Masimo Radical 7 Service manual

Page 1
Signal Extraction Pulse CO-Oximeter
Monochrome and Color Display
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
Page 2
The Radical-7 Pulse CO-Oximeter Service Manual is intended to provide the necessary information for proper servicing of all models of the Radical-7 Pulse Oximetry and CO-Oximetry systems. This manual is structured to support troubleshooting to the assembly or module level. This manual does not provide instructions for troubleshooting to the printed circuit board component level. There may be information provided in this manual that is not relevant for your system. Do not service the Radical-7 pulse oximeter without completely reading and understanding these instructions. Service training provided by Masimo is required prior to utilization of this manual.
NOTICE:
Purchase or possession of this device does not carry any express or implied license to use this device with replacement parts which would, alone or in combination with this device, fall within the scope of one of the patents relating to this device.
CAUTION:
Federal law (U.S.) restricts this device to sale by or on the order of a physician.
Masimo Corporation
40 Parker
Irvine, CA 92618
USA
Tel.: 949-297-7000
Fax.: 949-297-7001
www.masimo.com
EU Authorized Representative for Masimo Corporation:
EC REP
MDSS GmbH
Schiffgraben 41
30175 Hannover, Germany
Tel.: +49-511-62 62 86 30
Fax.: +49-511-62 62 86 33
MEDICAL ELECTRICAL EQUIPMENT WITH RESPECT TO ELECTRIC SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60601­1/CAN/CSA C22.2 No. 601.1
Covered by one or more of the following U.S. Patents: RE38,492, RE38,476, 7,221,971, 7,215,986, 7,215,984, 7,186,966, 6,979,812, 6,861,639, 6,850,787, 6,826,419, 6,816,741, 6,745,060, 6,699,194, 6,684,090, 6,654,624, 6,650,917, 6,643,530, 6,606,511, 6,515,273, 6,501,975, 6,463,311, 6,430,525, 6,388,240, 6,360,114, 6,263,222, 6,236,872, 6,229,856, 6,157,850, 6,067,462, 6,011,986, 6,002,952, 5,919,134, 5,769,785, 5,758,644, 5,685,299, 5,632,272, 5,490,505, 5,482,036, international equivalents, or one or more of the patents referenced at www.masimo.com/patents.htm Products containing Satshare® feature are also covered by U.S. Patent 6,770,028. Other patents pending.
© 2007 Masimo Corporation. LNOP, Masimo, Radical, the Radical logo, Rainbow, Satshare, SET, Signal IQ, and SpCO are federally registered trademarks of Masimo Corporation. FlashApp, PVI, Radical-7, RDS, SpMet and Signal Extraction Pulse CO-Oximeter are trademarks of Masimo Corporation.
ii
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
Page 3
Table of Contents
Page
1 Overview........................................................................................1-1
1.1 About this Manual....................................................................1-1
1.2 Warnings, Cautions and Notes................................................1-1
2 Maintenance ..................................................................................2-1
2.1 Introduction..............................................................................2-1
2.2 Cleaning...................................................................................2-1
3 Battery Operation and Maintenance ...........................................3-1
3.1 Overview..................................................................................3-1
3.2 Battery Charging......................................................................3-1
3.3 Replacing the Handheld Battery..............................................3-2
3.4 Replacing the Docking Station Battery....................................3-2
3.5 Replacing the Fuses................................................................3-3
4 Software.........................................................................................4-1
4.1 Overview..................................................................................4-1
4.2 Software Download Procedure................................................4-1
5 Performance Verification .............................................................5-1
5.1 Overview..................................................................................5-1
5.2 Power-On Self-Test .................................................................5-1
5.3 Key Press Button Test.............................................................5-1
5.4 Alarm Limit Test.......................................................................5-1
5.5 Display Contrast Test on Monochrome only............................5-2
5.6 Testing with Masimo SET Tester.............................................5-2
5.7 %SpCO (Carboxyhemoglobin) Testing (If Applicable) ............5-3
5.8 %SpMET (Methemoglobin) Testing (If Applicable) .................5-3
5.9 PVI (Pleth Variability Index) Testing (If Applicable).................5-4
5.10
Nurse Call Test ....................................................................5-4
5.11
Analog Output Test ..............................................................5-5
5.12
Serial Port ............................................................................5-5
5.13
Battery Test..........................................................................5-6
6 Repair.............................................................................................6-1
6.1 Safety Precautions:..................................................................6-1
6.2 General Procedures.................................................................6-1
6.3 Troubleshooting.......................................................................6-1
6.4 SatShare Diagnostics ..............................................................6-7
7 Repair and Returns.......................................................................7-1
7.1 Repair Policy............................................................................7-1
7.2 Return Procedure ....................................................................7-1
8 Analog Output/Nurse Call Test....................................................8-1
9 Docking Station Power Supply Power Measurement ...............9-1
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
iii
Page 4
10 Portable Functional Block Diagram..........................................10-1
11 Docking Station Functional Block Diagram.............................11-1
12 Radical-7 Isolation Block Diagram............................................12-1
13 Portable Wiring Interconnect.....................................................13-1
14 Docking Station Wiring Interconnect........................................14-1
15 Radical-7 Color Parts List ..........................................................15-1
16 Radical-7 Monochrome Parts List.............................................16-1
17 Docking Station Parts List (RDS-1, RDS-1B and RDS-3)........17-1
18 RDS2 Parts List...........................................................................18-1
iv
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
Page 5
1 Overview
1.1 About this Manual
This manual explains how to service the Radical-7 Pulse CO-Oximeter. Important safety information relating to general use of the Radical-7 Pulse CO-Oximeter appears in the Radical-7 CO-Oximeter Operator’s Manual. Other important safety information is located throughout this manual where appropriate.
1.2 Warnings, Cautions and Notes
Please read and follow any warnings, cautions and notes presented throughout this manual. An explanation of these labels is as follows:
A warning is provided when actions may result in a serious outcome (i.e., injury, serious adverse affect, death) to the patient or user. Look for text in a gray shaded box.
Sample of Warning: WARNING: THIS IS A SAMPLE OF A WARNING STATEMENT.
A caution is given when any special care is to be exercised by the patient or user to avoid injury to the patient, damage to this device or damage to other property.
Sample of Caution: CAUTION: This is a sample of a Caution statement.
A note is provided when extra general information is applicable. Sample of Note:
Note: This is a sample of a Note statement.
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
1-1
Page 6
2 Maintenance
2.1 Introduction
2.1.1 This chapter covers how to test the operation of the Radical­7 and SatShare interface (if applicable), how to properly clean the Radical-7 pulse oximeter, how to recharge and replace the batteries, how to replace the fuses, and how to obtain service under normal operation. No internal adjustment or recalibration is required. Service should be done by qualified personnel only. Safety checks should be performed at regular intervals or in accordance with local and governmental regulations.
WARNING: ELECTRICAL SHOCK AND FLAMMABILITY HAZARD. BEFORE CLEANING THE OXIMETER, ALWAYS TURN THE UNIT OFF AND DISCONNECT THE POWER CORD FROM THE BACK OF THE UNIT.
2.2 Cleaning
2.2.1 To clean the display panel use a cotton swab moistened with 70% isopropyl alcohol and gently wipe the panel.
2.2.2 To clean the outer surface of the oximeter, use a soft cloth moistened with a 70% isopropyl alcohol. Do not allow liquids to enter the interior of the instrument.
2.2.3 To decontaminate unit, refer to Section 7.2.2.
CAUTIONS:
Do not autoclave, pressure sterilize, or gas sterilize this oximeter.
Do not soak or immerse the monitor in any liquid.
Use the cleaning solution sparingly. Allow unit to air dry. Excessive
solution can flow into the monitor and cause damage to internal components.
Do not touch, press, or rub the display panels with abrasive cleaning compounds, instruments, brushes, rough-surface materials, or bring them into contact with anything that could scratch the panel.
Do not use petroleum-based or acetone solutions, or other harsh solvents, to clean the oximeter. These substances attack the device’s materials and device failure can result.
2-1
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
Page 7
3 Battery Operation and Maintenance
3.1 Overview
3.1.1 The Radical-7 Handheld includes a 2.0 Amp-hr Nickel Metal
Hydride (NiMH) battery. The Radical-7 Docking Station may include the optional 6.5 Amp/hr NiMH battery.
3.1.2 Before using the Radical-7 as a handheld or transport
monitor, the Handheld battery and the optional Docking Station battery need to be fully charged.
3.2 Battery Charging
3.2.1 To charge the battery(s), attach the Handheld unit to the
Docking Station.
3.2.2
Note: Some older units are configured with an AC power switch. Turn the AC power switch to the ‘I’ position.
Ensure that AC power is attached to the Docking Station.
3.2.3 The battery charging LED indicator(s) on the Docking Station
will initially flash, then remain illuminated while the battery(s) is (are) charging. A prolonged flashing battery charging LED indicates that the internal battery temperature exceeds recommended operating conditions for proper battery charging. Proper battery charging will proceed when the temperature returns to recommended operating conditions.
3.2.4 The Handheld battery requires approximately 2 to 3 hours
for charging. The optional Docking Station battery requires approximately 6 hours for charging.
3.2.5 When the battery charging LED indicators turn off, additional
trickle charging may occur to complete charging. Although battery charging can occur while the Handheld is docked and powered on, most efficient charge times are achieved with the Handheld unit turned off.
3.2.6 It is advisable to completely discharge and fully recharge the
batteries periodically to maintain optimum battery health. Refer to the Battery Deep Discharge function as described in the Radical-7 Operators Manual, Section 4, Service.
CAUTIONS: on next page.
See
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
3-1
Page 8
CAUTIONS:
If the Radical-7 has not been charged for 1 month or more, then recharge the battery prior to use.
At low battery warning, connect the Radical-7 to AC power to prevent interruption of power.
Additional information on the Radical-7 batteries may be found in the Radical-7 Operators Manual.
3.3 Replacing the Handheld Battery
3.3.1 Turn the Radical-7 Handheld off and remove the patient
cable connection. Detach the Radical-7 Handheld from the Docking Station (if docked).
3.3.2 Place the Handheld face down on a soft surface such that no
cosmetic damage occurs to the unit.
3.3.3 Loosen the closure screw on the battery, or on color models,
the battery compartment door, and remove the battery.
3.3.4 Position the replacement battery into the Handheld, bottom
first. On color models, the battery needs to be plugged in / unplugged.
3.3.5 Replace the enclosure (if applicable) and tighten the battery
screw until it is secure. If there is a grasp loop, make sure it is pointing towards the top of the Handheld and lays flat into the plastic housing.
3.3.6 Place Handheld into Docking Station. Turn on power and if
necessary, charge battery according to Section 9 of the Radical-7 Operator’s Manual: Battery Operation and Maintenance.
3.4 Replacing the Docking Station Battery
CAUTION: Do not over-tighten screws. Tighten the screws throughout
the Radical-7 Handheld and Docking Station to 80 in.- oz.
Note: Docking Station Battery can only be installed in RDS-1B or RDS­3B type replacement units.
Note: It will be necessary to perform safety testing after replacement.
3-2
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
Page 9
3.4.1 With the AC power cord disconnected, remove the Handheld
from the Docking Station.
3.4.2 Lay the Docking Station on its side on a soft surface, with
the screw holes facing up
3.4.3 Remove the five (5) screws from the unit using a #1 Phillips
screw driver.
3.4.4 Carefully turn the Docking Station over to allow the screws to
fall out of the unit in a manner that allows the technician to catch the screws.
3.4.5 Lay the Docking Station on its feet and remove the top
plastic cover.
3.4.6 For units with a battery installed, disconnect the battery
connector (J21) from the Docking Station system board. Remove the screw that holds the Docking Station System board in place. Lift this board to allow access to the Battery. Remove the battery.
3.4.7 Carefully install the replacement battery into its
compartment.
3.4.8 Reinstall the screw to secure the Docking Station system
board. Attach the battery connector.
3.4.9 Carefully install the top plastic cover. Make sure no internal
cables are pinched between top and bottom case.
3.4.10 Turn the unit over, and re-install the screws.
3.4.11 Attach the Handheld onto the Docking Station.
3.4.12 If necessary, charge the battery before using the Docking
Station according to Section 3.2 of this manual.
CAUTION: Follow local governing guidelines for proper disposal of
internal batteries. Do not incinerate.
3.5 Replacing the Fuses
3.5.1 Disconnect unit from AC power.
3.5.2 Use a small flat-blade screwdriver and gently pry loose the
fuse holder from the power entry module.
3.5.3 Note how the fuses are placed in the fuse holder for
installation of the new fuses.
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
3-3
Page 10
3.5.4 To remove the fuses from the fuse holder, use the edge of
the screwdriver blade to pry against the bottom of the metal portion of the fuse.
3.5.5 Place fuses as needed (1 Amp, Metric, fast acting, 5x20mm,
250V) in the fuse holder, observing the proper orientation.
3.5.6 Slide the fuse holder back into the power entry module and
press firmly to make sure it is firmly inserted.
3.5.7 The unit is ready to be reconnected to AC power.
Note: If the fuses blow after replacement, troubleshoot the unit.
WARNING: FIRE HAZARD: TO PROTECT AGAINST FIRE HAZARD,
REPLACE ONLY WITH FUSES OF THE SAME TYPE, CURRENT RATING, AND VOLTAGE RATING.
3-4
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
Page 11
4 Software
4.1 Overview
4.1.1 The Radical-7 product is configured to allow software
downloads directly from a PC. The Radical-7 product has three different boards that utilize software that may require upgrade. The Pulse Oximeter Board, Handheld System Board and Docking Station System Board all operate with software that is independent of each other. The software must be compatible between the boards for the Radical-7 product to operate properly. The boards communicate with each other through on board connections - in the case of the Oximeter Board to Handheld System Board. For the Handheld to Docking Station communication, this is accomplished through a Contact Pin/Interface Cable.
4.1.2 When software upgrades are affected, it is imperative that
FRM-1312 is completed and returned to Masimo to ensure each unit is tracked with the software version of each unit. This form can be found in this manual, in the contents of the software disk and can also be acquired from Masimo if necessary by contacting Masimo using the information in Section 7.2.4.
Note: The Pulse Oximeter Board (MX-1) is not upgradeable using the PC_FlashApp Program. Only the Docking Station and Portable System Board are upgradeable in this fashion. The upgrade of the MX board requires the use of a special field upgrade tool.
4.2 Software Download Procedure
Note: No software is installed in the RDS-2 and the unit does not require
programming.
4.2.1 On the PC, open PC_FlashApp.exe from your hard drive.
Masimo software can be obtained by Authorized Service providers by calling Masimo Technical Support Within the United States at (800) 326-4890, option #2, or From Outside the United States at country code (+1) 949-297-7498 option #2.
4.2.2 Set the default “Target”, “Port” and “Baud” parameters as
shown in the table below.
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
4-1
Page 12
Programming
Software PC_FlashApp Portable COM1 57600 PC_FlashApp Docking Station COM1 57600
Target Port Baud1
4.2.3 For programming the Radical-7 Handheld System Board, perform the following steps:
4.2.4 Connect a Docking Station to AC outlet and turn the AC Switch to ON (if applicable).
4.2.5 Connect a straight through serial cable between the PC and the Docking Station.
4.2.6 Install the Radical-7 Handheld to the Docking Station.
4.2.7 Turn off the Handheld, if it is on.
4.2.8
Start the FlashApp program. Click Browse and select the proper Image file to be loaded. It will be listed under the monitor’s PSA subdirectory.
If loading on a Radical-7C “Color” the proper file will be named PPSA.ROM.
If loading on a Radical-7 “Monochrome” the proper file will be named PPSAMX.ROM.
4.2.9 Select the proper Port (usually COM1), and Select Portable (Blue) as the Target.
4.2.10 Press the “Green” Program button. A ‘status message’ of “waiting for docking station” will be displayed.
4.2.11 Turn on the Radical-7 Handheld.
4.2.12 The green Program button will turn off, and the red Stop button will illuminate. Wait until the green button turns on again indicating a successful transfer.
4.2.13 If the message on the Status window is “Complete – Success”, continue with the next step. If programming is not successful, go back to 4.2.7 and repeat. If this cycle is repeated 3 times without success, troubleshooting is required.
4.2.14 Turn off the Handheld.
1
If a unit does not program with 57,600, set baud rate to 9,600.
4-2
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
Page 13
4.2.15 To program Radical-7 Docking Station, perform the following steps:
4.2.16 Connect a Docking Station to AC outlet and turn the switch to ON position (if applicable).
4.2.17 Connect a serial cable between computer and the Docking Station.
4.2.18 Install a known good Radical-7 Handheld to the Docking Station.
4.2.19 Turn off the Handheld, if it is on.
4.2.20 Start the FlashApp program. Click Browse and select the proper Image file to be loaded. It will usually be listed under a SDSSA subdirectory.
4.2.21 Select the proper Port (usually COM1), and Select Docking Station as the Target.
4.2.22 Press the “Green” Program button. A ‘status message’ of “waiting for docking station” will be displayed.
4.2.23 Turn on the Handheld.
4.2.24 The green Program button will turn off, and the red Stop button will illuminate. Wait until the green button turns on again indicating a successful transfer. .
4.2.25 If the message on the Status window is “Complete – Success”, continue with next step. If the message is “Complete – Fail”, go back to 4.2.19 and repeat. If this cycle is repeated 3 times without success, troubleshooting is required.
4.2.26 To check software version numbers, perform the following steps:
4.2.27 Turn off the Handheld.
4.2.28 Turn on the Handheld.
4.2.29 Press the Menu key.
4.2.30 Press the Down key and select “About”.
4.2.31 Verify that the Handheld (Handheld, Masimo SET) and Docking Station (D-Station) software version(s) are the intended version. The Handheld must be docked to the Docking Station in order to see the version of the Docking Station.
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
4-3
Page 14
4.2.32 Record software version in FRM-1312 and notify Masimo as per Section 4.1.2.
4-4
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
Page 15
5 Performance Verification
Note: Under normal operation, no internal adjustment or recalibration is
required.
5.1 Overview
5.1.1 To test the performance of the Radical-7 pulse oximeter following repairs or during routine maintenance, follow the procedure outlined in this section. If the Radical-7 fails any of the described tests, discontinue its use and correct the problem before returning the unit back to the user.
5.1.2 Before performing the following tests place the Radical-7 Handheld into the Docking Station, connect the Radical-7 to AC power and fully charge the Radical-7 Handheld battery (the AC mains switch must be on). Disconnect any patient cables or pulse oximetry sensors, as well as SatShare, serial or Analog Output/Nurse Call cables from the instrument.
5.2 Power-On Self-Test
5.2.1 Connect the unit to AC power and verify that the AC Power Indicator LED is illuminated. If it is not, verify that the AC mains switch is on.
5.2.2 Turn the unit on by depressing the Power/Standby Button. Within 5 seconds all available LEDs are momentarily illuminated, a 1-second beep tone sounds, and the Masimo SET logo is displayed.
5.2.3 The Docking Indicator LED, AC Power Indicator LED, Handheld Battery Charge LED and Docking Station Charge LED (if Docking Station is equipped with optional battery) are illuminated and the Radical-7 begins normal operation.
5.3 Key Press Button Test
5.3.1 With the exception of the Power/Standby Button, press each soft key button and verify that the Radical-7 acknowledges each key-press with an audible beep tone or by indicating an associated change on the display.
5.4 Alarm Limit Test
5.4.1 With the monitor turned on, select the Menu Access key and enter the Alarm menu. Change the desired parameter, below, to a value two points below the currently selected value, and accept the change.
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
5-1
Page 16
5.4.2 Verify that the newly set parameter is shown on the Saturation Alarm Limit Display, next to the SpO2 or pulse rate measurement display.
5.4.3 Return the Alarm parameter to its original setting.
5.4.4 Perform steps 5.4.1 through 5.4.3 with the High Saturation Alarm parameter.
5.4.5 Perform steps 5.4.1 through 5.4.3 with the Low Saturation Alarm parameter.
5.4.6 Perform steps 5.4.1 through 5.4.3 with the High Pulse Rate Alarm parameter.
5.4.7 Perform steps 5.4.1 through 5.4.3 with the Low Pulse Rate Alarm parameter.
5.4.8 Reset the alarm limits again to the original settings.
5.5 Display Contrast Test on Monochrome only
5.5.1 With the monitor turned on, select the Menu key and enter the Display menu. Change the Contrast parameter by scrolling through the contrast settings.
5.5.2 Return the Contrast setting to the original value, or a value that allows maximum viewing contrast.
5.5.3 Exit the Menu system and press and hold down the Backlight/Contrast button for several seconds. The display will scroll again through all the contrast settings.
5.5.4 Release the Backlight/Contrast button again when the display shows maximum viewing contrast.
5.6 Testing with Masimo SET Tester
5.6.1 Turn the Radical-7 off and then on again.
5.6.2 Set the alarm limits to:
% SpO2 Pulse Rate
High 100 140
Low 90 50
5.6.3 Connect the Masimo tester to the Radical-7. A patient cable
LNOP to Red PC cable adapter (4', 8' or 12')
required to connect a modular tester to the Radical-7.
5-2
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
may be
Page 17
5.6.4 Verify that within 20 seconds a plethysmographic and a Signal IQ waveform displays.
5.6.5 Verify that the SpO2 measurement is between 78% and 84%. (with tester values of 81% +/- 3%)
5.6.6 Verify that the pulse rate measurement is between 59 bpm and 63 bpm (with tester values of 61 +/- 1 bpm).
5.6.7 Verify that an audible alarm occurs and that the SpO2 measurement and the low SpO2 alarm are flashing
5.6.8 Press the Alarm Silence button once and verify that the alarm is silenced.
5.6.9 Wait 120 seconds and verify that the alarm silence times out and the audible alarm is activated again. Disable the Alarm before continuing.
5.6.10 Press the Increase Loudness button several times and verify that the loudness of the pulse beep tone increases.
5.6.11 Press the Decrease Loudness button and verify that the loudness of the pulse beep tone decreases. Press the Decrease Loudness button and verify that the loudness of the pulse beep tone can be turned off.
5.7 %SpCO (Carboxyhemoglobin) Testing (If Applicable)
Note: This is performed as a functional test only.
5.7.1 Connect a Masimo SET Rainbow sensor to the Radical-7 Handheld.
5.7.2 Insert the non-dominate hand’s ring finger into the finger sensor.
5.7.3 Under %SpCO, a healthy, non-smoking individual would have a typical reading between 0 – 5 on the display. (These numbers are for reference only and should not be used for diagnostic purposes)
5.8 %SpMET (Methemoglobin) Testing (If Applicable)
Note: This is performed as a functional test only.
5.8.1 Connect a Masimo SET Rainbow sensor to the Radical-7 Handheld.
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
5-3
Page 18
5.8.2 Insert the non-dominant hand’s ring finger into the finger sensor.
5.8.3 Under %SpMET, a healthy, non-smoking individual would have a typical reading between 0-2% on the display. (These numbers are for reference only and should not be used for diagnostic purposes)
5.9 PVI (Pleth Variability Index) Testing (If Applicable)
Note: This is performed as a functional test only.
5.9.1 Connect a Red Masimo SET Rainbow sensor to the Radical­7 Handheld.
5.9.2 Insert the non-dominant hand’s ring finger into the finger sensor.
5.9.3 PVI will fluctuate depending on motion, perfusion, physiology and other factors, so there is no ‘normal’ value. If a number is displayed, then the PVI feature is functioning properly.
5.10 Nurse Call Test
Note: Refer to Section 8 in this manual for connector pin outs noted in
Sections 5.10
5.10.1 Disconnect the patient cable or the Masimo SET Tester from the Radical-7 and turn the instrument on. Ensure that there are no audible alarms and that the audible alarms are not silenced.
5.10.2 Connect the common lead of a digital multimeter to pin 12 (Nurse Call - Common) of the Analog Output/Nurse Call Connector (see Service Manual, Troubleshooting, Section 8, Analog Output/Nurse Call Test) on the Radical-7. Connect the positive lead of the multimeter to pin 6 (Nurse Call ­Normally Open) of the Analog Output/Nurse Call Connector and measure that the resistance is greater than 1 M (open circuit).
5.10.3 Trigger an alarm on the monitor (e.g. by disconnecting a sensor after it was measuring data) and verify the resistance is < 20.
5-4
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
Page 19
5.11 Analog Output Test
5.11.1 Disconnect all patient cables and sensors from the Radical-
7. Turn the Radical-7 off and then on again.
5.11.2 Connect the common lead of a multimeter to the pin 2 (Ground) of the Analog Output/Nurse Call Connector. Connect the positive lead of the multimeter to pin 9 (Analog
1) of Analog Output/Nurse Call Connector.
5.11.3 Enter the menu system and set the "Output", "Analog 1" to "0v Output". Verify that the voltmeter measures a voltage of approximately between –10mV and 100mV.
5.11.4 Set “Analog 1” to 1v Output”. Verify that the voltmeter measures a voltage between 0.90V to 1.10V.
5.11.5 Reset “Analog1” to “Spo2 0 – 100%”.
5.11.6 Repeat steps 5.8.3 and 5.8.4 with the positive lead of the multimeter connected to pin 15 (Analog2) and set “Analog2” to “0V Output” and “1V Output”.
5.11.7 Reset “Analog2” to “Pulse Rate”.
5.12 Serial Port
5.12.1 The serial port outputs ASCII data using standard RS- 232 protocol via a straight through serial cable. The Handheld must be docked with AC power applied.
5.12.2 Validation of the port can be done by connecting a serial printer or using a PC with Hyper Terminal and the following setup:
5.12.3 Communication Parameters
baud rate 9600 bps
data bits 8
stop bit 1
parity none
hand shaking none
5.12.4
With the Radical-7 connected as per the above setup, attach a patient cable and sensor to the Radical-7. From the Output menu, select ASCII 1 and the printer (or PC monitor) will print in the following format every second.
05/15/07 08:12:21 SN=0000070986 SPO2=---% BPM=--- PI=--.--% SPCO=--.-% SPMET=--.-% DESAT=-- PIDELTA=+-- PVI=--­ALARM=0000 EXC=000000
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
5-5
Page 20
5.13 Battery Test
5.13.1 Fully charge the Radical-7 Handheld Battery by placing the Handheld into the Docking Station and connecting the AC power (the AC mains switch must be on).
5.13.2 Verify that the green Handheld Battery Indicator LED is illuminated.
5.13.3 When the Radical-7 is fully charged the green Handheld Battery Indicator turns off.
5.13.4 Turn the Radical-7 on and verify that the Handheld Battery indicator shows a full charge.
5.13.5 Fully charge the Radical-7 Docking Station Battery by placing the Handheld into the Docking Station and connecting the AC power (the AC mains switch must be on).
5.13.6 Verify that the green Handheld Battery and blue Connection Indicator LEDs, on the Docking Station, are illuminated.
5.13.7 When the Radical-7 is fully charged the green Docking Station Battery Indicator turns off.
5.13.8 Turn the Radical-7 on and verify that the Docking Station Battery indicator shows a full charge.
5-6
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
Page 21
6 Repair
6.1 Safety Precautions:
WARNING: DO NOT DISASSEMBLE OR ASSEMBLE THE UNIT WITH
THE AC POWER CORD ATTACHED.
WARNING: ENSURE THE UNIT HAS BEEN CLEANED PER THE
CLEANING INSTRUCTIONS.
6.2 General Procedures
6.2.1 Examine the unit prior to disassembly or troubleshooting and inspect for cosmetic damage. External damage may be helpful in determining the root cause of the failure.
6.2.2 Note any markings or labeling the end user may have placed on the unit that may need to be duplicated or transferred in the event the housing or ancillary components are to be replaced.
6.2.3 Make note of the unit serial number.
6.2.4 Make note of the software versions in the event the Oximeter board or system boards are replaced, as these boards require software updates to make them compatible with the end users version of software.
6.2.5 Contact Masimo Corporation using the information in Section
7.2.4 to research the warranty history of the unit to determine if the unit will be covered under warranty.
WARNING: A HI-POT SAFETY TEST IS REQUIRED IN THE EVENT
THE DOCKING STATION HAS BEEN DISASSEMBLED.
6.3 Troubleshooting
6.3.1 The troubleshooting procedures in this manual instruct the technician to isolate the failure down to the sub-assembly level. The intention is to have the Unit Under Test (UUT) repaired with minimal steps. A safety analyzer or Hi­pot/Current Leakage test unit will be needed in the event the Docking Station is disassembled. The Masimo SET Tester will be needed to perform the Performance Verification in Section 5 upon completion of repairs.
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
6-1
Page 22
Tools required are as follows:
Phillips head screwdriver Multimeter 10mm wrench or small pliers
6.3.2 The troubleshooting steps below are presented in order of the most probable cause first. Proceed to the next step only if unit continues to fail.
6.3.3 Always note the software versions of all three boards prior to repair.
6.3.4 All units must have Performance Verification in Section 5 performed after repairs have been affected.
6.3.5 All Docking Stations that have been opened must be safety tested after reassembly.
6.3.6
Service Messages
6.3.6.1 R2.0.x.x – When a fault is detected, the unit either
watchdogs or displays “Service Required XXX” (where “XXX” indicates one of the symptoms below), and generates high priority alarm sound. In either case the fault is latched until power is cycled or the alarm silence key is pressed.
6.3.6.2 R2.1.x.x – When a fault is detected, the unit either
watchdogs or displays “XXX” and generates high priority alarm sound. When the fault triggers watchdog alarm, it is latched until the power is cycled or the alarm silence key is pressed. But when the unit displays “XXX”, it also resets itself to clear the fault. If the fault is one-time event, software clears the fault and continues to operate normally. If the fault is permanent, the Radical-7 will display “XXX”, reset, display “XXX”, reset…
Symptom Cause and Corrective Action
Steps
001
002
003
004
SRAM Memory Failure
1. Replace Portable System Board
Flash Memory Failure
1. Replace Portable System Board
Low Speed Bus Timeout
1. Replace Portable System Board
Display or Low Speed Bus
Failure
1. Replace Portable System Board
6-2
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
Page 23
Symptom Cause and Corrective Action
Steps
2. Replace Display
005
006
007
Dual UART Failed at Power Up
1. Replace Portable System Board
Oximeter Board Failure
1. Replace Oximeter Board
Oximeter MX Board
Communication Error
1. Verify Oximeter Board (SET) Software is correct
a. If version is incorrect or not
displayed, return unit to an authorized service center or use a field upgrade tool to reload the oximeter board software.
b. If version is correct, replace
Oximeter Board.
c. Replace Portable System Board
008
009
Real Time Clock Failure
1. Replace Portable System Board
Watchdog Failure
1. Replace Portable System Board
010 Docking Station Watchdog
Failure
1. Replace Docking Station System
Board.
2. Replace Portable System Board
200 Invalid SatShare coefficients
1. Replace Docking Station System
Board.
Symptom Cause and Corrective Action
Steps
All Flashing LED's
1. Flashing alternates between all LED's and AC Power LED.
2. Refer to Radical-7 Operators
Manual for LED status Parameters.
Communication error between
Portable and Docking Station
1. Verify software versions of Portable and Docking Station System Boards using the Portable menu a. If versions are incorrect or not
displayed, reload System Application Software for that device utilizing procedures outlined in the Software
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
6-3
Page 24
Symptom Cause and Corrective Action
Alarm Condition But No
Steps
Download Procedure.
2. Replace The Interface Cable
3. Replace the Portable System Board
4. Replace the Docking Station System Board
No LED'S Illuminated
1. Both Portable and Docking Station are turned on.
2. Refer to Radical-7 Operators
Manual for LED status Parameters.
Power fault
1. Replace external Power Cable
2. Check fuses at AC Input Filter
3. Check 12VDC power (Sections 7 & 8)
4. Replace Docking Station System Board
No Battery Icon Displayed NiMH Battery Fault
1. Replace NiMH battery
2. Replace Portable System Board
Battery LED Not
Illuminated
1. No fault, battery is fully charged
2. Check 12VDC power (Sections 7
& 8)
3. Replace Docking Station System
Board
Battery LED Flashing
1. Ensure Service Mode – Deep
Discharge is not enabled
2. Allow NiMH Battery to cool down
3. Verify Fan operation
4. Replace NiMH Battery
5. Replace Docking Station System
Board.
Alarm LED does not
Illuminate
1. Unit must be in alarm state
1. Replace Docking Station System
Board
2. Replace LED Cable.
2. Audible alarm can be heard
Audible Indication.
1. Unit must be in alarm
Alarm Fault
1. Replace Speaker
2. Replace Portable System Board
condition.
2. Should hear sound upon button depression.
Display Missing Segments
1. Replace LCD
2. Replace Portable System Board
Display Backlight Will Not
Illuminate
Button Function Error
1. Replace Portable System Board
2. Replace Display
1. Replace Keypad
6-4
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
Page 25
Symptom Cause and Corrective Action
Steps
1. Any front panel button
2. Replace Portable System Board
affected
2. Buttons do not activate feature
3. Buttons hard to press
Rightmost Touch Key
Difficult to Depress
1. Open unit. If Swivel Connector flex
cable is interfering with button, replace Swivel Connector
2. Replace Keypad.
"No Sensor", “Defective
Sensor” or “Sensor Off” Displayed
Display Does Not Rotate
Automatically
1. Replace sensor.
2. Replace patient cable.
3. Replace oximeter board.
1. Ensure Docking Station is
powered up
2. Ensure Power Save feature is not
enabled
3. Replace Docking Station System
Board
Display Does Not Rotate
Manually
1. Flex interference? Replace Swivel
Connector.
2. Replace Keypad
3. Replace Portable System Board
Docking Station Fails
When Portable is Docked
1. Ensure software of Handheld and
Docking Station is compatible
2. Replace Portable System Board
3. Replace Docking Station System
Board
4. Replace Interface Cable
Fan Does Not Function
1. Ensure AC power is on
1. Replace Fan
2. Replace Docking Station System
Board
Portable Will Not Power
Up When Undocked
1. If Clicking is heard when On/Off button is depressed,
1. Check NiMH Battery charge
2. Replace Battery if charging is not
successful
3. Replace Portable System Board.
perform Step 3.
Portable will not Power Up
When Docked
1. Docking Station is
1. Replace Interface Cable
2. Replace Docking Station System
Board
functioning
2. Portable powers up undocked
Portable Will Not Turn Off
1. Replace Portable System Board
2. Adjust or replace Keypad
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
6-5
Page 26
Symptom Cause and Corrective Action
Steps
Portable Will Not Undock
From Docking Station
1. Ensure Portable Lock is not in
2. Latch failure, gently pry latch
down; once removed, replace latch and/or striker
Watchdog Alarm
1. Replace Portable System Board
1. Other than "Service" failures
WARNING: IF SYMPTOMS ARE PRESENT AND ARE NOT
CORRECTABLE, DO NOT USE MALFUNTIONING COMPONENT.
6-6
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
Page 27
6.4 SatShare Diagnostics
CAUTION: As SatShare Diagnostics simulates physiological information
(SpO2 and Pulse Rate values), it should be performed with caution, especially in areas with central monitoring. The attending nurse and supervisor should always be made aware that SatShare Diagnostic testing is being performed.
6.4.1
If the SpO2 and pulse rate values displayed on a multi- parameter monitor is in question, perform the following steps.
6.4.1.1 Press Menu – General – SatShare Numbers – Yes.
6.4.2 With the Handheld docked and AC power applied, disconnect the patient cable from Radical-7. Press the following key sequence: Menu – About – More – 2 – 1 – 1 (see Service menu for number key assignments). Five (5)
different SatShare values can be selected by pressing the Select key. The selected value will be displayed on the
multi-parameter monitor if the unit is working properly. The SatShare Diagnostics feature times out if a key is not pressed within 1 minute.
Note: During transient situations, the SpO2 and pulse rate values on the
Radical-7 and multi-parameter monitor may differ by more than the specified tolerance. The averaging time set on the multi-parameter monitor may also affect the delay during transient situations.
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
6-7
Page 28
7 Repair and Returns
7.1 Repair Policy
7.1.1 Masimo or an authorized Service Department must perform warranty repair and service. Do not use malfunctioning equipment. The product must be repaired.
7.1.2 Please clean contaminated/dirty equipment before returning. Make sure it is fully dry before packing the equipment.
7.1.3 To return the Radical-7 unit for service, please follow the Return Procedure outlined in this section.
7.2 Return Procedure
7.2.1 Please clean contaminated/dirty equipment before returning and make sure it is completely dry before packing the equipment.
7.2.2 Recommended decontamination procedure for Radical-7 equipment that has the potential for being exposed to blood or other potentially infectious materials:
7.2.2.1 Put on Lab coat or other protective garment.
7.2.2.2 Put on gloves.
7.2.2.3 Place equipment on the designated decontamination
table.
7.2.2.4 Apply cleaning agent to paper towel. Clean with
70% Isopropyl alcohol wipes, or 1:10 bleach to water solution. Wipe down all aspects of the equipment.
7.2.2.5 Let the equipment air dry completely.
7.2.2.6 Remove gloves.
7.2.2.7 Move equipment to the area designated for
equipment that has been through the decontamination process.
7.2.2.8 Dispose of all contaminated products properly.
7.2.3 Package the equipment securely – in the original shipping container if possible – and enclose the following information and items:
7.2.3.1 Call Masimo From within the United States at (800)
326-4890, option #2, or from outside the United States at country code (+1) 949-297-7498, option #2 and ask for an RMA number.
7-1
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
Page 29
7.2.3.2 Place a copy of the supplied RMA inside the box,
and write the RMA number on the outside of the shipping container
7.2.3.3 A letter or email describing in detail any symptoms
or difficulties experienced with the pulse oximeter.
7.2.3.4 Warranty information – a copy of the invoice or other
applicable documentation must be included
7.2.3.5 Purchase order number to cover repair if the
oximeter is not under warranty, or for tracking purposes if the warranty is in effect
7.2.3.6 Ship-to and bill-to information
7.2.3.7 Person (name, telephone/Telex/fax number, email
and country) to contact for any questions about the repairs
7.2.3.8 A document stating the oximeter has been
decontaminated for blood borne pathogens
7.2.4 Return the pulse oximeter and a copy of the RMA to the following shipping address:
Masimo Corporation
40 Parker
Irvine, California
USA 92618
(Domestic Toll Free): 800-326-4890, option #2 (International): 949-297-7498, option #2 FAX: 949-297-7499
Email: tech@masimo.com
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
13928/4649A-0307
7-2
Page 30
8
Analog Output/Nurse Call Test
8-1
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
Page 31
9 Docking Station Power Supply Power Measurement
Part Number 20241 and 20364
Ground point
-
JP2
JP1
+
AC
Part Number 20293
-
+
JP2
12VDC
12VDC
JP1
AC
Docking Station System Board Power Measurement
PIN1
12 VDC
Fan
J22
PIN1
12 VDC
PIN2
12 VDC
J23
J21
Optional Battery
DC
PIN 8
GROUND
J25
LED Status
PIN5
3.3VDC
PIN6
5.0VDC
J19
Interface
PIN4
7.5VDC
SUPPLY VOLTAGE PIN2 12VDC
BATTERY VOLTAGE PIN3 4.8VDC
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
9-1
Page 32
10 Portable Functional Block Diagram
Keypad
data/addr
Power Key
Alarm Inhibit Key
Real Time Clock Calendar
data/addr
interrupt
Docking Station Power Enable
data/addr
Backlight
Emulator
Portable NiMH
Low Speed Bus
data/clock/frame
data/addr
DSP
SPORT0 Irq2
SPORT1
TAP Irq0
data/addr
flag1
Irq1
Power Manager
off inhibit key
power key
V empty V7.5
Battery Status
Status
empty V7.5
flag2 flag3
flag0
FLASH
data/addr
SRAM
data/addr
Dual Uart
data/addr
Output Port
Input Port
Port A
Port B
interrupt
fail
status
I
V
data/addr
Contrast
data/addr
Watchdog Timer
reset
Waveform Generator/Volume
data/clock/frame
V
Square Wave Generator
Enable
Display Display Contrast Display
Voltage
V
Audio Amplifier
V
V
V
Speaker
MS-X MS-X Reset
Docking Station Docking Station Power
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
10-1
Page 33
11 Docking Station Functional Block Diagram
Low Speed Bus
data/clock/frame
data/addr
/Reset
LED Interface
V
data/addr
Serial EEPROM Interface
data/addr
CLK/Data
SatShare
I
data/addr
Nurse Call
data/addr
V
J25 LED Status Connector
1500VAC ISOLATION
J27 SatShare Connector
J28 Emulator Port Connector
J21 Docking Station Battery Pack Connector
J23 ACDC Connector
J19 Portable Interface Connector
7.5VDC Sepic Power Supply
12VDC
Portable Battery Charger
12VDC
Fan Controller
12VDC
DSP SPORT0
SPORT1
TAP
Irq0
data/addr
flag1 Irq1 /Reset
Power Supervisors
5VDC
Reset
3.3VDC
FLASH
data/addr
SRAM
data/addr
Dual UART
data/addr
Output Port
Input Port
Interrupt
Port A
Port B
Analog Output data/clock/frame
V
RS-232 Level Translator
RX/TX RS-232
RX/TX
RTS/CTS RS-232
RTS/CTS
J24 Analog Output/Nurse Call Connector
1500VAC ISOLATION
J20 Serial Port Connector
J22 Fan Connector
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
11-1
Page 34
12 Radical-7 Isolation Block Diagram
RADICAL ISOLATION BLOCK DIAGRAM
AC PLUG
Protective Earth
Equipotential Terminal
FILTER MODULE
SW
FUSE 1A
FUSE 1A
FERRITE
20241
POWER SUPPLY
L2
2.0mH
3.3M
0.047uF
1000pF
1000pF
100pF
100pF
Double Insulation 4kV
DOCKING STATION SYSTEM BOARD
Basic Insulation 1.5kV
12V-SELV
DC/DC PS
Satshare CF Rated
OPTO
Basic Insulation 1.5kV
FERRITE
L1
NOTE: Verify Power Supply Part Number
20293 POWER SUPPLY
680pF
DC/DC PS
SIP/SOP-Serial Port/Analog Out/Nurse Call IEC- 60950 Compliant Device
OPTO
12V-SELV
Handheld Interface
680pF
HANDHELD
Patient Cable
Basic Insulation 1.5kV
"BF" Type Applied Part
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
12-1
Page 35
13 Portable Wiring Interconnect
Displa
y
Front Panel
Patient
Connector
Speaker
J6
Portable System
Board
J1
J7
J16
J17
J3
J9
J8
MX- x
Board
J1
NiMH
Battery
Pack
J19
Connection
to
Docking Station
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
13-1
Page 36
14 Docking Station Wiring Interconnect
AC
Power Entry
Module
Equipotential Stud
Neutral
Blue
Brown
Yellow/ Green
GROUND
Yellow/
Green
AC Power Cable
or
20320
LINE
Blue
Brown
Yellow/
Green
(N) (L)
JP1
Power
Docking Front
Panel LED's
JP2
Fan
DC Power
J23
J22
J20
Serial Port
P1
Docking
System Board
J24
Analog Out/
Nurse Call
P2
J25
J19
J21
J27
SatShare
P3
LED Status
Handheld
Unit
Interface
Optional
Battery Pack
NOTE: P1, P2 and P3 are not available with RDS-2.
NOTE: P3 is not available with RDS-3.
See Page 11, Operators Manual, for
illustration of the back panel
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
14-1
Page 37
15 Radical-7 Color Parts List
1. Shell, Front, Radical-7 Color 10. Screw, #4-40 x 3/16” Phillips, pan head machine, 18-8 SS 19. Door, Battery, Radical-7 Color
2. Lens, Radical-7 Color 11. MX-1, Standard 20. Label, Radical-7 Color, HH, SN, 4-UP, MX-1
3. Keypad, Radical-7 Color 12. Speaker, Radical, Reinforcement 21. Battery, Lithium, 12.5MM coin, 3v, 48mAH
4. LCD Module, 480 x 272, Color TFT 13. Spring, Speaker, Radical-7 Color 22. Tape, Polyimide, Acrylic Adhesive, 0.75” w, .0025” THK
5. Bracket, LCD, Radical-7 Color 14. Assembly, FIC, Radical-7 Color 23. Label, Radical-7 Color Screen, connector with notch
6. Radical-7c User Interface circuit board 15. Shell, Back, Radical-7 Color
7. Screw #4-40 x 3/8” Phillips, pan head machine, 18-8 SS 16. Screw, #4-40 x 3/4” Phillips, pan head machine, 18-8 SS
8. Radical-7 Color system circuit board 17. Latch, Radical-7 Color
9. Standoff, Hex 3/16”, #4-40 IT, 5/16” L ACI, RoHS Compliant 18. Battery Pack, NiMH, AA, 4.8V, 2000mAH
WARNING: Disassembly or repair by authorized Service Personnel only.
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
15-1
Page 38
16 Radical-7 Monochrome Parts List
1.
2.
4.
7.
3.
8.
9.
10.
12.
22.
13.
15.
18.
20
11.
5.
6.
14.
21.
16.
19.
17.
1. Latch
2. Top Shell
3. Swivel Connector
4. Leash Pins
5. Speaker Gasket
6. Speaker
7. Keypad Adhesive
8. Keypad
9. Display.
10.
Display Retainer
11.
MX-x Standoff Mounting Hardware (4 pcs)
12.
Handheld System Board
13.
Portable System Board Mounting Hardware (2 pcs)85
14.
Standoff (4 pcs)
15.
MX-xx mounting Hardware (4 pcs)
16.
MX-1, MX-3
17.
Back Shell
18.
Plastics Hardware (6 pcs)
19.
NiMH Battery Pack
20.
Radical-7 Logo Label
21.
Lithium Coin Cell
22.
Foam Pad
Portable Board – MX-x board compatibility Table (lower case x indicates compatibility):
Portable System Board MX-12 20755 20699 Radical 7 Portable System Board Re-order Part Number 20841 MX-1 Board Re-order Part Number 2143
WARNING: Disassembly or repair by authorized Service Personnel only.
2
The Oximeter Board re-order part number is different than the part number label on the board. The re-order part number indicates that the board has been tested and programmed. The
portable Systems Board and the MX-x Circuit Board must be compatible. Re-order the part number as indicated in the table above.
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
16-1
Page 39
17 Docking Station Parts List (RDS-1, RDS-1B and RDS-3)
4
2
6
5
1
24
3
27
7
8
9
25
20
10
11
12
14
15
16
13
18
26
17
19
21
23
22
28
Note: Only RDS-1B configured Docking Stations will contain the 20284 Optional Battery Pack.
1. Front Shell
2. LED Status Label 11. AC Power Entry Module
3. Radical Logo Label 12. AC Filter Module Bracket
4. LED Status Cable 13. Fan
5. Interface Cable 14. AC Power Entry Module Bracket Hardware (5 pcs [1x2x2 ] )
6. Interface Cable Hardware (4 pcs [2x2] ) 15. AC Cable Insulator
7. Bottom Shell 16. Power Supply
8. Equi-potential Hardware (4 different pcs )
9. Plastics Hardware (5 pcs) 18. Insulator
10. AC Filter Module Fuses (2 pcs)
17. 20179 AC Power Cable or 20320 (required with 20293 Power Supply)
WARNING: Disassembly or repair by authorized Service Personnel only.
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
19.
DC Power Cable
20.
Top Shell
21.
Docking Station System Board (2 types)
- RDS-1 and RDS-1B (All ports available – P1, P2 and P3)
- RDS-3 (Only ports P1 and P2 available)
22.
Standoff
23.
Docking Station System Board Mounting Screw
24.
No Implied License Label
25.
Certification Label – (Note: Port P3 Closed
- RDS-1 and RDS-1B (All ports open)
- RDS-3 (P3 closed)
26.
Power Supply Hardware (4 pcs)
27.
Front Shell Mounting Hardware (4 pcs [2x2] )
28.
Docking Station Battery Pack (Optional)
17-1
Page 40
18 RDS2 Parts List
25
4
7
19
2
1
3
1.
Front Shell
2.
LED Status Label
3.
Radical Logo Label
4.
LED Status Cable
5.
Interface Cable
6.
Interface Cable Hardware (4 pcs [2x2] )
7.
Bottom Shell
8.
Equipotential Hardware (4 different pcs )
9.
Plastics Hardware (5 pcs)
26
10
15
11
13
14
16
17
24
18
19.
Top Shell
20.
RDS2 System Board
21.
Standoff
22.
Docking Station System Board Mounting Screw
23.
No Implied License Label
24.
Power Supply Hardware (4 pcs)
25.
Certification Label – (Port P1, P2 and P3 closed)
26.
Front Shell Mounting Hardware (4 pcs [2x2] )
20
21
22
6
5
12
23
8
9
10. AC Filter Module Fuses (2 pcs)
11. AC Power Entry Module
12. AC Filter Module Bracket
13. AC Power Entry Module Bracket Hardware (5 pcs [1x2x2 ] )
14. AC Cable Insulator
15. Power Supply
16. 20179 AC Power Cable or 20320 (required with 20293 Power Supply)
17. Insulator
18. DC Power Cable
WARNING: Disassembly or repair by authorized Service Personnel only.
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
18-1