Masimo Rad-87 Service manual

Page 1
Rad-87
Signal Extraction Pulse CO-Oximeter
Page 2
Rad-87
Signal Extraction Pulse CO-Oximeter
ii
Rad-87 Signal Extraction Pulse CO-Oximeter Service Manual
Page 3
The Rad-87 Pulse CO-Oximeter Service Manual is intended to provide the necessary information for proper servicing of all models of the Rad­87 Pulse CO-Oximetry systems. This manual is structured to support troubleshooting to the assembly or module level. This manual does not provide instructions for troubleshooting to the printed circuit board component level. There may be information provided in this manual that is not relevant for your system. Do not service the Rad-87 Pulse CO­Oximeter without completely reading and understanding these instructions.
NOTICE:
Purchase or possession of this device does not carry any express or implied license to use this device with replacement parts which would, alone or in combination with this device, fall within the scope of one of the patents relating to this device.
CAUTION:
Federal law (U.S.) restricts this device to sale by or on the order of a physician.
Masimo Corporation
40 Parker
Irvine, CA 92618
USA
Tel.: 949-297-7000
Fax.: 949-297-7001
www.masimo.com
EU Authorized Representative for Masimo Corporation:
EC REP
MDSS GmbH
Schiffgraben 41
30175 Hannover, Germany
Tel.: +49-511-62 62 86 30
Fax.: +49-511-62 62 86 33
CONFORMS TO UL STD 60601-1 AND CERTIFIED TO CAN/CSA STD C22.2 NO. 601.1
Covered by one or more of the following U.S patents: 5482036, 5490505, 5632272, 5685299, 5758644, 5769785, 5919134, 6002952, 6011986, 6067462, 6157850, 6229856, 6236872, 6263222, 6360114, 6388240, 6430525, 6463311, 6501975, 6515273, 6606511, 6643530, 6650917, 6654624, 6661161, 6684090, 6699194, 6745060, 6816741, 6826419, 6850787, 6861639, 6979812, 7186966, 7215984, 7215986, 7221971, 7254433, 7295866, 7328053, 7373194, 7376453, 7377899, 7467002, 7469157, 7471969, 7489958, 7496393, 7499741, 7509154, 7530955, RE38476, RE38492, international equivalents, or one or more of the patents referenced at www.masimo.com/patents. Other patents pending.
© 2010 Masimo Corporation. LNCS, Masimo, MX-1, PVI, Rad, the Radical logo, Rainbow, Satshare, SET, Signal I.Q., SpCO, and SpMet are federally registered trademarks of Masimo Corporation. APOD, Pulse CO­Oximeter, Pulse CO-Oximetry and Rad 87 are trademarks of Masimo Corporation.
iii
Rad-87 Signal Extraction Pulse CO-Oximeter Service Manual
Page 4
Table of Contents
Page
1 Overview........................................................................................1-1
1.1 About this Manual.....................................................................1-1
1.2 Warnings, Cautions and Notes.................................................1-1
2 Maintenance ..................................................................................2-1
2.1 Introduction...............................................................................2-1
2.2 Cleaning....................................................................................2-1
3 Battery Operation and Maintenance ...........................................3-1
3.1 Overview...................................................................................3-1
3.2 Replacing the Battery ...............................................................3-1
4 Performance Verification .............................................................4-1
4.1 Overview...................................................................................4-1
4.2 Power-On Self-Test..................................................................4-1
4.3 Key Press Button Test..............................................................4-1
4.4 Alarm Suspend Indicator ..........................................................4-1
4.5 Alarm Volume Test...................................................................4-1
4.6 Alarm Limit Test........................................................................4-2
4.7 Sensitivity Mode Test ...............................................................4-2
4.8 Button Volume Test..................................................................4-2
4.9 Brightness Test.........................................................................4-3
4.10 Testing with Masimo Rainbow SET Tester ..............................4-3
4.11 PVI (Pleth Variability Index) Testing (If Applicable)..................4-4
4.12 Wireless Radio Testing (If Applicable)......................................4-4
4.13 Serial Port.................................................................................4-4
4.14 Nurse Call Test.........................................................................4-5
5 Electrical Safety Tests..................................................................5-1
5.1 Test Equipment.........................................................................5-1
5.2 Ground Wire Leakage Current Tests .......................................5-1
5.3 Enclosure Leakage Current Tests............................................5-2
5.4 Source Leakage Current Tests.................................................5-3
6 Repair.............................................................................................6-1
6.1 Safety Precautions: ..................................................................6-1
6.2 General Procedures .................................................................6-1
6.3 Troubleshooting........................................................................6-2
6.4 Service Messages: ...................................................................6-3
7 Returns...........................................................................................7-1
7.1 Product Decontamination .........................................................7-1
7.2 Return Transportation...............................................................7-1
8 Drawings and Schematics ...........................................................8-1
iv
Rad-87 Signal Extraction Pulse CO-Oximeter Service Manual
Page 5
v
Rad-87 Signal Extraction Pulse CO-Oximeter Service Manual
Page 6
Page 7
1 Overview
1.1 About this Manual
This manual explains how to service the Rad-87 Pulse CO-Oximeter. Important safety information relating to general use of the Rad-87 Pulse CO-Oximeter appears in the Rad-87 CO-Oximeter Operator’s Manual. Other important safety information is located throughout this manual where appropriate.
1.2 Warnings, Cautions and Notes
Please read and follow any warnings, cautions and notes presented throughout this manual. An explanation of these labels is as follows:
A WARNING is provided when actions may result in a serious outcome (i.e., injury, serious adverse affect, or death) to the patient or user. Look for text in a gray shaded box.
Sample of WARNING: WARNING: THIS IS A SAMPLE OF A WARNING STATEMENT.
A CAUTION is given when any special care is to be exercised by the patient or user to avoid injury to the patient, damage to this device or damage to other property.
Sample of Caution: CAUTION: This is a sample of a Caution statement.
A NOTE is provided when extra general information is applicable. Sample of NOTE:
NOTE: This is a sample of a NOTE statement.
1-1
Rad-87 Signal Extraction Pulse CO-Oximeter Service Manual
Page 8
Page 9
2 Maintenance
2.1 Introduction
2.1.1 This chapter covers how to test the operation of the Rad-87 and SatShare interface (if applicable), how to properly clean the Rad­87 Pulse CO-Oximeter, how to recharge and replace the batteries, how to replace the fuses, and how to obtain service under normal operation. No internal adjustment or recalibration is required. Service should be performed by qualified personnel only. Safety checks should be performed at regular intervals or in accordance with local and governmental regulations.
Note: Opening or repairing the Rad-8 Pulse Oximeter will void the Masimo warranty. Products that require repair while under warranty should always be returned to Masimo for warranty service.
WARNING: ELECTRICAL SHOCK AND FLAMMABILITY HAZARD. BEFORE CLEANING THE PULSE CO-OXIMETER,
ALWAYS TURN THE UNIT OFF AND DISCONNECT THE POWER CORD FROM THE BACK OF THE UNIT.
2.2 Cleaning
2.2.1 To clean the display panel use a cotton swab moistened with
70% isopropyl alcohol and gently wipe the panel.
2.2.2 To clean the outer surface of the Pulse CO-Oximeter, use a soft
cloth moistened with a 70% isopropyl alcohol. Do not allow liquids to enter the interior of the instrument.
2.2.3 To decontaminate unit, refer to Section 6.
CAUTIONS:
Do not autoclave, pressure sterilize, or gas sterilize this Pulse CO-Oximeter.
Do not soak or immerse the monitor in any liquid.
Use the cleaning solution sparingly. Allow unit to air dry. Excessive solution
can flow into the monitor and cause damage to internal components.
Do not touch, press, or rub the display panels with abrasive cleaning
compounds, instruments, brushes, rough-surface materials, or bring them into contact with anything that could scratch the panel.
2-1
Rad-87 Signal Extraction Pulse CO-Oximeter Service Manual
Page 10
Do not use petroleum-based or acetone solutions, or other harsh solvents, to
clean the Pulse CO-Oximeter. These substances attack the instrument’s materials and instrument failure can result.
2-2
Rad-87 Signal Extraction Pulse CO-Oximeter Service Manual
Page 11
3 Battery Operation and Maintenance
3.1 Overview
3.1.1 The Rad-87 Pulse CO-Oximeter uses a 6 volt, 1.4AHr lead acid,
sealed, rechargeable battery.
3.1.2 Please reference the Rad-87 Operator’s Manual for
specifications on battery charge time and capacity.
CAUTIONS:
If the Rad-87 has not been charged for 1 month or more, then
recharge the battery prior to use.
At low battery warning, connect the Rad-87 to AC power to prevent
interruption of power.
Additional information on the Rad-87 batteries may be found in the
Rad-87 Operators Manual.
3.2 Replacing the Battery
3.2.1 Turn the Rad-87 off and remove the patient cable connection.
3.2.2 Place the unit on its side with the rubber footings facing up.
3.2.3 Remove the four screws from the housing assembly.
3.2.4 Carefully remove the speaker-side housing assembly. DO NOT
pull housing away as it may damage the speaker cable.
3.2.5 Remove the speaker connector from the system board and set
the housing aside.
3.2.6 Carefully lift the battery out of the housing and remove the
positive and negative battery connections.
3.2.7 Install the replacement battery. Verify proper positioning of the
(+) positive and (-) negative connectors. Reverse steps 3..3.1 through 3.3.6 to reassemble.
3.2.8 Verify the unit turns on, that all the LEDs momentarily light (AC
LED will not illuminate when powering on with battery power only), that the low battery LED is not illuminated, and that the unit stabilizes to a normal operation mode.
CAUTION: Follow local governing guidelines for proper disposal of the internal battery. Do not incinerate.
3-1
Rad-87 Signal Extraction Pulse CO-Oximeter Service Manual
Page 12
CAUTION: Do not over-tighten screws. Tighten the screws throughout
the Rad-87 to 80 in.- oz.
NOTE: It will be necessary to perform safety testing after replacement. Refer to
section 5.
3-2
Rad-87 Signal Extraction Pulse CO-Oximeter Service Manual
Page 13
4 Performance Verification
NOTE: Under normal operation, no internal adjustment or recalibration is
required.
4.1 Overview
4.1.1 To test the performance of the Rad-87 Pulse CO-Oximeter
following repairs or during routine maintenance, follow the procedure outlined in this section. If the Rad-87 fails any of the described tests, discontinue its use and correct the problem before returning the unit back to the user.
4.2 Power-On Self-Test
4.2.1 With the Rad-87 plugged into an AC power source and no cable
or sensor plugged into the instrument, turn the monitor on by depressing the Power Button. For about 2 seconds all available LEDs are illuminated and a brief beep tone sounds. The unit will then cycle through Saturation and Pulse Rate alarm limits on the top LCD.
4.2.2 After this initialization “boot” cycle, the No Sensor, Alarm
Suspend and Normal Sensitivity Mode LED indicators will light. Also, the SpO2 and BPM monitoring display will indicate “---“. The Rad-87 is ready to begin normal operation.
4.2.3 Turn power off and then back on as many times as necessary to
verify this Power-On cycle.
4.3 Key Press Button Test
4.3.1 With the exception of the Power/Standby Button, press each soft
key button and verify that the Rad-87 acknowledges each key­press with an audible beep tone or by indicating an associated change on the display.
4.4 Alarm Suspend Indicator
4.4.1 Induce an alarm condition.
4.4.2 Press the Alarm Suspend button and cycle through the 2
different indicator levels: the red indicator blinks red when an alarm condition is met. The yellow indicator will blink on when the alarms have been suspended.
4.5 Alarm Volume Test
4.5.1 Press the Enter button. Monitor displays should show “AL” and
“VOL”.
4-1
Rad-87 Signal Extraction Pulse CO-Oximeter Service Manual
Page 14
4.5.2 Press the UP arrow button and cycle through the three different
volume levels. Make sure the tone does not have a harsh or raspy tone due to the speaker rattling in the plastic housing. The tone heard should be louder with each press. Press the DOWN arrow button and cycle through the different volume levels. The tone heard should be quieter with each press. Verify both speakers are working.
4.5.3 Return the volume to the factory default level 3.
4.6 Alarm Limit Test
4.6.1 With the monitor turned on, depress the ALARM LIMITS button
and enter alarm menu. Change the High Saturation Alarm parameter to a value two points below the currently selected value, and accept the change by pressing the Enter button.
4.6.2 Verify that the newly set parameter is shown on the Saturation
Alarm Limit Display.
4.6.3 Return the High Saturation Alarm parameter to its original setting
and accept the change by pressing the Enter button.
4.6.4 Repeat steps 4.6.1 to 4.6.3 with the Low Saturation Alarm
parameter.
4.6.5 Repeat steps 4.6.1 to 4.6.3 with the Low and High Pulse Rate
Alarm parameter.
4.6.6 Reset the alarm limits again to the original settings and accept
the changes by pressing the Enter button.
4.7 Sensitivity Mode Test
4.7.1 Press the Sensitivity button to alternately light the APOD,
NORM, and MAX LEDs.
4.8 Button Volume Test
4.8.1 Press the Enter button x5. LED displays should show “but”, and
“VOL”.
4.8.2 Press the Up arrow button and cycle through the different
volume levels. Make sure the tone does not have a harsh or raspy tone due to the speaker rattling in the plastic housing. The tone heard should be louder with each press. Press the DOWN arrow button and cycle through the different volume levels. The tone heard should be quieter with each press. There are three volume levels and volume off.
4.8.3 Return the volume to the factory default level 2. 4-2
Rad-87 Signal Extraction Pulse CO-Oximeter Service Manual
Page 15
4.9 Brightness Test
4.9.1 Press the Display Brightness button. LED displays should show
“dlS” above and “brt” below.
4.9.2 Press the Display Brightness button and cycle through the
different brightness levels. There are four brightness levels. Factory default is level 2.
4.9.3 Press the Enter button to return to patient monitoring mode or
wait for display to time out.
4.10 Testing with Masimo Rainbow SET Tester
4.10.1 Turn the Rad-87 off and then on again.
4.10.2 Set the alarm limits to:
% SpO2 Pulse Rate % SpCO % SpMet
High 100 140 10 3
Low 90 50 -- --
4.10.3 Connect the Masimo tester to the Rad-87 (PN 2386). (See
Tester DFU for complete instructions.)
4.10.4 Verify that within 20 seconds a Signal I.Q. Bar Graph and SpO
2
& Pulse Rate numbers display.
4.10.5 Verify that the SpO2 measurement is between 78% and 84%.
(with tester values of 81% +/- 3%)
4.10.6 Verify that the pulse rate measurement is between 60 bpm and
62 bpm (with tester values of 61 +/- 1 bpm).
4.10.7 Verify that the SpCO measurement is between 4% and 10%.
(with tester values of 7% +/- 3%)
4.10.8 Verify that the SpMet measurement is between 2.4% and 4.4%.
(with tester values of 3.4% +/- 1%)
4.10.9 Verify that an audible alarm occurs and that the SpO2
measurement and the low SpO2 alarm limit display are flashing
4.10.10 Press the Alarm Silence button once and verify that the alarm is
silenced.
4.10.11 Wait 120 seconds and verify that the alarm silence times out and
the audible alarm is activated again. Disable the Alarm before continuing.
4-3
Rad-87 Signal Extraction Pulse CO-Oximeter Service Manual
Page 16
4.10.12 Press the Up Arrow button several times and verify that the
loudness of the pulse beep tone increases.
4.10.13 Press the Down Arrow button and verify that the loudness of the
pulse beep tone decreases. Press the Down Arrow button and verify that the loudness of the pulse beep tone can be turned off.
4.11 PVI (Pleth Variability Index) Testing (If Applicable)
NOTE
: THIS IS PERFORMED AS A FUNCTIONAL TEST ONLY.
4.11.1 Connect a Red Masimo SET Rainbow sensor to the Rad-87
Handheld.
4.11.2 Insert the non-dominant hand’s ring finger into the finger sensor.
4.11.3 PVI will fluctuate depending on motion, perfusion, physiology
and other factors, so there is no ‘normal’ value. If a number is displayed, then the PVI feature is functioning properly.
4.12 Wireless Radio Testing (If Applicable)
4.12.1 Press and hold the Brightness button and the Down button for 5
seconds.
4.12.2 Use the Up and Down arrows to turn the wireless radio on.
4.12.3 When the radio feature is enabled and configured with the
network information, the LCD display will show the SSID, IP Address, Subnet, Gateway, MAC Address, and Signal Strength for 20 seconds.
4.13 Serial Port
4.13.1 The serial port outputs ASCII data using standard RS-232
protocol via a straight through serial cable.
4.13.2 Validation of the port can be done by connecting a serial printer
or using a PC with Hyper Terminal and the following setup:
4.13.3 Communication Parameters
baud rate 9600 bps
data bits 8
stop bit 1
parity none
hand shaking none
4-4
Rad-87 Signal Extraction Pulse CO-Oximeter Service Manual
Page 17
4.13.4 With the Rad-87 connected as per the above setup, attach a
patient cable and sensor to the Rad-87. From the Output menu, select ASCII 1 and the printer (or PC monitor) will print in the following format every second:
05/15/10 08:12:21 SN=0000070986 SPO2=---% BPM=--­PI=--.--% SPCO=--.-% SPMET=--.-% DESAT=-­PIDELTA=+-- PVI=--- ALARM=0000 EXC=000000
4.14 Nurse Call Test
4.14.1 Disconnect the patient cable or the Masimo SET Tester from the
monitor and turn the instrument on. Ensure that there are no audible alarms and that the audible alarms are not silenced.
4.14.2 Connect the common lead of a digital multi-meter to the tip
(Nurse Call - Common) of the Nurse Call Connector.
4.14.3 Connect the positive lead of the multi-meter to the sleeve (Nurse
Call - Normally Open) of the Nurse Call Connector and measure that the resistance is greater than 1 M Ohm (open circuit).
4.14.4 Trigger an alarm on the monitor (e.g. by disconnecting a sensor
after it has started measuring data) and verify the resistance is < 20 Ohm.
4-5
Rad-87 Signal Extraction Pulse CO-Oximeter Service Manual
Page 18
Page 19
5 Electrical Safety Tests
Masimo recommends that you perform all safety tests presented in this section.
Upon receipt of the device
Each time the main enclosure is disassembled or a circuit
board is removed, repaired, tested, or replaced.
5.1 Test Equipment
The recommended test equipment required to perform electrical safety tests is listed below.
Fluke 180 Electrical Safety Analyzer or compatible.
5.2 Ground Wire Leakage Current Tests
Leakage-HI:
300.00 uA Probe: Ground To Line
Leakage-LO:
0.00 uA Neutral : Closed
Voltage HI:
120.0V
Voltage LO:
100.0V
Reverse: Off Ground: Open
5.2.1 Connect the power cord of the monitor to the power receptacle
on the leakage tester.
5.2.2 Turn the monitor being tested ON.
5.2.3 Read the current leakage on the Electrical Safety Analyzer.
5.2.4 Set the polarity on the Electrical Safety Analyzer to Reverse.
Leakage-HI:
300.00 uA Probe: Ground To Line
Leakage-LO:
0.00 uA Neutral : Closed
Voltage HI:
120.0V
Voltage LO:
100.0V
Reverse: On Ground: Open
5.2.5 Read the current leakage on the Electrical Safety Analyzer.
NOTE: The readings should not exceed 300 µA or 0.3 volts at 100-120V
/50-60Hz on the Electrical Safety Analyzer with the monitor powered on.
5.2.6 Turn the monitor off.
5-1
Rad-87 Signal Extraction Pulse CO-Oximeter Service Manual
Page 20
5.3 Enclosure Leakage Current Tests
Leakage-HI:
300.00 uA Probe: Ground To Line
Leakage-LO:
0.00 uA Neutral : Closed
Voltage HI:
120.0V
Voltage LO:
100.0V
Reverse: Off Ground: Open
5.3.1 Connect the probe of the Electrical Safety Analyzer to the
Equipotential lug on the back of the monitor.
5.3.2 Set the Ground switch to the Open position on the Electrical
Safety Analyzer.
5.3.3 Turn the monitor on.
5.3.4 Read the current leakage on the Electrical Safety Analyzer.
5.3.5 Set the polarity on the Electrical Safety Analyzer to Reverse.
5.3.6 Read the current leakage on the Electrical Safety Analyzer.
NOTE: The readings should not exceed 300 µA or 0.3 volts at 100-120V
/50-60Hz on the Electrical Safety Analyzer with the monitor powered on.
Leakage-HI:
100.00 uA Probe: Ground To Line
Leakage-LO:
0.00 uA Neutral : Closed
Voltage HI:
120.0V
Voltage LO:
100.0V
Reverse: Off Ground:
Closed
5.3.7 Set the Ground switch to the Closed position on the Electrical
Safety Analyzer.
5.3.8 Read the current leakage on the Electrical Safety Analyzer.
5.3.9 Set the polarity on the Electrical Safety Analyzer to Reverse.
5.3.10 Read the current leakage on the Electrical Safety Analyzer.
NOTE: The readings should not exceed 100 µA or 0.1 volts at 100-120V
/50-60Hz on the Electrical Safety Analyzer with the monitor powered on.
5.3.11 Turn the Monitor off.
5-2
Rad-87 Signal Extraction Pulse CO-Oximeter Service Manual
Page 21
5.4 Source Leakage Current Tests
Leakage-HI:
100.00 uA Probe: High To Line
Leakage-LO:
0.00 uA Neutral : Closed
Voltage HI:
120.0V
Voltage LO:
100.0V
Reverse: Off Ground: Open
5.4.1 Set the Ground switch to the Open position on the Electrical
Safety Analyzer.
5.4.2 Set the polarity on the Electrical Safety Analyzer to Normal.
5.4.3 Connect a patient cable with all the wires spliced together to the
Electrical Safety Analyzer.
5.4.4 Read the current leakage on the Electrical Safety Analyzer.
5.4.5 Set the polarity on the Electrical Safety Analyzer to Reverse.
5.4.6 Read the current leakage on the Electrical Safety Analyzer.
NOTE: The readings should not exceed 100 µA or 0.1 volts at 100-120V
/50-60Hz on the Electrical Safety Analyzer with the monitor powered on.
Leakage-HI:
100.00 uA Probe: High To Line
Leakage-LO:
0.00 uA Neutral : Closed
Voltage HI:
120.0V
Voltage LO:
100.0V
Reverse: Off Ground: Open
5.4.7 Set the Ground switch to the Closed position on the Electrical
Safety Analyzer.
5.4.8 Set the polarity on the Electrical Safety Analyzer to Normal.
5.4.9 Read the current leakage on the Electrical Safety Analyzer.
5.4.10 Set the polarity on the Electrical Safety Analyzer to Reverse.
5.4.11 Read the current leakage on the Electrical Safety Analyzer.
NOTE: The readings should not exceed 100 µA or 0.1 volts at 100-120V
/50-60Hz on the Electrical Safety Analyzer with the monitor powered on.
5.4.12 Turn the Monitor off.
5-3
Rad-87 Signal Extraction Pulse CO-Oximeter Service Manual
Page 22
Page 23
6 Repair
6.1 Safety Precautions: WARNING: DO NOT DISASSEMBLE OR ASSEMBLE THE UNIT WITH
THE AC POWER CORD ATTACHED.
WARNING: ENSURE THE UNIT HAS BEEN CLEANED PER THE
CLEANING INSTRUCTIONS.
WARNING: WHEN REMOVING OR INSTALLING ANY INTERNAL
ELECTRICAL COMPONENTS, BE SURE TO FIRST
DISCONNECT AT LEAST ONE TERMINAL OF THE
BATTERY. VERIFY RECONNECTION OF BATTERY
PRIOR TO REASSEMBLY OF UNIT FOR PATIENT
MONITORING.
6.2 General Procedures
Opening the Rad-8 Pulse Oximeter will void the Masimo warranty. Products that require repair while under warranty should always be returned to Masimo for warranty service.
6.2.1 Masimo or other qualified service personnel must perform
warranty repair and service. Do not use malfunctioning equipment until the cause of the malfunction has been identified and the unit has been repaired in accordance with the instructions set forth herein.
6.2.2 Inspect the unit for cosmetic damage prior to disassembly or
troubleshooting. External damage may be helpful in determining the root cause of the failure.
6.2.3 Note any markings or labeling on the unit that may have been
placed by the end user. Duplicate or transfer this information in the event the housing or ancillary components are replaced.
6.2.4 Make note of the unit’s serial number.
6.2.5 Make note of the software versions in the event the Rad-87
board or system boards are replaced. These boards require software updates to make them compatible with the end user’s version of software.
6-1
Rad-87 Signal Extraction Pulse CO-Oximeter Service Manual
Page 24
6.2.6 Contact Masimo Corporation using the information in Section 6
Alarm Condition But No
to research warranty history.
6.3 Troubleshooting
6.3.1 The troubleshooting procedures in this manual instruct the
technician to isolate the failure down to the sub-assembly level. The intention is to have the Unit Under Test (UUT) repaired with minimal steps. The Masimo SET Tester is needed to perform the Performance Verification in Section 1 upon completion of repairs.
NOTE: The Masimo SET Tester is not a calibrated device. It is to be used for
evaluation purposes only and not to determine if an instrument is “calibrated”.
6.3.2 Tools required are as follows:
Phillips head screwdriver Digital Multi-meter
6.3.3 The troubleshooting steps below are presented in order of the
most probable cause first. Proceed to the next step only if the unit continues to fail.
Symptom Cause and Corrective Action
Steps
Speaker Alarming (very
loud) and unit is not functioning or displaying.
Alarm LED does not
Illuminate
1. Unit must be in alarm state.
Probable Watchdog error.
1. Replace System board.
2. Replace User Interface board.
1. Replace User Interface board.
2. Replace Display board.
2. Audible alarm can be heard.
Audible Indication.
1. Replace Speaker.
2. Replace System board.
1. Unit is in alarm condition.
Display Missing Segments
1. Replace User Interface board.
2. Replace Display board.
Button Function Error
1. Any front panel button
1. Replace Keypad.
2. Replace User Interface board.
Affected.
2. Buttons do not activate Feature.
"No Sen", or “Sen Off”
1. Replace sensor.
6-2
Rad-87 Signal Extraction Pulse CO-Oximeter Service Manual
Page 25
Symptom Cause and Corrective Action
Steps
Displayed.
1. Cable is connected with Sensor attached.
2. Replace patient cable.
3. Replace flex cable.
4. Replace Pulse CO-Oximeter board.
Unit Will Not Turn Off
1. Replace Keypad.
2. Replace System board.
If unable to repair unit, then please arrange for
return to Masimo Corporation
refer to section 6.
6.4 Service Messages:
Upon detection of a system error, the main software alerts the user with an audible alarm indicator and displays the appropriate error code on the display. The MX-1 board is held in reset and no other runtime functionality is available to the user until the unit is power cycled or until power is completely removed and re-applied to the system (removal and/or replacement of battery).
System Error System
Error Code
MX-1 board Failure 1 MX-1 board Communication Failure 2 Board Host Failure 3 Speaker Failure 5 Vor Voltage Out of Range Error 11 Vbatt Voltage Out of Range Error 12 V5 Voltage Out of Range Error 13 Vcc Voltage Out of Range Error 14
6-3
Rad-87 Signal Extraction Pulse CO-Oximeter Service Manual
Page 26
Page 27
7 Returns
To return the unit for service, please follow the Return Procedure outlined in this section.
7.1 Product Decontamination
7.1.1 Please clean contaminated/dirty equipment before returning.
Make sure it is fully dry before packing the equipment.
7.1.2 Recommended decontamination procedure for equipment that
has the potential for being exposed to blood or other potentially infectious materials.
7.1.2.1 Put on Lab coat or other protective garment.
7.1.2.2 Put on gloves.
7.1.2.3 Place equipment on the designated decontamination table.
7.1.2.4 Apply cleaning agent (10% bleach to water solution) to paper
towel or use Cidex wipes.
7.1.2.5 Wipe down all aspects of the equipment.
7.1.2.6 Let the equipment air dry completely.
7.1.2.7 Remove gloves.
7.1.2.8 Move equipment to the area designated for equipment that has
been through the decontamination process.
7.1.2.9 Dispose of all contaminated products properly.
7.2 Return Transportation
7.2.1 Call Masimo “domestically” at (800)326-4890, option #2;
“internationally” at 949-297-7498, option #2 and ask for an RMA number (or Customer Feedback Number-CFB #)
7.2.2 Package the equipment securely – in the original shipping
container if possible – and enclose the following information and items:
7.2.2.1 A letter or email describing in detail any symptoms or difficulties
experienced with the Pulse CO-Oximeter. Please include the RMA number (CFB #) in the letter.
7.2.2.2 Warranty information – a copy of the invoice or other applicable
documentation must be included.
7.2.2.3 Purchase order number to cover repair if the instrument is not
under warranty, or for tracking purposes if the warranty is in effect.
7-1
Rad-87 Signal Extraction Pulse CO-Oximeter Service Manual
Page 28
7.2.2.4 Ship-to and bill-to information.
7.2.2.5 Person (name, telephone/Telex/fax number, email and country)
to contact for any questions about the repairs.
7.2.2.6 A document stating the Pulse CO-Oximeter has been
decontaminated for blood-borne pathogens.
7.2.3 Return Pulse CO-Oximeter to the following shipping address:
Masimo Corporation
40 Parker
Irvine, California
USA 92618
(Domestic Toll Free): 800-326-4890, option #2 (International): 949-297-7498, option #2
FAX: 949-297-7499
Email: tech@masimo.com
7-2
Rad-87 Signal Extraction Pulse CO-Oximeter Service Manual
Page 29
8 Drawings and Schematics
8-1
Rad-87 Signal Extraction Pulse CO-Oximeter Service Manual
Page 30
8-2
Rad-87 Signal Extraction Pulse CO-Oximeter Service Manual
Page 31
8-3
Rad-87 Signal Extraction Pulse CO-Oximeter Service Manual
Page 32
Item # Rad-87 Parts
1 KIT ENCLOSURE RAD-87 3 ASSY, CABLE, RS232 ADAPTER, KESTREL 4 CABLE ASSY, MICRO COAXIAL, PLG RIGHT BLK 5 ASSEMBLY, FLEX INTERCONNECT 6 LCD MODULE, 480 X 272, COLOR TFT 7 ASSEMBLY, PCB, USER INTERFACE 8 ASSEMBLY, PCB, ANTENNA
9 MX-1 BOARD 10 SYSTEM CIRCUIT BOARD 11 RADIO INTERFACE BOARD 12 WLAN MODULE 13 STUD, EQUIPOTENTIAL, POAG, 6/15 14 JACKSCREW, 3/16" HEX, #4-40 MALE-FEMALE, 15 BATTERY,LEAD-ACID 6V, 1.3 Ahr MIN 16 TAPE, FOAM, 5/8" WIDE, 1/16" THICK 17 NUT, HEX, M6, STAINLESS STEEL 18 NUT, #4, SMALL PATTERN, STAINLESS STEEL 19 NUT, #4, WITH TOOTH WASHER, #4-40, 18-8, STAINLESS STEEL 20 NUT, HEX, M2, 18-8SS 21 WASHER, M6, SERRATED, STAINLESS STEEL 22 WASHER, M6, FLAT, STAINLESS STEEL 23 #4 SPLIT LOCK WASHER, STAINLESS STEEL 24 WASHER, FLAT, No. 2 NYLON 25 WASHER, FLAT, .156 OD X .093 ID, NYLON 26 SCREW, PAN HEAD PHILLIPS, #2-56 X .312 L 27 SCREW, PAN HD PHILIPS, M2 x 18MM LG 18-8SS 28 SCREW, FLAT HEAD PHILLIPS, #2-56 X .187 29 SCREW, PAN HEAD PHILIPS SEMS, #4-40 X 5/ 30 SCREW, #4-40X 5/8", PHILLIPS, Pan Head 31 SCREW, #4-40 X.125", SHLDR.125" 32 SPACER, ROUND, 3/16” OD x 9/32” LG, ALUM 33 WASHER, FLAT, No. 2 STAINLESS TEEL
8-4
Item # Rad-87 Parts
34 STANDOFF, 3/16" HEX, MALE-FEMALE THREADED 35 STANDOFF, 1/4" ROUND, THREADED, FEMALE, 36 BRACKET-ANTENNA 37 LABEL, RAD-87 CONNECTOR 38 CABLE ASSY, MICRO COAXIAL, PLG RIGHT WHT 39 EJECTOR HEADER CAP, 13 PINS, ROHS 40 TAPE, FOAM, 5/8" WIDE, 1/16" THICK 41 FOAM, RUBBER, NEOPRENE, 0.50" X 0.1875" X 50.0" 42 LABEL,NO IMPLIED LICENSE 43 BRACKET, FAN, MOUNT 44 ASSEMBLY, FAN 45 CABLE TIE 46 MOUNT, ADHESIVE BACKED, NYLON 47 GASKET, SPEAKER
Rad-87 Signal Extraction Pulse CO-Oximeter Service Manual
Page 33
8-5
Rad-87 Signal Extraction Pulse CO-Oximeter Service Manual
Loading...