Masimo Rad-5, Rad-5v Operator's Manual

signal extraction pulse oximeters
OPERATOR’S MANUAL
TM
Rad-5/5v Signal Extraction Pulse Oximeter Operator’s Manual i
The Rad-5 and Rad-5v Operating Instructions intend to provide the necessary information for proper operation of all Rad-5 and Rad-5v pulse oximeter models. There may be information provided in this manual that is not relevant for your pulse oximetry system.
General knowledge of pulse oximetry and an understanding of the features and functions of the Rad-5 and Rad-5v Pulse Oximeter are prerequisites for proper use.
Do not operate the Rad-5 or Rad-5v Pulse Oximeter without completely reading and understand­ing these instructions.
NOTICE
Purchase or possession of this device does not carry any express or implied license to use with replacement parts which would, alone or in combination with this device, fall within the scope of one of the relating patents.
CAUTION:
FEDERAL LAW (U.S.) RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN.
For further information contact: Masimo Corporation 40 Parker Irvine, CA 92618 USA Tel.: 949-297-7000 Fax.: 949-297-7001 www.masimo.com
EU Authorized Representative for Masimo Corporation:
MDSS Burckhardtstr. 1
30163 Hannover, Germany Tel.: +49-511-62 62 86 30 Fax.: +49-511-62 62 86 33
MEDICAL ELECTRICAL EQUIPMENT WITH RESPECT TO ELECTRIC SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60601-1/CAN/CSA C22.2 No. 601.1
Covered by one or more of the following U.S. Patents: RE38,492, RE38,476, 6,850,787, 6,826,419, 6,816,741, 6,699,194, 6,684,090, 6,658,276, 6,654,624, 6,650,917, 6,643,530, 6,606,511, 6,501,975, 6,463,311, 6,430,525, 6,360,114, 6,263,222, 6,236,872, 6,229,856, 6,206,830, 6,157,850, 6,067,462, 6,011,986, 6,002,952, 5,919,134, 5,823,950, 5,769,785, 5,758,644, 5,685,299, 5,632,272, 5,490,505, 5,482,036, international equivalents, or one or more of the patents referenced at www.masimo.com/patents. Other patents pending.
© 2006 Masimo Corporation. All rights reserved. Masimo, SET, , Radical, Rad-5, Rad-5v, PRONTO, TrendCom, SatShare, Signal IQ, SIQ, APOD, and FastSat are trademarks of Masimo Corporation.
EC REP
The Rad-5/5v Handheld Pulse Oximeter’s software program is designed to minimize the pos­sibility of hazards from errors by following sound engineering design processes, Risk Analysis and Software Validation.
Explosion hazard. Do not use the Pulse Oximeter in the presence of flammable anes­thetics or other flammable substance in combination with air, oxygen-enriched environ­ments, or nitrous oxide.
The Pulse Oximeter is NOT intended for use as an apnea monitor.
A Pulse Oximeter should be considered an early warning device. As a trend towards patient hypoxemia is indicated, blood samples should be analyzed by laboratory instruments to completely understand the patient’s condition.
The Pulse Oximeter is to be operated by qualified personnel only.This manual, accessory directions for use, all precautionary information, and specifications should be read before use.
Electric shock hazard. Do not open the pulse oximeter cover except to replace the battery of the unit. Only a qualified operator may perform maintenance procedures specifically described in this manual. Refer servicing to Masimo for repair of this equipment.
As with all medical equipment, carefully route patient cabling to reduce the possibility of patient entanglement or strangulation.
Do not place the pulse oximeter or accessories in any position that might cause it to fall on the patient. Do not lift the pulse oximeter by the patient cable.
Interfering Substances: Carboxyhemoglobin may erroneously increase readings. The level of increase is approximately equal to the amount of carboxyhemoglobin present. Dyes, or any substance containing dyes, that change usual blood pigmentation may cause erroneous readings.
Severe anemia may cause erroneous SpO2readings.
Do not use the pulse oximeter or oximetry sensors during magnetic resonance imag­ing (MRI) scanning. Induced current could potentially cause burns. The pulse oximeter may affect the MRI image, and the MRI unit may affect the accuracy of the oximetry measurements.
If using pulse oximetry during full body irradiation, keep the sensor out of the irradia­tion field. If sensor is exposed to the irradiation, the reading might be inaccurate or the unit might read zero for the duration of the active irradiation period.
Always remove the sensor from the patient and completely disconnect the patient from the pulse oximeter before bathing the patient.
Do not place the pulse oximeter where the controls can be changed by the patient.
Do not place the pulse oximeter face against a surface. This will cause the alarm to be muffled.
Do not place the pulse oximeter on electrical equipment that may affect the pulse oximeter, preventing it from working properly.
Do not expose the pulse oximeter to excessive moisture such as direct exposure to rain. Excessive moisture can cause the pulse oximeter to perform inaccurately or fail.
ii Rad-5/5v Signal Extraction Pulse Oximeter Operator’s Manual Rad-5/5v Signal Extraction Pulse Oximeter Operator’s Manual iii
Do not place containers containing liquids on or near the pulse oximeter. Liquids spilled on the pulse oximeter may cause it to perform inaccurately or fail.
Failure of Operation - If the pulse oximeter fails any part of the setup procedures remove the pulse oximeter from operation until qualified service personnel have corrected the situa­tion.
Patient Safety - If a sensor is damaged in any way, discontinue use immediately.
The pulse oximeter can be used during defibrillation, but the readings may be inac­curate for up to 20 seconds.
This equipment has been tested and found to comply with the limits for medical devices to the IEC 60601-1-2, Medical Device Directive 93/42/EEC. These limits are designed to provide reasonable protection against harmful interference in a typical medical installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to other devices in the vicinity. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to other devices, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
Reorient or relocate the receiving device.
Increase the separation between the equipment.
Consult the manufacturer for help.
SAFETY INFORMATION, WARNINGS, CAUTIONS AND NOTES SAFETY INFORMATION, WARNINGS, CAUTIONS AND NOTES
SECTION 1 - OVERVIEW
About this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
Warnings, cautions and notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Features and Benefits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Indications for use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
Pulse Oximetry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
General Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
Principle of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
Functional vs. Fractional Saturation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
Measured vs. Calculated Saturation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
Masimo SET Signal Extraction Technology . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
Masimo SET Parallel Engines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-7
Masimo SET DST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-7
SECTION 2 - SYSTEM DESCRIPTION
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
Rad-5/5v front panel controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
Rad-5/5v rear panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
SECTION 3 - SETUP
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
Unpacking and inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
Preparation for monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
Power requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
Monitor setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
Initial Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
SECTION 4 - OPERATION
Basic operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1
General Setup and Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1
Default Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
Successful SpO
2
monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-4
Numeric Display - SpO
2
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-4
Masimo Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-4
Numeric Display - Pulse Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-5
Signal IQ and pulsebar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-6
Low Perfusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-7
Actions To Be Taken . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-7
Battery Level Indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-8
Low Battery Audible Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-8
Normal patient monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-9
Rad-5v front panel control operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-9
Rad-5 front panel control operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-10
Setup menu (Rad-5 only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-10
Menu Navigation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-10
Setup Menu Level 1 – Alarm Limits and Alarm Volume. . . . . . . . . . . . . . . . . . . .4-11
Setup Menu Level 2 – Averaging and Sensitivity . . . . . . . . . . . . . . . . . . . . . . . . . .4-11
table of contents
iv Rad-5/5v Signal Extraction Pulse Oximeter Operator’s Manual Rad-5/5v Signal Extraction Pulse Oximeter Operator’s Manual v
table of contents
Setup menu Level 3 - Trend Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-12
Setup Menu Level 4 - LED brightness and Factory Defaults . . . . . . . . . . . . . . . . .4-13
Trend Setup and Use (Rad-5 only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-14
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-14
TrendCom Utility Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-14
TrendCom Utility operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-14
Erasing Trend Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-15
Trend Data Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-15
Sample Trend Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-15
Special menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-16
Standard and Sleep mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-16
Sleep mode Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-16
SECTION 5 - ALARMS AND MESSAGES
Alarm Indication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-1
Alarm limits (Rad-5 only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-1
Alarm Silence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-2
Alarm Silenced Indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-2
Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-3
SECTION 6 - TROUBLESHOOTING
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-1
SECTION 7 - SPECIFICATIONS
Rad-5/5v specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-1
Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-1
Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-1
Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-1
Environmental . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-1
Physical characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-2
SECTION 8 - SENSORS & PATIENT CABLES
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-1
Masimo SpO
2
sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-1
Selecting a Masimo Set Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-1
Sensor Application Site . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-2
LNOP® Reusable Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-2
LNOP® Adhesive Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-3
LNOP® Specialty Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-3
LNCSTM Reusable Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-3
LNCSTM Adhesive Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-4
LNOPvTM Adhesive Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-4
Cleaning And Reuse Of Masimo Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-4
Reattachment of Single Use adhesive Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-4
Masimo SET patient cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-5
Cleaning And Reuse Of Masimo SET Patient Cables . . . . . . . . . . . . . . . . . . . . . . .8-5
SECTION 9 - SERVICE AND MAINTENANCE
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-1
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-1
Rad-5/5v Signal Extraction Pulse Oximeter Operator’s Manual 1-1
1
overview
About this Manual
This manual explains how to set up and use the Rad-5/5v Handheld Pulse Oximeter. Important safety information relating to general use of the Rad-5/5v Pulse Oximeter appears before this introduction. Other important safety information is located throughout the manual where appropriate.
Read the entire safety information section before you operate the monitor.
In addition to the safety section, this manual includes the following sections:
SECTION 1 OVERVIEW gives a general description of pulse oximetry.
SECTION 2 SYSTEM DESCRIPTION describes the Rad-5/5v Handheld Pulse
Oximeter system and its functions and features.
SECTION 3 SETUP describes how to setup the Rad-5/5v Handheld Pulse
Oximeter for use.
SECTION 4 OPERATION describes the operation of the Rad-5/5v Pulse
Oximetry system.
SECTION 5 ALARMS AND MESSAGES describes the alarm system
messages.
SECTION 6 TROUBLESHOOTING gives troubleshooting information.
SECTION 7 SPECIFICATIONS gives the detailed specifications of the
Rad-5/5v Handheld Pulse Oximeter.
SECTION 8 SENSORS AND PATIENT CABLES outlines how to use and
care for the Masimo SET LNOP and LNCS sensors and Masimo SET patient cables.
SECTION 9 SERVICE AND MAINTENANCE describes how to maintain,
service and obtain repair for the Rad-5/5v Handheld Pulse Oximeter.
SECTION 10 ACCESSORIES.
Battery Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-1
Performance verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-2
Service and repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-5
Repair Policy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-5
Return Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-5
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-6
Exclusions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-6
End-user license agreement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-6
SECTION 10 - ACCESSORIES
Rad-5 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-1
table of contents
vi Rad-5/5v Signal Extraction Pulse Oximeter Operator’s Manual
Rad-5/5v Signal Extraction Pulse Oximeter Operator’s Manual 1-3
1
overview
Product Description
The Rad-5 family of Handheld Pulse Oximeters are noninvasive, arterial oxygen saturation and pulse rate monitors. The Rad-5 family features a multicolored LED display that contin­uously displays numeric values for SpO
2
and pulse rate, as well as LED indicator bars for
Perfusion Index (PI) and Signal Identification and Quality Indicator (Signal IQ
TM
).
The Rad-5 family consists of two models:the full-featured Rad-5 and the Rad-5v entry-level spot checker. Both units are built on the same motion tolerant pulse oximetry technology, with the Rad-5 adding parameter alarming, three sensitivity settings and adjustable aver­aging times.
Features that apply to the Rad-5 only will be indicated with “(Rad-5)”.
FEATURES AND BENEFITS
These features are common to the Rad-5 family:
Clinically proven Masimo SET®technology performance
Applicable for use on neonate, pediatric and adult patients
Proven for accurate monitoring in motion and low perfusion environments
SpO2, pulse rate, alarm, and perfusion index displays
Signal I.Q.®for signal identification and quality indication
Lightweight, convenient handheld design
Long battery life: over 48 hours on 4 “AA” alkaline batteries
Audible Alarm for no sensor, sensor-off, interference detected and low battery
The Rad-5 adds these features:
Alarms for High/Low saturation and High/Low pulse rate
FastSat
®
User defineable alarm limit settings
Sleep study mode
Three sensitivity levels - Max, Normal and APOD
TM
Stores up to 72 hours of trending memory
Adjustable alarm volume
Adjustable averaging 2 to 16 seconds
Warnings, cautions and notes
Please read and follow any warnings, cautions and notes presented throughout this manual. An explanation of these labels are as follows:
A WARNING is provided when actions may result in a serious outcome (i.e., injury, serious adverse affect, death) to the patient or user. Look for text in a gray shaded box.
Sample of Warning:
A CAUTION is given when any special care is to be exercised by the patient or user to avoid injury to the patient, damage to this device or damage to other property.
Sample of Caution:
CAUTION: THIS IS A SAMPLE OF A CAUTION STATEMENT.
A
NOTE
is provided when extra general infor mation is applicable.
Sample of Note:
NOTE: This is a sample of a Note.
WARNING: THIS IS A SAMPLE OF A WARNING STATEMENT.
overview
1-2 Rad-5/5v Signal Extraction Pulse Oximeter Operator’s Manual
1
Rad-5/5v Signal Extraction Pulse Oximeter Operator’s Manual 1-5
1
overview
Pulse Oximetry
GENERAL DESCRIPTION
Pulse oximetry is a continuous and non-invasive method of measuring the level of arterial oxygen saturation in blood. The measurement is taken by placing a sensor on a patient, usually on the fingertip for adults, and the hand or foot for neonates. The sensor connects to the pulse oximetry instrument with a patient cable. The sensor collects signal data from the patient and sends it to the instrument. The instrument displays the calculated data in two ways:1) as a percent value for arterial oxygen saturation (SpO
2
), and 2) as a pulse rate
(PR). The following figure shows the general monitoring setup.
PRINCIPLE OF OPERATION
Pulse oximetry is governed by the following principles:
1. Oxyhemoglobin (oxygenated blood) and deoxyhemoglobin (non-oxygenated blood) differ in their absorption of red and infrared light (spectrophotometry).
2. The amount of arterial blood in tissue changes with your pulse (photoplethysogra- phy). Therefore, the amount of light absorbed by the varying quantities of arterial blood changes as well.
The Rad-5/5v Handheld Pulse Oximeter uses a two-wavelength pulsatile system to distin­guish between oxygenated and deoxygenated blood. Signal data is obtained by passing red (rd) (660 nm wavelength) and infrared (ir) (905 nm wavelength) light through a capil­lary bed (for example a fingertip, a hand, a foot) and measuring changes in light absorp­tion during the pulsatile cycle. This information may be useful to clinicians. The radiant power of the light is rated at 0.79mW (max.). See figure below. The Rad-5/5v utilizes a sen­sor with red and infrared light-emitting diodes (LEDs) that pass light through the site to a photodiode (photodetector). The photodetector receives the light, converts it into an elec­tronic signal and sends it to the Rad-5/5v for calculation.
Once the Rad-5/5v receives the signal from the sensor, it utilizes Masimo SET signal extraction technology for calculation of the patient’s functional oxygen saturation and pulse rate.
1
2
3
1. Instrument
2. Patient Cable
3. Sensor
2
1
1. Light Emitting Diodes (LEDs)
2. Recessed Photo Detector
INDICATIONS FOR USE
The Rad-5 family of Handheld Pulse Oximeters and accessories are indicated for the con­tinuous (Rad-5 only) or spotcheck, noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO
2
) and pulse rate (measured by an SpO2sensor). The Rad-5 family of Handheld Pulse Oximeters and accessories are indicated for use with adult, pedi­atric and neonatal patients during both motion and no motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile and home envi­ronments.
overview
1-4 Rad-5/5v Signal Extraction Pulse Oximeter Operator’s Manual
1
1-7
1
overview
Rad-5/5v Signal Extraction Pulse Oximeter Operator’s Manual
MASIMO SET DST
Digitized, Filtered &
Normalized
R/IR
R/IR
(Conventional
Pulse Oximetry)
Adaptive Filter with
DST
TM
SST
TM
Proprietary Algorithm 5
Saturation
Post
Processor
Confidence
Based
Arbitrator
MEAS
CONF
MEAS
CONF
MEAS
CONF
MEAS
CONF
MEAS
CONF
Proprietary
Algorithm 4
MASIMO SET PARALLEL ENGINES
1-6 Rad-5/5v Signal Extraction Pulse Oximeter Operator’s Manual
1
overview
FUNCTIONAL VS. FRACTIONAL SATURATION
The Rad-5/5v is calibrated to measure and display functional saturation:the amount of oxy­genated hemoglobin expressed as a percentage of the hemoglobin that can transport oxy­gen. The Rad-5/5v does not measure fractional saturation: oxygenated hemoglobin expressed as a percentage of all measured hemoglobin, including measured dysfunction­al hemoglobin such as carboxyhemoglobin or methemoglobin. To conver t fractional satu­ration to functional saturation, the fractional saturation measurements must be conver ted according to:
Functional saturation =
Fractional saturation
x 100
100 - (% carboxyhemoglobin + % methemoglobin)
MEASURED VS. CALCULATED SATURATION
Oxygen saturation measurements obtained from a pulse oximeter are commonly com­pared to saturations calculated from the partial pressure of oxygen (PO
2
) obtained from an arterial blood gas sample. When comparing the two measurements and interpreting val­ues, caution should be used, as the calculated value obtained from the blood gas sample may differ from the SpO
2
measurement of the pulse oximeter. Different results are usual­ly obtained from the blood gas sample if the calculated saturation is not appropriately cor­rected for the effects of variables that shift the relationship between PO
2
and saturation,
such as: pH, temperature, the partial pressure of carbon dioxide (PCO
2
), 2,3-DPG, and fetal hemoglobin. Also, as blood gas samples are usually taken over a period of 20 sec­onds (the time it takes to draw blood) a meaningful comparison can only be achieved if the core oxygen saturation of the patient is stable and not changing over the period of time that the blood gas sample is taken.
MASIMO SET SIGNAL EXTRACTION TECHNOLOGY
Masimo Signal Extraction Technology’s signal processing differs from conventional pulse oximeters. Conventional pulse oximeters assume that arterial blood is the only blood mov­ing (pulsating) in the measurement site. During patient motion, however, the non-arterial blood also moves, causing conventional pulse oximeters to read low values, because they cannot distinguish between the arterial and venous blood movement (sometimes referred to as noise). Masimo SET pulse oximetry utilizes parallel engines and adaptive digital fil­tering. Adaptive filters are powerful because they are able to adapt to the varying physio­logic signals and/or noise and separate them by looking at the whole signal and breaking it down to its fundamental components. The Masimo SET signal processing algorithm, Discrete Saturation Transform
TM
(DST), reliably identifies the noise, isolates it and, using adaptive filters, cancels it. It then reports the true arterial oxygen saturation for display on the monitor.
Rad-5/5v Signal Extraction Pulse Oximeter Operator’s Manual 2-1
2
system description
Introduction
The Rad-5 family of Handheld Pulse Oximeters are full featured pulse oximeters designed for ease of operation. All pulse oximetry measurement information, as well as device sta­tus data, is displayed on the front panel of the device. All user input is handled by control buttons on the front panel and the sensor cable connection is located at the top edge of the device.
The Rad-5/5v are powered by 4 “AA”alkaline batteries, which provide over 48 hours of bat­tery life.
Rad-5 family offers full Masimo SET technology in a small, handheld device
Rad-5 family supports the full line of Masimo sensors and patient cables (see Section 8,
sensors and patient cables
).
Rad-5 family supports standardization of sensors and pulse oximetry technology throughout the hospital
Rad-5v provides essential pulse oximetry features
Rad-5 includes all the features of the Rad-5v, plus
High/Low saturation alarms
High/Low pulse rate alarms
User defineable alarm limit settings
Sleep study mode
Stores up to 72 hours of trending memory
Adjustable averaging time
Three sensitivity levels - Max, Normal and APOD
FastSat
A LNOP DCSC Spot Check Sensor or Masimo Patient Cable and Masimo sensor attach to the connector on the top of the Rad-5/5v unit. The Rad-5/5v can be used either for spot checks or continuous (Rad-5) SpO2 monitoring.
system description
2-2 Rad-5/5v Signal Extraction Pulse Oximeter Operator’s Manual Rad-5/5v Signal Extraction Pulse Oximeter Operator’s Manual 2-3
2
2
system description
CONTROL / INDICATOR DESCRIPTION
1
Patient Cable Connector
Connects to LNOP DCSC Spot Check sensor or Masimo Patient Cable
2
FastSat Indicator
Illuminates when the FastSat mode has been enabled.
FastSat enables rapid tracking of arterial oxygen saturation changes
3
Saturation Display
The functional arterial hemoglobin oxygen saturation is dis­played in units of SpO2. When searching for a saturation and pulse, it will flash dashed lines.
4
Signal IQ / Pulse Bar
The Signal IQ provides an indication of the quality of the acquired signal as well as the timing of the pulse. A green vertical LED bar rises and falls with the pulse, where the height of the bar indicates the quality of the signal.
5
Pulse Rate Display
The pulse rate in beats per minute (bpm). When searching for a saturation and pulse, it will flash dashed lines.
6
Mode / Enter Button
Used to enter the setup menus and to select/activate certain entries within the menu/setup system.
7
Next Button
Used within the menu/setup system to move through setup options. Not active during normal patient monitoring.
CONTROL / INDICATOR DESCRIPTION
8
Power On / Off
Used to turn the unit on and off.
9
Battery Level Indicator
Four LED’s indicate the status of the battery. When the final indi­cator begins flashing, replace the batteries.
10
Visual Alarm Indicator
Illuminates when any alarm condition exists. This indicator may not be turned-off or otherwise over-ridden.
11
Perfusion Index
The Perfusion Index provides an indication of the percentage of pulsatile signal to non pulsatile signal. The bar is highest when the quality of the perfused site is best.
12
MAX Sensitivity Indicator
Illuminates when the MAX Sensitivity mode has been enabled.
Note:
When using the Maximum Sensitivity setting, the SENSOR
OFF detection performance may be compromised.
13
Alarm Silenced Indicator
Flashes to indicate the alarm is temporarily silenced (by pushing the Alarm Silence Button once) or is illuminated solid to indicate the alarms have been permanently muted (by pushing the Alarm Silence Button twice) .
14
Alarm Silence Button
Push once to temporarily silence the alarm for 120 seconds. Push a second time to permanently mute (suspend) the alarm. A third push will return the unit to standard alarm monitoring.
15
Up button Down button
During saturation monitoring, use these buttons to adjust the vol­ume of the pulse beep tone.
Within the menu/setup system, these buttons are used to select values within each menu option.
17
Speaker
Provides audible indication of alarm conditions, pulse tone and feedback for key-presses. Ensure the speaker is not covered or the unit is placed face-down on bedding or other sound absorbing surface.
18
Pulse Tone Volume
Provides control of the pulse tone volume. Cycles through three volume levels, and mute. At the loudest level, pressing the Pulse Tone Volume button will return the volume to mute.
19
Display Brightness
Provides control of the front panel indicator brightness. Cycles through four brightness levels. At the brightest level, pressing the Display Brightness button will return the display to the lowest brightness setting.
1
MAX
2
4
5
8
9
10
11
13
14
15
16
3
7
6
12
17
3
4
5
8
9
1
10
11
13
14
18
19
17
Rad-5
Rad-5v
Rad-5/5v front panel controls
16
Loading...
+ 22 hidden pages