Masimo RAD87A User Manual

Rad-87
Pulse CO-Oximeter
OPERATOR’S MANUAL
Rad-87
Pulse CO-Oximeter
OPERATOR’S MANUAL
The Rad-87 Operating Instructions provide the necessary information for proper operation of all models of the
Rad-87 device. General knowledge of pulse CO-Oximetry and an understanding of the features and functions of the
the instructions in this manual.
NOTICE:
Purchase or possession of this device does not carry any express or implied license to use this device with
replacement parts which would, alone or in combination with this device, fall within the scope of one of the patents
relating to this device.
CAUTION:
Federal law (U.S.) restricts this device to sale by or on the order of a physician.
Masimo Corporation
40 Parker
Irvine, CA 92618
USA
Tel.: 949-297-7000
Fax.: 949-297-7001
www.masimo.com
EU Authorized Representative for Masimo Corporation:
EC REP
MDSS GmbH
Schiffgraben 41
30175 Hannover, Germany
Tel.: +49-511-62 62 86 30
Fax.: +49-511-62 62 86 33
CONFORMS TO UL STD 60601-1, UL STD 763 UL STD 963 AND NSF STD 12;
3149433
Covered by one or more of the following U.S. Patents: RE38,492, RE38,476, 7,221,971, 7,215,986, 7,215,984,
7,186,966, 6,979,812, 6,861,639, 6,850,787, 6,826,419, 6,816,741, 6,745,060, 6,699,194, 6,684,090, 6,654,624,
6,650,917, 6,643,530, 6,606,511, 6,515,273, 6,501,975, 6,463,311, 6,430,525, 6,388,240, 6,360,114, 6,263,222,
6,236,872, 6,229,856, 6,157,850, 6,067,462, 6,011,986, 6,002,952, 5,919,134, 5,769,785, 5,758,644, 5,685,299,
5,632,272, 5,490,505, 5,482,036, international equivalents, or one or more of the patents referenced at www.masimo.
com/patents.htm. Other patents pending.
© 2008 Masimo Corporation. Masimo, Discrete Saturation Transform, DST, DCI, Signal Extraction Technology, SET,
Radical,
Rainbow and SpCO are federally registered trademarks of Masimo Laboratories.
Rad-87, Pleth Variability Index, PVI, Patient SafetyNet, RadNet, LNOPv and APOD are trademarks of Masimo
Corporation. Rainbow SET, SpMet, SpHb, SpOC, Pulse CO-Oximeter and Signal Extraction Pulse CO-Oximeter
are trademarks of Masimo Laboratories.
Rad-87 Pulse CO-Oximeter Operator’s Manual i
CERTIFIED TO CAN/CSA STD C22.2 NO. 601.1
, Signal IQ, SIQ, FastSat, LNOP and LNCS are federally registered trademarks of Masimo Corporation.
SAFETY INFORMATION, WARNINGS, CAUTIONS AND NOTES
The Rad-87 Pulse CO-Oximeter is designed to minimize the possibility of hazards from errors in the software program by following sound engineering design processes, Risk Analysis and Software Validation.
Explosion hazard. Do not use the Rad-87 in the presence of fl ammable anesthetics or other fl ammable
substance in combination with air, oxygen-enriched environments, or nitrous oxide.
High intensity extreme lights (including pulsating strobe lights) directed on the sensor, may not allow
the Pulse CO-Oximeter to obtain readings.
The Rad-87 is NOT intended for use as an apnea monitor.
The Pulse CO-Oximeter should be considered an early warning device. As a trend towards patient
hypoxemia is indicated, blood samples should be analyzed by laboratory instruments to completely understand the patient’s condition.
Pulse rate measurement is based on the optical detection of a peripheral flow pulse and therefore
may not detect certain arrhythmias. The pulse oximeter should not be used as a replacement or substitute for ECG based arrhythmia analysis.
The Rad-87 is to be operated by qualifi ed personnel only. This manual, accessory directions for use,
all precautionary information, and specifi cations should be read before use.
Electric shock hazard. Do not open the Rad-87 device. Only a qualifi ed operator may perform
maintenance procedures specifi cally described in this manual. Refer servicing to Masimo for repair of this equipment.
As with all medical equipment, carefully route patient cabling to reduce the possibility of patient
entanglement or strangulation.
Do not place the Rad-87 or accessories in any position that might cause it to fall on the patient. Do not
lift the Rad-87 by the power cord or any other cable.
Interfering Substances: Dyes, or any substance containing dyes, that change usual blood pigmentation
may cause erroneous readings.
SpO
is empirically calibrated to functional arterial oxygen saturation in healthy adult volunteers with
2
normal levels of carboxyhemoglobin (COHb) and methemoglobin (MetHb). A pulse oximeter can not measure elevated levels of COHb or MetHb. Increases in either COHb or MetHb will affect the accuracy of the SpO
For increased COHb: COHb levels above normal tend to increase the level of SpO
of increase is approximately equal to the amount of COHb that is present. NOTE: High levels of COHb may occur with a seemingly normal SpO laboratory analysis (CO-Oximetry) of a blood sample should be performed.
NOTE: High levels of COHb may occur with a seemingly normal SpO
COHb are suspected, laboratory analysis (CO-Oximetry) of a blood sample should be performed.
For increased MetHb: the SpO
to 15%. At higher levels of MetHb, the SpO levels of MetHb are suspected, laboratory analysis (CO-Oximetry) of a blood sample should be
measurement.
2
. The level
2
. When elevated levels of COHb are suspected,
2
. When elevated levels of
2
may be decreased by levels of MetHb of up to approximately 10%
2
may tend to read in the low to mid 80s. When elevated
2
performed.
Elevated levels of Methemoglobin (MetHb) will lead to inaccurate SpO
measurements.
and SpCO
2
Elevated levels of Carboxyhemoglobin (COHb) will lead to inaccurate SpO2 measurements.
Elevated levels of Total Bilirubin may lead to inaccurate SpO2, SpMet, SpCO, SpHb and SpOC
measurements.
Motion artifact may lead to inaccurate SpMet, SpCO, SpHb and SpOC measurements.
Rad-87 Pulse CO-Oximeter Operator’s Manualii
SAFETY INFORMATION, WARNINGS, CAUTIONS AND NOTES (CONTINUED)
Very low arterial Oxygen Saturation (SpO2) levels may cause inaccurate SpCO and SpMet
measurements.
Severe anemia may cause erroneous SpO
Hemoglobin synthesis disorders may cause erroneous SpHb readings.
readings.
2
Do not use the Rad-87 or sensors during magnetic resonance imaging (MRI) scanning. Induced
current could potentially cause burns. The Rad-87 may affect the MRI image and the MRI device may affect the accuracy of the Pulse CO-Oximetry parameters and measurements.
If using Rad-87 during full body radiation, keep the sensor out of the radiation fi eld. If the sensor is
exposed to the radiation, the reading might be inaccurate or the device might read zero for the duration of the active irradiation period.
For home use, ensure that the Rad-87’s alarm can be heard from other rooms in the house especially
when noisy appliances such as vacuum cleaners, dishwashers, clothes dryers, televisions, or radios are operating.
Always remove the sensor from the patient and completely disconnect the patient from the Rad-87
before bathing the patient.
Additional information specific to Masimo sensors including information about parameter/measurement
performance during motion and low perfusion, may be found in the sensor's Directions For Use (DFU).
Do not place the Rad-87 where the controls can be changed by the patient.
Do not place the Rad-87's face against a surface. This will cause the alarm to be muffl ed.
Do not place the Rad-87 on electrical equipment that may affect the Pulse CO-Oximeter, preventing it
from working properly.
Do not expose the Rad-87 to excessive moisture such as direct exposure to rain. Excessive moisture
can cause the device to perform inaccurately or fail.
Do not place containers with liquids on or near the Rad-87. Liquids spilled on the device may cause it
to perform inaccurately or fail.
If the Rad-87 fails any part of the setup procedures or leakage tests, remove the device from operation
until qualifi ed service personnel have corrected the situation.
Patient Safety - If a sensor is damaged in any way, discontinue use immediately.
Disposal of product - Comply with local laws in the disposal of the device and/or its accessories.
The Rad-87 can be used during defi brillation, but the readings may be inaccurate for up to 20
seconds.
This equipment has been tested and found to comply with the limits for medical devices to the
EN 60601-1-2: 2002, Medical Device Directive 93/42/EEC and Class B digital device, Part 15, FCC Rules/USA. These limits are designed to provide reasonable protection against harmful interference in a typical medical installation.
Rad-87 Pulse CO-Oximeter Operator’s Manual iii
SAFETY INFORMATION, WARNINGS, CAUTIONS AND NOTES (CONTINUED)
This equipment generates, uses and can radiate radio frequency energy and, if not installed
and used in accordance with the instructions, may cause harmful interference to other devices in the vicinity. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to other devices, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
Reorient or relocate the receiving device.
Increase the separation between the equipment.
Connect the equipment into an outlet on a circuit different from that to which the other
device(s) are connected.
Consult the manufacturer for help.
In order to connect wirelessly to a compatible interface system like Patient SafetyNet, the
Rad-87 should be placed in an environment free from RF shielding, which could hinder wireless reception.
To minimize radio interference, other electrical equipment that emits RF transmissions should
not be in close proximity to the RAD-87.
Changes or modifications to the wireless radio feature whether intentional or unintentional are
prohibited without written approval from Masimo Corporation.
The Rad-87 (device with optional radio) wirelessly transmits real-time sensor connectivity status,
indicating a connect and/or disconnect state. If the device is in a failure mode then the radio power is disabled and an error message is indicated on the device display. The device does not have a powered state where no information is transmitted.
In accordance with FCC requirements, the Rad-87 (device with optional radio) must be placed
greater then 20 cm from the patient's head.
In accordance with FCC requirements, radio accessories on the Rad-87 (device with optional
radio) cannot be attached directly to the patient using any accessory containing metal components.
In accordance with international telecommunication requirements, the frequency band of
5,150 MHz to 5,250 MHz is only for indoor usage to reduce potential for harmful interference to co-channel mobile satellite systems.
The battery should be adequately charged to ensure backup power in case of AC power
disruption.
A functional tester cannot be utilized to assess the accuracy of the Pulse CO-Oximeter or any
sensors.
To ensure safety, avoid stacking multiple devices or placing anything on the device during
operation.
Ensure the speaker is not covered or the device is placed face-down on bedding or other
sound absorbing surface.
To protect against injury from electric shock, follow the directions below:
Avoid placing the device on surfaces with visible liquid spills.
Do not soak or immerse the device in liquids.
Always turn off and disconnect the power cord from the AC power supply before cleaning
the device.
Use cleaning solutions sparingly.
Rad-87 Pulse CO-Oximeter Operator’s Manualiv
table of contents
SECTION 1 - OVERVIEW
About This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Warnings, Cautions and Notes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Optional Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Indications for Use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Pulse CO-Oximetry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
SpO2 General Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
SpCO General Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
SpMet General Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
SpHb General Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Total Arterial Oxygen Content (SpOC) General Description . . . . . . . . . . . . . . . . . 1-5
Principle of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Functional Saturation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Rad-87 vs. Drawn Whole Blood Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Masimo SET Signal Extraction Technology for SpO2 Measurements . . . . . . . . . 1-7
SpMet, SpCO, and SpHb and SpOC Measurements During Patient Motion . . . . 1-7
Fastat . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Masimo Rainbow SET Parallel Engines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Masimo SET DST
SECTION 2 - SYSTEM DESCRIPTION
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Rad-87 Front Panel Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Rad-87 Rear Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
LCD Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
® . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
SECTION 3 - SETUP
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Unpacking and Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Preparation for Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Rad-87 Power Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Initial Battery Charging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Initial Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
SECTION 4 - OPERATION
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Basic operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
General Setup and Use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Default Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Factory Default and User Configurable Settings . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Successful Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Masimo Pulse CO-Oximetry Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Numeric Display - SpO
Numeric Display - Pulse Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Numeric Display - SpCO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Numeric Display - SpMet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Numeric Display - SpHb . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Rad-87 Pulse CO-Oximeter Operator’s Manual v
2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4-5
table of contents
SECTION 4 - OPERATION (CONTINUED)
Numeric Display - SpOC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Numeric Display - PI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Pleth Variability Index (PVI) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
Low Perfusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
Signal SIQ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
Sensor Placement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
Sensitivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Low Battery Audible Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Normal Patient Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Paramater/Measurement Selection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
Setup Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
Menu Navigation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
Setup Menu Level 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
Parameter/Measurement Alarm Limits - Screen 1 . . . . . . . . . . . . . . . . . . . . . . . 4-13
Parameter/Measurement Alarm Limits - Screen 2 . . . . . . . . . . . . . . . . . . . . . . . 4-14
Parameter/Measurement Alarm Limits - Screen 3 . . . . . . . . . . . . . . . . . . . . . . . 4-15
LED Brightness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-15
Sensitivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-16
Setup Menu Level 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-17
Alarm Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-17
Alarm Silence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-17
Alarm Delay. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-18
Clear Trend . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-18
Button Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-19
FastSat . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-19
Trend Setup and Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-19
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-19
Trend Utility Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-19
Trendcom Utility Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-20
Erasing Trend Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-20
Trend Data Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-21
Sample Trend Output. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-21
Setup Menu Level 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-22
Averaging Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-22
Rapid Desat Limit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-23
Alarm On/Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-23
Out Put Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-24
Default Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-24
Device Profile Setup and Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-25
PVI Bar On/Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-26
Smart Tone On/Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-26
Year . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-26
Month . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-26
Day . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-27
Hour. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-27
Minute . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-28
Software Version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-28
Serial Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-29
Rad-87 Pulse CO-Oximeter Operator’s Manualvi
table of contents
SECTION 4 - OPERATION - (CONTINUED)
Interface Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-29
System Interfaces. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-29
Philips VueLink Set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-29
Radnet Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-30
Patient SafetyNet Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-30
Nurse Call . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-31
Polarity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-31
Line Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-31
Parameter/Measurement Select - Screen 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-32
Parameter/Measurement Select - Screen 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-32
Parameter/Measurement Select - Screen 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-33
LCD Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-33
Set Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-33
Home Mode Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-34
Sleep Mode Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-34
Enable/Disable Radio . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-34
LCD Display Function with Radio Configured and Enabled . . . . . . . . . . . . . . . . . . .4-35
SECTION 5 - ALARMS AND MESSAGES
Alarm Identification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Alarm Indication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Alarm Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Alarm Limit: User Configurable Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Alarm Limit: Factory Default Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Single Alarm Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Multi-Parameter/Measurement Alarm Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Alarm Priority for Display Screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Alarm Silence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Alarm Bell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
System Status Light . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Alarm Mute . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
SECTION 6 - TROUBLESHOOTING
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
SECTION 7 - SPECIFICATIONS
Rad-87 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Environmental . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Physical Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Serial Interface Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4
Serial Interface Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4
Serial Printer Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-5
Nurse Call Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-5
Nurse Call . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-5
Rad-87 Pulse CO-Oximeter Operator’s Manual vii
table of contents
SECTION 8 - SENSORS & PATIENT CABLES
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
Selecting a Masimo SET Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
Sensor Application Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
Masimo Rainbow Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2
Rainbow Reusable Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2
Rainbow Direct Connect Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2
Masimo SpO2 Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3
Red Direct Connect Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3
LNOP® Reusable Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3
LNOP® Adhesive Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4
LNOPvTM Adhesive Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4
LNOP® Specialty Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4
LNCS® Reusable Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4
LNCS® Adhesive Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-5
®
LNCS
Specialty Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-5
Sensor Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-6
Cleaning And Reuse Of Masimo Reusable Sensors and Cables . . . . . . . . . . . . . . . . . . 8-6
Reattachment of Single Use Adhesive Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-6
SECTION 9 - SERVICE AND MAINTENANCE
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1
Battery Service and Performance Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-2
Service and Repair. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-4
Repair Policy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-4
Return Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-4
Sales & End-user License Agreement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-5
Warranty. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-5
Exclusions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-5
End-user License . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-6
Restrictions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-6
SECTION 10 - PART NUMBERS
Part Numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-1
Rad-87 Pulse CO-Oximeter Operator’s Manualviii
overview
About This Manual
This manual explains how to set up and use the Rad-87 Pulse CO-Oximeter containing Masimo Rainbow SET technology. Important safety information relating to general use of the Rad-87 appears before this introduction. Other important safety information is located throughout the manual where appropriate.
Read the entire safety information section before you operate the monitor.
In addition to the safety section, this manual includes the following sections:
SECTION 1 OVERVIEW gives a general description of Rad-87 Pulse CO-Oximeter.
SECTION 2 SYSTEM DESCRIPTION describes the Rad-87 Pulse CO-Oximeter
system and its functions and features.
SECTION 3 SETUP describes how to setup the Rad-87 Pulse CO-Oximeter for use.
SECTION 4 OPERATION describes the operation of the Rad-87 Pulse CO-Oximeter
system.
SECTION 5 ALARMS AND MESSAGES describes the alarm system messages.
SECTION 6 TROUBLESHOOTING describes troubleshooting information.
SECTION 7 SPECIFICATIONS gives the detailed specifi cations of the Rad-87 Pulse
CO-Oximeter.
1
SECTION 8 SENSORS & PATIENT CABLES outlines how to use and care for Masimo
SECTION 9 SERVICE AND MAINTENANCE describes how to maintain, service and
SECTION 10 ACCESSORIES lists the available models of the Rad-87 Pulse
Rad-87 Pulse CO-Oximeter Operator’s Manual 1-1
Rainbow SET technology sensors, Masimo Rainbow SET technology patient cables, Masimo Red sensors and, Masimo Red PC cables.
obtain repair for the Rad-87 Pulse CO-Oximeter.
CO-Oximeter.
1
overview
Warnings, Cautions and Notes
Please read and follow any warnings, cautions and notes presented throughout this manual. An explanation of these labels are as follows:
A WARNING is provided when actions may result in a serious outcome (i.e., injury, serious adverse affect, death) to the patient or user. Look for text in a gray shaded box.
Sample of Warning:
WARNING: THIS IS A SAMPLE OF A WARNING STATEMENT.
A CAUTION is given when any special care is to be exercised by the patient or user to avoid injury to the patient, damage to this device or damage to other property.
Sample of Caution:
CAUTION: THIS IS A SAMPLE OF A CAUTION STATEMENT.
A NOTE is provided when extra general information is applicable.
Sample of Note:
NOTE: This is a sample of a Note.
1-2 Rad-87 Pulse CO-Oximeter Operator’s Manual
overview
Product Description
The Rad-87 Pulse CO-Oximeter Monitor is a noninvasive, arterial oxygen, carboxyhemoglobin and methemoglobin saturation, total hemoglobin concentration, total arterial oxygen content and pulse rate monitor. The Rad-87 features a multicolored LED display that continuously displays numeric values for SpO2, SpCO®*, SpMet™*, SpHb™*, total arterial oxygen content (SpOC*), perfusion index (PI), pleth variability index* (PVI) and pulse rate. It also provides bar graph displays for quick visual identifi cation of Signal Identifi cation Quality (SIQ®), perfusion index and pleth variability index.
The Rad-87 is available in four models: vertical Rad-87, horizontal Rad-87, vertical Rad-87 with radio and horizontal Rad-87 with radio.
FEATURES
These features are common to Rad-87 monitors:
Masimo SET is clinically proven to be the highest sensitivity and specifi city pulse oximeter
technology in the world.
Rainbow technology continuously and noninvasively measures arterial oxygen saturation (SpO2)
and pulse rate (BPM), as well as providing a reliable probe-off detection.
Perfusion Index (PI) with trending capability indicates arterial pulse signal strength during low
perfusion.
Accurate on cyanotic infants with congenital heart disease when used with an LNOP® Blue
Sensor.
Signal IQ® provides signal identifi cation and quality indication during excessive motion and low
signal to noise situations.
FastSat® tracks rapid changes in arterial O2 saturation with high fidelity.
Variable pitch provides tonal variance for every 1% change in saturation.
Remote alarming interface.
Up to 72 hours of trending. (See Section 4, Trends Setup and Use.)
Allows user to customize the default settings and set the device to retain these settings through a
power off/on cycle.
The LCD Display allows the user to view a scrolling marque of (installed) parameter/measurement
alarm limits, system information, and wireless radio communication (wireless radio model only).
1
OPTIONAL FEATURES
Rainbow technology uses 7+ wavelengths of light to continuously and noninvasively
measure carboxyhemoglobin (SpCO), methemoglobin (SpMet) and total hemoglobin (SpHb), as well as providing a reliable probe-off detection.
Pleth Variability Index (PVI) may show changes that refl ect physiologic factors such as
vascular tone, circulating blood volume, and intrathoracic pressure excursions.
Total arterial oxygen content (SpOC) provides a calculated measurement of the amount of
1
oxygen in arterial blood which may provide useful information for both oxygen dissolved in plasma and combined with hemoglobin.
Provides a 802.11a/b/g wireless radio which interfaces with compatible systems (wireless radio
model only)
Ability to connect to Masimo Patient SafetyNet through a wireless network (wireless radio
model only).
1
The utility of PVI is unknown at this time and requires further clinical studies. Technical factors that may affect PVI
include probe malposition and patient motion.
*Optional features: SpCO, SpMet, SpHb, SpOC, PVI
Rad-87 Pulse CO-Oximeter Operator’s Manual 1-3
1
overview
INDICATIONS FOR USE
The Rad-87 Pulse CO-Oximeter and accessories are indicated for the continuous, non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO methemoglobin concentration expressed in percentage (SpCO and SpMet) and total hemoglobin
), carboxyhemoglobin and
2
concentration expressed in grams per deciliter (SpHb).The Rad-87 Pulse CO-Oximeter and accessories are indicated for use with adult, pediatric and neonatal patients during both motion and no motion conditions, who are well or poorly perfused patients in hospitals, hospital-type facilities, mobile and home environments.
Pulse CO-Oximetry
SpO2 GENERAL DESCRIPTION
Pulse CO-Oximetry is a continuous and noninvasive method of measuring the level of arterial oxygen saturation in blood. The measurement is taken by placing a sensor on a patient, usually on the fi ngertip for adults and the hand or foot for neonates. The sensor is connected to the Pulse CO-Oximetry instrument with a patient cable. The sensor collects signal data from the patient and sends it to the instrument.
The following fi gure shows the general monitoring setup.
1
N O
R
M A P O
D
M A X
97
S
E N S I
T
I
V
I T
Y M O
D E
76
10.1
r b
c m
o n i t
o r
2
3
1. Instrument
2. Patient Cable
3. Sensor
SpCO GENERAL DESCRIPTION
Pulse CO-Oximetry is a continuous and noninvasive method of measuring the levels of carboxyhemoglobin concentration (SpCO) in arterial blood. It relies on the same basic principles of pulse oximetry (spectrophotometry) to make its SpCO measurement. The measurement is obtained by placing a sensor on a patient, usually on the fi ngertip for adults and the hand or foot for infants. The sensor connects either directly to the Pulse CO-Oximetry instrument or through an instrument patient cable. The sensor collects signal data from the patient and sends it to the instrument. The instrument displays the calculated data as percentage value for the SpCO, which refl ect blood levels of carbon monoxide bound to hemoglobin.
SpMet GENERAL DESCRIPTION
Pulse CO-Oximetry is a continuous and noninvasive method of measuring the levels of methemoglobin concentration (SpMet) in arterial blood. It relies on the same basic principles of pulse oximetry (spectrophotometry) to make its SpMet measurement. The measurement is obtained by placing a sensor on a patient, usually on the fi ngertip for adults and the hand or foot for infants. The sensor connects either directly to the Pulse CO-Oximetry instrument or through an instrument patient cable. The sensor collects signal data from the patient and sends it to the instrument. The instrument displays the calculated data as percentage value for the SpMet.
TOTAL HEMOGLOBIN (SpHb) GENERAL DESCRIPTION
1-4 Rad-87 Pulse CO-Oximeter Operator’s Manual
overview
Pulse CO-Oximetry is a continuous and noninvasive method of measuring the levels of total hemoglobin (SpHb) in arterial blood. It relies on the same principles of pulse oximetry to make the SpHb measurement. The measurement is taken by a sensor capable of measuring SpHb, usually on the fingertip for adults and pediatric patients. The sensor connects directly to the Pulse CO-Oximeter or with a patient cable. The sensor collects signal data from the patient and sends it to the instrument. The instrument displays the calculated data as measurement of total hemoglobin concentration. The Rad-87 can be configured to be a combined SpO
TOTAL ARTERIAL OXYGEN CONTENT (CaO2) GENERAL DESCRIPTION
Oxygen (O2) is carried in the blood in two forms, either dissolved in plasma or combined with hemoglobin. The amount of oxygen in the arterial blood is termed the oxygen content (CaO is measured in units of ml O whereas 100 ml of blood plasma may carry approximately 0.3 ml of oxygen. The oxygen content is determined mathematically as:
CaO
= 1.34 (ml O2/g Hb) x Hb (g/dl) x HbO2 + PaO2 (mm Hg) x (0.3 ml O2/ 100 mm Hg/dl)
2
Where HbO arterial oxygen.
For typical PaO Hg/dl] is approximately 0.3 ml/dl. Furthermore, for typical carboxyhemoglobin and methemoglobin levels, the functional saturation (SpO
2
Martin, Laurence. All You Really Need to Know to Interpret Arterial Blood Gases, Second Edition.
New York: Lippincott Williams & Wilkins, 1999.
monitor with other available parameters/measurements.
2
2
) and
/dl blood. One gram of hemoglobin (Hb) can carry 1.34 ml of oxygen,
2
is the fractional arterial oxygen saturation and PaO2 is the partial pressure of
2
values, the second part of the above equation [PaO2 (mm Hg) x (0.3 ml O2/ 100 mm
2
) as measured by a pulse oximeter is given by:
2
SpO
= 1.02 x HbO2
2
2
1
SpOC General Description (Pulse CO-Oximetry)
The above approximations result in the following reduced equation for oxygen content via the Pulse CO-Oximeter:
SpOC (ml/dl*) = 1.31 (ml O
*When ml O the gram unit in the numerator of g/dl resulting in ml/dl (ml of oxygen in one dl of blood) as the
/g Hb is multiplied by g/dl of Hb, the gram unit in the denominator of ml/g cancels
2
/g Hb) x SpHb (g/dl) x SpO2 + 0.3 ml/dl
2
unit of measure for SpOC.
Rad-87 Pulse CO-Oximeter Operator’s Manual 1-5
1
overview
PRINCIPLE OF OPERATION
Pulse CO-Oximetry is governed by the following principles:
1. Oxyhemoglobin (oxygenated blood), deoxyhemoglobin (non-oxygenated blood), carboxyhemoglobin (blood with carbon monoxide content), methemoglobin (blood with oxidized hemoglobin) and blood plasma constituents differ in their absorption of visible and infrared light (using spectrophotometry, see fi gure below).
Absorption Spectra
4.0
3.5
3.0
2.5
2.0
1.5
Absorption (1/mm)
1.0
0.5
0 600 800
Carboxyhemoglobin
Oxyhemoglobin
Methemoglobin
Deoxyhemoglobin
Plasma
1000 1200 1400 1600
Wavelength (nm)
2. The amount of arterial blood in tissue changes with your pulse (photoplethysography). Therefore, the amount of light absorbed by the varying quantities of arterial blood changes as well.
The Rad-87 Pulse CO-Oximeter uses a multi-wavelength sensor to distinguish between oxygenated blood, deoxygenated blood, blood with carbon monoxide, oxidized blood and blood plasma. The Rad-87 utilizes a sensor with various light-emitting diodes (LEDs) that pass light through the site to a photodiode (detector). See fi gure below. Signal data is obtained by passing various visible and infrared lights (LED’s, 500 to 1400nm) through a capillary bed (for example, a fi ngertip, a hand, a foot) and measuring changes in light absorption during the blood pulsatile cycle. This information may be useful to clinicians. The maximum radiant power of the strongest light is rated at 25mW. The detector receives the light, converts it into an electronic signal and sends it to the Rad-87 for calculation.
1
1. Light Emitting Diodes (LEDs) ( 7+ wavelengths )
2. Detector
2
Once the Rad-87 receives the signal from the sensor, it utilizes Masimo Rainbow SET signal extraction technology to calculate the patient’s functional arterial oxygen saturation, blood levels of carboxyhemoglobin (SpCO), methemoglobin (SpMet) and pulse rate. The SpCO and SpMet measurements rely on a multiwavelength calibration equation to quantify the percentage of carbon monoxide and methemoglobin in arterial blood. In an ambient temperature of 35º C the maximum skin surface temperature has been measured at less than 106º F (41º C), verifi ed by Masimo sensor skin temperature test procedure.
1-6 Rad-87 Pulse CO-Oximeter Operator’s Manual
overview
FUNCTIONAL SATURATION
The Rad-87 is calibrated to measure and display functional saturation (SpO expressed as a percentage of the hemoglobin that is available to transport oxygen.
RAD-87 vs. DRAWN WHOLE BLOOD MEASUREMENTS
When SpO compared to drawn whole blood (invasive) measurements by blood gas and/or laboratory CO-Oximetry
, SpCO, SpMet and SpHb measurements obtained from the Rad-87 (noninvasive) are
2
methods, caution should be taken when evaluating and interpreting the results. The blood gas and/or laboratory CO-Oximetry measurements may differ from the SpO of the Rad-87 Pulse CO-Oximeter. In the case of SpO arterial blood gas sample if the calculated measurement is not appropriately corrected for the effects of
, SpCO, SpMet and SpHb measurements
2
, different results are usually obtained from the
2
variables that shift the relationship between the partial pressure of oxygen (PO as: pH, temperature, the partial pressure of carbon dioxide (PCO the case of SpCO, different results are also expected if concentration of methemoglobin in the blood gas sample is elevated. High levels of bilirubin may cause erroneous SpO readings. As blood samples are usually taken over a period of 20 seconds (the time it takes to draw the blood) a meaningful comparison can only be achieved if the oxygen saturation, carboxyhemoglobin and methemoglobin concentration of the patient are stable and not changing over the period of time that the blood gas sample is taken. Subsequently, blood gas and laboratory CO-Oximetry measurements of SpO2, SpCO, SpMet and SpHb may vary with the rapid administration of fl uids and in procedures such as dialysis. Additionally, drawn, whole-blood testing can be affected by sample handling methods and time elapsed between blood draw and sample testing.
MASIMO SET SIGNAL EXTRACTION TECHNOLOGY FOR SpO
Masimo Signal Extraction Technology’s signal processing differs from conventional pulse oximeters. Conventional pulse oximeters assume that arterial blood is the only blood moving (pulsating) in the measurement site. During patient motion, however, the venous blood also moves, causing conventional pulse oximeters to read low values, because they cannot distinguish between the arterial and venous blood movement (sometimes referred to as noise). Masimo SET pulse oximetry utilizes parallel engines and adaptive digital fi ltering. Adaptive fi lters are powerful because they are able to adapt to the varying physiologic signals and/or noise and separate them by looking at the whole signal and breaking it down to its fundamental components. The Masimo SET signal processing algorithm, Discrete Saturation Transform
®
(DST
) reliably identifi es the noise, isolates it and, using adaptive fi lters, cancels it. It then reports the true
arterial oxygen saturation for display on the monitor.
SpMet, SpCO, AND SpHb MEASUREMENTS DURING PATIENT MOTION
The Rad-87 displays measurements of SpCO, SpMet and SpHb during patient motion. However, because of the changes in the physiological parameters such as blood volume, arterial-venous coupling, etc. that occur during patient motion, the accuracy of such measurements may not be reliable during excessive motion. When the Rad-87 does not have confidence in the value of a parameter due to poor signal quality caused by excessive motion or other signal interference, the measurement for the parameter will alternate with "---".
FASTSAT
FastSat enables rapid tracking of arterial oxygen saturation changes. Arterial oxygen saturation data is averaged using pulse oximeter averaging algorithms to smooth the trend. When the Rad-87 is set to FastSat “On”, the averaging algorithm evaluates all the saturation values providing an averaged saturation value that is a better representation of the patient’s current oxygenation status. With FastSat, the averaging time is dependent on the input signal.
): the amount of oxyhemoglobin
2
) and saturation, such
), 2,3-DPG, and fetal hemoglobin. In
2
2
, SpMet, SpCO and SpHb
2
MEASUREMENTS
2
®
1
Rad-87 Pulse CO-Oximeter Operator’s Manual 1-7
1
MASIMO RAINBOW SET PARALLEL ENGINES
This fi gure is for conceptual purposes only.
12
FST
MST
SpOC
MASIMO SET DST
1-8 Rad-87 Pulse CO-Oximeter Operator’s Manual
system description
Introduction
The Rad-87 Pulse CO-Oximeters are full featured devices designed for ease of operation. All pulse CO-Oximetry measurement information, as well as device status data, is displayed on the front panel of the device. All user input is handled by control buttons on the front panel. The sensor cable connections are located on the left side of the front panel for the Rad-87 horizontal device and the bottom of the front panel for the Rad-87 vertical device.
Rad-87 offers full Masimo SET technology in a small compact device
Rad-87 supports the full line of Masimo sensors and patient cables (see Section 8, Sensors
and Patient Cables)
Rad-87 supports standardization of sensors, and pulse CO-Oximetry technology throughout
the hospital
The LCD Display identifies system settings, monitoring modes, alarm limits and information
from Patient SafetyNet or Philips VueLink (when connected). The LCD is located on top of the device (Horizontal) or on the left of the device (Ver tical).
2
Rad-87 Pulse CO-Oximeter Operator’s Manual 2-1
2
1234567890123456
APOD
MAX
NORM
APOD
MAX
NORM
system description
RAD-87 PULSE CO-OXIMETER - HORIZONTAL
1 23 5678 91011
APOD
APOD
APOD
APOD
CONTROL / INDICATOR DESCRIPTION
Device Profile
1
LED
4
DISPLAY
NORM
NORM
NORM
NORM
SENSITIVITY
SENSITIVITY
MODE
MODE
SENSITIVITY
SENSITIVITY
MAX
MAX
MAX
MAX
10.1
DISPLAY
ENTER
ENTER
rbc monitor
rbc monitor
Rad-87
Rad-87
121314151617181920
The Device Profile LED illuminates when the device has been
set to user configured "default" settings. Upon power up, the
user configured default settings are retained and the Device
Profile LED remain lit.
When user configured default settings are active, any changes
to the default settings cause the Device Profile LED to turn
off until the device is returned to the user configured default
settings or powered off.
Sensitivity
2
Button/Indicator
Used to set the device into Maximum Sensitivity, Normal
Sensitivity, or APOD Mode.
The LCD display identifies system settings, monitoring modes, alarm
3
LCD Display
limits, and information from Patient SafetyNet or Philips VueLink (when
connected.)
The Signal IQ provides an indication of the quality of the acquired
4
Signal IQ Index
5
Perfusion Index
2-2 Rad-87 Pulse CO-Oximeter Operator’s Manual
signal as well as the timing of the pulse. A green vertical LED bar
rises and falls with the pulse, where the height of the bar indicates
the quality of the signal.
The Perfusion Index provides an indication of the percentage of
pulsatile signal to non pulsatile signal.
system description
CONTROL / INDICATOR DESCRIPTION
Off: No connection to Masimo Patient SafetyNet or other
compatible interface system.
6
Wireless Indicator
Flashing Green: Rad-87 attempts to connect to Patient
SafetyNet or other compatible interface system.
Solid Green: Rad-87 is connected to the Patient SafetyNet or
other compatible interface system.
2
Alarm Limits
7
Button
8
Display Button
9
Alarm Bell
Alarm Silence
10
Button
System Status
11
Light
DISPLAY
DISPLAY
Used to enter the alarm menu to adjust Hi/Low SpO
SpHb, PI, PVI and pulse rate alarm limits.
SpCO, SpMet,
2
,
The LED indicator (located above the Alarm Limits Button) will
illuminate when one or more of the factory default alarm settings
is changed to alert the user to verify alarm settings.
Allows movement through the 3 different display screens to view
sets of parameters and measurements.
Also used to exit setup menu screens and return the display to
screen 1.
Press and hold the button down for 5 seconds to scroll through
device settings on the LCD Display.
The Alarm Bell flashes red to indicate a high priority alarm.
Press the Alarm Silence Button to temporarily silence patient
and low battery alarms. Press the Alarm Silence Button
when the “SEN OFF” message is flashing (i.e. the sensor
is removed from the patient) to acknowledge the end of
monitoring. In this state, all further alarms are silenced until
the Pulse CO-Oximeter starts measuring patient parameters/
measurements again.
NOTE: The alarm silence time can be set for 120, 90, 60 and
30 seconds. See Section 4 - Setup Menu Level 2.
Solid Green: Collecting data, no alarms.
Solid Yellow:
1. low priority alarms.
2. Not monitoring and no alarms.
3. Sleep Mode.
4. Interface Alarms "Off".
Flashing Yellow:
1. Low parameter/measurement confidence.
2. Medium priority alarms.
Flashing Red: High priority alarms.
Rad-87 Pulse CO-Oximeter Operator’s Manual 2-3
2
system description
CONTROL / INDICATOR DESCRIPTION
12
Enter Button
Used to enter the setup menus and to select/activate certain
entries within the menu/setup system.
Brightness
13
Button
Up Button
14
Down Button
Pleth Variability
15
Index
Battery Charge
16
Level Indicator
AC Power
17
Indicator
Controls the level of the brightness for the LED display by
providing 4 levels of brightness. Each press of the button
increases the brightness one level. Once level 4 is accessed,
an additional press of the button returns the brightness to level 1.
Use these buttons to adjust the volume of the pulse beep tone.
Within the menu/setup system, these buttons are used to select
values within each menu option or the numeric value for the
parameter/measurement alarm feature.
Pressing and holding down these buttons allow for the rapid
scrolling of alarm limits.
PVI is displayed as a percentage.
The lower the height of the bar, the less variability there is in the
PI over a respiratory cycle.
Press the Display key to toggle to the PVI numeric
measurement.
Provides a visual representation of the battery charge status.
When unplugged, bars illuminate to indicate battery charge. As
the battery discharges power, bar illumination decreases from
right to left.
A low battery status is indicated by a low audible beep and the
first battery bar to the left flashing green.
The AC Power Indicator is illuminated when the Rad-87 is
connected to AC power and during battery charging.
18
Power Button
Used to turn the device on and off. Press the button once to
power on the device. Press the button for 2 seconds to power
off the device.
Pulse CO-Oximeter
19
Patient Cable
Connects to a Masimo Pulse CO-Oximeter sensor or Masimo
Pulse CO-Oximeter Patient Cable with a sensor.
Connector
20
Speaker
2-4 Rad-87 Pulse CO-Oximeter Operator’s Manual
Provides audible indication of alarm conditions, pulse tone and
feedback for key-presses.
system description
MAX
NORM
RAD-87 PULSE CO-OXIMETER - VERTICAL
2
11
8
4
15
2
18
3
17
16
888
888
888
97
88
88
76
10.1
88
8 8 8
888
NORM
MAX
8
8
8
9
10
5
7
13
12
6
14
1
20
19
Rad-87 Pulse CO-Oximeter Operator’s Manual 2-5
2
RAD-87 REAR PANEL
system description
NURSE CALL
1
CONNECTOR
SERIAL OUTPUT
2
CONNECTOR
2
4
1
RS-232
P1
P2
100-240v~50-60 Hz 15VA MAX100-240v~50-60 Hz 15VA MAX
Made in USA Manufactured by:
Masimo Corporation Irvine, CA 92618 USA
Covered by one or more of the following U.S. Patents: RE38,492, RE38,476, 7,221,971, 7,215,986, 7,215,984, 7,186,966, 6,979,812, 6,861,639, 6,850,787, 6,826,419, 6,816,741, 6,745,060, 6,699,194, 6,684,090, 6,654,624, 6,650,917, 6,643,530, 6,606,511, 6,515,273, 6,501,975, 6,463,311, 6,430,525, 6,388,240, 6,360,114, 6,263,222, 6,236,872, 6,229,856, 6,157,850, 6,067,462, 6,011,986, 6,002,952, 5,919,134, 5,769,785, 5,758,644, 5,685,299, 5,632,272, 5,490,505, 5,482,036, international equivalents, or one or more of the patents referenced at www.masimo.com/patents. Other patents pending.
Distributed by: Masimo Corporation Irvine, CA 92618 USA
Rx Only
32294/4778A-1207
3
Use the 1/4” round Connector to interface with a nurse call system. This is a stereo output and should be utilized with a stereo cable. All external device connections to the Nurse Call Connector must be IEC-60950 compliant.
Use the Serial Output Connector to connect a serial device, including a serial printer, RadNet Interface Module, or PC, to the Rad-87. See Section 7, Output Interface Specifications. All external device connections to the Serial Output Connector must be IEC-60950 compliant.
The power entry module contains the input connector for AC power.
POWER ENTRY
3
MODULE
EQUIPOTENTIAL
4
GROUND CONNECTOR
2-6 Rad-87 Pulse CO-Oximeter Operator’s Manual
The AC input provides power to the system from the AC line. Always connect the Rad-87 to the main power for continuous operation and/ or battery recharging.
Use the Equipotential Ground Connector for grounding.
system description
7
SYMBOLS
The following symbols are found on the Rad-87 or packaging and are defined below:
RS-232
Equipotential Ground Terminal
Consult accompanying documents
Nurse Call Interface
WEEE compliant
Mark of Conformity to European Medical Device Directive 93/42/EEC
2
Rx Only
5%-95% RH
+70 C
-40 C
+1060 hPa - +500 hPa 95 mmHg - 375 mmHg
EC REP
Federal law restricts this device to sale by or on the order of a physician (USA audiences only)
Year of manufacture
Storage humidity range: 5% to 95%
Storage temperature range: +70˚C to -40˚C Storage altitude range: +1600hPa to +500hPa
Keep dry
Fragile/breakable, handle with care
Indicates wireless Radio signal (wireless radio model only)
EU authorized representative
Defi brillation Proof Type BF
Caution
Electrical Testing Laboratory certification
Rad-87 Pulse CO-Oximeter Operator’s Manual 2-7
2
system description
LCD DISPLAY
The LCD Display shows radio communication information when radio communication is active (wireless radio model only). It also shows system information. All Rad-87 models are equipped with a LCD display which is located on the top panel of a horizontal model, or on the left side panel of a vertical model.
The LCD Display illuminates upon start up and displays the installed parameter/measurement's low and high alarm limits. Once the Rad-87 completes system initiation, the display light turns off.
As the front panel buttons are pressed, each menu selection is shown on the LCD Display.
When Rad-87 actively communicates with another system using the radio feature, the LCD Display shows the following:
Patient SafetyNet: The LCD Display shows the information sent from the Patient SafetyNet to the
Rad-87.
Philips VueLink: The LCD Display shows "VueLink Conn" and "SpO2 & PR Al On" or "SpO2 & PR Al
Off".
NOTE: When the Rad-87 is interfaced to the Philips VueLink and the LCD Display shows
"SpO2 & PR AL On", SpO2 and BPM audible alarms are active at the device and patient monitor. When the LCD Display shows "SpO2 & PR AL Off", SpO2 and BPM audible alarms are inactive at the device but active at the patient monitor.
Additionally, if the Display Button is pressed down for 5 seconds, the LCD Display shows the following settings three times and then returns to the default screen. The display cycle can be interrupted by pressing any button except for the Sensitivity or the Alarm Silence Buttons.
Label: System Settings
Monitoring Mode: Normal, Sleep or Home
Installed parameter/measurement's low and high alarm limits
Audible Alarm
Alarm Volume
Alarm Silence
Alarm Delay
Rapid Desat
Sensitivity
Averaging Time
2-8 Rad-87 Pulse CO-Oximeter Operator’s Manual
setup
Introduction
Before the Rad-87 Pulse CO-Oximeter can be used in a clinical setting, it needs to be inspected, properly setup and the batteries need to be fully charged.
Unpacking and inspection
Remove the instrument from the shipping carton and examine it for signs of shipping damage. Check all materials against the packing list. Save all packing materials, invoice and bill of lading. These may be required to process a claim with the carrier.
If anything is missing or damaged, contact the Technical Service Department. The contact address and phone numbers are listed in Section 9, Service and Repair.
Preparation for monitoring
The following sections of the manual describe the preparation, set-up and initial installation of the Rad-87 Pulse CO-Oximeter.
RAD-
87 POWER REQUIREMENTS
Always use a hospital grade, AC power cable to connect the Rad-87 to an AC power source.
CAUTION: DO NOT CONNECT THE RAD-87 TO AN AC OUTLET CONTROLLED BY A
SWITCH.
Verify the AC power voltage and frequency before use. Verify that the power source can provide adequate power rating as indicated on the rear panel of the Rad-87.
The Rad-87 is designed to operate on 100 to 240VAC, 47-83 Hz. The device is rated at 15 VA max.
Connect a hospital grade power cable to the power entry module of the Rad-87 device(IEC-320 connector type at the device). Connect the power cable to an AC power source. Ensure that the device is adequately powered by verifying that the AC power indicator on the Rad-87 is illuminated.
3
CAUTION:
CONNECT THE RAD-87 ONLY TO A HOSPITAL-GRADE RECEPTACLE (FOR HOSPITAL USE).
DO NOT UNDER ANY CIRCUMSTANCES REMOVE THE GROUNDING CONDUCTOR FROM THE
POWER PLUG.
DO NOT USE EXTENSION CORDS OR ADAPTERS OF ANY TYPE. THE POWER CORD AND PLUG
MUST BE INTACT AND UNDAMAGED.
USE THE POWER CORD AS THE MEANS TO DISCONNECT THE DEVICE FROM THE MAINS
POWER SUPPLY.
IF THERE IS ANY DOUBT ABOUT THE INTEGRITY OF THE PROTECTIVE EARTH
CONDUCTOR ARRANGEMENT, OPERATE THE RAD-87 ON INTERNAL BATTERY POWER
UNTIL THE AC POWER SUPPLY PROTECTIVE CONDUCTOR IS FULLY FUNCTIONAL.
TO ENSURE PATIENT ELECTRICAL ISOLATION, CONNECT ONLY TO OTHER
EQUIPMENT WITH ELECTRICALLY ISOLATED CIRCUITS.
DO NOT CONNECT TO AN ELECTRICAL OUTLET CONTROLLED BY A WALL SWITCH OR
DIMMER.
Rad-87 Pulse CO-Oximeter Operator’s Manual 3-1
3
Configure the Rad-87 for your regional power line frequency (50 or 60 hz) if needed. Default is 60 hz (standard for the United States). See Section 4,
Operation, Setup menu Level 3, Line
Frequency.
CAUTION:
HE DEVICE MUST BE CONFIGURED TO MATCH YOUR LOCAL POWER LINE FREQUENCY TO ALLOW FOR THE CANCELLATION OF NOISE INTRODUCED BY FLUORESCENT LIGHTS AND OTHER SOURCES.
CAUTION:
THE BATTERY SHOULD BE ADEQUATELY CHARGED TO ENSURE BACKUP
POWER IN CASE OF AC POWER DISRUPTION.
INITIAL BATTERY CHARGING
Before use, the Rad-
To charge the internal battery, connect the AC power cord to an AC outlet and to the Power Entry Module located on the back of the Rad­remain illuminated while the battery is charging. The Battery Charge Level Indicator will not be illuminated unless the device is operating on battery power. Once the battery is fully charged, the device has up to 4 hours of battery life. See Section 7-2, Specifications.
INITIAL INSTALLATION
Place the Rad-87 on a stable hard flat surface near the patient. Always place the Rad­dry surface. Maintain a minimum of 1 inch (2.54 cm) free space around the device. Make sure that Rad-87 loudspeaker is not covered to avoid a muffled alarm sound.
The Rad-
87 should not be operated outside the following environmental conditions:
TEMPERATURE
87 battery needs to be fully charged.
87. The AC Power Indicator illuminates. The AC Power Indicator will
OPERATING ENVIRONMENTAL CONDITIONS
+5°C to +40°C, +41°F to +104°F
setup
87 on a
HUMIDITY
OPERATING ALTITUDE
Configure the Rad-87 for your regional power line frequency (50 or 60 hz) if needed. Default is 60 hz (standard for the United States). See Section 4,
Frequency.
CAUTION: TT
CAUTION:
HE DEVICE MUST BE CONFIGURED TO MATCH YOUR LOCAL POWER LINE FREQUENCY TO ALLOW FOR THE CANCELLATION OF NOISE INTRODUCED BY FLUORESCENT LIGHTS AND OTHER SOURCES.
THE BATTERY SHOULD BE ADEQUATELY CHARGED TO ENSURE BACKUP
POWER IN CASE OF AC POWER DISRUPTION.
5% to 95%, non-condensing
500 mbar to 1060 mbar pressure
-1000 ft to 18,000 ft (-304 m to 5,486 m)
Operation, Setup menu Level 3, Line
3-2 Rad-87 Pulse CO-Oximeter Operator’s Manual
operation
Introduction
To operate the Rad-87 system effectively, the device must be set up correctly and the operator must:
Know how the Rad-
Be familiar with its controls, components and operation.
Understand its status and alarm messages (see Section 5, Alarm and Messages and
Section 6, Troubleshooting).
Basic operation
GENERAL SETUP AND USE
1. Inspect the Rad-
2. Connect a patient cable or a direct connect sensor to the Rad­connection and the cable is not twisted, sliced or frayed.
3. If utilizing a patient cable, select a sensor that is compatible with the Rad­before connecting it to the patient cable. See section 8, Sensors and Patient Cables. If using a single patient adhesive or disposable sensor, check that the emitter (red light) and the detector are properly aligned. Remove any substances that may interfere with the transmission of light between the sensor’s light source and detector.
4. Refer to the Directions for Use of the sensor before attaching the sensor to the patient.
5. Attach the sensor to the Patient. With a Masimo sensor, connect the sensor to the patient cable with the logos lining up; make sure it is a firm connection.
6. Press the Power button to turn the Rad-
7. Verify all front-panel indicators momentarily illuminate and a tone is heard.
8. Verify the front-panel display is free of alarm and system failure messages (see Section 5, Alarms and Messages).
NOTE: The number "0" scrolls across the screen as the system calibrates and obtains patient
data (approximately 20 seconds).
9. Verify the LED and the LCD displays shows the following (see Setup Menu Level 1;
Parameter/Measurement Alarm Limits - Screen 1, Parameter/Measurement Alarm Limits ­Screen 2, Parameter/Measurement Alarm Limits - Screen 3, and Setup Menu Level 3; Set Mode located in this chapter):
Mode setting: Standard (Std) or Sleep (SLP) or Home (Hnn)
SpO
Pulse Rate Low Alarm Limit and Pulse Rate High Alarm Limit,
SpCO Low Alarm Limit and SpCO High Alarm Limit,
SpMet Low Alarm Limit and SpMet High Alarm Limit,
SpHb Low Alarm Limit and SpHb High Alarm Limit,
PI Low Alarm Limit and PI High Alarm Limit,
PVI Low Alarm Limit and PVI High Alarm Limit.
Measurement Alarm Limits - Screen 2, Parameter/Measurement Alarm Limits - Screen 3 in this chapter).
NOTE: The number "0" scrolls across the screen as the system calibrates and obtains patient
2
data (approximately 20 seconds).
87 derives its readings (see Section 1).
87 case for damage.
Low Alarm Limit and SpO
High Alarm Limit,
2
87 device. Make sure it is a firm
87 and the patient
87 on.
4
Rad-87 Pulse CO-Oximeter Operator’s Manual 4-1
4
operation
Basic operation continued
10. On the LED and the LCD displays, verify the alarm limit settings (see Setup Menu Level 1,
Parameter/Measurement Alarm Limits - Screen 1, Parameter/Measurement Alarm Limits ­Screen 3 in this chapter).
NOTE: The number "0" scrolls across the screen as the system calibrates and obtains patient
data (approximately 20 seconds).
11. Verify that the patient alarms are functional by setting the high and low alarm limits beyond the patient readings. (see Setup Menu Level 1, Parameter/Measurement Alarm Limits -
Screen 1, Parameter/Measurement Alarm Limits - Screen 2, Parameter/Measurement Alarm Limits - Screen 3 in this chapter).
An alarm tone sounds.
The Alarm Bell flashes red for high priority alarms.
The System Status Light flashes red for high priority alarms, flashes yellow for medium
priority alarms and is solid yellow for low priority alarms.
The number value and parameter/measurement label for the violated alarm limit will
flash on the LED display.
12. Verify the sensor alarms are functional.
Remove the sensor from the sensor site.
The alarm tone sounds.
The Alarm Bell flashes red.
The System Status Light flashes red.
The display shows "SEN OFF" message.
Disconnect the sensor from the patient cable or Rad-87.
The alarm tone sounds.
The Alarm Bell flashes red.
The System Status Light flashes red.
The display shows "NO SEN" message.
NOTE: "NO SEN" or "SEN OFF" conditions will only generate a high priority alarm if the Rad-87 is
actively monitoring a patient when the sensor is disconnected.
13. Verify that the audible alarm can be silenced when a parameter/measurement alarm is exceeded.
Create an alarm condition by lowering the high alarm limit for the pulse rate so that it
is lower than the patient value.
Press the Alarm Silence button.
The alarm tone ceases for 120 seconds (default).
The Alarm Bell flashes red for a high pulse rate (high priority alarm).
The System Status Light flashes red.
14. To begin patient monitoring:
Adjust the alarm limits.
Adjust the alarm volume.
■ Adjust the pulse beep volume.
4-2 Rad-87 Pulse CO-Oximeter Operator’s Manual
operation
Basic operation continued
15. Verify the sensor is applied correctly and that the measured data is appropriate, see Section 4, Successful Monitoring.
16. Monitor the patient.
17. After monitoring is complete, remove the sensor from the patient and store or dispose of the sensor according to local laws. See the Directions for Use of the sensor.
18. Press and hold the Power Button for 2 seconds to turn the Rad-87 off [3 seconds in the Home Mode].
DEFAULT SETTINGS
The Rad-87 Pulse CO-Oximeter stores two types of default values that the device automatically retains after a power cycle.
1. Factory defaults set by Masimo.
2. Default settings that can be changed by the user which will be remembered after a power cycle.
FACTORY DEFAULT AND USER CONFIGURABLE SETTINGS
OPTION FACTORY DEFAULTS USER CONFIGURABLE DEFAULTS
high alarm limit "---" Off 2 to 99%
SpO
2
SpO
low alarm limit 90% 1 to 98%
2
Pulse rate high alarm limit 140 BPM 35 to 235 BPM
Pulse rate low alarm limit 50 BPM 30 to 230 BPM
SpCO high alarm limit 10
SpCO low alarm limit "---" Off "---", then 1 to 97
SpMet high alarm limit 3 1 to 99.5, then "---"
SpMet low alarm limit Off "---", then .1 to 99
SpHb high alarm limit 16 2 to 24.5, then "---"
SpHb low alarm limit 7 "---", then 1 to 24
PI high alarm limit "---" Off 0.04 to 19, then "---"
PI low alarm limit "---" Off "---", then 0.03 to 18
PVI high alarm limit "---" Off 2 to 99, then "---"
PVI low alarm limit "---" Off "---", then 1 to 98
Sensitivity APOD
Display brightness Level 2 Levels 1 thru 4
Pulse tone volume Level 2 Off, Levels 1 thru 3
Alarm Silence Time 120 seconds 30, 60, 90, or 120 seconds
Alarm Volume Level 3, 70 db min Levels 1 thru 4, 87 db max
Monitoring Mode Standard (Normal) Standard, Sleep, Home
Audible Alarm Off Alarms active (On) "On/Off or muted with reminder"
Alarm Delay 5 sec 0, 5, 10, or 15 seconds
Rapid Desat Alarm 5% 5, 10, Off
Serial out ASCII 2 Philips/ASCII 1/ASCII 2
Interface Alarm Alarm On Alarm, Off/On
2 to 98, then "---"
Max/Normal/APOD
NOTE: MAX sensitivity will default to APOD after a power cycle.
4
Rad-87 Pulse CO-Oximeter Operator’s Manual 4-3
4
operation
FACTORY DEFAULT AND USER CONFIGURABLE SETTINGS (CONTINUED)
Nurse Call Type Alarm
Nurse Call Polarity Normal Normal/Invert Line Frequency 60 Hz 60 Hz, 50 Hz
Averaging time 8 sec 2, 4, 8, 10, 12, 14, 16 SmartTone Off On/Off
FastSat Off Setting is retained after power cycle.
Device Profile settings N/A Light blue, purple, pink, orange or green
LCD Language English
PVI Bar On On/Off
Alarm and Low Signal IQ/ Low Signal IQ/ Alarm
English (Default) French, German, Italian, Spanish, Swedish, Dutch, Danish, Portuguese
Successful Monitoring
The following general points will aid in ensuring monitoring success.
NOTE: See Safety Information, Warnings, Cautions and Notes for additional information.
Place the sensor on a site that has sufficient perfusion and provides proper alignment of the
LED’s and detector.
Place the sensor on a site that has unrestricted blood flow.
Do not secure a sensor with tape.
Do not select a site near potential electrical interference (electro-surgical device, for example).
Read the sensor Directions for Use for proper sensor application.
MASIMO PULSE CO-OXIMETRY SENSORS
Before use, carefully read the Masimo sensor Directions for Use.
Use only Masimo sensors for pulse oximetry or pulse CO-Oximetry measurements.
Tissue damage can be caused by incorrect application or use of a sensor, for example by wrapping the sensor too tightly. Inspect the sensor site as directed in the sensor Directions for Use to ensure skin integrity and correct positioning and adhesion of the sensor.
CAUTIONS
DO NOT USE DAMAGED SENSORS. DO NOT USE A SENSOR WITH EXPOSED
OPTICAL OR ELECTRICAL COMPONENTS. DO NOT IMMERSE THE SENSOR IN WATER, SOLVENTS, OR CLEANING SOLUTIONS (THE SENSORS AND CONNECTORS ARE NOT WATERPROOF). DO NOT STERILIZE BY IRRADIATION, STEAM, AUTOCLAVE OR ETHYLENE OXIDE UNLESS OTHERWISE INDICATED IN THE SENSOR DIRECTIONS FOR USE. SEE THE CLEANING INSTRUCTIONS IN THE DIRECTIONS FOR USE FOR ALL MASIMO REUSABLE SENSORS.
DO NOT USE DAMAGED PATIENT CABLES. DO NOT IMMERSE THE PATIENT CABLES IN
WATER, SOLVENTS, OR CLEANING SOLUTIONS (THE PATIENT CABLE CONNECTORS ARE NOT WATERPROOF). DO NOT STERILIZE BY IRRADIATION, STEAM, AUTOCLAVE OR ETHYLENE OXIDE.
DO NOT ATTEMPT TO REPROCESS, RECONDITION OR RECYCLE MASIMO SENSORS
OR PATIENT CABLES AS THESE PROCESSES MAY DAMAGE THE ELECTRICAL COMPONENTS, POTENTIALLY LEADING TO PATIENT HARM.
4-4 Rad-87 Pulse CO-Oximeter Operator’s Manual
operation
4
NUMERIC DISPLAY - SpO
Stability of the SpO2 readings may be a good indicator of signal validity. Although stability is a relative term, experience will provide confidence in changes that are artifactual or physiological and the speed, timing, and behavior of each. The stability of the readings over time is affected by the averaging mode being used. The longer the averaging time, the more stable the readings tend to become. This is due to a dampened response as the signal is averaged over a longer period of time than during shorter averaging times. However, longer averaging times delay the response of the device and reduce the measured variations of SpO2 Inaccurate measurements may be caused by:
Elevated levels of Carboxyhemoglobin.
Elevated levels of Methemoglobin.
Severe anemia.
Elevated Total Bilirubin levels.
Low arterial perfusion.
Motion artifact.
NUMERIC DISPLAY - PULSE RATE
The Pulse Rate displayed on the Rad­displayed on ECG monitors due to differences in averaging times. There may also be a discrepancy between cardiac electrical activity and peripheral arterial pulsation. Significant differences may indicate a problem with the signal quality due to physiological changes in the patient or one of the instruments or application of the sensor or patient cable. The pulsations from intra-aortic balloon support can cause the pulse rate displayed on the Rad-87 to be significantly different than the ECG heart rate.
NUMERIC DISPLAY - SpCO
A stable SpCO reading is associated with correct sensor placement, small physiological changes during the measurement and acceptable levels of arterial perfusion in the patient’s measurement site. Physiological changes at the measurement site are mainly caused by fluctuations in the arterial oxygen saturation, blood concentration and perfusion. Inaccurate measurements may be caused by:
Elevated levels of methemoglobin.
Intravascular dyes such as indocyanine green or methylene blue.
Abnormal hemoglobin levels.
Low arterial perfusion.
Low arterial oxygen saturation levels.
Elevated Total Bilirubin levels.
Motion artifact.
NUMERIC DISPLAY - SpMet
A stable SpMet reading is associated with correct sensor placement, small physiological changes during the measurement and acceptable levels of arterial perfusion in the patient’s measurement site. Physiological changes at the measurement site are mainly caused by fluctuations in the arterial oxygen saturation, blood concentration and perfusion. Inaccurate measurements may be caused by:
Intravascular dyes such as indocyanine green or methylene blue.
Low arterial perfusion.
Low arterial oxygen saturation levels.
Elevated Total Bilirubin levels.
Motion artifact.
2
87 Pulse CO-Oximeter may differ slightly from the heart rate
Rad-87 Pulse CO-Oximeter Operator’s Manual 4-5
4
NUMERIC DISPLAY - SpHb
A stable SpHb reading is associated with correct sensor placement, small physiological changes during the measurement and acceptable levels of arterial perfusion at the measurement site. Physiological changes at the measurement site are mainly caused by fluctuations in the arterial
oxygen saturation, blood concentration and perfusion. Inaccurate measurements may be caused by:
Intravascular dyes such as indocyanine green or methylene blue.
Low arterial perfusion.
Low arterial oxygen saturation levels.
Elevated Total Bilirubin levels.
Motion artifact.
NUMERIC DISPLAY - SpOC
A stable SpOC reading is associated with stable readings for both SpO2 and SpHb which comes with correct sensor placement, small physiological changes during the measurement and acceptable levels of arterial perfusion at the measurement site. Physiological changes at the measurement site are mainly caused by fluctuations in the oxygen saturation, blood concentration and perfusion.
Inaccurate measurements may be caused by:
Intravascular dyes such as indocyanine green or methylene blue.
Low arterial perfusion.
Low arterial oxygen saturation levels.
Elevated Total Bilirubin levels.
Motion artifact.
Elevated levels of carboxyhemoglobin.
Elevated levels of methemoglobin.
Severe anemia may cause erroneous SpOC readings.
NUMERIC DISPLAY - (PI)
The perfusion index (PI) display and bar graph indicator provide a relative numeric indication of the pulse strength at the monitoring site. It is a calculated percentage between the pulsatile signal and non-pulsatile signal of arterial blood moving through the site. PI may be used to find the best perfused site and to monitor physiological changes in the patient. It displays an operating range of
0.02 percent to 20.00 percent. A percentage greater than 1.00 percent is desired. Extreme changes in the display number are due to motion artifact and changes in physiology and blood flow.
PLETH VARIABILITY INDEX - (PVI)
The pleth variability index (PVI) is a measure of the dynamic changes in the perfusion index (PI) that occur during the respiratory cycle. The calculation is accomplished by measuring changes in PI over a time interval where one or more complete respiratory cycles have occurred. PVI is displayed as a percentage (0-100%).
The PVI bar graph is a representation of the calculated value. The lower the height of the bar, the less variability there is in the PI over a respiratory cycle. Press the Display key to toggle to the PVI numeric measurement.
operation
4-6 Rad-87 Pulse CO-Oximeter Operator’s Manual
operation
LOW PERFUSION
The Rad-87 indicates perfusion on a 10-bar LED indicator. The lower two segments of the bar will turn red when the amplitude of the arterial pulsations is very low (low perfusion).
It has been suggested that at extremely low perfusion levels, pulse oximeters can measure peripheral saturation, which may differ from central arterial saturation may result from the metabolic demands of other tissues extracting oxygen proximal to the monitoring site under conditions of sustained peripheral hypoperfusion. (This may occur even with a pulse rate that correlates with the ECG heart rate.)
CAUTION: IF THE LOW PERFUSION MESSAGE IS FREQUENTLY DISPLAYED, FIND A
BETTER-PERFUSED MONITORING SITE. IN THE INTERIM, ASSESS THE PATIENT AND, IF INDICATED, VERIFY OXYGENATION STATUS THROUGH OTHER MEANS.
3
Severinghaus JW, Spellman MJ. Pulse Oximeter Failure Thresholds in Hypotension and Vasoconstriction.
Anesthesiology 1990; 73:532-537
SIGNAL IQ - (SIQ)
The Rad­when the displayed SpO indicator displayed on the Rad-
87 Pulse CO-Oximeter display provides a visual indicator of signal quality and an alert
values are not based on adequate signal quality. The signal quality
2
87 is called the Signal IQ. The Signal IQ can be used to identify the
occurrence of a patient’s pulse and the associated signal quality of the measurement.
The Signal IQ is represented as a bar indicator where the height of the bar coincides with the strength of an arterial pulsation. Even with a plethysmographic waveform obscured by artifact, the Rad-
87 locates the arterial pulsation. The pulse tone (when enabled) coincides with the rise of the
Signal IQ Index.
The height of the Signal IQ Index indicates the quality of the measured signal. A high Signal IQ Index indicates that the SpO Index indicates that the SpO signal quality is very low the accuracy of the SpO Signal IQ is low and the bar turns red, proceed with caution and do the following:
measurement is based on a good quality signal. A small Signal IQ
2
measurement is based on data with low signal quality. When the
2
measurement may be compromised. When the
2
Assess the patient.
Check the sensor and ensure proper sensor application. The sensor must be well secured
to the site for the Rad-87 to maintain accurate readings. Also, misalignment of the sensor’s emitter and detector can result in smaller signals.
Determine if an extreme change in the patient’s physiology and blood flow at the monitoring site occurred, (e.g. an inflated blood pressure cuff, a squeezing motion, sampling of an arterial blood specimen from the hand containing the pulse oximetry sensor, severe hypotension, peripheral vasoconstriction in response to hypothermia, medications, or an episode of Raynaud’s syndrome).
With neonates or infants, check that the peripheral blood flow to the sensor site is not interrupted. Interruption, for example, may occur while lifting or crossing their legs, during a diaper change.
After performing the above, if the Low Signal IQ message is displayed frequently or continuously obtaining an arterial blood specimen for CO-Oximetry analysis may be considered to verify the arterial oxygen saturation value.
3
. This “localized hypoxemia”
4
Rad-87 Pulse CO-Oximeter Operator’s Manual 4-7
4
SENSOR PLACEMENT
If the SpO
Make sure the emitter and detector are aligned directly opposite each other.
Select a site where the distance between the emitter and detector is minimized.
Wipe the sensor site with a 70% isopropyl alcohol pad or rubefacient cream (10-30% methyl
If possible, remove electrical noise sources such as electro-surgical devices or other electrical/
If artificial nails or excessive fingernail polish are present, select another site or remove the polish/
If possible, ensure that the sensor is placed in a location with low ambient light. Although the
CAUTION: IF ANY MEASUREMENT SEEMS QUESTIONABLE, FIRST CHECK THE
SENSITIVITY
The Rad-87 Pulse CO-Oximeter is equipped with the clinician to change the sensitivity settings of the device to meet the increased demands of the patient’s physiological condition or enable it to work during periods of low perfusion and/or motion.
They are as follows:
Normal Sensitivity (NORM) – This is the recommended mode for patients that are
Adaptive Probe Off Detection (APOD) – This is the recommended start-up monitoring mode
Maximum Sensitivity (MAX) - This mode is recommended for patients with low perfusion or
CAUTION: WHEN USING THE MAXIMUM SENSITIVITY SETTING, THE PERFORMANCE OF
SpCO, SpMet or SpHb readings are questionable or unavailable, do the following:
2,
salicylate and 2-10% menthol) for 20-30 seconds to increase perfusion. However, strong vasodilator creams, such as nitroglycerin paste, are not recommended.
electronic equipment. If these solutions are not possible, operate the Rad-87 on battery power, or try plugging the device into a different electrical outlet.
artificial nails.
Rad-
87 with integrated Masimo Rainbow SET technology has significant immunity to ambient
light, excessive ambient light may cause readings to be incorrect.
PATIENT’S VITAL SIGNS BY ALTERNATE MEANS AND THEN CHECK THE PULSE CO-OXIMETER FOR PROPER FUNCTIONING.
3 different sensitivity modes. Each mode allows
experiencing some compromise in blood flow or perfusion. It is advisable for care areas where patients are observed frequently, such as an intensive care unit (ICU).
for most patients with acceptable perfusion or where a more robust sensor off detection is desired. It is the suggested mode for care areas where patients are not visually monitored continuously. This mode delivers enhanced protection against erroneous pulse rate and arterial oxygen saturation readings when a sensor becomes inadvertently detached from a
patient.
when the low perfusion or low signal quality message is displayed on the screen in APOD or normal sensitivity mode. This mode is not recommended for care areas where patients are not monitored visually, such as general wards. It is designed to interpret and display data at the measuring site when the signal may be weak due to decreased perfusion. When a sensor becomes detached from a patient, it will have compromised protection against erroneous pulse rate and arterial saturation readings. Also, after a power off and on cycle, the sensitivity will change from the MAX to the factory default or user configured default setting of APOD or NORM.
THE SENSOR OFF DETECTION MAY BE COMPROMISED. IF THE DEVICE IS IN THIS SETTING AND THE SENSOR BECOMES DISLODGED FROM THE PATIENT, THE POTENTIAL FOR FALSE READINGS MAY OCCUR DUE TO ENVIRONMENTAL ‘NOISE’ SUCH AS LIGHT, VIBRATION AND EXCESSIVE AIR MOVEMENT.
operation
4-8 Rad-87 Pulse CO-Oximeter Operator’s Manual
operation
CAUTION:
THE BATTERY SHOULD BE ADEQUATELY CHARGED TO ENSURE BACKUP
POWER IN CASE OF AC POWER DISRUPTION.
LOW BATTERY AUDIBLE ALARM
If a low battery condition occurs the audible alarm can be silenced until power cycle by pressing the Alarm Silence Button. Refer to Setup Menu Level 2 in this section to change setting.
If a low battery condition occurs while not monitoring a patient, a low priority audible alarm will sound and can be silenced by pressing the Alarm Silence Button. The audible alarm is silenced until the power is cycled or patient monitoring begins.
While audible alarms are silenced, the first Battery Level Indicator bar to the left flashes green, and the System Status Light flashes yellow to provide a visual alert for the user.
When a low battery condition occurs, immediately discontinue patient monitoring and plug the Rad-87 into AC power. The AC Power Indicator on the Rad-87 illuminates and remains illuminated while the battery is charging, however, the Battery Charge Level Indicator does not illuminate. Once the battery is fully charged all Battery Charge Level Indicators illuminate green when unplugged.
During normal patient monitoring, the Battery Charge Bars (Battery Charge Level Indicator) illuminate green from left to right to indicate the approximate amount of battery charge when unplugged.
CAUTION:
POWER IN CASE OF AC POWER DISRUPTION.
Normal patient monitoring
When all optional parameters/measurements are installed and during normal operation, the Rad-87 display shows Screen minute (BPM) and total hemoglobin (SpHb g/dl)*. By pressing the Display Button once, the display changes to show Screen (PI) and pleth variability index (PVI)*. Pressing the Display Button again changes the display to show Screen additional press of the Display Button returns the display to Screen
*Optional parameters/measurements: SpCO, SpMet, SpHb, SPOC, PVI
Display Screens showing Parameters/Measurements -Default Locations
THE BATTERY SHOULD BE ADEQUATELY CHARGED TO ENSURE BACKUP
1 containing arterial oxygen saturation (%SpO
2 containing total arterial oxygen content (SpOC
3 containing methemoglobin (%SpMet)* and carboxyhemoglobin (%SpCO)*. An
) and pulse rate in beats per
2
ml/dl)* perfusion index
1, the home display screen.
4
1.60
.
76
10.1 3
Screen 1
Home display screen
Rad-87 Pulse CO-Oximeter Operator’s Manual 4-9
Screen
13
PVI
2
20
Screen 3
4
Normal patient monitoring continued
If SpMet and SpCO are/or not installed, the display will only move from Screen 1 to Screen 3 when the Display Button is pressed; Screen 2 will not show. By pressing the Display Button again, the display will move from Screen 3 to Screen 1.
Display Screens with All Parameters and Measurements
operation
1.60
76
10.1
Screen 1
Home display screen
Display Screens when SpMet, SpCO and PVI Are Not Installed
13
10.1
Screen
2
1376
20
%SpCO
Screen 3
No Screen 3
10.1
Screen 1
Home display screen
Display Screens with SpHb and PVI Installed and SpMet and SpCO Not Installed
Screen
1376
2
No Screen
3
3
10.1
Screen 1
Home display screen
4-10 Rad-87 Pulse CO-Oximeter Operator’s Manual
3
Screen 2
operation
PARAMETER/MEASUREMENT SELECTION
The bottom fi eld of any display screen may be confi gured to show SpHb or PVI (see Parameter Select menus). Once SpHb or PVI is confi gured to show on a screen different from the default location, the alarm limit menu for SpHb or PVI can be accessed from the new screen by pressing the Alarm Limits Button.
Display Screens Showing PVI on Screen 1
1.60
4
3
Screen 1
Home display screen
Display Screens Showing SpHb on Screen 1 and 2
76
10.1
Screen 1
Home display screen
10.1
Screen
13
10.1
Screen
2
2
201376
Screen 3
1.60 20
%SpCO
3
Screen 3
Rad-87 Pulse CO-Oximeter Operator’s Manual 4-11
4
SETUP MENU
This section gives an overview of the Rad-87 menu selections available. To access the menu levels and navigate through the menu selections, use the front panel buttons, Enter Button and Up/Down Buttons as indicated in the following sections. Sub-sections describe each menu item in more detail. The Rad-
MENU NAVIGATION
The Rad-87 set-up and configuration options are accessed through the menu system. Three levels of menus are available to the user. Once a menu level is accessed, a front panel button (Level 1 only) or the Enter Button (Level cycling through the options. The Up and Down Buttons are used to adjust values within each option. The parameter/measurement value is set when the Enter Button is pressed. Pressing the Display Button exits the menus and returns the device to Screen 1.
When accessing the Rad-87 menu, each selection will be communicated visually on the LED (front of device) and LCD (top of device) displays, simultaneously.
NOTE: The Rad-
SETUP MENU LEVEL
Setup Menu Level 1 contains the parameters/measurements and settings that are adjusted most often for patient monitoring; alarm limits, display brightness, and sensitivity settings. Pressing the Display Button while viewing alarm limit settings allows the user to exit the Setup Menu and return to Screen 1.
87 has options that allow user configuration to accommodate specific needs.
2 and 3) is used to move from one option to the next allowing repeated
87 will automatically ‘time out’ of the setup menu after 10 seconds with no button
presses.
1
operation
4-12 Rad-87 Pulse CO-Oximeter Operator’s Manual
operation
SETUP MENU LEVEL 1 (CONTINUED)
PARAMETER/MEASUREMENT ALARM LIMITS - SCREEN 1
To access alarm limits for parameters/measurements contained on Screen 1, press the Alarm Limits Button to access the Alarm Limits menu for Screen 1 parameters/measurements.
BUTTONS SETTINGS
Press once
%SpO2 LO
Press 2x
Use Up or Down Buttons to adjust the
value to the desired setting
AND
press the Alarm Limits Button or Enter
Button to accept the setting and move
to the next option. Once the last option
is accessed an additional press of the
Alarm Limits Button or Enter Button will
return the device to Screen 1.
OR
press the Display Button to exit at any
time and return to Screen 1.
Use the Alarm
Limits Button to
access the alarm
limits options and
move between
options.
%SpO2 HI
Press 3x
Pulse rate (BPM) LO
Press 4x
Pulse rate (BPM) HI
Press 5x
SpHb g/dl LO
4
Press 6x
SpHb g/dl HI
NOTE: User default settings can be changed for specific patient environments.
Rad-87 Pulse CO-Oximeter Operator’s Manual 4-13
4
operation
SETUP MENU LEVEL 1 (CONTINUED)
PARAMETER/MEASUREMENT ALARM LIMITS - SCREEN 2
To access alarm limits for parameters/measurements contained on Screen 2, press the Display Button to move from Screen 1 to Screen 2. From Screen 2, press the Alarm Limits Button to access
the Alarm Limits menu for Screen 2 parameters/measurements.
BUTTONS SETTINGS
Press once
Use the Alarm Limits
Button to access the
alarm limits options and
move between options.
PI LO
Press 2x
PI HI
Press 3x
PVI LO
Press 4x
PVI HI
Use Up or Down Buttons to adjust
the value to the desired setting
AND
press the Alarm Limits Button or
Enter Button to accept the setting
and move to the next option. Once
the last option is accessed an
additional press of the Alarm Limits
Button will return the device to
Screen 2.
OR
press the Display Button to exit at
any time and return to Screen 2.
NOTE: User default settings can be changed for specific patient environments.
4-14 Rad-87 Pulse CO-Oximeter Operator’s Manual
operation
PARAMETER/MEASUREMENT ALARM LIMITS - SCREEN 3
To access alarm limits for parameters/measurements contained on Screen 3, press the Display Button two times to move from Screen 1 to Screen 3. Press the Alarm Limits Button from Screen 3 to access the Alarm Limits menu for Screen 3 parameters/measurements.
BUTTONS SETTINGS
Press once
%SpMet LO
Use the Alarm
Limits Button to
access the alarm
limits options and
move between
options.
Press 2x
%SpMet HI
Press 3x
%SpCO LO
Press 4x
%SpCO HI
NOTE: User default settings can be changed for specific patient environments.
Use Up or Down Buttons to adjust the
value to the desired setting
AND
press the Alarm Limits Button or Enter
Button to accept the setting and move
to the next option. Once the High Pulse
rate limit is accessed an additional
press of the Alarm Limits Button or
Enter Button will return the device to
Screen 3.
OR
press the Display Button to exit at any
time and return to Screen 3.
4
LED BRIGHTNESS
The Display screen and all active LED indicators are effected while adjusting this setting.
BUTTONS SETTINGS
Use the Brightness
Button to access
the LED brightness
options and move
between options.
Press once
Press 2x
Press 3x
Press 4x
Default
Level 2
Level 3
Level 4
Level 1
Use the Brightness Button to move
between menu options and the Enter
Button to accept the setting and return
to the home display screen.
NOTE: User default settings can be changed for specific patient environments.
Rad-87 Pulse CO-Oximeter Operator’s Manual 4-15
4
SETUP MENU LEVEL 1 (CONTINUED)
SENSITIVITY
BUTTONS SETTING
Use the Sensitivity Button
to access the sensitivity
options and move between
options.
Press once
Press 2x
Default
APOD
NORM
operation
Use the Sensitivity Button to
move between menu options
and accept the setting.
Press 3x
MAX
(The MAX Indicator
flashes in this
mode.)
SETUP MENU LEVEL 2
Level 2 menu contains parameters and settings that are not changed as frequently as Level
1. These include alarm volume, alarm silence, alarm delay, clear trend and button volume parameters.
ALARM VOLUME
BUTTONS SETTINGS
Default
Level 3
Use the Enter Button
to access the Alarm
Volume menu and to
move between Level
2 menus.
Level 4
Level 1
Level 2
Use Up or Down Button to move between
settings and the Enter Button to accept
the setting and move to the next menu
screen.
OR
press the Display Button to exit without
saving the new setting and to return to
the home display screen.
4-16 Rad-87 Pulse CO-Oximeter Operator’s Manual
operation
SETUP MENU LEVEL 2 (CONTINUED)
ALARM SILENCE
BUTTONS SETTINGS
Default
120 seconds
Press the Enter Button
again to move to the
next menu.
2x
90 seconds
60 seconds
30 seconds
ALARM DELAY
Alarm delay allows the user to adjust the time in which the audible status indicator will occur after an alarm condition has been initiated.
BUTTONS SETTINGS
Use Up or Down Button to move between
settings and the Enter Button to accept the
setting and move to the next menu screen.
OR
press the Display Button to exit without
saving the new setting and to return to the
home display screen.
4
Default
5 seconds
Press the Enter Button
again to move to the
next menu.
0 seconds
15 seconds
3x
10 seconds
Rad-87 Pulse CO-Oximeter Operator’s Manual 4-17
Use Up or Down Button to move between
settings and the Enter Button to accept the
setting and move to the next menu screen.
OR
press the Display Button to exit without
saving the new setting and to return to the
home display screen.
4
operation
SETUP MENU LEVEL 2 (CONTINUED)
CLEAR TREND
The Rad-87 only stores data in the trend memory while the device is turned on. Trend data saves to the memory until the memory is full or cleared by the user.
NOTE: It is recommended that you clear the trend prior to performing a new patient data
collection procedure.
BUTTON SETTING
Press the Enter Button
again to move to the
next menu.
4x
BUTTON VOLUME
BUTTON SETTING
Press the Enter Button
again to move to the
next menu.
5x
Default
NO
YES
(Clear trend)
Default
Level 2
Level 1
Off
Level 3
Use Up or Down Button to move
between settings and the Enter
Button to accept the setting and move
to the next menu screen.
OR
press the Display Button to exit
without saving the new setting and to
return to the home display screen.
Use Up or Down Button to move between
settings and the Enter Button to accept the
setting and move to the next menu screen.
OR
press the Display Button to exit without
saving the new setting and to return to the
home display screen.
FASTSAT
FastSat enables rapid tracking of arterial oxygen saturation changes. Arterial oxygen saturation data is averaged using pulse oximeter averaging algorithms to smooth the trend. When the Rad-87 is set to FastSat “On”, the averaging algorithm evaluates all the saturation values providing an averaged saturation value that is a better representation of the patient’s current oxygenation status.
With FastSat, the averaging time is dependent on the input signal.
4-18 Rad-87 Pulse CO-Oximeter Operator’s Manual
operation
SETUP MENU LEVEL 2 (CONTINUED)
FASTSAT
BUTTON SETTING
4
Use Up or Down Button to move
between settings and the Enter Button
to accept the setting and move to the
Alarm Volume menu.
OR
press the Display Button to exit without
saving the new setting and to return to
the home display screen.
Press the Enter Button
again to move to the
next menu.
6x
Default
Off
On
Trend setup and use
INTRODUCTION
The Rad-87 can store up to 72 hours of trend data captured at 2 second intervals. The trend data can then be transferred to a PC for evaluation.
A serial cable is required to connect the Rad-87 to a PC. Patient monitoring is not possible while trend memory is being transferred to a PC.
Trend data is stored in non-volatile memory, so it is not erased when the device is shut off.
A trend data download is initiated using the TrendCom utility which downloads the trend data and saves it to an ASCII text (.out) file with an output delimiter option.
NOTE: Rad-87 Serial Ouput must be set to ASCII 2 for successful download of trending data. Refer
to the Serial Output menu and settings located farther in this chapter.
TRENDCOM UTILITY INSTALLATION
Copy the TrendCom utility from the TrendCom CD onto a PC running MS-Windows.
TRENDCOM UTILITY OPERATION
1. Turn Rad-87 off if not already off.
2. Connect serial cable to Rad-87 and other end to a com port on the PC.
3. Turn the Rad-87 on.
4. Start the TrendCom utility on the PC.
5. Select Rad-87 from the first pull-down menu.
6. Select the appropriate com port number from the second pull-down menu, if necessary.
7. Select the Output Delimiter Option (Tab, Comma or Space).
8. Select the RETRIEVE TREND button on the TrendCom utility. Select the desired location and assign a file name for the trend file. Select SAVE.
Rad-87 Pulse CO-Oximeter Operator’s Manual 4-19
4
operation
Trend setup and use continued
9. The Rad-87 will display “dat out” while trend data is being transferred. A progress bar will advance to indicate the status of the download. Larger trend files will take longer to download. Transfer time is approximately 20 seconds per hour of trend data.
NOTE: During download of trend information, all normal Rad-87 functions are unavailable and the
keypad is locked, except for the power button.
10 When trend data transfer is complete, close TrendCom and disconnect the Rad-87 from the
serial cable.
11. Turn the Rad-87 off to exit the trend download mode.
NOTE: Contact USB to serial port adapter manufacturer for assistance or support.
ERASING TREND MEMORY
To erase (clear) the trend memory, set the Clear Trend option to Yes and press the Enter Button to accept the setting and clear the data from the memory. Refer to the Clear Trend menu located before this Trend setup and use section.
The Rad-87 continuously trends data. When performing a new study and gathering data on a new patient, it is highly recommended the "clear function" be utilized in order for the results to be separate. Turning the Rad-87 off will not erase the trend data.
TREND DATA FORMAT
After a successful download of the trend data, a .out file will be created containing the trend-dump information in ASCII delimited format. The format is defined in the following table.
4-20 Rad-87 Pulse CO-Oximeter Operator’s Manual
operation
TREND DATA FORMAT (CONTINUED)
PARAMETER SPECIFICATION
Date
Time
Installed Parameter/ Measurement
Exception Messages
MM\DD\YY
HH:MM:SS
Numeric value (see the display ranges in the Factory and User Configurable Default
Settings table located at the beginning of this section)
The exceptions are displayed as a 3 digit, ASCII encoded, hexadecimal value. The binary bits of the hexadecimal value are encoded as follows:
000 = Normal operation; no exceptions 001 = No Sensor 002 = Defective Sensor 004 = Low Perfusion 008 = Pulse Search 010 = Interference 020 = Sensor Off 040 = Ambient Light 080 = Unrecognized Sensor 100 = reserved 200 = reserved 400 = Low Signal IQ 800 = Masimo SET. This flag means the algorithm is running in full SET mode. It
requires a SET sensor and needs to acquire some clean data for this flag to be set
4
SAMPLE TREND OUTPUT
07/21/08 09:56:08 SpO2=000 PR=000 PI=00.00 EXC=820:OffPat,SET 07/21/08 09:56:10 SpO2=000 PR=000 PI=00.00 EXC=828:Search,OffPat,SET 07/21/08 09:56:12 SpO2=097 PR=069 PI=04.69 EXC=800:SET 07/21/08 09:56:14 SpO2=096 PR=074 PI=02.28 EXC=C00:LowSigIQ,SET 07/21/08 09:56:16 SpO2=098 PR=078 PI=03.64 EXC=800:SET 07/21/08 09:56:18 SpO2=000 PR=000 PI=00.00 EXC=800:SET 07/21/08 09:56:20 SpO2=000 PR=000 PI=00.00 EXC=820:OffPat,SET 07/21/08 09:56:22 SpO2=096 PR=078 PI=02.68 EXC=800:SET
Rad-87 Pulse CO-Oximeter Operator’s Manual 4-21
4
operation
SETUP MENU LEVEL 3 ENTER BUTTON + DOWN BUTTON MENU SETTINGS
The Level 3 menu contains advanced parameter/measurement settings.
To access Level 3 parameters/measurements, hold down the Enter Button and press the Down Button for 5 seconds. After entering menu Level 3, use the Enter Button to save new settings and move to the next menu.
The user may cycle through the menu options by continuing to press the Enter Button. Pressing the Display Button will exit the menu and return the display to home display screen.
AVERAGING TIME
BUTTONS SETTINGS
Default
8 seconds
4 seconds
Hold down
the Enter
Button and
press the
Down Button for 5 seconds.
+
2 seconds
16 seconds
14 seconds
12 seconds
10 seconds
(The cycling function for the
menu options is not available.)
Use Up or Down Button to move
between settings
AND
press the Enter Button to accept the
setting and
move to the next menu option.
OR
press the Display Button to exit
without saving the new setting and to
return to the home display screen.
4-22 Rad-87 Pulse CO-Oximeter Operator’s Manual
operation
SETUP MENU LEVEL 3 (CONTINUED)
RAPID DESAT LIMIT
The Rapid Desat Limit is designed to detect rapid desaturations of 5% or 10% below the low alarm limit and overrides the Alarm Delay feature when activated.
BUTTONS SETTING
4
Press the Enter
Button again
to move to the
next menu.
Default
5%
Off
10%
Use Up or Down Button to move between settings
AND
press the Enter Button to accept the setting and
move to the next menu.
OR
press the Display Button to exit without saving the
new settings and to return to the home display
screen.
ALARM ON/OFF
BUTTONS SETTING
Default
Press the Enter
Button again
to move to the
next menu.
When Alarm On/Off is set to "Off rE", the audible alarm "beeps" twice every three minutes to
*
On
Off
Off rE
(Alarm off with
reminder.)
Use Up or Down Button to move between
settings
AND
press the Enter Button to accept the setting
and move to the next menu.
OR
press the Display Button to exit without saving
the new setting and to return to the home
display screen.
remind the user that the Rad-87 is currently in alarm status but the audible alarm is muted. Visual alarms are active in this mode. If an alarm limit is violated, the associated parameter/ measurement label and value flash, the alarm bell flashes red for high priority alarms and the System Status Light is solid yellow for low priority alarms, flashes yellow for medium priority
alarms or flashes red for high priority alarms.
Rad-87 Pulse CO-Oximeter Operator’s Manual 4-23
4
operation
SETUP MENU LEVEL 3 (CONTINUED)
DEFAULT SETTINGS
BUTTONS SETTING
No change
Press the Enter
Button again
to move to the
next menu.
(Do not adjust factory
default settings.)
User Default
(
Set to user settings.*)
Factory Default
(Restore factory default
settings.)
Set Device Profile
(One or more pre-defined
device profiles)
Use Up or Down Button to
move between settings
AND
press the Enter Button
to accept the setting and
move to the next menu.
OR
press the Display Button to
exit without saving the new
setting and to return to the
home display screen.
*Set the Factory Default to this setting when confi guring a Device Profi le. Refer to the Device Profi le Setup and Use section of this chapter for additional information and instructions.
4-24 Rad-87 Pulse CO-Oximeter Operator’s Manual
operation
Device Profile Setup and Use
The Rad-87 can be configured to save changes to the device settings as a Device Profile. Using the Rad-87 button menu or an external configuration application, users can adjust Rad-87 settings and parameter/measurement alarm limits. After changing settings, the user may save the settings as a Device Profile. This Device Profile becomes the new default settings and the saved (Device Profile) settings will be retained after a power cycle.
To save the settings as a profile from the Rad-87 button menu, the user must enter Setup Menu Level 3 by pressing and holding the Enter Button and Down Button at the same time for 5 seconds. Then, by pressing the Enter Button three times, the Default Settings screen is displayed. Press the Up or Down Arrow Button until “User Default – Set” is displayed on the LCD. The user can press the Enter Button again to save the settings.
Using an external configuration application, the user can save up to five profiles. The user may select a color for the Device Profile LED to associate with the saved profiles. The Device Profile LED (located on the front panel of the Rad-87 above the sensor connector) will illuminate with the selected color, allowing the user to verify at a glance that a Device Profile has been set on the Rad-87. If changes are made to the device settings after the Device Profile feature has been enabled, the Device Profile LED will turn off until the device is returned to the user configured default settings or powered off, indicating a change from the Device Profile settings.
To set a device profile, the user enters Menu Level 3 by pressing and holding the Enter Button and Down Arrow Button at the same time for 5 seconds. Then, by pressing the Enter Button three times, the Default Setting screen is displayed. The user can use the Up or Down Arrow Button to select the desired profile and press the Enter Button to save the setting.
4
Rad-87 Pulse CO-Oximeter Operator’s Manual 4-25
4
SETUP MENU LEVEL 3 (CONTINUED)
PVI BAR ON/OFF (IF AVAILABLE)
BUTTONS SETTING
operation
Press the Enter
Button again
to move to the
next menu.
Default
On
Off
(No PVI Bar
or parameter/
measurement
label displayed)
Use Up or Down Button to move between settings
AND
press the Enter Button to accept the setting and move to the next menu.
OR
press the Display Button to exit without saving the new setting and to return to the home display screen.
SMART TONE ON/OFF
The SmartTone feature uses a proprietary algorithm that will provide pulse tones during excessive motion and low perfusion conditions. The pulse tone is based on an averaged pulse rate measurement from the proprietary algorithm and may not identify irregular heart beat patterns when there is excessive artifact present. The Normal Tone feature uses a proprietary algorithm that will provide pulse tones during non motion and adequate perfusion conditions. In this mode, the pulse tone may not sound if excessive artifact is present.
BUTTONS SETTING
Press the
Enter Button
again to move
to the next
menu.
Default
Off
(Normal Tone)
On
(Smart Tone)
Use Up or Down Button to move between settings
AND
press the Enter Button to accept the setting and move to the next menu.
OR
press the Display Button to exit without saving the new setting and to return to the home display screen.
YEAR
BUTTONS SETTING
Press the Enter
Button again
to move to the
next menu.
4-26 Rad-87 Pulse CO-Oximeter Operator’s Manual
Default
year
Use Up or Down
Button to adjust
the setting.
Use Up or Down Button to move between settings
AND
press the Enter Button to accept the setting and move to the next menu.
OR
press the Display Button to exit without saving the new setting and to return to the home display screen.
operation
SETUP MENU LEVEL 3 (CONTINUED)
MONTH
BUTTONS SETTING
Press the Enter
Button again
to move to the
next menu
DAY
BUTTONS SETTING
Press the Enter
Button again
to move to the
next menu.
Default
Month
Use Up or
Down Button
to adjust the
setting.
Default
days
Use Up or
Down Button
to adjust the
setting.
4
Use Up or Down Button to move between
settings
AND
press the Enter Button to accept the setting
and move to the next menu.
OR
press the Display Button to exit without
saving the new setting and to return to the
home display screen.
Use Up or Down Button to move between
settings
AND
press the Enter Button to accept the setting
and move to the next menu.
OR
press the Display Button to exit without
saving the new setting and to return to the
home display screen.
HOUR
BUTTONS SETTING
Press the Enter
Button again
to move to the
next menu.
Rad-87 Pulse CO-Oximeter Operator’s Manual 4-27
Default
hour
Use Up or
Down Button
to adjust the
setting.
Use Up or Down Button to move between
settings
AND
press the Enter Button to accept the setting
and move to the next menu.
OR
press the Display Button to exit without
saving the new setting and to return to the
home display screen.
4
SETUP MENU LEVEL 3 (CONTINUED)
MINUTE
BUTTONS SETTING
Press the Enter
Button again
to move to the
next menu.
SOFTWARE VERSION
BUTTONS SETTING
Press the Enter Button
again to move to the
next menu
Default
30 minutes
Use Up or
Down Button
to adjust the
setting.
Displays software version.
operation
Use Up or Down Button to move between
settings
AND
press the Enter Button to accept the setting
and move to the next menu.
OR
press the Display Button to exit without saving
the new setting and to return to the home
display screen.
SERIAL OUTPUT
BUTTONS SETTING
Default
Press the Enter
Button again to
move to the next
menu.
4-28 Rad-87 Pulse CO-Oximeter Operator’s Manual
AS2 (ASCII 2)
AS1 (ASCII 1)
PHL
(Philips Vuelink)
Use Up or Down Button to move between
settings
AND
press the Enter Button to accept the
setting and move to the next menu.
OR
press the Display Button to exit without
saving the new setting and to return to the
home display screen.
operation
SETUP MENU LEVEL 3 (CONTINUED)
INTERFACE ALARMS
When Rad-87 is interfaced to another system and the Interface Alarms are set to "Off", SpO2 and BPM audible alarms are muted at the Rad-87 and active at the interfaced system. This prevents both systems from producing SpO2 and BPM audible alarms at the same time.
NOTE: The Rad-87 reverts to Interface Alarms "On" during power interruptions or when the
interface connection is lost. This ensures that the Rad-87 provides SpO2 and BPM audible
alarms when connection to the interfaced system becomes compromised.
BUTTONS SETTING
Press the Enter
Button again to
move to the next
menu.
Default
On
NOTE: LCD shows
SpO2 and BPM alarms
On.
Off (only SpO2/BPM)
NOTE: LCD shows
SpO2 and BPM alarms
Off.
Use Up or Down Button to move
between settings
AND
press the Enter Button to accept the
setting and
move to the next menu.
OR
press the Display Button to exit
without saving the new setting and to
return to the home display screen.
4
System interfaces
PHILIPS VUELINK SETUP
1. Select the Philips VueLink selection from the Serial Output menu on the Rad-87. After
selecting, choose the preferred settings by stepping through menu options.
2. Connect one end of the VueLink cable to the Serial Output connector on the back of the Rad-87.
3. Connect the other end of the VueLink cable to the VueLink module and insert the module into the Philips/Agilent monitor rack.
4. The SpO
5. In order for the pleth waveform to be displayed on the Philips/Agilent monitor and for the Philips/
Agilent monitor to indicate the alarm conditions measured by the pulse oximeter, the user must configure the Philips/Agilent monitor. Please see the Philips/Agilent Operator’s manual for complete instructions.
6. The Rad-87 Pulse CO-Oximeter can be set up to audibly indicate all patient alarms while communicating with the Philips/ VueLink module. Use the Interface Alarms setting in the Output menu to enable and disable audible alarms on the Rad-87.
Rad-87 Pulse CO-Oximeter Operator’s Manual 4-29
and pulse rate values will automatically appear on the HP/Agilent monitor.
2
4
SETUP MENU LEVEL 3 (CONTINUED)
RADNET SETUP
1. Connect one end of the serial cable to the Serial Output connector on the back of the Rad-87.
2. Connect the other end of the serial cable to the RadNet Interface Module connector.
3. Turn the RadNet Interface Module on.
4. Select the ASCII 2 selection from the Serial options on the Rad-87.
5. A proper connection is shown by the RadNet Interface Module's Online LED being solid.
6. With a properly configured RadNet Interface Module, the Rad-87 will automatically display the SpO2, PVI, PI and Pulse Rate parameters/measurements on the screen at the RadNet Central Station.
7. The Rad-87 Pulse CO-Oximeter can be set up to audibly indicate all patient alarms while communicating with the RadNet Interface module.
PATIENT SAFETYNET SETUP
1. Select the ASCII 2 selection from the Serial options on the Rad-87.
2. Contact Masimo installation personnel for proper installation guidance.
operation
4-30 Rad-87 Pulse CO-Oximeter Operator’s Manual
operation
SETUP MENU LEVEL 3 (CONTINUED)
NURSE CALL*
BUTTONS SETTING
4
Press the Enter
Button again
to move to
the next menu
option.
Default
Alarm
Signal IQ
Alarm and
Signal IQ
Use Up or Down Button to move between
settings
AND
press the Enter Button to accept the setting
and move to the next menu option.
OR
press the Display Button to exit without saving
the new setting and to return to the home
display screen.
*Refer to Section 7, Nurse call specifi cations for additional information.
POLARITY*
BUTTONS SETTING
Press the Enter
Button again
to move to
the next menu
option.
Default
Normal
Inverse
Use Up or Down Button to move between
settings
AND
press the Enter Button to accept the setting
and move to the next menu option.
OR
press the Display Button to exit without saving
the new setting and to return to the home
display screen.
*Refer to Section 7, Nurse call specifi cations for additional information.
LINE FREQUENCY
BUTTONS SETTING
Press the Enter
Button again
to move to
the next menu
option.
Rad-87 Pulse CO-Oximeter Operator’s Manual 4-31
Default
60
50
Use Up or Down Button to move between
settings
AND
press the Enter Button to accept the setting
and move to the next menu option.
OR
press the Display Button to exit without saving
the new setting and to return to the home
display screen.
4
SETUP MENU LEVEL 3 (CONTINUED)
PARAMETER/MEASUREMENT SELECT - SCREEN 1
BUTTONS SETTING
operation
Press the Enter
Button again
to move to
the next menu
option.
SpHb
(if available)
PVI
(if available)
(no parameter/
measurement
displayed)
Use Up or Down Button to move between
settings
press the Enter Button to accept the setting and
move to the next menu option.
press the Display Button to exit without saving
the new setting and to return to the home
display screen.
PARAMETER/MEASUREMENT SELECT - SCREEN 2
BUTTONS SETTING
SpHb
Press the Enter
Button again
to move to
the next menu
option.
(if available)
PVI
(if available)
(no parameter/
measurement
displayed)
AND
OR
Use Up or Down Button to move between
settings
AND
press the Enter Button to accept the setting
and move to the next menu option.
OR
press the Display Button to exit without saving
the new setting and to return to the home
display screen.
4-32 Rad-87 Pulse CO-Oximeter Operator’s Manual
operation
SETUP MENU LEVEL 3 (CONTINUED)
PARAMETER/MEASUREMENT SELECT - SCREEN 3
BUTTONS SETTING
4
SpHb
Press the Enter
Button again
to move to
the next menu
option.
(if available)
PVI
(if available)
(no parameter/
measurement
displayed)
LCD LANGUAGE
BUTTONS SETTING
Press the Enter
Button again
to move to
the next menu
option.
Default
English
Scrolls through
available
languages
displayed on
the LCD
ENTER BUTTON + UP BUTTON MENU SETTING
SET MODE
BUTTONS SETTING
Hold down the
Enter Button
and press the Up Button for 5
seconds.
+
Default
Standard
Sleep Mode*
Home Mode
Use Up or Down Button to move between
settings
AND
press the Enter Button to accept the setting
and move to the next menu option.
OR
press the Display Button to exit without saving
the new setting and to return to the home
display screen.
Use Up or Down Button to move between
settings
AND
press the Enter Button to accept the setting
and move to the next menu option.
OR
press the Display Button to exit without saving
the new setting and to return to the home
display screen.
Use Up or Down Button to move between
settings
AND
press the Enter Button to accept the setting.
OR
press the Display Button to exit without saving
the new setting and to return to the home
display screen.
*CAUTION: ALARMS ARE DISABLED IN THIS MODE.
Rad-87 Pulse CO-Oximeter Operator’s Manual 4-33
4
operation
HOME MODE OPERATION
The Rad-87 can be placed into the Home Mode to protect unqualified users from changing the Rad-87 alarm settings and operation. Only the following menu and front panel functions are available: display brightness, pulse beep volume adjustment and alarm silence. Alarm volume is at highest setting. All default and user defined default settings are locked to their current values when home mode is selected and return to those values after a power cycle. Upon power up, the Hnn mode will be displayed along with a 10 second display of parameters/measurements. To turn the device off the Power Button must be depressed and held for 3 seconds. The Enter and Up Buttons held simultaneously for 5 second will put it
back into the special menu to select a different mode.
SLEEP MODE OPERATION
The Rad-87 can be placed into the Sleep Mode to allow the device to capture normal and abnormal patient data without triggering the alarms. This mode will blank out the LED and the LCD Displays with the exception of the AC Power Indicator, the System Status Light, and the Battery Level Indicator and disables the alarms even after a power cycle. However, pressing any button illuminates the display and the System Status Light (solid yellow) for 10 seconds. Upon power up, the SLP mode will be displayed along with a 10 second display of parameterr/measurement settings. The Enter and Up Buttons held
simultaneously for 5 seconds will put it back into the special menu to select a different mode.
CAUTION: ALARMS ARE DISABLED IN THIS MODE
BRIGHTNESS BUTTON
Access the Enable/Disable Radio menu by holding down the Brightness Button and the Down Button for
ENABLE/DISABLE RADIO
5 seconds.
BUTTONS SETTING
Hold down the
Brightness Button
and press the Down Button for 5 seconds.
+
LCD DISPLAY FUNCTION WITH RADIO CONFIGURED AND ENABLED
When the Radio feature is enabled (set to "On") and configured with the required network information, the LCD Display shows the following information for 20 seconds:
SSID
IP address
Subnet
Gateway
MAC Address
+ DOWN BUTTON MENU SETTING
Default
Off
On
Use Up or Down Button to move between
settings
press the Enter Button to accept the setting
and exit to the home display screen.
AND
4-34 Rad-87 Pulse CO-Oximeter Operator’s Manual
Loading...