The Rad-5/5v Service Manual is intended to provide the necessary
information for proper servicing of all models of the Rad-5 class of pulse
oximetry systems. This manual is structured to support troubleshooting
to the assembly or module level. This manual does not provide
instructions for troubleshooting to the printed circuit board component
level. There may be information provided in this manual that is not
relevant for your system. Do not service the Rad-5 class of pulse
oximeters without completely reading and understanding these
instructions. Service training provided by Masimo is required prior to
utilization of this manual.
CAUTION:
Federal law (U.S.) restricts this device to sale by or on the order of a physician
Masimo Corporation
40 Parker
Irvine, CA 92618
USA
Tel: 949-297-7000
Fax:949-297-7001
www.masimo.com
Covered by one or more of the following U.S. Patents: RE38,492, RE38,476,
6,699,194, 6,684,090, 6,658,276, 6,654,624, 6,650,917, 6,643,530, 6,606,511,
6,584,336, 6,501,975, 6,463,311, 6,430,525, 6,360,114, 6,263,222, 6,236,872,
6,229,856, 6,206,830, 6,157,850, 6,067,462, 6,011,986, 6,002,952, 5,919,134,
5,823,950, 5,769,785, 5,758,644, 5,685,299, 5,632,272, 5,490,505, 5,482,036,
international equivalents, or one or more of the patents referenced at
www.masimo.com/patents.
Other patents pending.
Radical Signal Extraction Pulse Oximeter Service Manual
Radical Signal Extraction Pulse Oximeter Service Manual
UM-10026 –1204
1. Overview
1.1. About this Manual
This manual explains how to service the Rad-5 class of pulse oximeters.
However, when a differentiation between models is required, the actual
model type will be identified prior to the explanation (For example, the
difference between a Rad-5 or Rad-5v). Important safety information
relating to general use of the Rad-5 class of pulse oximeters appears in
the Rad-5 Operator’s Manual. Other important safety information is
located throughout this manual where appropriate.
1.2. Warnings, Cautions and Notes
Please read and follow any warnings, cautions and notes presented
throughout this manual. An explanation of these labels is as follows:
A warning is provided when actions may result in a serious outcome
(i.e., injury, serious adverse affect, death) to the patient or user. Look for
text in a gray shaded box.
Sample of Warning:
WARNING: THIS IS A SAMPLE OF A WARNING STATEMENT.
A caution is given when any special care is to be exercised by the
patient or user to avoid injury to the patient, damage to this device or
damage to other property.
Sample of Caution:
CAUTION: This is a sample of a Caution statement.
A note is provided when extra general information is applicable.
Sample of Note:
Note: This is a sample of a Note statement.
Rad-5 Signal Extraction Pulse Oximeter Service Manual
UM-10026 – 1204
1-1
2. Maintenance
2.1. Introduction
2.1.1. This chapter covers, how to test the operation of the Rad-5
oximeters, how to properly clean the Rad-5 pulse oximeters,
how to replace the batteries and how to obtain service under
normal operation. No internal adjustment or recalibration is
required. Service should be done by qualified personnel only.
Safety checks should be performed at regular intervals or in
accordance with local and governmental regulations.
WARNING: BEFORE CLEANING THE OXIMETER, TURN
THE UNIT OFF.
2.2. Cleaning
2.2.1. To clean the display panel, use a cotton swab moistened with
70% isopropyl alcohol and gently wipe the panel.
2.2.2. To clean the outer surface of the oximeter, use a soft cloth
moistened with a 70% isopropyl alcohol. Do not allow liquids to
enter the interior of the instrument.
2.2.3. To decontaminate unit, refer to Section 6.2.2.
CAUTIONS:
• Do not autoclave, pressure sterilize, or gas sterilize this oximeter.
• Do not soak or immerse the monitor in any liquid.
• Use the cleaning solution sparingly. Allow unit to air dry. Excessive
solution can flow into the monitor and cause damage to internal
components.
•Do not touch, press, or rub the display panels with abrasive cleaning
compounds, instruments, brushes, rough-surface materials, or bring
them into contact with anything that could scratch the panel.
•Do not use petroleum-based or acetone solutions, or other harsh
solvents, to clean the oximeter. These substances attack the
device’s materials and device failure can result.
2-1
Rad-5 Signal Extraction Pulse Oximeter Service Manual
UM-10026 – 1204