Masimo rainbow Operator's Manual

Page 1
rainbow® Series Patient Cables
For Pulse Oximeters with rainbow Technology
© 2016 Masimo Corporation
Images 2
en
English 3-5
Page 2
rainbow® Series Patient Cables
For Pulse Oximeters with rainbow Technology
1
2
1
3
2
1
2
3
1
2
4523G-eIFU-0516
Page 3
rainbow® Series Patient Cables
For Pulse Oximeters with rainbow Technology
DIRECTIONS FOR USE
en
Reusable
LATEX
Not made with natural rubber latex Non-Sterile
INDICATIONS
The rainbow® Series Patient Cable Series have the same indications for use as the applicable sensors. Please see the compatible sensor’s directions for use for indications for use and prescribed information.
DESCRIPTION
The rainbow Series Patient Cables are used with Masimo® rainbow and M-LNCS™ Sensors.rainbow Series Patient Cables are for use only with devices containing Masimo rainbow SET™ technology Version 7.1 or higher or licensed to use rainbow compatible sensors. Consult individual oximetry system manufacturers for compatibility of particular sensor and device models. Each device manufacturer is responsible for determining whether their devices are compatible with each sensor model.
rainbow Series Patient Cables were veried using Masimo rainbow SET technology. WARNING: Masimo sensors and cables are designed for use with devices containing Masimo SET® or Masimo rainbow
SET oximetry or licensed to use Masimo sensors.
WARNINGS, CAUTIONS, AND NOTES
• Always refer to the oximeter module operator’s manual for complete instructions or additional instructions.
• Ensure the cable is physically intact, with no broken or frayed wires or damaged parts. Visually inspect the cable and discard if cracks or discoloration are found.
• All sensors and cables are designed for use with specific monitors. Verify the compatibility of the monitor, cable and sensor before use, otherwise degraded performance and/or patient injury can result.
• Carefully route patient cable to reduce the possibility of patient entanglement or strangulation.
• Failure to properly connect the sensor or the oximeter module to the cable will result in intermittent readings, inaccurate results, or no reading.
• To avoid damage to the cable, always hold it by the connector rather than the cable when connecting or disconnecting either end.
• To prevent damage, do not soak or immerse the sensor or cable in any liquid solution. Do not attempt to sterilize the cable or sensor.
• Do not attempt to reprocess, recondition or recycle Masimo sensors or patient cables as these processes may damage the electrical components, potentially leading to patient harm.
• Caution: Replace the patient cable when a Replace cable or a persistent low SIQ message is displayed when the patient monitoring time is exhausted.
• Note: The patient cable is provided with X-Cal™ technology to minimize the risk of inaccurate readings and unanticipated loss of patient monitoring. The cable will provide 17,520 hours of patient monitoring time. Replace the cable when the patient monitoring time is exhausted.
CABLE AND SENSOR SELECTION
NOTE: The maximum length for the sensor, cable and extension cable combination is 15 feet. Refer to the table below for correct examples:
Cable Sensor
Length 18 inches 3 feet 6 feet
1 ft Yes Yes Yes
4 ft Yes Yes No
12 ft No No No
3
4523G-eIFU-0516
Page 4
INSTRUCTIONS
A. Connect the rainbow patient cable to the Device
1. Connect the device connector of the patient cable to the mating connector of the device.
2. Ensure the patient cable connector is fully engaged with the device. A tactile or audible click of connection should be heard.
B. Connect the rainbow patient cable to the Sensors
Connect the rainbow patient cable to an M-LNCS disposable sensor:
1. Refer to Fig. 1. Orient the sensor connector to the patient cable connector as shown.
2. Refer to Fig. 2. Insert the sensor connector completely into the patient cable connector until it locks in place as shown.
Connect the rainbow patient cable to a rainbow disposable sensor:
1. Refer to Fig. 3. Orient the sensor connector (1) to patient cable connector (2) as shown.
2. Refer to Fig. 4. Insert the sensor connector (1) completely into the patient cable connector (2) until it locks in place as shown.
Connect the rainbow patient cable to a reusable M-LNCS or a reusable rainbow sensor:
1. Refer to Fig. 5. Orient sensor connector (1) to patient cable connector (2) as shown.
2. Refer to Fig. 6. Insert the sensor connector (1) completely into the patient cable connector (2) until it locks in place..
C. Disconnecting the rainbow patient cable from the Sensors
NOTE: Pull on the connectors only, do not pull the cables.
Disconnect the M-LNCS disposable sensor:
1. Refer to Fig. 2. Pull firmly on the sensor connector to remove it from the patient cable.
Disconnect the rainbow disposable sensor:
1. Refer to Fig. 4. While holding the sides of cable connector, place thumb on latch (1) and press to disconnect sensor connector from patient cable connector.
Disconnect the reusable M-LNCS or reusable rainbow sensor:
2. Refer to Fig. 5. While holding the sides of patient cable connector (2), lift sensor connector cover (3) from patient cable connector (2). Disconnect sensor connector (1) from patient cable connector (2).
CLEANING
1. Remove the sensor from the patient and disconnect it from the patient cable.
2. Clean the surface of the cable by wiping it with a 70% isopropyl alcohol pad.
3. Wipe all surfaces of the cable.
4. Saturate another cloth or gauze pad with sterile or distilled water and wipe all surfaces of the cable.
5. Dry the cable by wiping all surfaces with a clean cloth or dry gauze pad.
CAUTION:
• Do not immerse the connectors on the cable in any liquid solution.
• Do not sterilize by irradiation, steam, autoclave, or ethylene oxide.
• Do not clean with chemicals not approved above.
ENVIRONMENTALS
Operating Temperature 41°F to 104°F (5°C to 40°C)
Storage Temperature -40°F to 158°F (-40°C to 70°C)
Relative Humidity 0% to 95% non-condensing
COMPATIBILITY
rainbow series patient cables are intended for use only with devices containing Masimo SET or Masimo rainbow SET technology and with compatible sensors. Consult individual oximetry system manufacturers for compatibility of particular devices and sensor models. Each device manufacturer is responsible for determining whether their devices are compatible with each sensor and cable model. Use of this cable with other devices may result in no or
improper performance. For Compatibility Information Reference: www.Masimo.com
WARRANTY
Masimo warrants to the initial buyer only that this product, when used in accordance with the directions provided with the Products by Masimo, will be free of defects in materials and workmanship for a period of six (6) months. Single use products are warranted for single patient use only.
THE FOREGOING IS THE SOLE AND EXCLUSIVE WARRANTY APPLICABLE TO THE PRODUCTS SOLD BY MASIMO TO BUYER. MASIMO EXPRESSLY DISCLAIMS ALL OTHER ORAL, EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION ANY WARRANTIES OF MERCHANTABILITY OR FITNESS FOR PARTICULAR PURPOSE. MASIMO’S SOLE OBLIGATION AND BUYER’S EXCLUSIVE REMEDY FOR BREACH OF ANY WARRANTY SHALL BE, AT MASIMO’S OPTION, TO REPAIR OR REPLACE THE PRODUCT.
4
4523G-eIFU-0516
Page 5
WARRANTY EXCLUSIONS
-40 C
+70 C
5%-95% RH
This warranty does not extend to any product that has been used in violation of the operating instructions supplied with the product, or has been subject to misuse, neglect, accident or externally created damage. This warranty does not extend to any product that has been connected to any unintended device or system, has been modied, or has been disassembled or reassembled. This warranty does not extend to sensors or patient cables that have been reprocessed, reconditioned or recycled.
IN NO EVENT SHALL MASIMO BE LIABLE TO BUYER OR ANY OTHER PERSON FOR ANY INCIDENTAL, INDIRECT, SPECIAL OR CONSEQUENTIAL DAMAGES (INCLUDING WITHOUT LIMITATION LOST PROFITS), EVEN IF ADVISED OF THE POSSIBILITY THEREOF. IN NO EVENT SHALL MASIMO’S LIABILITY ARISING FROM ANY PRODUCTS SOLD TO BUYER UNDER A CONTRACT, WARRANTY, TORT OR OTHER CLAIM EXCEED THE AMOUNT PAID BY BUYER FOR THE LOT OF PRODUCTS INVOLVED IN SUCH CLAIM. IN NO EVENT SHALL MASIMO BE LIABLE FOR ANY DAMAGES ASSOCIATED WITH A PRODUCT THAT HAS BEEN REPROCESSED, RECONDITIONED OR RECYCLED. THE LIMITATIONS IN THIS SECTION SHALL NOT BE DEEMED TO PRECLUDE ANY LIABILITY THAT, UNDER APPLICABLE PRODUCTS LIABILITY LAW, CANNOT LEGALLY BE PRECLUDED BY CONTRACT.
NO IMPLIED LICENSE
Purchase or possession of this rainbow patient cable does not carry any express or implied license to use this cable with any device that is not an authorized device or separately authorized to use rainbow patient cables.
CAUTION:
For professional use. See instructions for use for full prescribing information, including indications, contraindications, warnings, precautions and adverse events.
The following symbols may appear on the product or product labeling:
FEDERAL LAW (U.S.A.) RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN.
SYMBOL DEFINITION SYMBOL DEFINITION
i
Consult Instructions for Use
Separate collection for electrical and electronic equipment (WEEE).
(blue background)
LATEX
Rx Only
EC REP
Patents: http://www.masimo.com/patents.htm
Masimo, SET, M-LNCS, rainbow SET and X-Cal are trademarks of Masimo Corporation.
Follow Instructions for Use
Manufacturer
Date of Manufacture Masimo reference number
Use By YYY Y-MM-DD Body weight
Do not discard
Non-Sterile
Not made with natural rubber latex Storage Humidity Range
Federal law (USA) restricts this device to sale by or on the order of a physician
Mark of Conformity to European Medical Device Directive 93/42/EEC
Authorized representative in the European community
Do not use if package is damaged
, and rainbow are federally registered trademarks of Masimo Corporation.
LOT
REF
>
<
+1060 hPa - +500 hPa
795 mmHg - 375 mmHg
Lot code
Catalogue number (model number)
Greater than
Less than
Storage Temperature Range/ Storage Altitude Range
Keep Dry
Atmospheric Pressure Range
Instructions/Directions for Use/Manuals are available in electronic format @ http://www.Masimo.com/TechDocs Note: eIFU is not available for CE mark countries.
5
4523G-eIFU-0516
Page 6
© 2016 Masimo Corporation
Manufacturer:
Masimo Corporation 40 Parker Irvine, CA 92618 USA
www.masimo.com
EU Authorized Representative for Masimo Corporation:
EC REP
MDSS GmbH Schigraben 41 D-30175 Hannover, Germany
4523G-eIFU-0516