Masimo MSAT01A User Manual

Masimo Corporation

MightySat Fingertip Pulse Oximeter

changes in PI over a time interval where one or more complete respiratory cycles have occurred.
Operator's Manual
For Sale in the US A- Not Intended for Medical Use Notice: Purchase or possession of this device does not carry any express or implied license to use with
replacement parts which would, alone or in combination with this device, fall within the scope of one of the relating patents.

About this Manual

Do not operate the MightySat Fingertip Pulse Oximeter without completely reading and understanding the instructions.
Always use the MightySat precisely in accordance with the directions in this manual, including site selection and sensor placement. Failure to follow all of the directions in this manual could lead to inaccurate measurements.
Read and follow any warnings, cautions, and notes presented throughout this manual. The following are explanations of warnings, cautions, and notes.
A warning is given when actions may result in a serious outcome to the user (for example, injury, serious adverse effect, or death).
Warning: This is an example of a warning statement. A caution is given when any special care is to be exercised by the user to avoid injury to the user, damage to
this instrument, or damage to other property. Caution: This is an example of a caution statement. A note is given when additional general information is applicable. Note: This is an example of a note.

Product Description and Intended Use

Product Description

The MightySat Fingertip Pulse Oximeter is a noninvasive device that measures and displays arterial oxygen saturation (SpO2), Pulse Rate (PR), Perfusion Index (PI), and Pleth Variability Index (PVI®). The MightySat is available for people who weigh more than 30 kg (66 lbs).
The MightySat Fingertip Pulse Oximeter is available in the following models:
Model Product Features
9700 MightySat Measures and displays arterial oxygen saturation (SpO2), Pulse Rate (PR), and
9800 MightySat, Bluetooth Measures and displays arterial oxygen saturation (SpO2), Pulse Rate (PR), and
9900 MightySat, Bluetooth and PVI Measures and displays arterial oxygen saturation (SpO2), Pulse Rate (PR),

Intended Use

The MightySat Fingertip Pulse Oximeter by Masimo is for use by sports and aviation users who are interested in knowing their blood oxygenation level (SpO2), pulse rate (PR) and perfusion index (PI). The MightySat is not intended for medical use. The MightySat can be used in a wide range of settings, including extreme sports, and for passengers and pilots.

Safety Information, Warnings, and Cautions

Safety Warnings and Cautions

Warning: Do not use MightySat if it appears or is suspected to be damaged.
Warning: Do not repair, open, or modify MightySat. Injury or equipment damage could occur.
Warning: Do not use MightySat during magnetic resonance imaging (MRI) or in an MRI
environment.
Warning: Do not expose MightySat components to excessive moisture such as direct exposure to
rain. Excessive moisture can cause the components to perform inaccurately or fail.
Warning: Do not place MightySat or accessories in any position that might cause it to fall on the
user.
Warning: Electric Shock Hazard: Do not use MightySat during defibrillation.
Warning: Do not to use device during electrosurgery.
Warning: Do not use MightySat in the presence of flammable anesthetics or other flammable
substance in combination with air, oxygen-enriched environments, or nitrous oxide.
Warning: Do not use tape to secure MightySat to the site; this can restrict blood flow and cause
inaccurate readings. Use of tape can cause skin damage, and/or pressure necrosis or damage the sensor.
Warning: Keep the oximeter away from young children. Small items such as the battery door and
battery may become choking hazards.
Warning: When MightySat is applied for extended periods of time, the sensor site must be checked
frequently to ensure adequate circulation, skin integrity, and correct optical alignment.
Warning: If MightySat is applied too tightly or becomes tight due to edema it will cause inaccurate
readings and can cause pressure necrosis.
Warning: Do not use the lanyard with MightySat during activities where a risk of strangulation may
exist (i.e. sleeping).
Caution: Do not attempt to reprocess, recondition or recycle MightySat as these processes may
damage the electrical components, potentially leading to user harm.
Caution: Before changing batteries, make sure the device is off and is not applied to a finger.
Caution: MightySat contains a magnet which may interfere with other devices, including
pacemakers.
Note: The maximum skin surface temperature is measured to be less than 41°C (106°F) in a
minimum 35°C (95°F) environment. This was verified by measuring the skin interface temperature with MightySat operating under reasonable worst-case conditions.

Performance Warnings and Cautions

Warning: MightySat is not intended for medical use.
Warning: MightySat is not an apnea monitor and should not be used for arrhythmia analysis.
Warning: Do not self-diagnose or self-medicate on the basis of the measurements. Always consult
your doctor. MightySat is not intended for medical use.
Warning: MightySat does not include visual alerts or audible alarm features to support continuous
monitoring. It is intended for spot-check use only.
Warning: SpO2 is empirically calibrated in healthy adult volunteers with normal levels of
carboxyhemoglobin (COHb) and methemoglobin (MetHb).
Warning: Inaccurate SpO2 readings may be caused by:
Improper MightySat placement or alignment
Elevated levels of COHb and MetHb: High levels of COHb or MetHb may occur with a
seemingly normal SpO2. When elevated levels of COHb or MetHb are suspected, laboratory analysis (CO-Oximetry) of a blood sample should be performed.
Intravascular dyes such as indocyanine green or methylene blue
Externally applied coloring and texture such as nail polish, acrylic nails, glitter, etc.
Users attached to a high pressure cuff
Avoid placing the MightySat sensor on any extremity with an arterial catheter or blood
pressure cuff.
Elevated levels of bilirubin
Severe anemia
Venous congestion
Venous pulsation
Low arterial perfusion
Motion artifact
Perfusion Index (P I).
Perfusion Index (P I).
Bluetooth LE radio for transfer of parameter data to a comp atible smart device.
Perfusion Index (PI), and Pleth Variability Index (PVI).
Bluetooth radio for transfer of parameter data to a smart device.
Caution: To minimize radio interference, other electrical equipment that emits radio frequency
transmissions should not be in close proximity to MightySat.
Caution: High ambient light sources, fluorescent lights, infrared heating lamps and direct sunlight
can interfere with the performance of MightySat.
Caution: To prevent interference from ambient light, ensure that MightySat is properly applied.
Failure to take this precaution in high ambient light conditions may result in inaccurate or no measurements.
Caution: If no measurements are displayed, find a better perfused monitoring site.
Caution: When using MightySat Model 9800 or Model 9900 with a smart device, keep both devices
within the recommended range of each other (see Specifications for details); moving outside of this range may cause missing, lost, and/or inaccurate data to display on the smart device.
Note:The MightySat display may be difficult to view when exposed to direct sunlight or bright lights.
Note: A functional tester cannot be used to assess accuracy of MightySat.
Note: The MightySat display will shut off automatically if there are no readings.

Cleaning and Service Warnings and Cautions

Warning: Changes or modifications not expressly approved by Masimo shall void the warranty for
this equipment and could void the user’s authority to operate the equipment.
Warning: Use only AAA alkaline batteries.
Warning: Batteries may leak or explode if used or disposed of improperly. Remove batteries if the
device will be stored for more than 30 days.
Warning: Do not mix fully charged and partially charged batteries at the same time. These actions
may cause the batteries to leak.
Caution: A user may only perform maintenance procedures specifically described in the manual;
otherwise, return MightySat for servicing.
Caution: Do not use petroleum-based or acetone solutions, or other harsh solvents, to clean
MightySat. These substances affect the device’s materials and instrument failure can result.
Caution: Do not submerge MightySat in any cleaning solution or attempt to sterilize by autoclave,
irradiation, steam, gas, ethylene oxide or any other method. This will seriously damage the device.
Caution: Do not use undiluted bleach (5% - 5.25% sodium hypochlorite) or any cleaning solution
other than those recommended here because permanent damage to MightySat may occur.
Caution: Never immerse MightySat in water or any other liquid solution.

Compliance Warnings and Cautions

Caution: Disposal of Product: Comply with local laws in the disposal of the instrument and/or its
accessories, including batteries.
Note: When using MightySat with a device with wireless features, consideration should be taken to
local government frequency allocations and technical parameters to minimize the possibility of interference to/from other wireless devices.
Note: In accordance with international telecommunication requirements, the frequency band of 2.4
GHz is only for indoor usage to reduce potential for harmful interference to co-channel mobile satellite systems.
Note: This device complies with Part 15 of the FCC Rules. Operation is subject to the following two
conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
Note: This equipment has been tested and found to comply with the limits for a Class B digital
device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
Reorient or relocate the receiving antenna.
Increase the separation between the equipment and receiver.
Connect the equipment into an outlet on a circuit different from that to which the receiver is
connected.
Consult the dealer or an experienced radio/TV technician for help.
Note: This equipment has been tested and found to comply with the Class B limits for medical
devices according to the IEC 60601-1-2:2007. These limits are designed to provide reasonable protection against harmful interference in all establishments, including domestic establishments.

The Masimo Difference

From the leader in hospital pulse oximetry comes a complete pulse oximeter that is compact and portable to help you measure blood oxygen saturation through motion and low perfusion. The MightySat Fingertip Pulse Oximeter allows you to noninvasively track oxygen saturation (SpO2), pulse rate (PR), perfusion index (PI), and Pleth Variability Index (PVI). The MightySat features Masimo® Signal Extraction Technology® (SET®) technology, proven to provide accurate pulse oximetry measurements during motion and low perfusion.
For more than 20 years, Masimo has been focused on a singular mission–to take noninvasive patient monitoring to new sites and applications. Visit www.Masimo.com for more information about Masimo products.

About the MightySat Technology

Pulse oximetry is governed by the following principles: Oxyhemoglobin (oxygenated blood) and deoxyhemoglobin (non-oxygenated blood) differ in their
absorption of red and infrared light (spectrophotometry).
Display Parameter Description
Functional oxygen saturation (SpO2) Functional oxygen saturation (SpO2) is the amount of oxyhemoglobin expressed as a
Pulse Rate (PR) Pulse rate (PR), measured in beats per minute (BPM), is based on the optical detection of
Perfusion Index (P I) The Perfusion Index (PI) is the ratio of the pulsatile blood flow to the non-pulsatile or static
Pleth Variability Index (PVI ) (Model 9900 only)
Plethysmograph Waveform (Pleth) The amount of arterial bloo d in tissue changes with your pulse (photoplethysmography).
Signal I.Q. Indicators
(SIQ)
MightySat Fingertip Pulse Oximeter
Manufacturer:
percentage of the hemoglobin that is available to transport oxygen.
peripheral flow pulse.
blood in peripheral tissue. PI thus represents a noninvasive measure of peripheral perfusion that can be continuously and noninvasively obtained from a pulse oximeter.
The Pleth Variability Index (PVI) is a measure of the dynamic changes in t he Perfusion Index (PI) that occur during the respiratory cycle. The calculation is accomplished by measuring
PVI is displayed as a percentage (0-100%).
The utility of PVI is unknown at this time and requires further clinical studies. Technical factors that may affect PVI include malposition and user motion.
Therefore, the amount of light absorbed by the varying quantities of arterial blood changes as well. Clinicians have attempted to utilize this waveform and its variation for assigning signal integrity, physiological and artifactual changes such as perfusion changes, dysrhythmia, motion artifact, and electrical interference. For this reason, accurate and reliable presentation of the plethysmograph waveform is of importance.
The Signal I.Q.® (SIQ) provides an indicator of the assessment of the confidence in the displayed SpO2 value. The height of the vertical line of the SpO2 SIQ provides an assessment of the confidence in the measurement displayed. A high vertical line indicates higher confidence in the measurement. A small vertical line indicates lower confidence in the displayed measurement.
40 Parker Irvine, CA 92618 USA Tel.: 949-297-7000 Fax.: 949-297-7001 www.masimo.com
37633/8722A-0115

Getting Started with MightySat

Compatibility

Note: Compatibility is limited to Models 9800 and 9900 only.

Turning off MightySat

The MightySat automatically turns off after removing the finger from the device in the absence of device interaction or connection to a smart device.

Using the Touchpad

For select Android®-powered devices. For select Apple® devices.

Operation

MightySat Features

ID Description Function
1 Enclosure Clip Clip provided for ease of lanyard attachment.
2 Bluetooth Indicator (For Models 9800 and
9900)
3 Battery Status Indicator Indicates the remaining relative life of the battery.
4 Display Screen Display for measurements and indicators.
5 Waveform and SIQ Display of the pleth waveform and the SIQ provides an indicator of the relative
6 Touchpad User interface to allow for change of settings (see Using the Touchpad section in

Installing the AAA Batteries

The MightySat requires two alkaline AAA batteries to operate. To install batteries, follow the instructions below:
Warning: Before changing batteries, make sure the device is off and is not applied to a finger.
1. Orient the MightySat so that the display screen is facing downwards. Locate the battery button on the front of the sensor pad.
Indicates when Bluetooth is enabled on the device.
Note: Numbers will dim when confidence in the value is low.
signal strength.
this manual).
The Touchpad on MightySat device is located below the display screen. Note: The display is not a touch screen.
Action Description Function
Touch Tap and release. Action will be performed
Press and Hold Press and hold. Release finger once action
Swipe Touch and slide (left, and right) and
once finger is released.
has been performed.
release.
Select a menu item or action
Touch rotates the display clockwise
Enter and Exit the Main Screen
Scroll through all selectable menu options.

Navigating the Menu

From the Main Screen, press and hold the Touchpad to access the Main Menu. Use the Touchpad Swipe gesture to scroll through the Main Menu Options. Use the Touch gesture to select
the Main Menu Option. Use the same gestures to adjust settings. The Menu options are:
Main Menu Options Display Button Description Default Options
Back
Waveform
Return to Main screen N/A N/A
Allows the user to choose if the waveform will be displayed on the screen.
On On or Off
2. Push lightly on the battery button to release the battery cover and then remove the battery cover.
3. Locate the battery orientation labels (+ and -) provided inside the battery compartment.
4. Insert two new AAA alkaline batteries and carefully match the orientation labels (+ and -). Caution: MightySat will not work if the batteries are inserted in the incorrect orientation.
5. Once the batteries are correctly inserted, carefully snap the battery door back onto the device.
Warning: Ensure that the battery door is intact before use. Note: MightySat will automatically activate when the device is opened so that the sensor pads are
exposed as shown in the image below.

Using MightySat

To take readings with the MightySat, follow the instruction below:
Note: Before use, ensure batteries are correctly installed in the MightySat.
1. MightySat will automatically activate when the device is opened so that the sensor pads are exposed.
2. To open the MightySat squeeze the back portion of the device as shown in the image above.
3. Once the sensor pads are exposed, insert a finger (non-dominant, ring finger) so the sensor LED is above the fingernail.
Note: The display screen of the MightySat should be facing upwards as depicted in the image above.
4. Position the finger so that the tip of the finger touches the backstop provided at the end of the sensor pad as shown in the image below.
Brightness
About
Bluetooth (Models 9800 and 9900 only)
Change the brightness of the display screen.
Hardware and software information about the device including serial number, software version, and Bluetooth LE Mac Address.
Allows the user to connect to a smart device via Bluetooth LE.
100%
N/A N/A
On On or Off
25%, 50%, 75% and 100%

Cleaning and Service

Cleaning MightySat

Warning: Before cleaning, read Cleaning and Service Warnings and Cautions section in this manual. Warning: Before cleaning, make sure the device is off and is not applied to a finger. Note: Before cleaning, remove the batteries and make sure the battery cover is attached correctly.
To clean the MightySat, follow the instructions below:
The outer surfaces can be cleaned either with a soft cloth dampened with a mild detergent and warm
water solution or they can be wiped down with the following cleaning solutions:
Cidex Plus (3.4% glutaraldehyde)
10% bleach solution
70% isopropyl alcohol solution
Allow the MightySat to dry thoroughly before using again.
Caution: To avoid permanent damage to the MightySat, do not use undiluted bleach (5% - 5.25% sodium hypochlorite) or any other cleaning solution not recommended.

Service

Warning: Do not attempt to repair the MightySat as this may cause damage to the device and prevent it from operating properly.
If the device does not appear to be operating correctly, see Troubleshooting section in this manual. Note: To maintain the proper functionality of the battery compartment and avoid possible damage from
alkaline batteries that may leak, remove batteries from the device when not in use for long periods of time.
5. Once the finger is correctly positioned, gently close the MightySat by releasing the pressure on the back of the device.
Note: Ensure the finger is correctly positioned for accurate measurements.
6. Once the MightySat is correctly closed on the finger, the MightySat will display readings.
Note: If no readings are displayed, see Troubleshooting section in this manual. Warning: While on the finger, do not press the top of the device against any surface. Warning: Do not attempt to secure the MightySat to the finger using external pressure. The internal
spring provides the correct pressure; additional pressure may cause inaccurate readings.