Martel Electronics T150 User Manual

Model T150
Hand-held
Frequency
Calibrator
OPERATING INSTRUCTIONS
General Description
The T150 Frequency calibrator is designed to be a high accuracy, simple to use frequency standard for calibrating process control instrumentation. The T150 can generate or read frequencies ranging from 1 count per minute to 100 KHz allowing it to be used for a wide range of flow measuring instrumentation.
When operated in the source mode, the user can quickly step the output in 10% of range increments or make smaller changes using the scroll keys.
Customer Service
Corporate Office:
www.martelcorp.com
e-mail: sales@martelcorp.com
Tel: (603) 434-1433 800-821-0023
Fax: (603) 434-1653
Martel Electronics
3 Corporate Park Drive
Derry, NH 03038
Installation
Unpacking
Remove the Packing List and verify that all equipment has been received. If there are any questions about the shipment, please call Martel at 1-800-821-0023 or email sales@martelcorp.com.
When you receive the shipment, inspect the container and equipment for any signs of damage. Note any evidence of rough handling in transit. Immediately report any damage to the shipping agent.
1
2
NOTE: The carrier will not honor any claims unless all shipping material is saved for their examination. After examining and removing contents, save packing material and carton in the event reshipment is necessary.
Set-up
Before beginning, become familiar with the keypad layout and the configuration of the input/output jacks. Remember, these jacks are used in multiple configurations so pay careful attention to the jack identification label on back of unit and how the test leads are connected for the specific application.
Operating Procedure
1. Turn on power and select the desired range by depressing the range select key. Your range choices are CPM, Hz, and KHz.
2. Make connections as required. Refer to the operating mode section or the rear jack identification label for connection information.
3. The step and scroll keys adjust the output value up or down. Holding the scroll key continuously for several seconds will increase the scroll rate. The step 'key' allows the user to make larger changes quickly and is preset to step in 10% of range increments.
3
Operating Modes
1. When operating in the output mode the T150 has the capability to drive loads up to 1 kOhm with TTL levels. The standard output amplitude is a 5 volt ± 10% p-p square wave.
Figure 1
2. When operating in the read (input) mode the T150 can monitor frequencies with amplitudes ranging from 1 to 100 volts. The waveform shape can be square, sine or triangular.
Figure 2
Accuracy
The T150 is checked against an NIST Traceable reference before shipment to verify that each range falls within the 0.01% ±1count of full scale. Since the accuracy of this particular product is based on a very stable quartz crystal, the T150 should stay well within its operating specification without the need for calibration for periods of 1 year or more.
Specifications
Input/Output Ranges 0 to 1000 CPM
0 to 1000 Hz 0 to 100.0 kHz
Accuracy ±0.01% ± 1 count Max Load Driving 5 mA (1 kOhm load min) Input Signal Amplitude 1 to 100 Volts Output Waveform 5 V ± 10%, Square Wave Operating Temperature 0 to 50°C Storage Temperature -20 to 60°C Temperature Stability 0.005% of full scale/°C Power
1
9 V Alkaline Battery
Weight 12 oz. / 340 g
Notes:
1
A low battery indication is given at 5.5 volts.
4
5
Maintenance
Generally, with normal usage, the calibrator should hold its rated specifications for at least 12 months. Beyond this, it should be returned for re-calibration annually. If after the stated warranty period, the device falls out of calibration, or needs repair, it can be returned to Martel Electronics for re-calibration. Please use our website www.martelcorp.com under Customer Service for return information.
Warranty
Martel Electronics Corporation warrants all products against material defects and workmanship for a period of twelve (12) months after the date of shipment. Problems or defects that arise from misuse or abuse of the instrument are not covered. If any product is to be returned, a “Return Material Authorization” form can be obtained from our website www.martelcorp.com under customer service. You can also call 1-800-821-0023 to have a form faxed. Martel will not be responsible for damage as a result of poor return packaging. Out of warranty repairs and recalibration will be subject to specific charges. Under no circumstances will Martel Electronics be liable for any device or circumstance beyond the value of the product.
02C1977 Rev B 6/11
www.martelcorp.com
e-mail: sales@martelcorp.com
Tel: (603) 434-1433 800-821-0023
Fax: (603) 434-1653
Martel Electronics
3 Corporate Park Drive
Derry, NH 03038