Masimo Radical Signal Extraction Operator's Manual

signal extraction pulse oximeter
OPERATOR’S MANUAL
Signal Extraction Pulse Oximeter
OPERATOR’S MANUAL
Radical Signal Extraction Pulse Oximeter Operator’s Manual i
The Radical Operating Instructions provide the necessary information for proper operation of all models of the Radical pulse oximetry system.There may be information provided in this manual that is not relevant for your system.
General knowledge of pulse oximetry and an understanding of the features and functions of the Radical Pulse Oximeter are a prerequisite for its proper use.
Do not operate the Radical Pulse Oximeter without completely reading and understanding the instructions in this manual.
NOTICE
Purchase or possession of this device does not carry any express or implied license to use this device with replacement parts which would, alone or in combination with this device, fall within the scope of one of the patents relating to this device.
CAUTION:
FEDERAL LAW (U.S.) RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN.
Masimo Corporation 40 Parker Irvine, CA 92618 USA Tel.: 949-297-7000 Fax.: 949-297-7001 www.masimo.com
EU Authorized Representative for Masimo Corporation:
MDSS GmbH Schiffgraben 41 30175 Hannover, Germany Tel.: +49-511-62 62 86 30 Fax.: +49-511-62 62 86 33
MEDICAL ELECTRICAL EQUIPMENT WITH RESPECT TO ELECTRIC SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60601-1/CAN/CSA C22.2 No. 601.1
Covered by one or more of the following U.S. Patents: RE38,492, RE38,476, 6,850,787, 6,826,419, 6,816,741,
6,699,194, 6,684,090, 6,658,276, 6,654,624, 6,650,917, 6,643,530, 6,606,511, 6,584,336, 6,501,975, 6,463,311,
6,430,525, 6,360,114, 6,263,222, 6,236,872, 6,229,856, 6,206,830, 6,157,850, 6,067,462, 6,011,986, 6,002,952,
5,919,134, 5,823,950, 5,769,785, 5,758,644, 5,685,299, 5,632,272, 5,490,505, 5,482,036, international
equivalents, or one or more of the patents referenced at www.masimo.com/patents. Products containing Satshare®
feature are also covered by U.S. Patent 6,770,028. Other patents pending.
© 2007 Masimo Corporation. Masimo, Radical, Discrete Saturation Transform, DST, Satshare,SET, LNOP, LNCS
and LNOPv are federally registered trademarks of Masimo Corporation.
RadNet, RadicalScreen, Signal IQ, FastSat, FastStart and APOD are trademarks of Masimo Corporation.
EC REP
SAFETY INFORMATION,WARNINGS, CAUTIONS AND NOTES
The Radical Signal Extraction Pulse Oximeter is designed to minimize the possibility of hazards from errors in the software program by following sound engineering design processes, Risk Analysis and Software Validation.
Explosion hazard. Do not use the Pulse Oximeter in the presence of flammable anesthetics or other flammable substance in combination with air, oxygen-enriched environments, or nitrous oxide.
High intensity extreme lights (such as pulsating strobe lights) directed on the sensor, may not allow the Pulse Oximeter to obtain vital sign readings.
The Pulse Oximeter is NOT intended for use as an apnea monitor.
A Pulse Oximeter should be considered an early warning device. As a trend towards patient hypoxemia is indicated, blood samples should be analyzed by laboratory instruments to completely understand the patient’s condition.
The Pulse Oximeter is to be operated by qualified personnel only. This manual, accessory directions for use, all precautionary information, and specifications should be read before use.
Electric shock hazard. Do not open the pulse oximeter cover except to replace the battery of the Handheld unit. Only a qualified operator may perform maintenance procedures specifically described in this manual. Refer servicing to Masimo for repair of this equipment.
As with all medical equipment, carefully route patient cabling to reduce the possibility of patient entanglement or strangulation.
Do not place the pulse oximeter or accessories in any position that might cause it to fall on the patient. Do not lift the pulse oximeter by the power cord or any other cable.
Interfering Substances: SpO2is a functional calculation of arterial oxygen saturation. Carboxyhemoglobin and Methemoglobin may erroneously increase SpO2readings. The level of increase is approximately equal to the amount of carboxyhemoglobin and/or methemoglobin that is present. Dyes, or any substance containing dyes, that change usual blood pigmentation may cause erroneous readings.
Severe anemia may cause erroneous SpO2readings.
Do not use the pulse oximeter or oximetry sensors during magnetic resonance imaging (MRI) scanning. Induced current could potentially cause burns.The pulse oximeter may affect the MRI image, and the MRI unit may affect the accuracy of the oximetry measurements.
If using pulse oximetry during full body irradiation, keep the sensor out of the irradiation field. If the sensor is exposed to the irradiation, the reading might be inaccurate or the unit might read zero for the duration of the active irradiation period.
For home use, ensure that the pulse oximeter’s alarm can be heard from other rooms in the house especially when noisy appliances such as vacuum cleaners, dishwashers, clothes dryers, televisions, or radios are being operating.
Always remove the sensor from the patient and completely disconnect the patient from the pulse oximeter before bathing the patient.
Do not place the pulse oximeter where the controls can be changed by the patient.
Do not place the pulse oximeter face against a surface. This will cause the alarm to be muffled.
Do not place the pulse oximeter on electrical equipment that may affect the pulse oximeter, preventing it from working properly.
Do not expose the pulse oximeter to excessive moisture such as direct exposure to rain.Excessive moisture can cause the pulse oximeter to perform inaccurately or fail.
ii Radical Signal Extraction Pulse Oximeter Operator’s Manual Radical Signal Extraction Pulse Oximeter Operator’s Manual iii
Do not place containers containing liquids on or near the pulse oximeter. Liquids spilled on the pulse oximeter may cause it to perform inaccurately or fail.
Failure of Operation - If the pulse oximeter fails any part of the setup procedures or leakage tests, remove the pulse oximeter from operation until qualified service personnel have corrected the situation.
Patient Safety - If a sensor is damaged in any way, discontinue use immediately.
Disposal of product - Comply with local laws in the disposal of the unit and/or its accessories.
The Pulse Oximeter Module can be used during defibrillation, but the readings may be inaccurate for up to 20 seconds.
This equipment has been tested and found to comply with the limits for medical devices to the EN 60601-1-2: 2002, Medical Device Directive 93/42/EEC. These limits are designed to provide reasonable protection against harmful interference in a typical medical installation.This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to other devices in the vicinity. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to other devices, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
Reorient or relocate the receiving device.
Increase the separation between the equipment.
Connect the equipment into an outlet on a circuit different from that to which the other device(s) are connected.
Consult the manufacturer for help.
A functional tester cannot be utilized to assess the accuracy of the Pulse Oximeter or any sensors.
SAFETY INFORMATION,WARNINGS, CAUTIONS AND NOTES (CONTINUED)
Radical Signal Extraction Pulse Oximeter Operator’s Manual v
table of contents
First Page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-8
Second page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-8
Simplified user interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-9
First page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-9
Second Page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-9
Navigating the main menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-10
Main menu selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-10
Menu categories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-10
Editing a parameter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-11
Menu tree . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-11
Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-13
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-16
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-17
Clock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-17
About . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-17
Config . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-17
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-18
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-20
Trend display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-21
Navigating the trend display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-23
First page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-23
Second page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-23
Third page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-23
Trend setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-24
Backlight/contrast operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-25
Satshare operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-26
Home mode operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-27
Password operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-28
SECTION 5 - ALARMS/MESSAGES
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-1
Alarm identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-1
System messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-2
Alarm operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-3
Alarm messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-3
SECTION 6 - TROUBLESHOOTING
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-1
SECTION 7 - SPECIFICATIONS
Radical specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-1
Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-1
Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-1
Environmental . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-2
Physical characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-2
Serial interface specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-3
Serial Printer Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-4
Analog output / nurse call specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-5
Analog Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-6
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-6
Nurse Call . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-6
SECTION 1 - OVERVIEW
About this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
Product description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Features and benefits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Indications for use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Pulse oximetry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
General Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
Principle of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
Functional saturation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
Measured vs. Calculated Saturation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
Masimo SET signal extraction technology . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
SECTION 2 - SYSTEM DESCRIPTION
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
Handheld Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Handheld back Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
Radical standalone . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
Standalone Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
Standalone Back Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-9
Radical monitor interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-10
SECTION 3 - SETUP
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
Unpacking and inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
Preparation for monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
Radical power requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
Initial battery charging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
Initial installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
Factory Default Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Custom (user) Defined Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Adult/neo (hospital defined) settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-4
SatShare Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-5
Philips VueLink setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-6
Spacelabs universal flexport setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-7
RadNet Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-7
SECTION 4 - OPERATION
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1
Basic operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1
General Setup and Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1
Successful SpO
2
monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
Masimo Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
Numeric Display - SpO
2
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-4
Numeric Display - Pulse Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-4
Signal IQ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-5
Low Perfusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-6
Actions To Be Taken . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-6
Sensitivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-7
Touch key control button and icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-8
Traditional user interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-8
table of contents
iv Radical Signal Extraction Pulse Oximeter Operator’s Manual
SECTION 8 - SENSORS & PATIENT CABLES
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-1
Masimo SpO
2
sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-1
Selecting a Masimo Set Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-1
Sensor Application Site . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-2
LNOP
®
Reusable Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-2
LNOP®Adhesive Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-2
LNOP®specialty Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-3
LNOPvTMAdhesive Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-3
LNCS
TM
Reusable Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-3
LNCSTMAdhesive Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-4
Cleaning And Reuse Of Masimo Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-4
Reattachment of Single Use adhesive Sensors . . . . . . . . . . . . . . . . . . . . . . . . . .8-4
Masimo SET patient cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-5
Cleaning And Reuse Of Masimo SET Patient Cables . . . . . . . . . . . . . . . . . . . . . .8-5
SECTION 9 - SERVICE AND MAINTENANCE
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-1
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-1
Battery operation and maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-2
Replacing the Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-4
Replacing the Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-5
Performance verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-5
Service and repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-9
Repair Policy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-9
Return Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-9
Exclusions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-10
End-user license agreement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-10
SECTION 10 - ACCESSORIES
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-1
table of contents
vi Radical Signal Extraction Pulse Oximeter Operator’s Manual Radical Signal Extraction Pulse Oximeter Operator’s Manual 1-1
1
overview
About this Manual
This manual explains how to set up and use the Masimo SET Radical Pulse Oximeter, herein known as the Radical. Important safety information relating to general use of the Radical appears before this introduction. Other important safety information is located throughout the manual where appropriate.
Read the entire safety information section before you operate the monitor.
In addition to the safety section, this manual includes the following sections:
SECTION 1 OVERVIEW gives a general description of pulse oximetry.
SECTION 2 SYSTEM DESCRIPTION describes the Radical system and its
functions and features.
SECTION 3 SETUP describes how to setup the Radical for use.
SECTION 4 OPERATION describes the operation of the Radical Pulse
Oximetry system.
SECTION 5 ALARMS AND MESSAGES describes the alarm system messages.
SECTION 6 TROUBLESHOOTING describes troubleshooting information.
SECTION 7 SPECIFICATIONS gives the detailed specifications of the
Radical Pulse Oximeter.
SECTION 8 SENSORS AND PATIENT CABLES outlines how to use and
care for Masimo SET sensors and Masimo SET patient cables.
SECTION 9 SERVICE AND MAINTENANCE describes how to maintain,
service and obtain repair for the Radical Pulse Oximeter.
SECTION 10 ACCESSORIES lists the available Radical accessories.
Radical Signal Extraction Pulse Oximeter Operator’s Manual 1-3
1
overview
Product description
The Radical Pulse Oximeter is a noninvasive, arterial oxygen saturation and pulse rate monitor.The Radical can be used as either a Handheld or a Standalone monitor.The Radical features a backlit Liquid Crystal Display (LCD) that continuously displays numeric values for SpO
2
, pulse rate, plethysmographic waveform and Signal
Identification and Quality Indicator (Signal IQ
TM
). The Radical can be used to interface with a multiparameter patient monitor to provide Masimo SET pulse oximetry information to that monitor for display.
FEATURES AND BENEFITS
Clinically proven as the highest sensitivity and specificity pulse oximeter in the world
Signal IQ ®waveform for signal identification and quality indication during excessive motion and low signal to noise situation
FastSat ®tracks rapid changes in arterial O2with high fidelity unlike any other pulse oximeter
FastStartTMallows for rapid measurement from the time the instrument is first turned on
APODTM(Adaptive Probe Off Detection) offers the best probe off detection of Masimo's three sensitivity modes - APOD, Normal and MAX sensitivity
ClearVueTMtechnology delivers the highest resolution display available
SmartToneTMbeeps in sync with pulse, and with tonal changes for every 1% drop in saturation, even under most patient motion conditions
SatShare ®interface connection to existing multiparameter monitor
(models RDS-
1/1B only)
RS-232, analog output, and nurse call interfaces
(models RDS-1/1B and RDS-3 only)
User configurability maximizes operation for all applications
Facilitates economical sensor standardization throughout the hospital
SpO2, pulse rate, alarm, trending, perfusion index, Signal IQ and plethysmographic waveform display
Automatic screen rotation provides upright display for vertical or horizontal monitor positioning
(not available on the RDS-2 model)
Remote alarming interface
Designed for hospital, transport and home use
Up to thirty days of trending
INDICATIONS FOR USE
The Radical and accessories are indicated for the continuous, noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO
2
) and pulse rate (measured by
an SpO
2
sensor). The Radical and accessories are indicated for use with adult, pediatric
and neonatal patients during both motion and no motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile and home environments. In addition, the is indicated to provide continuous noninvasive monitoring data to validated multiparameter patient monitors* for display on those monitors.
*Contact Masimo for the latest list of SatShare validated multiparameter monitors.
Warnings, cautions and notes
Please read and follow any warnings, cautions and notes presented throughout this manual. An explanation of these labels are as follows:
A WARNING is provided when actions may result in a serious outcome (i.e., injury, serious adverse affect, death) to the patient or user. Look for text in a gray shaded box.
Sample of Warning:
A CAUTION is given when any special care is to be exercised by the patient or user to avoid injury to the patient, damage to this device or damage to other property. Sample of Caution:
CAUTION: THIS IS A SAMPLE OF A CAUTION STATEMENT.
A
NOTE
is provided when extra general information is applicable.
Sample of Note:
NOTE: This is a sample of a Note.
WARNING: THIS IS A SAMPLE OF A WARNING STATEMENT.
overview
1-2 Radical Signal Extraction Pulse Oximeter Operator’s Manual
1
Radical Signal Extraction Pulse Oximeter Operator’s Manual 1-5
1
overview
FUNCTIONAL SATURATION
The Radical is calibrated to measure and display functional saturation (SpO
2
): the amount of oxyhemoglobin expressed as a percentage of the hemoglobin that is available to transport oxygen.
MEASURED VS. CALCULATED SATURATION
Oxygen measurements that can be obtained from the Radical are commonly compared to invasive measurements obtained from blood gas samples. When comparing invasive and noninvasive measurements and interpreting values, caution should be used, as the calculated values obtained from the blood gas sample may differ from the SpO
2
measurements of the pulse oximeter. Different results are usually obtained from the arterial blood gas sample if the calculated measurement is not appropriately corrected for the effects of variables that shift the relationship between the partial pressure of oxygen (PO2) and saturation, such as: pH, temperature, the partial pressure of carbon dioxide (PCO2), 2,3-DPG, and fetal hemoglobin.As blood gas samples are usually taken over a period of 20 seconds (the time it takes to draw the blood) a meaningful comparison can only be achieved if the core oxygen saturation of the patient is stable and not changing over the period of time that the blood gas sample is taken.
MASIMO SET SIGNAL EXTRACTION TECHNOLOGY
Masimo Signal Extraction Technology’s signal processing differs from conventional pulse oximeters. Conventional pulse oximeters assume that arterial blood is the only blood moving (pulsating) in the measurement site. During patient motion, however, the non­arterial blood also moves, which causes conventional pulse oximeters to read low values, because they cannot distinguish between the arterial and venous blood movement (sometimes referred to as noise). Masimo SET pulse oximetry utilizes parallel engines and adaptive digital filtering. Adaptive filters are powerful because they are able to adapt to the varying physiologic signals and/or noise and separate them by looking at the whole signal and breaking it down to its fundamental components. The Masimo SET signal processing
algorithm, Discrete Saturation Transform
®
(DST)®, reliably identifies the noise, isolates it and, using adaptive filters, cancels it. It then repor ts the true arterial oxygen saturation for display on the monitor.
Pulse oximetry
GENERAL DESCRIPTION
Pulse oximetry is a continuous and non-invasive method of measuring the level of arterial oxygen saturation in blood. The measurement is taken by placing a sensor on a patient, usually on the fingertip for adults, and the hand or foot for neonates.The sensor is connected to the pulse oximetry instrument with a patient cable. The sensor collects signal data from the patient and sends it to the instrument. The instrument displays the calculated data in three ways:
1. As a percent value for arterial oxygen saturation (SpO
2
)
2. As a pulse rate (PR)
3. As a plethysmographic waveform
The following figure shows the general monitoring setup.
PRINCIPLE OF OPERATION
Pulse oximetry is governed by the following principles:
1. Oxyhemoglobin (oxygenated blood) and deoxyhemoglobin (non-oxygenated blood) differ in their absorption of red and infrared light (spectrophotometry).
2. The amount of arterial blood in tissue changes with your pulse (photoplethysography). Therefore, the amount of light absorbed by the varying quantities of arterial blood changes as well.
The Radical Pulse Oximeter uses a two-wavelength pulsatile system to distinguishbetween oxygenated and deoxygenated blood. Signal data is obtained by passing red(rd)(660 nm wavelength) and infrared (ir)(905 nm wavelength) light through a vascular bed (for example a fingertip, a hand, a foot) and measuring changes in light absorption during the pulsatile cycle. This information may be useful to clinicians. The radiant power of the light is rated at 0.79mW (max.). See figure below. The Radical utilizes a sensor with red and infrared light-emitting diodes (LEDs) that pass light through the site to a photodiode (photodetector). The photodetector receives the light, converts it into an electronic signal and sends it to the Radical for calculation.
Once the Radical receives the signal from the sensor, it utilizes Masimo SET signal extraction technology for calculation of the patient’s functional oxygen saturation and pulse rate.
Recessed Photo Detector
LED's
overview
1-4 Radical Signal Extraction Pulse Oximeter Operator’s Manual
1
3
2
1
1. Instrument
2. Patient Cable
3. Sensor
Radical Signal Extraction Pulse Oximeter Operator’s Manual 2-1
2
system description
Introduction
Radical provides the functionality of three pulse oximeters in one:
The Radical is a fully featured Handheld pulse oximeter
The Radical is a fully featured standalone pulse oximeter
The Radical interfaces to the SpO2input module of multiparameter patient monitors* to upgrade conventional pulse oximetry technology to Masimo SET technology.
CAUTIONS:
THE WAVEFORM DISPLAYED ON THE MULTIPARAMETER PATIENT MONITOR IS A SIMULATED SIGNAL (NON-NORMALIZED). REFER TO THE RADICAL DISPLAY FOR PATIENT WAVEFORM.
IF DISPLAYING THE SIMULATED WAVEFORM IS NOT DESIRABLE, IT IS RECOMMENDED TO TURN OFF THE PLETH WAVEFORM DISPLAY ON THE MULTIPARAMETER MONITOR
ONLY USE A SATSHARE CABLE THAT HAS A FERRITE BEAD INSTALLED.
Refer to Section 3,
SatShare Setup
and Section 4,
SatShare Operation
for additional
details.
*Contact Masimo for the latest list of SatShare validated multiparameter monitors.
Utilizing a SatShareTMcable, the standalone Radical also interfaces with the SpO
2
input
of a validated multiparameter patient monitor*, instantly upgrading the conventional pulse oximetry to Masimo SET pulse oximetry. The SatShare cable attaches to the back of the Radical Docking Station, and SatShare cables are available to interface with most multiparameter patient monitors*.
4
90
The Handheld portion of the Radical contains the majority of the pulse oximeter features. All pulse oximetry measurement information, as well as device status data is displayed on the Handheld LCD screen. All user input is handled through the control buttons on the front panel. The sensor cable connector is located on the Radical Handheld oximeter.
The Handheld oximeter snaps into the Radical Docking Station to provide a fully featured standalone pulse oximeter. The Docking Station connects to AC power for standalone operation or charging of the Handheld. An optional Docking Station battery is available. The standalone Radical features nurse call, analog output and interfaces to serial printers.
overview
1-6 Radical Signal Extraction Pulse Oximeter Operator’s Manual
1
Masimo SET DST
Masimo SET Parallel Engines
Digitized, Filtered &
Normalized
R/IR
R/IR
(Conventional
Pulse Oximetry)
Adaptive Filter with
DST
TM
SST
TM
Proprietary Algorithm 5
Saturation
Post
Processor
Confidence
Based
Arbitrator
MEAS
CONF
MEAS
CONF
MEAS
CONF
MEAS
CONF
MEAS
CONF
Proprietary
Algorithm 4
Radical Signal Extraction Pulse Oximeter Operator’s Manual 2-3
2
system description
1
HANDHELD RELEASE BUTTON
Press down the Handheld Release Button and pull the Handheld device off the Docking Station.
2
SpO
2
MEASUREMENT DISPLAY
The functional arterial hemoglobin oxygen saturation is displayed in units of percentage SpO2. The upper and lower SpO2alarm limits are also displayed next to the SpO2measurement. When a sensor is not connected to a patient and during pulse search, the display will show dashed lines. When the measured value is outside of the alarm limits, the SpO2Measurement Display flashes and an alarm will sound. The oxygen saturation is calculated and the display is updated at a frequency of once per second.
3
Masimo
SET
MASIMO SET
The Masimo SET symbol is shown on the Radical display when SET processing is active.
Note: Patient information, not Masimo SET, will be displayed when using a RadNet compatible unit connected to a RadNet central station.
4
FastSat
FASTSAT
The FastSat symbol is shown on the Radical display whenever the Radical is set to operate in the FastSat mode.
5
SATURATION ALARM LIMITS DISPLAY
The Saturation Alarm Limits Display shows the upper and lower saturation alarm limits. When an alarm limit is reached or exceeded, the SpO2value and the violated limit flashes.
6
ALARM STATUS INDICATOR
The alarm status indicator (a bell) can be shown with or without a slash. It flashes when an alarm condition is present. When the alarm is silenced using the Alarm Silence Button, an alarm status indicator with a slash and a timer is shown to indicate that the alarm is temporarily silenced. When the alarm is silenced through All Mute menu selection, which is permanent until power is cycled or deselected using menu, an alarm status indicator with a slash is shown to indicate that alarm has been silenced.
7
SYSTEM MESSAGE AREA
The system messages generated by the instrument are displayed in the System Message Area. See Section 5,
System messages
.
8
Max
APOD
SENSITIVITY
The Max or APOD sensitivity icon is shown on the Radical display to indicate if the Radical is set to operate in Normal sensitivity, Maximum sensitivity or Adaptive Probe Off Detection mode.When in Normal mode, this area will appear blank.
9
BATTERY STATUS INDICATORS
The Battery Status Indicators show the capacity of the Radical Handheld and optional Docking Station batteries. The indicator flashes when less than 15 minutes of battery life is left and the battery needs to be recharged. The Docking Station Battery Status indicator is not shown when the optional Docking Station battery is not present.
10
TOUCH KEY CONTROL BUTTONS
Press a Touch Key Control Button to select the corresponding touch key icon. See Section 4,
Touch Key Control Buttons and Icons
for more
details.
11
SPEAKER
The speaker indicates audio alarms. Care should be taken not to cover the speaker and muffle the audible alarm volume.
12
PATIENT CABLE CONNECTOR
Connect the patient cable to the Handheld Radical by plugging the cable into the Patient Cable Connector. Use only Masimo compatible sensors and cables with this oximeter. See Section 8, Sensors and Patient Cables for more details.
13
TOUCH KEY ICONS
The Touch Key Icons indicate the software menu items that can be selected through the Touch Key Control Buttons. Pressing a Touch Key Control Button next to an icon selects the option.
Radical handheld
The Handheld Radical provides most of the functionality of the pulse oximeter. All user input and displays are controlled by this part of the Radical system. The sensor cable connects into the swivel connector on the Handheld unit. The Handheld is battery powered and can be used either as a transport monitor or as a Handheld Pulse Oximeter for spot checks.
HANDHELD FRONT PANEL
The following figure and corresponding text outline all the features of the Handheld Radical:
Pleth + Signal IQ View
Numbers View
Masimo SET
11:42:46 03/27/03
10085%SpO
2
160
40
BPM
23
5
22
06:10:23pm 03/27/03
140
50
BPM
100
85
%SpO
2
No Sensor
1110
9
3
2
20
18 17 12
19
1
22423 151316 14
6
HPO1
Masimo SET FastSat
5
7
8
21
system description
2-2 Radical Signal Extraction Pulse Oximeter Operator’s Manual
2
14
SIGNAL IQ
The Signal IQ shows the acquired signal quality and the timing of the pulse. A tall vertical line indicates a high quality signal, while a small vertical line indicates a low quality signal. The Signal IQ may also be shown as a single, pulsating bar.
15
PULSE WAVEFORM DISPLAY
The Pulse Waveform Display shows the acquired plethysmograph waveform.The pleth waveform is scaled with signal strength. Signal strength is defined as the relation of arterial pulsatile signal to the non-pulsatile signal component.
16
TIME AND DATE INDICATOR
The Time and Date Indicator displays the current time and date. The time is displayed in 12 or 24 hour format. The date is displayed in dd/mm/yy or mm/dd/yy format.Select the date and time display formats in the Clock menu.
17
BACKLIGHT/ CONTRAST BUTTON
Press the Backlight Button to change the illumination level of the backlight. With the AC line power connected, four levels of illumination are available (in addition to the no illumination level). In the Handheld mode, three levels of illumination are available (again in addition to the no illumination level). Use the lowest illumination for most efficient battery usage. The backlight Button is also used to change the contrast of the LCD display. Press and hold the Backlight Button for longer than six seconds to change the contrast. Release the Backlight Button at the desired contrast setting.
18
ALARM SILENCE BUTTON
Press the Alarm Silence Button to temporarily silence patient and low battery alarms. Press the Alarm Silence Button when the SENSOR OFF or NO SENSOR messages are flashing (i.e. the sensor is removed from the patient) to acknowledge the end of monitoring. In these states, all further alarms are suspended until the pulse oximeter starts measuring SpO2and pulse rate again.
Note: System failure alarms can be silenced by pressing the Power/Standby or Alarm Silence Button. If the Power/Standby Button does not silence the system fault alarm, press the Alarm Silence Button
.
19
OUTPUT MODE INDICATOR
The Output Mode Indicator displays the output mode selected by the user (Philips Vuelink or Spacelabs Flexport). The Output Mode Indicator also displays the type of SatShare® cable. The Output Mode Indicator is only displayed when the Radical device actively outputs data other than ASCII text, or interfaces with a monitor through the SatShare cable. The Output Mode Indicator will take place of the Time and Date indicator when utilized.
Philips Vuelink
20
POWER/STAND BY BUTTON
Press the Power/Standby Button to turn the instrument on. Press, hold the button for more than 2 seconds and then release the button to turn the instrument off.
21
PI
PERFUSION INDEX
The Perfusion Index indicates numerically the percentage of pulsatile signal to non-pulsatile signal (pulse strength)
22
PULSE RATE ALARM LIMITS DISPLAY
The Pulse Rate Alarm Limits Display shows the upper and lower pulse rate alarm limits. When an alarm limit is reached or exceeded, the pulse rate value and the violated limit flashes.
23
PULSE RATE
The Pulse Rate Measurement Display shows the patient’s pulse rate in beats per minute. The upper and lower pulse rate alarm limits are also displayed next to the pulse rate measurement. The pulse rate is calculated and the display is updated at a frequency of once per second.
system description
2-4 Radical Signal Extraction Pulse Oximeter Operator’s Manual
2
Radical Signal Extraction Pulse Oximeter Operator’s Manual 2-5
2
system description
HANDHELD BACK PANEL
The Handheld back panel features the interconnection to the Docking Station, an accessory mount for the pole clamp accessory and access to the Handheld battery pack.
1
DOCKING STATION CONNECTOR
The Radical Handheld interfaces with the Docking Station through this connector.
2
POLE CLAMP ACCESSORY HOLDER
The optional Pole Clamp accessory attaches to this holder. See the Directions for Use of the Pole Clamp accessories for attachment instructions.
3
BATTERY PACK
The Radical Handheld is powered by a NiMH battery located in this compartment. For battery care and replacement please see Section 9,
Replacing the Batteries
.
1251 0000 536
MEDICAL ELECTRICAL EQUIPMENT
UL 2601-1/CAN/CSA C22.2 No. 601.1
80FK
11328/2269M-0805
Covered by one or more of the following U.S. Patents: RE38,492,
RE38,476, 6,850,787, 6,826,419, 6,816,741, 6,699,194,
6,684,090, 6,658,276, 6,654,624, 6,650,917, 6,643,530,
6,606,511, 6,584,336, 6,501,975, 6,463,311, 6,430,525,
6,360,114, 6,263,222, 6,236,872, 6,229,856, 6,206,830,
6,157,850, 6,067,462, 6,011,986, 6,002,952, 5,919,134,
5,823,950, 5,769,785, 5,758,644, 5,685,299, 5,632,272,
5,490,505, 5,482,036, international equivalents, or one or more
of the patents referenced at www.masimo.com/patents.
Products containing Satshare® feature are also covered by
U.S. Patent 6,770,028. Other patents pending.
1
2 3
Radical Signal Extraction Pulse Oximeter Operator’s Manual 2-7
2
system description
STANDALONE FRONT PANEL
The following figure and corresponding text review the features of the Radical Standalone device.
NOTE: When the Radical Standalone is turned on, all indicator LEDs initially turn on and off at start up.
1
DOCKING STATION BATTERY CHARGING INDICATOR
The Docking Station Battery Charging Indicator is illuminated when the Docking Station battery is charging. The indicator blinks just prior to charging. The charging indicator does not illuminate when the battery is fully charged or when the battery is not present.
2
HANDHELD BATTERY CHARGING INDICATOR
The Handheld Battery Charging Indicator is illuminated when the Handheld battery is charging. The indicator blinks just prior to charging. The Charging Indicator does not illuminate when the battery is fully charged or when the battery is not present.
3
VISUAL ALARM INDICATOR
The Visual Alarm Indicator is illuminated when an alarm condition is active and the Alarm Status Indicator is shown.
4
AC POWER INDICATOR
The AC Power Indicator is illuminated when the Radical Docking Station is plugged into AC line power.
5
DOCKING INDICATOR
The Docking Indicator is illuminated when the Handheld unit is turned on and is properly interfaced to a Docking Station.
Check Sensor
Masimo SET FastSat
11:42:46 03/27/03
100
85
%SpO
2
140
50
BPM
1
2
3
4
5
Radical standalone
When the Radical Handheld is placed into the Docking Station, the Radical becomes a full-featured standalone device.The Radical Standalone acts as a battery charger for the Handheld device and has AC power connection capabilities. The Standalone can also interface to serial devices, nurse call or analog output devices, and multiparameter patient monitors through a SatShare cable.
There are several models of Docking Stations available.The following table outline which features are available for each model of Docking Station.
The RDS-1 and RDS-3 are optionally available with RadNet capability. (Refer to Section 3
RadNet Setup for details).
DOCKING STATION FEATURES RDS-1 RDS-1B RDS-2 RDS-3 RDS-3B RDS-4
AC Power Input
SatShare Interface
Serial RS-232 Interface
Nurse Call/Analog Output Interface
12-hour Extended Battery
Automatic Display Rotation Support (Gravity Detector)
Docking Station Battery Charging Indicator
Handheld Battery Charging Indicator
Red Alarm Indicator
AC Power Indicator
Docking Indicator
Handheld Battery Deep Discharge Support
Docking Station Battery Deep Discharge Support
system description
2-6 Radical Signal Extraction Pulse Oximeter Operator’s Manual
2
Radical Signal Extraction Pulse Oximeter Operator’s Manual 2-9
2
system description
SYMBOLS
The following symbols are found on the Radical unit, Docking Station or packaging and are defined below:
NOTE: Some of the interfaces and symbols are not available in all versions of the
Docking Station.
STANDALONE BACK PANEL
1
SERIAL OUTPUT CONNECTOR
Use the Serial Output Connector with a ferrite bead installed to connect a serial device, including a serial printer, RadNet Interface Module or PC, to the Radical. The data is provided in standard RS­232C format. See Section 7,
Serial Interface Specifications
. All external device connections to the Serial Output Connector must be IEC-60950 compliant.
2
ANALOG OUTPUT / NURSE CALL CONNECTOR
Use the Analog Output Connector with a ferrite bead installed to interface with an analog output device, such as a chart recorder or nurse call system. All external device connections to the Analog Output / Nurse Call Connector must be IEC-60950 compliant.
3
SATSHARE CABLE CONNECTOR
Use the SatShare Cable Connector to connect a SatShare cable to the SpO2input connector of a multiparameter patient monitor. All external device connections to the SatShare Cable Connector must be IEC-60601-1 compliant. SatShare cables are available to interface with most major multiparameter patient monitors. Check the label on the SatShare cable and the SatShare Directions for Use to ensure that the correct cable is used for each type of patient monitor. Refer to the Masimo web site at www.masimo.com for the latest SatShare cables and validated instruments.
4
POWER ENTRY MODULE
The power entry module contains the input connector for AC power and two fuses. The AC input provides power to the system from the AC line. Always connect the pulse oximeter to the mains power for continuous operation and/or battery recharging.
NOTE: Use the power cord as the means to disconnect the device from the mains power supply.
5
EQUIPOTENTIAL GROUND CONNECTOR
Use the Equipotential Ground Connector for grounding.
1
2
3
5 4
P2 P3
Manufactured by: Masimo Corporation Irvine, CA 92618 USA Made in USA
Rx Only
Covered by one or more of the following U.S. Patents: RE38,492, RE38,476, 6,850,787, 6,826,419, 6,816,741, 6,699,194, 6,684,090, 6,658,276, 6,654,624, 6,650,917, 6,643,530, 6,606,511, 6,584,336, 6,501,975, 6,463,311, 6,430,525, 6,360,114, 6,263,222, 6,236,872, 6,229,856, 6,206,830, 6,157,850, 6,067,462, 6,011,986, 6,002,952, 5,919,134, 5,823,950, 5,769,785, 5,758,644, 5,685,299, 5,632,272, 5,490,505, 5,482,036, international equivalents, or one or more of the patents referenced at www.masimo.com/patents. Products containing Satshare®
P1P1
MEDICAL ELECTRICAL EQUIPMENT
UL 60601-1/CAN/CSA C22.2 No. 601.1
80FK
system description
2-8 Radical Signal Extraction Pulse Oximeter Operator’s Manual
2
SYMBOLS DEFINITION
RS-232
SatShare interface
Equipotential Ground Terminal
See Instructions for Use
Fuse Replacement
Analog Out Interface
Nurse Call Interface
WEEE Compliant
Mark of Conformity to European Medical Device Directive 93/42/EEC
Rx ONLY
Federal law restricts this device to sale by or on the order of a physician (USA audiences only)
Underwriter's Laboratories Inc. approved
NON
STERILE
5%-95% RH
Storage humidity range: 5% to 95%
-40 C
+70 C
+1060 hPa - +500 hPa
795 mmHg - 375 mmHg
Storage temperature range: +70C to -40C Storage altitude range: +1600hPa to +500hPa
Keep dry
Fragile/breakable, handle with care
Year of Manufacture
IPX1
Protection against liquid drops falling vertically.
Defibrillation Proof Type BF
Radical Signal Extraction Pulse Oximeter Operator’s Manual 3-1
3
setup
Introduction
Before the Radical Pulse Oximeter can be used in a clinical setting, it needs to be inspected, properly setup and the batteries need to be fully charged.
Unpacking and inspection
Remove the instrument from the shipping carton and examine it for signs of shipping damage. Check all materials against the packing list. Save all packing materials, invoice and bill of lading. These may be required to process a claim with the carrier.
If anything is missing or damaged, contact the Technical Service Department. The contact address and phone numbers are listed in Section 9,
Service and Repair
.
Preparation for monitoring
The following sections of the manual describe the preparation, set-up and initial installation of the Radical Pulse Oximeter.
RADICAL POWER REQUIREMENTS
Always use a hospital grade, AC power cable to connect the Radical Pulse Oximeter to an AC power source. Do not connect the Radical Pulse Oximeter Docking Station to an AC outlet controlled by a switch.
Verify the AC power voltage and frequency before use. Verify that the power source can provide adequate power rating as indicated on the rear panel of the Radical Docking Station.
The Radical Pulse Oximeter is designed to operate on 100 to 240VAC, 47-63 Hz. The device is rated at 55 VA max.
Connect a hospital grade power cable to the power entry module of the Radical unit (IEC-320 connector type at the Radical). Connect the power cable to an AC power source. Ensure that the unit is adequately powered by verifying that the AC power indicator on the Docking Station is illuminated.
CAUTION:
CONNECT THE OXIMETER ONLY TO A HOSPITAL-GRADE RECEPTACLE.
DO NOT UNDER ANY CIRCUMSTANCES REMOVE THE GROUNDING CONDUCTOR FROM THE POWER PLUG.
DO NOT USE EXTENSION CORDS OR ADAPTERS OF ANY TYPE. THE POWER CORD AND PLUG MUST BE INTACT AND UNDAMAGED.
USE THE POWER CORD AS THE MEANS TO DISCONNECT THE DEVICE FROM THE MAINS POWER SUPPLY.
Radical monitor interface
In addition to being a full-featured Handheld and Standalone Pulse Oximeter, Radical’s unique SatShare interface links the Radical to most existing multiparameter patient monitors through the pulse oximetry patient cable or SpO2input connector.
Upgrades any approved and validated monitor to Masimo SET performance by using the calculated SpO
2
and pulse rate determined by Radical to simulate an ideal
waveform, which is sent to the validated multiparameter patient monitor.
Connects into the SpO2patient cable or SpO2input connector of the multiparameter patient monitor.
Refer to Section 3,
SatShare Setup
and Section 4,
SatShare Operation
for additional
details.
CAUTION: THE WAVEFORM DISPLAYED ON THE MULTIPARAMETER PATIENT MONITOR IS A SIMULATED SIGNAL. REFER TO THE RADICAL DISPLAY FOR THE PATIENT WAVEFORM.
CAUTION: ONLY USE A SATSHARE CABLE THAT HAS A FERRITE BEAD INSTALLED.
system description
2-10 Radical Signal Extraction Pulse Oximeter Operator’s Manual
2
Radical Signal Extraction Pulse Oximeter Operator’s Manual 3-3
3
setup
Monitor setup
The Radical maintains three types of default values which will automatically reverts to after a power cycle:
Factory – these options are restored to factory values
Custom – these settings can be changed by the user and retained through the power cycle.
Adult / Neo – these settings can be selected to revert to factory or hospital-defined values (for Adult or Neonatal) after a power cycle.
The following outlines the default values the Radical reverts to after a power cycle.
FACTORY DEFAULT SETTINGS
CUSTOM (USER) DEFINED SETTINGS
This mode is indicated by “Mode Custom” on the Alar ms menu.
The following table outlines the settings that may be changed by the user and the Radical will remember after a power cycle.
OPTION DEFAULT SETTING
DISPLAY VIEW AND CONTRAST Set to pre-power down setting
AVERAGING TIME Set to pre-power down setting
FASTSAT Set to pre-power down setting
HOME USE Set to pre-power down setting
INTERFACE ALARMS Set to pre-power down setting
SATSHARE NUMBERS Set to pre-power down setting
POWER SAVE Set to pre-power down setting
DATE AND TIME FORMAT Set to pre-power down setting
LANGUAGE Set to pre-power down setting
ANALOG OUTPUT Set to pre-power down setting
SERIAL OUTPUT PORT MODE Set to pre-power down setting
OPTION DEFAULT SETTING
LCD SCREEN ILLUMINATION
AC Power Battery Power
Set to maximum, level 4 Set to off
SENSITIVITY Set to normal mode
IF THERE IS ANY DOUBT ABOUT THE INTEGRITY OF THE PROTECTIVE EARTH CONDUCTOR ARRANGEMENT, OPERATE THE OXIMETER ON INTERNAL BATTERY POWER UNTIL THE AC POWER SUPPLY PROTECTIVE CONDUCTOR IS FULLY FUNCTIONAL.
TO ENSURE PATIENT ELECTRICAL ISOLATION, CONNECT ONLY TO OTHER EQUIPMENT WITH ELECTRICALLY ISOLATED CIRCUITS.
DO NOT CONNECT TO AN ELECTRICAL OUTLET CONTROLLED BY A WALL SWITCH OR DIMMER.
INITIAL BATTERY CHARGING
Before use, the Radical Handheld battery and the optional Docking Station battery needs to be fully charged.
To charge the batteries, attach the Handheld unit to the Docking Station and plug in the AC power cord. Verify that the batteries are charging. The battery charging LED indicators on the Docking Station flash prior to charging and remain illuminated while the batteries are charging.
Refer to Section 9,
Battery Operation and Maintenance
, for proper battery charging.
INITIAL INSTALLATION
Place the Docking Station on a stable hard flat surface near the patient.Always place the Radical unit on a dry surface. Maintain a minimum of 3 cm (1 inch) free space around the Radical Standalone unit. Make sure that Radical loudspeaker is not covered to avoid a muffled alarm sound.
The Radical Handheld, Docking Station or Standalone should not be operated outside the following environmental conditions:
OPERATING ENVIRONMENTAL CONDITIONS
TEMPERATURE
+5°C to +40°C, +41°F to +104°F
HUMIDITY
5% to 95%, non-condensing
OPERATING ALTITUDE
1060 mbar to 500 mbar pressure
-1000 ft to 18,000 ft (-304 m to 5,486 m)
setup
3-2 Radical Signal Extraction Pulse Oximeter Operator’s Manual
3
Loading...
+ 33 hidden pages