Martel Electronics QM140V User Manual

DIMENSIONS
All dimensions in mm (inches)
FITTING THE EMV 1125
Drill a 5.5mm / " hole in the panel.
not toovertighten the nut as thismay damage the meter.
7
/
Fit the module to the panel by passing its screw threaded stud and the wires through the hole, fitting the washer and tightening the nut provided. Take care not to trap any of the connecting wires. Do
EMV 1125 isan easy to mountLCD digital panel meter. Theuser need only drill asingle 5.5mm / hole in thepanel. The module isfitted to the panelby locating its screw threaded stud through the hole, fitting the washer and tightening the nut provided. The module's 10 connecting wires pass through the hollow stud into the target application,where they canbe easily soldered into place. The module features 200mV fullscale input, auto-zero and auto-polarity. A separate screw terminal adaptor board (T/BLK-4) simplifies connectionandcalibration.
"
7
/
Single Hole Mounting
200mV Full Scale Reading
12.5mm / 0.5" LCD Digit Height
Auto-zero, Auto-polarity
10 Wire Connections
Low Battery Warning
Screw Terminal Board Option (T/BLK-4)
EMV 1125
Single Hole Mounting Meter
Stock Number
Easy Mounting Voltmeter EMV 1125 Optional Screw Terminal Adaptor Board T/BLK-4
Specification Min. Typ. Max. Unit
Accuracy (overall error)* 0.05 0.1 % (±1 count) Linearity ±1 count Displayed reading -1999 1999 Resolution 0.1 mV d.c. Sample rate 3 sample/sec Operating temperature range 0 50 °C Temperature stability 100 ppm/°C Supply voltage 7.5 9 15 V d.c. Supply current 150
* To ensure maximum accuracy, re-calibrate periodically.
µA
WEBSITE: http://www.lascarelectronics.com/
Shown Actual Size
REAR VIEW
EMV 1125 T/BLK-4
Panel 10 Wires (30cm)
Shakeproof
Washer
Nut
Nut
21.40 (0.84)
10.70 (0.42)
5.00 (0.20)
10.70 (0.42)
43.50 (1.71)
21.75 (0.85)
21.75 (0.85)
SAFETY
To comply with the Low Voltage Directive(LVD 93/68/EEC), input voltages to the module’s pinsmust not exceed 60Vd.c.If voltagesto the measuring inputs do exceed 60Vd.c., then fit scaling resistors externally to the module. The user must ensure that the incorporation of the meter into the user’s equipment conforms tothe relevantsections of BS EN61010 (Safety Requirements for ElectricalEquipment forMeasuring, Control and LaboratoryUse).
Module thickness 5.5mm (0.22)
If T/BLK-4 is not fitted, then do not fit the first nut.
9.0
(0.35)
Adding a calibration circuit when T/BLK-4 is not used
Note: The calibration circuit must always be fitted, even when scaling resistors are not fitted. This circuit is fitted as standard on T/BLK-4.
100k
3k
1k5
R
CAL
V+
(Red/2)
Vref
COM
(Orange/3)
(Brown/1)
WIRE CONNECTIONS and SCREW TERMINAL FUNCTIONS
(EMV 1125) (T.BLK-4)
0. Black (BK) V- Negative power supply connection.
1. Brown (BR) COM The ground for the analogue section of the A/D converter,held activelyat 2.8V(nom.)belowV+. COM must not be allowed to sink excessivecurrent (>100 A)byconnecting it directly to a higher voltage. Positivepower supplyconnection.
3. Orange (OR) Vref Connection forcalibration circuit(seediagrambelow).
4. Yellow (YL) IN HI Positive measuring differential input. IN HI and IN LO must be no closer than 1V to either the positiveor negativesupply.
5. Green (GN) INLO Negative measuringdifferentialinput. (On T/BLK-4,INLO is connectedtoCOMvianormally-closedLinkL2.)
6. Blue (BL) DP COM Commonconnection fordecimal pointsDP1, DP2andDP3,seebelow.
µ
2. Red (RD) V+
7. Violet(VI) DP1 199.9
8. Grey (GY) DP2 19.99 Connect to DP COM to display required decimal point.
9. White (WT) DP3 1.999 (or makeappropriate solderlink onT/BLK-4)
Do notconnect more than one meter to the same powersupply if the meters cannot use the same signalground. Taking any input beyond the power supplyrails will damage themeter. Keep leads shortto ensure noise-free operation.
Specifications liable to change without prior warning EMV 1125 Issue 3 January/2001 M.C. Applies to EMV 1125/2
Required F.S.R. Ra Rb CAL
200mV N/A N/A Adjust 2V 910k 100k 20V 1M 10k 200 A 0R 1k 2mA 0R 100R 20mA 0R 10R Ad 200mA 0R 1R
Adjust Adjust Adjust Adjust
just
Adjust
µ
SCALING and CALIBRATION
A calibrationcircuit andtwo resistors (Raand Rb)may be added on the EMV 1125 or T/BLK-4 board (cut Link La if fitting Ra) to alter the full scale reading of the meter - see table. Note that the meter will have to be (re-)calibrated byadjusting thepotentiometer.
APPLICATIONS
INPUT
+
-
Ra 100k
Rb
IN HI
(Yellow/4)
IN LO
(Green/5)
Adding an input filter and scaling resistors when T/BLK-4 is not used
100nF
Note: To reduce noise on the meter's input, fit the input filter components (1M and 10nF) shownabove. Theinput filter is fitted as standardon T/BLK-4.
1
CAL
Rb
V- V +
INHI
INLO
Ra
La
L1
L2
123
DP
BK BR RD OR YL GN BL VI GY WT
Solder wires from EMV 1125
Calibration potentiometer
Scaling resistors (values for 2V and 20V FSD Supplied with T/BLK-4)
Decimal point links
Connecting to T/BLK-4
Measuring current. Recalibrate the module. The current to be measured must be isolated from the module's power supply.
Rb =I0.2
Full Scale
+
+
-
-
Supply Voltage
Current to be
measured
1
CAL
Rb
V- V+
INHI
INLO
Ra
La
L1
L2
123
DP
BK BR RD OR YLGN BL VI GY WT
+
-
Supply Voltage
1
CAL
Rb
V- V+
INHI
INLO
Ra
La
L1
L2
123
DP
GN BL VI GY WT
Measuring a supply voltage. (min. 7.5V, max. 15V) Cut Links La and L2. Recalibrate the module.
510k
510k
10k
Measuring a floating voltage source of 200mV full scale. Add Ra and Rb and cut Link La to increase the measurement range.
+
+
-
-
Supply Voltage
+200mV
Input Voltage
1
CAL
Rb
V- V+
INHI
INLO
Ra
La
L1
L2
123
DP
BK BR RD OR YLGN BL VI GY WT
BK BR RD OR YL
LASCAR ELECTRONICS LTD. LASCAR ELECTRONICS INC. MODULE HOUSE, WHITEPARISH, PO BOX 50727, PALO ALTO, WILTSHIRE. SP5 2SJ UK CA 94303-0727 USA TEL: ++44 (1794) 884567 F TEL: ++1 (650) 838 9027AX: ++44 (1794) 884616 FAX: ++1 (650) 833 5432