Masimo NomoLine ISA CO2 Operator's Manual

NomoLine
ISA™ CO2 Sidestream Gas Analyzer
®
Operator's Manual
These operating instructions intend to provide the necessary information for proper operation of the NomoLine® ISA™ CO2 (NomoLine ISA CO2). General knowledge of capnography monitoring and an understanding of the features and functions of NomoLine ISA CO2 are prerequisites for proper use. Do not operate NomoLine ISA CO2 without completely reading and understanding these instructions.
Notice: Purchase or possession of this device does not carry any express or implied license to use with replacement parts which would, alone or in combination with this device, fall within the scope of one of the relating patents.
Note: Cleared Use Only: The device and related accessories are cleared by the Food and Drug Administration (FDA) and are CE Marked for noninvasive patient monitoring and may not be used for any processes, procedures, experiments, or any other use for which the device is not intended or cleared by the applicable regulatory authorities, or in any manner inconsistent with the directions for use or labeling.
CAUTION: Federal (USA) law restricts this device to sale by or on the order of a physician. See instructions for use for full prescribing information, including indications, contraindications, warnings and precautions.
For professional use. See instructions for use for full prescribing information, including indications, contraindications, warnings, and precautions.
Masimo Corporation 52 Discovery Irvine, CA 92618, USA Tel.: 949-297-7000 Fax.: 949-297-7001 www.masimo.com
EU authorized representative for Masimo Corporation:
MDSS GmbH Schiffgraben 41 D-30175 Hannover, Germany
MEDICAL ELECTRICAL EQUIPMENT
WITH RESPECT TO ELECTRIC SHOCK, FIRE AND MECHANICAL HAZARDS ONLY
IN ACCORDANCE WITH
3149433
ANSI/AAMI ES60601-1:2005/A1:2012, CAN/CSA C22.2 No. 60601-1:2014, and
applicable Particular (EN/ISO 80601-2-55:2018) Standards for which the
product has been found to comply by Intertek.
Patents: www.masimo.com/patents.htm
Masimo®, ®, and NomoLine® are registered trademarks of Masimo Corporation.
ISA™ is a trademark of Masimo Corporation.
All other trademarks and registered trademarks are property of their respective owners.
© 2019 Masimo Corporation
www.masimo.com 1 Masimo
Contents
About This Manual------------------------------------------------------------------------------------------ 5
Product Description, Features and Indications for Use ------------------------------------------------7
Product Description --------------------------------------------------------------------------------------7
Intended Use ---------------------------------------------------------------------------------------------7
Safety Information, Warnings and Cautions ----------------------------------------------------------- 9
Safety Warnings and Cautions ------------------------------------------------------------------------ 9
Performance Warnings and Cautions --------------------------------------------------------------- 10
Cleaning and Service Warnings and Cautions ---------------------------------------------------- 10
Compliance Warnings and Cautions ---------------------------------------------------------------- 11
Chapter 1: Description ------------------------------------------------------------------------------------ 13
General System Description -------------------------------------------------------------------------- 13
Features -------------------------------------------------------------------------------------------------- 13
NomoLine Sampling Lines --------------------------------------------------------------------------- 14
Chapter 2: Setting Up ------------------------------------------------------------------------------------- 15
System Setup ------------------------------------------------------------------------------------------- 15
Chapter 3: Operation -------------------------------------------------------------------------------------- 17
LEGI Indicator ------------------------------------------------------------------------------------------ 17
Capnography Display ---------------------------------------------------------------------------------- 17
Chapter 4: Troubleshooting ------------------------------------------------------------------------------ 19
Troubleshooting NomoLine ISA CO2 --------------------------------------------------------------- 19
Chapter 5: Specifications --------------------------------------------------------------------------------- 21
Accuracy ------------------------------------------------------------------------------------------------- 21
Environmental ------------------------------------------------------------------------------------------ 21
Physical Characteristics ------------------------------------------------------------------------------- 22
Electrical ------------------------------------------------------------------------------------------------- 22
Compliance ---------------------------------------------------------------------------------------------- 22
Additional Specifications ----------------------------------------------------------------------------- 23
Symbols -------------------------------------------------------------------------------------------------- 26
Chapter 6: Service and Maintenance------------------------------------------------------------------- 29
Cleaning ------------------------------------------------------------------------------------------------- 29
Maintenance -------------------------------------------------------------------------------------------- 29
www.masimo.com 3 Masimo
NomoLine ISA CO2 Contents
Repair Policy ------------------------------------------------------------------------------------------- 30
Return Procedure ------------------------------------------------------------------------------------- 30
Sales & End-User License Agreement ------------------------------------------------------------- 30
Limited Warranty -------------------------------------------------------------------------------------- 31
Exclusions ----------------------------------------------------------------------------------------------- 31
Limitation of Warranty -------------------------------------------------------------------------------- 31
End-User License -------------------------------------------------------------------------------------- 32
No Implied License ------------------------------------------------------------------------------------ 33
Index --------------------------------------------------------------------------------------------------------- 35
www.masimo.com 4 Masimo

About This Manual

This manual explains how to set up and use NomoLine® ISA™ CO2. Important safety information relating to general use of NomoLine ISA CO2 appears in this manual. Read and follow any warnings, cautions, and notes presented throughout this manual. The following are explanations of warnings, cautions, and notes.
A warning is given when actions may result in a serious outcome (for example, injury, serious adverse effect, death) to the patient or user.
WARNING: This is an example of a warning statement.
A caution is given when any special care is to be exercised by the patient or user to avoid injury to the patient, damage to this device, or damage to other property.
CAUTION: This is an example of a caution statement.
A note is given when additional general information is applicable.
Note: This is an example of a note.
www.masimo.com 5 Masimo

Product Description, Features and Indications for Use

Product Description

NomoLine® ISA™ CO2 is part of the ISA product family and is a sidestream gas analyzer intended to be connected to other medical backboard devices for monitoring of breath rate and carbon dioxide CO2 gas concentration. NomoLine ISA CO2 supports various types of host connectors for providing measured data.

Intended Use

NomoLine® ISA™ CO2 sidestream gas analyzer and accessories including NomoLine, is intended to be connected to other medical backboard devices for monitoring of breath rate and the following breathing gases:
CO
2
NomoLine ISA CO2 is intended to be connected to a patient breathing circuit for monitoring of inspired/expired gases during anesthesia, recovery and respiratory care. The intended environment is the operating suite, intensive care unit and patient room. NomoLine ISA CO2 is also intended to be used in road ambulances. The intended patient population is adult, pediatric, infant and neonatal patients.
www.masimo.com 7 Masimo

Safety Information, Warnings and Cautions

CAUTION: NomoLine ISA CO2 is to be operated by, or under the supervision of, qualified personnel only. Read this Operator's Manual, accessories directions for use, all precautionary information, and specifications before use. Refer to the host device operator's manual or user's guide for additional safety information, warnings and cautions.

Safety Warnings and Cautions

WARNING: Do not use NomoLine ISA CO2 if it appears or is suspected to be damaged. Damage to the device can result in exposed electrical circuits that may cause patient harm.
WARNING: Do not adjust, repair, open, disassemble, or modify the NomoLine ISA CO2. Damage to the device may result in degraded performance and/or patient injury.
WARNING: Do not start or operate the NomoLine ISA CO2 unless the setup was verified to be correct. Improper set-up of this device may result in degraded performance and/or patient injury.
WARNING: Do not place the NomoLine ISA CO2 or accessories in any position that might cause it to fall on the patient.
WARNING: Only use Masimo authorized devices with NomoLine ISA CO2. Using unauthorized devices with NomoLine ISA CO2 may result in damage to the device and/or patient injury.
WARNING: Carefully route the sampling line to reduce the risk of patient entanglement or strangulation.
WARNING: Do not lift the NomoLine ISA CO2 by the NomoLine capnography sampling line as it could disconnect from the NomoLine ISA CO2, causing the device to fall on the patient.
WARNING: Do not use the NomoLine ISA CO2 during magnetic resonance imaging (MRI) or in an MRI environment.
WARNING: Only use sample lines intended for anesthetic agents if N agents are being used.
WARNING: Do not re-use disposable single-patient use NomoLine Family sampling lines due to the risk of cross contamination.
WARNING: Do not use the NomoLine Infant/Neonate Airway Adapter Sets for adults/pediatrics as they may cause excessive flow resistance (0,7 ml dead space).
WARNING: Do not use the NomoLine Adult/Pediatric Airway Adapter Sets for infants/neonates as the airway adapter adds 6 ml dead space.
WARNING: Do not apply negative pressure to remove condensed water from the NomoLine Family sampling line.
WARNING: NomoLine ISA CO2 is not intended to be used for returning exhaust gases to the patient circuit. Exhaust gases should be returned to a scavenging system.
Note: Disconnect the device from AC mains by removing the device cable connection from the host device.
O and/or anesthetic
2
www.masimo.com 9 Masimo
NomoLine ISA CO2 Safety Information, Warnings and Cautions
Note: Use and store the NomoLine ISA CO2 in accordance with specifications. See the Specifications section in this manual.

Performance Warnings and Cautions

WARNING: NomoLine ISA CO2 should not be used as the sole basis for medical decisions. It must be used in conjunction with clinical signs and symptoms.
WARNING: Use of high-frequency electrosurgical equipment in the vicinity of NomoLine ISA CO2 may produce interference and cause incorrect measurements.
WARNING: Do not use the NomoLine ISA CO2 with metered-dose inhalers or nebulized medications as this may clog the bacteria filter.
WARNING: Properly apply sampling lines according to the sampling lines directions for use. Misapplied sampling lines that become partially dislodged may cause no or incorrect readings.
WARNING: Replace the sampling line if the sampling line input connector starts flashing red, or host device displays a check sampling line type of message.
WARNING: Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating properly.
WARNING: Portable and mobile RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30cm (12 inches) to any part of the NomoLine ISA CO2 including the cable. Otherwise, degradation of the performance of the NomoLine ISA CO2 could result.
WARNING: Too strong positive or negative pressure in the patient circuit might affect the sample flow.
WARNING: Strong scavenging suction pressure might affect the sample flow.
CAUTION: Do not operate NomoLine ISA CO2 outside of the specific operating environment.
CAUTION: NomoLine ISA CO2 should be mounted securely to avoid risk of damage to the
NomoLine ISA CO2.

Cleaning and Service Warnings and Cautions

WARNING: To avoid electric shock, always physically disconnect the NomoLine ISA CO2 and all patient connections before cleaning.
WARNING: Do not attempt to remanufacture, recondition or recycle the NomoLine ISA CO2 as these processes may damage the electrical components, potentially leading to patient harm.
CAUTION: Do not sterilize or immerse NomoLine Family sampling lines in liquid.
CAUTION: To avoid permanent damage to the NomoLine ISA CO2, do not use undiluted
bleach (5% - 5.25% sodium hypochlorite) or any other cleaning solution not recommended.
www.masimo.com 10 Masimo