Masimo RAD87A User Manual

Rad-87
Pulse CO-Oximeter
OPERATOR’S MANUAL
Rad-87
Pulse CO-Oximeter
OPERATOR’S MANUAL
The Rad-87 Operating Instructions provide the necessary information for proper operation of all models of the
Rad-87 device. General knowledge of pulse CO-Oximetry and an understanding of the features and functions of the
the instructions in this manual.
NOTICE:
Purchase or possession of this device does not carry any express or implied license to use this device with
replacement parts which would, alone or in combination with this device, fall within the scope of one of the patents
relating to this device.
CAUTION:
Federal law (U.S.) restricts this device to sale by or on the order of a physician.
Masimo Corporation
40 Parker
Irvine, CA 92618
USA
Tel.: 949-297-7000
Fax.: 949-297-7001
www.masimo.com
EU Authorized Representative for Masimo Corporation:
EC REP
MDSS GmbH
Schiffgraben 41
30175 Hannover, Germany
Tel.: +49-511-62 62 86 30
Fax.: +49-511-62 62 86 33
CONFORMS TO UL STD 60601-1, UL STD 763 UL STD 963 AND NSF STD 12;
3149433
Covered by one or more of the following U.S. Patents: RE38,492, RE38,476, 7,221,971, 7,215,986, 7,215,984,
7,186,966, 6,979,812, 6,861,639, 6,850,787, 6,826,419, 6,816,741, 6,745,060, 6,699,194, 6,684,090, 6,654,624,
6,650,917, 6,643,530, 6,606,511, 6,515,273, 6,501,975, 6,463,311, 6,430,525, 6,388,240, 6,360,114, 6,263,222,
6,236,872, 6,229,856, 6,157,850, 6,067,462, 6,011,986, 6,002,952, 5,919,134, 5,769,785, 5,758,644, 5,685,299,
5,632,272, 5,490,505, 5,482,036, international equivalents, or one or more of the patents referenced at www.masimo.
com/patents.htm. Other patents pending.
© 2008 Masimo Corporation. Masimo, Discrete Saturation Transform, DST, DCI, Signal Extraction Technology, SET,
Radical,
Rainbow and SpCO are federally registered trademarks of Masimo Laboratories.
Rad-87, Pleth Variability Index, PVI, Patient SafetyNet, RadNet, LNOPv and APOD are trademarks of Masimo
Corporation. Rainbow SET, SpMet, SpHb, SpOC, Pulse CO-Oximeter and Signal Extraction Pulse CO-Oximeter
are trademarks of Masimo Laboratories.
Rad-87 Pulse CO-Oximeter Operator’s Manual i
CERTIFIED TO CAN/CSA STD C22.2 NO. 601.1
, Signal IQ, SIQ, FastSat, LNOP and LNCS are federally registered trademarks of Masimo Corporation.
SAFETY INFORMATION, WARNINGS, CAUTIONS AND NOTES
The Rad-87 Pulse CO-Oximeter is designed to minimize the possibility of hazards from errors in the software program by following sound engineering design processes, Risk Analysis and Software Validation.
Explosion hazard. Do not use the Rad-87 in the presence of fl ammable anesthetics or other fl ammable
substance in combination with air, oxygen-enriched environments, or nitrous oxide.
High intensity extreme lights (including pulsating strobe lights) directed on the sensor, may not allow
the Pulse CO-Oximeter to obtain readings.
The Rad-87 is NOT intended for use as an apnea monitor.
The Pulse CO-Oximeter should be considered an early warning device. As a trend towards patient
hypoxemia is indicated, blood samples should be analyzed by laboratory instruments to completely understand the patient’s condition.
Pulse rate measurement is based on the optical detection of a peripheral flow pulse and therefore
may not detect certain arrhythmias. The pulse oximeter should not be used as a replacement or substitute for ECG based arrhythmia analysis.
The Rad-87 is to be operated by qualifi ed personnel only. This manual, accessory directions for use,
all precautionary information, and specifi cations should be read before use.
Electric shock hazard. Do not open the Rad-87 device. Only a qualifi ed operator may perform
maintenance procedures specifi cally described in this manual. Refer servicing to Masimo for repair of this equipment.
As with all medical equipment, carefully route patient cabling to reduce the possibility of patient
entanglement or strangulation.
Do not place the Rad-87 or accessories in any position that might cause it to fall on the patient. Do not
lift the Rad-87 by the power cord or any other cable.
Interfering Substances: Dyes, or any substance containing dyes, that change usual blood pigmentation
may cause erroneous readings.
SpO
is empirically calibrated to functional arterial oxygen saturation in healthy adult volunteers with
2
normal levels of carboxyhemoglobin (COHb) and methemoglobin (MetHb). A pulse oximeter can not measure elevated levels of COHb or MetHb. Increases in either COHb or MetHb will affect the accuracy of the SpO
For increased COHb: COHb levels above normal tend to increase the level of SpO
of increase is approximately equal to the amount of COHb that is present. NOTE: High levels of COHb may occur with a seemingly normal SpO laboratory analysis (CO-Oximetry) of a blood sample should be performed.
NOTE: High levels of COHb may occur with a seemingly normal SpO
COHb are suspected, laboratory analysis (CO-Oximetry) of a blood sample should be performed.
For increased MetHb: the SpO
to 15%. At higher levels of MetHb, the SpO levels of MetHb are suspected, laboratory analysis (CO-Oximetry) of a blood sample should be
measurement.
2
. The level
2
. When elevated levels of COHb are suspected,
2
. When elevated levels of
2
may be decreased by levels of MetHb of up to approximately 10%
2
may tend to read in the low to mid 80s. When elevated
2
performed.
Elevated levels of Methemoglobin (MetHb) will lead to inaccurate SpO
measurements.
and SpCO
2
Elevated levels of Carboxyhemoglobin (COHb) will lead to inaccurate SpO2 measurements.
Elevated levels of Total Bilirubin may lead to inaccurate SpO2, SpMet, SpCO, SpHb and SpOC
measurements.
Motion artifact may lead to inaccurate SpMet, SpCO, SpHb and SpOC measurements.
Rad-87 Pulse CO-Oximeter Operator’s Manualii
SAFETY INFORMATION, WARNINGS, CAUTIONS AND NOTES (CONTINUED)
Very low arterial Oxygen Saturation (SpO2) levels may cause inaccurate SpCO and SpMet
measurements.
Severe anemia may cause erroneous SpO
Hemoglobin synthesis disorders may cause erroneous SpHb readings.
readings.
2
Do not use the Rad-87 or sensors during magnetic resonance imaging (MRI) scanning. Induced
current could potentially cause burns. The Rad-87 may affect the MRI image and the MRI device may affect the accuracy of the Pulse CO-Oximetry parameters and measurements.
If using Rad-87 during full body radiation, keep the sensor out of the radiation fi eld. If the sensor is
exposed to the radiation, the reading might be inaccurate or the device might read zero for the duration of the active irradiation period.
For home use, ensure that the Rad-87’s alarm can be heard from other rooms in the house especially
when noisy appliances such as vacuum cleaners, dishwashers, clothes dryers, televisions, or radios are operating.
Always remove the sensor from the patient and completely disconnect the patient from the Rad-87
before bathing the patient.
Additional information specific to Masimo sensors including information about parameter/measurement
performance during motion and low perfusion, may be found in the sensor's Directions For Use (DFU).
Do not place the Rad-87 where the controls can be changed by the patient.
Do not place the Rad-87's face against a surface. This will cause the alarm to be muffl ed.
Do not place the Rad-87 on electrical equipment that may affect the Pulse CO-Oximeter, preventing it
from working properly.
Do not expose the Rad-87 to excessive moisture such as direct exposure to rain. Excessive moisture
can cause the device to perform inaccurately or fail.
Do not place containers with liquids on or near the Rad-87. Liquids spilled on the device may cause it
to perform inaccurately or fail.
If the Rad-87 fails any part of the setup procedures or leakage tests, remove the device from operation
until qualifi ed service personnel have corrected the situation.
Patient Safety - If a sensor is damaged in any way, discontinue use immediately.
Disposal of product - Comply with local laws in the disposal of the device and/or its accessories.
The Rad-87 can be used during defi brillation, but the readings may be inaccurate for up to 20
seconds.
This equipment has been tested and found to comply with the limits for medical devices to the
EN 60601-1-2: 2002, Medical Device Directive 93/42/EEC and Class B digital device, Part 15, FCC Rules/USA. These limits are designed to provide reasonable protection against harmful interference in a typical medical installation.
Rad-87 Pulse CO-Oximeter Operator’s Manual iii
SAFETY INFORMATION, WARNINGS, CAUTIONS AND NOTES (CONTINUED)
This equipment generates, uses and can radiate radio frequency energy and, if not installed
and used in accordance with the instructions, may cause harmful interference to other devices in the vicinity. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to other devices, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
Reorient or relocate the receiving device.
Increase the separation between the equipment.
Connect the equipment into an outlet on a circuit different from that to which the other
device(s) are connected.
Consult the manufacturer for help.
In order to connect wirelessly to a compatible interface system like Patient SafetyNet, the
Rad-87 should be placed in an environment free from RF shielding, which could hinder wireless reception.
To minimize radio interference, other electrical equipment that emits RF transmissions should
not be in close proximity to the RAD-87.
Changes or modifications to the wireless radio feature whether intentional or unintentional are
prohibited without written approval from Masimo Corporation.
The Rad-87 (device with optional radio) wirelessly transmits real-time sensor connectivity status,
indicating a connect and/or disconnect state. If the device is in a failure mode then the radio power is disabled and an error message is indicated on the device display. The device does not have a powered state where no information is transmitted.
In accordance with FCC requirements, the Rad-87 (device with optional radio) must be placed
greater then 20 cm from the patient's head.
In accordance with FCC requirements, radio accessories on the Rad-87 (device with optional
radio) cannot be attached directly to the patient using any accessory containing metal components.
In accordance with international telecommunication requirements, the frequency band of
5,150 MHz to 5,250 MHz is only for indoor usage to reduce potential for harmful interference to co-channel mobile satellite systems.
The battery should be adequately charged to ensure backup power in case of AC power
disruption.
A functional tester cannot be utilized to assess the accuracy of the Pulse CO-Oximeter or any
sensors.
To ensure safety, avoid stacking multiple devices or placing anything on the device during
operation.
Ensure the speaker is not covered or the device is placed face-down on bedding or other
sound absorbing surface.
To protect against injury from electric shock, follow the directions below:
Avoid placing the device on surfaces with visible liquid spills.
Do not soak or immerse the device in liquids.
Always turn off and disconnect the power cord from the AC power supply before cleaning
the device.
Use cleaning solutions sparingly.
Rad-87 Pulse CO-Oximeter Operator’s Manualiv
table of contents
SECTION 1 - OVERVIEW
About This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Warnings, Cautions and Notes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Optional Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Indications for Use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Pulse CO-Oximetry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
SpO2 General Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
SpCO General Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
SpMet General Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
SpHb General Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Total Arterial Oxygen Content (SpOC) General Description . . . . . . . . . . . . . . . . . 1-5
Principle of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Functional Saturation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Rad-87 vs. Drawn Whole Blood Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Masimo SET Signal Extraction Technology for SpO2 Measurements . . . . . . . . . 1-7
SpMet, SpCO, and SpHb and SpOC Measurements During Patient Motion . . . . 1-7
Fastat . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Masimo Rainbow SET Parallel Engines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Masimo SET DST
SECTION 2 - SYSTEM DESCRIPTION
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Rad-87 Front Panel Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Rad-87 Rear Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
LCD Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
® . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
SECTION 3 - SETUP
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Unpacking and Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Preparation for Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Rad-87 Power Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Initial Battery Charging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Initial Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
SECTION 4 - OPERATION
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Basic operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
General Setup and Use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Default Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Factory Default and User Configurable Settings . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Successful Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Masimo Pulse CO-Oximetry Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Numeric Display - SpO
Numeric Display - Pulse Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Numeric Display - SpCO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Numeric Display - SpMet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Numeric Display - SpHb . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Rad-87 Pulse CO-Oximeter Operator’s Manual v
2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4-5
table of contents
SECTION 4 - OPERATION (CONTINUED)
Numeric Display - SpOC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Numeric Display - PI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Pleth Variability Index (PVI) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
Low Perfusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
Signal SIQ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
Sensor Placement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
Sensitivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Low Battery Audible Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Normal Patient Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Paramater/Measurement Selection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
Setup Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
Menu Navigation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
Setup Menu Level 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
Parameter/Measurement Alarm Limits - Screen 1 . . . . . . . . . . . . . . . . . . . . . . . 4-13
Parameter/Measurement Alarm Limits - Screen 2 . . . . . . . . . . . . . . . . . . . . . . . 4-14
Parameter/Measurement Alarm Limits - Screen 3 . . . . . . . . . . . . . . . . . . . . . . . 4-15
LED Brightness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-15
Sensitivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-16
Setup Menu Level 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-17
Alarm Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-17
Alarm Silence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-17
Alarm Delay. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-18
Clear Trend . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-18
Button Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-19
FastSat . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-19
Trend Setup and Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-19
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-19
Trend Utility Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-19
Trendcom Utility Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-20
Erasing Trend Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-20
Trend Data Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-21
Sample Trend Output. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-21
Setup Menu Level 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-22
Averaging Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-22
Rapid Desat Limit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-23
Alarm On/Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-23
Out Put Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-24
Default Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-24
Device Profile Setup and Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-25
PVI Bar On/Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-26
Smart Tone On/Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-26
Year . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-26
Month . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-26
Day . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-27
Hour. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-27
Minute . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-28
Software Version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-28
Serial Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-29
Rad-87 Pulse CO-Oximeter Operator’s Manualvi
table of contents
SECTION 4 - OPERATION - (CONTINUED)
Interface Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-29
System Interfaces. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-29
Philips VueLink Set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-29
Radnet Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-30
Patient SafetyNet Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-30
Nurse Call . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-31
Polarity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-31
Line Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-31
Parameter/Measurement Select - Screen 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-32
Parameter/Measurement Select - Screen 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-32
Parameter/Measurement Select - Screen 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-33
LCD Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-33
Set Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-33
Home Mode Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-34
Sleep Mode Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-34
Enable/Disable Radio . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-34
LCD Display Function with Radio Configured and Enabled . . . . . . . . . . . . . . . . . . .4-35
SECTION 5 - ALARMS AND MESSAGES
Alarm Identification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Alarm Indication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Alarm Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Alarm Limit: User Configurable Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Alarm Limit: Factory Default Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Single Alarm Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Multi-Parameter/Measurement Alarm Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Alarm Priority for Display Screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Alarm Silence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Alarm Bell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
System Status Light . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Alarm Mute . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
SECTION 6 - TROUBLESHOOTING
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
SECTION 7 - SPECIFICATIONS
Rad-87 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Environmental . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Physical Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Serial Interface Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4
Serial Interface Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4
Serial Printer Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-5
Nurse Call Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-5
Nurse Call . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-5
Rad-87 Pulse CO-Oximeter Operator’s Manual vii
table of contents
SECTION 8 - SENSORS & PATIENT CABLES
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
Selecting a Masimo SET Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
Sensor Application Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
Masimo Rainbow Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2
Rainbow Reusable Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2
Rainbow Direct Connect Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2
Masimo SpO2 Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3
Red Direct Connect Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3
LNOP® Reusable Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3
LNOP® Adhesive Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4
LNOPvTM Adhesive Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4
LNOP® Specialty Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4
LNCS® Reusable Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4
LNCS® Adhesive Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-5
®
LNCS
Specialty Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-5
Sensor Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-6
Cleaning And Reuse Of Masimo Reusable Sensors and Cables . . . . . . . . . . . . . . . . . . 8-6
Reattachment of Single Use Adhesive Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-6
SECTION 9 - SERVICE AND MAINTENANCE
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1
Battery Service and Performance Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-2
Service and Repair. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-4
Repair Policy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-4
Return Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-4
Sales & End-user License Agreement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-5
Warranty. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-5
Exclusions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-5
End-user License . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-6
Restrictions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-6
SECTION 10 - PART NUMBERS
Part Numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-1
Rad-87 Pulse CO-Oximeter Operator’s Manualviii
overview
About This Manual
This manual explains how to set up and use the Rad-87 Pulse CO-Oximeter containing Masimo Rainbow SET technology. Important safety information relating to general use of the Rad-87 appears before this introduction. Other important safety information is located throughout the manual where appropriate.
Read the entire safety information section before you operate the monitor.
In addition to the safety section, this manual includes the following sections:
SECTION 1 OVERVIEW gives a general description of Rad-87 Pulse CO-Oximeter.
SECTION 2 SYSTEM DESCRIPTION describes the Rad-87 Pulse CO-Oximeter
system and its functions and features.
SECTION 3 SETUP describes how to setup the Rad-87 Pulse CO-Oximeter for use.
SECTION 4 OPERATION describes the operation of the Rad-87 Pulse CO-Oximeter
system.
SECTION 5 ALARMS AND MESSAGES describes the alarm system messages.
SECTION 6 TROUBLESHOOTING describes troubleshooting information.
SECTION 7 SPECIFICATIONS gives the detailed specifi cations of the Rad-87 Pulse
CO-Oximeter.
1
SECTION 8 SENSORS & PATIENT CABLES outlines how to use and care for Masimo
SECTION 9 SERVICE AND MAINTENANCE describes how to maintain, service and
SECTION 10 ACCESSORIES lists the available models of the Rad-87 Pulse
Rad-87 Pulse CO-Oximeter Operator’s Manual 1-1
Rainbow SET technology sensors, Masimo Rainbow SET technology patient cables, Masimo Red sensors and, Masimo Red PC cables.
obtain repair for the Rad-87 Pulse CO-Oximeter.
CO-Oximeter.
1
overview
Warnings, Cautions and Notes
Please read and follow any warnings, cautions and notes presented throughout this manual. An explanation of these labels are as follows:
A WARNING is provided when actions may result in a serious outcome (i.e., injury, serious adverse affect, death) to the patient or user. Look for text in a gray shaded box.
Sample of Warning:
WARNING: THIS IS A SAMPLE OF A WARNING STATEMENT.
A CAUTION is given when any special care is to be exercised by the patient or user to avoid injury to the patient, damage to this device or damage to other property.
Sample of Caution:
CAUTION: THIS IS A SAMPLE OF A CAUTION STATEMENT.
A NOTE is provided when extra general information is applicable.
Sample of Note:
NOTE: This is a sample of a Note.
1-2 Rad-87 Pulse CO-Oximeter Operator’s Manual
overview
Product Description
The Rad-87 Pulse CO-Oximeter Monitor is a noninvasive, arterial oxygen, carboxyhemoglobin and methemoglobin saturation, total hemoglobin concentration, total arterial oxygen content and pulse rate monitor. The Rad-87 features a multicolored LED display that continuously displays numeric values for SpO2, SpCO®*, SpMet™*, SpHb™*, total arterial oxygen content (SpOC*), perfusion index (PI), pleth variability index* (PVI) and pulse rate. It also provides bar graph displays for quick visual identifi cation of Signal Identifi cation Quality (SIQ®), perfusion index and pleth variability index.
The Rad-87 is available in four models: vertical Rad-87, horizontal Rad-87, vertical Rad-87 with radio and horizontal Rad-87 with radio.
FEATURES
These features are common to Rad-87 monitors:
Masimo SET is clinically proven to be the highest sensitivity and specifi city pulse oximeter
technology in the world.
Rainbow technology continuously and noninvasively measures arterial oxygen saturation (SpO2)
and pulse rate (BPM), as well as providing a reliable probe-off detection.
Perfusion Index (PI) with trending capability indicates arterial pulse signal strength during low
perfusion.
Accurate on cyanotic infants with congenital heart disease when used with an LNOP® Blue
Sensor.
Signal IQ® provides signal identifi cation and quality indication during excessive motion and low
signal to noise situations.
FastSat® tracks rapid changes in arterial O2 saturation with high fidelity.
Variable pitch provides tonal variance for every 1% change in saturation.
Remote alarming interface.
Up to 72 hours of trending. (See Section 4, Trends Setup and Use.)
Allows user to customize the default settings and set the device to retain these settings through a
power off/on cycle.
The LCD Display allows the user to view a scrolling marque of (installed) parameter/measurement
alarm limits, system information, and wireless radio communication (wireless radio model only).
1
OPTIONAL FEATURES
Rainbow technology uses 7+ wavelengths of light to continuously and noninvasively
measure carboxyhemoglobin (SpCO), methemoglobin (SpMet) and total hemoglobin (SpHb), as well as providing a reliable probe-off detection.
Pleth Variability Index (PVI) may show changes that refl ect physiologic factors such as
vascular tone, circulating blood volume, and intrathoracic pressure excursions.
Total arterial oxygen content (SpOC) provides a calculated measurement of the amount of
1
oxygen in arterial blood which may provide useful information for both oxygen dissolved in plasma and combined with hemoglobin.
Provides a 802.11a/b/g wireless radio which interfaces with compatible systems (wireless radio
model only)
Ability to connect to Masimo Patient SafetyNet through a wireless network (wireless radio
model only).
1
The utility of PVI is unknown at this time and requires further clinical studies. Technical factors that may affect PVI
include probe malposition and patient motion.
*Optional features: SpCO, SpMet, SpHb, SpOC, PVI
Rad-87 Pulse CO-Oximeter Operator’s Manual 1-3
1
overview
INDICATIONS FOR USE
The Rad-87 Pulse CO-Oximeter and accessories are indicated for the continuous, non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO methemoglobin concentration expressed in percentage (SpCO and SpMet) and total hemoglobin
), carboxyhemoglobin and
2
concentration expressed in grams per deciliter (SpHb).The Rad-87 Pulse CO-Oximeter and accessories are indicated for use with adult, pediatric and neonatal patients during both motion and no motion conditions, who are well or poorly perfused patients in hospitals, hospital-type facilities, mobile and home environments.
Pulse CO-Oximetry
SpO2 GENERAL DESCRIPTION
Pulse CO-Oximetry is a continuous and noninvasive method of measuring the level of arterial oxygen saturation in blood. The measurement is taken by placing a sensor on a patient, usually on the fi ngertip for adults and the hand or foot for neonates. The sensor is connected to the Pulse CO-Oximetry instrument with a patient cable. The sensor collects signal data from the patient and sends it to the instrument.
The following fi gure shows the general monitoring setup.
1
N O
R
M A P O
D
M A X
97
S
E N S I
T
I
V
I T
Y M O
D E
76
10.1
r b
c m
o n i t
o r
2
3
1. Instrument
2. Patient Cable
3. Sensor
SpCO GENERAL DESCRIPTION
Pulse CO-Oximetry is a continuous and noninvasive method of measuring the levels of carboxyhemoglobin concentration (SpCO) in arterial blood. It relies on the same basic principles of pulse oximetry (spectrophotometry) to make its SpCO measurement. The measurement is obtained by placing a sensor on a patient, usually on the fi ngertip for adults and the hand or foot for infants. The sensor connects either directly to the Pulse CO-Oximetry instrument or through an instrument patient cable. The sensor collects signal data from the patient and sends it to the instrument. The instrument displays the calculated data as percentage value for the SpCO, which refl ect blood levels of carbon monoxide bound to hemoglobin.
SpMet GENERAL DESCRIPTION
Pulse CO-Oximetry is a continuous and noninvasive method of measuring the levels of methemoglobin concentration (SpMet) in arterial blood. It relies on the same basic principles of pulse oximetry (spectrophotometry) to make its SpMet measurement. The measurement is obtained by placing a sensor on a patient, usually on the fi ngertip for adults and the hand or foot for infants. The sensor connects either directly to the Pulse CO-Oximetry instrument or through an instrument patient cable. The sensor collects signal data from the patient and sends it to the instrument. The instrument displays the calculated data as percentage value for the SpMet.
TOTAL HEMOGLOBIN (SpHb) GENERAL DESCRIPTION
1-4 Rad-87 Pulse CO-Oximeter Operator’s Manual
overview
Pulse CO-Oximetry is a continuous and noninvasive method of measuring the levels of total hemoglobin (SpHb) in arterial blood. It relies on the same principles of pulse oximetry to make the SpHb measurement. The measurement is taken by a sensor capable of measuring SpHb, usually on the fingertip for adults and pediatric patients. The sensor connects directly to the Pulse CO-Oximeter or with a patient cable. The sensor collects signal data from the patient and sends it to the instrument. The instrument displays the calculated data as measurement of total hemoglobin concentration. The Rad-87 can be configured to be a combined SpO
TOTAL ARTERIAL OXYGEN CONTENT (CaO2) GENERAL DESCRIPTION
Oxygen (O2) is carried in the blood in two forms, either dissolved in plasma or combined with hemoglobin. The amount of oxygen in the arterial blood is termed the oxygen content (CaO is measured in units of ml O whereas 100 ml of blood plasma may carry approximately 0.3 ml of oxygen. The oxygen content is determined mathematically as:
CaO
= 1.34 (ml O2/g Hb) x Hb (g/dl) x HbO2 + PaO2 (mm Hg) x (0.3 ml O2/ 100 mm Hg/dl)
2
Where HbO arterial oxygen.
For typical PaO Hg/dl] is approximately 0.3 ml/dl. Furthermore, for typical carboxyhemoglobin and methemoglobin levels, the functional saturation (SpO
2
Martin, Laurence. All You Really Need to Know to Interpret Arterial Blood Gases, Second Edition.
New York: Lippincott Williams & Wilkins, 1999.
monitor with other available parameters/measurements.
2
2
) and
/dl blood. One gram of hemoglobin (Hb) can carry 1.34 ml of oxygen,
2
is the fractional arterial oxygen saturation and PaO2 is the partial pressure of
2
values, the second part of the above equation [PaO2 (mm Hg) x (0.3 ml O2/ 100 mm
2
) as measured by a pulse oximeter is given by:
2
SpO
= 1.02 x HbO2
2
2
1
SpOC General Description (Pulse CO-Oximetry)
The above approximations result in the following reduced equation for oxygen content via the Pulse CO-Oximeter:
SpOC (ml/dl*) = 1.31 (ml O
*When ml O the gram unit in the numerator of g/dl resulting in ml/dl (ml of oxygen in one dl of blood) as the
/g Hb is multiplied by g/dl of Hb, the gram unit in the denominator of ml/g cancels
2
/g Hb) x SpHb (g/dl) x SpO2 + 0.3 ml/dl
2
unit of measure for SpOC.
Rad-87 Pulse CO-Oximeter Operator’s Manual 1-5
1
overview
PRINCIPLE OF OPERATION
Pulse CO-Oximetry is governed by the following principles:
1. Oxyhemoglobin (oxygenated blood), deoxyhemoglobin (non-oxygenated blood), carboxyhemoglobin (blood with carbon monoxide content), methemoglobin (blood with oxidized hemoglobin) and blood plasma constituents differ in their absorption of visible and infrared light (using spectrophotometry, see fi gure below).
Absorption Spectra
4.0
3.5
3.0
2.5
2.0
1.5
Absorption (1/mm)
1.0
0.5
0 600 800
Carboxyhemoglobin
Oxyhemoglobin
Methemoglobin
Deoxyhemoglobin
Plasma
1000 1200 1400 1600
Wavelength (nm)
2. The amount of arterial blood in tissue changes with your pulse (photoplethysography). Therefore, the amount of light absorbed by the varying quantities of arterial blood changes as well.
The Rad-87 Pulse CO-Oximeter uses a multi-wavelength sensor to distinguish between oxygenated blood, deoxygenated blood, blood with carbon monoxide, oxidized blood and blood plasma. The Rad-87 utilizes a sensor with various light-emitting diodes (LEDs) that pass light through the site to a photodiode (detector). See fi gure below. Signal data is obtained by passing various visible and infrared lights (LED’s, 500 to 1400nm) through a capillary bed (for example, a fi ngertip, a hand, a foot) and measuring changes in light absorption during the blood pulsatile cycle. This information may be useful to clinicians. The maximum radiant power of the strongest light is rated at 25mW. The detector receives the light, converts it into an electronic signal and sends it to the Rad-87 for calculation.
1
1. Light Emitting Diodes (LEDs) ( 7+ wavelengths )
2. Detector
2
Once the Rad-87 receives the signal from the sensor, it utilizes Masimo Rainbow SET signal extraction technology to calculate the patient’s functional arterial oxygen saturation, blood levels of carboxyhemoglobin (SpCO), methemoglobin (SpMet) and pulse rate. The SpCO and SpMet measurements rely on a multiwavelength calibration equation to quantify the percentage of carbon monoxide and methemoglobin in arterial blood. In an ambient temperature of 35º C the maximum skin surface temperature has been measured at less than 106º F (41º C), verifi ed by Masimo sensor skin temperature test procedure.
1-6 Rad-87 Pulse CO-Oximeter Operator’s Manual
overview
FUNCTIONAL SATURATION
The Rad-87 is calibrated to measure and display functional saturation (SpO expressed as a percentage of the hemoglobin that is available to transport oxygen.
RAD-87 vs. DRAWN WHOLE BLOOD MEASUREMENTS
When SpO compared to drawn whole blood (invasive) measurements by blood gas and/or laboratory CO-Oximetry
, SpCO, SpMet and SpHb measurements obtained from the Rad-87 (noninvasive) are
2
methods, caution should be taken when evaluating and interpreting the results. The blood gas and/or laboratory CO-Oximetry measurements may differ from the SpO of the Rad-87 Pulse CO-Oximeter. In the case of SpO arterial blood gas sample if the calculated measurement is not appropriately corrected for the effects of
, SpCO, SpMet and SpHb measurements
2
, different results are usually obtained from the
2
variables that shift the relationship between the partial pressure of oxygen (PO as: pH, temperature, the partial pressure of carbon dioxide (PCO the case of SpCO, different results are also expected if concentration of methemoglobin in the blood gas sample is elevated. High levels of bilirubin may cause erroneous SpO readings. As blood samples are usually taken over a period of 20 seconds (the time it takes to draw the blood) a meaningful comparison can only be achieved if the oxygen saturation, carboxyhemoglobin and methemoglobin concentration of the patient are stable and not changing over the period of time that the blood gas sample is taken. Subsequently, blood gas and laboratory CO-Oximetry measurements of SpO2, SpCO, SpMet and SpHb may vary with the rapid administration of fl uids and in procedures such as dialysis. Additionally, drawn, whole-blood testing can be affected by sample handling methods and time elapsed between blood draw and sample testing.
MASIMO SET SIGNAL EXTRACTION TECHNOLOGY FOR SpO
Masimo Signal Extraction Technology’s signal processing differs from conventional pulse oximeters. Conventional pulse oximeters assume that arterial blood is the only blood moving (pulsating) in the measurement site. During patient motion, however, the venous blood also moves, causing conventional pulse oximeters to read low values, because they cannot distinguish between the arterial and venous blood movement (sometimes referred to as noise). Masimo SET pulse oximetry utilizes parallel engines and adaptive digital fi ltering. Adaptive fi lters are powerful because they are able to adapt to the varying physiologic signals and/or noise and separate them by looking at the whole signal and breaking it down to its fundamental components. The Masimo SET signal processing algorithm, Discrete Saturation Transform
®
(DST
) reliably identifi es the noise, isolates it and, using adaptive fi lters, cancels it. It then reports the true
arterial oxygen saturation for display on the monitor.
SpMet, SpCO, AND SpHb MEASUREMENTS DURING PATIENT MOTION
The Rad-87 displays measurements of SpCO, SpMet and SpHb during patient motion. However, because of the changes in the physiological parameters such as blood volume, arterial-venous coupling, etc. that occur during patient motion, the accuracy of such measurements may not be reliable during excessive motion. When the Rad-87 does not have confidence in the value of a parameter due to poor signal quality caused by excessive motion or other signal interference, the measurement for the parameter will alternate with "---".
FASTSAT
FastSat enables rapid tracking of arterial oxygen saturation changes. Arterial oxygen saturation data is averaged using pulse oximeter averaging algorithms to smooth the trend. When the Rad-87 is set to FastSat “On”, the averaging algorithm evaluates all the saturation values providing an averaged saturation value that is a better representation of the patient’s current oxygenation status. With FastSat, the averaging time is dependent on the input signal.
): the amount of oxyhemoglobin
2
) and saturation, such
), 2,3-DPG, and fetal hemoglobin. In
2
2
, SpMet, SpCO and SpHb
2
MEASUREMENTS
2
®
1
Rad-87 Pulse CO-Oximeter Operator’s Manual 1-7
1
MASIMO RAINBOW SET PARALLEL ENGINES
This fi gure is for conceptual purposes only.
12
FST
MST
SpOC
MASIMO SET DST
1-8 Rad-87 Pulse CO-Oximeter Operator’s Manual
system description
Introduction
The Rad-87 Pulse CO-Oximeters are full featured devices designed for ease of operation. All pulse CO-Oximetry measurement information, as well as device status data, is displayed on the front panel of the device. All user input is handled by control buttons on the front panel. The sensor cable connections are located on the left side of the front panel for the Rad-87 horizontal device and the bottom of the front panel for the Rad-87 vertical device.
Rad-87 offers full Masimo SET technology in a small compact device
Rad-87 supports the full line of Masimo sensors and patient cables (see Section 8, Sensors
and Patient Cables)
Rad-87 supports standardization of sensors, and pulse CO-Oximetry technology throughout
the hospital
The LCD Display identifies system settings, monitoring modes, alarm limits and information
from Patient SafetyNet or Philips VueLink (when connected). The LCD is located on top of the device (Horizontal) or on the left of the device (Ver tical).
2
Rad-87 Pulse CO-Oximeter Operator’s Manual 2-1
2
1234567890123456
APOD
MAX
NORM
APOD
MAX
NORM
system description
RAD-87 PULSE CO-OXIMETER - HORIZONTAL
1 23 5678 91011
APOD
APOD
APOD
APOD
CONTROL / INDICATOR DESCRIPTION
Device Profile
1
LED
4
DISPLAY
NORM
NORM
NORM
NORM
SENSITIVITY
SENSITIVITY
MODE
MODE
SENSITIVITY
SENSITIVITY
MAX
MAX
MAX
MAX
10.1
DISPLAY
ENTER
ENTER
rbc monitor
rbc monitor
Rad-87
Rad-87
121314151617181920
The Device Profile LED illuminates when the device has been
set to user configured "default" settings. Upon power up, the
user configured default settings are retained and the Device
Profile LED remain lit.
When user configured default settings are active, any changes
to the default settings cause the Device Profile LED to turn
off until the device is returned to the user configured default
settings or powered off.
Sensitivity
2
Button/Indicator
Used to set the device into Maximum Sensitivity, Normal
Sensitivity, or APOD Mode.
The LCD display identifies system settings, monitoring modes, alarm
3
LCD Display
limits, and information from Patient SafetyNet or Philips VueLink (when
connected.)
The Signal IQ provides an indication of the quality of the acquired
4
Signal IQ Index
5
Perfusion Index
2-2 Rad-87 Pulse CO-Oximeter Operator’s Manual
signal as well as the timing of the pulse. A green vertical LED bar
rises and falls with the pulse, where the height of the bar indicates
the quality of the signal.
The Perfusion Index provides an indication of the percentage of
pulsatile signal to non pulsatile signal.
Loading...
+ 44 hidden pages