. Frequency detection: automatic (when harmonic
is serious, it is better to use manual settings for
testing frequency so as to assure the reading
stability)
Page 30
28
5. Over-range display: “OL”
6. Data holding: “HOLD” is shown on LCD
7. Power supply: 3 batteries of 1.5 V, AAA
8. Power consumption: 250 mW
9. Storage temperature: - 20 °C ~ 70 °C
10. Operating temperature: 0 °C ~ 40 °C
11. Temperature Coefficient:0.05 × (specified
accuracy)per°C
12. Electromagnetic Compatibility: In an RF field of
3V/M,accuracy=specified accuracy , Otherwise
accuracy is unspecifieced.
13. Operating Altitude: 2000m CAT 600V
3000m CAT 600V
14. Store Altitude: 12000m
15. Dimensions: 208mm×78mm×35mm
16. Weight: about 350 g (with battery)
Page 31
29
Technical specification
Accuracy: ±(% read + graduation #) ambient
temperature: 18°C ~ 28°C, Humidity 80%, frequency
for voltage, current: 45 Hz ~ 65 Hz
AC Voltage RMS
Range
Accuracy Resolution Input
impedance
10V-600V
±
(1.0%+5)
0.1V 1 MΩ //
10 pF
Max. allowed overload voltage: 750 V (RMS)
AC Current RMS
Range Accuracy Resolution
10mA-3999mA
±(2%+5) 0.01mA
4A-200A ±(2%+5) 0.01A
Max. allowed overload current: 200 A
Page 32
30
Single-phase active power (((W))))
Range Accuracy Resolution
0.01W-120KW ±(3%+5) 0.01W
Min. test current: 10mA; Min. test voltage: 10V
Power factor
Range Accuracy
Resolution
0.31 Capacitive ±(0.02+2) 0.001
0.31 Inductive ±(0.02+2) 0.001
Min. test current: 10mA; Min. test voltage: 10V
Passive power (VAr)
2
=((((VA) 2+W
2
Range Accuracy Resolution
0.01W-120KW ±(3%+5) 0.01VAr
Min. test current: 10mA; Min. test voltage: 10V
Passive power Var is calculated according to the
measured V, A, and kW value
Page 33
31
Frequency (Hz)
Range Accuracy
Resolution
30Hz1kHz 0.5%+1 0.1Hz
Min. test current: 10mA; Min. test voltage: 10V
Harmonic test
Harmonic number
Precision of harmonic voltage
1
±(3.0%+10)
2-6
±(3.5%+10)
7-8
±(4.5%+10)
9-10
±(5.0%+10)
11-15
±(7%+10)
16-20
±(10%+10)
Min. test current: 20mA; Min. test voltage: 20V
Page 34
32
Accessories
Item Quantity
Instruction
Manual
×1
Battery 1.5V AA A Battery×3
Test Leads ×1
Test clip
×2Red,Black
Package box ×1
Page 35
33
Quality Assurance
Thank you for using the product of our company;
this product has a warranty period of one year
starting from purchasing date.
This product has passed the strict quality test of our
company. Our company will exert our efforts to do
after-sales service for you in accordance with the
Warranty instructions.
In case there is any problem occurred during
warranty period, which is resulted from product
quality problem, please fill in the warranty card and
mail it in together with the product, and the service
department of our company will repair it or replace it
free of charge, while the user himself/herself shall not
take the meter apart.
Page 36
34
When warranty period is over, the repairing will
be charged.
Free service does not apply to the following cases:
Problems and damages due to improper use or use
under environment which is not stipulated for this
product, including overloading.
Problems and damages due to unauthorized
dismantling or refitting and misuse.
Problems and damages due to user reasons.
Problems and damages due to natural disasters.