The SP 300 features a 200mV d.c. measurement range with auto-zero and auto-polarity.Decimal points are user
selectable. The SP 300 features a negative rail generator which enables the meter to measure a signal referenced to its
own power supply 0V. The bright red LED display ensures excellent readability under low light conditions. It can be
blanked in applications requiring low power operation. The module is easily fitted into the panel, using the fixing clip
provided. The module's low cost means it will suit high and low volume applications. The design of the panel meter's
housing and seal ensures splash proofing in many applications.
FEATURES
•
9.4mm (0.37") Digit Height
• 200mV d.c. Full Scale Reading
• 50mA @ +5V d.c. Power Supply
•
Auto-zero and Auto-polarity
•
Programmable Decimal Points
• Bright Red
• Display Blanking Facility
• Simplified Connection
• Splash Proof
LED Display
TYPICAL APPLICATIONS
• Precision Instrumentation Systems
• Power Supply Monitoring
ORDERING INFORMATION
Stock Number
Standard MeterSP 300
• Test Boxes
• Panel-Mount Indication
ELECTRICAL SPECIFICATIONS
SpecificationMin.Typ.Max.Unit
Accuracy (overall error) *0.1% (±1 count)
Linearity±1count
Sample rate2.5samples/sec
Operating temperature range050°C
Warm-up time10minute
Temperature stability150ppm/°C
Supply voltage4.7555.25V
Supply current
Input leakage current (Vin = 0V)110pA
* To ensure maximum accuracy, re-calibrate periodically.
50mA
SAFETY
To comply with the Low Voltage Directive (LVD 93/68/EEC), input voltages to the module’s pins must not exceed
60Vdc. The user must ensurethattheincorporationofthe panel meter into the user ’s equipment conforms to the
relevant sections of BS EN 61010 (Safety Requirements for Electrical Equipment for Measuring, Control and
Laboratory Use).
3. COM Ground for analogue section of A/D converter.
It is actively held at 3.05V (nom) below V+ and must not be allowed to sink excessive current
µ
(>100 A)by,forinstance, connecting to a higher voltage.
4. IHPositivemeasuring input.
5. ILNegative measuring input.
6. DP1Connect to 0V to display DP1 (199.9).
7. DP2Connect to 0V to display DP2 (19.99).
8. DP3Connect to 0V to display DP3 (1.999).
Note:
A negative supply is generated internally and mirrors the positive supply.For example: if V+ is +5V, then
the internally generated V- is -5V. When measuring with the input referenced to the same supply rail as
that of the panel meter, then the limitations on the input range are (V- + 1.5V) to (V+ - 1.5V).
Solder Links:
LCOMNormallyOpen. ConnectsIL to COM.
PO Box 770, Londonderry, NH 03053 1-800-821-0023
www.lascarelectronics.com
www.martelcorp.com
Page3of4
Page 4
Page4of4
SP 300
3½ Digit LED Voltmeter Module
SCALING
Two resistors Ra and Rb
may be used to alter the
full scale reading (FSR) of
the meter - see table. The
meter will have to be recalibrated by adjusting the
calibration potentiometer
on the rear of the module.
FSRRaRb
2V910k**100k
Voltage
20V1M**10k
200V1M**1k
2000V* 1M**100R
200 A0R1k
µ
Current
2mA0R100R
20mA0R10R
200mA0R1R
* Ensure that Ra is rated for high voltage use.
** Ensure solder link LA is cut.
Vin or inI
LA
Rb
4
IH
5
IL
SP 300
+
Ra
-
APPLICATIONS
Do not connect more than one meter to the same power supply if the meters cannot use the same signal
ground. Taking any input beyond the power supply rails will damage the meter.
V+
1
V+
4
+
±200mV
-
5
3
IH
IL
COM
0V
2
0V
Measuring an input voltage referenced
to a floating supply, i.e. the input
voltage and the meter's power supply
are isolated from each other.
7805
V
supply
220nF470nF
+5V
10uF
3
COM
V+
0V
1
IH
IL
2
Measuring supply voltage and current to a load.
S selectsbetween voltage and current measurement.
1
V+
1
V+
4
+
±200mV
-
IH
5
IL
0V
2
0V
Measuring a single ended input
voltage referenced to supply, i.e. the
input voltage and the meter's power
supply share the same 0V rail.
I
Ra=1M
V
4
S
1
A
5
Rb=1k
Rs
Shunt
Load
R
L
Set Zero
4-20mA
current loop
220k
5k
I
+
R
I-
5
3
4
IL
COM
IH
V+
0V
Measuring a 4-20mA loop current.
R= Reading at 20mA
160
The meter's power supply must be
isolated from the 4-20mA current
loop.
V+
1
V+
4
5
3
IH
IL
COM
0V
2
0V
R = 0.2
R
I
FSR
Measuring current.
The meter's supply is isolated
from the current being measured.
V+
1
2
0V
Ra and Rb shown scaled for 200V
FSR
Rs = 200mV
I
(e.g. 0.1 /400mWfor 2A full scale)Ω
FSR
Display DP1, DP2 or DP3, by connecting to 0V, as required.